Spectrophotometers UV Visible
Spectrophotometers UV Visible
Spectrophotometers, UV/Visible
solution concentrations for quantitative analyses or,
Scope of this Product Comparison with more sophisticated instruments, for enzyme or
reaction kinetics assays.
This Product Comparison covers general-purpose
spectrophotometers having spectral wavelength Principles of operation
capabilities in the ultraviolet and visible (UV/Vis)
regions, most commonly 190 to 900 nanometers The basic components of a spectrophotometer are a
(nm). Although a few of the models listed can ana- light source, a wavelength monochromator, focusing
lyze into the near infrared (NIR) region, instru- devices, a cuvette, a photodetector, and a display de-
ments dedicated for infrared (IR) analysis differ in vice (see Fig. 1). Light passes through a monochroma-
design and optical materials and are generally tor that separates the light into its component
considered a separate class of device; therefore, wavelengths. Slits then isolate a bandpass (the seg-
they are not included in this report. Information on ment of the spectrum isolated by the monochromator)
atomic absorption spectrophotometers is available of the wavelength needed for measurement, improving
in the Product Comparison titled SPECTRO- its purity. Next, the light passes through the sample,
PHOTOMETERS, ATOMIC ABSORPTION.
and a portion of the radiant energy is absorbed; the
amount depends on the nature and concentration of
UMDNS information the sample. The portion of the bandpass that is not
absorbed is transmitted to a photodetector, which con-
This Product Comparison covers the following
verts light energy to electrical energy that can be
device term and product code as listed in ECRI’s
registered on a meter or digital readout. When a pho-
Universal Medical Device Nomenclature System™
todiode array detection system is used in place of a
(UMDNS™):
• Spectrophotometers, UV/Visible [15-083]
Purpose
A spectrophotometer optically determines the ab-
sorbance or transmission of characteristic wave-
lengths of radiant energy (light) by a chemical species
in solution. Each molecule absorbs light at certain
wavelengths in a unique spectral pattern because of
the number and arrangement of its characteristic func-
tional groups, such as double bonds between carbon
atoms. According to the Beer-Lambert law, the amount
of light absorbed at these wavelengths is directly pro-
portional to the concentration of the chemical species.
Spectrophotometers are generally used to determine Basic UV/Vis spectrophotometer
Monochromators
2 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
C236UN2B
Light Sample
Source Cuvette
photodiode arrays, offer better stability than single-
element, vacuum-based detectors such as vacuum pho-
Figure 2. Basic components of a diode-array todiodes and PMTs. Solid-state detectors also have less
spectrophotometer lag (persistence of a charge image on the camera tube)
and blooming (spreading of charge from channels hav-
ing a large signal to adjacent channels).
Cuvettes
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 3
Healthcare Product Comparison System
An LCC analysis can be used to compare high-cost • Cost for disposables = $2,160/year ($0.90/cuvette)
alternatives and/or to determine the positive or negative Total Operating Costs = $2,160 for year 1;
economic value of a single alternative. For example, $3,060/year for years 2 through 7
hospitals can use LCC analysis techniques to examine
the cost-effectiveness of leasing or renting equipment PV = ($27,698)
versus purchasing the equipment outright. Because it
Alternative 2 — Spectrophotometer/reusable glass cu-
examines the cash-flow impact of initial acquisition costs
vettes
and operating costs over a period of time, LCC analysis
is most useful for comparing alternatives with different Capital Costs
cash flows and for revealing the total costs of equipment • Spectrophotometer, standard configuration =
ownership. One LCC technique — present value (PV) $8,500
analysis — is especially useful because it accounts for
• Glass cuvettes, 300, various sizes = $900
inflation and for the time value of money (i.e., money
received today is worth more than money received at a Total Capital Costs = $9,400
later date). Conducting a PV/LCC analysis often demon- Operating Costs
strates that the cost of ownership includes more than just
the initial acquisition cost and that a small increase in • Service contract, years 2 through 7 = $900/year
initial acquisition cost may produce significant savings • Cost for replacing cuvettes every 2 years =
in long-term operating costs. The PV is calculated using $450/year (Note: in addition to the periodic replace-
the annual cash outflow, the dollar discount factor (the ment of reusable cuvettes, the cost associated with
cost of capital), and the lifetime of the equipment (in cleaning the cuvettes after each use should also be
years) in a mathematical equation. considered)
Total Operating Costs = $450 for year 1; $1350/year
Buyers can use a PV/LCC analysis to compare long-
for years 2 through 7
term costs involving accessories and disposables used
by spectrophotometers.
PV = ($17,207)
The following represents sample seven-year As illustrated by the above sample PV/LCC analy-
PV/LCC analyses comparing the costs of spectro- sis, the initial acquisition cost is only a fraction of the
photometers with reusable glass cuvettes versus units total cost of operation over seven years. Also, a decision
using disposable plastic cuvettes. to use disposable over reusable accessories should take
4 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
into account the long-term costs of these items. The laboratories that require spectrophotometers only to
analyses for plastic and glass cuvettes demonstrate measure and display solution concentrations should
that the cost of replacing all the used cuvettes every consider those models that only perform quantitative
two years is still significantly less than the cost of using analysis; these units typically range in cost from
disposable cuvettes over this period. Therefore, rather $3,000 to $7,000. More sophisticated units, which can
than making a purchase decision based solely on the cost from $10,000 to more than $70,000, come with
acquisition cost of a basic unit, buyers should consider more standard and optional features, such as autosam-
operating costs, especially costs of reusables versus plers, cuvette autochangers, temperature-regulated
disposables, over the lifetime of the equipment. cell holders, and either PCs or in-board computers with
large memory capacities, analytical and graphics pro-
For further information on PV/LCC analysis, read-
grams, video displays, and data printers, than do basic
ers should refer to ECRI’s Health Devices December
models. These spectrophotometers have powerful com-
1984-January 1985 Special Issue, “Devices & Dollars.”
putational ability and software for data acquisition,
For customized analyses and purchase decision sup-
manipulation, analysis, and presentation that can per-
port, readers should contact ECRI’s SELECT™ Group.
form enzyme and reaction kinetics assays and spectral
Hospitals can purchase service contracts or service on processing.
a time-and-materials basis from the supplier. Service
may also be available from a third-party organization. Purchasing several units from one supplier could
The decision to purchase a service contract should be result in a significant discount. Standardization of
carefully considered. The decision to purchase a service equipment can make staff training easier, simplify
contract can be justified for several reasons. Most suppli- servicing and parts acquisition, and provide greater
ers provide routine software updates, which enhance the bargaining leverage when negotiating new equipment
system’s performance, at no charge to service contract purchases and/or service contract costs.
customers. Furthermore, software updates are often cu-
mulative; that is, previous software revisions may be
required in order to install and operate a new perform- Stage of development
ance feature. Purchasing a service contract also ensures UV/Vis spectrophotometry use has been well estab-
that preventive maintenance will be performed at regu- lished since the early 1950s. Recent advances in micro-
lar intervals, thereby eliminating the possibility of unex- computers, detectors, and diffraction gratings have
pected maintenance costs. Also, many suppliers do not made today’s instruments capable of previously unat-
extend system performance and uptime guarantees be- tainable measurements. Also, a UV light source that
yond the length of the warranty unless the system is consists of a luminous crystal irradiated by a radioac-
covered by a service contract. tive element has been developed. This component has
ECRI recommends that, to maximize bargaining a lifetime of more than 10 years; conventional light
leverage, hospitals negotiate pricing for service con- sources may last for only 1 year.
tracts before the system is purchased. As a guideline, Key parameters for UV/Vis spectrophotometers are
full-service contracts typically cost approximately 10% reproducibility, accuracy, and wavelength range,
of the spectrophotometer’s purchase price. Additional which have been improved with features such as auto-
service contract discounts may be negotiable for matic baseline correction and lower levels of stray light
multiple-year agreements, or for service contracts (radiation outside the bandpass that is transmitted by
that are bundled with contracts on other spectro- the monochromator) in the instrument. Recent ad-
photometers in the department or hospital. vances in grating technology have resulted in high-
Buyers should make sure that applications training resolution, low-stray-light spectrophotometers for
is included in the purchase price of the system. Some many applications.
suppliers may offer more extensive on-site or off-site
training programs for an additional cost.
Bibliography
ECRI recommends that buyers consider the number
and types of tests performed before deciding on a Bennington JL, ed. Saunders dictionary & encyclope-
specific system configuration. Also, if multiple units dia of laboratory medicine and technology. Philadel-
are necessary to handle the test volume, hospitals phia: WB Saunders; 1984.
should consider the types of systems and capabilities that
need to be purchased to avoid paying for unnecessary Burtis CA, Ashwood ER, eds. Tietz textbook of clinical
analysis packages and assaying features. For example, chemistry. 2nd ed. Philadelphia: WB Saunders; 1994.
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 5
Healthcare Product Comparison System
Canadian Standards Association. Laboratory equip- Beckman Coulter Instruments Japan Ltd
ment [standard]. C22.2 No. 151-M1986(R1992). [175265]
1986 (reaffirmed 1992). Sasazuka Center Bldg 2-1-6 Sasazuka
Shibaya-ku
International Electrotechnical Commission. Medical Tokyo 151
electrical equipment — part 1: general require- Japan
ments for safety [standard]. 2nd ed. 60601-1. 1988. Phone: 81 (3) 53522820
Fax: 81 (3) 53522818
Medical electrical equipment — part 1: general re-
quirements for safety. Amendment 1 [standard]. bioMerieux
2nd ed. 60601-1-am1. 1991.
Marketed worldwide
Medical electrical equipment — part 1: general re-
quirements for safety. Amendment 2 [standard]. bioMerieux SA [140148]
2nd ed. 60601-1-am2. 1995. Marcy l’Etoile
F-69280 Charbonnieres Les Bains
Medical electrical equipment — part 1: general re- France
quirements for safety. Section 2. Collateral standard: Phone: 33 (478) 872000
electromagnetic compatibility — requirements and Fax: 33 (478) 872090
tests. 2nd ed. 60601-1-2. 1993. E-mail: [email protected]
6 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 7
Healthcare Product Comparison System
8 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
Perkin-Elmer Instruments Asia Pte Ltd [175323] Shimadzu Scientific Instruments Inc [103338]
111 N Bridge Rd No 29-01/02/03/04 7102 Riverwood Dr
Peninsula Plaza Columbia MD 21046-1284
Singapore 179098 Phone: (410) 381-1227, (800) 477-1227
Republic of Singapore Fax: (410) 381-1222
Phone: 65 (336) 0322
Fax: 65 (338) 3991 Spectronic
Marketed worldwide
Pharmacia Biotech
Marketed worldwide Spectronic Instruments [272853]
820 Linden Ave
Pharmacia Biotech Asia Pacific Ltd [175322] Rochester NY 14625
15th Floor North Cornwall House Taikoo Pl Phone: (716) 248-4000, (800) 654-9955
979 Kings Rd Fax: (716) 248-4014
Quarry Bay E-mail: [email protected]
Hong Kong SAR Internet: http://www.spectronic.com
People’s Republic of China
Phone: 852 28118693
Turner
Fax: 852 28115270
Barnstead/Thermolyne Corp [106642]
Pharmacia Biotech Inc [152712]
800 Centennial Ave 2555 Kerper Rd
PO Box 1327 PO Box 797
Dubuque, IA 52004-0797
Piscataway NJ 08855-1327
Phone: (732) 457-8000, (800) 526-3593 Phone: (319) 556-2241, (800) 553-0039
Fax: (800) 329-3659 Fax: (319) 589-0516
Internet: http://www.biotech.pharmacia.se
Varian
Secomam Marketed worldwide
Marketed in Africa, Asia, Europe, and North America
Varian Analytical Instruments
Secomam [152683] Div Varian Assoc Inc [161922]
4 rue des Charpentiers PO Box 9000
boite postale 106 Dept 87
F-95334 Domont Cedex San Fernando CA 91340
France Phone: (415) 424-4819, (800) 926-3000
Phone: 33 (1) 39354200 Fax: (281) 240-6752
Fax: 33 (1) 39913018 Internet: http://www.varian.com
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 9
Healthcare Product Comparison System
10 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
CE MARK (MDD) Not specified Not specified Not specified Not specified
CONFIGURATION Single beam Single beam Single beam Single beam
SCAN SPEED, nm/min NA 120, 240, 600, 120, 240, 600, 0.1 sec full-
1,200, 2,400 1,200, 2,400 spectrum scan
LAMP TYPE(S) Quartz halogen Tungsten; deuterium Tungsten; deuterium Tungsten : Deuterium
WAVELENGTH, nm
Range 335-825 190-1,100 190-1,100 190-800
ZERO DRIFT 0.003 A/hr <0.003 A/hr <0.003 A/hr <0.002 A/hr
Auto compensation No Not specified Not specified Not specified
RESPONSE TIME, sec <3 0.05 0.05 0.1
DISPLAY LED Color or monochrome, Color or monochrome, Graphics color CRT
VGA CRT VGA CRT
DATA PROCESSING Absorbance, trans- 20 readings/sec, up Same as DU-640, plus Scan vs. time or
mittance, concen- to 12 fixed wave- step programming for wavelength, %T or A,
tration, 0-100 mV lengths, scanning, generation of custom software for
analog out spectral manipula- programs for data kinetics, nucleic
tion, time-based en- acquisition, sampler acids, proteins,
zyme kinetics, rate control, data fraction read, file
and activity report- reduction, transfer, program-
ing, RS232 port for reporting, etc., mable, mouse, Quant,
data and remote con- without the need for Full Spectrum
trol, printer port, an external computer Quantitation (FSQ),
RediRead and Redi- enzyme mechanism,
Scan programs, can activity, and
also use programs multiwavelength
generated on DU-650 enzyme analysis
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 11
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE 2/year Not specified Not specified Not specified
H x W x D, cm (in) 41 x 34 x 22 58 x 69 x 53 58 x 69 x 53 53 x 64 x 56
(16 x 14 x 9) (22.8 x 27.2 x 20.9) (22.8 x 27.2 x 20.9) (20.9 x 25.2 x 22)
POWER REQUIREMENTS,
VAC 115/230 100-120/220-240 100-120/220-240 120/240
OPTIONS None 1-, 6-, or 12- 1-, 6-, or 12- Linear and quadratic
position heated or position heated or quantitative
unheated cell hold- unheated cell hold- analysis, FSQ for
ers, gel scanning, ers, gel scanning, multicomponent
sipper/batch sipper/batch analysis, enzyme
sampler, 5 µL/ sampler, 5 µL/ mechanics/activity
50 µL/100 µL cells, 50 µL/100 µL cells, software, multiwave-
50 and 100 µL 50 and 100 µL length, integrated
sipper cells, 3.5" sipper cells, 3.5" drive
disk drive, linear disk drive, linear
and quadratic and quadratic
quantitative analy- quantitative analy-
sis, FSQ for multi- sis, FSQ for multi-
component analysis, component analysis,
enzyme mechanism and enzyme mechanism and
activity software ** activity software **
OTHER SPECIFICATIONS None specified. Window/mouse/menu Window/mouse/menu Optional integrated
operation; storage operation; storage drive stores user
and protection of and protection of methods and data on
data and methods; data and methods; MS-DOS (Macintosh-
microfocused beam. microfocused beam. readable) diskettes.
Meets requirements Meets requirements
of CSA, FCC, IEC, of CSA, FCC, IEC,
ISO 9000, and UL. ISO 9000, and UL.
12 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
CE MARK (MDD) Not specified Not specified Not specified Not specified
CONFIGURATION Single beam 9-channel filter Single beam Single beam : Auto
photometer
SCAN SPEED, nm/min 500 NA NA NA : Auto
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 13
Healthcare Product Comparison System
POWER REQUIREMENTS,
VAC 110/220 110/220 110/220 110/220
OTHER SPECIFICATIONS Routine control 4 filters included Recorder output is RS232 and recorder
applications; are 405, 450, 492, standard. outputs are
teaching and and 620 nm; can hold standard; integrated
environment control; up to 8 filters. membrane LCD
microprocessor keyboard and
control; resident. display; M330W uses
Windows software and
requires a PC.
14 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
LAMP TYPE(S) Deuterium, quartz Tungsten halogen, Tungsten halogen, Tungsten halogen,
halogen deuterium deuterium deuterium
Life, hr 1,000 3,500 for deuterium 3,500 for deuterium 3,500 for deuterium
WAVELENGTH, nm
Range 190-1,100 325-1,000 : See footnote * 190-1,000 : 190-
200-1,000 1,000 : 200-750
Accuracy ±0.5 ±1 <±1 <1
Reproducibility ±0.25 <0.1 ±0.1 ±0.1
BANDPASS, nm 2 8 4 1.8
MONOCHROMATOR Holographic grating Holographic grating Holographic grating Holographic grating
ZERO DRIFT <0.001 A/hr ±0.001 A/hr ±0.001 A/hr ±0.001 A/hr
Auto compensation Yes Yes Yes Yes
RESPONSE TIME, sec 0.3 Not specified Not specified Not specified
DISPLAY Computer monitor LED Backlit graphic LCD Backlit graphic LCD
with scrolling with scrolling
DATA PROCESSING Multicomponent; Absorbance, Absorbance, Derivatives, absorb-
preprogrammed concentration, concentration, ance, concentration,
methods in unlimited transmittance, auto- transmittance, transmittance,
computer memory zero, automatic absorbance ratio absorbance ratio
peak seek with difference, 30-curve difference, 30-curve
baseline flattening, storage, cell/wave- storage, cell/wave-
bidirectional RS232 length program, length program, band
standard wavelength assays, area, peak seek with
reaction kinetics, baseline flattening,
peak seek with wavelength assays,
baseline flattening, reaction kinetics,
bidirectional RS232, bidirectional RS232,
protein assays spectral processing,
wavelength scanning
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 15
Healthcare Product Comparison System
OPTIONS Cell changer; Cell changer and Automatic cell 1-6 derivatives,
automatic sampling holder for 4 cells, changers, return long cell holders
sip flow-cell micropipette with thermostat,
system, long and system, temperature automatic cell
cylindrical cell cell holder, micro- changers, return
holders up to 100 cell holder, auto- micropipette
mm, cell holders matic batch samplers systems, batch
with thermostat, sampler
time course plot
OTHER SPECIFICATIONS None specified. None specified. Tablet dissolution Tablet dissolution
system with 8-cell system with 8-cell
changer and software changer and software
available. available;
reflectance-
measuring
accessories
available.
16 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
WAVELENGTH, nm
Range See footnote * 190-800 (optional 190-800 (opt 900) : 320-1,000
900) 190-900
Accuracy Not specified ±0.5 ±0.5 ±2
Reproducibility Not specified ±0.1 ±0.1 ±0.3%
BANDPASS, nm 4 2 2 : 0.5, 1, 2, 4 10
MONOCHROMATOR Holographic grating Holographic grating Holographic grating Reflection gratings
ZERO DRIFT ±0.001 A/hr ±0.0001 A/hr <0.0001 A/hr 0.002 A/hr
Auto compensation Yes Yes Yes Not specified
RESPONSE TIME, sec Not specified Not specified Not specified <0.3
DISPLAY Backlit LCD with Backlit LCD with Backlit LCD with LED
scrolling scrolling scrolling
DATA PROCESSING Wavelength scanning, Wavelength scanning, Wavelength scanning, Absorbance,
spectral reprocess, spectral manip- spectral reprocess, transmittance,
water analysis pre- ulation, time course time course plots, concentration,
programmed for 170 plot, superimposed superimposed plots, factor
tests, time course plot, method method and spectral
plots, peak seek, storage, multicell/ storage, multicell/
multiwavelength wavelength measure, wavelength scans and
assays, cell/wave- peak seek deriva- measurement, peak
length programming, tive, autozero, seek, derivatives
method storage reaction kinetics 1-6, autozero,
with plot, histogram reaction kinetics
plot, gel scanning, with plot, turbid
curve fit and stor- sample measurement,
age, formula entry gel scanning **
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 17
Healthcare Product Comparison System
POWER REQUIREMENTS,
VAC 110/250 110/250 110/250 110/220
OTHER SPECIFICATIONS None specified. Tablet dissolution Tablet dissolution None specified.
systems and systems and
software. software.
18 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
CE MARK (MDD) No No No No
CONFIGURATION Single beam Single beam, Single beam, micro- Single beam
PC based processor based
SCAN SPEED, nm/min Manual 200, 600, 900 900 Manual
LAMP TYPE(S) Deuterium, quartz Deuterium, quartz Deuterium, quartz Quartz halogen
halogen halogen halogen
Life, hr 500 500 500 500
WAVELENGTH, nm
Range 190-1,000 200-1,000 200-1,000 340-1,000
ZERO DRIFT ±0.003 A/hr ±0.003 A/hr ±0.003 A/hr ±0.001 A/hr
Auto compensation Not specified Yes Yes Not specified
RESPONSE TIME, sec Not specified Not specified Not specified Not specified
DISPLAY 3.5-digit, 7-segment Color or monochrome 24 x 2 dot-matrix 3.5-digit, 7-segment
LED VGA CRT LCD LCD
DATA PROCESSING Absorbance, trans- Absorbance, trans- Absorbance, trans- Absorbance, trans-
mittance, concentra- mittance, concentra- mittance, concentra- mittance, concentra-
tion tion, ratiometric, tion, scanning, tion
scanning, multi- multiwavelength,
wavelength, overlay, RS232C
1-4 derivative, time
scan
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 19
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE 1/year 1/year 1/year 1/year
H x W x D, cm (in) 14.5 x 55.5 x 41.5 14.5 x 55.5 x 41.5 14.5 x 55.5 x 41.5 13.5 x 41 x 31.5
(5.7 x 22 x 16) (5.7 x 22 x 16) (5.7 x 22 x 16) (5.3 x 16 x 12)
POWER REQUIREMENTS,
VAC 230, 50 Hz 230, 50 Hz 230, 50 Hz 230, 50 Hz
Delivery time, ARO 4-6 weeks 4-6 weeks 4-6 weeks 4-6 weeks
Training Not specified Not specified Not specified Not specified
Year first sold 1992 1992 1993 1995
OTHER SPECIFICATIONS None specified. None specified. None specified. None specified.
20 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
SCAN SPEED, nm/min 15, 30, 62, 125, 5-7,000 5-7,000 10;100;200;400;800;
250, 500, 1,000 1,200;2,400;3,600
LAMP TYPE(S) Tungsten halogen, Tungsten halogen, Tungsten halogen, Deuterium, tungsten
deuterium deuterium deuterium iodide
Life, hr Not specified Not specified Not specified Not specified
WAVELENGTH, nm
Range 190-1,000 190-1,000 : 190-900 190-900 190-1,100
ZERO DRIFT <0.002 A/hr <0.0003 A/hr <0.0003 A/hr ±0.003 A/hr
Auto compensation Yes Yes Yes Not specified
RESPONSE TIME, sec 0.1 0.0125 0.0125 Not specified
DISPLAY LED CRT CRT LCD
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 21
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE Not specified Not required Not required Not specified
H x W x D, cm (in) 74 x 37 x 29 27 x 68 x 64 27 x 68 x 64 50 x 56 x 25
(29.1 x 14.6 x 11.4) (10.7 x 26.8 x 25.4) (10.7 x 26.8 x 25.4) (20 x 22 x 9.9)
POWER REQUIREMENTS,
VAC 100/120/220/240 100/120/220/240 100/120/220/240 Not specified
OPTIONS Micro cell holder, Micro cell holder, Micro cell holder, None specified
water-thermostat- sipper, XYZ auto- sipper, XYZ auto-
able cell holders, sampler, 7 x 7 and sampler, 7 x 7 and
sipper, Scanmaster 6 x 6 dual carou- 6 x 6 dual carou-
software sels, 6 x 1 linear sels, 6 x 1 linear
movement module, movement module,
water-thermostated water-thermostated
cell holders, cell holders,
Peltier effect cell Peltier effect cell
holders, total holders, total
integrating sphere integrating sphere
OTHER SPECIFICATIONS 7-cell carousel. Windows-based soft- Windows-based soft- None specified.
ware. GLP/GMP and ware. GLP/GMP and
EMC compliant. Meets EMC compliant. Meets
requirements of BP, requirements of BP,
EP, USP, and TGA; EP, USP, and TGA;
ISO 9001 certified. ISO 9001 certified.
22 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
CE MARK (MDD) Not specified Not specified Not specified Not specified
CONFIGURATION Double beam Double beam Double beam Double beam with
single monochromator
SCAN SPEED, nm/min 10;100;200;400;800; See footnote ** 40; 100; 200; 400; 10-4,000
1,200;2,400;3,600 1,000; 2,000; 4,000
LAMP TYPE(S) Deuterium, tungsten Deuterium, tungsten Tungsten halogen, Tungsten iodide,
iodide iodide deuterium deuterium
Life, hr Not specified Not specified Not specified Not specified
WAVELENGTH, nm
Range 190-1,100 190-900 190-1,100 190-900
WARM-UP TIME, min Not specified Not specified Not specified Not specified
PATH LENGTH, mm Up to 100 Up to 100 Variable Variable
STRAY LIGHT
220 nm <0.5% <0.015% : <0.0005% <0.04% 0.03%
340 nm <0.5% <0.015% : <0.0005% <0.04% 0.03%
PHOTODETECTOR Silicon photodiode PMT Silicon photodiode PMT
ZERO DRIFT ±0.0003 A/hr ±0.0004 A/hr ±0.001 A/hr ±0.0004 A/hr
Auto compensation Yes Yes Yes Yes
RESPONSE TIME, sec Not specified Not specified Variable, 4-speed Variable, 4-speed
DISPLAY Color LCD, color CRT Color CRT Graphic LCD display CRT
DATA PROCESSING PC software for PC software for Quantitative analy- Automatic baseline
general purpose general purpose sis, calibration correction,
included; absorb- included; absorb- curve, wavelength quantitative
ance, transmittance, ance, transmittance, scan, baseline cor- analysis, time scan,
concentration modes; concentration modes; rection, time scan, fixed wavelength
programs for wave- programs for wave- fixed wavelength, smoothing, peak
length scanning, length scanning, absorbance/%T, scale picking, integra-
multiwavelength, multiwavelength, change, conversion, tion, derivatives,
time ratio, and time ratio, and auto scale, zoom, zoom, trace, scale
photometry; unlimit- photometry; unlimit- trace peak/valley change, area calcu-
ed method storage ed method storage picking, peak ratio lation
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 23
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE Not specified Not specified Not specified Not specified
H x W x D, cm (in) 50 x 22.5 x 56.5 33 x 67 x 62 20.5 x 48.4 x 47.7 25 x 46 x 59.5
(20 x 8.9 x 22) (13 x 26 x 24) (8.1 x 19.1 x 18.8) (9.8 x 18.1 x 23.4)
POWER REQUIREMENTS,
VAC Not specified Not specified 100/115/200/220/ 100/115/200/220/
230/240 230/240
Watts Not specified Not specified 130 130
Warranty Not specified Not specified 1 year, parts and 1 year, parts and
labor labor
Delivery time, ARO Not specified Not specified 30 days 30 days
Training Not specified Not specified Not specified Not specified
Year first sold Not specified Not specified Not specified Not specified
24 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
LAMP TYPE(S) Tungsten iodide, Tungsten iodide, Quartz halogen Tungsten halogen
deuterium deuterium
Life, hr Not specified Not specified Not specified Not specified
WAVELENGTH, nm
Range 190-900 190-2,500 320-920 320-950
WARM-UP TIME, min Not specified Not specified Not specified Not specified
PATH LENGTH, mm Variable Variable Up to 100 Not specified
STRAY LIGHT
220 nm 0.0003% 0.03% NA NA
340 nm 0.0003% 0.03% <0.3% 0.5%
PHOTODETECTOR PMT PbS photocell, PMT Photodiode Not specified
ZERO DRIFT ±0.0004 A/hr ±0.0004 A/hr ±0.002 A/hr 0.001 A/hr
Auto compensation Yes Yes Not specified Not specified
RESPONSE TIME, sec Not specified Variable, 4-speed Not specified Not specified
DISPLAY CRT LCD Custom LCD LCD
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 25
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE Not specified Not specified Not specified Not specified
H x W x D, cm (in) 25 x 46 x 59.5 25 x 46 x 59.5 16 x 36 x 27 18 x 52 x 33
(9.8 x 18.1 x 23.4) (9.8 x 18.1 x 23.4) (6.3 x 14 x 11) (7.1 x 20.5 x 13)
POWER REQUIREMENTS,
VAC 100/115/200/220/ 100/115/200/220/ 100/220 ±10%, 100/230
230/240 230/240 50/60 Hz
Watts 130 180 60 Not specified
Warranty 1 year, parts and 1 year, parts and 2 years, parts and 2 years
labor labor labor *
Delivery time, ARO 30 days 30 days 4-6 weeks 4-6 weeks
Training Not specified Not specified Not specified Not specified
Year first sold Not specified Not specified 1996 Not specified
OTHER SPECIFICATIONS Variety of cell Variety of cell Entry-level specifi- None specified.
holders. holders. cation; small foot-
print; push-button
wavelength selec-
tion; auto wave-
length calibration;
ideal for teaching.
26 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
SCAN SPEED, nm/min 166, 400, 1,200 10; 20; 50; 100;200; 10; 20; 50; 100;200; 10; 20; 50; 100;200;
500; 1,000; 2,000 500; 1,000; 2,000 500; 1,000; 2,000 *
LAMP TYPE(S) Deuterium, tungsten Tungsten halogen; Tungsten halogen; Tungsten halogen;
halogen deuterium deuterium deuterium
Life, hr Not specified 2,000; 500 2,000; 500 2,000; 500
WAVELENGTH, nm
Range 190-950 190-900 190-900 180-900
ZERO DRIFT 0.001 A/hr <0.0007 A/hr <0.0004 A/hr <0.0002 A/hr
Auto compensation Not specified Yes Yes Yes
RESPONSE TIME, sec Not specified Automatic Automatic Automatic
DISPLAY LCD CRT, LCD CRT, LCD CRT, LCD
DATA PROCESSING Absorbance, Parameter and file Parameter and file Parameter and file
transmittance, storage, kinetics, storage, kinetics, storage, kinetics,
concentration, differences, stan- differences, stan- differences, stan-
analog, Centronics dard curves with dard curves with dard curves with
parallel port, recalculation, recalculation, recalculation,
RS232 serial port, zooming, peak and zooming, peak and zooming, peak and
data acquisition valley pick, wave- valley pick, wave- valley pick, wave-
software length programming, length programming, length programming,
ratios, overlay, ratios, overlay, ratios, overlay,
log A, 1-4 deriva- log A, 1-4 deriva- log A, 1-4 deriva-
tives tives, multi- tives, multi-
component, Quick component, Quick
BASIC and Turbo BASIC and Turbo
PASCAL macroprogram PASCAL macroprogram
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 27
Healthcare Product Comparison System
POWER REQUIREMENTS,
VAC 100/230 100/120/220/240, 100/120/220/240, 100/120/220/240,
50/60 Hz 50/60 Hz 50/60 Hz
Watts Not specified 350 350 350
Delivery time, ARO 4-6 weeks Not specified Not specified Not specified
Training Not specified Not specified Not specified Not specified
Year first sold Not specified 1993 1995 1995
OTHER SPECIFICATIONS None specified. Standard thermostat Standard thermostat Standard thermostat
cell holder capable cell holder capable cell holder capable
of microvolume of microvolume of microvolume
samples. Meets samples; runs under samples; runs under
requirements of CSA MS Windows 9x. Meets MS Windows 9x. Meets
and ISO 9001; has requirements of CSA requirements of CSA
CE mark. and ISO 9001; has and ISO 9001; has
CE mark. CE mark.
28 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
WAVELENGTH, nm
Range 200-900 330-999 330-999 330-1,000
ZERO DRIFT Not specified 0.003 A/hr 0.003 A/hr <0.003 A/hr
Auto compensation Yes Not specified Not specified Yes
RESPONSE TIME, sec Not specified Automatic Automatic Not specified
DISPLAY Alphanumeric LED Meter : LED LED : LCD LED
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 29
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE None 1/year 1/year None
H x W x D, cm (in) 16 x 41 x 45 40 x 30 x 14.5 40 x 30 x 14.5 16.5 x 44 x 36
(6.3 x 16 x 18) (16 x 12 x 5.7) (16 x 12 x 5.7) (6.5 x 17.1 x 14) :
16.5 x 39 x 32
(6.5 x 15.4 x 12.6)
WEIGHT, kg (lb) 21 (46) 9 (20) 9 (20) 9.5 (21) :
8 (17.7)
POWER REQUIREMENTS,
VAC 220 ±10% 115/230, 50/60 Hz 115/230, 50/60 Hz 110/220
30 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
WAVELENGTH, nm
Range 330-1,000 190-1,100 190-850 340-750
Accuracy ±1 ±1 ±2 ±2
Reproducibility ±0.2 ±0.2 ±0.2 ±0.2
BANDPASS, nm 8 2 5 5
MONOCHROMATOR Blazed holographic Blazed holographic Holographic grating Holographic grating
grating grating
Linearity ±1%, -0.5 to 4.5 A ±1%, -0.5 to 4.5 A ±1% (0-2 A) ±1% (0-2 A)
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 31
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE None None Not specified Not specified
H x W x D, cm (in) 12.5 x 42 x 35 (4.9 11.8 x 52 x 42 22 x 58 x 38 22 x 58 x 38
x 16.4 x 13.7) : (4.6 x 20.5 x 16.5) (8.7 x 23 x 15) (8.7 x 23 x 15)
12.5 x 46.5 x 35
(4.9 x 18.1 x 13.7)
WEIGHT, kg (lb) 9 (19.9) : 18 (39.7) 13.3 (29) 13.3 (29)
10.5 (23.2)
POWER REQUIREMENTS,
VAC 110/220 110/220 110/220, autorange 110/220, autorange
OTHER SPECIFICATIONS None specified. None specified. Reads 96-well Reads 96-well
microplates; temper- microplates; temper-
ature control; ature control;
robotics compatible; robotics compatible.
PathCheck sensor.
32 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
LAMP TYPE(S) Xenon flash Prealigned deuterium Prealigned deuterium Prealigned deuterium
and tungsten-halogen and tungsten-halogen and tungsten-halogen
Life, hr 1 billion flashes Not specified Not specified Not specified
WAVELENGTH, nm
Range 190-1,000 200-900 190-1,100 190-1,100
WARM-UP TIME, min <20 sec Not specified Not specified Not specified
PATH LENGTH, mm 1-10 Variable Variable Variable
STRAY LIGHT
220 nm NA 0.02% T 0.01% T 0.01% T
340 nm <0.01% T 0.02% T 0.01% T 0.01% T
PHOTODETECTOR Silicon photodiode Photodiode Photodiode Photodiode
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 33
Healthcare Product Comparison System
POWER REQUIREMENTS,
VAC 90-240, autorange 100-250, 50/60 Hz 100-250, 50/60 Hz 100-250, 50/60 Hz
OTHER SPECIFICATIONS Reads cuvettes, test Stand-alone versions Stand-alone versions Stand-alone versions
tubes, and 96-well have preprogrammed have preprogrammed have preprogrammed
microplates; temper- application methods application methods application methods
ature control; built in and ready- built in and ready- built in and ready-
robotics compatible; to-run methods for to-run methods for to-run methods for
PathCheck Sensor. instrument perfor- instrument perfor- instrument perfor-
mance validation and mance validation and mance validation and
automatic self- automatic self- automatic self-
check. * check. * check. *
34 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
CE MARK (MDD) Not specified Not specified Not specified Not specified
CONFIGURATION Double beam Single beam Single beam Single beam
LAMP TYPE(S) Tungsten, deuterium Tungsten, deuterium Tungsten halogen Tungsten halogen,
deuterium
Life, hr Not specified Not specified Not specified Not specified
WAVELENGTH, nm
Range 185-3,300 200-900 325-900 200-900
WARM-UP TIME, min Not specified Not specified Not specified Not specified
PATH LENGTH, mm Variable Variable Up to 10 Up to 50
STRAY LIGHT
220 nm <0.00008% <0.03%T NA <0.2% T
340 nm <0.00008% <0.03%T <0.5% T <0.2% T
PHOTODETECTOR PMT, UV/Vis, PbS,NIR Photodiode Silicon photodiode Silicon photodiode
DATA PROCESSING PC controller via Extensive stand- Absorbance ratio, Absorbance concen-
UV Winlab, Windows- alone absorbance tration, time
based PC software difference, intervals, scan to
kinetics, water chart recorder, %T,
analysis, scan to standard curve
chart recorder
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 35
Healthcare Product Comparison System
POWER REQUIREMENTS,
VAC 90-250 115/230 100-120/200-240 90-265
±10%
Watts Not specified Not specified 50 VA 100 VA
Delivery time, ARO Not specified Not specified Not specified Not specified
Training Not specified Not specified Not specified Not specified
Year first sold Not specified Not specified 1991 1997
36 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
LAMP TYPE(S) Tungsten halogen, Tungsten halogen, Tungsten halogen, Deuterium, quartz
deuterium deuterium deuterium halogen
Life, hr Not specified Not specified Not specified 1,000
WAVELENGTH, nm
Range 190-1,100 190-1,100 190-1,100 195-1,000
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 37
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE Not specified Not specified Not specified 1/year
H x W x D, cm (in) 19 x 50 x 36 19 x 50 x 36 19 x 50 x 36 20 x 56 x 52
(7.5 x 19.7 x 14.2) (7.5 x 19.7 x 14.2) (7.5 x 19.7 x 14.2) (7.9 x 22 x 20.5)
POWER REQUIREMENTS,
VAC 90-265 100-240 ±10% 90-265 110/220
OTHER SPECIFICATIONS 6-position automatic 6-position automatic 8-position automatic Dedicated to water
cell changer, cell changer is cell changer is control, teaching,
recorder output, standard; built-in standard. Meets re- routine control
RS232C, stored applications soft- quirements of applications; auto
methods for RNA/DNA ware. Meets require- ISO 9001. 9-cell holder; PC or
quantification are ments of ISO 9001. membrane keyboard;
all standard. Meets microprocessor-
requirements of controlled; resident
ISO 9001; CSA programs. ISO 9001
pending. certified.
38 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
WHERE MARKETED Africa, Asia, Eu- Africa, Asia, Eu- Africa, Asia, Eu- Africa, Asia, Eu-
rope, North America rope, North America rope, North America rope, North America
CE MARK (MDD) Not specified Not specified Not specified Not specified
CONFIGURATION Single beam Single beam Single beam Single beam
SCAN SPEED, nm/min 0.5, all spectra 0.5, all spectra Not specified 500
WAVELENGTH, nm
Range 400-700 205-320 334-900 330-900
Accuracy ±2 2 ±1 ±1
Reproducibility ±0.5 0.5 ±0.5 ±0.2
BANDPASS, nm ±6 Not specified ±5 ±5
MONOCHROMATOR Static grating Static mounting Original mounting Original mounting
spectrograph spectrograph grating grating
Noise at 0 A <0.001 A @ 450 nm, Not specified <0.001 A, 450 nm, <0.5 A @ 340 nm
<0.002 A @ 2A, 450nm <0.002 A @ 2A, 450nm
Linearity <2% @ 0-2 A Not specified <2% @ 0-2 A ±1% @ 0-2 A
ZERO DRIFT <0.004 A/hr Not specified <0.004 A/hr <0.004 A/hr
Auto compensation Yes Not specified Yes Yes
RESPONSE TIME, sec 1 1 Not specified Not specified
DISPLAY LCD LCD LED, video LED
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 39
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE None 1/year 1/year 1/year
H x W x D, cm (in) 5 x 12 x 30 44 x 37 x 20 41 x 31 x 30 40 x 14 x 31
(2 x 4.7 x 11.8) (17 x 15 x 8) (16.1 x 12.2 x 11.8) (15.7 x 5.5 x 12.2)
POWER REQUIREMENTS,
VAC 110/220, 9 VDC 110/220 110/220, 12 VDC 110/220
Watts 25 VA 50 VA 25 VA 25 VA
OTHER SPECIFICATIONS Dedicated to routine Dedicated to water Dedicated for For applications in
applications in testing/pollution routine applications diagnostics, bio-
diagnostics, water monitoring; clean in industry, technology, water
control, education; analysis; no reagent diagnostics, control, industry,
microprocessor- needed. ISO 9001 environment, or others; micropro-
controlled programs; certified. education; micro- cessor-controlled;
autowavelength sel- processor-controlled autozero; autowave-
ector; portable; PC programs; portable; length selector;
software; membrane autowavelength 24/50 programmable
keyboard. ISO 9001 selector; autozero. methodologies. ISO
certified. ISO 9001 certified. ** 9001 certified. **
40 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
WARM-UP TIME, min 10 Immediate for survey ** Immediate for survey ** Immediate Vis, 20 UV
PATH LENGTH, mm Up to 50 Up to 100 Up to 100 1-100
STRAY LIGHT
220 nm <0.3% <0.05% T NA <0.05% T
340 nm <0.1% <0.05% T >0.05% T <0.05% T
PHOTODETECTOR Silicon photocell Silicon photodiode Silicon photodiode Silicon photodiode
ZERO DRIFT <0.004 A/hr ±0.001 A/hr ±0.001 A/hr <0.001 A/hr
Auto compensation Yes Turnkey operation Not specified Yes
RESPONSE TIME, sec Not specified 0.1-0.4 0.1-0.4 Not specified
DISPLAY LED Backlit LCD Backlit LCD Backlit LCD/color
monitor
DATA PROCESSING Absorbance, Photometric mode All program packs Windows environment,
transmittance, with interactive LCD for UV-1201 except zooming, peak pick,
concentration, screen commands; DNA analysis point pick (to 10),
kinetics, end point, optional single- smoothing, arith-
nonlinear wavelength metic between files
standards, scanning, quantitation, time or factor, log, 1-4
multiwavelength, course, kinetics derivatives,
multicomponent, 1-4 differentiation,
derivatives, peak automatic wavelength
maximum, and photometric
turbidimetry, 36 value at cursor,
preprogrammed scale expansion,
methodologies in repeat scan and
residence, RS232C radar ***
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 41
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE 1/year 1/year None 1/year
H x W x D, cm (in) 40 x 14 x 43 18 x 43 x 30 18 x 43 x 30 20 x 55 x 47
(15.7 x 5.5 x 16.9) (7.1 x 16.9 x 11.8) (7.1 x 16.9 x 11.8) (7.9 x 21.7 x 18.5)
POWER REQUIREMENTS,
VAC 110/220 100/115/220/240 110/220 100/115/220/240
(switch-selectable) (switch-selectable) (switch-selectable)
Watts 75 VA 100 VA 100 VA 160 VA
OPTIONS Printer, thermostat Sipper, automatic Sipper, automatic Ambient and thermo-
cell holder, rotary sample changer, sample changer, electric 6-cell
cell holder, auto- micro flow-through micro flow-through positioner, sipper,
matic suction, auto- cells, thermo- cells, thermo- super micro cells,
matic sampler and electric cell electric cell multi micro cells
diluter, data positioner, analog positioner, analog (8/16 in), program
acquisition soft- recorder, data pack, recorder, data pack, packs, Windows
ware, analog output, 18 program packs 16 program packs software, integra-
software for PCs ting sphere
OTHER SPECIFICATIONS Routine control Includes 6-cell 1-cell holder Video display,
applications, holder; LCD has standard. copier/printer, and
teaching, water adjustable contrast all software are
control; micropro- and viewing angle; standard; the
cessor control; RS232 digital BioSpec includes
resident programs; output. software for DNA
autozero; autowave- quantitation and
length selector. protein analysis.
ISO 9001 certified.
42 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
LAMP TYPE(S) Tungsten, deuterium Tungsten, deuterium Tungsten, deuterium Tungsten halogen
WAVELENGTH, nm
Range 190-900 190-900, extendable 190-3,200 325-1,100
to 1,100
Accuracy ±0.3 ±0.3 ±0.3 ±2
Reproducibility ±0.1 ±0.1 ±0.1 0.5
BANDPASS, nm 0.1-5 0.1-5 0.1-30 8
MONOCHROMATOR High-performance See footnote * 2 sets of triple Holographic grating
holographic grating holographic gratings
ZERO DRIFT <0.0004 A/hr <0.0004 A/hr <0.0004 A/hr 0.003 A/hr
Auto compensation Yes Yes Yes Yes
RESPONSE TIME, sec 0.1 0.1 0.1 Not specified
DISPLAY Color monitor Color monitor Color monitor 2-line, 20-character
LCD
DATA PROCESSING Windows environment, Windows environment, Windows environment, Absorbance, trans-
zooming, peak pick, zooming, peak pick, zooming, peak pick, mittance, concentra-
point pick (to 10), point pick (to 10), point pick (to 10), tion, standard
smoothing, arith- smoothing, arith- smoothing, arith- RS232, bidirectional
metic calc between metic calc between metic calc between
files or factor, files or factor, 1-4 files or factor, 1-4
log, 1-4 deriv- derivatives, derivatives,
atives, different- differentiation, differentiation,
iation, automatic automatic wavelength automatic wavelength
wavelength and and photometric and photometric
photometric value at value at cursor, value at cursor,
cursor, scale scale expansion, scale expansion,
expansion, repeat repeat scan and repeat scan and
scan and radar ** radar ** radar **
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 43
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE 1/year 1/year 1/year Not specified
H x W x D, cm (in) 27.5 x 57 x 66 27.5 x 57 x 66 25 x 102 x 69 30 x 33 x 19
(11 x 22 x 26) (11 x 22 x 26) (9.8 x 40.2 x 27.2) (12 x 13.2 x 7.6)
WEIGHT, kg (lb) Not specified Not specified Not specified 4.5 (10)
POWER REQUIREMENTS,
VAC 100/117/200/240 100/120/220/240 100/117/200/240 115, 50/60 Hz;
230, 50/60 Hz
Watts 400 VA 250 VA 400 VA Not specified
OTHER SPECIFICATIONS Includes computer, Includes computer, Includes computer, Languages: English,
software, MS Windows software, MS Windows software, MS Windows Spanish, German,
package. package. package, photo- Portuguese, Czech,
metrics, kinetics, French; total micro-
color analysis, film processor control;
thickness. advanced utility
functions; user
prompts. Meets re-
quirements of CSA.
44 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
WAVELENGTH, nm
Range 340-950 200-1,000 325-900 400-900
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 45
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE Not specified Not specified Not specified Not specified
H x W x D, cm (in) 20 x 41 x 33 22 x 51 x 25 25 x 41 x 36 21 x 41.3 x 33
(7.9 x 16.1 x 13) (8.7 x 20.1 x 9.8) (9.8 x 16.1 x 14.2) (8.25 x 16.25 x 13)
POWER REQUIREMENTS,
VAC 110-115/220-240 110-115/220-240 115/220 115, 50/60 Hz;
220, 50 Hz
Watts 90 VA 45 : 60 115 Not specified
OTHER SPECIFICATIONS Analog output (20), 3-position Built-in RS232C Designed for high-
test tube adapters, sensitivity switch; interface; movable school teaching
20D has RS232C universal test tube keyboard; lamp- labs. Meets require-
interface. Meets holder. Meets saver feature. Meets ments of CSA.
requirements of CSA. requirements of CSA. requirements of CSA.
46 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
WAVELENGTH, nm
Range 200-1,100 200-1,100 200-1,100 330-999
Accuracy ±1 ±1 ±1 ±2
Reproducibility ±0.5 ±0.5 ±0.5 ±0.5
BANDPASS, nm 2 2 5 8
MONOCHROMATOR Holographic grating Holographic grating Holographic grating Ebert diffraction
grating
Noise at 0 A 0.001 A @ 220 and 0.001 A @ 220 and 0.001 A @ 220 and Not specified
340 nm 340 nm 340 nm
Linearity Not specified Not specified Not specified 1%
ZERO DRIFT ±0.002 A/hr ±0.002 A/hr ±0.002 A/hr ±0.003 A/hr
Auto compensation Yes Yes Yes Yes
RESPONSE TIME, sec Not specified Not specified Not specified 5
DISPLAY Graphic, backlit Computer CRT, Graphic, backlit LCD 3.5-digit LED
color LCD PC-driven unit w/adjust contrast
DATA PROCESSING RS232C interface, Absorbance, trans- RS232C interface, Absorbance,
Centronics port, mittance, concentra- Centronics- % transmittance
advanced scanning, tion, scanning, compatible printer
optional SoftCards: kinetics, absorbance port, memory Soft-
biochemistry, Soft- ratio, absorbance Cards (accessory)
Drink, applications difference, multiple for storage of test
II SoftCard, wavelengths, deriva- information, appli-
standard curves, tions, area-under- cations SoftCards
kinetics, absorbance curve, 3-point net (accessory) for sur-
ratio, absorbance vey scan kinetics,
difference, perform- absorbance ratio,
ance validation and linear/segmented
standard curve pro-
grams
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 47
Healthcare Product Comparison System
PREVENTIVE
MAINTENANCE Not specified Not specified Not specified 1/year
H x W x D, cm (in) 21 x 47.9 x 55 21 x 47.9 x 55 21 x 47.9 x 55 14.5 x 40 x 30
(8.3 x 18.9 x 22.3) (8.3 x 18.9 x 22.3) (8.3 x 18.9 x 22.3) (5.7 x 15.7 x 11.8)
POWER REQUIREMENTS,
VAC 115, 50/60 Hz; 115, 50/60 Hz; 115, 50/60 Hz; 120, 230; 50/60 Hz
230, 50/60 Hz 230, 50/60 Hz 230, 50/60 Hz
Watts Not specified Not specified Not specified 63
OPTIONS Sipper and tempera- Sipper and tempera- Internal printer, Micro UV quartz
ture control, ture control, adjustable cell cuvettes
Peltier cell holder, Peltier cell holder, holder platform,
long-path cell long-path cell single cell holder,
holders, short-path holders, short-path test tube holder,
adapters, glass adapters, glass long-path cell
filter and glass filter and glass holders, adjustable
sample holders, sample holders, filter holders,
10 µL cells 10 µL cells 10 µL cells
OTHER SPECIFICATIONS Pictogram status PC-driven Windows 95 Pictogram status of Auotzero; digital
indicators; standard operating system; lamps, accessories, dial wavelength
automated 8-cell Winspec application SoftCard; alpha- selection; error
holder; 2 SoftCard software. numeric sample message codes; 0-1
slots. identification; VDC analog output
8-position multi- for chart recorder.
cell holder; 2
SoftCard slots;
reference manual.
48 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
Noise at 0 A Not specified Not specified <0.0001 @ 500 nm, 0.000085 A @ 500 nm
<0.00015 @ 260 nm
Linearity 1% 1% Not specified Not specified
ZERO DRIFT ±0.003 A/hr ±0.003 A/hr <0.0004 A/hr 0.0003 A/hr
Auto compensation Yes Yes Not specified Yes
RESPONSE TIME, sec 5 5 0.0125-999 0.033-999
DISPLAY 4.5-digit LED 20 x 2 LCD CRT CRT
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 49
Healthcare Product Comparison System
POWER REQUIREMENTS,
VAC 120, 230; 50/60 Hz 120, 230; 50/60 Hz Powered from 100/120/220/240
connecting PC
Watts 63 63 NA Not specified
OTHER SPECIFICATIONS Keypad for data Automatic stray Full 32-bit software Full 32-bit software
input; decimal key; light filter system running under system running under
autozero; digital selection; keypad Windows 95; fully Windows 95/NT; fully
dial wavelength for data input; GLP compliant with GLP compliant with
selection; error autozero; push- password-protected password-protected
message codes; 0-1 button wavelength file security file security. Meets
VDC analog output selection; error system. Meets requirements of BP,
for chart recorder. message codes; 0-1 requirements of BP, EP, ISO 9001, TGA,
VDC analog output EP, ISO 9001, TGA, and USP.
for chart recorder. and USP.
50 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
CARY 300 Bio : CARY 400 Bio : CARY 500 Scan SK110
300 Conc : 300 Scan 400 Conc : 400 Scan Spectrophotometer
WHERE MARKETED Worldwide Worldwide Worldwide Worldwide
WAVELENGTH, nm
Range 190-900 175-900 175-3,300 320-900
ZERO DRIFT 0.0003 A/hr 0.0002 A/hr 0.0002 A/hr Not specified
Auto compensation Yes Yes Yes Not specified
RESPONSE TIME, sec 0.033-999 0.033-999 0.033-999 3
DISPLAY CRT CRT CRT 4-digit LED
DATA PROCESSING Complete with Complete with Complete with Absorbance, kinetics
applications applications applications via S110K module,
development language development language development language 0-2 V x 0-2 A,
(ADL) for user (ADL) for user (ADL) for user 2 x 4 mm sockets
programming and programming and programming and
applications applications applications
analyzer; kinetics, analyzer; kinetics, analyzer; kinetics,
concentration, concentration, concentration,
thermal melt thermal melt thermal melt
samples samples samples
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 51
Healthcare Product Comparison System
CARY 300 Bio : CARY 400 Bio : CARY 500 Scan SK110
300 Conc : 300 Scan 400 Conc : 400 Scan Spectrophotometer
PRINTER Optional Optional Optional Optional
PREVENTIVE
MAINTENANCE Not specified Not specified Not specified 1/year
H x W x D, cm (in) 64 x 33 x 66 99 x 33 x 66 99 x 33 x 66 12 x 37 x 30
(25.2 x 13 x 26) (39 x 13 x 26) (39 x 13 x 26) (4.7 x 15 x 12)
POWER REQUIREMENTS,
VAC 100/120/220/240 110/120/220/240 100/120/220/240 220/240, 110
OTHER SPECIFICATIONS Full 32-bit software Full 32-bit software Full 32-bit software Applications for
system running under system running under system running under serum biochemistry,
Windows 95/NT; fully Windows 95/NT; fully Windows 95/NT; fully routine QC applica-
GLP compliant with GLP compliant with GLP compliant with tions, environmental
password-protected password-protected password-protected monitoring,
file security. Meets file security. Meets file security. Meets teaching.
requirements of BP, requirements of BP, requirements of BP,
EP, ISO 9001, TGA, EP, ISO 9001, TGA, EP, ISO 9001, TGA,
and USP. and USP. and USP.
52 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.
Spectrophotometers, UV/Visible
UV1101 Filter
Photometer
WHERE MARKETED Worldwide
WAVELENGTH, nm
Range 220-700, filters
Accuracy ±2
Reproducibility <±0.1
BANDPASS, nm 10
MONOCHROMATOR Interference filters
©1998 ECRI. Duplication of this page by any means for any purpose is prohibited. 53
Healthcare Product Comparison System
UV1101 Filter
Photometer
PRINTER Optional
PREVENTIVE
MAINTENANCE 1/year
H x W x D, cm (in) 10 x 25 x 25
(3.9 x 9.9 x 9.9)
POWER REQUIREMENTS,
VAC 220/240, 110
Warranty 2 years
54 ©1998 ECRI. Duplication of this page by any means for any purpose is prohibited.