IT7900 User Manual-En
IT7900 User Manual-En
Supply
IT7900 Series User Manual
Model: IT7900
Version: V1.0/2023.09
Notices
Warranty Safety Notices
© Itech Electronic, Co., Ltd. 2023
No part of this manual may be The materials contained in this
reproduced in any form or by any document are provided “as is”, and
means ( including electronic storage is subject to change, without prior
and retrieval or translation into a foreign notice, in future editions. Further, to A CAUTION sign denotes a
language) without prior permission and the maximum extent permitted by hazard. It calls attention to an
written consent from Itech Electronic, applicable laws, ITECH disclaims all operating procedure or practice
Co., Ltd. as governed by international warrants, either express or implied,
copyright laws. that, if not correctly performed or
with regard to this manual and any
information contained herein, adhered to, could result in
Manual Part Number including but not limited to the damage to the product or loss of
IT7900 implied warranties of important data. Do not proceed
merchantability and fitness for a beyond a CAUTION sign until
Revision particular purpose. ITECH shall not the indicated conditions are fully
1st Edition: Sep.01, 2023. be held liable for errors or for understood and met.
incidental or indirect damages in
Itech Electronic, Co., Ltd.
connection with the furnishing, use
Trademarks or application of this document or of
any information contained herein.
Pentium is U.S. registered trademarks Should ITECH and the user enter
of Intel Corporation. into a separate written agreement A WARNING sign denotes a
Microsoft, Visual Studio, Windows and with warranty terms covering the hazard. It calls attention to an
MS Windows are registered trademarks materials in this document that operating procedure or practice
of Microsoft Corporation in the United conflict with these terms, the that, if not correctly performed or
States and/or other countries and warranty terms in the separate adhered to, could result in
regions. agreement shall prevail. personal injury or death. Do not
Technology Licenses proceed beyond a WARNING
The hardware and/or software described
sign until the indicated
herein are furnished under a license and conditions are fully understood
may be used or copied only in and met.
accordance with the terms of such
license.
Warranty
ITECH warrants that the product will be free from defects in material and
workmanship under normal use for a period of one ( 1) year from the date of
delivery ( except those described in the Limitation of Warranty below).
For warranty service or repair, the product must be returned to a service center
designated by ITECH.
⚫ The product returned to ITECH for warranty service must be shipped
PREPAID. And ITECH will pay for return of the product to customer.
⚫ If the product is returned to ITECH for warranty service from overseas, all
the freights, duties and other taxes shall be on the account of customer.
Limitation of Warranty
This Warranty will be rendered invalid if the product is:
⚫ Damaged resulting from customer-wired circuits or customer-supplied parts
or accessories;
⚫ Modified or repaired by customer without authorization;
⚫ Damaged resulting from customer-wired circuits or use in an environment
not designated by us;
⚫ The product model or serial number is altered, deleted, removed or made
illegible by customer;
⚫ Damaged as a result of accidents, including but not limited to lightning,
moisture, fire, improper use or negligence.
Safety Symbols
Direct current ON ( power)
A chassis terminal - -
Safety Precautions
The following safety precautions must be observed during all phases of
operation of this instrument. Failure to comply with these precautions or specific
warnings elsewhere in this manual will constitute a default under safety
standards of design, manufacture and intended use of the instrument. ITECH
assumes no liability for the customer’s failure to comply with these precautions.
Environmental Conditions
The instrument is designed for indoor use and an area with low condensation.
The table below shows the general environmental requirements for the
instrument.
Environmental Conditions Requirements
Operating temperature 0°C~40°C
Operating humidity 20%~80%( non-condensation)
Storage temperature -10°C~70 °C
Altitude Operating up to 2,000 meters
Installation category II
Pollution degree Pollution degree 2
Note
To make accurate measurements, allow the instrument to warm up for 30 min.
Regulatory Markings
The CE mark indicates that the
product complies with all the relevant
European legal directives. The specific
year ( if any) affixed refers to the year
when the design was approved.
The instrument complies with the
WEEE Directive ( 2002/96/EC)
marking requirement. This affix
product label indicates that you must
not discard the electrical/electronic
product in domestic household waste.
EMC Standard
IEC 61326-1:2012/ EN 61326-1:2013 ¹²³
Reference Standards
CISPR 11:2009+A1:2010/ EN 55011:2009+A1:2010 (Group 1, Class A)
IEC 61000-4-2:2008/ EN 61000-4-2:2009
IEC 61000-4-3:2006+A1:2007+A2:2010/ EN 61000-4-
3:2006+A1:2008+A2:2010
IEC 61000-4-4:2004+A1:2010/ EN 61000-4-4:2004+A1:2010
IEC 61000-4-5:2005/ EN 61000-4-5:2006
IEC 61000-4-6:2008/ EN 61000-4-6:2009
IEC 61000-4-11:2004/ EN 61000-4-11:2004
Safety Standard
IEC 61010-1:2010/ EN 61010-1:2010
Chapter1 Overview
This chapter introduces the front panel, the rear panel, key functions and LCD
display function of the IT7900 series power supply, make sure that you can
quickly know the appearance, instruction and the key function before you
operate the power supply. Help you make better use of this series of power
supply.
Note
Naming rules for this series of model are as follows: IT79XXX-YYY-ZZZ,
wherein, XXX means rated power, YYY means rated voltage, and ZZZ
means rated current. Take IT7915-350-90 for an example, the rated
power is 15kVA, rated voltage is 350V and rated current is 90A. In
addition, ATE indicates that this model is a system model without an
operator panel.
IT7902-350-10U/IT7904-350-20U/IT7906-350-30 机型
IT7905-350-30U/IT7906-350-90/IT7909-350-90/IT7912-350-90/IT7915-350-90 Model
IT7930-350-180 Model
IT7945-350-270 Model
IT7960-350-360/IT7975-350-450/IT7990-350-540/IT79105-350-630 Model
IT79120-350-720/IT79135-350-810/IT79150-350-900/IT79165-350-990 Model
Transportation
If you need to transport non-cabinet products, you must pay attention to the fol-
lowing to ensure the safety of equipment and personnel.
⚫ Before moving, make sure that the cabinet or stand where the
equipment will be placed has been fixed and meets the load-bear- ing
conditions to avoid tilting and collapsing, causing personnel to be
injured, and equipment broken.
⚫ Due to the weight of the product, transport by hand should be avoided
where possible. If unavoidable, carry it with two people and holding
the product shell and not external parts (such as han- dles, electrodes,
knobs, etc.).
⚫ When carrying, be prepared to bear the weight to avoid sprains or being
crushed by heavy objects.
⚫ Use suitable safety clothing, especially safety shoes, when carrying the
equipment, as due to its weight a fall can have serious consequences.
After unpacking the cabinet product, if you need to move it to other places, you
must pay attention to the following matters to ensure the safety of equipment
and personnel.
⚫ The cabinet product is very heavy. Before moving to another loca- tion,
confirm whether the ground load is in compliance.
⚫ During the process of moving the cabinet, it is recommended that two
or more people cooperate and push it slowly and at a constant speed.
If you encounter a pit, you need to pay special attention. It is forbidden
to push it quickly, otherwise it will easily cause excessive inertia and
cause the casters at the bottom of the cabinet to jam and the cabinet
to fall.
⚫ It is not advisable to push down the slope to prevent the cabinet from
falling down due to the shift of the center of gravity. It is rec- ommended
Copyright ©ITECH Electronic Co., Ltd. 19
Instrument Introduction
to use a forklift or crane to move the cabinet.
⚫ ITECH 27U and 37U cabinets are equipped with hoisting rings as
standard on the top. It is recommended to use a crane equipped with
a four-leg hoisting belt structure for horizontal hoisting and moving,
and ensure that the four hoisting belts are the same length to avoid
cabinet skew during movement. As shown below.
⚫ After moving to the destination, please lock the four casters to se- cure
the cabinet.
⚫ The cabinet should be placed on a level ground. It is forbidden to place
the cabinet on a sloped ground.
NOTE
After confirming that package contents are consistent and correct, please appropriately
keep package box and related contents. The package requirements should be met when
the instrument is returned to factory for repair.
The IT7930-350-180 model comes with a yellow optical fiber cable. The fiber cable need
to install by yourself. For details, see 3.3 chapter.
8
1 Power Switch 2 USB interface /Print/Trig/Menu
3 LCD touch screen 4 Function key
5 Number key 6 Up, down, left and right key and enter key
7 Rotary knob 8 Vent hole
2.4 Keyboard
The keyboard introduction of IT7900 series Power Supply is shown as follows.
Keys Description
Print Used for saving screen images
Trig Used for manual trigger
Power Power Switch
Menu Used for going back to menu page
Composite key [Shift], combined with other keys to realize functions marked
above keys. Firstly, press [shift] and the shift key will be lighted, and then press
the function key, the detailed functions are listed as follows.
Keys Description
[Shift]+[Set](List) Enter the List function menu.
[Shift]+[F-set] (Sweep) Enter the Sweep function menu.
Confirm settings
After completing the value setting or selecting a menu item, pushing the knob
acts like pressing [Enter] key to confirm the operation.
The rear panel of IT7900 series 15U and above series cabinets are the same.
Take IT7945-350-270 as an example, the rear panel is introduced as shown
below.
1 2 3
1: Output socket, which can be used when users need single-phase lower than
220V/10A power supply
2: Output terminals and protective cover, remove the protective cover to show
ABCN terminals and sense terminals
3: Input terminal, connected to the instrument's power supply
4: The communication interface for cabinet, the detailed diagram as follows, and
the descriptions are the same as 3U model.
Chapter3 Installation
3.1 Connecting the Power Cord
Connect power cord of standard accessories and ensure that the power supply
is under normal power supply.
⚫ Before connecting power cord, be sure to confirm that the power voltage
matches with the rated input voltage of the instrument.
⚫ Before connecting power cord, be sure to switch off the instrument. Verify
that there is no dangerous voltage on the connection terminals.
⚫ To avoid fire or electric shock, Make sure to use the power cord supplied by
ITECH.
⚫ Be sure to connect the power cord to the AC distribution box with protective
grounding. Do not use terminal board without protective grounding.
⚫ Do not use an extended power cord without protective grounding, otherwise
the protection function will fail.
⚫ Ensure that the power cord connection terminals are either insulated or
covered by the supplied protective cover so that no accidental contact with
lethal voltage can occur.
NOTE
IT7930-350-180 is standard equipped with two power cords.
NOTE
The T7930-350-180 consists of two 15kVA in parallel mode. The output terminals
and Remote sense terminals use the master interface. The wiring method is the
same as other models.
Please confirm that the Remote Sense function in the menu is set to Off,
otherwise the instrument will report an error in the present connection mode.
⚫ The connection diagram of single phase is shown as follow:
NOTE
When the output voltage has DC voltage, the output terminal L is positive, and N is
negative.
NOTE
When the output voltage has DC voltage, the output terminal A is positive, and B is
negative.
NOTE
When the output voltage has DC voltage, the output terminal L is positive, and N is
negative.
1. Confirm that the power switch is in the OFF position and verify that there is
no dangerous voltage on the connection terminals.
2. Remove the output terminals cover of the power system.
3. Loosen the screws of the output terminals and connect the red and black
test cables to the output terminals. Re-tighten the screws.
When maximum current that one test cable can withstand fails to meet the
rated current, use multiple pieces of red and black test cables. For example,
the maximum current is 1,200A, then 4 pieces of 360A red and black cables
are required.
4. Thread the red and black test cables through the output terminals cover of
the power system and install the cover.
Copyright ©ITECH Electronic Co., Ltd. 31
Installation
5. (Optional) According to the actual situation of DUT, connect the grounding
terminal on the rear panel of the instrument to the DUT to ensure the safe
grounding.
For the location information, see 1.5 Rear Panel Introduction.
6. Connect the other end of the red and black cables to the DUT. The positive
and negative poles must be properly connected and fastened when wiring.
NOTE
The T7930-350-180 consists of two 15kVA in parallel mode. Connect the Remote
sense terminals of master module. The wiring method is the same as other
models.
NOTE
When the output voltage has DC voltage, the output terminal L is positive, and N is
negative.
NOTE
When the output voltage has DC voltage, the output terminal A is positive, and N is
negative.
1. Confirm that the power switch is in the OFF position and verify that there is
no dangerous voltage on the connection terminals.
2. Remove the output terminals cover of the power system.
3. Refer to the wiring diagram and connect the Vs+ and Vs- with armored
twisted-pair cables. Loosen the screws of the output terminals and connect
the red and black test cables to the output terminals. Re-tighten the screws.
When maximum current that one test cable can withstand fails to meet the
rated current, use multiple pieces of red and black test cables. For example,
the maximum current is 1,200A, then 4 pieces of 360A red and black cables
are required.
4. Thread the red and black test cables through the output terminals cover of
Copyright ©ITECH Electronic Co., Ltd. 33
Installation
the power system and install the cover.
5. (Optional) According to the actual situation of DUT, connect the grounding
terminal on the rear panel of the instrument to the DUT to ensure the safe
grounding.
For the location information, see 1.5 Rear Panel Introduction.
6. Connect the other end of the remote sense cables to the DUT.
7. Connect the other end of the red and black cables to the DUT. The positive
and negative poles must be properly connected and fastened when wiring.
8. Power on the instrument and turn on the Sense function of the instrument.
⚫ Before connecting the cables, ensure that the instrument power switch is off
and the main switch of the AC power input (distribution box) is off.
⚫ Fiber optic cables cannot be flexed or folded. When the cable is too long
and needs to be arranged, gently wrap the cable in a circle and gently tie it.
As shown in the following figure, connect the System Bus between the master
and slave by fiber cords.
Precautions
To prevent electric shock and damage to the instrument, please observe the
following precautions.
Key operation is
Output is on
locked
Over temperature
Over voltage protection
protection
AC/AC+DC mode AC/AC+DC mode
Peak OCP RMS OCP
Menu Introduction
Press the [Menu] key on the front panel and enter to the menu interface. Menu
interface will display all of function icon, user can rotate the knob or direction key
to select, or click the screen to enter the function interface.
⚫ The [On/Off] key is used to turn the output on or off under normal
circumstances. Even if the instrument is in control by PC or the keyboard is
locked, the [On/Off] is still valid.
⚫ The [On/Off] key light is off and turning the output off does not place the
instrument in a safe state. Hazardous voltages may be present on all output
and guard terminals. Putting the equipment into an output-off state does not
guarantee that the outputs are powered off if a hardware or software fault
occurs. See the cautions about connecting the test lines before connecting
test lines.
You can press the [On/Off] key on the front panel to control the output status of
the power supply. If the [On/Off] key light is on, indicates that the output is turned
Copyright ©ITECH Electronic Co., Ltd. 40
Getting Started
on. The VFD displays the meter value such as voltage, current, power and so
on. If the [On/Off] key light is off, indicates that the output is turned off. The VFD
displays that the power supply state is OFF.
The relationship between output voltage, current, power and frequency is shown
in the figure below.
Single Phase
Under the single phase, the power supply works as a single phase power supply.
the output mode can be set to AC/DC/ACDC/DCAC.
Three Phase
Under the three phase mode, the power supply works as a three phase power
supply. the output mode can be set to AC/ACDC.
Reverse Phase
Under the reverse phase, the power supply works as a single phase power
supply, and the rated voltage will be extended to 200%. For example, AC power
supply rated voltage is 350V, under the reverse phase mode, the reted output
voltage is 700V. And the output mode can be set to AC/DC/ACDC/DCAC.
Multi-channel
Under the Multi-channel phase, the power supply works as a single phase power
supply and have three channels.
The multi-channel function of the IT7900 series allows users to test 3
independent DUTs at the same time without adding additional hardware
configuration. In the traditional solution, the user needs to configure 3 AC power
supplies; and one IT7900 device can meet multi-channel testing requirements.
Under multi-channel mode, parameters of each channel can be set
independently. The output mode can be set to AC/DC/ACDC/DCAC.
Under AC+DC mode, Set the output voltage and frequency of the AC power
supply in the main interface, and set the DC voltage in the configure menu.
Vac: you can set the Vac under the main interface or under the config interface.
DC: set the DC component under the config interface.
NOTE
The AC + DC mode not only expands the application range of pure AC voltage, but also expands
the application range of DC component in laboratory testing.When using AC+DC for testing, please
first understand the ripple parameters of the instrument when it is used as a DC power suppl y. If
there are strict noise requirements, additional DC noise filters are needed to obtain low noise and
good stable DC voltage for testing.
Under DC+AC mode, Set the Vdc in the main interface, and set the Vac in the
configure menu.
Vdc: you can set the Vdc under the main interface or under the config interface.
Ripple control: set the wave,Vac component and frequency under the config
interface.
⚫ Sine
⚫ Square
⚫ Sawtooth
⚫ Triangle
⚫ Trapezoid
⚫ Clipped-sine
⚫ Rectifier
⚫ THD wave
⚫ User-defined
When Trapezoid, Clipped-sine, Rectifier, THD Wave and User-define are
selected, the user should configure the parameter of waveform.
Operating steps
1. Press [Shift] + [F-set](Sweep)on the front panel to enter the sweep
interface, as shown in the figure below.
2. Insert the plug of the fiber optic cable into the fiber optic module and hear a click
sound to indicate that it is inserted in place. The fiber optic cable connection
schematic is as follows.
If in balance mode, the Phase Delay between the first instrument and the
second intrument is set to 60° to achieve six-phase balanced output of 60°
between six phases.
NOTE
For parameters of corresponding curves and related regulations, refer to
IEC standards.
Interface Instruction
Take IEC 61000-4-11 as an example to introduce the interface display and how
to use the regulation test.The usage method of other regulatory interfaces is the
same, but the displayed parameters are different.
⚫ Configure
Click Configure to enter the configuration interface of other parameters, and
General Menu:
Buzzer Set the keyboard sound
Key Buzzer on/off for key
General Protect Buzzer on/off for protect
Brightness Set the screen brightness.
1-10 Set the screen brightness level
Select Language
Users can select the instrument language type from the menu.
AC
Configuration setting for output.
Config
Output current.
Iac: output current of AC, range from 0-full
Current AC scale
Slew Rate: Slope , range from 0.0001-
5000A/ms
Freq: frequency of output, rang from 16-
Frequency
2400Hz
Copyright ©ITECH Electronic Co., Ltd. 69
System-Related Functions
Slew Rate: slew rate of frequency,
0.00001-5000Hz/ms
Selece and edit the output wave type. For
Waveform detailed introduction please refer to 4.3
waveform selection.
Status: Dimming function switch.
Dimming Edge: select front edge or back adge
Phase: phase control, range from 0-180°
AC+DC
Configuration setting for output.
Config
Output current.
Iac: output current of AC, range from 0-full
Current AC scale
Slew Rate: Slope , range from 0.0001-
5000A/ms
Idc: DC current value, range from 0-full
scale
Slew Rate: slew rate, 0-5000A/ms
Current DC It is not displayed in three-phase balanced
mode, if you need to set the three-phase
DC component, balanced mode needs to
be turned off.
Freq: frequency of output, rang from 16-
2400Hz
Frequency
Slew Rate: slew rate of frequency,
0.00001-5000Hz/ms
Selece and edit the output wave type. For
Waveform detailed introduction please refer to 4.3
waveform selection.
Status: Dimming function switch.
Dimming Edge: select front edge or back adge
Phase: phase control, range from 0-180°
Dimming Function
The front and back edge of the waveform can by concealed and the phase
angle set with Dimmer function to regulate the active power, thus
adjusting the lighting intensity.
Select Front edge, and phase set to 90° in Configuration menu.
Save Operation
The save operation procedures are as follows:
1. Press the composite keys [Shift]+[4] (Save) to enter the parameter save
interface.
2. Select the storage location. up to 10 position can be select.
Recall Operation
You can recall the parameters you saved in the specified memory location as
the setting values.
1. Press the composite keys [Shift]+[5] (Recall) to enter the parameter
recallinterface.
2. Set the storage location.
Press the direction keys to set the storage location, and then, the saved
parameters will be display at the bottom of the interface.
3. Press [Enter] to recall the parameters.
AC/ACDC Mode
Press [Shift]+[Config] (Protect) and enter to Protect configure menu, where
you can set the following protection.
Current RMS protection RMS current for OCP protection
Rms Protection point
Time delay time of protection
Type Protection type.
Limit
Output Disable
Current peak protection Current Peak OCP
Peak Protection point
Time delay time of protection
Voltage peak range Voltage peak range setting
V limit Maximum voltage setting
Power limit range Power limit setting
P limit Power limit point
Time delay time of protection
Type Protection type.
Limit
Output Disable
DC/DCAC Mode
Press [Shift]+[Config] (Protect) and enter to Protect configure menu, where
you can set the following protection.
Current limit range Current limit setting
I+ limit Maximum current setting, the
current setting will be limited to
this range.
How to Set
1. Press [Shift]+[Config] (Protect) keys and enter to Protection menu.
2. Press the up/down key or rotate the knob to select Current RMS protection
and press [Enter].
3. Set the protection RMS Level, the delay time and protection type in
sequence, and press [Enter] to confirm.
How to Set
1. Press [Shift]+[Config] (Protect) keys and enter to Protection menu.
2. Press the up/down key or rotate the knob to select Current peak protection
and press [Enter].
3. Set the protection peak Level and the delay time in sequence, and press
[Enter] to confirm.
⚫ Before connecting the system bus, you must ensure that each instrument is
in single mode (Single).
⚫ Fiber optic cables cannot be flexed or folded. When the cable is too long
and needs to be arranged, gently wrap the cable in a circle and gently tie it.
⚫ Before connecting the cables, ensure that the instrument power switch is off
and the main switch of the AC power input (distribution box) is off.
⚫ Before connecting 3 single instruments to the AC distribution box, ensure
that the distribution box capacity is sufficient. Refer to the corresponding
specifications for the AC input parameters of a single instrument.
For 3U models, the same model can be connected in parallel. Taking three 3U
single instruments (with operation panel) as an example, the steps for parallel
operation are as follows.
1. Ensure that the power switches of the three units and the main switch of the
AC power distribution box are off.
2. Refer wiring connection diagram to connect three units.
For long distance testing requirements:
3. Connect the AC input terminals of the three units separately, and connect
them to the AC distribution box.
4. Connect the output terminals of the three units in parallel and connect them
to the DUT.
5. Refer to the blue wiring legend in the figure, connect the System Bus (i.e.,
the fiber outer ring interfaces TX and RX) for fiber-optic communication
between the master and slaves.
a) Insert the fiber optic module into the hole corresponding to TX RX.
b) Insert the plug of the fiber optic cable into the fiber optic module and hear a
click sound to indicate that it is inserted in place. The fiber optic cable
connection schematic is as follows.
⚫ Before connecting the system bus, you must ensure that each instrument is
in single mode (Single).
⚫ Fiber optic cables cannot be flexed or folded. When the cable is too long
and needs to be arranged, gently wrap the cable in a circle and gently tie it.
⚫ Before connecting the cables, ensure that the instrument power switch is off
and the main switch of the AC power input (distribution box) is off.
⚫ Before connecting 3 single instruments to the AC distribution box, ensure
that the distribution box capacity is sufficient. Refer to the corresponding
specifications for the AC input parameters of a single instrument.
For cabinet models, the same model can be connected in parallel. Taking two
15U instruments (with operation panel) as an example, the steps for parallel
operation are as follows.
1. Ensure that the power switches of the instrument and the main switch of the
AC power distribution box are off.
Refer wiring connection diagram to connect two units.
The configuration method is the same as 3U model instrument.
Notice:
⚫ The connection cables from the cabinet terminals to the mid-node are
as short as possible and bundled up after the connection is completed.
The cable length from the two cabinets must be the same.
⚫ Connect TX, RX fiber cable of two cabinets for communication,
connected in the same way as a 3U model instrument.
⚫ Each cabinet has a red role selection button that allows the user to
switch the red button to select the present instrument as a master or as
a slave.
Pins Introduction
Different I/O implements different functions. The detailed functions description
are shown in the figure below:
Digital-IO
Digital IO-1 Input/Output Turn off the output under Level signal
emergency status
⚫ Input/Output Function
The IO-1 ~ IO-7 pins are featured default function, the user can setting the
function of pin according to requirement. The Input and Output are the
general digital I/O function, and the parameter settings and functions of the
Digital IO-1
IO-1 pin can be set to 【Inhibit-latch】,【Inhibit-living】, 【Input】, 【Output】
The default function is inhibit output. When the IO pin is configured for a Inhibit
function and the level signal is low, the output of the machine is forbidden. At
this point, Pin 1 has a bi-directional I/O function, which can both receive the level
signal input from the external instrument and output the level signal
outward.default level is high, and low is valid when entering. Outgoing output
also generates low level signals.
Inhibit function has two mode: Latch and Living.
⚫ Living: When input an inhibit signal and the instrument output is turned OFF.
The status bar of the LCD screen displays INH warning icon and the output
is marked as OFF. If power supply output is ON state before, the ON/OFF
button will be lit. When the input signal undoes, the output returns to normal.
This function can be used to control the output of the power supply.
⚫ Latch: When input an inhibit signal and the instrument output is turned OFF.
The ON/OFF button will be lighted off, the status bar of the LCD screen
displays INH warning icon. In this case, user need to remove the input signal
and press [Shift]+[Esc] to cleare protection, then manually turn on [On/Off]
again.
Digital IO-2
IO-1 pin can be set to 【PS-clear】, 【Input】, 【Output】
The default function is to clear the protected state. When the protection occurs,
the protection state can be cleared through this pin, so that the instrument can
continue to output normally.
IO-2 is bi-directional, that is, when the power supply is in a protected state, the
instrument can receive a pulse signal from an external input through IO-2 for
clean protection operation, or when the power supply is in a protected state, the
clean protection can generate a pulse signal from IO-2.
Digital IO-3
IO-3 pin can be set to 【PS】, 【Input】, 【Output】
Digital IO-4
IO-4 pin can be set to【Sync-in】, 【Sync-out】,【Input】,【Output】
This function can be used to simulate the six-phase output mode, in which one
IT7900 outputs a synchronous signal to another IT7900, and the second power
supply runs synchronously with the frequency and phase of the first power
supply to realize the six-phase output function.
【Sync-in】: Synchronous input function, which is used to output frequency lock
or phase lock with the external signal. At this time, the machine synchronizes
the frequency or phase information input from the IO-4 pin.
【Sync-out】: Synchronous output function, the IT7900 produces synchronous
signal to the outside, which is AC zero crossing pulse signal sent from the IO
port.
Digital IO-5
IO-5 can be set to 【OnOff-status】, 【Input】, 【Output】
The default function is to indicate the output state of the power supply, in case
of output is ON, output 5V, otherwise, output 0V.
When pin5 is set to Invert, the output level is completely opposite.
Digital IO-6
IO-6 can be set to【Trigger1-in】, 【Trigger1-out】, 【Input】,【Output】
【Trigger1-in】: The input trigger signal, the pulse signal sent to the IO-6 pin can
be used as the trigger source. Users can select as the trigger source of the
corresponding function in the menu.
【Trigger1-out】: The output trigger signal, when the instrument generates a
trigger signal, the pin6 generates an pulse signal.
Digital IO-7
IO-7 can be set to【Trigger2-in】, 【Trigger2-out】, 【Input】,【Output】
【Trigger2-in】: The input trigger signal, the pulse signal sent to the IO-6 pin can
be used as the trigger source. Users can select as the trigger source of the
corresponding function in the menu.
【Trigger2-out】: The output trigger signal, when the instrument generates a
trigger signal, the pin7 generates an pulse signal.
Remote Control
Through the analog input interface, the analog signal can be input to remotely
set the output voltage value or power amplification function. For the detailed
function definition of the pin, please refer to the definition of the analog pins
description. The following is an example of amplitude modulation to introduce
how to connect and how to use.
When the voltage setting is controlled through the analog interface, the external
voltage (-10V ~ 10V) is connected to program the voltage value between 0 and
full scale(AC mode).
For example, analog control AC range of 0~350V voltage, when the analog
signal voltage is set to 5V, the voltage ratio set to 50V/1 under source menu, the
instrument output voltage is set to 5*50=250V.
When parallel machine operation, can be controlled through the host analog
interface.
Single Phase
Under the single phase, the power supply works as a single phase power supply.
the output mode can be set to AC/ AC+DC.
Reverse Phase
Under the reverse phase, the power supply works as a single phase power
supply, and the rated voltage will be extended to 200%. For example, AC power
supply rated voltage is 350V, under the reverse phase mode, the reted output
voltage is 700V. And the output mode can be set to AC/AC+DC.
Under AC+DC mode, Set the output voltage and frequency of the AC power
supply in the main interface, and set the DC voltage in the configure menu.
⚫ Vac: you can set the Vac under the main interface or under the config
interface.
DC: set the DC component under the config interface.
Press on the front panel and the following waveform display interface
will appear. Different modes display different interfaces. the Oscilloscope
interface is shown in the figure below.
Trigger level
Measuring Wave
parameters
Vertical calibration
The voltage range and current range are subject to vertical calibration
(voltage/scale and current/scale). Press the soft key [Function] and select Volt-
Range or Curr-Range. Rotate the knob to set the voltage or current range of
each interval.
Horizontal calibration
If the soft key [Function] is pressed and “Trig-TimeBase” is selected, you can
rotate the knob to adjust the horizontal scale (scanning speed). When the
horizontal (time/scale) setting is changed by rotating the knob, you can observe
the change of time/scale on the screen. During data collection, the sampling
speed can be changed by adjusting the horizontal calibration knob. After
collection is stopped, the collected data can be amplified by adjusting the
horizontal calibration knob.
Trigger waveform
When the specified trigger conditions are satisfied, the trigger waveform will be
displayed. The triggering time is the trigger point, generally on the right of the
screen. When the trigger point is reached, the screen will display the waveform
from left to right over time. The user should set the following parameters before
using the trigger function.
⚫ Trigger mode
The trigger mode refers to the condition to update the contents on the
screen. It is divided into the Auto mode and Normal mode. In the Auto mode,
the displayed waveform will be updated when triggering occurs in the
suspension time; otherwise, the displayed waveform will be updated
automatically.
In the Normal mode, the displayed waveform will be updated in the case of
triggering and not updated in the case of no triggering.
⚫ Trigger source
The trigger source is used for generating trigger conditions. The user can
select the trigger source in the input signal of the input unit.
⚫ Trigger slope
Copyright ©ITECH Electronic Co., Ltd. 93
Measurement Functions
The slope refers to the change of the signal from low level to high level
(rising edge) or from high level to low level (falling edge). The slope used as
a trigger condition is referred to as the trigger slope.
⚫ Trigger level
The trigger level refers to the level which the trigger slope passes through.
If the signal of the trigger source passes through the set trigger level
according to the specified trigger slope, triggering occurs. When the soft key
[Function] is pressed and “Trig-Level” is selected, you can rotate the knob
to adjust the trigger level. In this case, the trigger level can be changed by
rotating the knob and you can observe trigger level changes on the screen.
Print data
In the advanced menu, users can select the Print Data item and select the data
logging mode. Data of oscilloscope interface will be recorded to U disk.
Print data mode:
⚫ Off: turn off the print data function.
⚫ Post: The recorded data is consistent with the data displayed on the
oscilloscope interface
⚫ Raw: The recorded data is original data, The default data sampling interval
is 10us
⚫ Both: Post and Raw, record two data file.
Press key on the front panel, and the following initial interface of
harmonic measurement will appear.
⚫ Harmonic bar chart screen
Harmonic number list: the signal data of 0-50th harmonic(s) will be shown in
the LIST. You can press the Up and Down key to display the hidden rows,
i.e. hidden data of single harmonic data.
⚫ Introduction to vector interface
When vector mode is selected in the harmonic measurement mode, to enter
the vector measurement interface, as shown in the figure below.
Item Description
Total THD observe setting THD interface observation switch, this setting is globally
effective and takes precedence over each regulation
module setting item.
Enable: Harmonic analysis according to conditions
Disable: Does not perform harmonic analysis
U-THD Total harmonic distortion of voltage
I-THD Total harmonic distortion of current
Copyright ©ITECH Electronic Co., Ltd. 96
Measurement Functions
Item Description
THC total harmonic current (The total rms value of the
harmonic current components of order 2 to 40)
POHC Partial odd harmonic current
PWHC Partial weighted harmonic current
Catagory Category of Regulations
Subset (IEC61000-3-2) Class A:
Harmonic order h
Max harmonic current Ih/Iref
Class B
Harmonic order h
Max harmonic current Ih/Iref
Class C
Harmonic order h
Max Permissible Ithd%
Class D
Harmonic order h
Max Permissible Ithd per watt mA/W
Max harmonic current Ih/Iref
Note: Parameter items are displayed only and cannot be
edited
Subset (valid for Self- I THD
defined) U THD
U/I THD
1.Form parameter items can be edited, 51 parameters can
be edited
2. I THD or U THD title suffix %f/%r dynamically changes
according to the harmonic interface configuration items
3. I THD/ U THD, The parameter value is set to -1 to delete
the current edit line parameter
4. After editing the parameters, sort them dynamically
from smallest to largest by harmonic order h settings
Type (IEC61000-3-1) Phase type
Non-balance 3-Phase
Balance 3-Phase
Balance 3-Phase(a,b,c)
Balance 3-Phase(d,e,f)
Note: Table parameter entries are displayed only and are
not editable.Iref: reference current, Ih: harmonic current
component
Rise (IEC61000-3-12) Minimum short-circuit ratio limit value
After setting, press Esc to return to the harmonic measurement interface, and
the instrument measurement interface will display the results after harmonic
analysis.
4. Insert the USB disk into the USB interface of the front panel. Press
[Shift]+[V-set](list) on the front panel to enter the List function configuration
interface.
5. Press [Open]. Select the List02.csv file and open it. The List file will be
imported.
NOTE
When using the dimming function, this function will automatically turn off, and when using this
function, dimming function is turned off.
Operating steps
9.3.1 THD
The distorted waveform can simulate voltage harmonic wave in the circuit. The
user can set the extent to which the output voltage waveform deviates from the
Sin wave voltage and test the usage of the DUT under this circumstance.
THD includes built-in 30 waveforms and user - defined waveforms.The interface
is shown below.
DST wave: Select waveforms which built into the instrument by default. When
the waveform name is selected, the parameters and waveform diagram are
displayed on the right side of the interface.
User THD wave: THD wave of user defined.
“+” Create a new THD wave.
“-“ Delete the seleted THD wave.
“ ” Edit the THD wave.
Press the “+” or “ ” enter to the edit interface.
Users can edit a custom harmonic waveform file in .csv format and import it into
the instrument. User defined harmonic waveform template files can be obtained
by contacting ITECH, or an empty file can be exported directly from the
instrument.
The harmonic waveform template description as follows:
Model: instrument model, Keep default.
Device_operation: operation mode of instrument, 0: voltage source, 1: load, 2:
current source, Users are cautious to modify, the waveform files are not common
in different modes.
Usage: Usage, keep default
Name: keep default
Type: Waveform type, harmonic custom waveform is 8, which must be 8 in this
file.
Editable: keep default
Formula: Harmonic calculation formula, 0 represents f%, 1 represents r%
Unit_number: The number of data points edited needs to be consistent with the
number of data rows.
Fund_phase: phase of fundamental waves, range from 0-360°
Order: harmonic order, range from 2-50
Edit interface:
Users can edit a user-defined waveform file in .csv format and import it into the
instrument. User defined harmonic waveform template files can be obtained by
contacting ITECH, or an empty file can be exported directly from the instrument.
The user-defined waveform template description as follows:
Model: keep default
operation mode of instrument, 0: voltage source, 1: load, 2: current source,
Users are cautious to modify, the waveform files are not common in different
modes.
Usage: Usage, keep default
Name: keep default
Type: Waveform type, user-defined waveform is 9, which must be 9 in this file.
Editable: keep default
Point_number: The number of data points to be edited, which needs to be
consistent with the data row in the table. Associated with the Origin_symm
parameter.
Origin_symm: symmetry option, 0 represents 512 non-origin symmetry, 1
represents 512 origin symmetry, 2 represents 1024 points.
Index/fval: Data points and detailed parameter values.
Connecting method
When the voltage signal is sampling, the wiring method is the same as the test
line connection method when outputting. For single-phase, connect the two
terminals A and N, and for three-phase, connect the four terminals ABCN. For
details, refer to section 2.5 Connecting the Test Lines, and take the sampling
DUT supply voltage signal as an example, the wiring diagram is shown below.
Operation method
Click Menu > Volt sample and enter to Voltage signals sampling interface, the
display is shown below.
⚫ File: Input the name of the sampling file, the default display is null, you need
to input the file name before sampling.
⚫ 4: Represents the number of sampling. If you click Run once, the number
of sampling will be increased once and a test-1 file will be formed.
⚫ Tdms: the file format, can be select to Tdms or CSV.
⚫ Export: export the sampling file.
⚫ Run: start to sampling
⚫ Sampling rate: Sampling rate, indicating the sampling interval of data points.
Setting range 1-50, unit kHz
⚫ Start Condition: setting the condition of sampling
Imme: Sampling is executed immediately after clicking Run.
Trigger1: After clicking Run, wait for the Trigger1 signal to be received to
start executing sampling.
Trigger2: After clicking Run, wait for the Trigger2 signal to be received to
start executing sampling.
Capture-Vac: The voltage signal sampling after compared. When this mode
is selected, additional parameters need to be set.
Trigger delay: Set the sample data delay time. Set a positive value to record
data after the capture event occurs. A negative value means that data before
the capture event occurs is recorded.
Compare: Select the object of comparison, either Vac/Rms or Vac/Peak
Up Limit: AC voltage upper limit value, grid voltage greater than the upper
limit value will start sampling.
Down Limit: AC voltage lower limit value, grid voltage less than the lower
limit value will start sampling.
Display mode: Auto/Normal
Parameter introduction when Sweep mode is selected for the function category
Selection: Calculation method selection, %f is equivalent to base wave
percentage mode, Voltage direct voltage setpoint mode.
Thd levels: The interharmonic distortion rate setting, depending on Selection,
can be a percentage and a voltage setting.
Start freq: Interharmonic sweep start frequency setting value.
End freq: Interharmonic sweep end frequency setting value.
Timer type: Select the timing reference, Time method or Cycle method.
Copyright ©ITECH Electronic Co., Ltd. 113
Configuration Arbitrary Waveform
Step time: Interharmonic scanning step time. If Cycle mode, it is the number of
simple harmonic operation cycles (Step cycles).
Step freq: Interharmonic scanning step frequency.
Start angle: Interharmonic start phase angle.
Regenerative
Max.
Regenerativ 5k VA
e power
Output Current < 5%
THD
Others
Efficiency typ 88%
Protection OVP, OCP, OPP, OTP, FAN,ECP,Sens e
dimension HxWxD 483.00mmW *151.3mmH*700mmD (841.6mm include cover and handle)
Weight 26.4kg
Working
temperature 0℃-50℃
Programming
response time 2ms
Remote
Sense 20V
Compensatio
n Voltage
Communication
interface Built-in USB/CAN/LAN/Digital IO interface, optional GPIB /
Analog&RS232
*1 supports single phase 220V input, power cord connecting refer to manual.
*2 Max. Output derated to 60% when input with ( 200~220 ) ±10% Vac 1 ph or 3 ph
*3 Under the output frequency of 50Hz/60Hz, the maximum CF is 6 without exceeding the peak
current; under the condition of full current and full power, the maximum CF is 3.
*4The dynamic response time test condition is: DC high-speed mode, and the capacitance of the
Copyright ©ITECH Electronic Co., Ltd. 116
Technical Specifications
DUT to be less than 10uF
All the above parameters are subject to change without prior notice from ITECH.
When you use the remote interface to send SCPI instructions, if you use the
programming commands that involve modifying the instrument settings,
such as modifying the output voltage value, after completing the
communication connection between the instrument and the host computer,
and after the communication settings are completed, you must execute the
SYST:REM command firstly.
The operation steps to change the USB interface type in System Menu are as
follows.
Connect Interface
Use the following steps to quickly connect your instrument to your LAN and
con- figure it. Two typical LAN interface systems are described below: private
LAN and site LAN.
LAN Config:
⚫ Mode: IP Address seting method, user can selectautomatically configure the
address of the instrument(DHCP) or manually.
⚫ IP: This value is the Internet Protocol (IP) address of the instrument. An IP
address is required for all IP and TCP/IP communications with the instru-
ment. An IP Address consists of 4 decimal numbers separated by periods.
Each decimal number ranges from 0 through 255 with no leading zeros (for
example, 169.254.2.20).
⚫ Mask: This value is used to enable the instrument to determine if a client IP
address is on the same local subnet. The same numbering notation applies
as for the IP Address. When a client IP address is on a different subnet, all
packets must be sent to the Default Gateway.
⚫ Gateway: This value is the IP Address of the default gateway that allows the
instrument to communicate with systems that are not on the local subnet, as
determined by the subnet mask setting. The same numbering notation ap-
plies as for the IP Address. A value of 0.0.0.0 indicates that no default gate-
Copyright ©ITECH Electronic Co., Ltd. 119
Inspection and Installation
way is defined.
⚫ Socket Port: This value indicates the port number corresponding to the
service.
How to Configure
3. Press the Left/Right key to select LAN and set the mode to Manual.
4. Set the IP, Mask and the other parameters in turns, and press [Enter].
You can select different pages by clicking the buttons shown in the navigation bar
on the left side of the window. The detailed descriptions are as follows.
Pins Description
H CAN_H
L CAN_L
CAN Configuration
The user needs to configure the CAN interface parameters in the system
menu before using the remote control. The CAN interface parameters are as
follows.
Name Description
Address Range: 0 – 127
Baud rate Select the baud rate from the following options:
5k/10k/20k/40k/50k/80k/100k/125k/200k/250k/400k
/500k/600k/800k/1000k
CAN Troubleshooting
If you meet some problems when communicating with PC by CAN interface,
Copyright ©ITECH Electronic Co., Ltd. 121
Inspection and Installation
GPIB Configuration
Each device on the GPIB (IEEE-488) interface must have a unique whole num- ber
address between 1 and 30. Your computer’s GPIB interface card address must not
conflict with any instrument on the interface bus. This setting is nonvo- latile; it will
not be changed by *RST.
When you purchase the interface accessory and successfully insert it into the
corresponding position on the rear panel of the instrument, the menu item for
changing the GPIB address appears in the System menu. The specific steps are as
follows:
1. Ensure that the instrument's power switch is off, that is, the instrument is
in Power Off state.
2. Insert the separately purchased GPIB interface card into the card slot on the
rear panel of the instrument.
3. Connect the instrument with the computer via the GPIB cable. After the con-
nection is successful, turn on the power switch of the instrument.
When using the RS-232 interface for communication, connect the pin 1, pin 2, and
pin 3 of the IT-E177 to the PC. The pin description is as follows:
Pins Description
3 DGND, ground
RS–232 Configuration
When you purchase the interface accessory and successfully insert it into the
corresponding position on the rear panel of the instrument, the RS–232 menu
item will appear in the System menu. The specific steps are as follows:
1. Ensure that the instrument's power switch is off, that is, the instrument is
in Power Off state.
2. Insert the separately purchased RS–232 interface card into the card slot on
the rear panel of the instrument.
3. Connect the instrument to the computer via an RS–232 cable. After the con-
nection is successful, turn on the power switch of the instrument.
RS-232 Troubleshooting
If you meet some problems when communicating with PC by RS-232 interface,
please check the following items:
• Check that whether the baud rate of the computer and instrument are
the same;
• Make sure the correct cable and adapter are connected. Note that
internal wiring may not be correct even if the cable has a suitable plug;
• The cable must be connected to the correct serial ports (COM1, COM2,
etc) of PC.
90A
CURRent:PROTection:PEAK 270 //Set the current Peak protect value to
270A
OUTPut ON //Turn on/off the output
MEASure:VOLTage? //Measure the output Vrms
MEASure:CURRent? // Measure the output Irms
MEASure:POWer? // Measure the output power
SYSTem:ERRor? //Query the error information of the instrument
SYSTem:CLEar // To clear the error queue
OUTPut:PROTection:CLEar //To clear the protection status
Appendix
Specifications of Red and Black Test Lines
ITECH provides you with optional red and black test lines, the user can choose
the company's test line for testing. For specifications of ITECH test lines and
maximum current values, refer to the table below.
Model Specifi Length Description
cations
IT-E30110-AB 10A 1m Alligator clips-Banana plugs A
pair of red and black test line
IT-E30110-BB 10A 1m Banana plugs - Banana plugs A
pair of red and black test line
IT-E30110-BY 10A 1m Banana plugs - Y-type terminals
A pair of red and black test line
IT-E30312-YY 30A 1.2m Y-type terminals - A pair of red
and black test line
IT-E30320-YY 30A 2m Y-type terminals - A pair of red
and black test line
IT-E30615-OO 60A 1.5m Ring terminals - A pair of red and
black test line
IT-E31220-OO 120A 2m Ring terminals - A pair of red and
black test line
IT-E32410-OO 240A 1m Ring terminals - A pair of red and
black test line
IT-E32420-OO 240A 2m Ring terminals - A pair of red and
black test line
IT-E33620-OO 360A 2m Ring terminals - A pair of red and
black test line