HPLC Communication Protocol
HPLC Communication Protocol
GLUQUANT HBA1C HPLC glycated hemoglobin analyzer marketed by Meril Diagnostics Pvt. Ltd. can upload the analytical
data automatically/manually to the laboratory information system (LIS) by the serial communication protocol.
GLUQUANT HBA1C HPLC glycated hemoglobin analyzer can examine A1c contained in the blood.
The analyzer can analyze the whole blood/diluted sample, display the peak map of glycated and non-glycated hemoglobins in
the sample, and figure out %A1ab, %F, %La1c, %A1c and %A0 by the area ratio of peak shape.
Before transferring the data to the LIS system, it is necessary for the professional personnel to check the analyzed peak shape
figure of the sample, confirm acceptance of the experimental result and then transmit it to the LIS system.
Two-way communication is used between the GLUQUANT HBA1C HPLC software and LIS system, so correct serial signal
communication must be ensured. However, transmission is one-way only for data transmission.
The data of LIS system, operation directive, and GLUQUANT HBA1C HPLC glycated hemoglobin analyzer.
The analysis result of each sample is transmitted to the LIS system, including machine model, sample information and sample
data, as well as one pair of start and stop characters.
At the back of the GLUQUANT HBA1C HPLC glycated hemoglobin analyzer, there is an RS-232 serial port with 9 pins,
connected to the LIS system, which is used for connecting a computer hyperterminal.
The user may at option choose the baud rate, data bit, parity and stop bit.
The default setting of the system: 115200 baud rate, 8 data bits, no parity, 1 stop bit.
100
Segments
Sample, a fixed English word of sample
Closing time of analysis, time format: YYYY-MM-DD HH:MM
Serial number of sample: Sample ID
(Where, for suffix #NN#n, NN indicates number of row, n indicates number of bits)
Sample type
(Where, 0 indicates the whole blood
sample, 1 indicates the quality control
sample, 2 indicates the calibration
sample,3 indicates the diluted sample)
Part 4 (sample data)
Included symbols and tags: <R>, |, </R>
The contents of this part occupy 3 or 4 lines, each line corresponds to one data of one sample
The content of data item on the left of the separator in each line, the value of data item on the right, accurate to one decimal
place
To be respectively: %A1ab, %F, %La1c, %A1c and %A0
Transmission case
The following is a sample of data transmitted to the computer through the serial port by the GLUQUANT HBA1C HPLC glycated
hemoglobin analyzer:
If closing an EAG item:
<SEND>
<M>MQ2000|Q2000JCA006</M>
<I>
sample|2010-06-02 15:10|4|--#01#2|0
</I>
<R>
HbA1ab|0.0
HbF|0.0
HbLa1c|0.0
HbA1c|0.0
HbA1|0.0
HbA0|0.0
</R>
</SEND>