Standard Operating Procedure Manual Integration
Standard Operating Procedure Manual Integration
INTEGRATION
Applicable Matrix or Matrices: NA
Standard Compound List and Reporting Limits: NA
Organics Technical
Director:
_______________________ Date:________
Bryce E. Stearns
Inorganics Technical
Director:
_______________________ Date:________
Kristine A. Dusablon
This documentation has been prepared by STL Burlington (STL) solely for STL's own use and the use of STL's
customers in evaluating its qualifications and capabilities in connection with a particular project. The user of this
document agrees by its acceptance to return it to STL upon request and not to reproduce, copy, lend, or
otherwise disclose its contents, directly or indirectly, and not to use if for any other purpose other than that for
which it was specifically provided. The user also agrees that where consultants or other outside parties are
involved in the evaluation process, access to these documents shall not be given to said parties unless those
parties also specifically agree to these conditions.
1.0 PURPOSE
1.1 This Standard Operating Procedure provides guidance on the proper way to
integrate chromatographic peaks. Integration of peak areas is commonly required to
relate instrument response to component concentration in Gas and Liquid
Chromatography. Chromatographic data systems can be set up to detect and
measure peak areas automatically. Automatic integration is preferred because it is
fast and in most cases, very consistent. Proper instrument and integration parameter
set-up facilitates automatic integration.
1.2 Consistency in integration between standards and samples is one of the most
important considerations in quantitative chromatographic analysis. Baseline upsets,
coelution and peak shape variation can sometimes complicate automatic integration
and analyst intervention through the process of manual integration may be required
to assure consistency between area assignment for standards and samples.
Examples of proper and improper integration can be found in Section 4.
2.0 POLICIES
2.3 Inappropriate peak integration used to meet quality control criteria is grounds for
dismissal.
2.5 Analysts are required to have knowledge of these commands for their particular
instruments.
3.1 The toxicity or carcinogenicity of each reagent used in this method has not been
fully established. Each chemical should be regarded as a potential health hazard and
exposure should be as low as reasonably achievable. Cautions are included for
known extremely hazardous materials or procedures.
3.2 STL maintains a current awareness file of OSHA regulations regarding the safe
handling of the chemicals specified in these methods. Material Safety Data Sheets
(MSDS) are made available to all personnel involved in the chemical analysis. STL
also has a written environmental health and safety plan.
3.3 Please note chemicals that have the potential to be highly toxic or hazardous, the
appropriate MSDS must be reviewed by the employee before handling the chemical.
3.0 PROCEDURES
3.1 Chromatograms are plots of detector signal as a function of the time.
Compounds are detected as peaks. It is necessary to determine the chromatographic
baseline signal and the start and stop points of the peaks along the baseline. Each
chromatographic data system has a particular set of integration instructions to
manually assign start and stop points once they have been determined. Assignment
of proper peak start and stop times along the baseline are illustrated in this SOP.
4.4 Properly integrated skimmed peak and properly integrated tailing peak.
4.8 Properly integrated fused tailing peak (very difficult to accurately quantify). Do
maintenance on chromatography system to avoid if possible).
* Peak has small amount of splitting and the whole peak area was not integrated.
* Peaks close in retention times utilizing the same quantitation ions, (ethylbenzene,
xylenes; dichloro-benzenes) often integrate together as one peak.
4.17 Procedure for Manual Integration. The data processing system (Target for
GC/MS, and multi-chrom for GC) has an integration option and can locate any peak
specified by the user. The analyst can choose to integrate the peak as needed and
examine the mass spectra at each part to verify proper identification. The area count
or height is automatically calculated when a peak has been integrated.
5.0 RESPONSIBILITIES
5.1.1 The reason supplied for the manual integration may be numerically designated
using the list below.
Number Reason
M1 Peak missed
Mis-identification of the peak (i.e. 1,4-
M2
dichlorobenzene identified as 1,3-dichlorobenzene)
M3 Poor automated baseline construction
Special requirement (i.e. have to calculate surrogate
M4
from a hydrocarbon envelope
M5 Other - Explain in narrative
5.2 A second review and approval of the manual integration for any compound of a
quality criteria standard must be performed, dated and initialed. Quality criteria
standards are MDLs, initial calibrations, continuing calibrations, second source
standards, surrogates, internal standards, matrix spikes, matrix spike duplicates,
blank spikes, blank spike duplicates and laboratory control standards. This second
review may be carried out by the Section Supervisor, Data Reviewer or Laboratory
Manager.
5.3 The QA Manager or designee will spot check manual integrations during internal
and data audit process. This review must also be dated and initialed.
6.0 DEFINITIONS
6.1 Integration - The process of determining area under a curve (ie a peak).
6.4 Peak - An increase in signal from baseline to a maximum and then back to
baseline.
6.5 Coelution - The concurrent elution of two or more compounds to the detector at
the same time.
6.8 Negative Spikes: Negative spikes in chromatograms are sudden upsets in the
chromatographic signal below the normal baseline.
6.9 Sudden Rise in Baseline: A rise in the baseline is caused by several factors
including high concentrations of organics in the sample and by normal column bleed.
6.12 Interference Peaks or Fused Peaks: Peaks that partially or totally coelute with
the peak of interest.
7.0 REFERENCES
7.1 US Army Corps of Engineers, August 1996, Shell For Chemical Analytical
Services, Draft Revision 4.2.
7.2 USEPA Contract Laboratory Program, Statement of Work for Organic Analysis
OLM03.2, EPA-540/R-94/073, August 1994, 1996.