Operating Focus Ellipsometer System(Tutorial)
Operating Focus Ellipsometer System(Tutorial)
Introduction This guide describes how to set up and use the FOCUS Ellipsometer
(FE) System for performing wafer thickness measurements and analy-
ses, and is made up of the following sections:
Glossary
Provides definitions of commonly used terms.
Intended This guide is mainly intended for those people who are responsible for
setting up wafers and cassettes for measuring and testing (Process
Audience Engineers), as well as those responsible for performing the actual
wafer measurements (Operators).
Usage The following table shows some of the conventions used in this guide.
Conventions
Usage
Meaning
Example
FOCUS Window names and areas of a dialog box appear in
Setup window initial capitals followed by the word “window” or “area”.
Bold Bold text identifies menu selections, window buttons,
and other text that may require particular attention.
NOTE
Cautions indicate the presence of a hazard that will or can cause prop-
erty damage (such as equipment damage, loss of software/data, or
service interruption) if the hazard is not avoided. Cautions are pre-
sented in the following manner:
CAUTION
Warnings indicate the presence of a hazard that will or can cause per-
sonal injury if the hazard is not avoided. Warnings are presented in the
following manner:
WARNING
Introduction The purpose of this chapter is to provide descriptions of the main hard-
ware and software components that make up the FOCUS Ellipsometer
System. The information in this section applies to the FOCUS Ellipsome-
ter (FE) III, IV, and VII systems, and includes the following topics:
1
3 2
• Control electronics
• Computer System (including keyboard and trackball pointing
device).
Measurement The measurement system, located behind the optics door on top of the
System measurement console, includes all of the hardware that actually performs
the film measurements.
Once a wafer is loaded onto the stage, it is detected and held in place by
the vacuum control system. The stage moves the wafer along the x and y
axes to position the desired measurement location under the laser spot.
Once properly positioned, the optical components perform the measure-
ments as programmed by the Engineer.
The FE System may be configured with either one laser (single wave-
length) or two lasers (dual wavelength).
Control The control electronics, located behind the left door panel on the bottom
Electronics of the measurement console, includes all of the electronic control mod-
ules (including rectifiers, transformers, and motion control circuitry) that
powers the FE System and contains the logic and drivers for all motion of
the robot, stage, and vacuum chuck.
Computer System The computer system, located behind the right door panel on the bottom
of the measurement console, is an IBM-compatible PC running OS/2.
The computer system is used to set up measurements and calculate the
data results. A VGA monitor (with live video overlay) and a trackball
pointing device are also part of the computer system.
Wafer Handler The Wafer Handler Console provides a storage area and contains the fol-
lowing components:
Console
• Wafer handler robot arm
• Cassette plates
• Printer (optional)
• External flat/notch finder (FE IV only)
Robot Arm The robot arm is used to transport wafers between a cassette and the
stage in the Measurement Console. Single arm and dual arm robots are
available.
Cassette Plates The cassette plates hold the wafer cassettes and contain sensors which
are used to automatically determine the size of the wafer cassette that
has been placed upon the plate.
FE System Printer A compartment located behind the door panel on the Wafer Handler Con-
sole provides for additional storage space for the FE System and may
also contain the optional printer for the system.
The printer allows you to generate hard copy of wafer maps and reports
that may be created when measurements are performed.
External Flat/ The external flat/notch finder, available only on FE IV systems, performs
Notch Finder centering and flat/notch orientation on a wafer prior to the wafer being
loaded onto the stage.
The external flat/notch finder takes up the space normally used by cas-
sette plate 3, therefore the FE IV system has only two cassette plates.
Monitor The optional Monitor Console is typically used to provide a mounting sur-
face for the computer monitor and for working counter space. The interior
Console of the console can be used for the storage of cassettes, tools, books and
other materials used with the FE System.
Optional This section describes the following FE System options that can be pur-
chased from Rudolph Technologies:
Equipment
• Dual wavelength
• SMIF unit
• Queued loading
• GEM compliant SECS-II
• Pattern Recognition (vision system)
• Printer
• Floor mounting
• Signal Tower
Dual Wavelength Dual wavelength uses a second laser to provide additional measurement
capability.
SMIF Unit The Standard Mechanical Interface (SMIF) unit provides a Class I mini
clean environment for the containment of cassettes on the FE System.
Queued Loading Queued loading improves throughput, productivity, and tool availability by
allowing data review, cassette removal, cassette setup and recipe setup
to occur while the system is measuring another wafer. In addition, the first
wafer from the next cassette may be pre-aligned while the last wafer of
the first cassette is being measured.
GEM-compliant The GEM-compliant SECS II option uses software and a standard 25-pin
SECS II connector to allow PC to remote mainframe communications.
Pattern The Pattern Recognition (or vision) system allows automatic alignment
Recognition for small site measurement.
(Vision System)
Printer An optional color printer may be connected to the FE System for printing
out wafer mapping information and measurement data reports.
2. Hard Feet
Levelers that screw into the bottom of the unit to lift and level the unit.
Signal Tower The signal tower option is an audible alarm and array of colored lights
which provide an easy indication of the status of the FE System. What
each light in the signal tower indicates is programmable through the
Focus Setup program.
DANGER
VISIBLE AND INVISIBLE LAS ER RADIATION
W HEN OPEN AND INTERLOCKS DEFEATED.
AVOID DIRECT EXPOSURE TO BEAM
Focus Operator The Focus Operator Queued Loading program is used by the Operator to
make measurements on production wafers using recipes created by the
Queued Engineer. After placing a cassette of wafers to be measured on a cas-
Loading sette plate, the Operator can select the desired measurement process to
measure the cassette of wafers, review system status, and start wafer
measurements.
Detailed instructions for using the Focus Operator Queued Loading pro-
gram are provided in Chapter 2, “Operator Interface”.
Detailed instructions for using the Focus Interactive program are pro-
vided in Chapter 3, “Developing Film Applications”.
Focus Recipe The Recipe Creator program is used by the Engineer to create wafer rec-
ipes which are incorporated into process steps. A “building block”
Creator approach is possible, where recipe components (templates) can be cre-
ated and stored for later assembly into complete recipes, or data can be
entered to create complete individual recipes.
After the necessary recipes have been created and stored by the Engi-
neer, the Operator can call up the desired process step in the Focus
Operator program and perform the wafer measurements.
Detailed instructions for using the Focus Recipe Creator program are
provided in Chapter 4, “Creating Recipes”.
Focus Mapping The Focus Mapping program permits the contents of the wafer informa-
tion database, which stores mapping files created during wafer measure-
ments, to be displayed as contour (2D) or topographic (3D) maps.
Detailed instructions for using the Focus Mapping program are provided
in Chapter 5, “Wafer Mapping and Data Reporting”.
Focus Browser The Focus Browser permits stored measurement data to be viewed and
organized as desired. The Focus Browser can be used to perform statis-
tical analyses and database queries and generate output in the form of
suitable reports, charts and graphs.
Detailed instructions for using the Focus Browser program are provided
in Chapter 5, “Wafer Mapping and Data Reporting”.
Focus Setup The Focus Setup program is typically used by Rudolph support person-
nel to change the FOCUS Ellipsometer software configuration, including
all the default parameters and settings for the various FE programs. The
Focus Setup program also provides for the creation of FE System login
names, data file backup, data file restore, and database maintenance
functions.
Focus SECS-II Using the optional SECS-II/GEM protocol makes it possible for the FE
System to communicate with a host computer. The SECS-II program pro-
vides a programmable interface where the SECS-II controls, communica-
tions and host monitor controls can be specified. Complete host control of
the FE System is possible using SECS-II.
When the FE System is powered up, the SECS-II program begins to exe-
cute in the background, allowing the Operator to run another program in
the foreground simultaneously. The program provides a control panel that
allows the Operator to configure a number of SECS-I and SECS-II inter-
face characteristics and parameters.
NOTE:
Introduction The purpose of this chapter is to provide information to both the Operator
and the Engineer on how to run a cassette of wafers and retrieve data.
The information in this section pertains to the FOCUS Ellipsometer (FE)
III, IV, and VII systems, and includes the following topics:
The tasks described in this chapter are performed with the use of the
Focus Operator Queued Loading program. This is the program that runs
the process steps created by the Recipe Creator and, if your FE System
is so equipped and configured, allows the “queuing” of additional cassette
wafers while a current cassette is running. Queuing allows for uninter-
rupted measurements, thus increasing overall system throughput.
Starting To start the Focus Operator Queued Loading program, perform the fol-
lowing procedure:
Focus
1. Position the trackball pointer over the Focus Operator Queued
Operator and Loading icon, located in the Focus Ellipsometer folder, and double
Logging In click the left trackball button.
A Focus Ellipsometer Operator Interface splash screen is displayed
briefly followed by one of the screens shown below.
If your FE System has been configured in such a way as to require
Operators to log in to the system, a screen such as the one shown in
Figure 2-1 is displayed. Continue the procedure with Step 2.
2. With the Login screen displayed as shown in Figure 2-1, position the
trackball cursor over the Login button on the screen and click the left
trackball button one time.
A list of the login names that are available on your system is dis-
played as shown in Figure 2-3.
3. Select a login name by clicking the left trackball button on the desired
name and then clicking on Enter, or by double clicking the left
trackball button on the name.
The Focus Operator Password Entry dialog box is displayed as
shown in Figure 2-4.
NOTE
Focus The Focus Operator program consists of three main views or windows.
These are the Cassette View, Wafer View, and Data View windows.
Operator
Windows The following sections provide descriptions of the parts that make up
each of the windows.
Cassette View When the Focus Operator software is first started, the Cassette View win-
dow is displayed. The appearance of the Cassette View window will vary
depending on the system setup and the task that is currently running.
Figure 2-5 shows the Cassette View window as it first appears if the FE
System is configured to require Operator login. Descriptions of the vari-
ous parts of the screen follow the figure.
SECS-II Status
Indicator
Robot
Arm
Cassette
Cassette
Plate Status
Window
Figure 2-6 shows the Cassette View window during process specification
selection. Items unique to this window are described below.
Figure 2-7 shows the Cassette View window during process step selec-
tion. Items unique to this window are described below.
Figure 2-8 shows the Cassette View window when the process is ready
to run on the selected cassette plate. Items unique to this window are
described below.
Cassette
Icon
Figure 2-9 shows the Cassette View window when the process is running
on the selected cassette plate. Items unique to this window are
described below.
• Control Panel Button — Displays the Wafer View window and the
FE System Control Panel. With the control panel, the Operator may
abort the entire measurement process run, skip the wafer currently
being measured, or select a different view to be displayed (video,
data, or cassette). The measurement process is paused while the
control panel is displayed.
Wafer View When a measurement process specification and process step have been
selected and assigned to a cassette plate and the run is started, the
Wafer View window is displayed. When all measurements in the run have
been completed, the Wafer View window is automatically closed and you
are returned to the Cassette View window.
Figure 2-10 shows an example of the Focus Operator Wafer View win-
dow. Descriptions of the various parts of the screen follow the figure.
- Focus — Clicking on this menu item will focus the laser spot to
a point. When measurements are performed, focus is
automatically performed.
Data View When a measurement process is running, the Data View window may be
displayed (in addition to the Wafer View window) by clicking on Control
Panel in the Wafer View window, selecting Data View, then clicking
on Enter.
Figure 2-11 shows an example of the Focus Operator Data View window.
Descriptions of the various parts of the screen follow the figure.
The Data View window displays the measurement data for each point on
each wafer as it is taken. When all measurements in the run have been
completed, the Data View window is automatically closed and you are
returned to the Cassette View window.
• Exit Menu — Exits the Data View window and returns you to the
Wafer View window.
Using Focus The FE System can measure both patterned and unpatterned wafers.
Operator From the operating point of view, pattern recognition (vision) recipes are
not different from non-pattern recognition recipes. The Operator selects
the process specification, selects the process step, and presses Start at
the confirmation screen. The Focus Operator can then perform the nor-
mal wafer load, wafer centering and flat/notch routines.
After the wafer is centered and the flat/notch is found, the vision system
registers the wafer by searching for the two registration points previously
trained. Once the sites are found, the system proceeds to each stepper
group to measure the selected sites.
The display is slightly different during operation. The site locator window
displays a simulated map of the chip and measurement sites on the
wafer. In the live video window, vision measurement sites are repre-
sented by squares. Non-vision sites are represented by an “X”. When the
system moves to a site, the box will be drawn at the location where the
site was predicted to be (by the numbers). Once the vision system cor-
rects the position, the blue diamond will separate from the box showing
the difference between the predicted location (box) and the measurement
location (blue diamond).
This process and the various windows that make up the Focus Operator
program will be discussed in more detail in the sections that follow.
Figure 2-12 shows the Focus Operator Queued Loading screen after the
Operator has logged in and prior to the selection of any process specifi-
cations.
Setting Up a 1. Place the cassette(s) that contain the wafer(s) to be measured on the
desired cassette plate(s) on the FE System.
Process Run
2. Select a Process Specification from the window list either by double
clicking on the desired name, or by single clicking on the name and
then clicking the Enter button.
A Process Step window will be displayed (as shown below) with a list
of the process steps defined for that process specification.
3. Select a Process Step from the window list by clicking on the name
of the desired Process Step then clicking on the appropriate cassette
plate. You may also click on Process Spec to return to the Process
Specification window to select a different name.
NOTE
NOTE
Cancel the Once the selected cassette plate says Ready and the Start button
appears, you may cancel out of the process run by clicking one time on
Process Run the cassette plate.
A confirmation box will appear asking if you wish to clear the process pro-
gram on the selected cassette plate. Click on Yes to clear the program,
No to return to the Focus Operator screen and continue with the pro-
gram. Return to Step 1 of “Setting Up a Process Run” on page 2–18 to
define a new process to run.
Start the When the cassette plate says Ready and the Start button appears (as
shown in Figure 2-14) click on Start.
Process Run
Typically, the Wafer View window is now displayed as shown in
Figure 2-15. However, your FE System may have been configured to dis-
play either the Cassette View or Data View window.
For a detailed description of the Wafer View, Cassette View, and Data
View windows, and of the information that is provided by these windows,
refer to “Focus Operator Windows” on page 2–6.
When the process run has started, from the Wafer View window you will
be able to observe the FE System register and center the selected wafer,
and then perform the specified measurements as defined by the process
specification and steps selected previously.
Upon completion of the process run, the run data is available for retrieval
and a screen as shown in Figure 2-16 is displayed.
Process Run When the process run has completed, the Focus Operator Cassette View
window is redisplayed and the cassette plate on which the process was
Completion run turns into an icon that says PRESS FOR DATA.
Click on PRESS FOR DATA to view the report from the process run that
has just completed. The report data, in text format, is then displayed on
the screen.
This report may be printed and/or saved to a file where it may then be
imported into a text viewer or spreadsheet.
For more information on data retrieval and report files, refer to “Retrieving
Run Data” on page 2–30.
Skipping a While the process is running, it is possible to skip the wafer currently
being measured by performing the following:
Wafer
Measurement 1. From the Wafer View window (Figure 2-15) click on Control Panel.
The Focus Ellipsometer Control Panel window is displayed as shown
in the figure below.
Aborting a While the process is running, it is possible to abort the entire session by
performing the following:
Measurement
Session 1. From the Wafer View window (Figure 2-15) click on Control Panel.
The Focus Ellipsometer Control Panel window is displayed as shown
in the figure below.
Setting Up If the queued loading option has been enabled on your FE System, the
Focus Operator program allows “queuing” of one or two additional cas-
Queued sette(s) of wafers (depending upon the configuration of your FE System)
Loading while a current cassette is running. This allows uninterrupted measure-
ments, thus increasing overall system throughput.
Cassettes may be queued either prior to starting the first process run or
while a process is already running.
Queuing Multiple To queue up multiple cassettes prior to starting the first run, use the fol-
Cassettes Prior to lowing procedure:
Starting the Run
1. Set up the process run for the first cassette following the instructions
provided in “Setting Up a Process Run” on page 2–18 of this manual
but DO NOT press Start after the cassette plate indicates Ready.
2. Set up the process run for the second and/or third cassette also by
following the instructions provided in “Setting Up a Process Run” on
page 2–18 of this manual.
3. Once each of the cassettes have been set up, the Focus Operator
Cassette View window shows which cassette plates are ready and
the run information for each cassette. Figure 2-19 shows an example
of the Cassette View screen with two cassettes queued and ready to
run their programs.
NOTE
5. Once the run has completed on all cassettes, the data is available for
review. Proceed to “Queued Run Completion” on page 2–29.
Queuing Another To queue up another cassette while a program is running on the first cas-
Cassette While a sette, perform the following:
Run is in Progress
1. In the Wafer View window click on the Cassette View button.
The Focus Operator Cassette View window is displayed as shown in
Figure 2-20.
3. Set up the process run for the second and/or third cassette by
following the instructions provided in “Setting Up a Process Run” on
page 2–18 of this manual.
The Focus Operator Cassette View window will now display the infor-
mation for the cassette that is currently running, as well as the cas-
sette that is queued (as shown in Figure 2-21).
4. You may either remain in the Cassette View window or return to the
Wafer View window by clicking on the Wafer View button.
5. Once the run has completed on all cassettes, the data is available for
review. Proceed to “Queued Run Completion” on page 2–29.
Queued Run When the process runs have completed, the Focus Operator Cassette
Completion View window is redisplayed and the cassette plates on which the pro-
cesses were run turn into an icon that says PRESS FOR DATA.
Click on PRESS FOR DATA on one of the cassette plates to view the
report from the process run that has just completed. The report data, in
text format, is then displayed on the screen.
This report may be printed and/or saved to a file where it may then be
imported into a text viewer or spreadsheet.
For more information on data retrieval and report files, refer to “Retrieving
Run Data” on page 2–30.
Retrieving After each process run, a data file (or report) is generated by the FE Sys-
tem for that run. The data in the report is available for viewing on-line
Run Data either immediately after the completion of the run, or from the history log
file. The reports may then be saved as a text file and/or printed directly to
the printer connected to the FE System.
Figure 2-23 illustrates the two icons on the Focus Operator Cassette
View window that may be used for data retrieval:
• The “PRESS FOR DATA” Icon — Used to view the data associated
with the measurements just completed on that cassette plate.
• The History Log File Icon — Used to call up a history log which lists
the reports available for viewing on the system.
When you have finished viewing the report on-line, you may:
• Save the data in the report to a text file by clicking on the File
menu, then clicking on Save As. Enter the desired destination path
and filename in the dialog box that is displayed and click on Enter.
The data is saved in the specified file and may be imported into a
text editor or spreadsheet.
• Upload the data to a host by clicking on the Upload menu item. The
SECS-II interface must be on-line.
• Quit the report viewer by clicking on the Exit menu item. This exits
the report viewer, however the data contained in the report still
resides on the FE System in the history log file.
Accessing The FE System maintains a history log of previously run data. The num-
ber of reports that will be saved in the history log is determined by your
Previously Run system setup. Typically, the default is 10.
Data (History
Log) CAUTION
1. Click on the History Log File icon (as shown in Figure 2-23).
The History Log File Selection window is displayed as shown in the
figure below. This selection window shows all of the reports currently
saved in the history log, the date and time the measurement run was
completed, the Operator who ran the process, as well as the process
specification and step used to perform the measurements.
2. Click on the desired report and then click on Enter, or double click on
the report.
The selected report is displayed similar to the sample report shown in
Figure 2-24. The data that makes up the report is explained in greater
detail in “Sample Report Data” on page 2–34.
• Save the data in the report to a text file by clicking on the File
menu, then clicking on Save As. Enter the desired destination path
and filename in the dialog box that is displayed and click on Enter.
The data is saved in the specified file and may be imported into a
text editor or spreadsheet.
• Quit the report viewer by clicking on the Exit menu item. This
exits the report viewer, however the data contained in the report still
resides on the FE System in the history log file.
Sample Report This section provides a brief description of the information that makes up
a report file.
Data
Figure 2-26 provides a sample of a basic report file.
Wafer number 1
Lot ID: L-13288
2
Wafer ID: W-13288A
Time: Wed Dec 4 15:14:45 1996
Points Measured: 5
3 Filmstacks: 1
F1 = aa_SiO2_<1um_T
Parameters Measured: 1 L1T
Wafer Statistics
5
F1 L1 T : Avg 2183.50 Min 2183.5 Max 2183.5 Std. Dev 0.000
Wafer number 2
Lot ID: L-13288
Wafer ID: W-13288B
Time: Wed Dec 4 15:14:47 1996
6
.
.
.
End Of Report
Creating a The FE System has the capability of creating contour and topographical
(3D) maps of a wafer. In addition, etch rate and difference maps may also
Wafer Map be created.
Each of the wafer maps may be displayed in black and white, greyscale,
or color.
NOTE
2. Double click on a map name, or click once on the desired name and
click on Modify.
The selected slot and map appear in the Slot: and Map: fields.
3. Enter a new map name by typing over the information in the Map:
and/or Slot: fields
CAUTION
4. Click on Modify to change just the one map name or click on Modify
All to change the name of all of the wafer maps to same name with
the slot number as the file suffix.
The new map name(s) appear in the Wafer Map Name window.
5. If you are specifying individual names for each wafer map (that is,
clicking on Modify after entering the map name), repeat Steps 3 and
4 for each of the wafer maps until all map names have been entered.
6. Once all of the map names have been modified, click Enter.
The Cassette View window is displayed and the selected cassette
plate says Ready.
NOTE
Logging Out To log out and exit the Focus Operator Queued Loading program, per-
form the following procedure:
and Exiting
1. On the Focus Operator Interface screen, click on Exit.
Focus
Operator If Login Exit has not been enabled on your FE System, the Focus
Operator Queued Loading program will exit and you will be returned
to the OS/2 desktop.
If Login Exit has been enabled on your FE System, a list of the
names that are available on your system is displayed as shown in
Figure 2-3.
NOTE
NOTE
Applications
The tasks described in this chapter are performed with the use of the
Focus Interactive program. This is the program that allows the Process
Engineer to develop and test filmstack models that are used in the cre-
ation of wafer recipes and process specifications run by the Operator.
Introduction The FOCUS System provides a simple and straight forward method for
application development. Standard materials can be easily arranged into
to Film layers to produce filmstacks. Parameters such as thickness (T), refractive
Application index (N), and extinction coefficient (K) can be calculated for
these layers.
Development
NOTE
Figure 3-1 shows a basic block diagram of the filmstack development and
testing process for the measuring of film layers.
In Focus Interactive
Choose Existing Filmstack Model
If Necessary
Modify or Develop
New Filmstack Model
Analyze Results
for Accuracy & Repeatability
- Fit Error
- Error Estimates
- Order Resolution
Filmstacks are the interface between the user and the computation rou-
tine (algorithm). The parameters that describe a wafer are contained in
the filmstacks. The descriptive parameters, and the film parameters to be
calculated can be changed when the filmstack creation window is open.
When creating a new application, the filmstack is made first, then tested
by measuring the wafer to which the filmstack applies. The first measure-
ment should reveal whether or not the desired results can be obtained. It
may be necessary to adjust some variables before the desired results
are obtained.
After the results of the first measurement have been analyzed, the film-
stack is revised to improve the expected results, and the wafer is mea-
sured again. This loop continues until no further improvement can be
achieved (or the required results are obtained), then the filmstack
is saved.
Starting To start the Focus Interactive program, perform the following procedure:
Focus 1. Position the trackball pointer over the Focus Interactive icon, located
in the Focus Ellipsometer folder, and double click the left trackball
Interactive button.
and Logging In If your FE System does not require a Login for the Focus Interac-
tive program, the Focus Interactive main window is displayed as
shown in Figure 3-2.
2. Select a login name by clicking the left trackball button on the desired
name and then clicking on Enter, or by double clicking the left
trackball button on the name.
NOTE
NOTE
Focus The Focus Interactive main window is divided into the following areas:
the Wafer View, Site Locator (including the Location information window),
Interactive Filmstack, and Calculated Parameters area.
Main Window These areas are described in the sections that follow.
NOTE
Wafer View The Wafer View area provides a graphical representation of the wafer. It
may be used to move around on the wafer by using the scroll bars or by
Area double clicking the right trackball button directly on a spot on the wafer.
Site Locator Also called the Live Video window, this window shows a live video image
of the wafer surface as it appears in the FE System. It may also be used
Area to move around on the wafer by using the scroll bars or by double clicking
the right trackball button directly on a spot on the wafer. The small blue
spot in the window indicates where the laser is focused on the wafer.
The Location information in the Site Locator area shows the X and Y
coordinates for the location on which the laser is currently focused.
Site Locator Menu The Site Locator area also consists of the following menu selections:
• Focus — Clicking on this menu item will focus the laser spot to a
point. When measurements are performed, focus is automatically
performed.
- Center — Moves the wafer so that the laser is located over the
center of the wafer.
- Left — Moves the wafer so that the laser is located at the left
outer edge of the wafer.
- Top — Moves the wafer so that the laser is located at the top
outer edge of the wafer.
Indicator Meaning
Asterisk (✱) Shown next to a parameter in a filmstack,
indicates that parameter is being calculated.
@ symbol in a red An order search is being performed on that
highlighted layer layer. Order search is a method of obtaining
absolute thickness with no order ambiguity.
Classic single-wavelength, or single angle, ellip-
sometry is unable to resolve order due to the
cyclical behavior of the ellipsometric parame-
ters. The FOCUS multiple angle and multiple
wavelength technique reduces (improves) order
ambiguity. For more information, see “Order
Searching and Order Resolution” on page 3–34.
Filmstack Menu The Filmstack area also consists of the following menu selections:
• Ranges — Allows you to set the maximum fit error allowed and the
action the FE System will perform if a measurement is out of range.
Options include: warn and continue, warn and pause, reject, and
switch to a different filmstack.
This allows you to assess the validity of the filmstack for your application.
Focus The following menu items are available in the Focus Interactive Main win-
dow:
Interactive
Main Menu • Load Wafer — Used to load and unload a test wafer either
manually or automatically using the FE robot.
NOTE
NOTE
NOTE
NOTE
NOTE
NOTE
NOTE
NOTE
CAUTION
Load a Test A sample or test wafer is used to ensure that the filmstack model and rec-
ipe that you are developing will work on similar production wafers.
Wafer
Use the following procedure to load wafers to the stage, either manually
or from a cassette, define the wafer size, loading action, wafer orientation
and wafer registration:
2. Determine the method you wish to use to load the wafer and perform
one of the following:
Select the cassette and slot from which to load the wafer and click
on Enter. The Wafer Specification window is displayed as shown in
Figure 3-10.
3. Select the information that applies to the test wafer (wafer size,
flat/notch information, centering, and wafer orientation) then click
on the Enter button.
NOTE
NOTE
Choosing a This section describes how to choose a filmstack model that will later be
incorporated into a wafer recipe used for measuring production wafers.
Filmstack
Model There are three general procedures that can be used to choose
a filmstack:
NOTE
NOTE
NOTE
Modifying an The FE System software database contains many Rudolph supplied film-
Existing Filmstack stack models covering a wide range of possible applications. For special
applications, these Rudolph supplied filmstacks can be copied and the
new copy modified as follows:
CAUTION
NOTE
NOTE
NOTE
Creating a New The FE System allows for the creation of entirely new filmstack models to
Filmstack suit your particular application. The new filmstack is built layer by layer on
the default substrate. The default substrate is typically silicon (Si).
2. Enter a name for the new filmstack in the Name field and click on the
Enter button.
NOTE
NOTE
Logging After the filmstack has been developed, you may wish to save measure-
ment data to a log file. Use the procedures below to start and stop data
Measurements logging.
NOTE
3. Enter the desired log file name in the Name field (eight characters or
less with an extension of .log) and click on Enter.
CAUTION
5. Stop data logging once all measurement data has been collected.
- Click on the icon in the upper left corner beside the Logging
Measurement Data title then click on Close from the menu that
is displayed.
Testing the Once the filmstack has been developed, it may be tested by making sam-
ple measurements of the wafer. By default, the FE System will perform a
Filmstack single measurement at the center of the wafer. For higher confidence in
Model the test results, and to determine how robust the filmstack is, you may
perform test measurements on multiple locations.
NOTE
Focusing
Setting intensity
Measuring wafer
Modeling data
Analyzing Test The FE System provides three indicators to help you determine the accu-
racy and repeatability for using a filmstack model on your application:
Results
• Fit error
• Error estimates
• Order resolution
The degree of correlation between the model and the actual physical film-
stack is interpreted by the fit error, which is based on a χ2 fit between the
measured and modeled ellipsometric angles. Error estimates may also
be used to qualitatively determine the accuracy of the measured parame-
ters.
Fit Error After a measurement is made, the calculated parameters and the fit error
are returned. When performing calculations, the FE software searches
for the solution with the lowest fit error. The fit error is a measure of how
well the measured data fits the theoretical Delta (∆) and Psi (Ψ) data of
the film model. The statistical quantity “Chi squared” (χ2) is used to mea-
sure fit error.
Usually, the measured Delta (∆) and Psi (Ψ) data are correct to within the
accuracy of the measurement.
Therefore, a large fit error means there is a problem with the filmstack
model (one or more of the assumed parameter values is incorrect). You
must change the filmstack parameters and re-test the filmstack model.
Fit error can also be used to detect a measurement related problem. For
example, if the laser spot is hitting the edge of a measurement site on a
product wafer, the fit error will be much larger than when the laser spot is
well centered on the site.
Error Estimates Error estimates provide an assessment of the accuracy of the computed
film parameters. As a general rule, the error estimates should be approx-
imately 1% of the measured value. Thinner films (<100Å) may have
higher error estimates. Usually, as more parameters are calculated, the
error of each parameter is increased, while the fit error is decreased.
Fewer calculations provide better accuracy.
Order Searching Because of the cyclic nature of the interference of light when measuring
and Order thin films, the ∆/ψ pair obtained from a given measurement is not unique.
Resolution These values are periodic with the thickness of the film. Every material
has an associated order thickness, or “Cycle Thickness.” Cycle Thick-
ness (Tc) is a function of refractive index, wavelength of light and angle of
incidence. Figure 3-23 expresses the mathematical definition of this rela-
tionship.
λ
T c = --------------------------------
2 2
2 n – sin θ
NOTE
With other variables remaining constant, ∆/ψ values will be the same for
any multiple of Tc. For example, at a wavelength of 632.8 nm, AOI of 70
and an RI of 1.462, the Tc is 2825Å. Measurement of this material (SiO2
in this case) indicates a thickness of 6000Å as well as 6000Å ± mTc
where m =1,2,3.... . All have the same ∆/ψ pair.
The best three answers (by fit error) are generated and a calculation is
performed to determine whether the order is resolved.
If the next best fit is ≥1.5 times the best fit, and the next best fit error is
greater than 2.0, the order is resolved. If the next best fit is ≥1.5 times the
best fit, and the next best fit error is less than 2.0, then order is not
resolved.
Let:
x = Best Fit
y = Next Best Fit
If order is not resolved by fit error, and index or refraction (N) or absorp-
tion (K) are being computed, then the algorithm resorts to “index match-
ing”. Index matching chooses the solution whose measured index values
are within specified tolerances so that order can be resolved. If order
cannot be resolved, the solution whose thickness is closest to the starting
value will be returned.
Remodeling a Once a measurement is completed, if you are not satisfied with the
results of the measurement, you may want to modify the filmstack model
Modified and recalculate the measurement. To modify the filmstack model, you
Filmstack may change the parameters selected for calculation, choose a different
material, add or delete a layer, or switch from dual to single wavelength.
By using the ReModel option, the filmstack model can be quickly recalcu-
lated without making another measurement.
NOTE
2. Once the filmstack has been modified and saved, in the Focus
Interactive main menu, click on ReModel.
The ReModel menu item is highlighted.
Unloading the When you have completed measurements on the wafer, use the following
procedure to unload the wafer:
Wafer
1. In the Focus Interactive main menu, click on Load Wafer.
The Load Wafer menu is displayed.
• If the wafer was loaded using Cassette Load: the robot returns
the wafer to the cassette and slot from which it was initially taken.
• If the wafer was loaded using Manual Load: the stage moves
forward and prompts for the wafer to be manually removed.
A filmstack model has now been created and can be incorporated into a
wafer recipe as described in Chapter 4, “Creating Recipes”
Creating and The following procedures should be used to change the filmstack sub-
strate, add, delete, or modify filmstack layers, select filmstack measure-
Modifying ment controls, and to set range specifications.
Filmstacks
Changing the To change the substrate in the filmstack model, you must have already
selected, copied, or created a filmstack using the procedures in “Modify-
Filmstack ing an Existing Filmstack” on page 3–22 or “Creating a New Filmstack”
Substrate on page 3–26 respectively.
2. Enter the values for the substrate in the Substrate View window using
the information below as a guide.
NOTE
• Spec
- N — Displays the N value (index of refraction) based on the
selected material.
NOTE
Adding Film To add a layer to the filmstack model, you must have already selected a
filmstack to modify or specified the name for a new filmstack using the
Layers procedures in “Modifying an Existing Filmstack” on page 3–22 or “Creat-
ing a New Filmstack” on page 3–26 respectively.
NOTE
2. Select the material for the new layer from the list by clicking once on
the item then clicking Enter, or by double clicking on the item.
A Layer View window is displayed as shown in Figure 3-27 and
Figure 3-28. The Layer View window in Figure 3-27 shows a single
wavelength measurement, the Layer View window in Figure 3-28
shows a dual wavelength measurement. The window displayed will
depend upon the type of filmstack being built.
NOTE
3. Enter the values for the selected material in the Layer View window
using the information below as a guide.
• Spec
- Thickness — Enter the approximate thickness (in Angstroms)
for the layer.
NOTE
- None — The FE System will find the closest local minima of fit
error.
4. When all entries have been made for this layer, click the Enter button
in the Layer View window (Figure 3-28).
You are returned to the Filmstack Model window (Figure 3-12) with
the new layer displayed as the top layer of the filmstack model.
Creating New The FE System contains an extensive database of film materials. How-
ever, you may create new materials to suit your application by using the
Materials procedure in this section.
NOTE
1. With the Material Selection window displayed, enter a name for the
new material in the Name field (such as Poly15%) and click on the
Enter button.
A window is displayed asking if this is an EMA (Effective Medium
Approximation) composite material.
Creating Non- 1. In the window asking if this is an EMA composite material, click on
EMA Composite No.
Materials A Material window is displayed that allows you to enter the N, K, dN,
and K2 values for the newly created material.
2. Enter the N, K, dN, and K2 values for the material and click on the
Enter button.
The Material Selection window is displayed with the newly created
material highlighted.
4. In the Layer View window, enter the desired information for the
material using the procedure given in “Modifying Film Layers” on
page 3–50.
Creating EMA 1. In the window asking if this is an EMA composite material, click
Composite on Yes.
Materials The EMA Material window appears with the material name specified
as shown in Figure 3-30.
3. In the Constituent window, click on the Material button and select the
first component of the composite material from the Material Selection
window.
The selected material is displayed in the Constituent window.
8. In the Layer View window, enter the desired information for the
material (such as spec thickness, ranges, and the parameters to be
calculated) using the procedure given in “Modifying Film Layers” on
page 3–50.
Deleting Film To delete the top layer of the filmstack model, you must have already
selected or copied a previously defined filmstack using the procedures in
Layers “Modifying an Existing Filmstack” on page 3–22.
You are returned to the Filmstack Model window and the top layer
is removed from the filmstack.
3. Repeat this procedure if you wish to delete another layer from the
filmstack model.
Modifying Film To modify an existing layer in the filmstack model, you must have already
selected, copied, or created a filmstack using the procedures in “Modify-
Layers ing an Existing Filmstack” on page 3–22 or “Creating a New Filmstack”
on page 3–26 respectively.
1. In the Filmstack Model window, click on the layer that you wish to
modify.
A Layer View window is displayed that contains the information for
the selected layer. In the examples below the Layer View window in
Figure 3-34 shows a single wavelength measurement, the Layer
View window in Figure 3-35 shows a dual wavelength measurement.
The window displayed will depend upon the type of filmstack being
modified.
2. Enter the values for the selected layer in the Layer View window
using the information below as a guide.
• Layer Material — Click on the button with name of the material that
makes up the layer to display the Material Selection window as
shown in Figure 3-26. Select the desired material from the list by
clicking once on the item then clicking Enter, or by double clicking
on the item. The name of the selected material is displayed in the
Layer Material button in the Layer View window and the specs for
N, K, dN, and/or K2 (as applicable) are updated.
NOTE
• Spec
- Thickness — Enter the approximate thickness (in Angstroms)
for the layer.
NOTE
- None — The FE System will find the closest local minima of fit
error.
3. When all entries have been made for this layer, click the Enter button
in the Layer View window (Figure 3-35).
You are returned to the Filmstack Model window with the layer modifi-
cations displayed in the filmstack model.
4. Repeat this procedure if you wish to modify another layer from the
filmstack model.
Selecting To select a measurement control for the filmstack model, you must have
already selected, copied, or created a filmstack using the procedures in
Measurement “Modifying an Existing Filmstack” on page 3–22 or “Creating a New Film-
Controls stack” on page 3–26 respectively.
2. Select a measurement control for the layer from the list by clicking
once on the item then clicking Enter, or by double clicking on the
item. Select DEFAULT to use the standard measurement mode
(equivalent to thin mode).
You are returned to the Filmstack Model window. The selected mea-
surement control is displayed in the Meas: button.
NOTE
Setting Range To set the maximum fit error for the filmstack and the action that the FE
System will take when a measurement is out of range, you must have
Specifications already selected, copied, or created a filmstack using the procedures in
“Modifying an Existing Filmstack” on page 3–22 or “Creating a New Film-
stack” on page 3–26 respectively.
2. Enter the desired maximum fit error and select the out-of-range
action to be taken when a measurement exceeds the maximum fit
error.
4. When you have completed the range specifications for the filmstack,
click on the Enter button in the Range Specifications window.
You are returned to the Filmstack Model window.
Faxing Data In the event that you encounter difficulties developing your filmstack
model, or if you need assistance in modeling a wafer, measurement data
to Rudolph may be faxed to the Rudolph Applications Support group. This will allow
Technologies the simulation of your filmstack modeling at the Rudolph Technologies
facilities.
6. Click on OK.
A Measured Data Selection window is displayed as shown in
Figure 3-39.
7. Enter lpt1 in the Name field and click on the Enter button.
The filmstack and measurement data is output to the printer that is
connected to the FE System.
From:
Name:
Company:
Phone
Fax
Comments:
Filmstack:
Film Stack: aa_SiO2_<1um_T
Exiting To exit the Focus Interactive program, perform the following procedure:
• Single click the left trackball button on the icon located in the upper
left-hand corner of the screen (beside the Focus Interactive
program name). Select Close from the menu that is displayed.
The Focus Interactive program will exit and you will be returned to the
OS/2 desktop.
Applications The following system applications and Fab areas are covered in the sec-
tions that follow. Examples of common applications, or “application notes”
by Fab Area are provided.
NOTE
Diffusion This section provides examples of the system applications within the Dif-
fusion area of the Fab. These include standard filmstacks such as Thin
Applications Oxide, Thick Oxide, Thin Nitride, and Thick Nitride, as well as custom
filmstacks including Poly Variations, and Poly-Si using EMA.
NOTE
Diffusion Introduction
Application:
The formation of silicon dioxide (SiO2) on a silicon substrate (oxidation)
Oxides/Nitrides is a popular process in the diffusion area of the Fab. Some functions of
thermally-grown oxide films are:
Silicon nitride (Si3N4) plays an important role in the diffusion area of the
Fab. It acts as a diffusion barrier, and is useful as a mask for the selective
oxidation of exposed silicon.
The following example will describe the procedure for developing and
testing a filmstack for measuring the thickness of a single layer of SiO2 or
Si3N4 on a silicon substrate.
Procedure
The expected fit error should be in the range of 0 to 0.5 for thin film
dielectrics, and 0 to 1 for thick film dielectrics.
NOTE
As shown in the calculated parameters window, the fit error for the mea-
sured thickness of the oxide/nitride layer, 1T, falls within the acceptable
range. Therefore, the filmstack chosen for this application can be reliably
used for production measurements in a Wafer Recipe.
Summary
Diffusion Introduction
Application:
Polycrystalline silicon (Poly-Si) films are generally used as gate elec-
Poly Variations trodes and interconnections in MOS circuits. Optimum resistivity for these
products can be achieved by the doping of polycrystalline films in the dif-
fusion area of the Fab. In most cases, an SiO2 layer is grown before the
Poly-Si is deposited, for purposes of electrical isolation. The FE System
can be used to measure multi-layer filmstacks and multi-parameters for
optical characterization of Poly-Si.
The following example will show the procedure for optimizing the film-
stack required for measuring Poly-Si thickness of a wafer with 1500Å of
Poly-Si deposited on 1000Å thermally grown SiO2. The interlayer feature
for improved modeling is explained below and the EMA feature for optical
characterization will be explained later in this section.
About Interlayers
Interlayers serve to reduce fit errors and improve order resolution. Some
applications model better with one or more interlayers. Poly films often
require these interlayers.
Although these rules allow some flexibility, very few applications will use
more than one interlayer.
Procedure
NOTE
This filmstack model has an initial oxide layer with a fixed value of
1000Å. Better knowledge of the bottom oxide thickness will provide
more accurate results for the calculated parameters.
The POLY-SI layer shows a spec thickness of 2500Å with optical con-
stants initially set at fixed values. The thickness of the Poly-Si layer is
being calculated with a Restricted Search applied.
NOTE
The fit error for this application can range from 0 to 10, depending on the
quality of the film. Higher fit errors may be expected for Poly-Si films with
disordered crystal structures. The fit error shown above falls within the
acceptable range, and the calculated thickness is close to the target
value. If needed, the fit error may be improved by calculating the absorp-
tion coefficient (K) of the Poly-Si layer. An incorrect fixed value of K may
introduce inaccuracies in the calculated thickness of the Poly-Si.
2. In the Layer View window, check the Calc. box for K and click on the
Enter button.
The fit error has decreased, and is approximately one half the value of
the fit error using the original filmstack. This indicates that by modifying
the filmstack to calculate the Poly-Si thickness and K, the results have
been improved.
Summary
Diffusion Introduction
Application:
Deposition and doping conditions can vary the optical properties. There-
Poly-Si Using fore, you might produce inaccurate results if you calculate a Poly-Si thick-
EMA ness while assuming values for the optical constants N (refractive index)
and K (absorption coefficient). EMA (Effective Medium Approximation)
provides the added capability of characterizing the optical constants, N
and K, of the Poly-Si.
The following simplified assumption is the basis for using EMA modeling:
The N and K values of the Poly-Si layer are determined by calculating the
volume fraction (Vf) of the A-Si content. With EMA, the N and K of the
material is assumed to be in proportion to the N and K of the composite
materials. If a Poly-Si film is 80% crystalline silicon and 20% A-Si, the N
and K should be 80% like silicon and 20% like A-Si.
Use EMA to characterize the optical constants of the Poly-Si, and create
a new material according to the calculated Vf of A-Si. Although the rules
for using EMA are application dependent, this feature works best with a
minimum Poly-Si thickness of 1500Å and a bottom oxide of at least 500Å.
NOTE
The following example will show the procedure for using EMA and creat-
ing a new material for measuring Poly-Si thickness of a wafer with 5000Å
of Poly-Si deposited on 1000Å thermally grown SiO2.
Procedure
NOTE
The fit error for this application can range from 0 to 10, depending on
the quality of the film. Higher fit errors may be expected for Poly-Si
films with disordered crystal structures. The fit error shown above is
acceptable, and the calculated thickness close to the target value.
11. Enter a name for the new material in the Name field (in this example
enter Poly15%) in the Material Selection window and click on the
Enter button.
A window is displayed asking if this is an EMA composite material.
12. Click Yes to create an EMA material that is created by combining the
two component materials, A-Si and Si.
The EMA Material window appears with the material name specified.
14. In the Constituent window, click on the Material button and select the
first component of the composite material (in this example, A-Si) from
the Material Selection window.
The selected material is displayed in the Constituent window.
16. In the EMA Material window, click on Add and select the second
component for the composite material by clicking on the Material
button in the Constituent window and selecting SI from the material
list.
The Constituent window is displayed with the selected material
shown. The FE System automatically determines that the composi-
tion is made up of 85% of the second component and displays 85 in
the Volume Fraction field.
20. In the Layer View window, click on the Calc. box for Vf to calculate
thickness only, then click on the Enter button.
The Filmstack Model window is displayed with the updated informa-
tion.
The fit error is the same as when calculating Vf and thickness simulta-
neously, but the filmstack will be more reliable for production measure-
ment purposes.
Summary
CVD This section provides examples of Thick Oxide, Amorphous Si, and
TiN/Si applications within the CVD area of the Fab.
(Deposition)
Applications NOTE
CVD Introduction
Application:
This example will describe setting up a filmstack model for wafers with a
Thick Oxide, thick oxide layer on top of silicon. This is a simple film structure which
T&N shows the FE System’s ability to calculate both thickness and refractive
Measurements index simultaneously. This procedure applies for filmstacks in the range
of 1000Å to 10000Å.
Procedure
Summary
CVD Introduction
Application:
This example will describe setting up a filmstack model for wafers with an
Amorphous Si amorphous silicon (A-Si) on an oxide layer. This is not a simple film struc-
ture; however, it is one often encountered in CVD.
All A-Si wafers have a native oxide layer present on the surface. This
layer is often considered negligible, but results have shown that by add-
ing a native oxide layer to the filmstack more stable measurement results
can be obtained.
Procedure
4. In the box under Thickness, type in your thickness value and click
Enter. If necessary repeat this step for the bottom oxide layer.
The thickness returned in the report is very good. As the A-Si layer’s
thickness increases, the fit error may increase, however, less than one is
usually acceptable. If the fit error is poor, one should question either the
accuracy of the assumed thickness on the bottom layer or the optical
constants of the amorphous silicon. For thicker amorphous silicon layers,
it might also be necessary to iterate on the extinction coefficient, K, if
there is difficulty resolving order.
Summary
The FOCUS Ellipsometer has the ability to accurately model and mea-
sure materials with multiple layers.
CVD Introduction
Application:
This example will describe setting up a filmstack model for wafers with a
TiN/Si non opaque Titanium Nitride layer on silicon (TiN/Si). The thickness value
of a TiN layer can be determined for thicknesses less than about 500Å,
depending on how absorbent the material is to light.
Procedure
NOTE
2. Enter an appropriate name for the filmstack in the Name field (such
as 250_TiN/Si) and click on the Enter button.
The Filmstack Model window is displayed as shown in Figure 3-60.
4. Select TiN from the list of available items by single clicking the left
trackball button on the item name and clicking on Enter, or by double
clicking on the name.
The Layer View window is displayed.
5. Enter the desired Thickness Spec (for this example, enter 250) and
click the Calc. checkbox under Thickness. When you have
completed entering the specifications, click the Enter button in the
Layer View window.
7. Select ad_Dual from the list of available items and click on the Enter
button.
The Filmstack Model window is displayed with the newly created film-
stack as shown in Figure 3-62.
Summary
Etch Introduction
Application:
Presently, few nondestructive metrology production tools are capable of
Oxide Etch to accurately detecting a successful etch, specifically the complete removal
Clear of a transparent oxide film on silicon substrate. Any remaining oxide can
adversely effect subsequent process steps by inhibiting interfacial reac-
tions of films deposited on the substrate, or by causing high contact resis-
tance.
When over etch occurs substrate damage causes the SiO2/Si model to
erroneously report the presence of oxide. Hence, this application requires
calculation of a more sensitive parameter in the film model to reliably
detect oxide etch to clear. Simultaneous calculation of both thickness (T)
and absorption (K) of the oxide reliably indicates both the complete
removal of the oxide and the over etch condition by a dramatic change
in K.
The following example will describe the procedure for developing and
testing a filmstack for detecting both the over etch and under etch condi-
tions.
Procedure
2. Enter the name for the new filmstack (in this example, Oxide_Clear)
and click on the Enter button.
The Filmstack Model window is displayed with the default substrate
material, SI.
5. Type in the target value in the Thickness field (in this example, 100)
and click on the Calc. checkboxes for both Thickness and K
(absorption). Click on the Enter button when you have completed the
selections.
The Filmstack Model window is displayed with the new material.
8. In the Filmstack Model window menu, click on Enter to save the new
filmstack in the database.
Point # T Only T K
1 145.24 145.59 0.002
2 137.13 136.53 0.001
3 74.88 74.13 0.007
4 148.19 148.14 0.002
5 169.22 169.64 0.007
Point # T Only T K
1 39.62 0.0 0.86
2 40.18 0.0 0.83
3 31.71 0.0 0.85
4 35.87 0.0 0.91
5 43.22 0.0 0.84
Table 3-4 shows data from a five point measurement of an over etched
wafer. The standard SiO2/Si only film model shows the presence of about
40Å of oxide in spite of the fact that the wafer has been over etched. The
proposed film model which calculates both T and K indicates a dramatic
change in K due to the damaged silicon induced by over etch. Notice the
difference in K in Table 3-4 as compared with K in Table 3-3 of the under
etched wafer. The magnitude of the change in K between the two condi-
tions allows the FOCUS Ellipsometer to accurately detect the complete
removal of the oxide layer without incorrectly reporting the presence of
oxide when over etch occurs.
Software features provided with the FOCUS Ellipsometer allow for a film
model which indicates to the user a complete and successful etch while
displaying the true oxide thickness if the wafer is under etched.
Summary
CMP This section provides examples of Teos on Metal, BPSG on Metal, and
PMD on Silicide applications within the CMP area of the Fab.
Applications
NOTE
CMP Introduction
Application:
This example will describe how to measure the thickness of a Teos layer
Teos on Metal on metal. In this case the AlCu is optically opaque and is considered to
be a substrate. The TiN layer can be either transparent or opaque. The
TiN and AlCu will be considered to be one effective metal substrate hav-
ing pseudo optical constants Ns and Ks.
It is well understood that the underlying metal layers can vary in thick-
ness and composition. Other measurement techniques can suffer from
order skipping problems in the ILD (Inter-Layer Dielectric – a layer placed
between metal films that acts as an insulating layer) thickness. This is
due to variations in the metal’s pseudo optical constants, and techniques
other than ellipsometry cannot compute the metal optical constants. With
FOCUS Ellipsometry, a user can measure the thickness of the ILD layer,
and the substrate optical constants Ns and Ks simultaneously. The Ns
and Ks parameters account for the process variations in the metal layers
and will stabilize the ILD thickness, thus avoiding order skipping.
NOTE
Procedure
3. Click on the TEOS layer in the Filmstack Model window to modify the
thickness of the Teos layer.
NOTE
4. In the Layer View window, enter the spec thickness of the Teos layer
and click on the Enter button.
The Filmstack Model window is displayed as shown in Figure 3-64.
In this example the substrate is the TiN layer. This represents the
combined TiN/AlCu substrate. For other applications the substrate
can be changed to other materials such as Ti, W-Si, or TiS2. All these
materials are present in the material library.
The thickness returned in this report is very good. Typically the fit error
should be less than 1 for Oxide measurements on Silicon. This sample
however is on metal and the fit error can range between 0-10. This is due
to the roughness and variation of the metal substrate.
Summary
To measure ILD layers, you must compute the thickness of the ILD along
with the pseudo-optical constants of the metal substrate. The Ns and Ks
parameters yield stable thickness measurements that will not order skip.
Only with ellipsometry can you measure optical constants.
CMP Introduction
Application:
For information pertaining to the measurement of a BPSG layer on metal,
BPSG on Metal refer to “CMP Application: Teos on Metal” on page 3–91.
Lithography This section provides an example of a Photo Resist application within the
Lithography area of the Fab.
Applications
NOTE
Lithography Introduction
Application:
This example will describe setting up a filmstack model for wafers with a
Photo Resist Resist on silicon. Resist is a straightforward application well suited for
measurement by the FOCUS Ellipsometer.
Procedure
2. Select the filmstack that is most appropriate for your application (such
as aa_Resist_<1um_T, and aa_Resist_>1um_T). The thickness of
the Resist will determine which filmstack should be chosen. In this
example the sample wafer has approximately 11000Å of Resist,
therefore the aa_Resist_>1um_T is applicable.
The Filmstack Model window is displayed as shown in Figure 3-66.
Summary
The FE System easily calculates the thickness and, if desired, the refrac-
tive index of photoresists on oxide. Choose from one of three wide thick-
ness ranges for accurate measurements.
Introduction The purpose of this chapter is to provide the Engineers and Supervisors
who are responsible for setting up measurement processes, with the pro-
cedures necessary to create wafer recipes. The information in this sec-
tion pertains to the FOCUS Ellipsometer (FE) III, IV, and VII systems, and
includes the following topics:
• How to log in, use, and exit the Focus Recipe Creator.
• Describe the components of the Focus Recipe Creator.
• How to create wafer recipes for monitor (bare) and product
(patterned) wafers to include:
- Wafer information
- Filmstacks
- Measurement patterns
- Wafer reports
• How to customize monitor wafer recipes.
• How to customize product wafer recipes to include:
- Pattern recognition
- Deskew
- Site-by-site pattern recognition
• How to create wafer mapping recipes.
• Vision system modes and thresholds.
• Vision system diagnostics.
The tasks described in this chapter are performed with the use of the
Focus Recipe Creator program. This is the program that allows the Pro-
cess Engineer to perform such tasks as; create process specifications
and recipes, and train and automate recognition of patterns on wafers.
NOTE
Figure 4-1 shows a basic block diagram of the recipe creation process for
both monitor and product wafers.
Choose Measurement
Site and Specify
Direct Drive
or Site by Site
Monitor recipes can be created from standard
recipes supplied with the program or custom
recipes may be developed
Select Measured
Chips
Wafer A Process Engineer will use the Focus Recipe Creator to develop a mea-
surement process using the filmstack created using the Focus Interactive
Recipes program (see Chapter 3, “Developing Film Applications” for information).
This measurement process will later be accessed by an Operator using
the Focus Operator Queued Loading program (see Chapter 2, “Operator
Interface” for information).
CAUTION
Recipes are combined with transfer and control instructions to form pro-
cess steps which in turn are grouped to form process specifications. This
organization is designed to match the process used at a semiconductor
Fab.
Process There may be several steps in the process specification, each step
involves measurements done on a wafer film layer after a production pro-
Specifications cess. Each process step contains one or more recipes. This relationship
and Process is shown in the figure below:
Steps
Process Specification
Process Step 1
Recipe(s)
Transfer
Control
Process Step 2
Recipe(s)
Transfer
Control
.
.
.
Starting To start the Focus Recipe Creator program, perform the following proce-
dure:
Recipe
1. Position the trackball pointer over the Focus Recipe Creator icon,
Creator and located in the Focus Ellipsometer folder, and double click the left
Logging In trackball button.
The Focus Recipe Creator main window is displayed as shown in
Figure 4-3.
If your FE System does not require a Login for the Recipe Creator,
you are now ready to begin the recipe creation process.
2. Select a login name by clicking the left trackball button on the desired
name and then clicking on Enter, or by double clicking the left
trackball button on the name.
NOTE
NOTE
Recipe The Recipe Creator Main window is divided into three areas: the Wafer
View, Site Locator (including the Location information window), and
Creator Main Wafer Recipe area.
Window These areas are described in the sections that follow.
Wafer View The Wafer View area provides a graphical representation of the wafer. It
may be used to move around on the wafer by using the scroll bars or by
Area double clicking the right trackball button directly on a spot on the wafer.
This window also indicates all the specified measurement sites and, dur-
ing measurement, the site currently being measured.
Wafer Recipe Displays information about the wafer recipe (including wafer information,
wafer report information, the measurement pattern to be used, and the
Area filmstack to be used for the wafer) that is currently being programmed.
This information is updated as the recipe is built.
Site Locator Also called the Live Video window, this window shows a live video image
of the wafer surface as it appears in the FE System. The small blue spot
Area in the window indicates where the laser is focused on the wafer.
The Location information in the Site Locator area shows the X and Y
coordinates for the location on which the laser is currently focused.
Site Locator Menu The Site Locator area also consists of the following menu selections:
• Focus — Clicking on this menu item will focus the laser spot to a
point. When measurements are performed, focus is automatically
performed.
- Center — Moves the wafer so that the laser is located over the
center of the wafer.
- Left — Moves the wafer so that the laser is located at the left
outer edge of the wafer.
- Top — Moves the wafer so that the laser is located at the top
outer edge of the wafer.
Recipe Creator The following menu items are available in the Recipe Creator Main win-
dow:
Main Menu
• Process — Named by the Engineer, a process specification
includes multiple process steps which can be added to or deleted
from a list of Steps (maximum of 25). Each process step includes a
Transfer, Recipe, and Control which identify the wafer(s) to be
measured, the recipes to be used, and how the data will be stored.
There are Transfer files supplied with the Focus software which
have significant meaning. For example: aaX1-1S1,12,25 indicates
cassette 1 as the source and destination of the wafers in slots 1,
12, and 25; aaX1-1S1-5Reject2 indicates cassette 1 as the source
and destination of the wafers in slots 1 through 5, sending the
rejected wafers to cassette 2. The DEFER Transfer allows the
Operator to specify which cassette and slot holds the wafer of
interest.
NOTE
NOTE
Creating a An FE wafer recipe specifies the composition and instructions for mea-
suring a wafer.
Wafer Recipe
To create a recipe, the Process Engineer must complete each step of the
Wafer Recipe menu as described below, ultimately saving the recipe and
using it in a process.
Many of the initial steps taken to create a wafer recipe are common to
both monitor (bare) and product (patterned) wafers. The information pre-
sented in this section is applicable to both monitor and product wafers.
CAUTION
Use the following procedure to create a new wafer recipe. For information
on how to select and modify existing wafer recipes, refer to “Modifying a
Wafer Recipe” on page 4–43.
Recipe Selection 1. In the Recipe Creator main menu, click on Wafer Recipe.
The Wafer Recipe menu is displayed.
3. Create a new wafer recipe by typing the desired name in the Name
field and a description (if desired) in the Description field. Click on
Enter when you have completed the entry.
The pertinent wafer recipe information (if any) is displayed in the
Wafer Recipe area of the Recipe Creator.
Wafer Information 1. In the Recipe Creator main menu, click on Wafer Recipe.
The Wafer Recipe menu is displayed.
3. Select the item that matches the information for the wafer to be
measured by clicking once on the item then clicking Enter, or by
double clicking on the item. You may also click on DEFAULT to select
the default wafer parameters based on the previously selected wafer
recipe.
You may also modify an existing Wafer Selection entry by clicking
once on the entry name, then clicking on Modify. A Wafer Specifica-
tion window is displayed as shown in Figure 4-11. Specify the wafer
size, flat/notch information, and orientation as desired and click on
Enter.
CAUTION
NOTE
Film Stack 1. In the Recipe Creator main menu, click on Wafer Recipe.
Specification
The Wafer Recipe menu is displayed.
3. Select the desired filmstack from the list by clicking once on the item
then clicking Enter, or by double clicking on the item. You may also
click on DEFAULT to select the default filmstack specification.
You may also modify an existing filmstack entry by clicking once on
the entry name, then clicking on Modify. A Filmstack Model window
is displayed in which you may make changes to the layer and mea-
sured parameters as desired. For information on how to use this win-
dow, refer to Chapter 3, “Developing Film Applications”.
CAUTION
Wafer Registration 1. In the Recipe Creator main menu, click on Wafer Recipe.
The Wafer Recipe menu is displayed.
Creating a After completing the steps in “Creating a Wafer Recipe” on page 4–16,
and selecting None from the Registration sub-menu under the Wafer
Monitor Wafer Recipe menu, select a measurement pattern and report style using the
Recipe information in the sections that follow.
NOTE
CAUTION
NOTE
Wafer Report 1. In the Recipe Creator main menu, click on Wafer Recipe.
Selection
The Wafer Recipe menu is displayed.
CAUTION
When you have completed modifying the wafer report, the report
name is displayed in the Wafer Recipe area of the Recipe Creator.
NOTE
• Create a new report: type the desired name in the Name field, and
a description (if desired) in the Description field. Click on Enter
and the Wafer Report window is displayed in which you may specify
the data source, where the data will be saved and/or displayed,
how the data is saved and/or displayed, and what data will be
included in the report.
When you have completed creating the new report, the report name
is displayed in the Wafer Recipe area of the Recipe Creator.
Saving the Recipe Once all of the recipe selections have been made, save the current rec-
ipe in the database by performing the following:
2. Click on Save in the Wafer Recipe menu. You may also click on
Cancel in the Wafer Recipe menu at any time while creating the
recipe to cancel the current wafer recipe and start over.
The monitor wafer recipe has now been created. Before the wafer
recipe can be run by the Operator, the recipe (together with transfer
and control information) must be inserted into a process step under a
process specification.
Creating a After completing the steps in “Creating a Wafer Recipe” on page 4–16,
and selecting Automatic from the Registration sub-menu under the
Patterned Wafer Recipe menu, you must now train the FE System to automatically
Wafer Recipe recognize (register) the patterns on the wafer using one of three methods
(Cognex (direct drive, site-by-site, and deferred), and select the measured chips.
System)
4. Select the cassette and slot from which to load the wafer and click on
Enter.
The selected wafer is loaded onto the stage and the Site Locator win-
dow now shows a live video image of the wafer. An Automatic Regis-
tration window is displayed instructing you to input the die pitch and
click on the Train button to determine chip spacing parameters.
NOTE
6. If the Die Pitch (the size of the stepper groups) is known, enter the
value in millimeters in the Die Pitch boxes. If the Die Pitch is not
known, leave the default values and proceed with the next step.
7. Click on the Train button in the Wafer Layout window to begin the
stepper group training.
A Die Pitch Training window is displayed instructing you to move to
the lower left-hand corner of a die in the Site Locator window.
9. In LoMag, drive the laser spot to the lower left-hand corner of the first
die (corner 1), switch to HiMag and refine the positioning to the
precise lower left corner of the first die. Positioning the wafer under
the laser spot (blue diamond) may be done by moving the trackball
pointer to the desired location in the live video image of the Site
Locator window, and double clicking the right trackball button.
NOTE
10. Once the laser spot has been positioned over the lower left corner of
the first die, click on the Corner 1 button.
A window is displayed instructing you to choose the lower left-hand
corner of the next die DOWN.
11. Switch to LoMag, then drive the wafer so that the laser spot (blue
diamond) is positioned over the lower left-hand corner of the die
directly below the first die (corner 2), switch to HiMag and refine the
positioning. If the correct die pitch was chosen, the stage will
automatically drive to the approximate position of the second corner,
then you may refine the position.
NOTE
12. Once the laser spot has been positioned over the lower left corner of
the second die, click on the Corner 2 button.
A window is displayed instructing you to choose the lower left-hand
corner of the next die to the LEFT.
13. Switch to LoMag, then drive the wafer so that the laser spot (blue
diamond) is positioned over the lower left-hand corner of the die
directly to the left of the second die (corner 3), switch to HiMag and
refine the positioning. If the correct die pitch was chosen, the stage
will automatically drive to the approximate position of the third corner,
then you may refine the position.
NOTE
14. Once the laser spot has been positioned over the lower left corner of
the third die, click on the Corner 3 button.
The Wafer Layout window is redisplayed with the rough X and Y val-
ues for the Die Pitch.
NOTE
16. Position the trackball pointer over the corner of the image you want to
train, click down and hold the left trackball button and simultaneously
move the trackball to drag a rectangular box over the desired area.
When the box is the correct size, release the left trackball button.
NOTE
NOTE
• If the score for the teach site is acceptable: Click on the Enter
button in the Vision System Setup window.
Once you have accepted the score for the teach site and clicked on
Enter, the system will refine the die pitch using the pattern taught in
the previous steps. At each site, depending on the version of the soft-
ware, a window may be displayed asking if the site shown is a good
refine site.
19. If prompted to confirm the refine site, perform one of the following:
• If the refine site displayed closely matches the teach site: click
on Yes.
The system moves to the next refine site and you are once again
prompted to confirm if this is a valid refine site.
• If the refine site displayed does not closely match the teach
site: click on No.
The system moves to a new refine site and you are once again
prompted to confirm if this is a good refine site.
Once the system has found the refine sites on the wafer, a window is
displayed showing the refined die pitch information.
20. Click on the OK button in the Automatic Die Pitch Refine window.
The system will now deskew the wafer to ensure that it is centered on
the stage and the pattern is aligned correctly. The number of sites
checked will depend on the version of the software. At each location
a window is displayed asking you to verify that this is a valid registra-
tion site.
21. Each time you are prompted to verify the registration site, perform
one of the following:
You may now train the measurement patterns for the wafer.
Pattern Training The steps provided in this section will allow you to identify the measure-
ment sites within a die on the wafer. The trained measurement pattern
may then be duplicated onto other selected dies. Modifications may be
made to trained measurement patterns using the procedure in “Changing
Measurement Sites in a Registered Pattern” on page 4–58.
4. In the Site Locator Wafer View window located in the upper left-hand
portion of the screen, move to the stepper group that will be used as
a model to train the measurement sites by positioning the trackball
pointer over the desired stepper group and double clicking the right
trackball button.
The stage moves the wafer to the selected stepper group.
NOTE
8. Click the Enter button in the lower right corner of the Site Locator
window.
The Point Selection window is displayed as shown in Figure 4-24. A
default filmstack is displayed in the Filmstack field.
• Allow the Operator to fine tune the position of the measurement site
at run time by clicking on the button beside Defer Location. This is
particularly useful if you have a very small measurement site and
should only be used if the FE System has no pattern recognition
hardware, or when the pattern recognition hardware will not work
and the direct drive alignment is not acceptable.
NOTE
When the teach site has been selected, you are returned to the
Recipe Creator Main window with the Site Pattern window
displayed. A square box is displayed on the selected
measurement location to indicate that pattern recognition will
be used for this site.
12. When all measurement sites have been selected, click on the Enter
button in the Site Pattern window.
A window is displayed indicating that training is completed and
instructing you to select the Select Measured Chips item from the
Patterned Wafer sub-menu of the Wafer Recipe menu.
You may now select the chips to be measured on the wafer. Once the
pattern has been trained, it may be modified using the procedure
provided in “Changing Measurement Sites in a Registered Pattern” on
page 4–58.
Select Measured The steps provided in this section will allow you to identify the stepper
Chips groups on which to perform the previously defined measurement. The
stepper groups will be measured in the order in which they are selected
in this procedure.
NOTE
4. Click on the Group # button in the upper left Site Locator Wafer View
window.
The selected stepper group is now highlighted in the Site Locator
window and the number in the Group # button increments by one.
The Total Selected Groups field of the Stepper Group Map window
displays the number of stepper groups that have currently been
selected.
NOTE
5. Repeat Step 3 and Step 4 above for each desired stepper group.
NOTE
6. Once all of the desired stepper groups have been selected, click on
Enter in the Stepper Group Map window.
You are returned to the Recipe Creator Main window. You may now
select the wafer report to complete the recipe.
Wafer Report 1. In the Recipe Creator main menu, click on Wafer Recipe.
Selection
The Wafer Recipe menu is displayed.
When you have completed modifying the wafer report, the report
name is displayed in the Wafer Recipe area of the Recipe Creator.
CAUTION
• Create a new report: type the desired name in the Name field, and
a description (if desired) in the Description field. Click on Enter
and the Wafer Report window is displayed in which you may specify
the data source, where the data will be saved and/or displayed,
how the data is saved and/or displayed, and what data will be
included in the report.
When you have completed creating the new report, the report name
is displayed in the Wafer Recipe area of the Recipe Creator.
Saving the Recipe Once all of the recipe selections have been made, save the current rec-
ipe in the database by performing the following:
2. Click on Save in the Wafer Recipe menu. You may also click on
Cancel in the Wafer Recipe menu to cancel the current wafer recipe
being created and start over.
CAUTION
Modifying a To modify a recipe, the Process Engineer first selects the desired recipe.
Once selected modifications may either be made to that recipe or to a
Wafer Recipe copy of the selected recipe. The changes are then saved and the recipe
may then be used in a process.
CAUTION
Use the following procedure to select and modify existing wafer recipes.
For information on creating new wafer recipes, refer to “Creating a Wafer
Recipe” on page 4–16.
Recipe Selection 1. In the Recipe Creator main menu, click on Wafer Recipe.
The Wafer Recipe menu is displayed.
4. Once the wafer recipe is selected and displayed in the Wafer Recipe
area, perform any of the following:
Creating a This section provides information on how to create a wafer recipe that will
generate a wafer map file. The map file may be viewed either using the
Wafer Mapping Focus Mapping software (see Chapter 5, “Wafer Mapping and Data
Recipe Reporting”) or the Focus Operator Queued Loading software (see
Chapter 2, “Operator Interface”).
Any wafer recipe can be a wafer mapping recipe providing at least five
points are measured. Perform the following to configure a wafer recipe to
generate a wafer map file:
CAUTION
• Create a new report by entering the desired name for the report in
the Name field and a description for the report (if desired) in the
Description field and click on Enter.
The Wafer Report window is displayed as shown in Figure 4-37 on
page 4–61.
CAUTION
5. In the Recipe Creator main menu, click on Wafer Recipe then click
on Save when the Wafer Recipe menu is displayed.
The wafer recipe has now been saved.
Creating and The FE System allows the Process Engineer to create new wafer mea-
surement patterns, or modify measurement patterns that already exist in
Modifying the database, in the event that a default pattern does not meet the needs
Measurement of the current application.
Patterns Once you have entered a new measurement pattern name, or selected a
pattern for modification, as described in Step 3 of “Measurement Pattern
Selection” on page 4–22, the Site Pattern window is displayed as shown
in Figure 4-29.
NOTE
The following buttons are available from the Site Pattern window:
Pattern To create a wafer measurement pattern point by point, perform the follow-
ing:
Creation by
Point 1. In the Site Locator window, change to Wafer View by selecting the
Control menu. Click on Mag from the Control menu items and select
Wafer View from the sub-menu that is displayed
The Site Locator window is now in Wafer View mode.
7. When all measurement points have been selected, click on the Enter
button in the Site Pattern window.
You are returned to the Recipe Creator main window.
NOTE
5. When all measurement points have been entered, click on the Enter
button in the Site Pattern window.
You are returned to the Recipe Creator main window.
The following buttons are available from the Shape Pattern window:
Line Patterns To create a linear wafer measurement pattern, perform the following:
2. Position the trackball pointer over the desired starting point for the
line, press and hold the left trackball button and move the trackball
pointer to the ending point for the line and release the left trackball
button.
The Linear Selection window is displayed as shown in Figure 4-33.
This window provides information about the number of measurement
points along the line.
5. When all linear patterns have been completed, click on the Enter
button in the Shape Pattern window.
You are returned to the Site Pattern window.
6. Click on the Enter button in the Site Pattern window to save the
measurement pattern information.
You are returned to the Recipe Creator main window.
Circular Patterns To create a circular wafer measurement pattern, perform the following:
1. Click on the Circle button in the Draw section of the Shape Pattern
window.
A window is displayed instructing you to drag the center and radius of
the circle in the Site Locator window. Click on OK .
2. Position the trackball pointer over the desired center point of the
circle, press and hold the left trackball button and move the trackball
pointer the desired distance for the radius of the circle and release
the left trackball button.
The outline of a circle is displayed in the Site Locator window as you
drag the trackball. When the left trackball button is released, the Lin-
ear Selection window is displayed as shown in Figure 4-34. This win-
dow provides information about the number of measurement points
along the perimeter of the circle.
5. When all circular patterns have been completed, click on the Enter
button in the Shape Pattern window.
You are returned to the Site Pattern window.
6. Click on the Enter button in the Site Pattern window to save the
measurement pattern information.
You are returned to the Recipe Creator main window.
Rectangular Area To create a rectangular area wafer measurement pattern, perform the fol-
Patterns lowing:
3. Click on the Enter button in the Shape Pattern window when all
shape patterns have been completed.
You are returned to the Site Pattern window.
You may also click on the Cancel button to reject this information.
You are returned to the Site Pattern window where you may make
another selection.
4. Click on the Enter button in the Site Pattern window to save the
measurement pattern information.
You are returned to the Recipe Creator main window.
Circular Area To create a circular area wafer measurement pattern, perform the follow-
Patterns ing:
1. Click on the Circle button in the Area section of the Shape Pattern
window.
The Circular Area window is displayed as shown in Figure 4-36.
2. Enter the desired Grid Style and/or circle radius for the pattern. Click
on the desired number of measurement points for the pattern and
then click on the Enter button.
The area pattern is displayed on the wafer view and you are returned
to the Shape Pattern window.
You may also click on the Cancel button to reject this information.
You are returned to the Shape Pattern window where you may make
another selection.
3. Click on the Enter button in the Shape Pattern window when all
shape patterns have been completed.
You are returned to the Site Pattern window.
You may also click on the Cancel button to reject this information.
You are returned to the Site Pattern window where you may make
another selection.
4. Click on the Enter button in the Site Pattern window to save the
measurement pattern information.
You are returned to the Recipe Creator main window.
Changing A Registered Pattern is one in which the stepper groups have already
been trained.
Measurement
Sites in a To change a Registered Pattern, but keep the selected chips to be mea-
sured the same, use the following procedure:
Registered
1. In the Recipe Creator main menu, click on Wafer Recipe.
Pattern
The Wafer Recipe menu is displayed.
8. In the Site Locator Wafer View window located in the upper left-hand
portion of the screen, move to the stepper group that you wish to
modify by positioning the trackball pointer over the desired stepper
group and double clicking the right trackball button.
The wafer is moved to position the selected stepper group under the
laser spot.
10. Click on the Die 1 button in the Stepper Group Pattern Training
window.
The measurement sites that were previously chosen are highlighted
and the Site Pattern Window is displayed as shown in Figure 4-23.
• Click on Last under Erase in the Site Pattern window to clear any
unwanted measurement sites.
12. When you are finished erasing and adding sites, click on Enter in the
Site Pattern window.
You are returned to the Recipe Creator Main window.
14. Click on Patterned Wafer in the Wafer Recipe menu, then click on
Select Measured Chips when the Registration sub-menu is
displayed.
The Site Locator Live Video window is displayed in half screen,
LoMag mode. A Stepper Group Map window is displayed as shown in
Figure 4-25 and the Site Locator Wafer View window appears as
shown in Figure 4-26.
15. Click on Update in the Stepper Group Map window so that the new
measurement sites will appear in the selected chips to be measured.
A window is displayed informing you when the selected stepper
groups have been updated with the new measurement pattern. Click
on OK.
17. In the Stepper Group Pattern Training window menu, click on View,
then click on Wafer Pattern from the View menu.
The Wafer View will show all of the updated measurement sites.
18. In the Stepper Group Pattern Training window menu, click on Close.
The Stepper Group Pattern Training window is closed and you are
returned to the Recipe Creator Main window.
20. Click on Save in the Wafer Recipe menu. You may also click on
Cancel in the Wafer Recipe menu to cancel the current wafer recipe
being created and start over.
CAUTION
Wafer Once you have entered a new measurement report name, or selected a
report for modification, as described in Step 3 of “Wafer Report Selection”
Reports on page 4–24 for monitor wafers, or Step 3 of “Wafer Report Selection”
on page 4–41 for patterned wafers, the Wafer Report window is displayed
as shown in Figure 4-37.
This window is used to specify how the measurement data will be pre-
sented and where it will be stored.
The Wafer Report window is made up of the following sections that allow
you to customize the report format and destination.
NOTE
NOTE
NOTE
Transfer Wafer transfer control allows the Process Engineer to specify the source
and destination cassette(s) and or slot(s) for the wafer to be measured,
Control as well as how the wafers will be transferred and the destination for
rejected wafers. These options may also be deferred at this time allowing
the Operator to make these decisions when the process is run using the
Focus Operator Queues Loading program (see Chapter 2, “Operator
Interface” for information).
Once you have created and saved a wafer recipe, specify transfer control
by performing the following:
• xxX1-1S1,12,25
Indicates this transfer will take wafers from cassette 1 and return
them to cassette 1 after measurement. Wafers in slots 1, 12, and
25 will be measured.
• xxX1-2S1-25
Indicates this transfer will take wafers from cassette 1 and return
them to cassette 2 after measurement. Wafers in slots 1 through 25
will be measured.
• xxX1-1S1-5Reject2
Indicates this transfer will take wafers from cassette 1 and return
them to cassette 1 after measurement. Wafers in slots 1 through 5
will be measured. Wafers that are rejected will be placed in
cassette 2.
• xxX1-3S1-25_B2T
Indicates this transfer will take wafers from cassette 1 and return
them to cassette 3 after measurement. Wafers in slots 1 through 25
will be measured. As the measurements are completed, the wafers
are transferred between the two cassettes in a “bottom to top”
manner. That is the wafer taken from cassette 1 slot 1 will be
returned to cassette 3 slot 25 and so on.
CAUTION
• Create a new transfer control item: type the desired name in the
Name field, and a description (if desired) in the Description field
then click on Enter.
The Transfer Specification window is displayed as shown in
Figure 4-39.
• To accept the default values: select the desired wafer slots then
click on Enter in the Transfer Specification window.
CAUTION
CAUTION
Creating and The process specification consists of those steps (the wafer recipe(s),
transfer, and control information) that have been previously defined by
Modifying a the Process Engineer in this chapter. The process specification is then
Process run by the Operator and the actual wafer measurements are performed
(see Chapter 2, “Operator Interface” for information).
Specification
Once you have created and saved a wafer recipe and specified the trans-
fer control information, create a process specification by performing the
following:
• To create a new process step: continue with the next step in this
procedure.
8. Select a wafer recipe from the list or create or modify a wafer recipe
using the procedure in “Creating a Wafer Recipe” on page 4–16.
Click on Enter when you have selected the desired name.
The selected name appears on the Recipe field of the Step Compo-
nents in Figure 4-40.
• Select a recipe control item: click on the item name then click on
Enter, or double click on the item name. Selecting DEFAULT will
defer no options to the Operator.
• Modify a recipe control item: click on the item name and then
click on Modify.
Once all desired entries are made in the Recipe Control window,
click on the Enter button then click on Enter again when you are
returned to the Recipe Control Selection window.
The selected name appears on the Control field of the Step Compo-
nents.
When all process steps have been added to the process specification,
the recipe creation process has been completed. The Operator may now
run the process specification using the Focus Operator Queued Loading
program to perform wafer measurements. Refer to Chapter 2, “Operator
Interface” for information.
Modifying a To modify a process step that has already been created and is part of a
process specification, perform the following:
Process Step
NOTE
NOTE
5. When all changes to the process step have been completed, click on
the Add button in the Process Specification window.
A window is displayed to inform you that the selected process step
has been successfully modified. Click on OK .
Exiting To exit the Focus Recipe Creator program, perform the following proce-
dure:
Recipe
1. On the Recipe Creator screen, perform one of the following:
Creator
• Double click the left trackball button on the icon located in the upper
left corner of the screen (beside the Focus Recipe Creator program
name).
• Single click the left trackball button on the icon located in the upper
left corner of the screen (beside the Focus Recipe Creator program
name). Select Close from the menu that is displayed.
The Focus Recipe Creator program will exit and you will be returned
to the OS/2 desktop.
Data Reporting
- Contour maps
- Topographical maps
- Difference maps
- Etch Rate maps
• How to use the Focus Browser software to access tabular data
reports.
The tasks described in this chapter are performed with the use of the
Focus Mapping and Focus Browser software (with the exception of
importing and exporting data, which is performed using the Focus Recipe
Creator software). Wafer mapping and report data is also available
through the Focus Operator Queued Loading program. For information
on how to access this information using the Focus Operator Queued
Loading program, refer to Chapter 2, “Operator Interface”.
It should be noted that certain steps must be performed using the Focus
Recipe Creator and/or Focus Operator Queued Loading programs in
order to create the wafer mapping and report data. Refer to the appropri-
ate chapter in this guide for more information.
Wafer The Focus Mapping program makes it possible to display wafer measure-
ment data graphically. Any or all of the three optical parameters (T, N,
Mapping and K) can be presented on the monitor in either topographical (3D) or
contour (2D) form. In addition, etch maps (showing etch rate or difference
information) may also be generated.
NOTE
1. The Map box must be checked in the report section of the wafer
recipe. Refer to “Creating a Wafer Mapping Recipe” on page 4–45 of
Chapter 4, “Creating Recipes” for information.
Once the wafer map file is generated, the Focus Mapping program may
be used to create a wafer map from the wafer map file and to view, scale,
manipulate, and print the map.
Starting To start the Focus Mapping program, position the trackball pointer over
the Focus Mapping icon, located in the Focus Ellipsometer folder, and
Focus double click the left trackball button.
Mapping The Focus Mapping main window is displayed as shown in Figure 5-1.
Focus Mapping The following menu items are available in the Focus Mapping Main win-
dow:
Main Menu
• Maps — Used to select the wafer map to be viewed. Clicking on
the Maps menu item causes the Map Files Selection window to be
displayed.
• Clear All — Used to close all open maps and clear the screen.
Map Window This window displays the generated wafer map file. Wafer maps may be
viewed in Contour (2D) or Topographic (3D) mode, and each may be dis-
played in either color, greyscale, or black and white.
Map Window Menu The Wafer Map window also consists of the following menu selections:
• View — Used to alter the viewpoint from which the map is seen.
Displays a View Control Window that allows you to modify the
angle of elevation and rotation of the view. These modifications
only apply to Topographic wafer maps.
• Color — Used to display the map in Color, Grey, or Black & White
mode. Also allows you to determine whether the color divisions will
be set automatically or in a user specified range.
- Draw Grid — Displays a black wire grid over the wafer map.
- Draw Marker — Displays markers on each measurement point.
• Edit — Allows you to edit the data that makes up the wafer map.
• Text — Displays the wafer map data in text format.
• Update — When selected, the wafer map is automatically updated
each time a display mode is selected.
Viewing Wafer Once a wafer mapping recipe has been created, incorporated into a pro-
cess step, then run by the Operator using the procedures provided in
Maps Chapter 4, “Creating Recipes” and Chapter 2, “Operator Interface”
respectively, the wafer map may be viewed.
2. Select the wafer map you wish to view by clicking once on the
desired map file name in the list of available items and then clicking
on Enter, or double click on desired map file name.
The selected map is displayed on the screen. The type of map that is
displayed (contour or topographic) color mode (black and white, grey,
or color) is determined by your FE System configuration.
NOTE
Figure 5-5 shows an example of a black and white contour wafer map.
Note that the wafer statistics are also displayed with the map.
NOTE
Difference and The FE System’s mapping software has the capability to create the fol-
lowing types of etch maps:
Etch Rate Maps
• Difference Map — Compares a wafer before and after undergoing
an etch or CMP (Chemical Mechanical Polishing) process that
removes material from the wafer. Difference maps show how much
material was removed between measurements.
Creating a To create a difference map, you should first have created a map of the
Difference Map wafer both before and after the etch process. Once both maps have
been created, enter the mapping software using the procedure in
“Starting Focus Mapping” on page 5–3.
4. Click on OK.
A window is displayed instructing you to select the first wafer map.
5. Click on OK.
The Map File Selection window is displayed as shown in Figure 5-3
on page 5–6.
6. Click once on the desired map file name in the Map File Selection
window then click on Enter, or double click on the desired map
file name.
A window is displayed instructing you to select the wafer map that will
be subtracted from the first map.
7. Click on OK.
The Map File Selection window is displayed.
8. Click once on the desired map file name in the Map File Selection
window and then click on Enter, or double click on desired map file
name.
A window is displayed asking if you wish to see the map.
- Click on No and the difference map data is not saved. The map
remains on the screen.
CAUTION
Creating an Etch To create an etch rate map, you should first have created a map of the
Rate Map wafer both before and after the etch process. Once both maps have
been created, enter the mapping software using the procedure in
“Starting Focus Mapping” on page 5–3.
5. Click on OK.
A window is displayed instructing you to select the map before etch.
6. Click on OK.
The Map File Selection window is displayed as shown in Figure 5-3
on page 5–6.
7. Click once on the desired map file name in the Map File Selection
window and then click on Enter, or double click on desired map file
name.
A window is displayed instructing you to select the map after etch.
8. Click on OK.
The Map File Selection window is displayed.
9. Click once on the desired map file name in the Map File Selection
window and then click on Enter, or double click on desired map file
name.
A window is displayed asking if you wish to see the map.
• Click on Yes and the wafer etch rate map is displayed on the
screen along with a window asking if you wish to save the map.
- Click on No and the etch rate map data is not saved. The map
remains on the screen.
CAUTION
If you do not save the etch rate map, when you clear
the map or exit the Mapping software, the data is lost.
Updating Wafer There are two ways in which to update the wafer map that is currently
displayed on the screen. These methods are Automatic and Manual
Map Views updating.
Automatic In Automatic update mode, the Update button on the menu bar is high-
Updating lighted. Any change made to the display, such as changing the color
mode, map orientation, or switching between contour and topographical
displays will cause the wafer map to be updated immediately. This
method can be time consuming if several changes are to be made to the
wafer map view. Automatic update is the system default.
Manual Updating In Manual update mode, you may make several changes to the wafer
map view and the changes will not take effect until the Operator manually
updates the view. To enter Manual update mode, perform the following:
1. Click on the Update button in the menu bar so that the button is no
longer highlighted.
2. Make other menu selections (such as changing the color mode, map
orientation, switching between contour and topographical displays, or
modifying other display modes) as desired using the procedures
given in this chapter. Note that the wafer map view is not updated.
3. Click on the Update button in the menu bar. The button is now
highlighted and the wafer map view is updated with the new display
selections.
Switching To change the wafer map view to/from black and white, color, or grey-
scale, perform the following:
Color Modes
1. Click on Color in the menu bar.
The Color menu is displayed (with a checkmark beside the current
color mode).
Changing To change the orientation of the wafer map on the screen, perform the
following:
Wafer
Orientation 1. Click on View in the menu bar.
View The View Control window is displayed as shown in the figure below.
2. Click in the Rotation: and/or Elevation: fields and enter the desired
values, or use the slider bars to display the desired rotation and
elevation of the view.
NOTE
Switching To change the wafer map view between topographical and contour map-
ping, perform the following:
Between
Topographic 1. Click on Mode in the menu bar.
and Contour The Mode menu is displayed (with a checkmark beside the current
Mapping display mode).
Clearing the When you have completed viewing the wafer maps, you may clear the
screen by clicking on Clear All in the menu bar.
Wafer Map View
All currently displayed maps are closed. You may now exit the Mapping
software (see “Exiting Wafer Mapping” on page 5–19), or select another
wafer map to view (see “Opening a New or Additional Wafer Maps” on
page 5–16).
Opening a New You may open a new wafer map to view after clearing the screen, or open
an additional wafer map while the current map is still displayed by per-
or Additional forming the following:
Wafer Maps
1. Click on Maps in the menu bar.
The Map File Selection window is displayed as shown in
Figure 5-3 on page 5–6.
Editing the Erroneous points, which may be caused by processing, debris, or edge
effects on the wafer, may distort the appearance of a wafer map.
Measured Data
You may edit the measurement data for the displayed wafer map by per-
forming the following:
1. With a wafer map displayed, click on Edit in the menu bar then click
on Data from the Edit menu.
The Editing Measured Data window is displayed as shown in the fig-
ure below.
NOTE
CAUTION
Exiting Wafer To exit the wafer mapping software, double click the icon in the upper-left
corner of the screen, or single click the icon to display a menu. With the
Mapping menu displayed, click on the Close menu item.
Any wafer maps currently displayed will be closed and the Mapping soft-
ware will exit. You are returned to the OS/2 desktop.
Wafer The Focus Browser program makes it possible to view the wafer mea-
surement data collected by the FE System. In order to view the data you
Reports must select the wafers that you wish to view and, within the selected set
of wafers, the exact point(s), film layer(s), and film parameter(s) must be
selected.
The process of “Browsing” the data consists of defining a query and dis-
playing data. A query is the means used to collect the data. An example
query might be: “Select all of the 16 MEG-DRAM Gate oxide wafers and
display only the thickness of the first point on each wafer.” The data can
be displayed as a graph, an SPC trend chart or text.
1. A wafer recipe must be created by the Engineer that sends the wafer
measurement data to the database using the Focus Recipe Creator
program. The wafer recipe is then incorporated into a process step.
Refer to “Wafer Report Selection” on page 4–24 of Chapter 4,
“Creating Recipes” for information.
2. The process specification and step containing the wafer recipe are
run by the Operator using the Focus Operator Queued Loading
program. This will generate the database report file. Refer to
Chapter 2, “Operator Interface” for information.
Once the wafer database file is generated, the Focus Browser program
may be used to query the database and display the measurement
report data.
Starting To start the Focus Browser program, position the trackball pointer over
the Focus Browser icon, located in the Focus Ellipsometer folder, and
Focus double click the left trackball button.
Browser The Focus Browser main window is displayed as shown in Figure 5-9.
Focus Browser The following menu items are available in the Focus Browser Main win-
dow:
Main Menu
• Define Query — Used to select a predefined database query or to
define a new database query for retrieving the desired data from
the FE System database.
Cassette Icon The Cassette Icon, displayed in the Focus Browser main window, pro-
vides an easy method for reviewing data by cassette groups. Cassette
groups are identified by cassette ID and measurement data. The most
recently measured cassettes appear at the top of the list.
Browsing the The following methods may be used to browse the database:
NOTE
Cassette The cassette mode makes it possible to view the measurement data in its
“as measured” format. The data cannot be reorganized.
Browsing
To start the cassette browsing mode, perform the following:
1. Click the left trackball button on the Cassette Icon located on the
Focus Browser main window.
A window is displayed with a list of the available cassette IDs as
shown in Figure 5-10. The date and time during which the cassettes
were measured is also given. The cassette IDs are listed in reverse
time order, most recent cassette measured at the top of the list.
2. Select a cassette by single clicking on the desired item from the list
and clicking on the OK button or by double clicking on the desired
item. Scroll the list up or down using the Page Up and Page Down
buttons if necessary. The display may be changed to show the
cassette ID, the recipe that was run on the cassette, or the Operator
who ran the process step by clicking on the Cassette ID, Recipe, or
Person button respectively.
A measurement report is displayed as shown in Figure 5-11. The cas-
sette entry will be shown in a format similar to those produced by
Focus Operator Queued Loading program. One difference between
the formats is that multiple parameters create duplicate lines
(for example, two lines for point 1).
3. When when the report is displayed, you may perform any of the
following:
• Save the data in the report to a text file by clicking on the File
menu, then clicking on Save As. Enter the desired destination path
and filename in the dialog box that is displayed and click on Enter.
The data is saved in the specified file and may be imported into a
text editor or into Microsoft Excel.
• Quit the report viewer by clicking on the Exit menu item. This exits
the report viewer, however the data contained in the report still
resides in the FE System database.
Database A query is a set of rules used to retrieve data from the database. You
must indicate which wafers you wish to view and then decide which infor-
Queries mation from each wafer you wish to see.
You are returned to the Focus Browser main window with the
selected query name displayed as shown in Figure 5-14. Proceed
to “Viewing Report Data” on page 5–30 to perform the desired
query (graph, text, or text dump).
NOTE
• Create a new query: type the desired name in the Name field, type
a description for the new query in the Description field, then click
on the Enter button.
- Process Specification
- Process Step
- Wafer Recipe
- FilmStack
- Layer
- Film Parameter
- Wafer Points
- Wafer Slot
This information is required to avoid mixing different types of informa-
tion. For example, nitride film thicknesses cannot be compared to
oxide film thicknesses. However, there are cases where it may be
desirable to mix information. For example, the measurement points
listed on the wafer can be taken from any wafer. In this case, it is per-
missible to mix the choices.
The items in the above list comprise all of the information required to
extract one type of information from the database. The information
itself can be separated into three categories:
- Recipe information
- Film information
- Position information
3. Perform one of the following:
The system will automatically step you through the query definition
process. Each time you are prompted, select the desired process
spec, process step, wafer recipe, filmstack, layer, points, and
wafer slot.
NOTE
Viewing Data can be displayed as a graph, text report, or text dump (suitable for a
spread-sheet program). In addition, the query statistics (query name, pro-
Report Data cess specification and step, recipe name, and so on) can be displayed.
The current query, displayed in either in the Focus Browser main window
(Figure 5-14) or in the Query Definition Screen, will be used to select the
data for display.
Displaying Once a query has been defined and selected, display the data retrieved
by the query as a graph by performing the following:
Data as a Graph
1. Perform one of the following:
2. When you have finished viewing the graph, exit the viewer by double
clicking the icon in the upper left corner of the viewer window (to the
left of the Measurement Value Graph name), or single click the icon
to display a menu. With the menu displayed, click on the Close item.
You are returned to the screen from which you opted to display the
graph (the Focus Browser window or the Query Definition Screen).
Displaying The data retrieved by the selected query may be displayed in either text
mode or data dump format.
Data in Text
Format
Text Mode The text display presents the data in the format similar to that generated
by the Focus Operator Queued Loading program, except that the points
do not necessarily represent consecutive data.
Once a query has been defined and selected, display the data retrieved
by the query as a text report by performing the following:
2. When you have finished viewing the report on-line, you may:
• Save the data in the report to a text file by clicking on the File
menu, then clicking on Save As. Enter the desired destination path
and filename in the dialog box that is displayed and click on Enter.
The data is saved in the specified file and may be imported into a
text editor or into Microsoft Excel.
• Quit the report viewer by clicking on the Exit menu item. This exits
the report viewer, however the data contained in the report still
resides in the FE System database.
You are returned to the screen from which you opted to display the
report (either the Focus Browser main window or the Query Definition
Screen).
Text Dump Mode The text dump mode displays the data in columns separated by tabs to
facilitate its use in a spreadsheet program.
Once a query has been defined and selected, display the data retrieved
by the query as a data dump report by performing the following:
1. From the Focus Browser window, click on Display Report from the
main menu, then click on Data Dump.
A database query is performed and a text report is displayed (in data
dump mode) similar to the example shown in Figure 5-17.
2. When you have finished viewing the report on-line, you may:
• Save the data in the report to a text file by clicking on the File
menu, then clicking on Save As. Enter the desired destination path
and filename in the dialog box that is displayed and click on Enter.
The data is saved in the specified file and may be imported into a
text editor or into Microsoft Excel.
• Quit the report viewer by clicking on the Exit menu item. This exits
the report viewer, however the data contained in the report still
resides in the FE System database.
You are returned to the Focus Browser main window.
Query Statistics To display the parameters that make up the current query, perform the
following:
2. When you have finished viewing the query information, click on the
OK button.
You are returned to the Focus Browser main window.
Exiting To exit the Focus Browser software, double click the icon in the upper-left
corner of the screen, or single click the icon to display a menu. With the
Focus menu displayed, click on the Close menu item.
Browser Any reports currently displayed will be closed and the Focus Browser will
exit. You are returned to the OS/2 desktop.
NOTE
CAUTION
1. Start the Focus Recipe Creator program and log in using the
procedures provided in “Starting Recipe Creator and Logging In” on
page 4–5 of Chapter 4, “Creating Recipes”.
4. With the media disk inserted in drive A:\, click on the OK button.
The Recipe Import/Export window is displayed as shown in
Figure 5-19.
• Recipe Type
- Process Spec — Displays all process spec information that is
in the Focus database and on the media disk in the Focus
Database and Media Database lists.
• Initialize Media Button — Clicking this button will cause the media
disk to be erased. You will be prompted to confirm this action before
the disk is actually erased.
CAUTION
The selected item(s) are copied from the media disk to the FE data-
base. If Enable overwrite was not selected, you are prompted
before items that already exist in the database are overwritten.
Click on Yes to overwrite the file, No to keep the existing file.
The selected item(s) are copied from the FE database to the media
disk. If Enable overwrite was not selected, you are prompted
before items that already exist on the media disk are overwritten.
Click on Yes to overwrite the file, No to keep the existing file.
9. Repeat Step 5 through Step 8 above for each of the Recipe Types
you wish to import or export.
10. When all of the desired information has been imported or exported,
click on the Exit button.
The message Deregistering Export Media is displayed.
CAUTION
Creating an A media disk is a Focus database formatted disk which has all of the
default database filmstacks, materials, recipes, and so on.
Import/Export
Media Disk After inserting a blank, formatted floppy in disk drive A:\ as described in
Step 3 of the procedure provided in “Importing and Exporting Data” on
page 5–37, use the following procedure to create an import/export
media disk.
NOTE
3. With a blank, formatted floppy in drive A:\, click on the Yes button.
A window is displayed indicating that a Media Disk is being created.
Once the process has completed, the Recipe Import/Export window
is redisplayed as shown in Figure 5-19. Notice that the Make Media
Disk button is now labeled Initialize Media.
NOTE
4. With the media disk now created, return to Step 5 of the procedure
given in “Importing and Exporting Data” on page 5–37.
Introduction The purpose of this appendix is to provide the operator and the engineer
with detailed instructions for starting and stopping the system. The infor-
mation in this section pertains to the FOCUS Ellipsometer (FE) III, IV, and
VII systems, and includes the following topics:
NOTE
FE System All FE System models have a standard control panel (as shown in the fig-
ure below) which contains the following:
Control Panel
1. Power On/Off switch — Green switch used to turn the system on
and off under normal operating conditions.
1 2
Normal To startup your FE System, ensure that power is applied to the system
(that is, the power cord is plugged into the wall outlet, the main power
System breaker for that outlet is turned on, and the circuit breaker behind the
Startup large door on the measurement module is turned on) then press the
green Power On/Off switch on the control panel.
NOTE
The system software will boot up and the hardware will initialize automat-
ically.
C irc u it B re ak e r
1/4" O .D. Va cu um tubin g
con nec tion
S E C S -II C on ne c tion
A C r ece ptac le for m onitor and
op ti onal p rinter P ri nter/P lo tte r
C on n ec tio n
FE 7 U T IL 1 . W M F
System 1. Remove all wafers from the stage, external flat/notch finder, and the
robot.
Shutdown
2. Exit the Focus program that is currently running (such as: Focus
Operator, Focus Interactive, and so on).
You are returned to the OS/2 desktop.
NOTE
5. Press the green Power On/Off switch on the FE System control panel
to turn off the system.
The FE System is powered off.
NOTE
6. Once the FE System is powered off, and if desired, the power cord
can be removed from the outlet or the main power breaker may be
shut off.
3. The laser shutters will be activated and the tilt LED will turn off.
4. For an FE III or FE IV System, the last two LEDs on the rectifier board
(the Positive Rotating Compensator and Positive Motor LEDs) will
turn off.
5. For an FE VII System (with or without SMIF), the LEDs on the rectifier
board corresponding to Positive Chuck Power, Positive Motor, and
Positive Converter Input will turn off.
Recovering Once the EMO/EPO switch has been pressed and the FE System is shut-
down, perform the following:
from an
1. Clear any wafers from the robot or external flat/notch finder.
Emergency
System 2. Shutdown the software using the standard shutdown procedure given
in “Normal System Shutdown” on page A–4.
Shutdown
3. Press the green Power On/Off switch on the control panel.
NOTE
The FE System may now be used. If any wafers were left on the stage,
robot, or flat/notch finder, the system will sense them and will ask where
you wish to unload them.
Configuration
Introduction The purpose of this appendix is to provide information that will allow you
to configure the FOCUS Ellipsometer system and customize the default
parameters and settings for the programs that make up the FOCUS Ellip-
someter software package. The information in this section pertains to the
FOCUS Ellipsometer (FE) III, IV, and VII systems, and includes the fol-
lowing topics:
• How to start up, use, and exit the Focus Setup program.
• Describe the configuration options available through the Focus
Setup program.
• How to create user login names and passwords and configure the
FE System login requirements.
The tasks described in this appendix are performed with the use of the
Focus Setup program. This is the program that allows the user to change
the FOCUS Ellipsometer software configuration. The Focus Setup pro-
gram also provides data file backup, data file restore, and database main-
tenance functions.
NOTE
Starting To start the Focus Setup program, perform the following procedure:
Focus Setup 1. Position the trackball pointer over the Focus Setup icon and double
click the left trackball button.
The Focus Setup main window is displayed as shown in Figure B-1.
Focus Setup The following menu items are available in the Focus Setup Main window:
Main Menu
CAUTION
NOTE
NOTE
- Data Files — Specifies the default filenames for the History log
and the log that contains the data from the measurement that
was last run. Also specifies the log file path, the number of logs
that will be stored, the font style for the log viewer, and how the
pattern recognition data will be grouped in the log.
Setting FE The FE System can be configured to require a user to enter a login name
and password in order to access the following programs:
System
• Focus Operator Queued Loading
Login
Requirements • Focus Interactive
• Focus Recipe Creator
For each of the three programs, you may specify whether a user must log
in, if the user must enter a password, and the security level the user must
possess in order to run the program. In addition, for the Focus Operator
Queued Loading program, you may specify if the user must log in to view
measurement data, and if the user must log out in order to exit the Focus
Operator Queued Loading program.
Use the following procedure to specify the login requirements for each of
the FE programs:
1. If it is not already running, start the Focus Setup program using the
procedure given in “Starting Focus Setup” on page B–2.
The Focus Setup main window is displayed.
2. From the Focus Setup main menu, click on Setup then click on
Login when the Setup menu is displayed.
The Focus Login window is displayed as shown in Figure B-2.
3. Use the following information to set the login requirements for each of
the FE System programs (Focus Operator, Recipe Creator, and
Focus Interactive):
• Access — Specify the privilege level the user must have in order to
access the program by clicking on the checkbox for either Operator
or Engineer.
NOTE
6. You may now exit the Focus Setup program by using the procedure
provided in “Exiting Focus Setup” on page B–10.
7. If user records have not already been created for the FE System,
proceed to “Setting Up User Logins and Passwords” on page B–11.
Exiting To exit the Focus Setup program, perform the following procedure:
Focus Setup 1. On the Focus Setup screen, perform one of the following:
• Double click the left trackball button on the icon located in the upper
left corner of the screen (beside the Focus Recipe Creator program
name).
• Single click the left trackball button on the icon located in the upper
left corner of the screen (beside the Focus Recipe Creator program
name). Select Close from the menu that is displayed.
The Focus Setup program will exit and you will be returned to the
OS/2 desktop.
Creating New Use the following procedure to define a new user record:
User Records 1. Start the Focus Recipe Creator program using the procedures
provided in Chapter 4, “Creating Recipes”.
The Focus Recipe Creator main window is displayed.
2. From the Focus Recipe Creator main menu, click on Template then
click on Operator when the Template menu is displayed.
The Person Selection window is displayed as shown in Figure B-3.
3. Enter the new login name in the Name: field. If desired you may also
enter a brief description of this login name in the Description: field.
Click on the Enter button when you have completed the entries.
4. Enter the desired password for this user in the Password: field.
NOTE
5. Select the security level for this user by clicking the left trackball
button on the desired privilege button. The security level will
determine which programs the user is allowed to access (as
determined by the Focus Setup program). Refer to “Setting FE
System Login Requirements” on page B–7 for information.
NOTE
• If you wish to specify the process specs the user may access:
click on one of the Process Spec. Quick Buttons.
Repeat this step for each Quick Button you wish to assign then
proceed to the next step.
• If you do not wish to specify the process specs the user may
access: proceed to the next step.
9. When you have completed specifying login names, you may exit the
Focus Recipe Creator using the procedures provided in Chapter 4,
“Creating Recipes”.
Renaming, In addition to creating a new user record, you may rename an existing
user record, delete user records from the database, or copy an existing
Deleting, or user record to a new name by using the following procedure:
Copying User
Records 1. Start the Focus Recipe Creator program using the procedures
provided in Chapter 4, “Creating Recipes”.
The Focus Recipe Creator main window is displayed.
2. From the Focus Recipe Creator main menu, click on Template then
click on Operator when the Template menu is displayed.
The Person Selection window is displayed as shown in Figure B-5.
3. Click the left trackball button once on the login name you wish to
change, delete, or copy then click on the Options button.
The Person Selection Options window is displayed as shown in
Figure B-6.
• To delete the login name and its database information: click the
Delete button.
The login name is changed and you are returned to the Person
Selection window.
The login name and password are copied to the new name and you
are returned to the Person Selection window.
• To print the login name and its database information: click the
Print button.
The login name and its database information are printed and you
are returned to the Person Selection window.
5. Repeat Step 3 and Step 4 for each login name you wish to delete,
rename, copy, or print.
6. When you have completed all of the desired changes, in the Person
Selection Window click the Enter button.
The Focus Recipe Creator main window is displayed.
Modifying User Once a user record is created, you can change the user’s password,
security level, and/or process specification quick buttons by using the fol-
Records lowing procedure:
2. From the Focus Recipe Creator main menu, click on Template then
click on Operator when the Template menu is displayed.
The Person Selection window is displayed as shown in Figure B-7.
3. Select the login name to modify by single clicking the left trackball
button on the desired name then clicking the Enter button, or by
double clicking the left trackball button on the name.
The Operator Information window is displayed as shown in
Figure B-8.
NOTE
NOTE
7. When you completed all modifications, you may exit the Focus
Recipe Creator using the procedures provided in Chapter 4, “Creating
Recipes”.
Database In order to protect against the possible loss of database information, the
FE System provides the capability to back up and restore the following
Operations database information and files:
All of these database operations are performed with the use of the Focus
Setup program.
Backing Up the Use the following procedure to back up selected FE System files:
FE Database 1. If it is not already running, start the Focus Setup program using the
procedure given in “Starting Focus Setup” on page B–2.
The Focus Setup main window is displayed.
NOTE
5. Click on OK .
A window is displayed instructing you to insert a blank, formatted
floppy disk in drive A:\.
6. Insert a blank, formatted floppy disk in drive A:\ and click on OK.
A window is displayed indicating that the selected database files are
being written to the disk. You will be prompted to insert each addi-
tional floppy disk that is required to complete the backup.
7. When all files have been written to the selected backup path, a
window will be displayed indicating that the backup is complete.
Click on the OK button.
One of the following will occur:
Restoring the Use the following procedure to backup selected FE System files:
FE Database 1. If it is not already running, start the Focus Setup program using the
procedure given in “Starting Focus Setup” on page B–2.
The Focus Setup main window is displayed.
6. Enter the number of floppy disks that make up the backup set and
click on Enter.
A window is displayed instructing you to insert the first disk in
drive A:\.
7. Insert the first backup disk in drive A:\ and click on OK.
A window is displayed indicating that the files are being copied from
the floppy disk. You will be prompted to insert each additional floppy
disk that is required to complete the restore.
8. When all files have been copied from the selected restore path, a
window will be displayed indicating that the restore is complete. Click
on the OK button.
You are returned to the Focus Setup main window.
• Deleting the image files that are left behind in the database when
associated pattern recognition recipes are deleted.
Delete Vision Use the following procedure to delete the image files that are no longer
System Files associated with a current recipe:
1. If it is not already running, start the Focus Setup program using the
procedure given in “Starting Focus Setup” on page B–2.
The Focus Setup main window is displayed.
2. From the Focus Setup main menu, click on Database then click on
Vision when the Database menu is displayed.
The Vision Database Maintenance window is displayed as shown in
Figure B-9. This window will show all of the image files in the data-
base that are no longer associated with a recipe.
4. Click on the Delete button to delete the selected image files from the
database.
Database
Compression NOTE
1. If it is not already running, start the Focus Setup program using the
procedure given in “Starting Focus Setup” on page B–2.
The Focus Setup main window is displayed.
2. From the Focus Setup main menu, click on Database then click on
Database Compression when the Database menu is displayed.
A window is displayed informing you that this process may take a
long time to complete.
NOTE
Recommended The following table provides the recommended intervals for database
backup and maintenance procedures:
Intervals for
Database
Backup and
Table B-1. Recommended Intervals for Database Operations
Maintenance
Database File Operation Frequency
Recipe Database Backup User determined based on
frequency of Recipe creation.
Log0 File Backup and Delete Quarterly
Configuration Files Backup On installation and after
calibration.
Map Files Backup User determined based on
size of Map file database.
Wafer Database Backup User determined based on
use of the database.
Compression Compression should only be
performed when instructed
to do so by Rudolph Support
personnel.
Vision Files Purge Quarterly
and Filmstacks
Introduction The purpose of this appendix is to provide a listing of the filmstacks, reci-
pes, and measurement controls that are provided by Rudolph Technolo-
gies with your FE System. The information in this section applies to the
FOCUS Ellipsometer (FE) III, IV, and VII systems.
These listings are provided in tabular form with a brief description of each
column following the table.
NOTE
632.8/ Computes thickness of the A-Si and bottom oxide layer. Edit
ad_Amorphous_TT
780 spec thicknesses for your application.
632.8/ Measures thickness of the TEOS layer and the Tin layer
ad_CMP-Ox/TiN/Al-TT
780 simultaneously. Edit spec thicknesses for your application.
632.8/ Measures thickness of the TEOS layer along with the optical
ad_CMP-Ox/Metal-T-Ns
780 constants of the “substrate” underneath.
632.8/ Computes thickness of the top oxide and poly layers. Modify all
ad_Ox/Poly/Ox_TT
780 spec thicknesses for your application.
632.8/ Calibration filmstack for use with Rudolph’s 1100Å SiO2 check
zCal-1100
780 sample wafer.
632.8/ Calibration filmstack for use with Rudolph’s 1000Å SiO2 check
zCal-1um
780 sample wafer.
632.8/ Calibration filmstack for use with Rudolph’s 2300Å SiO2 check
zCal-2300
780 sample wafer.
632.8/ Calibration filmstack for use with Rudolph’s 400Å SiO2 check
zCal-400
780 sample wafer.
632.8/ Calibration filmstack for use with Rudolph’s 950Å Si3N4 check
zCal-950-Nit
780 sample wafer.
NOTE
Default The following wafer recipes are supplied by Rudolph Technologies and
should not be modified or deleted. If necessary, create a copy of the rec-
Recipes ipe and modify the copy.
Recipe Description
Used to measure 49 points of an SiO2 film on a 6" wafer in an
aa6inOxCir49pt-Map
octal circular pattern. A wafer map file is generated.
Used to measure 5 points in a standard pattern of an SiO2 film
aa6inOxStd5pt
on a 6" wafer.
Used to measure 49 points of an SiO2 film on an 8" wafer in an
aa8inOxCir49pt-Map
octal circular pattern. A wafer map file is generated.
Used to measure 5 points in a standard pattern of an SiO2 film
aa8inOxStd5pt
on an 8" wafer.
Used to measure 20 points along a line to create a linescan
aa8inOx_Linescan_20pt
graph of an SiO2 film on an 8" wafer.
Throughput test recipe using an 8" wafer and a circular 49 point
tp8inCir49pt_Fast-Map measurement pattern in a high speed measurement mode. A
map file is generated.
Throughput test recipe using both IR and HeNe lasers to
tp8inStd5pt_Dual
measure 5 points in a standard pattern on an 8" wafer.
Throughput test recipe using high speed mode to measure 5
tp8inStd5pt_Fast
points in a standard pattern on an 8" wafer.
Throughput test recipe applying rough film measurement
tp8inStd5pt_Rough techniques to measure 5 points in a standard pattern on an 8"
wafer.
Throughput test recipe using high repeatability measurement
tp8inStd5pt_Slow
mode to measure 5 points in a standard pattern on an 8" wafer.
Throughput test recipe using high speed mode to measure 9
tp8inStd9pt_Fast
points in a standard pattern on an 8" wafer.
Throughput test recipe using high repeatability measurement
tp8inStd9pt_Slow
mode to measure 9 points in a standard pattern on an 8" wafer.
Used to measure 5 points in a standard pattern of an SiO2 film
xx5in_Notch_Std5pt
on an 5" notched wafer.
Used to measure 10 points in a center pattern of an SiO2 film
xx5in_Ox_10pt_cnt
on a 5" wafer.
Recipe Description
Used to measure 10 points in a center pattern of a Poly layer
xx6in_2.7KPoly_10pt_cnt
with a thickness of 2.7kÅ on a 6" wafer.
Used to measure 10 points in a center pattern of a Poly layer
xx6in_4KPoly_10pt_cnt
with a thickness of 4kÅ on a 6" wafer.
Used to measure 10 points in a center pattern of a Poly layer
xx6in_550Poly_10pt_cnt
with a thickness of 550Å on a 6" wafer.
Used to measure 10 points in a center pattern of an SiO2 film
xx6in_Ox_10pt_cnt
on a 6" wafer.
Used to measure 5 points in a standard pattern of an SiO2 film
xx8in_Ox_5pt_Fast
on an 8" wafer using high speed mode.
NOTE
NOTE
Quick Reference
Introduction The purpose of this appendix is to provide menu maps for each of the
programs that make up the FOCUS software package. The information in
this section pertains to the FOCUS Ellipsometer (FE) III, IV, and VII sys-
tems, and includes the following:
Detailed descriptions for the selections from each of the menus are
included in the chapter that pertains to that particular program.
Focus Figure D-1 through Figure D-4 show the menu maps for the Focus Inter-
active program. Descriptions of the menu items and instructions for using
Interactive the Focus Interactive program are provided in Chapter 3, “Developing
Menus Film Applications”.
Control Menu
Mag >> Mag Menu
Zoom >>
Wafer View
Live Off/Live On
Low Mag
Laser Off/Laser On
High Mag
Illumination
Site Locator Menu
Digital/NTSC
LoMag/HiMag Scale
Focus
Control >> Zoom Menu
Move >> Move Menu
One Quarter
Keyboard One Half
Center One X
Left
Right
Top
Bottom
Layers Menu
Filmstack Model Menu Add Layer
Delete Top
Layers >>
Error Est. >>
Enter
Cancel Error Est. Menu
On
Off
Add
Delete
Enter
Cancel
Focus Recipe Figure D-5 through Figure D-7 show the menu maps for the Focus Rec-
ipe Creator program. Descriptions of the menu items and instructions for
Creator using the Focus Recipe Creator program are provided in Chapter 4, “Cre-
Menus ating Recipes”.
Registration Menu
Wafer Recipe Menu
None
Name Manual
Wafer Information Automatic
Film Stack Reteach
Registration >>
Unpatterned Wafer
Patterned Wafer >> Patterned Wafer Menu
Report
Save Train Patterns
Cancel Select Measured Chips
Load Menu
SMIF Arm Menu
Manual Load
Cassette Load Smif1 Load Cassette
Unload Wafer Smif1 Unload Cassette
SMIF Arm >> Smif2 Load Cassette
SMIF Indexer >> Smif2 Unload Cassette
Vision Menu
SMIF Indexer Menu
Setup
SMIF1 Load Cassette
SMIF1 Unload Cassette
SMIF2 Load Cassette
SMIF2 Unload Cassette
Control Menu
Mag >> Mag Menu
Zoom >>
Wafer View
Live Off/Live On
Low Mag
Laser Off/Laser On
High Mag
Illumination
Site Locator Menu
Digital/NTSC
LoMag/HiMag Scale
Focus
Control >> Zoom Menu
Move >> Move Menu
One Quarter
Keyboard One Half
Center One X
Left
Right
Top
Bottom
View Menu
Die Pattern
Stepper Group Wafer Pattern
Pattern Training Menu
View >>
Modify >> Modify Menu
Reteach >>
Close Die Pattern
Done Stepper Group
Reteach Menu
One Die
Stepper Group
Figure D-7. Focus Recipe Creator Stepper Group Pattern Training Menu
Focus Figure D-8 through Figure D-9 show the menu maps for the Focus Oper-
ator program. Descriptions of the menu items and instructions for using
Operator the Focus Operator program are provided in Chapter 2, “Operator Inter-
Menus face”.
Control Menu
Mag >> Mag Menu
Zoom >>
Wafer View
Live Off/Live On
Low Mag
Laser Off/Laser On
High Mag
Illumination
Site Locator Menu
Digital/NTSC
LoMag/HiMag Scale
Focus
Control >> Zoom Menu
Move >> Move Menu
One Quarter
Keyboard One Half
Center One X
Left
Right
Top
Bottom
Focus Figure D-10 through Figure D-12 show the menu maps for the Focus
Browser program. Descriptions of the menu items and instructions for
Browser using the Focus Browser program are provided in Chapter 5, “Wafer Map-
Menus ping and Data Reporting”.
Define Query
Display Graph
Display Report >> Display Report Menu
Display SQL
Text Report
Data Dump
View Menu
Zoom Out
Dependent Menu
Zoom In
Ind vs Dep1 Multiply 10
Ind vs Dep2 Divide 10
Ind vs 2Dep Statistics
Histogram N ✱ Sigma
Parametric Markers
SPC Join Points
Time Area
Error Bar Limits
Graph Menu
View >>
Dependent >>
Independent >> Independent Menu
Mouse >>
Data to File Multiply 10
Mouse Menu
Editor Divide 10
SPC Zoom Drag Join Points
Scale >> Digitize
Text Drag
Scale Menu
User Scale
Auto Scale
Focus Figure D-13 through Figure D-16 show the menu maps for the Focus
Mapping program. Descriptions of the menu items and instructions for
Mapping using the Focus Mapping program are provided in Chapter 5, “Wafer
Menus Mapping and Data Reporting”.
Focus Mapping
Main Menu
Maps
Clear All
NoFix
Etch >> Etch Menu
LineScan
Difference Map
Etch Rate Map
Etch Menu
Color
Grey
Black + White
Divisions >> Divisions Menu
Wafer Map Menu Automatic
User Range
View
Scale
Color >>
Mode >> Mode Menu
Edit >>
Text Topograph
Update Contour
Spec Line
Draw Box
Edit Menu
Draw Contour
Data Draw Grid
Save Draw Marker
View Menu
Zoom Out
Dependent Menu
Zoom In
Ind vs Dep1 Multiply 10
Ind vs Dep2 Divide 10
Ind vs 2Dep Statistics
Histogram N ✱ Sigma
Parametric Markers
SPC Join Points
Time Area
Error Bar Limits
Graph Menu
View >>
Dependent >>
Independent >> Independent Menu
Mouse >>
Data to File Multiply 10
Mouse Menu
Editor Divide 10
SPC Zoom Drag Join Points
Scale >> Digitize
Text Drag
Scale Menu
User Scale
Auto Scale
Focus Setup Figure D-17 shows the menu map for the Focus Setup program. Descrip-
tions of the menu items and instructions for using the Focus Setup pro-
Menus gram are provided in Appendix B, “System Configuration”.
Setup Menu
Model Number
Login
Simulation
Initialization
Mapping
Vision & Video
Measurement
Wafer Handler
Focus Operator
Database
Data Files
Miscellaneous
PM >> PM Menu
Parallel Solver
D-Series Maintenance Activation 1
Signal Tower Maintenance Activation 2
Throughput Maintenance Activation 3
Maintenance Activation 4
Setup >>
Backup >> Backup Menu
Restore >>
Database >> Recipe Database
Log0
Configuration Files
Restore Menu
Map Files
Recipe Database Wafer Database
Configuration Files
Map Files
Wafer Database
Database Menu
Vision
Database Compression
In general, the equations which relate the polarization state change to the
sample properties cannot be solved for the sample properties analytically.
Consequently, the sample properties are determined through an exten-
sive modeling process.
Interaction of The index of refraction (N) and extinction coefficient (K) are the constants
that characterize the optical properties of a material.
Light with
Material Refractive index is defined by the equation N=C/V, which describes the
ratio of the velocity of light in vacuum to the velocity of light when passing
through another medium. N≅1 for light propagating through air and other
gasses.
The extinction coefficient (K) describes how absorbing a material is. For
dielectric materials such as glass and transparent films, little visible light
is absorbed and K≅0. This is also the case for light in vacuum.
Polarized The polarization state of light is described by the orientation of the elec-
tric fields of its components. The components that constitute the resultant
Light electric field are represented as Ex and Ey. Polarized light may be
obtained by passing the light beam through polarizing optics or by caus-
ing the beam to be reflected under certain conditions.
Unpolarized light has components with electric fields oriented in all direc-
tions perpendicular to the direction of propagation.
Linear Linearly or plane polarized light occurs when all of the photons in a light
beam have the electric field oriented in one direction. If two linearly polar-
Polarization ized light beams, one vertical and one horizontal, with the same fre-
quency, amplitude, and phase are combined, the resultant electric field
vector is also linearly polarized 45° from each of the original beams. If the
amplitudes were not the same, the angle of polarization would be deter-
mined by the vector sum of the electric field vector.
Ey
Ex
Circular Circularly polarized light results from the combination of two linearly
polarized light beams whose electric field components are of equal mag-
Polarization nitude and are 90° out of phase. Looking down the direction of propaga-
tion, the tip of the resultant electric field vector would appear to trace a
circle.
Ey
Ex
Elliptical Elliptically polarized light, the most general form of polarized light, exhib-
its electric field components with any phase difference and unequal mag-
Polarization nitudes.
Ellipsometry ∆ (delta) is the change in phase difference between the P and S compo-
nents of the incident light and the P and S components of the reflected
Parameters ∆ light. ∆ ranges from zero to 360°.
and Ψ Ψ (psi) is the angle whose tangent is the ratio of the magnitudes of the
total reflection coefficients Rp and Rs . Ψ represents the change in ampli-
tude upon reflection, and can be between zero and 90°.
Figure E-7. Two Polarization States (S&P) for Light Reflected from a Surface
It can be seen that for materials with similar N values, it is difficult to dif-
ferentiate between their trajectories near the cycle thickness. By measur-
ing ∆ and Ψ at multiple angles of incidence, N can be calculated more
precisely.
Instrumentation The following sections discuss the equipment needed to perform the
measurements of the film materials.
Null Ellipsometers The Null Ellipsometer is based on the principal that by fixing the com-
pensator at 45°, the polarizer and analyzer are rotated until a null is
found, or the light has been extinguished. Therefore, upon reflection,
the ellipticity created by the compensator is canceled.
∆ and Ψ are related to the settings of the polarizer and analyzer, and
can be easily calculated.
The advantages of RCE are that it is fast, precise, has data processing
flexibility, operates at high detected light levels, and eliminates the hand-
edness ambiguity problem (the inability to distinguish between comple-
mentary polarization states).
Fas t A xis
z
S lo w A x is
x
Inc id en t L ig ht P olarize d
|| to F a st A xis
C o m p ensa tor
y
I= Ioco s2q
z
Fa st A xis
x
S low A x is
Incid ent L igh t P olarize d
45 o to F as t A xis
C o m pe nsa to r
I= Io co s2q
Fit Error and Fit error is a χ2 (Chi squared) statistical quantity that indicates how well
the measured ∆ and Ψ data fit the theoretical data of the film model.
Error
Estimates A small fit error means that the measured and modeled data agree to
within the accuracy of the measurement. A large fit error means that the
measured data is incompatible with the theoretical data of the film model.
Usually, the measured ∆ and Ψ data are correct to within the accuracy of
the measurement. Therefore, large fit errors can be attributed to a prob-
lem with the theoretical data of the film model, such as incorrect N, K , or
T of a fixed underlying layer.
Film The Engineer first creates a filmstack to compare the measured data with
the theoretical modeled data.
Parameter
Computation Upon making a measurement, the modeled ∆ and Ψ values are com-
pared with those that were measured.
Algorithm
The program searches for the best solution by finding that which has the
lowest fit error. The program is constantly updating the film model by
varying the calculated parameters until the measured and modeled data
converge, and the lowest fit error is found.
If the solution found by the program is not acceptable (the fit error is too
high), the user must then change the filmstack parameters.
Order Search There are two types of order search modes offered by the FOCUS soft-
ware to find the minimum fit error and to resolve order.
Algorithm
In Standard Order Search mode, using a single wavelength measure-
ment, the region searched varies with the application but is typically ± 5
order cycles. This option is most appropriate for low refractive index
films, such as SiO2 and Si3N4. When using a dual wavelength measure-
ment, the program starts from the spec thickness entered by the user,
and searches ± 4 order cycles.
Single λ Dual λ
Search Type
(632.8 nm) (632.8 nm + 788 nm)
Restricted Spec thickness ± 1/2 Spec thickness ± 2
Search order cycles. order cycles.
Standard Spec thickness ± 5 Spec thickness ± 4
Order Search order cycles for most order cycles.
applications.
Getting Help
Introduction The purpose of this appendix is to provide the user with information nec-
essary to perform simple troubleshooting and maintenance tasks. The
information in this section applies to the FOCUS Ellipsometer (FE) III, IV,
and VII systems.
NOTE
The troubleshooting tables that follow are not an all inclusive list of every
possible trouble resolution and should be used for reference only. If using
the procedures given in this section do not resolve the problem, contact
Rudolph Technologies for assistance and/or support using the informa-
tion provided at the end of this appendix.
WARNING
FE System Use the information in the following table to troubleshoot possible hard-
ware problems with your FE System. Locate the applicable symptom in
Hardware the left column of the table to determine probable causes and recom-
Symptoms mended actions to rectify the problem.
Software Use the information in the following table to troubleshoot possible soft-
ware problems with your FE System. Locate the applicable error mes-
Error sage in the left column of the table to determine probable causes and
Messages recommended actions to rectify the problem.
NOTE
Err Error
Probable Causes Recommended Actions
# Message
Alignment Matrix Not Align.dat file corrupted. Contact Rudolph Technologies.
6
Calibrate: PC.
ILLEGAL CASSETTE Incorrect wafer cassette identifier specified. Enter correct cassette identifier.
10
SPECIFIED: PC.
ILLEGAL WAFER Incorrect wafer location specified. Enter correct wafer location.
11
SPECIFIED: PC.
ILLEGAL OPTICAL BIAS Invalid optical bias selected. Enter a valid optical bias.
22
LEVEL.
Err Error
Probable Causes Recommended Actions
# Message
EEPROMB.DAT FILE EEPROMB.DAT file missing. Create file and enter command
23 MISSING. “upload base” or “upload_b” in
FOCTEST.
EEPROMO.DAT FILE EEPROMO.DAT file missing. Create file and enter command
24 MISSING. “upload optic” or “upload_o” in
FOCTEST.
WAFER REPOSITION Unable to center the wafer. Make sure flat/notch setting is
25 ERROR: PC. correct.
Recalibrate flat/notch finder.
Illegal Light Bulb Value entered that was outside the valid Specify value that is within the
28 Illumination. Valid Range range. specified range.
(0-31)
Base2 Control Board Not Invalid FOCUS configuration. Check FOCUS configured
29
Installed. correctly in SETUP program.
37 Extfn Not Installed. External flat/notch finder not installed. Check configuration of FOCUS.
Err Error
Probable Causes Recommended Actions
# Message
EEPROMIF.DAT file First time installation. Run FOCTEST and type
missing. Please run “upload intf”.
42 FOCTEST and upload
eeprom values from
interface control board.
Interface Control Board PC to Interface Control Board Cable not Check cables.
44 Not Communicating with connected.
PC. Interface Control Board failure. Replace interface control board.
Wafer found on robot. Wafer on robot arm when robot is Manually remove wafer from
Cannot initialize robot initialized. On SMIF units, the robot cannot robot.
45 while wafer present. be initialized when wafer is on robot.
Please manually remove
wafer from robot.
The base control board Failure to align optic head to wafer because Check wafer.
has been requested to lifter moved to limit.
109 Check alignment system and
move a lifter past its
recalibrate it.
limits.
The wafer FLAT was not Failure to find flat on wafer. Check wafer has a flat.
110
found.
The wafer NOTCH was Failure to find notch on wafer. Check wafer has a notch.
111
not found.
112 The lifters are not homed. Lifters not initialized. Initialize unit.
A wafer has been found Wafer unexpectedly found on robot. Remove wafer from robot by
113
on the robot. unloading it into a cassette.
No Wafer is loaded on Wafer not found on stage when a wafer Check vacuum system.
114 the stage. unload is attempted. Check wafer is on stage.
Err Error
Probable Causes Recommended Actions
# Message
Illegal Cassette Attempt to unload wafer to invalid cassette Retry command.
115
Destination. plate.
The wafer transfer cannot Cassette sizes are different. Place identical cassette sizes
be performed because for wafer transfer.
116
the two cassette sizes do
not match.
The Wafer was Not Wafer not found in cassette slot. Check wafer present.
124 Found in the Cassette Check vacuum system.
Slot.
STAGE VACUUM Stage vacuum sensor failure. Replace stage vacuum sensor.
127
ERROR.
ROBOT VACUUM Robot vacuum sensor failure. Replace robot vacuum sensor.
128
ERROR.
Vision Process unable to Possible OS/2 or memory problems. Exit from the program and
633
open semaphores. reboot OS/2.
Vision Process unable to Possible OS/2 or memory problems. Exit from the program and
634
set semaphores. reboot OS/2.
Vision Process unable to Possible OS/2 or memory problems. Exit from the program and
635
close semaphores. reboot OS/2.
Err Error
Probable Causes Recommended Actions
# Message
Vision Process unable to Possible OS/2 or memory problems. Exit from the program and
636
clear semaphores reboot OS/2.
Vision Process unable to VSP.cnf is missing from Retrieve vsp.cnf from FOCUS
638
find configuration file. d:\focus\vision\setup. vision install disk.
Vision Process unable to Filenum.str is missing from Find out the largest pattern
find number generation d:\focus\vision\setup. number in d:\focus\vision\pat,
639 file.
then edit filenum.str, enter the
number and save it.
Poor vision pattern teach Poor features or improper threshold setup. Select different site or re-adjust
641
quality. thresholds.
Not enough successful Poor feature site or repeat pattern site. Select another site.
642
alignments to calibrate.
Vision Calibration Figure Bad calibration site. Select another site and re-
643
Of Merit too high. calibrate.
Vision Teach unable to Poor teach site or hardware sync problem. Select another site or check the
644 find enough patterns to hardware.
qualify for alignment.
Vision Teach unable to No space on hard disk. Check the hard disk space on
645
write patterns. D:\ drive.
Vision Teach unable to Pattern is missing or hardware failure. Check the pattern in
648
read patterns. d:\focus\vision\pat directory.
Vision Process unable to OS/2 memory management problem. Shutdown OS/2 and reboot.
649
allocate.
Vision Search unable to Color variation or pattern is outside FOV. Contact Rudolph Service
650 find 2 patterns in site Department.
pattern file.
Vision Search unable to Color variation or pattern is outside FOV. Contact Rudolph Service
652 find pattern alignment Department.
quality above threshold.
Err Error
Probable Causes Recommended Actions
# Message
Vision Search coupling Possible repeat patterns or similar patterns Use restricted search.
653 exceed Coupling in FOV.
Difference Threshold.
Protect Device not EEPROM is missing on Imaging Tech Contact Rudolph Service
656 installed in Imaging Board. Department.
Technology Board.
Ill formed Vision System Incomplete localib.par or hicalib.par. Check the files. Re-calibrate
657
calibration matrix. vision system if necessary.
Vision Process made call Possible software bugs. Contact Rudolph Service
658 with parameter out of Department.
range.
Vision Process unable to Poor search image or pattern is outside Contact Rudolph Service
659
find any valid pattern. FOV. Department.
Vision Process unable to OS/2 API failure. Shutdown OS/2 and reboot or
660 free shared data. consult with Rudolph Software
Group.
Vision child process is Hardware security key is not installed. Contact Rudolph Service
661 not running, unable to Department to install the
register wafer. security key.
Vision Process unable to Communication failure between vision Re-plug in the board or replace
662
initialize vision board. board and PC. the board.
Vision child process is Communication failure between PC and Check Cognex device driver is
unable to open pseudo Cognex board. properly installed, board is
663
serial port. properly plugged in and
functions normally.
Vision child process is Communication failure between PC and Check Cognex device driver is
unable to close pseudo Cognex board. properly installed, board is
664
serial port. properly plugged in and
functions normally.
Vision child process is Communication failure between PC and Check Cognex device driver is
unable to reset. Cognex board. properly installed, board is
665
properly plugged in and
functions normally.
Vision child process is Cognex functions abnormally. Check the board and run
666
not in polling mode. reset_vp.
Parent Vision Process OS/2 API failure. Shutdown OS/2 and reboot or
675 unable to free shared consult with Rudolph Service
data. Department.
Err Error
Probable Causes Recommended Actions
# Message
Parent Vision Process OS/2 failure. Contact Rudolph Service
677 unable to create Department.
semaphores.
Parent Vision Process The vismain could be still running. Terminate vismain.
682 unable to execute Child
Vision Process.
NOTE
Glossary
Analog to Digital Board (A/D) Chip and Die
That part of the detector system which changes the Used interchangeably in this document. They refer
output of the detector from an analog to a digital to one integrated circuit on the wafer.
form.
Collimated Diode Laser
Analysis Used to generate the probe beam required by the
The data reduction routines applied to the measure- tilt control system.
ments.
Collimating Lens
Analyzer Makes parallel the diverging rays of light emerging
A manually rotatable Glan-Taylor polarizer assem- from the spatial filter.
bly.
Compensator
Analyzer Arm A mica quarter-wave plate.
An assembly of all the components mounted on the
analyzer arm plate. Used to “analyze” the polariza- Computer Bay
tion state of the light reflected from the sample. The lower right-hand side of the measurement con-
sole that contains the computer CPU and disk
Application drives.
A film measurement requirement.
Database
Barcode Reader Information stored in a structured, easily retrievable
A device used to scan the identification code on the fashion. Consists of the measured data, calculated
bottom of each wafer. parameters, thickness, index of refraction and wa-
fer data.
Batch
A numbered group of wafers. Delta (∆)
The relative phase shift between the S and P polar-
Border izations after reflection from the sample.
The boundary of a window. Clicking on the border
and dragging it changes the size of the window. Die
The region on a wafer that defines one integrated
Browser circuit (chip).
A means for accessing database information.
Dual Wavelength
Cassette Combined 633 nm HeNe laser and 780 nm diode la-
A container for transporting and storing wafers. ser measurement system.
Microscope Report
Used to provide magnified views of the wafer on the A template used to control the format of the mea-
video monitor. sured data.
Process Specification
A group of process steps.
Index
A C
Suggestions:
(Please be specific and provide marked up copies of pages where necessary.)
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
General Comments:
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________