Lecture SIP 2025
Lecture SIP 2025
Practice
Dr. Rohini. P
Department of ECE,
IIITDM Kancheepuram
Measuring process is the one in which the property of an object or system under
consideration is compared to an accepted standard unit.
Requirement of Measurement
The standard used for comparison must be accurately defined and commonly accepted.
Eg. Heavy or light weight doesn't have sense until its compared with a standard.
Several sets of units emerged for measuring the same physical variable - lead
to confusion
Agreed set of standard units (SI units or Syste`mes internationales d’unite´s) has
been defined.
Dr Rohini P, Department of ECE, IIITDM Kancheepuram. (Email: [email protected])
Measurement and its significance
Measurement play a significant role in achieving goals and objective of engineering
because of feedback information supplied by them. e.g. Each branch of engineering have
two functions :
Design of equipment and process,
The value of the quantity being measured is displayed in terms of the amount of movement of a pointer.
Null-Type Instruments:
Weights are put on top of the piston until the downward force balances the fluid pressure.
Weights are added until the piston reaches a datum level, known as the null point.
Pressure measurement is made in terms of the value of the weights needed to reach this null position.
It is far simpler to read the position of a pointer against a scale than to add and subtract
weights until a null point is reached.
Instruments with a Signal Output - output in the form of a measurement signal whose
magnitude is proportional to the measured quantity.
Measure of how close the output reading of the instrument is to the correct value
A pressure gauge with a measurement range of 0e10 bar has a quoted inaccuracy of 1.0%
of the full-scale reading.
(a) What is the maximum measurement error expected for this instrument?
(b) What is the likely measurement error expressed as a percentage of the output reading
if this pressure gauge is measuring a pressure of 1 bar?
Measuring quantities that are substantially less than the full-scale reading - the possible
measurement error is amplified.
It is an important system design rule that instruments are chosen - their range is
appropriate to the spread of values being measured, so that the best possible accuracy is
maintained in instrument readings.
If we are measuring pressures with expected values between 0 and 1 bar, we would not
use an instrument with a measurement range of 0e10 bar.
If a large number of readings are taken of the same quantity by a high-precision
instrument, then the spread of readings will be very small.
Reproducibility describes the closeness of output readings for the same input when
there are changes in the method of measurement, observer, measuring instrument,
location, conditions of use, and time of measurement.
Problem:
A particular micrometer is designed to measure dimensions between 50 and 75 mm. What
is its measurement range?
If the input is gradually increased from zero, the input will have to reach a certain minimum level
before the change in the instrument output reading is of a large enough magnitude to be
detectable.
Some quote absolute values, whereas others quote threshold as a percentage of full-scale
readings.
As an illustration, a car speedometer typically has a threshold of about 15 miles/h. This means
that if the vehicle starts from rest and accelerates, no output reading is observed on the
speedometer until the speed reaches 15 miles/h.
Lower limit on the magnitude of the change in the input measured quantity that
produces an observable change in the instrument output.
A major factor influencing the resolution of an instrument is how finely its output scale
is divided into subdivisions.
Using a car speedometer, this has subdivisions of typically 10 miles/h. This means that
when the needle is between the scale markings, we cannot estimate speed more
accurately than to the nearest 5 miles/h. This value of 5 miles/h thus represents the
resolution of the instrument.
For example, a pressure of 2 bar produces a deflection of 10 degrees in a pressure transducer, the sensitivity
of the instrument is 5 degrees/bar
Problem:
The following resistance values of a platinum resistance thermometer were measured at a range of
temperatures. Determine the measurement sensitivity of the instrument in ohms/C.
All calibrations and specifications of an instrument are valid only under controlled conditions
of temperature, pressure, and so on.
These standard ambient conditions are usually defined in the instrument specification.
Such environmental changes affect instruments in two main ways, known as zero drift and
sensitivity drift.
This causes a constant error that exists over the full range of measurement of the instrument.
This is often called the zero drift coefficient related to temperature changes.
If the characteristic of an instrument is sensitive to several environmental parameters, it will have several zero drift
coefficients, one for each environmental parameter.
Problem:
This table shows the output measurements of a voltmeter under two sets of conditions: (a) use in an environment kept at
20C, which is the temperature at which it was calibrated; and (b) use in an environment at 50C. Determine the zero drift
when it is used in the 50C environment, assuming that the measurement values when it was used in the 20C environment
are correct. Also calculate the zero drift coefficient.
It is then used in an environment at 30C and the following deflection/load characteristic
is measured. Determine the zero drift and sensitivity drift per degrees Celsius change in
ambient temperature.
Two parallel metal plates in which the dielectric between the plates is either air or some other medium
The capacitance C is given by C = ε0εrA/d, where ε0 is the absolute permittivity, εr is the relative
permittivity of the dielectric medium between the plates, A is the area of the plates and d is the distance
between them.
Used as displacement sensors - moveable capacitive plate relative to a fixed one changes the
capacitance
Fixed plate capacitors can also be used as sensors, in which the capacitance value is changed by
causing the measured variable to change the dielectric constant of the material between the plates –
moisture sensor
For certain pairs of materials, the terms involving squared and higher powers of T are approximately 0
and the emf-temperature relationship is approximately linear
Wires of such pairs of materials are connected at one end, and in this form they are known as
thermocouples
Voltage is measured between the two alloys (Small voltage, less than 10 mV)
Seebeck effect: when any conductor subjected to thermal gradient, generates a voltage.
If k is positive, the resistance increases with increasing temperature, and device is
called a positive temperature coefficient (PTC) thermistor, some time also known as
Resistance Temperature Detector (RTD)
If k is negative, the resistance decreases with increasing temperature, and the device
is called a negative temperature coefficient (NTC) thermistor.
Resistors that are not thermistors are designed to have the smallest possible k, so
that their resistance remains almost constant over a wide temperature range.
where:
• T is temperature (in Kelvin),
• TRef is the reference temperature, usually
at room temp (25°C;77°F; 298.5 K),
• R is the resistance of the thermistor (W),
• RRef is the resistance at TRef,
• b is a calibration constant depending on
the thermistor material, usually between
3,000 and 5,000 K.
Industrial RTDs are very accurate: the accuracy can be as high as ±0.1°C.
“Supply” R1 R2
Voltage
Vo
Vs
R4 RRTD
Pressure is defined as a force per unit area, and can be measured in units such as psi (pounds per
square inch), inches of water, millimeters of mercury, Pascal (Pa, or N/m²) or bar. Until the
introduction of SI units, the 'bar' was quite common.
Static Pressure: In the atmosphere at any point, static pressure is exerted equally in all
directions. Static pressure is the result of the weight of all the air molecules above that point
pressing down.
Static pressure does not involve the relative movement of the air
b) The density of the air. The dynamic pressure depends also on the
density of the air. If the flow rate was the same, and the air was less
dense, then there would be less force and consequently a lower
dynamic pressure
Advantages
Disadvantages
- Very expensive
The primary coil is located between the two secondary coils and is
energized with an alternating current.