Elb 11 05 01
Elb 11 05 01
net/publication/340333818
CITATIONS READS
8 565
4 authors:
All content following this page was uploaded by W. Z. Wan Hasan on 01 April 2020.
421
422 A. M. M. ALMASSRI, W. Z. W. HASAN, C. WADA AND K. HORIO
Subsequently, the calibration process was done by using CT3 Texture Analyser machine
to apply compressive loads to the sensing area of FlexiForce sensor. However, in order to
ensure the compatibility of the proposed calibration method with the desired application
which is grasping robotic hand glove, three experimental calibration tests were performed.
During each test, different loads of maximum pressure and holding times were applied.
By this calibration method, the features of force pattern are effectively identified and the
relationship between applied load and sensor output is sufficiently established. Table 1
explains the differences between three calibrations executed. Based on these experiments,
the dynamic evaluations of multi force sensors were investigated.
In the first experiment, a dynamic load ranging from zero to 44.13 N was applied to
each of the FlexiForce sensor via standard cylinder probe (diameter 52.4 mm, sized to fit
within the active area of the sensors (9.53 mm) and the puck (8.75 mm)). The puck put
between the probe and the sensor’s active area as well as the cylinder probe made from
plastic clear acrylic were inserted to ensure that the load was evenly distributed across
the active sensing area, avoiding local high pressures at edges, for example. The larger
load values were chosen because the typical maximum finger force produced by humans
is no higher than 30 N [11].
Consequently, and similarly, with previous calibration, the experiment was repeated
but with different applied forces and holding times. In the second experiment, an applied
424 A. M. M. ALMASSRI, W. Z. W. HASAN, C. WADA AND K. HORIO
force from zero to 19.61 N was loaded while more than 30 N was performed in the third
experiment as a random applied force. A set of holding times was also selected to imitate
the desired robotic hand grasping as well as to find an optimum holding time limit for
the material of sensor. The selected holding times were 10, 5 and zero sec at maximum
load for experiments 1, 2 and 3 respectively. The sensor output was recorded by readout
circuit at 100 Hz with multiple load-unload cycles tests run on each FlexiForce sensor.
Simultaneously, the load cell output (reference) was recorded by DAQ at the same sample
rate to match the sensor output. Then the sensor output was aligned with the applied
load. During each experimental test, load-unload cycles were applied and repeated with
more than 20 min and the data points output were recorded as outlined in Table 1.
4. Results. In this section, we present the real time calibration data of the force sensor
and its evaluation. Figure 2 shows the calibrated data of force sensor obtained from
three experiments within one hour of calibration as discussed previously. The blue, green
and red colours were related to the data of experiment 1, experiment 2 and experiment
3 respectively. It can be observed that the force patterns were clearly different from
experiment 1, experiment 2 and experiment 3 due to the dissimilar features of applied
forces and holding time. Indeed, different input force patterns and holding times for the
same sensor enhance the accuracy and efficiency towards identification of the behaviour
of force sensor over time so that it is easy to compensate the systematic errors in later
process.
Evaluation. To additionally clarify the evaluation of FlexiForce sensor over time, Figure
3 represents the behaviour of sensor output and its corresponding load cell reference
based on experiment 2 as example. Throughout the repeated load-unload cycles tests,
a decrease with fluctuation in the output voltage over time was observed even the same
pressure forces were applied in each cycle. This causes the systematic errors that would
be evaluated in this paper. On the whole, for evaluation and performance purpose of
calibration method, the Mean Absolute Deviation (MAD), Mean Square Error (MSE),
Root Mean Square Error (RMSE) and Mean Absolute Percentage Error (MAPE) were
0.288, 0.098, 0.313 and 79.07% respectively. The results were based on the calibration
data of experiment 2 while 39 cycles tests were repeated. However, the compensation of
systematic errors for FlexiForce sensor can be found in the paper published in [7] under
the same author.
ICIC EXPRESS LETTERS, PART B: APPLICATIONS, VOL.11, NO.5, 2020 425
Figure 3. The relative measured voltage changes in the sensor over time
based on experiment 2