Laboratory 5!
Laboratory 5!
Data Analysis
Faraday's Law states that a changing magnetic flux induces an electromotive force (EMF), and
the induced EMF is proportional to the change of magnetic flux over time:
E_emf = - dΦ / dt (1)
For a coil with N turns and area A in a uniform field B, the magnetic flux is given by:
Φ = NBA (2)
For the pendulum's mechanical damping (resistor disconnected), the equation of motion is:
I d²θ / dt² = -mgdθ - b dθ / dt (3)
The solution yields angular frequency ω and damping coefficient (b / 2I):
ω = sqrt(mgd / I - (b / 2I)²) (4)
The Faraday damping coefficient created by an electromagnetic induction is given by the
formula:
c = (NAB)² / (2L(R + r)) (5)
Calculations:
Relative error in Magnetic Field Strength:
ΔB/B * 100% = 10.14 / 55.92 * 100% = 18.13%
Moment of Inertia (I):
Using (4) and avoiding (b / 2I)² due to it being negligible(10-8):
I = mgd / ω² = (0.0903 * 9.81 * 0.11) / (5.99)² = 2.71 × 10⁻³
Error propagation:
ΔI = I * sqrt((Δm / m)² + (Δd / d)² + (2Δω / ω)²) = 0.02 × 10⁻³
Damping Coefficient (b):
b / 2I = 6.19 × 10⁻⁴ s⁻¹. (from exp fit)
b = 2I * 6.19 × 10⁻⁴ = 3.36 × 10⁻⁶
Error:
Δb = b * sqrt((ΔI / I)² + (Δ(b / 2I) / (b / 2I))²) = 0.15 × 10⁻⁶
Using the (5) theoretical value of Faraday damping is:
c = ((200 * 1.96 × 10⁻³ * 0.05076)²) / (2 * 0.255 * (1.9 + 50))
c = ((200 * 9.95 × 10⁻⁵)²) / (2 * 0.255 * 51.9)
c = (0.0199)² / 26.469 ≈ 0.000396 / 26.469 ≈ 1.496 × 10⁻⁵
Formula for error estimation in multiplication and division operations:
Δc / c = sqrt((2 ΔB / B)² + (ΔR / (R + r))²)
Δc / c = sqrt((2 × 0.0001 / 0.05076)² + (0.5 / 51.9)²)
≈ sqrt((0.00394)² + (0.00963)²) ≈ 0.0103
Δc = c × 0.0103 ≈ 1.54 × 10⁻⁷
Thus, final value of theoretical Faraday damping is ctheor = (1.50 ± 0.02) × 10⁻⁵
Discussion
In part I, measurement of the Magnetic Field Strength has relative error of 18%, which greatly
exceeds the acceptable range. It may have been caused by a man-made mistake of choosing
wrong range of values within PASCO graph(Appendix A). It is also estimated that the sensors
have issue with calibration as it is stated in manual that sensor requires preventive ‘tare’ reset
every time that it is used.