Genie 2000 Tutorials Manual
Genie 2000 Tutorials Manual
9235912E
9236243E
The remaining tutorials are brief introductions; their functions are covered more
thoroughly in the following documents:
• The Gamma Acquisition and Analysis chapter of the Genie 2000 Operations
Manual
• The Model S509 Genie 2000 Alpha Analysis User’s Manual, which is included
at the back of the Genie 2000 Operations Manual.
Glossary
A comprehensive glossary of nuclear spectroscopy terms is included at the back of this
document.
Starting with Genie 2000 version 3.2, the VDM application has been converted to a
Windows Service process, i.e. it runs in the background and does not appear in the
Task bar.
The Genie 2000 application will not be able to open any CAM
datasources until the VDM service is started again.
An MCA Input Definition follows a standard so that it can be saved, reused, and even
copied to other systems. The MCA Input Definition is saved as a single file called a
MID file. You can also save a copy of a MID file to a removable drive if need be, to
move that setup to another system or to easily set it up at a later time.
This chapter explains the basic concepts for the Mid Editor and the Mid Setup Wizard.
For most systems, you’ll use the MID Wizard to help you set up your Input Definition
quickly and easily. But if your Input Definition is more complex than the MID Wizard
was designed to handle, you’ll have to create your definition in the MID Editor.
For instructions on how to create the definition with both the MID Setup Wizard and
the MID Editor for your specific MCA, you may also refer to the your specific MCA
user’s manual.
Genie includes the InSpector 2000 Simulator, software which approximates many of
the functions of a real InSpector 2000. It is designed to be used as a tutorial for and
demonstration of Genie 2000, without the need for an actual MCA, a detector, and
associated front-end electronics. The Simulator mimics a fully functioning InSpector
2000 and works within Genie 2000 as though it were an actual MCA, with working
Adjust Controls and simulated data acquisition. Refer to the InSpector 2000 Simulator
appendix in the Genie 2000 Operations Manual for full details.
Note: When you open the MID Editor on a remote VDM server while it’s in use,
Genie 2000 will prevent certain operations with the VDM.
• Select the MCA and front end hardware to be used for each input through the
Device Menu. To do so, you add the MCA through the MID Editor’s Edit
Menu.
• Define the operating characteristics and modes for each MCA through the
Settings Menu.
• After your desired settings for this MCA Definition have been achieved, save
the MCA Input Definition as a MID file.
Although MID files can reside in your PC’s disk storage, to establish communication
with a MCA, the device’s MCA definition(s) must be loaded into Genie’s run-time
configuration database managed by the Virtual Data Manager (VDM).
A previously-loaded configuration can be edited with the MID Editor, but must first be
unloaded from the database, edited, then re-loaded back into the database.
The MID Wizard can create and load a new MCA definition; it cannot unload a
definition once loaded into the database.
Before you start an experiment you simply pick the pre-defined configuration you want
to use by selecting the appropriate MID file and loading it into the run-time database
with the MID Editor. Once loaded into the database, the definitions will remain in the
database even after rebooting the PC.
Because an MCA definition file can include any or all of the MCAs that are available
to your system, you can use more than one MCA at a time. However, each MCA can be
included only once in a given definition file and each MCA in that file must have a
unique name and/or IP address.
Next, we will discuss how to start the MID Editor, how to add (and delete) the MID
editor’s MCA entries, explain what the definition entry consists of, and how to build an
MCA Definition list.
Start the MID Editor by double clicking on its icon ( ) in the Genie 2000
Program Group (or by selecting the desktop icon, if provided) or from the Start menu
select Programs | GENIE-2000 | MCA Input Definition Editor.
To use the MID Editor, you build the MCA Definition list and you may then add to or
delete the desired MID files from the list.
Note: The phrase “local:Untitled” in the title bar, means that the MID Editor is
connected to the local VDM and that no file is currently open; this is the default
condition in a non-networked environment.
Editing an MCA
The Edit menu, at the top of the MCA Input Definition Editor's main window, is used
to add an MCA to or delete an MCA from an MCA Input Definition.
Adding an MCA
To add an MCA, select the Add MCA command in the Edit menu. You’ll see the Add
MCAs to Definition Table dialog box.
This typical Add MCA tree view shows the MCAs grouped by MCA type. Clicking on
the ‘+’ next to a MCA type lets you choose a specific MCA.
You can add as many MCAs to the definition as are necessary for your system by
highlighting one or more MCAs and clicking the Add button to add them to the MCA
Definition Table. When you’ve finished adding MCAs, click on the Done button.
Deleting an MCA
If you want to remove an MCA that you have added to the definition, you can do it
easily by selecting the table entry you want to delete and then selecting Delete MCA.
From the Delete MCA dialog, click on OK or click twice on the entry in the table to
remove the it.
The selected MCAs entries and connections will be deleted from the MCA Input
Definition Table.
Defining an MCA
Once an MCA has been added to the Input Definition Table, defining the MCA is a two
step process.
• Assigning the devices by using the Devices Menu to tell the system about the
MCA’s front end electronics.
• Setting the Controls by using the Settings Menu to define the settings for the
physical devices selected in the Device Menu.
To begin, click on the MCA entry in the Definition Table that you want to set up.
MCA
This is the type of MCA device being used for this particular entry in the table.
Input
This is the name that will be used to refer to this specific hardware entry in the
table. The MID Editor automatically assigns these names sequentially as DETnn,
starting with nn=01. You can easily edit this assignment to an input name of your
choice.
The rest of the settings are described in further detail in the Genie 2000 Operations
manual.
Device Setup
The Devices menu contains selections to allow defining the physical configuration of
the MCA and front end devices. Some Devices options are disabled (grayed out)
because they do not have any adjustable parameters or may be adjusted later in the
GAA application.
Parameter Settings
The Settings menu contains selections to allow setting the programmable controls of
the MCA and front end devices. Some Settings options are disabled (grayed out)
because they do not have any adjustable parameters. These are described in further
detail by the Genie 2000 Operations manual.
To remind you to save a changed definition, you’ll see an asterisk (*) next to the name
of the current definition in the Title Bar.
The Summary menu has two commands: By MCA and By Input, which change the
order in which the information in the Input Definition Table is displayed. By MCA
means that the first column of the table will display the MCA type that is being used for
each entry. If you choose By Input, the MCA and Input columns will be reversed in the
display and the Inputs will be sorted alphabetically.
You can choose either method, but in the case of systems with a large number of
inputs, By Input is an easier display to understand than By MCA.
Because New is a destructive operation, selecting it will cause the program to ask for a
confirmation in one of two ways:
• If the Definition currently being displayed has not been changed since it was
last saved (no asterisk in the Title Bar), you will be asked if you want to erase
the current Definition. Click on OK to erase it, or Cancel to return to the Input
Definition Editor.
• If the Definition currently being displayed has been changed but not saved,
you will be given a chance to save it.
The key piece of information here is the File Descriptor, which you added when the
file was first saved along with the File name and type. This should help you decide if
the file you selected is the one you want to edit.
This database is shared by all of the programs that make up the Genie 2000 software
package, and is used by those programs to gain access to the actual MCA hardware in
your system. In this section we’ll take a look at the procedures used for setting up that
database.
The Load/Unload functions will be disabled while any Acquisition and Analysis
applications are running and have open datasources. This prevents one user from
altering the runtime database while another user is accessing it.
Wizard Definitions – An MCA Definition created with the MID Wizard will
automatically be loaded into the MCA Runtime Configuration Database when you
finish the definition. You don’t need to do anything more.
Editor Definitions – If you created or edited the MCA Definition with the MID
Editor, you’ll have to manually load the saved file into the database yourself. The next
section tells you how.
If you try to load an Input Definition which contains a duplicate Input name, you’ll
see an error message telling you that the requested Definition can’t be loaded. A
unique name must be selected.
By default, the MID Wizard does not include a Sample Changer device into an Input
Definition. Only the MID Editor can include the Sample Changer device into an Input
definition.
If your Input Definition is more complex than the MID Wizard was designed to handle,
you'll have to use the MID Editor to create your definition. It is covered in detail
starting in the MCA Input Definition (MID) Editor starting on page 3.
Note: At any time you can click Next to proceed to the next screen, click Back to
return to the previous screen or click Cancel to exit the wizard. You can also
access the help topic for each screen by clicking on the Help button.
Note that if you didn’t enter an Input Name, you won’t be allowed to exit the Step 3
screen. If the name you entered is the same as the name of an existing MID file, the
system informs you and goes back to Step 3 to let you enter another name.
The Step 7 Mid FileName defaults to UNTITLED, which you’ll probably want to
change to something more meaningful. If the name you enter is the same as that of an
existing MID file, the system will ask if you want to overwrite the existing file.
The following figure shows the summary for the Inspector 2000 simulator.
This tutorial is intended to be a brief overview of the basic functionality of the Gamma
Spectroscopy Analysis window. The Alpha Spectroscopy Analysis window is very
similar, so Alpha users can benefit from this tutorial as well.
The first step in using your Genie 2000 system is to create at least one MCA Input
Definition (MID) so the system knows what kind of MCA is installed or connected to
your system. This is done using either the Genie MCA Input Definition Editor or the
Mid Setup Wizard.
The next step is to acquire some data to work with. To access the hardware in Genie
you will need to open the datasource that was previously created in the MID Wizard or
MID Editor, where the source for the data can be a Detector or a data file. For this
tutorial, we will open a data file.
2. Type: Select CAM File. CAM stands for Configuration Access Method, a type
of data file which contains a spectrum, its Regions of Interest (ROIs), its
calibration information, etc.
5. Click Open.
The VFS being used is always displayed in the upper right-hand corner of the spectral
display as either VFS=nnnn for a Linear scale or LOG=nnnn for a Logarithmic scale,
where nnnn is the vertical full scale value.
• Click on the upper or lower part scroll bar at right of the data display.
• Use the keyboard’s up and down arrow keys. Each press of an arrow key is the
same as clicking once on the corresponding end of the scroll bar.
Move the mouse cursor to the middle of the spectrum, hold down the left mouse button
and move the cursor to the left and right. You will see a short vertical bar, the spectrum
cursor, following your mouse’s movement.
• Dragging it across the data by holding the left mouse button down while you
move the mouse.
• Using the keyboard’s left and right arrow keys to move the spectrum cursor
through the data.
As you move the spectrum cursor, watch the Display Status Line. You’ll see the
information displayed there, the current channel number, its energy, and the number of
counts in that channel, change with the cursor’s movement. The word Idle at the far left
lets you know that no data acquisition is under way.
The region markers look like an “upside down L”, with the horizontal part pointing to
the left for the left marker, and right for the right marker. The figure on page 18
illustrates the region markers.
2. From the keyboard: First, use the keypad’s arrow keys to move the spectrum
cursor to the new location for a region marker. Press CTRL+L for the Left
marker or CTRL+R for the Right marker and the selected marker will jump to
the cursor location.
While the region markers can be moved individually, we’ve found the following to be
the easiest way to quickly place the markers on either side of a peak of interest when
the markers are close together:
1. Move the mouse cursor (not the spectrum cursor) to the region markers. When
it reaches a marker, it will change into a vertical bar with an arrow attached to
it. Dragging the mouse will move that marker.
The following figure shows the mouse cursor on the left (with an arrow), the
two region markers and the spectrum cursor, the vertical line at the top of the
peak.
Figure 9 The Mouse Cursor, the Two Region Markers, and the Spectrum Cursor
2. When the mouse cursor has a left-pointing arrow, it means that you have
selected the left marker for movement. Similarly, a right-pointing arrow
affects only the right marker.
3. Drag the left marker toward the right until it is just to the right of the peak you
want to analyze. You’ll notice that both markers will move together when you
do this, because the left marker will “push” the right marker along in front of it.
4. Now drag the left marker back to the left until it is at the left edge of the region
you want to analyze. When you do that, the right marker will stay where it was;
the markers are now bracketing the region.
Creating an ROI
In the previous section, you moved the markers so that they would bracket a peak,
which is the first step in creating an ROI. The second step is to press the keyboard’s
INSERT key to create an ROI between the two markers.
• Clicking one of the Control Panel’s ROI Index buttons. The (+) button moves
one ROI to the right and the (–) button moves one ROI to the left.
Deleting an ROI
Deleting an ROI is just as easy as entering one. Just Index the markers to the ROI you
want to remove, then press the keyboard’s DELETE key.
The result will be a display like the one in the following figure. The Reference
Spectrum, the lower half of the display, contains the full spectrum and the Expanded
Region, the upper half of the display, shows the expanded data.
Turning the expand mode on changes the button’s name to Expand Off, so the next
time you click on it (or press F8), the display will change back to normal.
1. Move the mouse cursor into the rectangle; it will change into the four-headed
2. Press the left mouse button and drag the rectangle to the new location. The
Expanded Region will track the movement so you can view the data as the
Expand Rectangle is moved.
2. Click and drag with the mouse, which moves the rectangle’s corner, changing
the size of the rectangle in two directions at once.
Using either of these two methods, you can change the size and shape until the data you
want to view is displayed in the Expanded Region.
This will add two kinds of peak information to the spectral display: Peak Labels,
identifying the nuclide for each peak, and a Peak Information Bubble for the current
peak, listing the peak’s: Nuclide ID, Energy, Net Area with percent error and the
nuclide’s Activity (optional).
Note: The nuclide information will only be displayed after an analysis to identify
nuclides in the spectrum has been executed. The stored NBSSTD.CNF file
with the Genie 2000 has this analysis performed from install.
The Sample Information editor lets you add a description of the sample to the data file.
Note this information does not have to be entered at this time; it can be added at
analysis time.
Many of the fields in this dialog box are for descriptive purposes only. That is, the text
you type in is stored with the data and will be included with the analysis reports to
provide a better description of the sample.
You can also enter information in the sample Quantity, Uncertainty, Units, Sample
Geometry, percent Random Error and percent Systematic Error fields, all of which are
used during sample analysis. Their use is beyond the scope of this tutorial; for
complete information, refer to “The Information Fields” section in the Genie 2000
Operations or Alpha Analysis Manual.
The Sample Date indicates the time that the sample was acquired, and during analysis
the activity will be decay-corrected to this date and time.
If the Sample Info page is not available, you can enable it through the Display |
Preferences menu selection.
• Select either the menu’s File | Save command or the Toolbar’s save icon
• If there is no existing file, you’ll see the “Save as” dialog box shown below,
which lets you save to a new file name
• Select the menu’s File | Save As command. You’ll see the “Save as” dialog
box.
Description
For CAM files, you can enter an optional description of up to 32 characters, making it
easier to identify this data file at a later date.
Save as Type
Use the default file type: CAM Files (*.CNF).
File Name
All you need to do is type in the file’s name. The dot and the extension (file type) will
be automatically added by Genie 2000 when you click Save.
Though Genie 2000 Spectroscopy Analysis includes several methods for Energy
Calibrating a spectrum, we’ll use only one of these methods in the tutorial.
Calibration Setup
Before we create a new energy calibration, we’ll take a look at a system-wide
parameter used in energy calibration. Select Calibrate | Setup to see the Calibration
Setup screen.
You can see two Tolerance parameters in the upper right corner of the window. The
first parameter, Energy Cal, is used in Energy Calibration, discussed below; the second
parameter, Eff Match, is used in Efficiency Calibration. Refer to Basic Efficiency
Calibration Techniques on page 30 for more information.
If you use a higher value, more of the spectrum’s peaks will be seen, which could lead
to false peaks being accepted. On the other hand, using a lower value may cause valid
peaks to be overlooked. Determining the best setting for a given spectrum is beyond
the scope of this tutorial, so we’ll leave the parameter set to its default value. It is
important to note that the appropriate value for this parameter will change depending
on the type of detector you are working with, i.e., NaI vs Germanium.
• For details on the other Energy Calibration methods, refer to the Calibrate
Menu options “Energy Only Calibration”, “Energy Coefficients”, and “Energy
Full” located in the Gamma Acquisition and Analysis chapter of the Genie
2000 Operations Manual or in the Calibrate Menu section of the Alpha
Analysis Manual.
• For information on how Certificate Files are created and maintained, refer to
the Using the Certificate File Editor chapter in your Genie 2000 Operations
Manual.
The energies of the nuclide lines in the certificate file will be loaded into the calibration
list for you.
Now scroll down the list to the 1836 line (88Y), place the spectrum cursor on the 88Y
peak at the high end of the spectrum, then click the Cursor button to add the cursor’s
channel position.
Below the graph, you’ll see the Datasource’s file name and the equation’s values for
Energy, FWHM and Low Tail.
The shape calibration is also being generated, using the FWHM and Low Tail values.
Select the Shape radio button to view these curves. While two points is all that is
required to create and Energy and Shape calibration, it is recommended to use as many
reliable peaks as you have available.
Assuming you have gone through the procedures on Energy Calibration and the use of
Certificate Files, you’ll find Efficiency Calibration to be a similar process. Similarly,
we will discuss only one of several different methods to perform an efficiency
calibration in this section.
The default value of 1.00 keV means that if a spectrum’s peak is within 1.00 keV of a
matching energy in the certificate file, it will be accepted as valid.
If you use a higher value, more of the spectrum’s peaks will be seen, which could lead
to false peaks being accepted. On the other hand, using a lower value may cause valid
peaks to be overlooked. Determining the best setting for a given spectrum is beyond
the scope of this tutorial, so we will leave the parameter set to its default value.
• For details on the other Efficiency Calibration methods, refer to the Calibrate
Menu's Efficiency option located in the Gamma Acquisition and Analysis
chapter of the Genie 2000 Operations Manual or in the Calibrate Menu
section of the Alpha Analysis Manual.
• For information on how Certificate Files are created and maintained, refer to
the Using the Certificate File Editor chapter in your Genie 2000 Operations
Manual.
2. In the Open Certificate File box, double click on Nbsstd.ctf (the same
Certificate File we used in Energy Calibration).
The energy lines in this file will populate the list box, as shown in Efficiency
Calibration Dialog.
• If you know that the current datasource contains peak area analysis results, you
can click the Use-results button to populate the list box with that data. If
you’re not sure whether the current datasource contains peak area data, you’ll
have to use the second method.
• Click the Auto button to have the system automatically perform a Peak Locate
and Peak Area analysis, then calculate and display the efficiency and percent
error values for each energy in the list box. This method will use other
parameters defined in the Calibrate Setup screen, including the Continuum
selections.
With either method, for each peak found that matches one in the list box, ±1.00 keV,
the Efficiency Match Tolerance value, an efficiency is calculated based on the peak net
area and the calibration data contained in the Certificate File.
To use two curves, highlight an energy, then click the Cross-over button to specify
the point where you want the low energy curve to end and the high energy curve to
start. If no crossover point is specified, a single equation is used across the entire
energy range.
2. Click the Cross-over button to put an X to the right of the peak in the list.
This will mark it as the energy used as the crossover point.
Peak Edits
The Peak Edits section lets you enter or change the Energy, Efficiency, and Error
(%) data by hand. You can use up to 45 calibration triplets; any more than that will be
ignored. The Accept button verifies the entered values and displays them in the list
box; the Delete button deletes the highlighted entry from the list box.
Cascade Correction
The Cascade Correction function, which is enabled only under certain conditions,
produces an efficiency calibration that is free from cascade summing effects. Its use is
beyond the scope of this tutorial. You will find a full description of this function in the
Correcting for Cascade Summing appendix of the Genie 2000 Operations Manual.
In addition to the default Dual Curve, you can display the calibration as an Empirical
Curve, a Linear Curve or an Interpolated Curve by selecting the appropriate button in
the Curve section of the data window. The order of these polynomials can be adjusted
in this screen to create a good fit.
All of the Genie 2000 analyses use modules: programs that take one data set – such as
spectral data – perform a process on it, then output the results – such as peak centroid
locations. Depending on the options installed on your system, you’ll find modules for
functions like Peak Location, Peak Area Calculation and Nuclide Identification.
• All look to the datasource for their primary input data. They may also
reference libraries, certificate files, or geometry files.
• All (with few exceptions) place their results back into the datasource so other
modules can use that data. For example, the output from the Peak Location
Module – Peak Centroids – is used as input data by the Peak Net Area and
Nuclide ID Modules. Their outputs, in turn, are used as inputs to other
modules.
• None, with the one exception, are capable of generating a printed report; all
they can do is get data from a datasource, process it, and send results back to
that datasource. However, all the module setup screens have a Generate Report
check box option that will run the Report Module after its analysis is complete
to automatically generate a report for that module.
• One exception is the Report Module. It doesn’t analyze data; it uses the data
generated by the other modules to create a report.
These modules can be run interactively, as demonstrated in the next section, or stored
as an Analysis Sequence File (ASF) that lets you automatically implement a sequence
of analysis modules as discussed in How to Execute an Analysis Sequence on page 39.
• Several days (weeks, months) later you’re asked “Since Nuclide X did
not appear to be present, what was its Minimum Detectable Activity
(MDA)?”.
The results from step 1 were stored in the datasource with the data, so all you
have to do is apply the MDA Module to the datasource and run the Report
Module again.
• Are you sure that’s right? What calibration did you use?
Since the calibration equations are also stored in the datasource, all you need to
do is use the Report Module to generate a Calibration Report.
In short, by running the modules you need in the order you need them, you can easily
tailor your analyses to answer your questions. In addition, as you’ll see in How to Edit
an Analysis Sequence on page 40, a module’s parameters can easily be modified to
refine the results.
Depending on which modules you use and how you modify them, you might have
several automatic analysis sequences for various types of analyses.
For this, we’ll look at an individual module, which you can interactively modify and
execute so you can see just what their impact will be. We’ll see how this works by
modifying the Peak Locate module.
To see the effects of a higher setting (less sensitivity), change the Significance
Threshold value to 12.00, check the Generate Report checkbox, then click on
Execute. When you look at the results in the Report Window, you will see that a
higher threshold (12.00) results in a lower sensitivity: only 17 peaks have been found,
nine less than with a Threshold setting of 5.00.
Each of the modules on the Analyze menu do functions on the spectrum, and in general
must be performed in order. Under each heading, there are a selection of algorithms to
choose from. These are described in detail in the Genie 2000 Customization Tools
Manual.
The above options will send the report to your computer's default printer. You may also
select Export Report to PDF under the File menu, which will create a PDF file of
the report in a location of your choice.
In addition, your custom sequences can be added to the Analyze Menu, allowing you to
tailor your system’s automatic analysis to your specific application needs.
When you select Analyze | Execute Sequence, you’ll see a list of the sequences
installed on your system (see the following figure).
Remember that this tutorial assumes that the Model S501 Gamma Analysis option is
installed on your system, so the list in the figure below includes the option’s sequences,
as well as those for the Basic Spectroscopy software.
To use any sequence, there must be a spectrum in the Gamma window. If a spectrum is
not displayed, open C:\GENIE2K\CAMFILES\Nbsstd.cnf.
The Report Window will display a report of the Peak Analysis of the spectrum (at the
bottom of the following figure).
As an example of how to use this editor, we will modify the standard Peak Analysis
w/Report sequence to make it less sensitive, just as we did interactively.
The Seq. Descriptions list, in the previous figure, shows all of the Sequences in the
Analyze Menu. The one we want to change is “Peak Analysis w/Report”. Double click
on its name to load its steps into the Edit Analysis Sequence window’s Current Steps
list, as shown in the following figure.
In the Current Steps window, you’ll see the four steps in this sequence:
• Reporting – Standard
• Reporting – Standard.
The report module is run twice because each instance uses a different section of the
Analysis template to create its part of the report. When the first instance runs, the report
Header is created. When the second instance runs, the Peak Analysis report is sent to
the Report Window.
The second step in the sequence is the Peak Locate module we ran interactively. Since
we already know how to use this module interactively, we will use it again as to see
how it can be modified and run in a sequence.
Since we want to change the sensitivity, we will edit the “Significance Threshold”
parameter to show 12.00 as the new value, as we did for the interactive version. When
you’re finished, click on OK to return to the Edit Analysis Sequence dialog box.
3. Since you are writing back to the original file, a confirmation box will appear
to confirm overwriting the existing file. Click Yes to complete saving the file.
2. Type a new file name, such as Peak_NEW, in the File name text box.
3. Type a new description in the Seq. Description box. Since this description will
be the sequence’s name in your Execute Sequence Menu, it should be both
meaningful and fairly short. For this example, the revised sequence has been
named Peak Analysis w/Less Sensitivity.
4. After you have completed your entries, click on OK to save the definition.
Using the procedures we have just covered, you can create any number of custom
sequences or edit existing sequences. For further information on defining a sequence,
refer to the “Edit Menu | Analysis Sequence” section of the Genie 2000 Operations
Manual and Alpha Analysis Manual.
Of course, it’ll take a little study to determine which modules will be best for your
sequence. You’ll find that referring to either of the following documents will be very
useful for that study.
All of the modules are fully described in the “Analyze Menu” section of the Genie
2000 Operations Manual and Alpha Analysis Manual. The algorithms behind the
modules are described in detail in the Genie 2000 Customization Tools Manual.
Note: You must install model S574 LabSOCS on your system to complete this
training course.
The [approximate] dimensions of a 125 ml wide mouth Nalgene bottle are as follows:
Wall thicknes = 1 mm
Outside diameter = 50 mm
Inside diameter = 48 mm
As we shall see, the size of the sample can be entered either as a volume or a fill height
(H):
100cm3 100cm3
H= = = 5.526213cm ≈ 55.262mm
mR 2 m (2.4)2
Note that this fill height is approximate as this number has been rounded. The
Geometry Composer performs a similar calculation when the sample size is entered as
a volume and will [as we shall see in step 8] issue a warning that there is a loss of
precision due to this rounding.
Note: A ‘Tip of the Day’ screen may appear over the main window.
Acknowledge this screen by clicking Close.
4. The Select Template window will appear. Select the LabSOCS button and
then “Simplified Beaker” from the LabSOCS list. The window should appear
as shown. When done click OK.
Note: The ISOCS radio button will be grayed out if S573 is not installed.
5. The Select Detector window will appear on top of the Geometry Composer's
main screen. Select “LabSOCS” from the drop down list for Body style, and
the specific detector characterization that corresponds to the detector you will
use for your measurement (i.e., the detector you intend to use to make this
measurement). In this case, select the “Dia=60mm_Coaxial” from the list of
detectors. When done click OK. The figure below shows the window only.
6. The Edit Dimensions – Simplified Beaker window will now appear on top of
the Geometry Composer's main screen. The figure below shows the window
only.
7. Click on the View Drawing button to view the template drawing. The drawing
shows the meaning of each dimension parameter for the box. Enter the
parameters previously determined as follows:
Fill Height of sample = dimension 2.1 = leave blank for now and select
Volume to enter the volume amount.
[The bottle is sitting directly on the face of the detector can – hence no
absorbers and the distance is zero.]
When all of the data has been entered, the screen should appear as shown.
When completed, click OK. Since the sample size was entered as a volume (as
opposed to a fill height) that produced a fill height that had to be rounded
(55.262 vs 55.26213…).
Figure 40 The Efficiency Parameters window shows the energies at which the
efficiency will be calculated
10. From this screen you can either edit the energy list or load an energy list from
a file by:
a. Click the Load from File button to display the Open dialog as shown.
b. Select the file that contains the desired energy list. The energy list should
be consistent with the intended sample measurement. If you are looking
for low energy events, be sure to have more data points in the lower energy
range. Conversely, if you are measuring for NORM materials, making
sure you have data points spanning to 3 MeV is recommended.
c. The Efficiency Parameters window will again appear - this time with the
new list of energies from the selected energy list file - as shown.
Figure 42 The Efficiency Parameters window after loading the Wide Energy.txt File
When satisfied with the energy list, click OK. You will be returned to the main
Geometry Composer window.
11. Click the icon or select the Edit | Efficiency Curve | Validate
Geometry menu option to validate the geometry. The message in the figure
below should appear. Click OK to continue.
12. On the main window, click the icon or select the File | Save As menu
option to save the file. Save the file in the default folder with the name:
Example_1_Simplified_Beaker.geo.
13. Click the icon or select the Edit | Efficiency Curve | Generate data
point menu option. The ISOCS Efficiency Calibration Calculation window
should briefly appear while the calculation is running. It will disappear upon
completion, and will be replaced with a message “Efficiency data points have
been generated: ....”. Click OK to continue.
Note that the file path may be different depending on the where the Geometry
Composer and Genie2k are installed.
14. Exit the Geometry Composer by selecting File | Exit to return to Genie.
b. Click the Select button. An Open screen will now appear. Navigate to the
file specified.
Click the Finish button. The Calibration is now saved to the spectrum stored and
ready to be used. It is recommended to use Interpolated during the efficiency
correction step for results with ISOCS or LabSOCS efficiency calibrations.
Note: You must install model S573 ISOCS on your system to complete this training
course.
Note: A ‘Tip of the Day’ screen may appear over the main window.
Acknowledge this screen by clicking Close.
4. The Select Template window will appear. Select the ISOCS button and then
“Simple Box” from the ISOCS list. The window should appear as shown.
When done click OK.
5. The Select Detector window will appear on top of the Geometry Composer's
main screen. Select “ISOCS” from the drop down list for Body style, and the
appropriate detector for your measurement (i.e., the detector you intend to use
to make this measurement). The figure below shows the window only.
The selected entry tells the ISOCS software where the characterization data for
the specific detector being used is stored. Canberra has used MCNP, a Monte
Carlo program, to develop a comprehensive characterization for each ISOCS
customer’s specific detector. Mathematical models of the detector were
created, and analyzed. The result is a set of formulae that when evaluated
define the efficiency for any point in space 360 degrees around the detector, at
all energies from either 10 or 45 to 7000 keV, and at all distances from the
endcap out to 500 meters.
All characterized detectors that the customer has can be listed here, and
therefore are available for selection here.
6. The Edit Dimensions – Simple Box window will now appear on top of the
Geometry Composer's main screen.
When all of the data has been entered, the screen should appear as shown.
Click on the View Drawing button to view the template drawing. The
drawing shows the meaning of each dimension parameter for the box. Copy
the blank sample sheets from the manual and use them in the field to record the
various container and sample dimensions. The completed sheets can then
serve as part of the permanent record of measurement process.
8. When completed, click OK. A graphic of the geometry will appear in the main
window as shown. Note the view has been modified to allow easier viewing of
the detector and source. You can use combinations of the mouse, Shift key,
and Control key to rotate, zoom, and pan the display to better view your
geometry.
10. On the main window, click the icon or select the File | Save As menu
option to save the file. Save the file in the default folder with the name:
Example_1_Simple_Box.geo as shown.
11. Click the icon or select the Edit | Efficiency Curve | Generate data
point menu option. The ISOCS Efficiency Calibration Calculation window
should briefly appear while the calculation is running. It will disappear upon
completion, and will be replaced with a message “Efficiency data points have
been generated: ....”. Click OK to continue.
Note that the file path may be different depending on the where the Geometry
Composer and Genie2k are installed.
12. Exit the Geometry Composer by selecting File | Exit to return to Genie.
b. Click the Select button. An Open screen will now appear. Navigate to the
file specified as shown in the following figure.
c. Click the Open button. The name of the file containing the ISOCS results
computed above will be displayed in the dialog box. Click the Next
button.
However, before storing it, one might choose to view the calibration,
adjust the order of the polynomial fit, etc. It is recommended to use
Interpolated fit. This is done by clicking the Show button - the results of
which is shown in the following figure.
One can also generate a calibration report by clicking the Report button -
the results of which is shown.
g. Finally, we will store the calibration by clicking the Store button. This
will cause the Store window to appear as shown. Name the calibration file
as shown in the figure below and click Store.
i. Click the Finish button. The Calibration is now stored and ready to be
used.
Upon selecting the “Efficiency Correction by ISOCS” option, the user will then be
presented with the ISOCS Efficiency Corrrection Setup dialog as shown below.
To select a geometry file to be used for the efficiency correction step, the user must
click on the Browse button, navigate to the GEO file of interest, and open it. Upon
selecting a geometry file, the analysis engine will ask the user if they would like to
copy the selected geometry to the current data source (CNF file). In order to proceed
with the Efficiency Correction by ISOCS step, the user must select “Yes” to this
question.
Before the geometry information is copied into the spectrum file, all of the auxiliary
information contained within the GEO file is checked against what is currently
available on the users system. The software will attempt to add any unique materials,
collimators, etc. to the user's local files. If there is already an item with the same name
as the object you are trying to add, then the user will be prompted to either cancel the
action, or to change the name of the item that is causing the conflict.
In addition, the energy list that is used in the ISOCS calculation is created at runtime.
The engine adds the energies from the peak search results to the energy list included in
the GEO file. The uncertainty due to the interpolation between the calculated
efficiency points is reduced by including the energies from the peak search results.
However, any points in the energy list that lie outside of the peak area channel range
limits (set in a previous analysis set up step) will be dropped (e.g. if the energy list has
3 MeV, but the peak area range only goes up to 2.5 MeV, then the 3 MeV point will not
be used in the efficiency calculation). Also, if the combined number of efficiency
points from the GEO file energy list and the peak search records exceeds 45, the
algorithm starts dropping points until the number of energies is 45.
For more details on the algorithm, please refer to the Genie 2000 Customization Tools
manual.
Press Execute, and now your peak efficiency results are available for further nuclide
analyses.
Example
Presented below are example scenarios which can be considered typical for ISOCS and
LabSOCS users, respectively. These examples demonstrate the features of the
Efficiency Correction by ISOCS analysis step in Genie 2000.
ISOCS
Consider a case where users are assaying 55-gallon drums. It is known that these drums
have some sort of hot spot contained in them, but the position of the hot spot can vary
from drum to drum.
The first step to solve this problem is to generate a geometry file in Geometry
Composer. To accommodate a hot spot in a drum, the geometry must be created using
the “Complex Cylinder” ISOCS template.
For this example, select the “Example” detector that comes with every ISOCS /
LabSOCS installation.
Enter the following dimensions for the drum and the hot spot parameter values.
Note that parameters 2.2, 2.3, and 2.4 determine the position of the hot spot within the
drum container. In this example, it is these parameters which can vary for different
drums, and will consequently be adjusted at runtime in Genie. Validate the geometry
and save the geo file as “G2K_Addendum_ISOCS.geo”.
Open up GAA, and load the “NBSSTD.CNF” spectrum file, which will act as our
measured drum spectrum. This spectrum comes with the Genie 2000 installation and
already has peak search records saved in the file. Click on Analyze | Efficiency
Correction | ISOCS… to set up the ISOCS calculation. The following dialog box
will appear.
Here the user can select the geometry file that will be used in the ISOCS efficiency
correction.
Note: If there is already geometry information contained in the spectrum file, the
Geometry File Name and Geometry Description text boxes would be
populated.
Next, the user should click the Browse button and navigate to where the
“G2K_Addendum_ISOCS.geo” file is stored. Upon selecting this file, the user will be
asked whether or not to copy the geometry to this datasource. In order to continue the
user must select “Yes”.
The set up dialog should then be populated with the geometry file path and description.
In this example, the hot spot location parameters will be adjusted at runtime. In order to
do this, the user must set up which parameters are variable. Pressing the Add
Variable Parameter button brings up the runtime variable parameters set up dialog.
The hotspot location parameters are 2.2, 2.3, and 2.4. These parameters can be made
variable by double-clicking the parameter dimension boxes. When they are
double-clicked a comment dialog box will pop up. Enter the following comments for
parameters 2.2, 2.3, and 2.4 respectively. These comments are simple text strings that
are meant to help the user identify the parameter they are providing values for during
run time.
Once these comments have been entered, press the OK button on the runtime variable
parameter setup dialog.
The ISOCS Efficiency Correction Setup dialog will then display the variable
parameters comments that were entered in the Run-time Parameter Query List.
Select the “Generate Report” checkbox, and press Execute. The user will then be
prompted to enter the runtime values for the variables.
Change the hot spot vertical height to 500, the radial offset to 150, and the angular
offset to 90, and press OK. The results of the efficiency calculation will be shown in
the Genie2K reporting window.
Note that the energy list sent to the ISOCS algorithms during the Efficiency Correction
by ISOCS analysis step is a combination of the default energy list and the peaks found
during peak search and is cut off above the upper limit of the peak analysis step set up.
LabSOCS
A typical problem encountered in the laboratory might be where the material
composition and the density of a sample might not be known until the time of
measurement. To set up this example, we must first generate a geometry file in
Geometry Composer using the “Custom General Purpose Beaker” LabSOCS template.
For this example, select the “Example“ detector that comes with every ISOCS /
LabSOCS installation.
Select the 20cc.bkr contour file from the beaker list. A beaker file (*.BKR) file is a set
of contours that define a cylindrical container. This is useful when working with a
container that has rounded or finely detailed walls, such as many standard laboratory
containers. You can create your own or use on the of the predefined containers, as is
done here.
Enter the following dimensions in the “Edit dimensions” dialog window. This will
define a sample inside the custom beaker (fill height is from the lowest point of the
beaker) as well as the position of the beaker with respect to the detector. Note that is is
important that the density of the sample be as accurate as possible to the true sample
density.
Note that that the source parameter is No. 3. Validate the geometry and save the geo
file as “G2K_Addendum_LabSOCS.geo”.
Open up Gamma Acquisition & Analysis and load the NBSSTD.CNF spectrum file,
which will act as our measured spectrum. This spectrum comes with the Genie 2000
installation and already has peak search records saved in the file. Click on Analyze |
Efficiency Correction | ISOCS… to set up the ISOCS calculation. The following
dialog box will appear.
Here the user can select the geometry file that will be used in the ISOCS efficiency
correction.
Note: If there is already geometry information contained in the spectrum file, the
Geometry File Name and Geometry Description text boxes would be
populated.
Next, the user should click the Browse button and navigate to where the
“G2K_Addendum_LabSOCS.geo” file is stored. Upon selecting this file, the user will
be asked whether or not to copy the geometry to this datasource. In order to continue
the user must select “Yes”.
The set up dialog should then be populated with the geometry file path and description.
To see the output of the efficiency calculation, the user should click on the “Generate
Report” check box. To execute the Efficiency Correction step, press the Execute
button. The results of the efficiency calculation should be shown in the Genie2K
reporting window.
Note that the energy list sent to the ISOCS algorithms during the Efficiency Correction
by ISOCS analysis step is a combination of the default energy list and the peaks found
during peak search.
Please refer to the Genie 2000 Operations Manual for full details on Cascade
Summing Correction.
Prerequisites
Before you execute a Cascade Correction analysis, you must have performed these
steps:
1. Create a Geometry Description for the sample analysis configuration that will
be used, as described in the Geometry Composer User's Manual.
• In order to make a positive correlation, the nuclide entries in the NID library
must following the naming convention XX-AAA[m] or XX-AAA[m2], where
XX is the element name, AAA is the mass number, and m or m2 is an optional
letter indicating decay from a metastable state (for example: Bi-212,
Te-131m, or Ir-194m2). For a complete list of available nuclides, refer to
Genie 2000 Coincidence Summing Library.pdf, either on your Genie 2000
CD-ROM or, if documentation was installed on your PC, in its
Genie2K\pdfs\docs folder.
• The gamma-ray energies do not have to match precisely, but rather agree
within the Tolerance defined in the “NID plus Interference Correction Setup”
window.
• It is not necessary that all the gamma-rays in the Coincidence library be listed
in the NID library, nor that the NID library be limited to those in the
Coincidence library. The proper correction will still be applied for the cases
listed.
To complete the analysis, select the Perform Cascade Correction checkbox and
select the Execute button.
Note: For the special case of cascade summing correction with Well Detectors: If
the referenced geometry file contains activity inside the well of a detector,
nuclides suffering from electron capture and positron emission events will not
be corrected. Instead, these will be reported as a MISS at the completion of
the analysis.
Press the Execute button in order to start the NID engine. The Interference Corrected
Report will create a table similar to the following.
On this report, the far right column indicates the results of the cascade summing
correction. If a value is indicated, this is the COI value, or the amount of correction
performed due to summing in or summing out. If the nuclide indicates a “Miss”, as is
shown for K-40, this indicates that the nuclide line was not in the cascade summing
library and no correction was calculated. If the nuclide indicates “Free”, this nuclide
is in the library and has been determined to not suffer from cascade summing effects.
MDA analysis can also be run by clicking Analyze | Detection Limits | Currie
MDA… which will bring up the following set up dialog.
Click the Execute button to launch the Currie MDA engine. The MDA engine will
also correct for cascade summing (using the same geometry as defined for the NID
step), and will generate a report with coincidence correction included in the same
manner as the NID step.
A. Glossary
Underlined terms are defined elsewhere in this glossary. These terms and their cross
references can also be researched online at http:/www.canbera.com.
A
ABSORBED DOSE
Absorbed dose is the amount of energy deposited in any material by ionizing
radiation. It is a measure of energy absorbed per gram of material. The SI unit of
absorbed dose is the gray. The special unit of absorbed dose is the rad.
ACCURACY
The degree of agreement between an individual measurement or average of
measurements and the accepted reference value of the quantity being measured.
See also precision.
ACTIVATION ANALYSIS
A method of chemical analysis (for small traces of material) based on the detection
of characteristic radionuclides in a sample after it has been subjected to nuclear
bombardment.
ADC
Analog to Digital Converter. A device which changes an analog signal to a digital
signal.
AIM
Acquisition Interface Module: a type of multichannel analyzer.
ALARA
Since exposure to radiation always carries some risk, the exposure should be kept
"As Low As Reasonably Achievable", as defined by 10 CFR 20.
ALGORITHM
A set of well-defined rules for solving a problem.
AMPLIFICATION
The process by which weak signals, such as those from a detector are magnified to
a degree suitable for measurement.
ANALOG MULTIPLEXER
An electronic instrument that accepts several inputs and stores each one in a
separate section of MCA memory. Also called a mixer/router.
ANNIHILATION RADIATION
Radiation produced by the annihilation of a positron and an electron. For particles
at rest, two photons with an energy of 511 keV each are produced.
ANTICOINCIDENCE CIRCUIT
A circuit with two inputs. The circuit delivers an output pulse if one input receives
a pulse within a predetermined time interval, usually on the order of milliseconds,
but not if both inputs receive a pulse. A principle used in pulse height analysis. See
also coincidence circuit.
AREA
The number of counts in a given region of a spectrum that are above the continuum
level.
ASCII
An acronym for American Standard Code for Information Interchange, a method
for encoding alphabetical, numeric, and punctuation characters and some
computer control characters.
ATTENUATION CORRECTION
Correction to the observed signal for the attenuation of radiation in a material
between the sample and the detector or within the sample itself.
B
BACKGROUND RADIATION
Radiation due to sources other than the sample, such as cosmic rays, radioactive
materials in the vicinity of a detector or radioactive components of the detection
system other than the sample.
BACKGROUND SUBTRACTION
The statistical process of subtracting the background level of radiation from a
sample count.
BACKSCATTERING
The process of scattering or deflecting into the sensitive volume of a measuring
instrument radiation that originally had no motion in that direction. The process is
dependent on the nature of the mounting material, the shield surrounding the
sample and the detector, the nature of the sample, the type of energy of the
radiation, and the geometry. See also scattering.
BASELINE
In biology, a known base state from which changes are measured. In electronics, a
voltage state (usually zero volts) from which a pulse excursion varies.
BREMSSTRAHLUNG
Radiation produced by the sudden deceleration of an electrically charged particle
when passing through an intense electrical field.
C
CASCADE SUMMING
Also referred to as true coincidence summing, it occurs when two or more pulses
from the same decay are summed because they deposit energy in the detector at the
same time. It is a function of the measurement efficiencies and occurs only with
susceptible cascading nuclides (60Co, 88Y, 152Eu, 133Ba, etc.)
CENTROID
The center of a peak; usually not an exact channel number.
CHANNEL
One of an MCA's memory locations for storage of a specific level of energy or
division of time.
CHERENKOV RADIATION
Photons emitted from polarized molecules when returning to their ground state
following excitation by charged particles traveling faster than the speed of light in
a transparent medium.
CHI-SQUARE TEST
A general procedure for determining the probability that two different
distributions are actually samples of the same population. In nuclear counting
measurements, this test is frequently used to compare the observed variations in
repeat counts of a radioactive sample to the variation predicted by statistical
theory.
COINCIDENCE CIRCUIT
A circuit with two inputs. The circuit delivers an output pulse if both inputs receive
a pulse within a predetermined time interval, usually on the order of milliseconds,
but not if just one input receives a pulse. A principle used in pulse height analysis.
See also anticoincidence circuit.
COINCIDENCE SUMMING
A process where the signal from two or more gamma rays emitted by a single
decay of a single radionuclide occur within the resolving time of the detector end
up being recorded together as a single event so that the recorded event is not
representative of the original decay. Typically causes counts to be lost from the
full energy peaks, but may also cause addition to the full energy peaks.
Coincidence summing is a function of the sample-to-detector geometry, and the
nuclide's decay scheme. It is not a function of the overall count rate.
COLLECT
An MCA function that causes storage of data in memory.
COMPTON SCATTERING
Elastic scattering of photons in materials, resulting in a loss of some of the
photon's energy.
CONFIDENCE FACTOR
It is common practice when reporting results to assign them a confidence level: the
value plus or minus one standard deviation. Radiation protection measurements
are usually reported at the 95% confidence level, meaning that the results would be
expected to be within plus or minus that range 95 out of 100 times. Also called
Confidence Level.
CONTINUUM
A smooth distribution of energy deposited in a gamma detector caused by the
partial absorption of energy from processes such as Compton scattering or
bremsstrahlung.
CONVERSION GAIN
The number of discrete voltage levels (or channels) that the ADC's full scale input
is divided into.
CONVERSION TIME
The time required to change an input signal from one format to another, such as
analog to digital, or time difference to pulse amplitude; contributes to dead time.
COSMIC RAYS
Radiation, both particulate and electromagnetic, that originates outside the earth's
atmosphere.
COUNT
A single detected event or the total number of events registered by a detection
system.
CROSSOVER ENERGY
In some efficiency calibration models, the energy at which one calibration curve is
changed into a second calibration curve. This is used in the Dual Efficiency
Calibration in Genie software.
D
DATASOURCE
A hardware device or a file which stores data acquisition parameters and spectral
data.
DAUGHTER NUCLIDE
A radionuclide produced by the decay of a parent nuclide.
DEAD TIME
The time that the instrument is busy processing an input signal and is not able to
accept another input; often expressed as a percentage. See also live time.
DECAY
The disintegration of the nucleus of an unstable atom by spontaneous fission, by
the spontaneous emission of an alpha particle or beta particle, isomeric transitions,
or by electron capture.
DEFAULT
The value of a parameter used by a program in the absence of a user-supplied
value.
DETECTION LEVEL
The level of net signal that can be predicted to be detectable. See also critical level.
DETECTOR
A device sensitive to radiation which produces a current or voltage pulse which
may or may not correspond to the energy deposited by an individual photon or
particle.
DIGITAL STABILIZATION
The monitoring of one or two reference peaks in a spectrum, one for gain and one
for zero, to correct for drift in the system electronics.
DISCRIMINATOR
An electronic circuit which distinguishes signal pulses according to their pulse
height or voltage so that unwanted counts can be excluded.
DISINTEGRATION
See decay.
DPM
Disintegrations per minute. One DPM equals 60 becquerels.
DOSE
The radiation delivered to the whole human body or to a specified area or organ of
the body. This term is used frequently in whole body counting applications.
E
EFFECTIVE HALF-LIFE [Symbol: Teff]
The time required for a radioactive element in a biological system, such as the
human body, to be reduced by one-half as a result of the combined action of
radioactive decay and biological elimination. Compare half-life and biological
half-life.
EFFICIENCY
The fraction of decay events from a standard sample seen by a detector in the peak
corresponding to the gamma ray energy of the emission, and stored by a detection
system. Also called Peak Efficiency. Used to calibrate the system for quantitative
analyses. Also used to specify germanium detectors, where the relative efficiency
of the germanium detector is compared to a standard (3 x 3 in.) NaI(Tl) detector.
Compare total efficiency.
EFFICIENCY CALIBRATION
A function, a lookup table, or series of functions, which correlate the number of
counts seen by the detection system in specific peaks with known activity
corresponding to such emission energies in a radioactive sample.
ELASTIC SCATTERING
See scattering.
ELECTRODEPOSITION
A process for coating the surface of samples being prepared for alpha spectroscopy
and alpha/beta counting.
ELECTROMAGNETIC RADIATION
A general term to describe an interacting electric and magnetic wave that
propagates through vacuum at the speed of light. It includes radio waves, infrared
light, visible light, ultraviolet light, X rays and gamma rays.
ENERGY CALIBRATION
A function which correlates each channel in the displayed spectrum with a specific
unit of energy. Allows peaks to be identified by their location in the calibrated
spectrum.
ESCAPE PEAK
A peak in a gamma ray spectrum resulting from the pair production process, the
subsequent annihilation of the photons produced, and escape from the detector of
the annihilation photons. If both annihilation photons escape, and the rest of the
original gamma energy is fully absorbed, a double escape peak is produced at an
energy equal to the original gamma ray energy minus 1.022 MeV. If only one of
the photons escapes, a single escape peak is produced at an energy equal to the
original gamma ray energy minus 511 keV.
EXCITED STATE
The state of molecule, atom, or nucleus when it possesses more than its ground
state energy. Excess molecular or atomic energy may be reduced through emission
of photons or heat. Excess nuclear energy may be reduced through emission of
gamma rays or conversion electrons or by further decay of a radionuclide.
eV
See electron volt.
F
FACTORS
The parameters used by an algorithm for its calculations.
G
GAIN, ADC
See conversion gain.
GAIN, AMPLIFIER
The ratio of the amplifier's output signal to its input signal.
GAIN CONTROL
A control used to adjust the height of a pulse received from the detecting system.
GAUSSIAN FIT
Calculating the parameters of a Gaussian (or Normal) function to best match a set
of empirical data (in spectroscopy, the acquired photopeak histogram). This
calculation is typically performed using a least squares method after subtracting
the Compton continuum underlying the peak.
GEOMETRY
The detector to sample distance, the sizes and shapes of the detector, the sample,
and any shielding, all of which affect the radiation seen by the detector. The
geometry helps define the efficiency of the detector.
GROUND STATE
The state of a nucleus, atom or molecule at its lowest energy level.
H
HALF-LIFE [Symbol: Tl/2]
The time in which one half of the atoms of a particular radioactive substance decay
to another nuclear form. Half-lives vary from millionths of a second to billions of
years.
HISTOGRAM
A representation of data by vertical bars, the heights of which indicate the
frequency of energy or time events.
I
INELASTIC SCATTERING
See scattering.
ION
An atom or molecule that has become electrically charged by having lost or gained
one or more electrons. Examples of an ion are an alpha particle, which is a helium
atom minus its two electrons, and a proton, which is a hydrogen atom minus its
single electron.
INDEX
An MCA function that jumps the cursor from one region of interest to another.
INPUT/OUTPUT
The process of loading data into or copying data from an MCA or computer using
a peripheral device, such as a computer, a floppy disk, or a printer.
IN SITU COUNTING
Measurement and analysis of radioactivity performed at the sample's location.
INTEGRAL
The total sum of counts in the region of interest.
INTENSIFY
To change the contrast of a displayed region of interest to set it off from data
regions of lesser importance.
INTERFERING PEAK
A peak due to background radiation which is produced at the location of a peak in
the sample spectrum or due to a peak produced by a radionuclide in the sample at
the location of another radionuclide's peak.
IN VIVO COUNTING
In vivo counting refers to directly measuring and analyzing radionuclide activity
levels in a living body.
IN VITRO COUNTING
In vitro counting refers to samples, such as tissue or blood, being analyzed for
radionuclide activity levels in an artificial environment (outside of a living body).
I/O
See input/output.
IONIZATION
The process by which an electrically neutral atom acquires a charge (either
positive or negative).
IONIZING EVENT
Any process whereby an ion or group of ions is produced. As applied to nuclear
spectroscopy, this refers to the passing of radiation through a gas, a crystal, or a
semiconductor.
ISOMERIC TRANSITION
The de-excitation of an elevated energy level of a nucleus to the ground state of the
same nucleus by the emission of a gamma ray or a conversion electron.
ISOTOPE
One of two or more atoms with the same atomic number (the same chemical
element) but with different atomic weights. An equivalent statement is that the
nuclei of isotopes have the same number of protons (thus the same chemical
element) but different numbers of neutrons (thus the different atomic weight).
Isotopes usually have very nearly the same chemical properties, but somewhat
different physical properties. See also nuclide and stable isotope.
K
keV (kiloelectron volt)
One thousand electron volts.
KEY LINE
Designated in nuclide libraries for reporting purposes only. It is intended to
indicate the highest abundance photopeak energy for nuclides with multiple
energy lines, or the line that is the least likely to have interferences.
L
LAN
Local area network: a network of two or more computers connected together.
Lc
See critical level.
LIVE TIME
The time that the ADC is not busy processing a signal. See also dead time and real
time.
M
MARINELLI BEAKER
A standard sample container that fits securely over a detector cryostat's endcap and
is used when calibrating voluminous samples (usually soil or water solutions).
MASS NUMBER
The sum of the neutrons and protons in a nucleus. It is the nearest whole number to
an atom's atomic weight. For instance, the mass number of 235U is 235.
MCA
See multichannel analyzer.
MCS
See multichannel scaling.
MDA
Minimum detectable activity. See lower limit of detection.
MEAN
The average of a group of numbers.
METASTABLE ISOTOPE
A long-lived energy state of a particular nuclide that is not its ground state. Some
nuclides have more than one isomeric state. An isomeric state has the same mass
number and atomic number as the ground state, but possesses different radioactive
properties.
MIXER/ROUTER
See analog multiplexer.
MONITORING, PERSONNEL
Periodic or continuous observation of the amount of radiation or radioactive
contamination present in or on an individual.
MULTISPECTRAL SCALING
Multispectral scaling acquisition mode, also called “ping-pong” mode, alternately
collects data in two separate memory regions, quickly collecting many spectra
with extremely low latency between acquisitions.
MULTIPLET
Peaks in a spectrum which overlap each other. Compare singlet.
N
NATURALLY OCCURRING RADIOACTIVE MATERIAL (NORM)
Radioactivity that is naturally present in the earth.
NEUTRON [Symbol: n]
An uncharged elementary particle with mass slightly greater than that of the
proton, and found in the nucleus of every atom heavier than hydrogen.
NID
Nuclide Identification, the process of identifying radionuclides by comparing peak
energies detected with entries in a nuclide library.
NIM
Nuclear Instrumentation Module. A nuclear instrument conforming to the
DOE/ER-00457T standard.
NOISE
Unwanted signals on or with a useful signal which can distort its information
content.
NUCLEAR SAFEGUARDS
The general topic of maintaining control and accountability of special nuclear
materials.
NUCLEUS
The positively charged core of an atom, which contains nearly all of the atom's
mass. All nuclei contain both protons and neutrons, except the nucleus of ordinary
hydrogen, which consists of a single proton.
NUCLIDE
A general term applicable to the isotopes of all elements, including both stable and
radioactive forms (radionuclides).
NUCLIDE LIBRARY
A file listing nuclides, their names, half-lives, types, energies/lines, and line
abundances. These files are used with library directed peak searches, nuclide
identification (NID) and as aids in performing calibrations.
O
OFFSET, ADC
A digitally performed shift in the ADC's channel zero. Shifts the entire spectrum
by the selected amount.
OVERLAP
An MCA function allowing one section of memory to be displayed over another.
P
PAIR PRODUCTION
Creation of an electron- positron pair by gamma ray interaction in the field of a
nucleus. For this process to be possible, the gamma ray's energy must exceed 1.022
MeV, twice the rest mass of an electron.
PARAMETER
A variable that is given a constant value for a specific application.
PARENT NUCLIDE
A radionuclide that produces a daughter nuclide during decay.
PEAK
A statistical distribution of digitized energy data for a single energy.
PEAK CHANNEL
The channel number closest to the centroid of a peak.
PEAK FIT
The optimization of parameters to match an expected model shape to empirical
data (see also gaussian fit). This optimization is typically performed using a least
squares method.
PEAK-TO-TOTAL RATIO
The ratio of the observed counts in a full energy peak to the counts in the entire
spectrum, caused by the interaction of radiation with the detector at that emission
energy only.
PHA
See pulse height analysis.
PHOTOELECTRIC ABSORPTION
The process in which a photon interacts with an absorber atom, the photon
disappears completely, and the atom ejects a photoelectron (from one of its bound
shells) in place of the photon.
PHOTOELECTRON
An electron released from an atom or molecule by means of energy supplied by
radiation, especially light.
PHOTON
In quantum theory, light is propagated in discrete packets of energy called
photons. The quantity of energy in each packet is called a quantum.
PHOTOPEAK
See Peak.
PHYSICAL HALF-LIFE
See half-life.
POLE/ZERO (P/Z)
A method of compensating the preamplifier's output signal fall-time and the
amplifier's shaping time constant. Its use improves the amplifier's high count rate
resolution and overload recovery.
PMT
See photomultiplier tube.
POSITRON ANNIHILATION
A process where a positron combines with an electron, producing two annihilation
photons of 511 keV each.
PRECISION
The degree of agreement between several measurements of the same quantity
under specific conditions. See also accuracy.
PRIMORDIAL NUCLIDE
A nuclide as it exists in its original state.
PROGENY
See daughter nuclide.
PROTON
An elementary particle with a single positive electrical charge and a mass
approximately 1837 times that of the electron. The atomic number (Z) of an atom
is equal to the number of protons in its nucleus.
PULSE PILEUP
A condition, where two energy pulses arrive at nearly the same time, which could
produce false data in the spectrum.
Q
QUANTUM
The unit quantity of energy according to quantum theory. It is equal to the product
of the frequency of the electromagnetic radiation and Planck's constant (6.626 x
10-34 J/s).
R
RAD
A special unit of absorbed dose. Equal to 0.01 gray.
RADIATION
The emission or propagation of energy through matter or space by electromagnetic
disturbances which display both wave-like and particle-like behavior. Though in
this context the “particles” are known as photons, the term radiation has been
extended to include streams of fast-moving particles. Nuclear radiation includes
alpha particles, beta particles, gamma rays and free neutrons emitted from an
atomic nucleus during decay.
RADIOACTIVITY
The emission of radiation from the spontaneous disintegration (decay) of an
unstable nuclide.
RADIONUCLIDE
A radioactive isotope. See also nuclide.
RANDOM SUMMING
A process where the signal from two or more separate decays of the same
radionuclide or different radionuclides that occur within the resolving time of the
detector end up being recorded together as a single event so that the recorded event
is not representative of the original decays. Typically causes counts to be lost from
the full energy peaks. Random summing is a function of the overall count rate, or
the activity of the sample being measured.
RANGE, ADC
The full-scale address (number of channels) of the ADC's assigned memory
segment.
REAL TIME
Elapsed clock time; also called true time. Compare live time.
RECOILING NUCLEUS
A nucleus that gains significant kinetic energy from its decay.
RESOLUTION
The ability of a spectroscopy system to differentiate between two peaks that are
close together in energy. Thus, the narrower the peak, the better the resolution
capability. Measured as FWHM.
ROENTGEN
The Roentgen, the international unit of X radiation or gamma radiation, is the
amount of radiation producing, under ideal conditions in one cc ionization of
either sign equal to one electrostatic unit of charge.
ROI
See region of interest.
S
SCA
Single Channel Analyzer. A device which recognizes events (pulses) occurring
between the settings of the lower level discriminator and the upper level
discriminator. In an MCA, each event within these limits is counted; events
outside of these limits are discarded.
SCATTERING
A process that changes a particle's trajectory. Scattering is caused by particle
collisions with atoms, nuclei and other particles or by interactions with electric or
magnetic fields. If there is no change in the total kinetic energy of the system, the
process is called elastic scattering. If the total kinetic energy changes due to a
change in internal energy, the process is called inelastic scattering. See also
backscattering.
SCINTILLATOR
A type of detector which produces a flash of light as the result of an ionizing event.
See also photomultiplier tube.
SELF ABSORPTION
Absorption of the photons emitted by the radioactive nuclides in the sample by the
sample material itself.
SHADOW SHIELD
An attenuating enclosure that shields the detector from direct background
radiation without being a 4m shield. Typically used in whole body counting.
SHAPE CALIBRATION
The process of establishing a relationship between the expected peak shape and
energy. A shape calibration can be established by using two or more peak
FWHM/energy (or FWHM/channel) pairs or by using a least squares fit algorithm.
SINGLET
A single peak in a spectrum, well separated from other peaks. Compare multiplet.
SMOOTHING
To decrease the effects of statistical uncertainties in computerized spectrum
analysis, the content of each channel is replaced by a weighted average over a
number of adjacent channels.
SPECIFIC ACTIVITY
The quantity of radioactivity per unit mass; for example, dpm/g or Bq/g.
SPECTRUM
A distribution of radiation intensity as a function of energy or time.
SPECTROMETER
A device used to count an emission of radiation of a specific energy or range of
energies to the exclusion of all other energies. See also multichannel analyzer.
STABLE ISOTOPE
An isotope that does not undergo radioactive decay.
STRIPPING
Subtracting a specified fractional part of the data in one section of memory from
the data in another section of memory.
T
TOTAL DETECTOR EFFICIENCY
All pulses from the detector are accepted, and all interactions (regardless of how
low in energy) are assumed counted.
TOTAL EFFICIENCY
The ratio of all pulses recorded in the MCA's memory (in all channels) to the
gamma quanta emitted by the sample. Compare efficiency.
TRANSURANIC (TRU)
Possessing an atomic number higher than that of uranium (92).
TRUE TIME
See real time.
U
UNCERTAINTY
In a nuclear decay measurement, uncertainty refers to the lack of complete
knowledge of a sample's decay rate due to the random nature of the decay process
and the finite length of time used to count the sample.
W
WHOLE BODY COUNTING (WBC)
In vivo determination of radionuclide activity levels in the human body. Used to
determine compliance with the regulations of various governmental bodies
regarding radiation exposure.
WINDOW
A term describing the upper and lower limits of radiation energy accepted for
counting by a spectrometer.
X
X RAY
A penetrating form of electromagnetic radiation emitted during electron
transitions in an atom to a lower energy state; usually when outer orbital electrons
give up some energy to replace missing inner orbital electrons
Z
ZERO, ADC
An ADC control which aligns its zero energy output with a specific channel in the
MCA's memory (usually channel zero).