Asb RG Method-Validation v2 2021-05-24
Asb RG Method-Validation v2 2021-05-24
Permission to Reproduce
Except as otherwise specifically noted, the information in this publication may be reproduced, in
part or in whole and by any means, without charge or further permission from the Standards
Council of Canada, provided that due diligence is exercised in ensuring the accuracy of the
information reproduced; that the Standards Council of Canada is identified as the source
institution; and that the reproduction is not represented as an official version of the information
reproduced, nor as having been made in affiliation with, or with the endorsement of, the
Standards Council of Canada.
For permission to reproduce the information in this publication for commercial purposes, please
contact [email protected].
© 2021, Standards Council of Canada
Aussi offert en français sous le titre Exigences et lignes directrices du CCN – Validation des
méthodes dans les laboratoires d’essais
This document sets out the requirements for, and gives guidance to, applicant and accredited
laboratories about the interpretation of the validation requirements for all types of test methods
as described in clause 7.2 “Selection, verification and validation of methods” of ISO/IEC
17025:2017.
1. Scope
These general guidelines shall apply to all types of laboratory activities including sampling,
regardless of field for the interpretation of the validation and verification requirements for testing
methods and additional requirements applicable to most common types of testing. Additional
program or sector-specific requirements for validation as outlined in specific Program Specialty
Area (PSA) documents will also apply.
2. Normative References
• ISO/IEC 17025:2017 General requirements for the competence of testing and calibration
laboratories.
• ISO/IEC Guide 99:2007 International Vocabulary of metrology- Basic and general concepts
and associate terms (VIM)
• International vocabulary of metrology – Basic and general concepts and associated terms
(VIM), JCGM 200:2012, GUM 2008 with minor corrections
• SCC Requirements and Guidance for the Accreditation of Testing Laboratories.
• SCC Requirements and Guidance for Accreditation of Laboratories Engaged in Test Method
Development and Non-Routine Testing
• SCC Requirements and Guidance for the Accreditation for Forensic Testing Laboratories.
• SCC Guidelines for the Presentation of Laboratory Scopes of Accreditation.
• Eurachem/CITAC Guide, The Fitness for Purpose of Analytical Methods. A Laboratory
Guide to Method Validation and Related Topics. 2nd edition, 2014.
• Eurachem/CITAC Guide, Quality Assurance for Research and Development and Non-
routine Analysis, 1st edition, 1998.
• Eurachem/CITAC Guide CG4, Quantifying uncertainty in analytical measurement, 3rd
edition, 2012.
3. Informative References
• Supplement to Eurachem Guide on the Fitness for Purpose of Analytical Methods, Planning
and Reporting Method Validation Studies, Planning and Reporting Method Validation
Studies –, 1st edition,2019).
• Eurachem/CITAC Guide, Setting and Using Target Uncertainty in Chemical Measurement,
1st edition,2015).
• ILAC G19:08/2014 Modules in a Forensic Science Process.
• AOAC INTERNATIONAL Methods Committee Guidelines for Validation of Qualitative and
Quantitative Food Official Methods of Analysis. J AOAC Int’l 85(5) 2002, p.1187-1200.
• Protocol for the Design, Conduct and Interpretation of Collaborative Studies. Pure Appl.
Chem, 60(6), 1988, p 855-864.
4. Definitions
Method: similar to “measurement procedure” defined in Guide 99.
Measurement procedure: a set of operations, described specifically, used in the performance
of particular measurements according to a given method. It is usually recorded in a document
that sometimes itself called a “measurement procedure” (or a measurement method) and is
usually insufficient detail to enable an operator to carry out a measurement without additional
information. (Guide 99)
Measurement procedure (JCGM 200, item 2.6): detailed description of a measurement
according to one or more measurement principles and to a given measurement method, based
on a measurement model and including any calculation to obtain a measurement result
NOTE 1: A measurement procedure is usually documented in sufficient detail to enable an
operator to perform a measurement.
NOTE 2: A measurement procedure can include a statement concerning a target measurement
uncertainty.
Verification (referred to as Method Verification): The process of ensuring that the laboratory
has the ability to perform a previously validated method consistent with the requirements of the
method and the needs of the laboratory.
Validation (referred to as Method Validation): The process of ensuring that a method has the
capabilities and performance characteristics consistent with specified requirements and that it’s
fit for the intended use and designed application.
Fitness for Purpose: (IUPAC Harmonized Guidelines for Single Laboratory Validations of
Methods of analysis.): It is the extent to which the performance of a method matches the
criteria, agreed between the analyst and the end-user of the data, that describe the end-user’s
needs. Fitness-for-purpose criteria will ultimately be expressed in terms of acceptable combined
uncertainty.
Modified method: standard and non-standard methods used outside their intended scope or
otherwise modified.
In-house method: A method in-house developed by the laboratory for its own use.
(e.g. laboratory-developed method)
Validation does not apply to standard published methods in terms of the requirement of ISO/IEC
17025 clause 7.2.2. Verification does apply to standard published methods as described in
clause 7.2.1. of ISO/IEC 17025.
The laboratory shall verify that it can properly perform methods before introducing them by
ensuring that it can achieve the documented performance characteristics according to clause
7.2.1.5 of ISO/IEC 17025.
When standard published methods are adopted by the laboratory without significant deviations,
verification is required under following examples but may not be limited to:
• Confirmation that the implementation of the standard published method is fit for the
intended purpose (e.g. precision, bias, linearity, interference, robustness, grounding,
linearity, electrical connections, physical stability are consistent with the ones published
in the reference method);
• Estimation of key performance parameters such as method detection limit (MDL),
reporting limit (RL) or signal-to-noise ratio for representative matrices, or types of
samples under test (Annex A for examples of performance parameters).
• Evaluation of measurement uncertainty;
SCC Requirements and Guidance for Method Verification
Page 7 of 14 and Validation of Testing Labatories
• Confirmation that the method works reliably and accurately with the typical samples
processed (e.g. acceptable spiked sample recovery, or comparison against an
independent reference or other validated test method);
• Confirmation that any decision rule associated with the method is valid for the
laboratory's implementation.
If guidelines within a particular PSA specify verification requirements, these shall also be
followed.
7. Validation Process
7.1. Extent of Validation
Validation shall apply for the following categories of method according to clause 7.2.2.1 of
ISO/IEC 17025:
• non-standard methods (in-house);
• laboratory-designed/developed methods (in-house);
• standard methods used outside their intended scope (modified);
• significant modifications of standard methods that could impact the reliability of results
(modified).
The stated purpose of the validation is to confirm that the methods are fit for the intended use.
In addition, clause 7.2.2. states that:
“The validation shall be as extensive as is necessary to meet the needs of the given application
or field of application”.
Changes in implementation of a previously Verification will suffice. The extent will vary in
validated method (e.g. changes to order to demonstrate changes do not have a
equipment, reagents, lab environment or significant impact on performance
staff) characteristics.
In some disciplines, the guidelines available to the laboratory testing community are extensive.
It is therefore the responsibility of the laboratory, with input from customers, to seek out the
relevant characteristics to be evaluated with respect to the laboratory’s specific situation and the
customer’s needs. The laboratory must have a documented validation plan, either to be used
generally or applied to a specific project or customer. Test method performance characteristics
to be evaluated will vary with the type of test and its intended use. Discipline-specific or
customer required performance criteria are to be applied to demonstrate fitness for purpose.
For detailed discussions of uncertainty evaluations in validation studies see the following:
• Eurachem / CITAC Guide, Quantifying Uncertainty in Analytical Measurement
• IUPAC’s Harmonized Guidelines for Single-Laboratory Validation of Methods of Analysis
(Appendices A14 and B)
• Evaluation of measurement data – guide to the expression of uncertainty in measurement,
JCGM 100:2008, GUM 1995 with minor corrections.
For quantitative and semi-quantitative method, an estimate of uncertainty is required, while for
qualitative methods an identification of sources of uncertainty and under which conditions they
are controlled is necessary.
8. Documentation
The laboratory must have available for review indexed records, summarizing the detailed
method validation / verification data for all non-standard, in-house developed or modifications
and amplifications of standard published methods. These records shall include:
• The test method including information about equipment, reagents, calibration etc.
• Reference to the validation procedure or plan used to generate method performance
characteristics.
• Test method performance characteristics and how these were calculated or defined.
The raw data should be available for review.
• Test method performance criteria against which the characteristics were evaluated and
whether or not the method is fit for purpose.
• Review and authorization of the report by a competent authorized person.
• The intended use of the method including relevant matrices.
• Estimates of uncertainty.
If a method that is not a standard published method is used routinely, often over time there will
be modifications or improvements made. This information needs to be documented and
available for assessment. Ongoing proficiency testing data and quality control data should be
reviewed by the laboratory to confirm the fitness of the method. The validation/verification
information shall be kept according to the timeline established by the laboratory under the
requirements of clause 8.4.2 of ISO 17025.
the intended use of the test method including customer needs: “the extent of validation, and
the consequences in time and cost, are one of the key issues to be agreed upon between
analyst and customer when commissioning method development.” (Eurachem/ CITAC
Guide - Quality Assurance for Research and Development and Non-routine Analysis,
section 6.8.5.4)
the discipline: some characteristics, including the definitions, are discipline or test type
specific.
the type of test (qualitative, quantitative) and purpose (screening, confirmatory).
Characteristics or criteria to be considered may include, but are not limited to:
• Accuracy
• Action levels (for regulatory analyses)
• Analytical specificity* (selectivity, exclusivity, inclusivity)
• Calibration (see linearity, range)
• Confirmation
• Diagnostic selectivity*, Diagnostic sensitivity*
• Efficacy
• False positive rate, False negative rate (paired studies)
• Ion Suppression for LC/MS or LC/MSMS assay
• Limit of detection (detection capability, decision limit, level of detection)
• Limit of determination (limit of quantitation)
• Linearity (calibration, range)
• Measurement bias
• Measurement uncertainty
• Precision
• Predictive value
• Probability of detection (unpaired studies)
• Range (see calibration, linearity)
• Recovery
• Relative Sensitivity, Relative Specificity
• Repeatability
• Reporting limit
• Reproducibility
• Ruggedness/robustness
• Selectivity*, Sensitivity*, Specificity*
SCC Requirements and Guidance for Method Verification
Page 12 of 14 and Validation of Testing Labatories
• Stability (of analyte, organism, sub-samples, extracts)
• Target organisms
• Test controls (including reference populations)
• Thresholds
* Definitions for some terms may vary among disciplines. For example, the term, “selectivity,” is
recommended for analytical chemistry instead of “specificity” (see IUPAC definition). Some
definitions such as analytical specificity, diagnostic selectivity and diagnostic sensitivity are used
in biological testing.
The AOAC food matrix triangle is a common tool used to assist laboratories in selecting
matrices to be included in their studies. The AOAC food triangle is constructed based on the
relative levels of fat, protein and carbohydrate divided into nine sectors. Consequently,
validation is required to be performed over a wide variety of foods since subsequent steps of
preparation, extraction and testing may be different for a sample containing 100% carbohydrate
compared to a sample that is high in lipids. While planning method validation studies, samples
that are reflective of the types of samples received by the laboratory should be considered
including samples that might be a challenge to analyze (e.g. high fat, high protein, high
moisture). Analysts’ professional judgement will be required when validating a certain segment
of the food triangle; other factors should be considered such as extraction technique, solids vs.
liquid ratio and matrix interferences.
SCC Requirements and Guidance for Method Verification
Page 14 of 14 and Validation of Testing Labatories