Emi Upload 1649315569
Emi Upload 1649315569
POLYTECHNIC
ROURKELA
LECTURE NOTES
ON
ELECTRICAL MEASURING
INSTRUMENT
DEPARTMENT OF ELECTRICAL
ENGINEERING
4TH SEMESTER
PREPARED BY
MAMATA MOHANTY
CONTENTS
Measuring instruments
Analog ammeters and voltmeters
Wattmeter and measurement of power
Energy meters and measurement of energy
Measurement of speed, frequency and power factor
Measurement of Resistance, Inductance& Capacitance
Sensors And Transducer
Oscilloscope
CHAPTER-1
MEASURING INSTRUMENTS
1. ACCURACY- The closeness with which an instrument reading approaches the true value
of the quantity being measured is called accuracy. Accuracy is determined as the maximum
amount by which the result differs from the true value.
2. PRECISION- The term precise means means clearly or sharply defined. Precision is a
measure of reproducibility of measurement.
3. ERRORS- The deviation or change of the value obtained from measurement from the desired
standard value.
Mathematically,
Error = Obtained Reading/Value – Standard Reference Value.
GROSS ERRORS-This are the error due to humans mistakes such as careless reading
mistakes in recoding observation incorrect application of an instrument.
3. OBSERVATIONAL ERROR- The errors introduced by the observer. These errors are
caused by habits of the observers like tilting his/her head too much while reading a
“Needle – Scale Reading”.
Measuring instruments are classified according to both the quantity measured by the
instrument and the principle of operation.
Electric measuring instruments and meters are used to indicate directly the value of
current,voltage, power or energy.An electromechanical meter (input is as an electrical signal
results mechanical force or torqueas an output) that can be connected with additional suitable
components in order to act as anammeters and a voltmeter.The most common analogue
instrument or meter is the permanent magnet moving coilinstrument and it is used for measuring
a dc current or voltage of an electric circuit.
5. RESOLUTION- Resolutions the least increment value of input or output that can be
detected, caused or otherwise discriminated by the measuring device.
6. TOLETANCE- Tolerance refers to the total allowable error within an item. This is
typically represented as a +/- value off of a nominal specification. Products can become
deformed due to changes in temperature and humidity, which lead to material expansion and
contraction, or due to improper feedback from a process control device. As such, it's
necessary to take errors into consideration with regard to design values in the manufacturing
and inspection processes.
Absolute instrument measures the process variable directly from the process without the use of
conversion. Such instruments do not require comparison with any other standard. The tangent
galvanometer is an example for the absolute instrument. These instruments are used as standards in
labs and institution.
Secondary instrument: These instruments are so constructed that the deflection of such instruments
gives the magnitude of the electrical quantity to be measured directly. These instruments required
to calibrated with respect to the standard instrument. These instruments are usually used in practice.
1. Indicating instrument: Those instruments that measure and indicates the magnitude of the
electricity. The indications are given by a pointer moving over a calibrated scale. Ordinary
ammeters, voltmeters, wattmeters, frequency meters, power factor meters, etc., fall into this
category.
2. Integrating instrument: Integrating instruments are those which measure the total amount of
either quantity of
electricity (ampere-hours) or electrical energy supplied over a period of time. The ampere-hour
meters and energy meters fall in this class.
3. Recording instrument: These instruments record continuously the variation of the magnitude of
the electric quantity for a definite period of time. Such instruments are generally used in
powerhouses where the current, voltage, power, etc., are to be maintained within a certain
acceptable limit.
Analog instrument: The signals of an analog unit vary in a continuous fashion and can take on an
infinite number of values in a given range. Fuel gauge, ammeter and voltmeters, wristwatch,
speedometer fall in this category.
Digital Instruments: Signals that vary in discrete steps and that take a finite number of different
values in a given range are digital signals and the corresponding instruments are of digital
type. Digital instruments have some advantages over analog meters, in that they have high accuracy
and high speed of operation. Digital multimeter is an example for the digital instrument.
Mechanical instrument: Mechanical instruments are very reliable for static and stable conditions.
As they use mechanical parts these instruments cannot faithfully follow the rapid changes which
are involved in dynamic instruments. But they are cheaper in cost and durable.
Electrical Instruments: When the instrument pointer deflection is caused by the action of some
electrical methods then it is called an electrical instrument. The time of operation of an electrical
instrument is more rapid than that of a mechanical instrument. This mechanical movement has
some inertia due to which the frequency response of these instruments is poor.
Electronic Instruments: Electronic instruments use semiconductor devices. They are very fast in
response. In electronic devices, since the only movement involved is that of electrons, the
response time is extremely small owing to very small inertia of the electrons. With the use of
electronic devices, a very weak signal can be detected by using pre-amplifiers and amplifiers.
A Self contained instrument has all its different elements in one physical assembly.
In a Remote Indicating Instrument the primary sensing element may be located at an adequate
long distance from the secondary indicating element. Such type of instrument are finding wide use
in the modern instrumentation technology.
Self-operated instruments don’t need any outside power for its working. The output energy is
supplied wholly or almost wholly by the input measurand. Dial indicating type instruments belong
to this category.
Power operated instrument need external power for its working. External power can electric
current, hydraulic or pneumatic energy. In such cases, the input signal supplies only an
insignificant portion of the output power.
In a deflection-type instrument, the deflection of the instrument indicates the measurement of the
unknown quantity. The measurand quantity produces some physical effect which deflects or
produces a mechanical displacement in the moving system of the instrument.
An opposite effect is built in the instrument which opposes the deflection or the mechanical
displacement of the moving system. The balance is achieved when opposing effect equals the
actuating cause producing the deflection or the mechanical displacement. Permanent Magnet
Moving Coil (PMMC), Moving Iron (MI), etc., type instruments are examples of this category.
In Null type instruments, a zero or null indication leads to the determination of the magnitude of
the measurand quantity. The null condition depends upon some other known conditions. These are
more accurate and highly sensitive as compared to deflection-type instruments. A dc potentiometer
is a null- type instrument.
1. DEFLECTING TORQUE/FORCE:
The defection of any instrument is determined by the combined effect of thedeflecting
torque/force, control torque/force and damping torque/force.The value of deflecting torque must
depend on the electrical signal to be measured.This torque/force causes the instrument movement
to rotate from its zero position.
MAGNITUDE EFFECT
When a current passes through the coil, it produces a imaginary bar magnet. When a soft-iron
piece is brought near this coil it is magnetized. Depending upon the current direction the poles
are produced in such a way that there will be a force of attraction between the coil and the soft
iron piece. This principle is used in moving iron attraction type instrument.
If two soft iron pieces are place near a current carrying coil there will be a force of repulsion
between the two soft iron pieces. This principle is utilized in the moving iron repulsion type
instrument.
When a current carrying coil is placed under the influence of magnetic field produced by a
permanent magnet and a force is produced between them. This principle is utilized in the moving
coil type instrument.
When two current carrying coils are placed closer to each other there will be a force of repulsion
between them. If one coil is movable and other is fixed, the movable coil will move away from
the fixed one. This principle is utilized in electrodynamometer type instrument.
2. CONTROLLING TORQUE/FORCE:
This torque/force must act in the opposite sense to the deflecting torque/force, and themovement
will take up an equilibrium or definite position when the deflecting andcontrolling torque are
equal in magnitude.The Spiral springs or gravity usually provides the controlling torque.
When the external signal to be measured by the instrument is removed, the pointer should return
back to the zero position. This is possibly due to the controlling force and the pointer will be
indicating a steady value when the deflecting torque is equal to controlling torque.
Td = Tc
SPRING CONTROL
Two springs are attached on either end of spindle. The spindle is placed in jewelled bearing, so
that the frictional force between the pivot and spindle will be minimum. Two springs are
provided in opposite direction to compensate the temperature error. The spring is made of
phosphorous bronze. When a current is supply, the pointer deflects due to rotation of the spindle.
While spindle is rotate, the spring attached with the spindle will oppose the movements of the
pointer. The torque produced by the spring is directly proportional to the pointer deflectionθ.
TC∝θ
The deflecting torque produced Td proportional to ‘I’. When TC Td , the pointer will come
to a steady position. Therefore
I
3. DAMPING TORQUE/FORCE:
A damping force is required to act in a direction opposite to the movement of the moving
system.This brings the moving system to rest at the deflected position reasonably quickly
withoutany oscillation or very small oscillation.
CONSTRUCTION:
The moving coil and permanent magnet are the main part of the PMMC instrument. The parts of
the PMMC instruments are explained below in details.
Moving Coil – The coil is the current carrying part of the instruments which is freely moved
between the stationary field of the permanent magnet. The current passes through the coil deflects it
due to which the magnitude of the current or voltage is determined. The coil is mounted on the
rectangular former which is made up of aluminium. The former increases the radial and uniform
magnetic field between the air gaps of the poles. The coil is wound with the silk cover copper wire
between the poles of a magnet.
The coil is mounted on the rectangular former which is made up of aluminium. The former
increases the radial and uniform magnetic field between the air gaps of the poles. The coil is wound
with the silk cover copper wire between the poles of a magnet.
Magnet System – The PMMC instrument using the permanent magnet for creating the stationary
magnets. The Alcomax and Alnico material are used for creating the permanent magnet because
this magnet has the high coercive force (The coercive force changes the magnetisation property of
the magnet). Also, the magnet has high field intensities.
Control – In PMMC instrument the controlling torque is because of the springs. The springs are
made up of phosphorous bronze and placed between the two jewel bearings. The spring also
provides the path to the lead current to flow in and out of the moving coil. The controlling torque is
mainly because of the suspension of the ribbon.
Damping – The damping torque is used for keeping the movement of the coil in rest. This damping
torque is induced because of the movement of the aluminium core which is moving between the
poles of the permanent magnet.
Pointer & Scale – The pointer is linked with the moving coil. The pointer notices the deflection of
the coil, and the magnitude of their deviation is shown on the scale. The pointer is made of the
lightweight material, and hence it is easily deflected with the movement of the coil. Sometimes the
parallax error occurs in the instrument which is easily reduced by correctly aligning the blade of the
pointer.
PRINCIPLE OF OPERATION
When D.C. supply is given to the moving coil, D.C. current flows through it. When the
current carrying coil is kept in the magnetic field, it experiences a force. This force produces a
torque and the former rotates. The pointer is attached with the spindle. When the former rotates,
the pointer moves over the calibrated scale. When the polarity is reversed a torque is produced in
the opposite direction. The mechanical stopper does not allow the deflection in the opposite
direction. Therefore the polarity should be maintained with PMMC instrument. If A.C. is
supplied, a reversing torque is produced. This cannot produce a continuous deflection. Therefore
this instrument cannot be used in A.C.
TORQUE DEVELOPED BY PMMC
The deflecting torque induces because of the movement of the coil. The deflecting torque is
expressed by the equation shown below.
The spring provides the restoring torque to the moving coil which is expressed as
The above equation shows that the deflection torque is directly proportional to the current passing
through the coil.
1. Magnet – The heat and vibration reduce the lifespan of the permanent magnet. This treatment
also reduced the magnetism of the magnet. The magnetism is the property of the attraction or
repulsion of the magnet. The weakness of the magnet decreases the deflection of the coil.
2. Springs – The weakness of the spring increases the deflection of moving coil between the
permanent magnet. So, even for the small value of current, the coil show large deflection. The
spring gets weakened because of the effect of the temperature. One degree rise in temperature
reduces the 0.004 percent life of the spring.
3. Moving Coil – The error exists in the coil when their range is extended from the given limit by
the use of the shunt. The error occurs because of the change of the coil resistance on the shunt
resistance. This happens because the coil is made up of copper wire which has high shunt resistance
and the shunt wire made up of Magnin has low resistance.
To overcome from this error, the swamping resistance is placed in series with the moving coil. The
resistor which has low-temperature coefficient is known as the swamping resistance. The
swamping resistance reduces the effect of temperature on the moving coil.
1. The PMMC instruments are only used for the direct current. The alternating current varies with
the time. The rapid variation of the current varies the torque of the coil. But the pointer can not
follow the fast reversal and the deflection of the torque. Thus, it cannot use for AC.
2. The cost of the PPMC instruments is much higher as compared to the moving coil instruments.
The moving coil itself provides the electromagnetic damping. The electromagnetic damping
opposes the motion of the coil which is because of the reaction of the eddy current and the
magnetic field.
One of the most accurate instruments used for both AC and DC measurement is moving iron
instrument. There are two types of moving iron instrument.
Attraction type
Repulsion type
PRINCIPLE OF OPERATION
The current to be measured is passed through the fixed coil. As the current is low through
the fixed coil, a magnetic field is produced. By magnetic induction the moving iron gets
magnetized. The north pole of moving coil is attracted by the south pole of fixed coil. Thus the
deflecting force is produced due to force of attraction. Since the moving iron is attached with the
spindle, the spindle rotates and the pointer moves over the calibrated scale. But the force of
attraction depends on the current flowing through the coil.
REPULSION TYPE MOVING IRON INSTRUMENT
CONSTRUCTION:
The repulsion type instrument has a hollow fixed iron attached to it .
The moving iron is connected to the spindle. The pointer is also attached to the spindle in
supported with jeweled bearing.
PRINCIPLE OF OPERATION:
When the current flows through the coil, a magnetic field is produced by it. So both fixed
iron and moving iron are magnetized with the same polarity, since they are kept in the same
magnetic field. Similar poles of fixed and moving iron get repelled. Thus the deflecting torque is
produced due to magnetic repulsion. Since moving iron is attached to spindle, the spindle will
move. So that pointer moves over the calibrated scale. Damping: Air friction damping is used to
reduce the oscillation. Control: Spring control is used.
Errors
Error due to variation in temperature.
Error due to friction is quite small as torque-weight ratio is high in moving coil instruments.
Stray fields cause relatively low values of magnetizing force produced by the coil. Efficient
magnetic screening is essential to reduce this effect.
Error due to variation of frequency causes change of reactance of the coil and also changes
the eddy currents induced in neighbouring metal.
Deflecting torque is not exactly proportional to the square of the current due to non-linear
characteristics of iron material.
Dynamometer type measuring instruments are similar to PMMC instrument. Except that the permanent
magnetic field coil is replaced by a coil which carries the current to be measured. They have precision
grade accuracy both for ac and dc measurements
CONSTRUCTION:
A fixed coil is divided in to two equal half. The moving coil is placed between the
two half of the fixed coil. Both the fixed and moving coils are air cored. So that the hysteresis
effect will be zero. The pointer is attached with the spindle. In a non metallic former the
moving coil is wounded.
Control: Spring control is used.
Damping: Air friction damping is used.
PRINCIPLE OF OPERATION:
When the current flows through the fixed coil, it produced a magnetic field, whose flux
density is proportional to the current through the fixed coil. The moving coil is kept in between
the fixed coil. When the current passes through the moving coil, a magnetic field is produced by
this coil. The magnetic poles are produced in such a way that the torque produced on the
moving coil deflects the pointer over the calibrated scale. This instrument works on AC and DC.
When AC voltage is applied, alternating current flows through the fixed coil and moving coil.
When the current in the fixed coil reverses, the current in the moving coil also reverses. Torque
remains in the same direction. Since the current i1 and i2 reverse simultaneously. This is because
the fixed and moving coils are either connected in series or parallel.
Frictional Error: Since the coils are air-cored, therefore the magnetic field produced is of
small strength. So they require a large number of ampere-turns to create necessary
deflecting torque. This result in the heavy moving system. Therefore small torque-weight
ratio. Thus the frictional losses in dynamo type instruments are somewhat larger as
compared to other instruments.
Temperature errors: Since the operation of dynamo type instrument required considerable
power, self heating in these instrument is appreciable. The error due to self heating may be
much as 1% of full scale deflection.
Error Due to Stray Magnetic field: Since the operating magnetic field produced by
the fixed coil. In these instruments is somewhat weaker in comparison to that in
the instrument of other type. The operation of these instruments is more sensitive to the
stray magnetic field.
Frequency error: The change in frequency causes error
Due to change in reactance of operating coil.
Due to change in magnitude of Eddy current setup in the metal part of the instrument
near to operating portion.
As the instrument has Square Law response so can be used on both the dc as well as on AC.
These instruments are free from hysteresis and Eddy current errors. It is because of absence
of iron in the operating part of the instrument.
Ammeter up to 10A and voltmeter up to 600V can be constructed with precision grade
accuracy.
Dynamo type voltmeter are useful for accurate measurement of rms value of voltage
irrespective of waveform.
Because of Precision grade accuracy and same calibration for DC and AC measurement
instruments are used as transfer and calibration instruments.
Disadvantage of Dynamometer type instrument
The scale is not uniform as the instrument uses Square Law response. These instruments
have small torque-weight ratio so the friction error is considerable.
Owing to heavy moving system friction losses in these instruments are somewhat more than
those in other instruments.
As a result of measures taken to reduce the frictional errors, their cost is more in
comparison to moving iron and PMMC instruments. They are more sensitive to overload
and mechanical impact and are to be handled with care.
Adequate screening of the movements against the stray magnetic field is essential.
The sensitivity of the instrument is typically very low due to poor deflecting torque. The
sensitivity of dynamo type wattmeter is 10 to 30 per volt in comparison to the sensitivity of
20-kilo-ohm per volt in case of D’Arsnoval movement.
The power consumption of this instrument is comparatively high because of their
construction.
Ranges:
Ammeter: 1. With fixed and moving coil in series- 0/0.01A-0/0.05 A
2. With moving coil shunted or parallel connection- upto 0/30A.
Voltmeter: Upto 0-750 volts
or S = ImRm/(I – Im)
VOLTMETER CONNECTION
Multipliers are used for the range extension of voltmeters. The multiplier is a non-inductive high-
value resistance connected in series with the instrument whose range is to be extended. The
combination is connected across the circuit whose voltage is to be measured.
Example: A moving coil voltmeter reading upto 20 mV has a resistance of 2 ohms. How this
instrument can be adopted to read voltage upto 300 volts.
R Im = V – ν
or R = (V – v)/Im
A rectifier type instrument measures alternating electrical signal by means of D.C measuring
instrument. As the name implies, this instrument first rectifies an A.C signal to D.C then measures.
Although it measures the rectified A.C signal (D.C signal), but the scale of the instrument is
calibrated for A.C. The sensitivity of D’ Arsonval instrument is quite high. But a D’Arsonval
instrument can only measure to D.C. So, to utilize the sensitivity of D’Arsonval movement for A.C,
we use a rectifier type instrument.
Rectifier Elements for Rectifier Type Instrument
To convert A.C. to D.C, a rectifier instrument must have some rectifier elements.
At the low-frequency range, the instrument uses copper oxide or selenium cells for rectification
purpose. Again at higher frequencies, we use germanium or silicon diodes.
The copper oxide rectifier element has a peak inverse voltage (PIV) about 2 volts. On the other
hand, the selenium element has PIV of 10 volts. Also, both of these rectifier elements have very
low current handling capacity. Therefore, for rectification purpose, these elements have become
obsolete in the modern era.
Besides it, a germanium diode has a peak inverse voltage (PIV) of about 300 volts. In addition to
that, the current carrying capacity of a germanium diode is about 100 mA. Also, PIV of silicon
diode is about 1000 volts with a current rating of 5000 mA. Therefore these germanium and silicon
diodes have become most suitable choice as rectifier elements for the purposes.
The operation of Induction type instruments depends on the production of torque due to reaction
between two magnetic fluxes having some phase difference or reaction between flux of an AC
magnet and the eddy current induced by this flux. This instrument having an aluminum disc
(or aluminum drum) in the magnetic field. Hence, the changing flux links with the
aluminum disc. As a result, the flux induces an eddy current on the disc. This eddy current
interacts with the flux which has induced it. Consequently, there is a mechanical torque
acting on the disc. This mechanical torque rotates the disc. These type of instruments are used
only for AC measurements.
So, the torque depends on two factors. The first one is the strength of the field of the electromagnet.
The second one is the value of eddy current on the disc. Of course, the torque is proportional to the
strength of the magnetic field. Also, it is proportional to the eddy current. Again, the strength of the
magnetic field depends on the current of the electromagnet. On the other hand, the value of eddy
current depends on the strength of the magnetic field. So, we can say, the value of eddy current also
depends on the current of the electromagnet.
So, the torque acting on the disc is directly proportional to the square of the current of the
electromagnet. In an induction type instrument, we directly feed the measuring current into the coil
of the electromagnet. Therefore, the deflecting torque is directly proportional to the square of the
measuring current.
The phase angle between that flux and induced eddy current is α. Hence, we can write the
expression of the eddy current as
Again, the instantaneous torque is directly proportional to the instantaneous eddy current
and the flux. Hence, we can write,
So, the mean torque is as follows,
The above expression tells that the torque is zero if α is 90°. Hence to obtain resulting torque it is
necessary to produce an eddy current which is either less than or more than 90ᵒ out of phase with
flux ∅. ∝ is the phase angle between the flux and eddy current.
So, there must be some means in induction type instrument to prevent this phase angle
from being 90°. We can achieve this by two methods listed below.
Shaded pole type induction instrument uses a single winding to produce flux. The flux
produced by this winding is split up into two fluxes, having phase difference with respect to each
other. The phase difference is usually 40 to 50 degrees and can be varied by varying the size of
shading band. This is done by making a narrow slot in the poles of electromagnet. A copper strip is
placed around the smaller of the two areas formed by t he slot. This copper shading band acts as a
short circuited secondary winding.
The exciting coil is placed on the poles and a current proportional to current or voltage
being measured is passed through it. An aluminium disc which is mounted on a spindle is inserted
in the air gap of the electromagnet. The spindle carries a pointer and has a control sring attached to
it. The controlling torque is provided by this spring only.
Damping is provided by a permanent magnet placed at the opposite side of the
electromagnet, so that the disc can be used for production of both deflecting and damping torque.
CHAPTER-2
MEASUREMENT OF POWER
DYNAOMETER TYPE WATTMETER
A dynamometer type wattmeter primarily consists of two coils called fixed coil and moving
coil. The fixed coil is splitted into two equal parts, which are placed parallel to each other. The two
fixed coils are air-cored to avoid hysteresis effects when used on AC.
The fixed coil is connected in series with the load and carries the circuit current. It is,
therefore, called the current coil. The moving coil is pivoted between the two parts of the fixed coil
and is mounted on a spindle.
A pointer is attached to the spindle, which gives deflection. The moving coil is connected in
parallel with the load and carries the current proportional to the voltage. It is, therefore, called the
potential coil.Generally, a high resistance is connected in series with the moving coil to limit the
current through it. By limiting the current, the moving coil is made lightweight, which in turn
increases the sensitivity of the instrument.
The springs provide the controlling torque. They also serve the additional purpose of leading the
current into and out of the moving coil. Air friction damping is employed in such instruments.
Dynamometer Type Wattmeter Working
We use the wattmeter for power measurements. Its current coil is connected in series with the load,
carries the load current, and the potential coil, connected in parallel with the load, carries the
current proportional to the voltage across the load.
The fixed coil produces a field Fm, and moving coil creates a field Fr. The field Fr tries to come in
line with the main field Fm, which provides a deflecting torque on the moving coil.
Thus, the pointer attached to the spindle of the moving coil deflects. This deflection is controlled by
the controlling torque produced by the springs. Also read Power Measurements in Three Phase
Circuits.
Advantages:
It can be used both on AC and DC circuits.
It has a uniform scale.
We can obtain a high degree of accuracy through careful design.
Disadvantages:
At low power factors, the inductance of the potential coil causes serious errors.
The reading of the instrument may be affected by stray fields acting on the moving coil. To
prevent it, magnetic shielding is provided by enclosing the instrument in an iron case.
Errors in Dynamometer Type Wattmeter
1. Error due to potential coil inductance: The inductance of the potential coil is liable to cause an
error in the reading of the wattmeter. Because of this error, the wattmeter gives a high reading on
lagging power factor and low reading on leading power factor.
The high non-inductive resistance connected in series with the coil swamps the phasing effect of
the potential coil inductance.
2. Error due to power loss in the potential coil or current coil: Another possible error in the
indicated power may be due to some voltage drop in the current coil or the current taken by the
potential coil.
We can overcome this defect by using an additional compensating winding. This winding is
connected in series with the potential coil and so placed that it produces a field in the opposite
direction to that of the current coils.
3. Error due to eddy currents: The alternating field of fixed or current coil induces eddy currents in
the solid metal parts which set up their own magnetic field. This alters the magnitude and phase of
the magnetic field, causing deflection.
Thus an error is introduced in the instrument reading. To reduce this error, the solid metal parts are
placed far away from the current coil as possible.
4. Error due to the stray magnetic field: The dynamometer type wattmeter has a relatively weak
operating field; therefore, stray fields affect the reading of this instrument considerably and cause
serious errors.
Hence, this type of instrument must be shielded against stray magnetic fields try using iron cases or
providing thin iron shields over the working parts.
Range
The principle of operation of an induction wattmeter is same as that of induction ammeters and
voltmeters i.e. induction principle. However, it differs from induction ammeter or voltmeter in so
far that separate two coils are used to produce the rotating flux in place of one coil with phase split
arrangement.
The principle parts of an induction wattmeter are as shown in the fig below. It consists of two
laminated electromagnets. One electromagnet, called shunt magnet is connected across supply and
carries current proportional to the applied voltage. The coil of this magnet is made highly inductive
so that the current in it lags behind the supply voltage by 90 degrees. The other electromagnet,
called series magnet is connected in series with supply and carries the load current. The coil of this
magnet is made highly non inductive so that the angle of lag or lead is determined fully by the load.
A thin aluminium disc mounted on the spindle is placed in between the two magnets so that it cuts
the fluxes of both the magnets. The controlling torque is provided by spiral springs. The damping is
electromagnet and is usually provided by a permanent magnet embracing the aluminium disc. Two
or more closed copper rings, called shading rings are provide on the central limb of the shunt
magnet. By adjusting the position of these rings, the shunt magnet flux can be made to lag behind
supply voltage by exactly 90degrees.
When the wattmeter is connected in the circuit to measure a.c power, the shunt magnet carries
current proportional to the supply voltage and the series magnet carries the load current. The two
fluxes produced by the magnets induce eddy currents in the aluminium disc. The interaction
between the fluxes and eddy currents produce the deflecting torque on the disc, causing the pointer
connected to the moving system to move over the scale.
Vector diagram
Deflecting torque of Induction type wattmeter:
The vector diagram of this wattmeter is shown in the fig below. The current Iv in the shunt magnet
lags the applied voltage V by 90 degrees and so does the flux av produced by it. The current Ic in
the series magnet is the load current and hence lags behind the applied voltage by a'. The flux
ac produced by this current Ic is in phase with it. Therefore the two currents Ic in the current coil
and Iv in the voltage coil and also corresponding fluxes av and ac are (90 - a') apart.
The flux ac induces the eddy currents iv in the aluminium disc which lags behind the flux by
90degrees. Similarly, flux ac induces eddy currents ic which again lags behind flux ac by 90
degrees.
Mean deflecting torque, Td proportional ac sin (90 - a )
Td proportional V I cos a
Td proportional a.c power
Figure 1
BASIC OPERATION:
Figure 2
ERROS IN ENERGYMETER:
CHAPTER-5
DIGITAL TACHOMETER
The technique employed in measuring the speed of a rotating shaft is similar to the
technique used in a conventional frequency counter, except that the selection of the gate period is
in accordance with the rpm calibration. Let us assume that the rpm of a rotating shaft is R. Let P
be the number of pulses produced by the pickup for one revolution of the shaft. Therefore, in one
minute the number of pulses from the pickup is R x P. Then, the-frequency of the signal from the
pickup is (R x P)/60. Now, if the gate period is G s the pulses counted are (R x P x G)/60. In order
to get the direct reading in rpm, the number of pulses to be counted by the counter is R. So we
select the gate period as 60/ P, and the counter counts (Rx P x 60)/ 60P = R pulses and we can
read the rpm of the rotating shaft directly. So, the relation between the gate period and the number
of pulses produced by the pickup is G = 60/P. If we fix the gate period as one second (G= 1 s),
then the revolution pickup must be capable of producing 60 pulses per revolution. Figure shows a
schematic diagram of a digital tachometer.
MECHANICAL RESONANCE TYPE FREQUENCYMETER:
CHAPTER -6
MEASUREMENT OF RESISTANCE, INDUCTANCE, CAPACITANCE
The general arrangement of Wheatstone bridge circuit is shown in the figure below. It
is a fourarms bridge circuit where arm AB, BC, CD and AD are consisting of electrical
resistances P, Q, Sand R respectively. Among these resistances P and Q are known fixed
electrical resistances andthese two arms are referred as ratio arms. An accurate and sensitive
Galvanometer is connectedbetween the terminals B and D through a switch S2. The voltage
source of this Wheatstonebridge is connected to the terminals A and C via a switch S1 as shown.
A variableresistor S isconnected between point C and D. The potential at point D can be varied
by adjusting the valueof variable resistor. Suppose electric current I1 and electric current I2 are
flowing through thepaths ABC and ADC respectively. If we vary the electrical resistance value
of arm CD the valueof electric current I2 will also be varied as the voltage across A and C is
fixed. If we continue toadjust the variable resistance one situation may comes when voltage drop
across the resistor Sthat is I2.S is becomes exactly equal to voltage drop across resistor Q that is
I1.Q. Thus thepotential at point B becomes equal to the potential at point D hence potential
difference betweenthese two points is zero hence electric current through galvanometer is nil.
Then the deflection inthe galvanometer is nil when the switch S2 is closed.
Here in the above equation, the value of S and P ⁄ Q are known, so value of R can easily
be determined. The electrical resistances P and Q of the Wheatstone bridge are made of definite
ratio such as 1:1; 10:1 or 100:1 known as ratio arms and S the rheostat arm is made continuously
variable
from 1 to 1,000 Ω or from 1 to 10,000 Ω.
MAXWELLS BRIDGE:
This bridge is used to find out the self inductor and the quality factor of the circuit. As it
is basedon the bridge method (i.e. works on the principle of null deflection method), it gives
veryaccurateresults. Maxwell bridgeis an AC bridge so before going in further detail let us know
moreaboutthe ac bridge.Let us now discuss Maxwell's inductor bridge. The figure shows the
circuit diagram ofMaxwell's inductor bridge.
Maxwells Bridge
In this bridge the arms bc and cd are purely resistive while the phase balance depends on the
arms ab and ad.
Here l1 =Unknown inductor of r1.
l2 =Variable inductor ofresistanceR2.
r2 =variable electricalresistance
As we have discussed in ac bridge according to balance condition, we have at balance point
We can vary R3 and R4 from 10 ohms to 10,000 ohms with the help of resistance box.
In this Maxwell Bridge, the unknown inductor is measured by the standard variable capacitor.
Circuit of this bridge is given below.
Maxwell's Inductance Capacitance Bridge
Advantages of Maxwell's Bridge
(1) The frequency does not appear in the final expression of both equations, hence it
isindependent of frequency.
(2) Maxwell's inductor capacitance bridge is very useful for the wide range of measurementof
inductor at audio frequencies.
This bridge is used to measure to the capacitance of the capacitor, dissipation factor
andmeasurement of relative permittivity. Let us consider the circuit of Schering bridge as
shownbelow
Schering Bridge
Here, c1 is the unknown capacitance whose value is to be determined with series
electricalresistance r1.
c2 is a standard capacitor.
c4 is a variable capacitor.
r3 is a pure resistor (i.e. non inductive in nature).
And r4 is a variable non inductive resistor connected in parallel with variable capacitor c4.
Now the supply is given to the bridge between the points a and c. The detector is
connectedbetween b and d. From the theory of ac bridges we have at balance condition
Application:
This bridge is used to measure to the capacitance of the capacitor, dissipation factor
andmeasurement of relative permittivity.
1) Ranges available
2)Squaring System
3)Sensitivity
4) Maximum working temperature
5) Method of cooling employed
6) Mounting details
7) Maximum depth
8) Linearity
and hysteresis
9) Output for zero input
10) Temperature co-efficient of zero drift
11) Natural Frequency.
1. Very small power is required for controlling the electrical or electronic system
2. The electrical output can be amplified to any desired level
3. Mass inertia effects are reduced to minimum possible.
4. The size and shape of the transducers can be suitably designed to achieve the
optimum weight and volume
5. The output can be indicated and recorded remotely at a distance from the sensing
medium .
6. The outputs can be modified to meet the requirements of the indicating or controlling
equipment.
RESISITIVE TRANSDUCERS
STRAIN GAUGE
INTRODUCTION
When a metal conductor is stretched or compressed , its resistance changes on account of
the fact thatboth length and diameter of conductor change . The value of resistivity of conductor
also changes.When it is strained it’s property is called piezo-resistance .Therefore , resistance
strain gauges arealso known as piezo- resistive gauges .The strain gauge is a measurement
transducer for measuring strain and associated stress inexperimental stress analysis.
TYPES
WORKING PRINCIPLE
Strain gauges work on the principle that the resistance of a conductor or a semiconductor
changeswhen strained .This property can be used for measurement of displacement, force and
pressure .When a strain gauge is subjected to tension (positive strain) it’s length increases while
it’s crosssectional area decreases. Since the resistance of a conductor is proportional to it’s length
and inverselyproportional to it’s area of cross section, The resistance of the gauge increases with
positive strain .Strain gauges are most commonly used in wheat –stone bridge circuits to measure
the change ofresistance of grid of wire for calibration proposes; the ‘GAUGE FACTOR’ is
defined as the ratio ofper unit change in resistance to per unit change in length.
i.e , Gauge factor (Gf) = ΔR/R ÷ ΔL/L
Where, ΔR = corresponding change in resistance, R
ΔL = Change in length per unit length, L
R= ρL/A
Where, R= resistance, Ω
ρ = Resistivity of conductor materials, Ω-m
L= Length of conductor, m
A = Cross sectional area of the conductor, m2
L.V.D.T
LVDT is a passive inductive transducer and is commonly employed to measure force(or
weight,pressure and acceleration etc. Which depend on force )in terms of the amount and
direction ofdisplacement of an object.
WORKING PRINCIPLE
When the core is in the centre (called reference position ) the induced voltages E1 and E2 are
equal andopposite. Hence they cancel out and the output voltages V0 is zero.When the external
applied force moves the core towards the coil S2 ,E2 is increased but E1 isdecreased in
magnitude though they are still antiphase with each other. The net voltage available is(E2-E1)
and is in phase with E2.
Similarly , When movable core moves towards coil S1, E1>E2 and Vo = E1-E2 and is in phase
with E1.
ADVANTAGES
1. It gives a high output and therefore many a times there is no need for intermediate
amplification
devices.
2. The transducer possess a high sensitivity as high as 40V/mm
3. It shows a low hysteresis and hence repeatability is excellent under all conditions.
4. Most of the LVDTs consume a power of less than 1W.
5. Less friction and less noise
DISADVANTAGES
1. These transducers are sensitive to stray magnetic fields but shielding is possible .This is done
byproviding magnetic shields with longitudinal slots.
2. Relatively large displacements are required for appreciable differential output.
3.Several times, the transducer performance is affected by vibrations.
APPLICATIONS
The block diagram of dual trace oscilloscope which consist of following steps,
1. Electronics gun (single)
2.Separate vertical input channels ( Two)
3. Attenuators
4.pr-amplifiers
5. Electronic switch.
The two separate input signals can be applied to single electron gun with the helpof
electronic switching it Produces a dual trace display .Each separate vertical inputchannel are uses
separate attenuators and pr-amplifier stages, so the amplitude of eachsignal can be
independentlycontrolled. Output of the pr-amplifiers is given to theelectronic switch, which
passes one signal at a time into the main vertical amplifier of theoscilloscope.The time base-
generator is similar to that of single input oscilloscope.By using switch S2 the circuit can be
triggered on either A or Bchannel, waveforms, or an external signal, or on line frequency. The
horizontalamplifier canbefed from sweep generator or from channel B by switching S1. When
switch S, is in channelB, itsoscilloscope operates in the X-Y mode in which channel A acts as
the vertical inputsignal andchannelBasthe horizontal inputsignal.
From the front panel several operating modes can be selected for display, like channel
Bonly,channel A only, channels B and A as two traces, and signals A + B, A - B, B ~ A or - (A +
B)as a single trace. Two types of common operating mode are there for the
electronicswitch,namely,
1.Alternatemode
2.Chopmode.