Troubleshooting

The document outlines the alarm and indicator system for an oxygen concentrator, detailing various alarms related to power issues, oxygen concentration, air pressure, temperature, and voltage. Each alarm includes an audible and visual indicator, priority level, and recommended actions for troubleshooting. It also notes that the oxygen sensor is optional and will not trigger alarms if absent.

Uploaded by

Medis MEDIS
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
8 views2 pages

Troubleshooting

The document outlines the alarm and indicator system for an oxygen concentrator, detailing various alarms related to power issues, oxygen concentration, air pressure, temperature, and voltage. Each alarm includes an audible and visual indicator, priority level, and recommended actions for troubleshooting. It also notes that the oxygen sensor is optional and will not trigger alarms if absent.

Uploaded by

Medis MEDIS
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

4 Troubleshooting

4.1 Oxygen Concentrator Alarm and Indicator System


Alarm system design aims at monitoring the working oxygen concentrator
in case of such situations as power off, abnormal pressure or indicator of
running condition of the equipment. All alarms of this equipment are
technical alarms.
It includes an acoustic alarm system and a visual alarm system.
The list of alarm messages is as follows:
Alarm Alarm reason Audible Visual Priority Measure
Turn off the power
immediately. If an
The alarm still exists after
Triple +
The network indicator confirming that the
double
power is quickly power supply and
+ triple
Power off disconnected flashes High connection are
+
during red, normal, please turn
double
operation frequency: off the oxygen
beep
1.4 2.8Hz concentrator and
contact local dealer
or manufacturer.
Contact your local
distributor or
The
Oxygen manufacturer. Spare
Low Single indicator
concentration is Low oxygen should be
concentration beep lights
lower than 82% prepared for those
yellow
who are in urgent
need of oxygen.
The Turn off the power
indicator immediately. Check
lights and make sure the
The internal
yellow and cleanness of outlet
pressure of the
the screen and inlet of air
oxygen Low
displays without any blocks.
Abnormal air concentrator is Single
code "E05" And then restart the
pressure higher than beep
if the air equipment, please
260kPa or
pressure is inform the
lower than
high. distributor or
20kPa
The manufacturer and
indicator Low shut down the
lights concentrator if the

17
yellow and alarm is still on.
the screen
displays
code "E02"
if the air
pressure is
low.
Turn off the power
immediately. .Check
and make sure the
The cleanness of outlet
The oxygen indicator and inlet of air
concentrator lights without any blocks.
High system Single yellow and And then restart the
Low
temperature temperature is beep the screen equipment, please
higher than displays inform the
70°C. code distributor or
"E35". manufacturer and
shut down the
concentrator if the
alarm is still on.
The
Contact your local
The oxygen indicator
distributor or
concentrator lights
Oxygen sensor manufacturer. Spare
did not receive Single yellow and
communication Low oxygen should be
communication beep the screen
failure prepared for those
from the displays
who are in urgent
oxygen sensor. code
need of oxygen.
"E31".
The Turn off the
The supply indicator concentrator
voltage of the lights immediately, and
concentrator is Single yellow and restart it after
Low Voltage Low
85% lower than beep the screen confirming that the
the standard displays voltage of the
voltage. code network power
"E03". supply is normal.

NOTE:
The oxygen sensor is an optional accessory. The oxygen concentrator
without an oxygen sensor will not trigger alarm for low oxygen
concentration or oxygen sensor communication failure.