Troubleshooting
Troubleshooting
17
yellow and alarm is still on.
the screen
displays
code "E02"
if the air
pressure is
low.
Turn off the power
immediately. .Check
and make sure the
The cleanness of outlet
The oxygen indicator and inlet of air
concentrator lights without any blocks.
High system Single yellow and And then restart the
Low
temperature temperature is beep the screen equipment, please
higher than displays inform the
70°C. code distributor or
"E35". manufacturer and
shut down the
concentrator if the
alarm is still on.
The
Contact your local
The oxygen indicator
distributor or
concentrator lights
Oxygen sensor manufacturer. Spare
did not receive Single yellow and
communication Low oxygen should be
communication beep the screen
failure prepared for those
from the displays
who are in urgent
oxygen sensor. code
need of oxygen.
"E31".
The Turn off the
The supply indicator concentrator
voltage of the lights immediately, and
concentrator is Single yellow and restart it after
Low Voltage Low
85% lower than beep the screen confirming that the
the standard displays voltage of the
voltage. code network power
"E03". supply is normal.
NOTE:
The oxygen sensor is an optional accessory. The oxygen concentrator
without an oxygen sensor will not trigger alarm for low oxygen
concentration or oxygen sensor communication failure.
18