ML 101B2 - Signal Generator
ML 101B2 - Signal Generator
1. Introduction
The signal generator produces an electrical signal which can be used to characterize or analyze in
an electrical or electronic circuit. They are widely used in electronic design, testing, and
troubleshooting applications. They are also described as function, tone, or waveform generators.
Signal generators can be either analog or digital devices. Digital signal generators allow you to
program and synthesize various waveforms that are not generally available on the analog signal
generator. They also provide more convenient control features. Both digital and analog signal
generators are used in labs. Here we focus on digital signal generators.
Figure 1 - Tektronix AFG 1022 signal generator
4. Message display area: This part displays the load value
5. Message display area: This part displays the current channel.
6. Parameter display area: This part displays the period.
7. Message display area: This part displays the type of the current signal or the current mode.
4. Setting up the signal generator
To connect the signal generator to other circuits you can use a variety of cables. Normally BNC
coaxial cables are used for this purpose. Usually, signal generators use BNC cables in the
following two formats.
1. BNC connectors at both ends
2. BNC connector at one end and crocodile (alligator) clips at the other end.
Coaxial cable has an inner core insulated from an outer shield. The shield is connected to the
black clip which is the external ring of the BNC cable. This connects to the ground or zero
voltage point in the circuit. The central core is connected to the center pin of the BNC connector.
It is associated with the red clip. In a test circuit, this is the connection made in the point of
interest. Usually, signal generators are calibrated to use test leads which have an impedance of 50
Ω. Before making connections, check if the leads are correct. Normally the impedance value is
marked on the cable.
Figure 5 - Use these buttons to select the waveform type. The selected waveform type is sine.
6. After setting the amplitude value correct unit should be selected. There are few options as
mVpp (peak to peak voltage value in mV), Vpp (peak to peak voltage value in V), mVrms
(root mean square voltage value in mV), or Vrms (root mean square voltage value in V).
Figure 11 - Step 6: Selecting the amplitude unit
Likewise, Offset, High and low values can be changed in the same way. Usually, the offset value
is set to zero.