6310/6315/6330/6360/6390/63120/63150/63180/
63210/63240
E2.01
WARRANTY
EEC certifies that the instrument listed in this manual meets or exceeds published manufacturing
specifications. This instrument was calibrated using standards that are traceable to the National
Institute of Standards Taiwan.
Your new instrument is warranted to be free from defects in workmanship and material for a
period of (2) year from date of shipment. During the warranty period, you must return the
instrument to EEC or its branches or its authorized distributor for repair. EEC reserves the right
to use its discretion on replacing the faulty parts or replacing the assembly or the whole unit.
Except as provided herein, EEC makes no warranties to the purchaser of this instrument and all
other warranties, express or implied (including, without limitation, merchantability or fitness for
a particular purpose) are hereby excluded, disclaimed and waived.
• This product and its related documentation must be reviewed for familiarization with
safety markings and instructions before operation.
• Before applying power verify that the instrument is set to the correct line voltage and the
correct fuse is installed.
• When using an oscilloscope to measure DUT waveform, please refer description below to
avoid DUT, instrument and oscilloscope damages.
• When the output of AC source has N-G or L-G shorted, customer must use differential
isolation type of oscilloscope probe or using isolated oscilloscope.
• In order to avoid the interference, please DON'T combine the cables of input and output
from instruments in bundle with GPIB or RS232 cable.
To prevent accidental injury or death, these safety procedures must be strictly observed
when handling and using the test instrument.
OPERATING ENVIRONMENT
Temperatures:0° - 40° C (32-104F)
Relative humidity: 20% - 80%
Attitude: 2,000 meters (6,500 inches)
1
1.4 Storage and Packing
OPERATING ENVIRONMENT
Temperatures:0° - 40° C (32-104F)
Relative humidity: 20% - 80%
Altitude:: 2,000 meters (6,500 inches)
STORAGE ENVIRONMENT
This instrument may be stored or shipped in environments with the following limits:
Temperature......................... -40° to +75°C
Altitude: …………………… 7,620 meters (25,000 inches)
The instrument should also be protected against temperature extremes, which may cause
condensation within the instrument.
PACKING
Original Packing
Please keep all original shipping carton and packing materials and if it is needed to send for
repair, then it can be used for repacking purpose. Kindly contact the EEC Customer Support
department for service assistance. Upon delivering goods for repair, please attach along the
power cord cable and other accessories along with the instrument and state clearly the actual
problem encountered. Nevertheless, it is necessary to stick on the FRAGILE stickers on the
packing carton.
Other Packing
For using other packing cartons other than the original shipping carton, please follow the
packing requirement accordingly if possible:
1. Pack the instrument with polystyrenes or air-compressed plastics.
2. The spaces between the instrument and carton must be occupied with anti-shock sponges
and the thickness should at least 70mm-100mm (3-4 inches).
3. Seal neatly on top of the carton filled with instrument.
4. Stick on FRAGILE stickers for delivery.
2
CHAPTER 2: SPECIFICATIONS
2.1 Specification
Model 6310 6315 6330 6345 6360 6390 63120
OUTPUT
Phase 3Ø
1Ø 3.33 kVA 5 kVA 10 kVA 15 KVA 20 kVA 30 kVA 40 kVA
Power Rating
Total 9.99 kVA 15 kVA 30 kVA 45 KVA 60 kVA 90 kVA 120 kVA
Power 1Ø 3.33 kVA 5 kVA 10 kVA 15 KVA 20 kVA 30 kVA 40 kVA
Rating(Option
1 Ø 3W) Total 6.66 kVA 10 kVA 20 kVA 30 KVA 40 kVA 60 kVA 80 kVA
0 - 150 V / 0 - 300 V or 0 - 175 V / 0 - 350 V Selectable (Option 0 - 350 V) or 0 - 300 V / 0
Range
- 600 V Selectable (Option 0 - 600 V)
Voltage1 Resolution 0.1 V
Accuracy ± (1% of setting + 2counts)
Max. Current 0 - 150 V 28 A 42.0 A 84.0 A 126.0 A 168.0 A 252.0 A 336.0 A
(r.m.s)2- 1 Ø 0 - 300 V 14.0 A 21.0 A 42.0 A 63.0 A 84.0 A 126.0 A 168.0 A
Max. Current 0 - 175 V 24.0 A 36.0A 72.0A 108.0A 144.0A 216.0A 288.0 A
(r.m.s)3 for
option 0 - 350
0 - 350 V 12.0 A 18.0A 36.0A 54.0A 72.0A 108.0A 144.0 A
V-1Ø
Max. Current 0 - 300 V 14.0 A 21.0 A 42.0 A 63.0 A 84.0 A 126.0 A 168.0 A
(r.m.s)4 for
option 0 - 600 0 - 600 V 7.0 A 10.5A 21.0 A 31.5 A 42.0 A 63.0 A 84.0 A
V-1Ø
45.0 - 70.0 Hz / 45.0 - 500.0 Hz (Option 45 - 500 Hz) / 360 - 440, 760 - 840 Hz (Option
Range
400 / 800 Hz) Full Range Adjust
Frequency5
Resolution 0.1 Hz / 1Hz
Accuracy ± 0.02% of setting
Total Harmonic Distortion < 1% for 110 / 220V, 50 / 60Hz for 300V and Option 350V at Resistive Load
(THD)6 < 1.2% for 220 / 440V, 50 / 60Hz for Option 600V at Resistive Load
Inrush Current 3 times rated Current(r.m.s)
Crest Factor 3 times rated Current(r.m.s)
Line Regulation ± 0.1 V
Load Regulation L - N : ± (0.5% of output + 0.5 V) / L - L : ± (1% of output + 1 V) at Resistive Load
Deviation of phase angle 120° ≤ ± 1° at Load Balance, 120° ≤ ± 2.5° at 100% Load Unbalance
Setting
Range 0.1 - 999.9s
Ramp up /
Resolution 0.1s
Ramp down
Accuracy ± (0.1% of setting + 0.2s)
INPUT
Phase 3Ø 4W
Voltage 220 (L - N) / 380 Vac (L - L) ±10%
Max. Current7 25 A 40 A 75 A 115 A 150 A 230 A 300 A
Frequency 47 - 63 Hz
Power Factor 0.8
MEASUREMENT
L - N Range 0.0 - 300.0 V / 0.0 - 600.0 V
L - L Range 0.0 - 520.0 V / 0.0 - 1040 V
Voltage
Resolution 0.1 V / 0.2 V
Accuracy ± (1% of reading + 2counts) at Voltage > 5 V
3
0.000 - 3.500
L 0.00 - 35.00 A
Range A
H 3.00 - 35.00 A 30.0 - 350.0 A
L 0.001 A 0.01 A
Resolution8
H 0.01 A 0.1 A
Current ± (1% of
L reading + ± (1% of reading + 5 counts)9
5counts)6
Accuracy
± (1% of
H reading + ± (1% of reading + 1 count)9
1count)6
0.000 - 3.500
L 0.00 - 35.00 A
Range A
H 3.00 - 35.00 A 30.0 - 350.0 A
L 0.001 A 0.01 A
Resolution8
Current H 0.01 A 0.1 A
(Option 0 - ± (1% of
600 V) L reading + ± (1% of reading + 5 counts)9
5counts)6
Accuracy
± (1% of
H reading + ± (1% of reading + 1 count)9
1count)6
Range 45.0 - 850.0 Hz
Frequency Resolution 0.1 Hz
Accuracy ± 0.1 Hz
L 0.0 - 350.0 W 0.000 - 3.500 kW
Range
H 300 - 4000 W 3.00 - 40.00 kW
L 0.1 W 0.001 kW
Power Resolution
H 1W 0.01 kW
L ± (1.5% of reading + 5 counts) at Voltage > 5 V
Accuracy
H ± (1.5% of reading + 1 count) at Voltage > 5 V
GENERAL
Remote Input Signal
ON / OFF, P1, P2, P3
Interface
8 Memories, 5steps per memory for Voltage, Frequency, Test Time, Delay time and
Memory
Current, Power Hi, Lo Limit setting
Sync Output Signal Yes
Display 4 group pf digital LED displays
Efficiency ≥80% ≥ 85%
Protection OCP, OVP, OPP, OTP, LVP, Short Circuit ; Alarm and shutdown
Interface10 Standard USB & RS232, Option GPIB & Analog card
Operation Environment 0 - 40 ℃ / 20 - 80% RH
800 x 1200 x
430 x 732 430 x 732 600 x 842 800 x 1557 1200 x 1557
Dimension (W x H x D), 1557 1557
(839) x 650 (839) x (949) x 900 (1662) x (1805) x 900
mm11 (1662) x (1805) x
(730) 650 (730) (980) 900 (980) (980)
900 (980) 900 (980)
600 x 800 x 1200 x
600 x 600 x 842 800 x 1557 1200 x 1557
Dimension (PFC)(W x H x D), 732(839) 1557 1557
732(839) x (949) x 900 (1662) x (1805) x 1100
mm11, 12 x (1662) x (1805) x
900(980) (980) 900 (980) (1180)
900(980) 900 (980) 900 (980)
Weight 160 kg 199 kg 373 kg 622 kg 645 kg 1139 kg 1659 kg
Weight (PFC) 299 kg 362 kg 547 kg 774 kg 909 kg 1505 kg 2139 kg
4
Model 63150 63180 63210 63240
OUTPUT
Phase 3Ø
1Ø 50 kVA 60 kVA 70KVA 80KVA
Power Rating
Total 150 kVA 180 kVA 210KVA 240KVA
Power 1Ø 50 kVA 60 kVA 70KVA 80KVA
Rating(Option
1 Ø 3W) Total 100 kVA 120 kVA 140KVA 160KVA
0 - 150 V / 0 - 300 V or 0 - 175 V / 0 - 350 V Selectable (Option 0 - 350 V) or 0 - 300 V /
Range
0 - 600 V Selectable (Option 0 - 600 V)
Voltage1 Resolution 0.1 V
Accuracy ± (1% of setting + 2counts)
Max. Current 0 - 150 V 420.0 A 504.0 A 588.0A 672.0A
(r.m.s)2- 1 Ø 0 - 300 V 210.0 A 252.0 A 294.0A 336.0A
Max. Current 0 - 175 V 360.0A 432.0A 504.0A 576.0A
(r.m.s)3 for
option 0 - 350
0 - 350 V 180.0A 216.0A 252.0A 288.0A
V-1Ø
Max. Current 0 - 300 V 210.0 A 252.0 A 294.0A 336.0A
(r.m.s)4 for
option 0 - 600 0 - 600 V 105.0 A 126.0 A 147.0A 168.0A
V-1Ø
45.0 - 70.0 Hz / 45.0 - 500.0 Hz (Option 45 - 500 Hz) / 360 - 440, 760 - 840 Hz (Option
Range
400 / 800 Hz) Full Range Adjust
Frequency5
Resolution 0.1 Hz / 1Hz
Accuracy ± 0.02% of setting
Total Harmonic Distortion < 1% for 110 / 220V, 50 / 60Hz for 300V and Option 350V at Resistive Load
(THD)6 < 1.2% for 220 / 440V, 50 / 60Hz for Option 600V at Resistive Load
Inrush Current 3 times rated Current(r.m.s)
Crest Factor 3 times rated Current(r.m.s)
Line Regulation ± 0.1 V
Load Regulation L - N : ± (0.5% of output + 0.5 V) / L - L : ± (1% of output + 1 V) at Resistive Load
Deviation of phase angle 120° ≤ ± 1° at Load Balance, 120° ≤ ± 2.5° at 100% Load Unbalance
Setting
Range 0.1 - 999.9s
Ramp up /
Resolution 0.1s
Ramp down
Accuracy ± (0.1% of setting + 0.2s)
INPUT
Phase 3Ø 4W
Voltage 220 (L - N) / 380 Vac (L - L) ±10%
Max. Current7 375 A 450 A 525A 600A
Frequency 47 - 63 Hz
Power Factor 0.8
MEASUREMENT
L - N Range 0.0 - 300.0 V / 0.0 - 600.0 V
L - L Range 0.0 - 520.0 V / 0.0 - 1040 V
Voltage
Resolution 0.1 V / 0.2 V
Accuracy ± (1% of reading + 2counts) at Voltage > 5 V
L 0.0 - 350.0 A
Current Range
H 300 - 504A 300 - 840A
5
Resolution L 0.1 A
8 H 1A
L ± (1% of reading + 5 counts)9
Accuracy
H ± (1% of reading + 1 count)9
L 0.0 - 350.0 A
Range
H 300 - 504A 300 - 840A
Current
(Option 0 - Resolution L 0.1 A
600 V) 8 H 1A
L ± (1% of reading + 5 counts)9
Accuracy
H ± (1% of reading + 1 count)9
Range 45.0 - 850.0 Hz
Frequency Resolution 0.1 Hz
Accuracy ± 0.1 Hz
L 0.00 - 35.00 kW
Range
H 30.0 - 60.0 kW 30.0 - 100.0kW
L 0.01 kW
Power Resolution
H 0.1 kW
L ± (1.5% of reading + 5 counts) at Voltage > 5 V
Accuracy
H ± (1.5% of reading + 1 count) at Voltage > 5 V
GENERAL
Remote Input Signal Interface ON / OFF, P1, P2, P3
8 Memories, 5steps per memory for Voltage, Frequency, Test Time, Delay time and
Memory
Current, Power Hi, Lo Limit setting
Sync Output Signal Yes
Display 4 group pf digital LED displays
Efficiency ≥ 85%
Protection OCP, OVP, OPP, OTP, LVP, Short Circuit ; Alarm and shutdown
Interface10 Standard USB & RS232, Option GPIB & Analog card
Operation Environment 0 - 40 ℃ / 20 - 80% RH
6
【Ordering Information】
• 6310 High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (9kVA)
• 6310P High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (9kVA) with PFC
• 6315 High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (15kVA)
• 6315P High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (15kVA) with PFC
• 6330 High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (30 kVA)
• 6330P High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (30 kVA) with PFC
• 6345P High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (45 kVA) with PFC
• 6360 High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (60 kVA)
• 6360P High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (60 kVA) with PFC
• 6390 High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (90 kVA)
• 6390P High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (90 kVA) with PFC
• 63120 High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (120 kVA)
• 63120P High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (120 kVA) with PFC
• 63150 High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (150 kVA)
• 63150P High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (150 kVA) with PFC (2 case)
• 63180 High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (180 kVA)
• 63180P High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (180 kVA) with PFC (2 case)
• 63210 High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (210 kVA)
• 63210P High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (210 kVA) with PFC (2 case)
• 63240 High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (240 kVA)
• 63240P High Power Programmable 3φ AC Power Source 0 - 300 V / 45 - 70Hz (240 kVA) with PFC (2 case)
• Opt.608 600 V AC Output for models 6310, 6315 & with PFC models
• Opt.609 600 V AC Output for models 6330, 6345, 6360 & with PFC models
• Opt.610 600 V AC Output for models 6390, 63120 & with PFC models
• Opt.615 Output Frequency 45 - 500 Hz Full Range Adjust for models 6310 & with PFC models
• Opt.616 Output Frequency 45 - 500 Hz Full Range Adjust for models 6315 & with PFC models
• Opt.617 Output Frequency 45 - 500 Hz Full Range Adjust for models 6330 & with PFC models
• Opt.637 Single Phase 3 Wire Output (L1 / L2 / N)
• Opt.649 600 V AC Output for models 63150, 63180 & with PFC models
• Opt.650 Remote Sensing
• Opt.668 suppression Input Voltage
• Opt.671 RCP function for model 6310, 6315, 6330 & with PFC models
• Opt.672 RCP function for models 6345, 6360, 6390, 63120, 63150, 63180, 63210, 63240 & with PFC models & 63300P
• Opt.673 Earthquake proof for models 6310, 6315, 6330, 6345, 6360 & with PFC models
• Opt.674 Output voltage 350V
• Opt.675 Output frequency 400 / 800Hz only for without PFC models
• Opt.676 600V AC Output for models 63210, 63240 & with PFC models & 63300P
7
CHAPTER 3: PANELS
1 21 22 2 23 24 3 25 26 4 27 28
36 35 34 33 38 37 32 31 33 34 30 29
20 19 18 17 16 15 14 13 12 11 10 9 7 8 6 5
1. Frequency Display. When the output is OFF, it shows the frequency setting. Otherwise, it
shows the frequency of the output.
2. Voltage Display. When the output is OFF it shows the output voltage setting. Otherwise it
shows the voltage of the output.
3. Current Display. . When the output is OFF it shows the output current setting. Otherwise
it shows the current of the output.
5. Meter Key. Select the display of output wattage (P) or power factor (PF) value.
6. LOCK/LOCAL Key. To disable all the keys on the front panel (ON/OFF) and to escape from
programmable operation mode.
7. P1 Button. Store setting memory [Press and hold for a second or above]
8. P2 Button. Store setting memory [Press and hold for a second or above]
9. ︿ Key. To increase high limit of current, as well as SYSTEM and PROGRAM parameters.
10. ﹀ Key. To decrease high limit of current, as well as SYSTEM and PROGRAM parameters.
11. SYSTEM Key. Enter or exit from the setting of SYSTEM parameter.
12. PROGRAM Key. Enter or exit from the setting of PROGRAM parameter.
8
13. ︿ Key. To increase the output voltage higher than the display value.
14. ﹀ Key. To decrease the output voltage lower than the display value.
15. Range Key. High Voltage Range: 0-300V, Low Voltage Range: 0-150V.
(If output voltage is 0-600V, the High Voltage Range is 0-300V and Low Voltage Range is 0-
150V.)
16. PHASE Key. To select either phase of R,S,T or Σ for display and setting.
17. ︿ Key. To increase the output frequency higher than the display value.
18. ﹀ Key. To decrease the output frequency lower than the display value.
19. 50/60 Hz Key. If the rating output frequency is 45~70Hz or optional 45~500Hz, press it will
be set the output frequency to 50/60 Hz. If the rating output frequency is optional 400Hz,
it will be disabled.
20. TEST/RESET Key. To turn on/off the output, or start/abnormal operation occurs.
21. TEST Indicator. When this LED is ON, the output voltage is turned ON.
22. PASS Indicator. When this LED is ON, an automatic output operation has passed.
23. R Indicator. When this LED is ON, the display shows the R-phase output.
24. S Indicator. When this LED is ON, the display shows the S-phase output.
25. LOCK Indicator. When this LED is lit, all the keys are disabled.
26. REMOTE Indicator. When this LED is ON, the AC Power Source is controlled by GPIB
(RS232) or PLC control.
27. P Indicator. When this LED is ON, the MULTIMETER display shows the power value.
28. PF Indicator. When this LED is ON, the MULTIMETER display shows the power factor
value.
29. T Indicator. When this LED is ON, the MULTIMETER display shows the time duration.
30. P-S Indicator. When this LED is ON, the MULTIMETER display shows the current memory
of Program-Step.
9
33. T Indicator. When this LED is ON, the display shows the T-phase output.
34. Σ Indicator. When this LED is ON, the display shows the Σ-phase output.
36. PROTECT Indicator. When the LED is ON, there is a protection at abnormal
37. SYSTEM Indicator. Indicates the current setting mode is at SYSTEM parameter setting.
38. PROGRAM Indicator. Indicates the current setting mode is at PROGRAM parameter
setting.
10
3.2 Panel Description
Panel Description
2 5 6 4 3 2
1. Handle
2. Caster Wheels
3. Front Panel Output Switch
4. Universal AC Output Socket (15A)
5. Input Power Switch
6. Ventilation Inlet
11
1
2 6 10 7 12 9 8 11 2
12
CHAPTER 4: MANUAL OPERATION
1. Power On Condition
Every time when turning on the AC Power Source, the voltage and current displays will
indicate model and version respectively for the particular AC Power Source. After that the
CPU will provide a beep sound to show the pervious setting.
Model Version
Press and hold “︿”, decimal (clear to 0) → 1st Integer (0.3sec / step) → 2nd Integer (0.3sec / step) →
3rd Integer (0.1sec / step)
Press and hold “﹀”, decimal (clear to 0) → 1st Integer (0.3sec / step) → 2nd Integer (0.3sec / step) →
3rd Integer (0.1sec / step)
Note: Option 45.0-500.0Hz, for frequency setting in the range of 45-99.9 Hz, each change on
the setting are 0.1Hz/step for normal setting and 1Hz/step for coarse setting. In 100-500
Hz range, each change on the setting are 1Hz/step for normal setting and 10Hz/step for
coarse setting.
13
4. Setting Of Voltage Range
If the output voltage is from 0 to 300 V and the desired voltage is lower than 150V, press the
0 – 150 V Range Key first. The LED is lighted to indicate the AC source is in 0-150 Volt range
and allow a higher range of current limit. If the output voltage is from 0 to 600 V and the
desired voltage is lower than 300V, press the 0 – 300 V Range Key first. The LED is lighted to
indicate the AC source is in 0-300 Volt range and allow a higher range of current limit. For the
higher voltage range, the range of current limit drops to half as comparing the current limit at
the lower range (Refer to the specification table). Setting of voltage range may not influence
the existing output voltage setting. Anyhow, changing the voltage range during the OUTPUT
indicator is ON will lead to the output voltage cutoff (about 20 ms) and this should be avoided
if possible. The instrument will also not accept any invalid change of range such as changing to
low voltage range when the output voltage is set to 200V.
5. Setting Of Current
During the RESET or TEST condition for the AC Power Source, the desire current limit can be
set by pressing “︿” or “﹀” key below the Current display. While the OUTPUT is not activated,
pressing the “︿” or “﹀” key will enter to the setting mode of low-limit current. Press further
“︿” or “﹀” key to continue scrolling the display in order to change the current limit. If the
setting is hold after 2 minutes, it will return to a standby mode from current setting mode.
Meanwhile, at the standby mode, the current limit is based on output capacity (refer to 2.1
Specification) to limit the output current for protection. Upon exceeding 110% of output
capacity, an error message, OCP, will be showed at Frequency display. When the actual
current has exceeded the high limit of current setting, the AC Power Source will cut off the
output immediately and a HI-A error message is shown.
6. Meter Key
Pressing the METER key will enable the MULTIMETER display to indicate Power (Watts) or
Power Factor measurement, Testing Time (T) and Programmable Memory-Step (P-S), for
example, P2-3, represents 2nd group of memory and 3rd step).
7. Executing Output
Pressing TEST key will turn ON the output voltage or initiate an automatic output operation.
First ensure all settings have been done accordingly before pressing the TEST button. When
the TEST indicator is lit, the desire output voltage is being supplied to the universal socket on
front panel and output terminal on rear panel.
If the indicator is blinking that shows an abnormal operation is encountered, thereby the
output voltage will be cut off immediately. By pressing the OUTPUT/RESET key will reset the
audible alarm, an error message is indicated.
8. Terminating Output
Upon reaching to complete the testing time, the output operation will terminate
automatically, and while the testing time (test - t) is set to "Cont", the output operation will
14
continue until a RESET key from the front panel has been pressed. When the output operation
is terminating, the TEST indicator will turn OFF, and meanwhile, as the automatic output
operation is tested successfully when the Auto-run-On is turning ON, a beep sound will alert.
9. Alarm Reset
Whenever the AC Power Source experiences abnormal operations such as overload, short
circuit, over temperature or exceeding high and low limits of current, power or power factor,
the CPU will cut off the output voltage and alert an alarm to user. In this case, the FAIL
indicator will blink while the PROTECT indicator will lit (for overload, short circuit and over
temperature). The alarm can be reset by pressing RESET key for once. Pressing the key again
will then return the error message display to the standby screen. Note: Kindly verify the cause
of error before pressing the TEST key to execute the output operation again.
10.Front-Panel Lockout
Press the LOCK/LOCAL key will lead to the LOCK indicator to be lit, in order to disable all keys
on the front panel (except METER key). Further pressing the LOCK/LOCAL key again it will
reactivate all keys to function normally. The Lockout feature is to prevent any unauthorized
alteration.
11.Plc Remote
Able to control the output voltage ON/OFF by using PLC Remote from the 9-pins connector
available on the rear panel. It can also perform the output control for 3 programmable
memories, which are P1, P2 and P3.
At RESET condition, press “SYSTEM” key to enter SYSTEM parameter setup and SYSTEM
indicator will light up. The indication of SYSTEM setup menu will be shown at the Voltage
display and the setup parameter for each SYSTEM setup will be shown at Current display. By
pressing the “︿” or “﹀” key from the Voltage display, it then allows to scroll the SYSTEM
parameter setup menu; Auto run, PLC Remote, GPIb Addr, Auto Adj, P–UP, t Unit, LooP rAtE,
FrEq HI, FrEq LO, Volt HI, and Volt LO settings. Pressing the “︿” or “﹀” key from Current
display allows varying the parameter value of the system setup. Press SYSTEM key again to
return to standby condition after the system parameter setup has completed.
15
System Parameter Menu
16
2. PLC Remote
The Voltage display will indicate “PLC” and by pressing the “︿” or “﹀” key from the Current
display to select ON/OFF for the PLC Remote. As the PLC Remote setting is OFF, the control of
operation is via the front panel. Meanwhile, when the setting is ON, then the operation will be
controlled by a remote control box from the rear panel. Once the setting is ON, the Current
display will indicate “PLC-ON” and the buzzer will sound beeps twice whenever a key is being
pressed. At this time, all keys at front panel are disabled except the LOCK/LOCAL, SYSTEM and
METER keys are functioning as normal.
6. Power- Up Status
While the Voltage display indicates “P-UP”, press the “︿” or “﹀” key from the Current
display to select ON/OFF/LAST for the output status. As the Power-Up is turned OFF, the
output is set to be at OFF condition after turn OFF or ON again the AC source. When the
Power-Up is ON, then the output is instantly ON after the AC Power Source is being OFF and
ON again. When the Power-Up is set to LAST, the display will indicate last POWER OFF setting
status. Note: This setup will only in effective after turning off and on the AC Power Source.
17
9. Frequency Hi Limit Setting
The Frequency display will indicate “FrEq” and the Voltage display shows “HI”. Press the “︿”
or “﹀” key from the Current display to adjust the high limit frequency within the range of
47.0-63.0Hz. This feature disallows user to vary the output frequency during RESET and TEST
conditions, which may cause to exceed the acceptable range of frequencies for any particular
DUTs. Note: Else two ranges: 45.0-500.0Hz and 400Hz for option.
18
4.3 Program Parameter Setting
At RESET condition, press “PROGRAM” key to enter PROGRAM parameter setup whereby the
“PROGRAM LED” indicator is lit. In the meanwhile, the indication of PROGRAM setup menu
will be shown at the Voltage display and the setup parameter for each PROGRAM setup will
be shown at Current display. By pressing the “︿” or “﹀” key from the Voltage display, it then
allows to scroll the PROGRAM parameter setup menu; Prog, Loop CyCL, StEP, Volt, FrEq, tESt t,
dLAy t, A-HI, A-LO, P-HI, P-LO, PF-H, PF-L, Conn StEP. Pressing the “︿” or “﹀” key from
Current display allows varying the parameter value of the PROGRAM setup. Press PROGRAM
key again to return to standby condition after the PROGRAM parameter setup has completed.
Program Parameters
19
1. Program Memory Selection
When the voltage display indicates “Prog” under PRORAM parameter setup, press “ ︿ “ or
“ ﹀ “ key to select “1 ~ 8” of memories. If the program memories (P1~P8) have to be
interconnected, it can be achieved by turning ON the “Auto run”. Each step of a memory can
be linked to the next step up to 400 steps (8 Programs × 5 Steps) in sequence by turning ON
the “Conn StEP”. All different settings on voltage, frequency, testing time, current, power, and
power factor high/low limits can be stored at each step individually.
20
2. Auto Loop Cycle Setting
The Frequency display will indicate “LooP” and the Voltage display shows “CyCL”. Press the
“︿” or “﹀” key from the Current display to select “Cont”, “OFF” and “2”~”999”. Setting of
“Cont” means continuous, and “OFF” is single operation. For “2”~“9” setting, it becomes the
number of sequence to be repeated for running operation. This selected loop cycle will then
be multiplied by the “LooP CyCL” of the SYSTEM parameter, which acts as a multiplier (×1, ×10
or ×100) to the selected value of ”LooP CyCL”. (For example, Loop Cycle is set to “2” and the
Loop Rate has selected “×10”. Therefore, the Auto run operation will execute for 20 times.)
21
judgment time is set to assign time period that will not make any judgment to all limits (Ex:
A,P,PF High/Low Limit Setting).
Note: The delay judgment time may not longer than the testing time. If the testing time has
been selected to “Cont”, the delay judgment time will be fixed to “0”, and therefore, the run
operation is always carried out with the limit judgments.
22
4.4 Displayed Messages
Below are the descriptions of error messages that may occur at abnormal conditions:
At any abnormal conditions, there are several error messages to be indicated on the display
LEDs. Thereafter the output will be disabled and the alarm will sound. The “FAIL” and
“PROTECT” indicators will also light up at the same time. If the indicator is blinking that shows
an abnormal operation is encountered, thereby the output voltage will be cut off immediately.
Press the RESET key to reset the audible alarm. The error message is eliminated and returns to
the RESET condition when the RESET key is pressed again.
All error messages are occurred at any abnormal conditions and therefore
must be recorded. Check the cause of error to ensure the problem is
eliminated before restarting the operation, or contact EEC or our official distributors for
further assistance.
If the input voltage is below the CPU setting level when turning on the AC Power Source, the
Voltage display will indicate “FAIL” and the alarm will sound. There is no effect to reset this
error message by pressing any keys. It needs to turn off the input power in order to reset the
alarm.
If the output voltage has exceeded 5V of setting voltage at 0-150V range or has exceeded 10V
of setting voltage at 0-300V range, the Frequency display will indicate “OVP” and the alarm
will sound. Consequently, the “FAIL” indicator will blink while the “PROTECT” indicator will
light up. The Voltage and Current displays will show the overloaded voltage and current
respectively.
If the output current has exceeded 110% of maximum current rating, the Frequency display
will indicate “OCP” and the alarm will sound. Consequently, the “FAIL” indicator will blink
while the “PROTECT” indicator will light up. The Voltage and Current displays will show the
overloaded voltage and current respectively.
If the output voltage is under 5V of setting voltage at 0-150V range or under 10V of setting
voltage at 0-300V range for 3 second, the Frequency display will indicate “LVP” and the alarm
will sound. Consequently, the “FAIL” indicator will blink while the “PROTECT” indicator will
light up. The Voltage and Current displays will show the overloaded voltage and current
respectively.
23
Model(6310、6315、6330、6360、6390、63150、63180、63210、63240、63300)
If the output power has exceeded 105-110% of rating power for 5 second or the output power
has exceeded 110% under 1 second, the Frequency display will indicate “OPP” and the alarm
will sound. Consequently, the “FAIL” indicator will blink while the “PROTECT” indicator will
light up. The Voltage and Current displays will show the overloaded voltage and current
respectively.
Model(63120)
If the output power has exceeded 100% of rating power for 10s, the Frequency display will
indicate “OPP” and the alarm will sound. Consequently, the “FAIL” indicator will blink while
the “PROTECT” indicator will light up. The Voltage and Current displays will show the
overloaded voltage and current respectively.
If the output current has exceeded the high limit current, the Frequency display will indicate
“HI-A” and the alarm will sound. Consequently, the “FAIL” indicator will blink and the Voltage
or Current displays will show the overloaded voltage or current respectively.
If the output current is below the low limit current, the Frequency display will indicate “LO-A”
and the alarm will sound. Consequently, the “FAIL” indicator will blink and the Voltage or
Current displays will show the overloaded voltage or current respectively.
If the output power has exceeded the high limit power, the Frequency display will indicate
“HI-P” and the alarm will sound. Consequently, the “FAIL” indicator will blink and the Voltage
or Current displays will show the overloaded voltage or current respectively.
If the output power is below the low limit power, the Frequency display will indicate “HI-P”
and the alarm will sound. Consequently, the “FAIL” indicator will blink and the Voltage or
Current displays will show the overloaded voltage or current respectively.
24
If the output power factor has exceeded the high limit power factor, the Frequency display
will indicate “HI-PF” and the alarm will sound. Consequently, the “FAIL” indicator will blink
and the Voltage or Current displays will show the overloaded voltage or current respectively.
If the output power factor is below the low limit power factor, the Frequency display will
indicate “LO-PF” and the alarm will sound. Consequently, the “FAIL” indicator will blink and
the Voltage or Current displays will show the overloaded voltage or current respectively.
If the input voltage has reached below the 20% of the operating voltage, the Frequency
display will indicate “Lodc” and the alarm will sound, since the input voltage has gone below
the requirement range of operating voltage for this AC Power Source. Consequently, the
“FAIL” indicator will blink while the “PROTECT” indicator will light up. The Voltage and Current
displays will show the overloaded voltage and current respectively.
If the input voltage exceeds the 20% of the operating voltage, the Frequency display will
indicate “HIdc” and the alarm will sound, since the input voltage has reached above the
requirement range of operating voltage for this AC Power Source. Consequently, the “FAIL”
indicator will blink while the “PROTECT” indicator will light up. The Voltage and Current
displays will show the overloaded voltage and current respectively.
If the input voltage cuts off for a short period and return as normal, the Frequency display will
indicate “AcLP”. This shows that an abnormal input voltage has occurred and thus the alarm
will sound. Consequently, the “FAIL” indicator will blink while the “PROTECT” indicator will
light up. The Voltage and Current displays will show the overloaded voltage and current
respectively.
If the fuse is blown, the Frequency display will indicate “FUSE”. This shows that the input
voltage has exceeded the fuse protection rating and causes the fuse to blow and thus the
alarm will sound. Consequently, the “FAIL” indicator will blink while the “PROTECT” indicator
will light up. The Voltage and Current displays will show the overloaded voltage and current
respectively.
25
If the IGBT is overloaded, the Frequency display will indicate “IGbt” and the alarm will sound.
Consequently, the “FAIL” indicator will blink while the “PROTECT” indicator will light up. The
Voltage and Current displays will show the overloaded voltage and current respectively.
If the heat sink of the instrument itself has exceeded 110 C, the Frequency display will
indicate “OtP”. This shows that the heat sink is overheated and thus the alarm will sound.
Consequently, the “FAIL” indicator will blink while the “PROTECT” indicator will light up. The
Voltage and Current displays will show the overloaded voltage and current respectively.
26
CHAPTER 5: DESCRIPTION OF OPERATION
Introduction
The RS232 connection is configured as follows for a 9 pins Serial Port interface:
The COM port should have the following configuration: 9600 baud, 8 data bits, 1 stop bit, no
polarity, and etc. This interface does not support XON/XOFF protocol or any hardware
handshaking. The controller should be configured to ignore the Handshaking Lines DTR (PIN 4),
DSR (PIN 6) and RTS (PIN 9). If the port cannot be configured through software to ignore the
lines, the handshake lines should be then jumped together in two different sets. The PIN 4 and
6 jumped together while PIN 7 and 8 jumped together at the controller end of the cable.
When sending command over the RS232 bus the AC Power Source will send a response string
of 06 Hex or 6 decimal and the Acknowledge (ACK) ASCII control code if the transfer was
recognized and completed by the AC Power Source. If there is an error with the command
string that is sent, the AC Power Source will respond with 15 Hex or 21 decimal and the Not
Acknowledge (NAK) ASCII code. The ACK or NAK response allows for software handshaking, in
order to monitor and control data flow. When requesting data from the AC Power Source, it
will automatically send the data back to the controller input buffer. The controller input buffer
will accumulate data being sent from the AC Power Source including the ACK and NAK
response strings, until the controller has read it. When the strings or command has been sent,
it must be terminated with EOI LF=(0AH), such as “TEST”+LF.
27
5.2 GPIB Interface
This interface of a device connected to the BUS is specified by its interface functions. These
functions provide the means for a device to receive, process, and send messages over the bus.
The interface functions are listed in the chart below:
GPIB Address
All the above commands are required by the IEEE-488.2 GPIB Interface BUS. The factory
default address of the 6300 Series AC Power Source is set to 8.
28
5.3 RS-232/GPIB Interface Command List
The following is the commands list for ON/OFF output operation and all other parameter
settings. When programming these commands, it does not require any parameter values to be
included with the command. Before using the commands of specific functions into a program,
it must first create a file name (COMMAND: FN nn,xxxx) and the selected step to load into
RAM (COMMAND:SS nn), in order to enter into each specific parameter that can be edited.
When the strings or command has been sent, it must be terminated with EOI.
Basic Commands
Command Value Description
TD? All present displayed values Query *
TDR? Readings of phase R Query *
TDS? Readings of phase S Query *
TDT? Readings of phase T Query *
TDC? Readings of phase Σ Query *
TEST Power Output
RESET Shut down
PHASE X 0 Display readings of phase R
1 Display readings of phase S
2 Display readings of phase T
3 Display readings of phase Σ
PHASE? Displayed phase Query
METER X 0 Power Reading
1 Power Factor Reading
2 Timer Reading
3 Program -Step
METER? Displayed parameter of meter Query
RANG X 0 Low 0 - 150V (or Low 0-300V for Option 0-600V)
1 High 0 - 300V (or High 0-600V for Option 0-600V)
RANG? Setting Range Query
29
Program Commands
COMMAND Value Description
PROG X 1-8 Program Number
PROG? Program Number Query
AR X 0 Auto Run Off
1 Auto Run On
AR? Auto Run setting Query
LC XXXX 0 Loop Cycle “Continuous”
1 Loop Cycle “Once”
2 - 9999 Loop Cycle 2 - 9999
LC? Loop Cycle Query
30
System Commands
COMMAND Value Description
31
IEEE 488.2 Common Commands
Command Value Description Acknowledgement
Company, Model Number, Serial
*IDN? Identification Query
Number, Firmware Revision
*RST Reset Command
*TST? Self-Test Query 01H=OK
Clear Standard Event Status Register and
*CLS
Service Request Register
The instrument saves each parameter in non-volatile memory when the file save "FS"
command is used to save the file. The non-volatile memory has a limited write cycle life,
32
therefore for programmers who wish to send all parameters before executing each test the
"FS" command should not be used. The parameters will be stored in the CPU's Random
Access Memory (RAM) until another memory location is selected. However, settings written
to RAM from GPIB/RS-232 mode will be lost when power is shut down. Parameter changes to
RAM are unlimited and will not effect the life of the internal non-volatile memory chip.
33
CHAPTER 6: DESCPIPTION OF APPLICATION
There is a remote interface connector attached on the rear panel of 6300 Series AC Power
Source and thus can control any test operation via remote once the PLC Remote setting from
SYSTEM setup parameter is turned on. The 9-pin D-type connector consists of I/O signals for
ON/OFF, 3 memories (P1, P2, P3) input control and “SYNC” output signal.
D4148
PLC Remote
34
6.2 Analog Card (Option)
This function provides two BNC connections, one for the control voltage and the other for the
control frequency. User can input 0 - 10Vdc from outside to control the output voltage and
frequency. If user want to control the voltage, then user must to connect the external input
signal to EXT. V. When voltage range at high range, user input 5Vdc that AC source will output
150Vac. The frequency is the same way.
35
CHAPTER 7: CALIBRATION PROCEDURE
The calibration procedure is with model 6330 and output voltage 0-300V for its example.
1. All the AC Power Sources are calibrated at the factory before delivery. Unless
necessarily, kindly do not re-calibrate within the first 12 months.
2. Ensure the model version is correct before turn OFF the AC Power Source. Press and
hold “LOCK/LOCAL” key while turn ON the AC Power Source. After two seconds, the AC
Power Source will indicate model number and version. It is now at calibration mode
and press “︿” or “﹀” key under the Frequency display to select a calibration menu.
3. Low Range Voltage (V-LO) Calibration: Press “︿” or “﹀” key from the Frequency
display to select “V-LO". Connect a calibrated true RMS Voltmeter on one of the
output socket and press TEST key in order to activate the CPU to read a low range
offset voltage and send a 150VAC output voltage. The accurate RMS Voltmeter will
indicate an actual voltage value while the Current display on the AC Power Source will
indicate a necessary value needed to be calibrated. Using the voltage value from the
RMS Voltmeter, press “︿ “or “﹀“ key from the Current display to scroll the display
which indicate the current voltage value, in order to set the voltage according to the
RMS Voltmeter. After it is done, press “LOCK/LOCAL” key to save the data stored and
the low range voltage calibration is completed.
4. High Range Voltage (V-HI) Calibration: Press “︿” or “﹀” key from the Frequency
display to select “V-HI". Connect a calibrated true RMS Voltmeter on one of the output
socket and press TEST key in order to activate the CPU to read a high range offset
voltage and send a 300VAC output voltage. The accurate RMS Voltmeter will indicate
an actual voltage value while the Current display on the AC Power Source will indicate
a necessary value needed to be calibrated. Using the voltage value from the RMS
Voltmeter, press “︿” or “﹀” key from the Current display to scroll the display, which
indicates the current voltage value, in order to set the voltage according to the RMS
Voltmeter. After it is finished, press “LOCK/LOCAL” key to save the data stored and
the high range voltage calibration is completed
36
5. Low Range Current (A-LO) Calibration: Press “︿” or “﹀” key from the Frequency
display to select “A-LO". Connect a variable resistor as load and a calibrated true RMS
Ammeter on one of the output socket and press TEST key in order to activate the CPU
to read a low range offset current and send a 120VAC output voltage. The accurate
RMS Ammeter will indicate an actual current value while the Current display on the AC
Power Source will indicate a necessary value needed to be calibrated. Adjust the load
or voltage to get the reading from the ammeter to be 30A. Using the current value
from the RMS Ammeter, press “︿” or “﹀” key from the Current display to scroll the
display, which indicates the current value, in order to set the current according to the
RMS Ammeter. After it is done, press “LOCK/LOCAL” key to save the data stored and
the low range current calibration is completed.
6. High Range Current (A-HI) Calibration: Press “︿” or “﹀” key from the Frequency
display to select “A-HI". Connect a variable resistor as load and a calibrated true RMS
Ammeter on one of the output socket and press TEST key in order to activate the CPU
to read a high range offset current and send a 120VAC output voltage. The accurate
RMS Ammeter will indicate an actual current value while the Current display on the AC
Power Source will indicate a necessary value needed to be calibrated. Adjust the load
or voltage to get the reading from the ammeter for the output current value. Using the
current value from the RMS Ammeter, press “︿” or “﹀” key from the Current display
to scroll the display, which indicates the current value, in order to set the current
according to the RMS Ammeter. After it is done, press “LOCK/LOCAL” key to save the
data stored and the high range current calibration is completed.
37
7. Low Range Power (P-LO) Calibration: Press “︿” or “﹀” key from the Frequency
display to select “P-LO". Connect a variable resistor as load and a calibrated true RMS
Power Meter on one of the output socket and press TEST key in order to activate the
CPU to read a low range offset power and send a 120VAC output voltage. The accurate
RMS Power meter will indicate an actual power value while the Current display on the
AC Power Source will indicate a necessary value needed to be calibrated. Adjust the
load or voltage to get the reading from the RMS Power Meter to be 3KW. By using the
power value from the RMS Power Meter, press “︿” or “﹀” key from the Current
display to scroll the display, which indicates the power value, in order to set the power
according to the RMS Power Meter. After it is done, press “LOCK/LOCAL” to save the
data stored and the low range power calibration is completed.
8. High Range Power (P-HI) Calibration: Press “︿” or “﹀” key from the Frequency
display to select “P-HI". Connect a variable resistor as load and a calibrated true RMS
Power Meter on one of the output socket and press TEST key in order to activate the
CPU to read a high range offset power and send a 120VAC output voltage. The
accurate RMS Power Meter will indicate an actual power value while the Current
display on the AC Power Source will indicate a necessary value needed to be calibrated.
Adjust the load or voltage to get the reading from the power meter for the output
power value. Using the power value from the RMS Power Meter, press “︿” or “﹀”
key from the Current display to scroll the display, which indicates the power value, in
order to set the power according to the RMS Power Meter. After it is done, press
“LOCK/LOCAL” to save the data stored and the high range power calibration is
completed.
9. Press “︿” or “﹀” key from the Frequency display to select the R, S or T phase for
calibration. Each calibration items for each phase (R, S, or T) from step 3~7 of the
38
above is non-related to one another. If the calibration has to be terminated half way,
user can just press RESET key to exit the calibration mode.
10. After the calibration is completed, turn OFF the AC Power Source.
39
CHAPTER 8: APPENDIX
USERS’ PROTECTION
To avoid electric shock, do not dismantle the cover of the AC Power Source. When any
abnormal symptom happens on the instrument, please kindly contact EEC office or the
authorized local distributor for further assistance. Attached block diagram and wiring
diagrams of 6300 Series AC Power Source are for reference purpose.
CONSISTANCY OF SERVICE
6300 Series AC Power Source with the input circuit and all related parts are required to be
checked and calibrated at least once every year. This is to protect user in terms of safety and
to ensure a high accuracy of this AC Power Source all times.
USERS’ MODIFICATION
Modification by user on the AC power Source internal circuits and parts are not allowed. The
warranty is void if the warranty seal is broken or the AC Power Source has been opened by
unauthorized person. EEC reserves the right to convert as original circuit and charge the
customer.
6300 Series Input and Output Conductor (Wire) Size Specifications (Below environment
temperature 35ºC)
N.F.B Conductor Size N.F.B Conductor Size
10 A 1.25 mm 2 150 A 50 mm2
15 A 2.0 mm 2 175 A 80 mm2
20 A 3.5 mm2 200 A 80 mm2
30 A 5.5 mm2 225 A 125 mm2
40 A 5.5 mm2 250 A 125 mm2
50 A 14 mm2 275 A 150 mm2
60 A 22 mm 2 300 A 200 mm2
75 A 22 mm2 350 A 250 mm2
100 A 38 mm2 400 A 300 mm2
125 A 50 mm2 500 A 400 mm2
1. Conductors are suggested to have multi-core wires.
2. The cores must be twisted when using the conductors. If the length of conductor has
reached above 3 meters, it needs to add higher cross-sectional area from 3.5 mm2 to 5.5
mm2.
3. N.F.B :Non-fuse Breaker.
40
8.3 Input and output terminal size
41
6390 6390P 63120 63120P
Input Output Input Output Input Output Input Output Input Output
Voltage Voltage Terminal Terminal Terminal Terminal Terminal Terminal Terminal Terminal
33200 300V M10 M10 M16 M10
600V
33220 300V M16 M10 M10 M10 M16 M10 M16 M10
600V
33380 300V
33400 300V M16 M10
33440 600V M10 M10
34208 300V
34380 300V M10 M10 M10 M10 M10 M10 M10 M10
600V M11 M8 M10 M10 M10 M10
34400 300V M12 M10
34415 300V M10 M10 M10 M10
63150 63180
Input Output Input Output Input Output
Voltage Voltage Terminal Terminal Terminal Terminal
33200 300V M16 M16
600V
33220 300V M16 M16 M16 M16
600V
33380 300V
33400 300V
33440 600V
34208 300V
34380 300V M12 M16 M16 M16
600V M16 M10
34400 300V
34415 300V M12 M16
42