This document details the minimum hardware and software requirements that
need to be met to run OpenLab CDS. It is valid for Workstation, Workstation Plus,
Client, or Analytical Instrument Controller (AIC) components. It also provides
information on supported Agilent and Non-Agilent instruments.
For server components please refer to the OpenLab Server and OpenLab ECM XT
Hardware and Software Requirements Guide
(ECM_XT_v2.7_HardwareSoftwareRequirements_en.pdf), or the Requirement
guide for ECM, respectively.
Table 1 Terms and abbreviations used in this document
Term Description
Shared Services Set of administrative services that control, for example, the security
policy and the central configuration of OpenLab CDS. Shared services
are accessed via the Control Panel.
1 Hardware Requirements
This chapter contains the hardware requirements for the different components of
an OpenLab CDS system.
2 Software Requirements
This chapter contains the software requirements for the different components of
an OpenLab CDS system.
3 Network Specifications
This chapter provides network specifications for an OpenLab CDS system.
5 Instrument Information
This chapter provides information on the instruments supported by the current
revision of OpenLab CDS and the required respective instrument drivers and
firmware revisions.
6 Software Compatibility
This chapter contains information on compatibility with other Agilent or
non-Agilent software.
1 Hardware Requirements 7
OpenLab CDS Topologies 8
OpenLab CDS Client/Server System Topologies 10
PC Recommendations 15
Configuration Capacity 18
Disk Space 21
OpenLab Server / OpenLab ECM XT configurations 22
2 Software Requirements 23
General Software Requirements 24
Supported Operating Systems 25
Supported Databases 28
Virtualization 29
Licensing 33
3 Network Specifications 34
Introduction 35
Network Specifications 36
About LAN Communications 39
Power Management 39
Specific Requirements for Compliant Systems 40
Firewall Settings 41
5 Instrument Information 61
Instrument Drivers 62
6 Software Compatibility 91
OpenLab CDS System Compatibility 92
Supported Content Management configurations 93
Compatible Libraries and Databases 95
Supported Agilent Software Add-Ons 96
This chapter contains the hardware requirements for the different components
of an OpenLab CDS system.
OpenLab CDS is available in different topologies. They differ on where the various
software components (Clients, Shared Services and, if relevant, Content
Management) are located. For details on the topologies refer to the respective
installation guide or check with your Agilent representative.
Workstation
Figure 1 All components on the same PC
Networked Workstation
Networked Workstation is defined as a PC machine that supports both user
interaction (i.e. sample submission and data review and processing) as well as
automated functions (i.e. data acquisition and automated processing and
printing). For OpenLab CDS, a Networked Workstation is an AIC used interactively
for sample submission and data processing. The Client components are installed
as part of the AIC installation.
Based on your expected system load, that depends on the number and type of
instruments and users, OpenLab CDS Client/Server systems can be deployed
with different topologies:
1-server all-in-one system (see Figure 2 on page 11),
as a Scalable System.
For more information on scalable systems, please refer to the Agilent
OpenLab Server and OpenLab ECM XT Scalable System Installation Guide.
AIC
Web/Repository
Database
Clients Server Index
File Storage
AIC
Clients
Web/Repository
Server Index Database
File Storage Server
Figure 3 2-Server topology: Clients and AIC link to two dedicated servers.
AIC
Private cloud providers AWS EC2, AWS S3 or AWS RDS Azure SQL
OpenLab or Azure VM1 Azure File (managed)
components storage2
Workstations - - - -
Database Server c
Legend
supported AWS: Amazon Web Services
Cloud deployments with OpenLab ECM 3.x as back-end are not supported under
a standard support agreement.
PC Recommendations
The following tables are intended to be guidelines for minimum hardware and
NOTE may need to be adjusted based on your intended load. Consult with your Agilent
support representative to decide which hardware and topology is appropriate for
your needs. Factors to consider include number of logical instruments,
concurrent users, and other connection points. See also calculations for “Disk
Space” on page 21 and “Configuration Capacity” on page 18.
Client
Table 3 Minimum hardware configuration for a client
Item HW requirements
Workstation
OpenLab CDS is available in two workstation flavors: With storage in the local file
system (OpenLab CDS Workstation software), or with a built-in Content
Management database (OpenLab CDS Workstation Plus software).
Based on the number of instruments OpenLab CDS Workstation software has
two recommended PC configurations: A small (default) with 8GB to control up to
2 instruments, and with upgraded memory (16GB) that can be used to control up
to 4 instruments. The table below provides the minimum hardware configuration
for Workstations. It also applies to OpenLab CDS VL (Note: VL only supports 1
instrument).
Table 4 Minimum hardware configuration for workstations
Hard disk 1 x 500 GB 7200 RPM SATA drive 2 x 500 GB or 1 TB 7200 RPM
minimum SATA drive minimum1
SSD drive recommended for better SSD drive recommended for better
performance performance
See Table 12 on page 21 See Table 12 on page 21
RS-232 port 1 serial port required for selected instruments that are still using RS-232
communication. See instrument specifications for details.
Win 11 Microsoft has specific chipset requirements for Win 11. Refer to Microsoft
for details
Item HW requirements
RS-232 port 1 serial port required for selected instruments that are still using RS-223
communication. See instrument specifications for details.
Configuration Capacity
Instrument An analytical system, made up of one or more modules that are configured
together to achieve the desired analytical functionality. Also known as a system,
i.e. an LC/MSD system.
Points The relative level of load (load value) that an instrument or module puts on a
physical or virtual system, aggregating CPU and RAM (memory) consumption.
Different types of detectors require more processing power and memory than
others, so they have been assigned higher “point” values.
Concurrent The number of simultaneous connections to the system. This includes user
sessions connections and instrument connections. For OpenLab CDS, use points as the
measure of instrumentation as this reflects both number of instruments and the
amount of data generated by each instrument system. See below for examples
of number of points per instrument type.
Instrument Connection Licenses refer to the licenses required in the software in order to run
licenses a specific instrument configuration. Note: Points do not equal instrument
connection licenses.
Standalone PC Up to 6 points
1 For customers migrating from OpenLab ChemStation their existing computer hardware is supported. In
this case it is possible to configure up to 10 points per AIC. See OpenLab ChemStation documentation for
ChemStation AIC specifications
Load Approximations
The following tables provide some guidance how to approximate load-based
requirements for different instruments. Note that the load is also depending on
actual instrument utilization and number of users.
Table 8 Points vs Licenses – OpenLab CDS system with Agilent LC Instruments
LC 2D detector 0 0
2 Additional Agilent DAD (if same DAD model) or a 3D-FLD, does not consume an additional license.
Non-Agilent 2D or 3D detectors require a specific own connection license.
Instrument Connection
Points Licenses
A2D instrument1 1 12
3 For a system that includes an A2D module as part of a supported Agilent instrument configuration,
e.g. a 8890 with additional detector connected through A2D
Disk Space
Disk space requirements should be adjusted based on the number and type of
instruments and archival periodicity. Agilent recommends providing disk space
for one year of lab operation in addition to the operating system and OpenLab
CDS requirements.
Table 12 Typical expected file sizes
Up to 70 Up to 30 All-in-One server
Up to 90 Up to 50 Two-server
1 See section "Configuration Capacity" for details on how to calculate Concurrent Sessions for OpenLab
CDS
This chapter contains the software requirements for the different components of
an OpenLab CDS system.
Component Details
3 The listed antivirus software has been tested and is recommended by Agilent. Support is not limited to
this software. Check the specific requirements and support of each product.
You may use a PDF viewer to open PDF manuals in OpenLab Help & Learning.
The PDF viewer is not required for the system to function correctly.
Language Compatibility
User interfaces are displayed in the language of the Windows operating system
for the following languages:
• English
• Chinese
• Japanese
• Brazilian Portuguese
• Russian (Workstations only)
The English language OpenLab CDS software is also supported with Western
European language operating systems, provided the OS Region settings are
configured correctly.
OpenLab Server/ECM XT, as well as Test Services (QualA) are supported with
HINT English, Chinese, Japanese and Brazilian Portuguese Windows OS languages
only.
Non-localized instrument drivers are supported; They will appear in English even
when running localized versions of OpenLab CDS.
Customized locale settings might be required for Non-Agilent drivers. Please
check the localization statement in the driver documentation.
1 Networked Workstation is defined as a PC machine that supports both user interaction (i.e. sample sub-
mission and data review and processing) as well as automated functions (i.e. data acquisition and auto-
mated processing and printing). For OpenLab CDS a Networked Workstation is an AIC used interactively
for sample submission and data processing.
3 virtual environment
Windows 8.1
Pro/Enterprise
1 Agilent supports the versions supported at release per Windows life-cycle fact sheet (https://docs.micro-
soft.com/en-us/lifecycle/faq/windows). Agilent expects, but cannot guarantee, that newer minor product
versions will be compatible.
2 Agilent does not support use on Windows Home Edition, or Windows Education Edition.
3 (https://www.agilent.com/en/support/software-informatics/analytical-software-suite/chromatography-dat
a-systems/openlab-cds/faq-openlab-cds-ltsc) . Any system using LTSC must be managed under a cus-
tom support agreement.
4 All Windows Server compatibility is provided per ECM XT/OpenLab Server compatibility. Other content
management compatibility is defined by that product.
Agilent may ship any supported Windows version. Updating to a different version
is the responsibility of the end user and is not part of standard installation
process.
Supported Databases
OpenLab CDS supports PostgreSQL databases for hosting the Data Repository.
OpenLab CDS Workstation Plus uses a PostgreSQL database for Shared
Services, the Data Repository and Content Management. It is installed and
configured automatically during installation. Agilent supports only the use of the
PostgreSQL version as installed by the OpenLab software.
If you are using OpenLab Server/ECM XT or OpenLab ECM to store data, please
refer to the respective product documentation for information on supported
databases.
For information on Databases for Shared Services Server please check the
manual Configuring OpenLab CDS with OpenLab ECM.
Virtualization
Terminating client sessions when there is data in the File Upload Queue will
NOTE result in data loss (for example, when reprocessing a large data set).
To avoid the risk of data loss do not use aggressive session management
settings that automatically terminate client sessions as a way to save server
resources. Session management should only disconnect idle sessions but not
terminate them.
(Thin)CDSClients
CDSClient(Host) RDSClient
RDSServer CitrixClient
OL/DBServer CitrixXenApp VMHorizonClient
VMHorizon HyperͲVClient
ECMXT2.xServer
Database(s) HyperͲV
Corporate LAN
WAN
Corporate LAN
CDS2.xAIC
Domain/DNS
Domain/DNS
NonͲAgilent
AgilentLC’s AgilentGC’s Instruments
Agilent Agilent
LCMS GCMS
More Information
See the Technical Overview Virtualizing OpenLab CDS Client/Server Systems
(5994-3609EN) for more detail.
For details on virtualization of OpenLab servers contact your Agilent support
representative.
Licensing
OpenLab CDS uses FlexNet Publisher (v. 11.12) for the distribution and tracking
of license entitlements. This software is installed with the OpenLab CDS
components.
Introduction 35
Network Specifications 36
About LAN Communications 39
Power Management 39
Specific Requirements for Compliant Systems 40
Firewall Settings 41
OpenLab Server /OpenLab ECM XT 42
ECM XT Add-ons 45
OpenLab CDS AICs 46
OpenLab CDS Clients 48
Instrument Communication 50
OpenLab ECM 3.x Server 51
OpenLab ECM 3.x Clients 52
Dynamic Ports 53
Introduction
Network Specifications
+
" $ " ,
! -.
% !' *
/$ 0 * 1"2
! ! "
MTU size MTU size on all segments of the packet path is the same
Connectivity from server component to All queries are executed successfully. Execution time is < 1
Active Directory through LDAP: s per query.
• Retrieve 2000 users from AD
• Get all available domains
• Get global catalog
All domains trusted from the currently All domain controllers are available, have the correct
running domain: version and global catalog is enabled. All LDAP checks for
• Domain Controller is available and each domain controller are within specification.
the tool can establish connection
• Requires domain functional levels
Windows Server 2008 or higher
• Global catalog is enabled
• Run all standard LDAP Checks
Table 17 Additional network requirements for Shared Services Server and Domain Controller
Server
Check Specification
Check Specification
The J4100 Jet Direct Card is not supported. For example, use a G1369 LAN
NOTE interface card instead.
See the separate driver installation guides for further information regarding
NOTE vendor specific instrument connections. GPIB or RS232 might be required.
Power Management
1
View the items by icon to see a list of all items.
If you intend to use your system in a compliant environment, ensure the following
settings related to time synchronization:
• Your network must have a time synchronization service to make sure that all
systems are using a consistent and valid time.
• To ensure that users cannot change the time, users must not operate using
an administrator account.
Firewall Settings
If you are using a third party firewall or anti-virus software on a network with
OpenLab CDS, the firewall ports listed in this section may not be in use by other
applications to allow communication between the system components of
OpenLab CDS. These apply to workstations as well as to Client/Server systems
as component communications rely on these communication channels.
For any mixed distributed configurations (clients or server) also the ports
required for v2.6 need to be opened and available. The System Preparation Tool
(SPT) of OpenLab installer will check the ports during installation, and set up
firewall rules that apply including v2.6 to allow for mixed mode configurations
during upgrade.
The ports listed only in the column for v2.6 and earlier, are not needed for pure
NOTE new installations of v2.7. They may be closed manually.
Local ports are used for internal communications only and do not need to be
opened in the firewall.
CM Content Management
DR Data Repository
OpenLab Reverse HTTP :80/ HTTP :80/ Any OpenLab Reverse Proxy
Proxy (Apache
HTTPD) HTTPS :443/ HTTPS :443/ Any OpenLab Reverse Proxy
OLSS Diagnostics HTTPS 443 TCP 3424 Clients, Used for collecting diagnostic logs
AICs,
Servers
Content TCP 5432 TCP 5432 Alfresco Required for secure system on
Management PostgreSQL systems
PostgreSQL Server For database access
Data Collection HTTPS :443/openlab/dcs HTTPS 52088 ChemSta DCS (until CDS 2.4, ChemStation)
Service2 tion ECM XT (may or may not be
HTTP 6328 (used by HTTP 6328 ECM XT remote)
ECM XT) (used by Server
ECM XT)
OLSS Server TCP 6570 TCP 6570 Clients, OpenLab Licensing (Flexera)
AICs Server
OLSS Server (REST HTTP 66253 HTTP 6625 Clients, (SSL Termination) Shared Services
API) HTTPS 443 HTTPS 443 AICs REST API, Licensing Support
service REST API
CM Server1 HTTP localhost:9083 HTTP 9083 Internal OpenLab Server website and REST
APIs (accessed via Reverse Proxy
only)
Test Services HTTPS :443/testservices/ HTTPS 9092 Any The port number can be changed
(QualA) Web Service :443/openlab/ca/ using QualA Config tool
(REST APIs and
website)
Test Services HTTPS :443/testservicess HTTPS :52088/o Any Central Management Service
(QualA) Central erver/ penlab/te manages scheduling and email
Management stservice notifications for Test Services
Service sserver/
Reverse Proxy HTTP 12876 HTTP 12876 Internal Reverse Proxy Configuration
Configuration Service hosts REST APIs to
Service5 configure the Reverse Proxy Server
Disabled after install
DCS2 HTTPS :443/openlab/dcs HTTPS 52088 Any Required in secure systems for
backwards compatibility with older
CertSvc6
HTTPS /openlab/certservi HTTPS 52088 clients/servers
ATS
ce/ Certificate Service
OLSS Server
• Required for internal
communication on secure
systems
• Not required in a secure
configuration for incoming
traffic
Not required for secure systems:
• Data Collection Service,
• Audit Trail Service
RabbitMQ Server TCP 5671 TCP 5671, Any AMQP Ports (HTTPS)
5672
6 No program configured in Windows Firewall - exe path is: C:\Program Files\Agilent Technologies\OpenLab Certificate Service\
Bin\Agilent.OpenLab.CertService.CertServiceCore.exe
OLSS Server TCP 25 Email Server If email server uses a different port, or uses secure
ports, the destination port will be different.
OLSS Server TCP 137 NetBios WINS For NetBios/Name resolution for NT Share
-139
OLSS Server TCP 445 NAS/Share Server Server Message Block (SMB). Used for storage on a
remote NAS share.
CM Server, OLSS TCP 1433 SQL Server Only when using MS SQL Server. Configurable
CM Server, OLSS UDP 1434 SQL Server Only when using MS SQL Server. UDP
CM Server, OLSS TCP 1521 Oracle Server Only when using Oracle Server; Configurable
OLSS Server TCP 3269 LDAP Server Global Catalog LDAP SSL
CM Server, OLSS TCP 5432 PostgreSQL Server Only when using external PostgreSQL Server.
Configurable
ECM XT Add-ons
Table 21 ECM XT Add-ons - Inbound Rules
Import Scheduler HTTP 9091 Server, Services for Import Scheduler communication port for Web UI
CM and REST API
Import Scheduler HTTPS 9093 Server, Services for Import Scheduler communication port for Web UI
CM and REST API
OLSS Storage Client TCP 2886 localhost Local traffic only, does not require
open port
OpenLab Automation Service
(Work Area, Buffered Upload)
OLSS Diagnostics HTTPS 443 TCP 3424 Clients, WCF. Used for collecting
(WCF) (WCF) AICs, troubleshooting logs
Server
OLSS Storage Client HTTPS 443 HTTP 6628 Clients Remote Work Area REST API
Test Services HTTPS :443/testservices/ HTTPS 9092 Any Test Services Web Service hosts
(QualA) Website & :443/openlab/ca REST APIs and website on this
REST APIs port. The port number can be
changed using Test Services
Config tool.
Sample Scheduler HTTPS 443 HTTPS 52088 Clients CDS 2.7 or later - messaging
Agent communication4
3 Reverse proxy is installed and active; all incoming connections are routed through the proxy
4 Reverse proxy is installed and active; all incoming connections are routed through the proxy
Content TCP 80 OpenLab Server OpenLab Server website and REST APIs
Management
TCP 443 OpenLab Server OpenLab Server secure website & secure REST APIs.
Needed only if HTTPS is used
OLSS Client API HTTPS 443 OpenLab Server OpenLab Shared Services WCF APIs
OLSS Licensing API TCP 6570 OpenLab Server OpenLab Licensing (Flexera) Server
Control Panel HTTPS 443 OpenLab Server Shared Services REST API, Licensing Support service
REST API
Sample Scheduler HTTPS 443 OpenLab Server Sample Scheduler, connection to Orchestrator
service
Control Panel TCP 3424 TCP 3424 Clients, Used for collecting diagnostics
AICs, logs
Servers
Test Services HTTPS 9092 (optional)1 HTTPS 9092 Any Test Services Web Service hosts
(QualA) REST APIs and website on port
9092 1
1 It is not necessary to open this port in the firewall for the tool to work. Users can load the web UI and access REST APIs using
https://localhost:9092/ from local system (client) itself. However if remote access it required, then this port should be open in the
firewall and users can access https://<client-fqdn>:9092/ from remote systems. PS 1: Reverse Proxy is not available on Client
systems. PS 2: The port number can be changed using QualA Config tool.
TCP 443 OpenLab Server OpenLab Server secure website and secure REST
APIs. Needed only if HTTPS is used.
OLSS Client API HTTPS 443 OpenLab Server OpenLab Shared Services WCF APIs
OLSS Licensing API TCP 6570 OpenLab Server OpenLab Licensing (Flexera) Server
Acquisition HTTPS 443 AIC Agilent OpenLab remote work area. Client talks to
AICs on this port
OLCF Data HTTPS 443 OpenLab Server Data Collection Service, 6328 used as fallback only if
Collection API, HTTPS is not available
Data Collection HTTP 6328
Agent
Instrument Communication
Table 26 Instruments - Inbound Rules
TCP, UDP 20 AIC, Workstation (FTP) GC MSD Firmware installation (SQ 597*,
Triple-Quad 70**)
TCP 21 AIC, Workstation
TCP, UDP 23 AIC, Workstation (Telnet) GC MSD Firmware installation (SQ 597*,
Triple-Quad 70**)
TCP 2883-2886 AIC, Workstation GC MSD instrument control (5975, 5973 MSD)
3068, 3071 (proprietary/SunRPC/TCP)
TCP 7972, 7973 AIC, Workstation GC MSD instrument control (597* MSD): Slick
protocol
TCP 10000-10020 AIC, Workstation Instrument communication (GC 78xx, 88xx, 9000)
TCP 443 Clients OpenLab ECM Server secure website & REST APIs
operate on this port.
Reverse Proxy HTTP 12876 Internal (localhost Reverse Proxy Configuration Service hosts REST
Configuration only) APIs to configure the Reverse Proxy Server's
Service1 configuration file. Currently, this modifies the Apache
HTTPD server's httpd.conf file.
TCP 1433 SQL Server Only when using MS SQL Server. Configurable.
UDP 1434 SQL Server Only when using MS SQL Server. UDP
TCP 1521 Oracle Server Only when using Oracle Server. Configurable.
Dynamic Ports
Dynamic Ports: used for temporary communications between clients. The ports
used depend on the operating system in use and are configurable. See the
operating system documentation for more information.
The System Preparation Tool (SPT) checks and applies Windows settings on
your machine.
The SPT only checks that minimum requirements are met. Work with your
NOTE Agilent representative to ensure your system is configured sufficiently for your
projected number of users, instruments, and load.
The settings are also applied automatically when you run the OpenLab installer.
Running the SPT in advance helps you to shorten the installation process and
gives you an overview of both mandatory and recommended settings.
1 Optional: Copy the entire content of the USB media to a local drive or
centralized folder, then remove the USB media from the PC.
2 To open the installer, right-click the setup.exe file, and run it as administrator.
If User Account Control (UAC) is switched on, this step requires active
NOTE confirmation to continue.
You can clear the check boxes for recommended settings. Mandatory
settings cannot be cleared. Recommended actions are selected by default
and will be applied unless they are cleared.
For more information on mandatory and recommended settings, see
“Reference of SPT Checks” on page 58.
7 Click Apply Fixes to apply the correct settings.
The System Preparation Tool attempts to fix the selected settings and
displays the new status on the Update Configuration page. All actions are
saved to a log file. A link to the log file is provided at the bottom of the page.
8 Click Next to proceed to the System Preparation Report page.
The System Preparation Report is displayed. It lists the new status for all
selected settings.
The System Preparation Report is saved to disk. Its location is shown at the
top of the page.
9 Click Print Report to print the System Preparation Report.
You may print to a file, for example, using the Adobe PDF printer, and add
comments.
10 The System Preparation Report lists any mandatory or recommended
settings that are not automatically updated by the System Preparation Tool.
Follow the instructions provided in the Actions Required section of the System
Preparation report to manually update operating system settings.
11 Click Finish.
12 Reboot your system if requested to do so.
Telnet Client
TFTP Client
System
HTTP service
Power options
1 W10/W11 only
2 Local Group Policy Requirements for OpenLab: - Set "Hide entry points for Fast User Switching" to Enabled - Adds “Users” group to
"Access this computer from the network" setting
CPU
Minimum Memory
OS Compatibility
OS Minimum Version 1
Screen Resolution
Language Compatibility
Ports - configuration
1 W10/W11 only
1 W10/W11 only
2 Windows 10 only
Table 35 Actions Required section of the SPT report (Settings need to be checked and updated manually)
Windows activation
Instrument Drivers 62
Agilent LC, SFC, and CE Instrument support 64
Recommended Firmware 64
Supported LC modules 65
Agilent Supercritical Fluid Chromatography (SFC) Modules 74
Capillary Electrophoresis (CE) Instruments 75
Agilent LC/MS Instrument support 76
Agilent GC System and Sampler support 78
Agilent GC System support 78
Agilent GC Autosampler support 80
Agilent Headspace Sampler support 81
Agilent CTC Sampler support 82
Mini Thermal Desorber 83
Agilent GC/MS Instrument support 84
Other supported Agilent Instruments 85
Non-Agilent Instruments 86
OpenLab CDS VL Workstation and OpenLab CDS VL Workstation Plus
Instruments 88
Instrument Drivers
The following Agilent instrument drivers are packaged with the OpenLab CDS 2.7
software.
Table 36 Agilent driver packages shipped with OpenLab CDS 2.7
Agilent GC 3.7
1 NOTE: LC&CE driver version 3.5 or higher is required to use the extended CE workflow, for
non-RC.Net Method Import (2DLC), and for the whole feature set provided by the Modify instrument
configuration in CDS Client permission.
Drivers that are not automatically installed by the OpenLab CDS installer can be
found on the media under Setup\Packages\Add-ons. To install these drivers see
section Install or Upgrade Driver Software in chapter 2 of your OpenLab CDS
Workstation, or Clients and Instrument Controller guide.
Agilent driver software is forward compatible with respect to firmware, i.e. the
firmware can be updated without the need of updating the driver or CDS. Note
that Agilent and other vendors release drivers and firmware independent of the
OpenLab CDS releases.
Recommended Firmware
OpenLab CDS 2.7 is shipped with Agilent LC and CE Drivers 3.4. With the release
of this driver version it is recommended to use the following firmware revisions:
Device Recommended
Firmware
Agilent 1100 Series, 1200 Series and 1200 Infinity A.07.01 or later
Agilent 1200 Series, 1200 Infinity and 1120 Compact LC B.07.34 or later
Note that some driver features like valve-thermostat clusters and new
temperature control modes require current firmware versions. Agilent
recommends that you always use the most recent firmware revisions with your
OpenLab CDS to have access to the latest firmware features and improvements.
Download current LC/CE Firmware from
www.agilent.com/en-us/firmwareDownload?whid=69761.
Firmware revisions are grouped into sets for each module or system. Firmware
sets include just the latest firmware of each module.
Do not mix firmware revisions from one set with older or newer sets. A firmware
NOTE update within set A/B/C/D.07.xx is required for all modules in that stack, not only
new modules.
For detailed driver information please refer to the LC driver release notes. A
recent version of Release Note for Agilent LC and CE Drivers
(LC-and-CE-Driver-Release-Note-3-4.pdf) is available in the Docs/EN folder of the
OpenLab CDS media.
Supported LC modules
Most Agilent LC modules can be controlled with the current version of OpenLab
CDS. LC & CE driver release 3.4 has been tested with this revision and is installed
by default with the software.
LC & CE Drivers 3.4 come “ready for data acquisition” with 2D-LC systems. 2D-LC
release 1.1 together with OpenLab CDS v2.7, provides extensive functionality
around 2D-LC. Several new features have been added to the 2D-LC functionality.
Activating the 2D-LC System driver will require a separate dongle-based license.
The Agilent LC drivers are backwards compatible. Modules with identical product
NOTE numbers are supported. The tables below list the name of the current model
version. Look for the product number at the lower right of each module or
system. 1100 Series models are supported on a best effort basis only.
For more detail please refer to the release notes of the driver revision you are
using.
1 Pump valve clusters are possible for marked pumps with up to 2 valves of type G1160A and/or G1170A
with 5067-4159 or 5067-4147.
VTC Valve Combinations of G7116B, G1170A and G1316C (valve See LC Driver Release
Thermostat or column hosts) and G1316A/B and G7130A Notes: Valve-Thermostat
Cluster Cluster
G1156A 1200 Series 6-Position/7-Port Valve (400 bar) A module to hold valve
part of purge kit head
host required
G4286B 1220 Infinity LC System Isocratic, Man. Inj., VWD, 600 bar
G4287B 1220 Infinity LC System Isocratic, ALS, VWD and Oven 600 bar
G4288C 1220 Infinity LC System VL, Gradient, Man. Inj. VWD, 400 bar
G4289B 1220 Infinity LC Gradient, Man. Inj., VWD and Oven 600 ba
G4289C 1220 Infinity LC System VL, Gradient, Man. Inj. VWD, 400 bar
G4290C 1220 Infinity LC System VL, Gradient, ALS, TCC, VWD, 400 bar
G4291B 1220 Infinity LC System Isocratic, Man. Inj., VWD and Oven 600 bar
G4293C 1220 Infinity LC System VL, Gradient, ALS, VWD, 400 bar
1 These are manual modules which do not have a driver. However, they are supported within an 1200 LC
instrument.
Agilent 1120 and 1220 Compact LC Systems are not supported with Fraction
NOTE Collection.
Recommended Firmware
Always use the most recent Firmware installation package that comes with the
driver package.
LC/MS Modules
Agilent Single Quad 6100 Series instruments may be controlled with OpenLab
CDS.
Table 48 Compatibility statements for Agilent LC/MS instruments
Installs of Headspace driver version 3.x will uninstall B.01.09 or earlier headspace
drivers. Any instrument with a G1888 configured will encounter an error message
that the G1888 driver is not installed and the G1888 will be removed from the
configuration. In this case, install the Agilent OpenLab CDS – Agilent G1888
Headspace driver from the media and reconfigure the instrument with the
G1888.
G6501B Agilent GC Sampler 80 for Liquid Not compatible with 8860 Series, 8890
Injection Series, or Intuvo 9000 GC systems.
Not supported with any GC ALS.
G6502B Agilent GC Injector 80 for Liquid Injection
Recommended Firmware
Always use the most recent MS Firmware installation package that comes with
the driver package.
The supported GC/MS firmware is available after installation of OpenLab CDS in
Program Files (x86) >Agilent Technologies >OpenLab
Acquisition >GCMS >Firmware. Open the appropriate MS folder (5977 or 5975)
and run msupdate.exe to upgrade the MS firmware.
GC/MS Modules
The table below summarizes the compatibility of Agilent GC/MS Single Quad
Series models with OpenLab CDS2.7.
For GCs supported as part of a GC/MS system - see “Agilent GC System and
Sampler support” on page 78.
For Autosamplers, CTC Samplers and Headspace supported as part of a GC/MS
system - see “Agilent GC Autosampler support” on page 80, “Agilent Headspace
Sampler support” on page 81, or “Agilent CTC Sampler support” on page 82
respectively.
Table 62 Compatibility statements for Agilent GC/MS Single Quad Series Models
5975A
5975B MS System supported
5975C CI added with GC/MS Driver A.01.02
5975E
5977A
5977B MS System supported
5977E CI added with GC/MS Driver A.01.02
Other Agilent instruments or modules that you can connect with the latest
revision of OpenLab CDS:
Table 63 Other Agilent modules
Non-Agilent Instruments
OpenLab CDS offers comprehensive multi-vendor instrument control for LC, GC,
LC/MS and GC/MS systems. The support matrix for non-Agilent drivers is subject
to frequent changes.
n/a Axcend1 LC
M8223BA CTC LC
M8227BA Hitachi LC
n/a Metrohm IC
M8229BA PerkinElmer LC
M8242BA GC
M8240BA Shimadzu LC
M8232BA GC
Table 64 Non-Agilent instrument drivers available for use with OpenLab CDS v2.7
M8505BA Waters3 LC
1
planned
2
Not supported on Windows 11 at time of release of OpenLab CDS v2.7
3
Waters Driver Pack does not support Windows 11 at time of release of OpenLab CDS
v2.7
OpenLab CDS VL and VL Plus licenses include one instrument connection, and
provide a limited support for Agilent instruments. The following subset of Agilent
Chromatography instruments is configurable with a OpenLab CDS VL
Workstation or OpenLab CDS VL Workstation Plus.
Table 65 VL-configurable 1260 Infinity LC series modules
G4286B 1220 Infinity LC System Isocratic, Man. Inj., VWD, 600 bar
G4287B 1220 Infinity LC System Isocratic, ALS, VWD and Oven 600 bar
G4288C 1220 Infinity LC System VL, Gradient, Man. Inj. VWD, 400 bar
G4289B 1220 Infinity LC Gradient, Man. Inj., VWD and Oven 600 ba
G4289C 1220 Infinity LC System VL, Gradient, Man. Inj. VWD, 400 bar
G4290C 1220 Infinity LC System VL, Gradient, ALS, TCC, VWD, 400 bar
G4291B 1220 Infinity LC System Isocratic, Man. Inj., VWD and Oven 600 bar
G4293C 1220 Infinity LC System VL, Gradient, ALS, VWD, 400 bar
1 These are manual modules which do not have a driver. However, they are supported within an 1200 LC
instrument.
1 Instrument add-ons, like Headspace and RTL, are supported per the availability of that feature on the asso-
ciated instrument.
OpenLab ECM 3.5 Update 6, or greater Details of CDS data usage in ECM are
3.6 (all) documented within ECM.
3.6 update 2 or greater is required for certain
search related features.
OpenLab Shared 3.6 For OpenLab CDS with OpenLab ECM 3.x
Services Server configurations only.
Standalone Deployment of OpenLab Shared
Services Server without a content
management solution (aka Distributed
WorkStations) with OpenLab CDS is not
supported.
For Networked Workstations (Client Co-installed on AIC) not all add-ons are
NOTE supported on all configurations. Check add-on Software product documentation
for details.
OpenLab ECM OpenLab Shared See ECM requirements for additional details
Services Server regarding capacity requirements
1 See OpenLab ECM XT/ OpenLab Server Documentation for list of supported configurations.
Mixed configurations
Environments combining OpenLab CDS with other Agilent CDS products are
supported with the following storage backends:
• OpenLab Server/OpenLab ECM XT
Mixed configurations are supported if OpenLab CDS and the other CDS
product can both connect to the same OpenLab Server/OpenLab ECM XT
version. See Agilent white-paper Mixed Environment Support in OpenLab CDS
for details on the Mixed Environment design.
Check with your Agilent contact for details on the CDS versions supported in
mixed configurations. At time of release the following versions can be
connected with OpenLab CDS in mixed configurations, per compatibility of
ChemStation or EZChrom with OpenLab Server/ECM XT respectively:
• OpenLab Chemstation C.01.09
• OpenLab Chemstation C.01.10
• OpenLab EZChrom A.04.09
• OpenLab EZChrom A.04.10
• OpenLab ECM
OpenLab CDS can be configured with OpenLab ECM v3.x via an Agilent
OpenLab Shared Services Server to operate in an environment containing
different Data Systems. OpenLab ChemStation or OpenLab EZChrom may be
connected to OpenLab ECM via a separate Shared Services server.
For more details, refer to the Configuring OpenLab CDS with OpenLab ECM
guide (CDS_v2.7_configure-with-ECM_en.pdf).
Standalone deployment of OpenLab Shared Services Server without a content
management solution (aka Distributed Workstations) with OpenLab CDS is
not supported.
The libraries and databases listed below are compatible with version 2.7 of
OpenLab CDS.
Backwards compatibility to
• 2.3 / NIST17
• 2.2 / NIST14
• Any library in the appropriate NIST format can be opened in NIST MS Search,
and a basic search, e.g. name search, can be run against it.
NIST MS Search Software and Demo Library are available for download at
https://chemdata.nist.gov/
Please check the following web site for your local sales and support contact:
https://www.agilent.com/en/support
Agilent Community
To get answers to your questions, join over 10,000 users in the Agilent
Community. Review curated support materials organized by platform
technology. Ask questions to industry colleagues and collaborators. Get
notifications on new videos, documents, tools, and webinars relevant to your
work.
https://community.agilent.com/
www.agilent.com
Agilent Technologies Inc. 2015-2022
Edition: 05/2022
Document No: D0013817 Rev. A