Nano Scan
Nano Scan
PC PET/CT
in vivo molecular and preclinical imager
Preclinical Line
PET/CT Introduction
To unveil fine and accurate details of living organisms, scientists need high resolution, high-sensitivity, time-dependent
quantitation. Definition of kinetic constants of biological processes also became a basic scientific aim.
Offering solutions for modern-day science, Molecular Imaging has gone a long way in a relatively short time to become
an indispensable tool for virtually any biologist. One of the most established methods in the field, Positron Emission
Tomography (PET) remains the cutting edge and it is still the gold standard in Molecular Imaging.
The reasons of this long-lasting premier position are robust quantitation, high time resolution and real three-
dimensional results combined with an adequate observational time window. With PET, fast and easy translation of a
result from the laboratory bench to its application in the clinic is also imminent.
In the last years almost every PET imaging result is supported by information about anatomical structures of live animals
through X-ray Computed Tomography (CT) imaging.
In the recent past Mediso developed innovative technologies that provided new perspectives to small animal imaging.
Now, the new nanoScan® PET/CT system offers user-friendly imaging and a large scope of applications in one simple to
use, high resolution, high-end PET/CT system. nanoScan® PET/CT is equipped with an imaging technology widely
considered as the most advanced PET and CT detector construction, data processing and reconstruciton chain in the
industry.
Pursuit of perfection
Always eager to find an even better solution, Mediso constantly strives to develop the highest
level imaging technology possible.
We wish to serve the scientific community with our core value: supreme image quality.
2
Relying on proven excellence
Mediso's unique PET detector technology enables you to detect femtomolar quantities of
proteins per milligram of tissue with high resolution and exquisite in vivo image quality.
nanoScan® PET/CT
MultiCellTM
anesthesia / imaging bed
3
NanoSPECT/CT ®
nanoScan® PET/MRI
PET/CT
TM
MultiCell Animal Anesthesia / Imaging Bed
- Automated positioning with µm accuracy
- Closed circuit anesthesia system integrated
into a pathogen-free imaging chamber
- One-click dockable PET / CT / SPECT / MRI
compatible imaging chamber
- Integrated heating / temperature control and
monitoring
- Respiratory and ECG gating and monitoring
®
- Compatible with NanoSPECT/CT systems
4
No compromise in image quality
X-Ray CT System
- Wide energy range X-Ray tube
- 7 µm focal spot size
- High DQE at low dose detector
- Variable zoom (35 mm - 120 mm FOV)
- Closed cabinet X-Ray system
Structure
Mediso's traditions in nuclear detection technology and fine precision mechanics date back almost a century.
Leveraged by our expertise gathered across generations of patented, cutting-edge instrument innovations, the company
has designed and built the most advanced PET detector present in the market.
Mediso nanoScan® PET/CT's LYSO crystal pins are the most tightly packed (92%) and smallest in the industry,
minimizing dead detector space. Large-surface modular detector design together with the large detector ring diameter
(largest among currently available PET/CT imagers) minimizes parallax error. Mediso's patented proprietary TeraTomoTM
3D PET reconstruction algorithm offers attenuation correction, models the whole system matrix with Monte Carlo
simulations and removes parallax error. This achieves a uniform imaging resolution of ≤ 0.8 mm even at 3 cm off the
radial center. With a resolution of 0.7 mm at 1 cm off center using the very advanced TeraTomoTM 3D PET algorithm -
a value unseen in PET up to now - nanoScan® PET/CT is the first and only PET imager in the world with sub-half mm3
volumetric resolution (0.3 mm3 volumes are resolved).
The large bore size of the detector allows for a ”PET-only” or ”CT-only” mode of use in the nanoScan® PET/CT system.
High throughput imaging of larger animals is also possible. Marmoset heads or two rats can be imaged with the
standalone PET mode.
Interchangeable collimator
for large PET bore (16 cm)
6
3 TM
0.3 mm resolution by Tera-Tomo 3D PET engine
Mediso always uses state-of-the art computers and acquisition electronics to optimize data processing with the
nanoScan® PET/CT.
The combination of Mediso high-end PET detector with a very advanced 3D Teraflop Computing for Tomography:
Tera-TomoTM 3D PET reconstruction engine, also developed by Mediso leads to a PET resolution very near the physical limits.
For any tomographic detector, the acquired image is blurred and degraded due to the distortions of the imaging system.
This blurring is characterized by the Point Spread Function (PSF) or impulse response of the system.
The Tera-TomoTM 3D PET reconstruction engine incorporates both projection-space (or data-space) and image-space PSF
modeling in order to faithfully recover the original spatial resolution of the imaged objects.
Using corrections for physical factors such as detector geometry, Monte Carlo DOI estimation, object attenuation
and scatter, randoms and dead time to even positron range, a quantitative three-dimensional PET reconstruction called
Tera-TomoTM 3D PET has been developed and applied by Mediso in collaboration with prestigious Hungarian universities.
TM
Tera-Tomo 3D PET reconstruction engine principle
with on the fly system matrix generation
High-speed transfer and teraflops computing speed provides you with ultra-fast reconstructions for enhanced PET study
throughput.
7
PET Subsystem
10
6
4x10
Reconstructed activity (Bq)
6
2x10
4
1x106 2
0
0 1x106 2x106 3x106 4x106 -60 -40 -20 0 20 40 60
Activity concentration linearity for quantification Absolute sensitivity according to NEMA NU 4-2008: > 8.0%
Quantification error: < 3% Maximum sensitivity: > 9.0% (150-750 keV window)
Representation of quantitative accuracy of the nanoScan® PET/CT's PET component over 2 orders of magnitude. The
radioactivity measured in the syringes is well repeated by the reconstructed values.
1.1 0.7
1.0
0.8
RMS FWHM (mm)
0.6
0.2
Mediso 0.9 nanoScan® PC
0 5 10 15 20 25 30
Radial (mm)
CT PET
Imaging performance with Tera-TomoTM
3D PET Reconstruction Engine using 10 MBq
18
F-FDG and 30 min image acquisition
in an ultra micro-Derenzo phantom
Size of the rods: 0.7 mm – 1.2 mm
8
Exclusive PET Imaging Performance
Three plane sections of a Nude mouse bearing FaDu xenograft tumor. 34 MBq of 64Cu-labelled antibody fragment
injected i.v., imaging 24 h post injection for 20 min. Note the accumulation in kidney cortex and the tumor uptake
inhomogeneities. MIP image of the same Nude mouse.
18
F-fluoro-deoxy-glucose uptake in rat brain, PET/CT MIP image. 30 MBq of 18F-FDG injected into awake rat and imaged
50 min p.i. for 20 min.
9
CT Subsystem
CT image quality depends on pixel number, focal point size and image zoom. To prevent trade-offs in these fields,
nanoScan® PET/CT features the industry's largest surface, largest pixel number and smallest focal point CT where the
user can choose the zoom by setting it from 1.4 to 5.0. The results are a superb soft tissue contrast, low dose to the
animal, and breath-taking details. With the highest resolution mode 9.6 micron isotropic voxels are defined whereas in
overview mode large detector surface means image stitching is avoided. Mediso InterViewTM FUSION software effectively
complements the CT by providing advanced Volume-of-Interest statistical tools of Hounsfield Units. Full radiation
shielding of the closed X-ray cabinet type approved in more than 16 countries complement the CT system. The open
back-door of the CT part ensures animal access but can also be used as a traditional closed CT.
CT images of a rat skull and a mouse whole body. Imaging time 3.4 min for the rat, 6 min for
the mouse study.
10
Image courtesy of CROmed Ltd. Budapest
TM
MultiCell Animal Anesthesia / Imaging Bed
Continuous digital temperature control: by closed circuit airflow integrated into the wall of the chamber – avoiding the
side effect of the open airflow (dehydration of the eyes, contamination by pathogens etc.)
Embedded anesthetic gas connection: for any isoflurane system through dockable connection to the mouse /rat nose
cone via closed circuit tubes integrated into the wall of the chamber
Integrated animal head mounting: for precise and reproducible animal positioning
4D/5D imaging accessories: dockable connections for ECG and respiratory gating
One-click connection imaging cells: for easy and fast connection of mouse /rat imaging cells to the PET/CT, PET/MRI and
SPECT/CT scanners or dual bed docking station
PrepaCell™ Preparation Station: for complete preparation of the animal before the scan (“click and scan”)
transparent cover
- Intel® Core™ i7 platform @ 3.4 GHz - Intel® Core™ i7 platform up to 4.5 GHz
- Non-stop operating WS through liquid cooling - Cutting edge computing with liquid cooling
- Safe data handling by 0,5 TB SSD - Safe raw data backup by 0.5 TB SSD
- GPU based on the fly volume rendering 1/10 Gb Ethernet - GPU cluster with 6 GB memory 10 Gb Ethernet
- Ease of use CT scout for positioning - Mainstream data transfer through dual 10 GbE
- 24” + 17” (on the gantry) LCD monitor - Realtime 64 bit OS by Linux
- Multitask 64 bit OS by MS Windows 7
Researcher`s Room
Management
13
Post-processing by
InterViewTM FUSION multi-modal application developed by Mediso is an essential part of nanoScan® PET/CT system. The
application provides a wide range of functionalities to evaluate preclinical PET/CT data. 2D single, orthogonal and tiled,
as well as 3D MIP and Volume Rendering viewers represent fast and flexible visualization techniques built on GPU
acceleration. Viewers provide dual, triple and quadruple fusion to accurately compare and enhance multi-modal single
and follow-up studies. Dynamic PET images together with CT can be fused, and PET images can be studied over time.
Multiple Time Activity Curves (TAC) of dynamic PET studies over time can be visualized and evaluated.
Calculations and statistical evaluations for PET are available on voxel and ROI or VOI level. A wide range of ROI and VOI
tools are available for evaluation (e.g. freehand, polygon, ellipse, rectangle, sphere, box and isocount).
CT based body-air-bone-lung segmentation methods are provided for effective PET attenuation correction.
An in-built state of the art automated rigid, affine and non-linear image registration framework provides a quick and
accurate way to superimpose different studies for comparison.
Advanced segmentation methods for different modalities, tissues and organs are available for feature extraction (e.g.
lung, vessel, bone, CT body-air-bone-lung, PET lesion detection)
Arithmetic operations help differentiating follow-up studies on voxel level by several methods (e.g. sum, difference,
absolute difference, average, minimum, maximum, multiply).
Visualization of 18F-fluoro-deoxy-glucose uptake in rat brain. After the intravenous injection of 30 MBq of 18F-FDG rats
were kept awake for 50 min. The static PET image was collected during 20 min in isoflurane anesthesia. Cortical gyri,
thalamus, cerebellum and brain stem uptake is visualized. Note the FDG uptake in brown fat tissue over the scapular
region. Image courtesy of Karolinska Institute.
To enable researchers with all necessary image visualization, analysis and quantification tools, Mediso offers a uniquely
complete and flexible software package for the nanoScan® PET/CT system. Beside Mediso's InterViewTM FUSION both
inviCRO's VivoQuant providing advanced 3D segmentation, data sharing and atlas tools and PMOD's industry
standard pharmacokinetic analysis and post-processing modules are supported.
14
Applications
input, response and fit response contributions cg ~ fdgmodel(tt, ca, metfrac, k1, k2, k3, k4, fbv, delay)
Fit: k1 = 0.4423, k2 = 0.06304, k3 = 0.02874, fbv = 0.03741 Fix: k4 = 0, delay = 0, ca = 8.384e+04, metfrac = 0
5e+05
5e+05
5e+05
4e+05
4e+05
l l
l
l l
l
3e+05
3e+05
l l
input + response (Bq/ml)
3e+05
l
response (Bq/ml)
response (Bq/ml)
l l
l
l l
l
l l
2e+05
2e+05
l
2e+05
l
ll ll l l l l
l l l l l
l l l
ll l
ll ll
l l l l l l l
l l l
l residuals
ll lll 0 10 20 30 40 50
ll
ll lll l
l ll
l
l l l
llll ll
1e+05
1e+05
ll
1e+05
l l
ll ll ll
l
20000
l l l
l l l
l l
l l l
l l
0
l l l l
l l
l
l
-20000
l l
0e+00
0e+00
ll ll
0e+00
l ll
l l
ll ll
ll
ll
l
l
ll l
l
ll
l
ll
l
ll
ll
l
l llllllll l l l l l l l l
l
llll
0 10 20 30 40 50
0 10 20 30 40 50 0 10 20 30 40 50
blood, myocard time (min) time (min) blood, myocard time (min)
For exact details and up to date information please contact your local
distributor or Mediso Medical Imaging Systems.