LabMedUK Training Day Morning Workshop
LabMedUK Training Day Morning Workshop
You have been asked to lead on the verification of the new method
1. Discuss and decide in your group which experiments are required for
this verification.
2. Consider which samples or other material are required, how you will
obtain appropriate samples for this verification and any issues you
anticipate.
3. Decide what resources you would need for this verification.
4. Making use of the background information provided, decide on
appropriate acceptance criteria for precision and method comparison.
5. Calculate Grubbs’ limits for the imprecision data and determine if any
replicate is an outlier.
6. Determine the upper verification limit for repeatability and within-
laboratory imprecision. Decide if imprecision is acceptable.
a. If accepted, consider if mitigations are required and what these
would be.
b. If not accepted, consider what potential actions or further
experiments are required.
7. Calculate the expanded uncertainty from within Laboratory
Imprecision (ignoring any calibration uncertainty) and compare this to
the Royal College of Pathologists of Australasia Acceptable
Performance Limits.
8. Review raw data, charts and summary statistics for the method
comparison study. Compare the data to your acceptance criteria and
decide whether this is acceptable.
a. If accepted, consider if mitigations are required and what these
would be.
b. If not accepted, consider what potential actions or further
experiments are required.
9. Use the raw data to determine concordance (PPA, PNA, POA) between
the methods at a cut-off of 5 mg/L.
10. Decide an overall conclusion including potential clinical significance of
changing to the new method.
Background Information:
From: https://www.antimicrobialcompanion.scot/nhs-lothian/vancomycin-gentamicin-
other-antibiotics-dosing-guidance/gentamicin-adult-standard-dosing/
Information from the new assay Manufacturer’s Instructions for Use (IFU):
Current assay QC performance:
EQA Method Comparison:
Lower limit applies to values equal to or below the quoted concentration and upper limit applies to
concentrations above the quoted concentration
Imprecision Data:
Run
1 2 3 4 5
Replicate
Repeatability:
Within-laboratory imprecision:
Claims ratio (ρ) = Manufacturer’s within-Lab CV (or SD)/repeatability CV (or SD) – see kit
insert
Second, determine F:
Use the dfR and dfWL to find F in table 2 (assume only 1 sample)
Finally, determine separate UVL values for repeatability and within-laboratory imprecision:
From CLSI EP15-A3 User Verification of Precision and Estimation of Bias; Approved Guidelines – Third
edition.
TABLE 2: UVL factors (F) as a function of DF and number of samples
From CLSI EP15-A3 User Verification of Precision and Estimation of Bias; Approved Guidelines – Third
edition.
Measurement Uncertainty (MU):
35
y = 1.2457x + 0.4506
30 R² = 0.9948
New Method (mg/L)
25
20
15
10
0
0 10 20 30 40
Current Method (mg/L)
4
difference (mg/L)
0
0 5 10 15 20 25 30
-1
Gentamicin Concentration
450.0
400.0
350.0
300.0
Difference (%)
250.0
200.0
150.0
100.0
50.0
0.0
0 5 10 15 20 25 30
-50.0
Gentamicin Concentration
4
Difference (mg/L)
0
0 5 10 15 20 25 30 35 40 45
-1
Sample rank
450.0
400.0
350.0
300.0
Difference (%)
250.0
200.0
150.0
100.0
50.0
0.0
0 5 10 15 20 25 30 35 40 45
-50.0
Sample rank
mg/L %
Mean Difference 1.9 57.8
Mean difference ≤5 mg/L 0.9 85.6
Mean difference >5 mg/L 3.0 31.4
Concordance is very similar to diagnostic accuracy but does not assume that
the comparator method is a gold standard or true diagnosis. It can be used to
determine clinical agreement between two tests. This is often used for
qualitative tests but can be used for a quantitative test at a set cut-off.
Comparative Method
Candidate Positive Negative
Method Positive a b
Negative c d