MRI Questions and Explanations
MRI Questions and Explanations
True
False
True
False
True
False
d. Reducing the TR
True
False
e. Using gadolinium
True
False
imaging time
An MRI scan consists of a number of different sequences (for example T1w, T2w, FLAIR,
Pre-contrast, post-contrast, delayed contrast). To complete one image sequence (for
example the T1weighted images), the entirety of k-space must be filled for each slice
required. The time taken to fill k-space for one slice relates to the number of phase-
encoding steps needed to cover the whole of k-space. The number of phase encoding
steps relates to the pixels in the image. For example, for a 256x256 pixel image there
must be 256 different repetitions of the sequence, each using a different phase
encoding gradient. Each different repetition takes 'TR' (time to repeat) length of time to
occur. This is usually measured in milliseconds (tens to hundreds). So the overall image
acquisition time is determined by multiplying together each of the following:
The imaging time is therefore usually around 10-60 minutes depending on exactly what
is required. Using contrast requires more sequences to be performed (pre, post, and
sometimes many different delayed sequences) so naturally will increase the scanning
time.
In the field of MRI, which of the following regarding the field of view is true?
True
False
True
False
True
False
True
False
e. A rectangular FOV has a reduced signal to noise ratio
True
False
Field of view
The physics behind changing the field of view is difficult to understand but in general,
smaller field of view:
• Higher resolution
• Higher gradient strengths needed
• The returning signal strength is lower
Rectangular fields of view can be used, for example in spine imaging. The number of
phase encoding steps is halved but the difference between each phase encoding step
is doubled such that the remaining steps cover the same overall range. There are two
problems which arise due to this technique however. Firstly, the signal to noise ratio is
reduced and secondly aliasing can occur.
Which of the following statements are true regarding MRI signal decay?
True
False
True
False
c. T2 decay occurs due to interactions with the magnetic fields of individual nuclei
True
False
True
False
e. T1* decay occurs due to magnetic inhomogeneities of the earth’s magnetic field
True
False
Which of the following statements are true regarding relaxation times for a 1.5T
magnet?
True
False
True
False
True
False
True
False
True
False
Spin-spin relaxation (T2 decay, or transverse relaxation) - T2 decay occurs due to spins
dephasing from one another. This occurs due to field inhomogeneities. The
electromagnet produces a very strong magnetic field but no field is 'perfect' i.e. the
strength of the field is not completely uniform throughout. There exist miniscule
fluctuations in the strength of the field which are called 'inhomogeneities'. These
inhomogeneities affect the tiny magnetic fields produced around each spinning proton
to different extents and as the protons move around within the water of the tissues,
the different amounts of energy they have is shared out (hence spin to spin). T2 decay
times are therefore faster in tissues with very little water in them (like bone or air). The
protons in these cannot move around as much, and therefore cannot share their
energies around; they dephase faster. The higher the water content of a tissue, the
longer its T2 decay time is; hence fluid shows as high signal on T2 weighted sequences.
T2 decay is only relevant to the Mxy plane. T1 is also relevant to the M xy plane but also
describes the signal gain in the Mz direction too.
Each proton has a property called 'spin'; they spin like the earth spins on its north-
south axis. The axis is also called the dipole. The direction of the dipole is represented
by an arrow. Normally within a substance there is no overall direction but when that
substance is placed inside a magnetic field the dipoles all align along the field-lines of
the magnetic field pointing either in the same direction or the opposite direction. If the
dipole points along the direction of the field it is described either as 'spin-up' or
parallel. If it points against it, it is termed 'spin-down' or antiparallel.
The electromagnet of an MRI machine is very powerful. Due to the high currents
passing through the coil the apparatus would overheat if it was not supercooled.
Cooling is achieved by bathing the coil in liquid helium at temperatures of -270oC or
4Kelvin. Normally helium is a gas. At 5Kelvin it forms a liquid.
Magnetic fields are measured in either 'Gauss' or 'Tesla' where 1 Gauss is the strength
of the earth's own magnetic field, 1Tesla is equivalent to 10,000Gauss and a typical
diagnostic MRI scanner is 1.5 or 3Tesla in strength.
How many dipoles are in the spin-up state?
The images we produce in MRI depend on the dipoles that are in the spin-down state.
The majority of spins align with along the direction of the magnetic field but a few align
in the opposite direction. At body temperature, about 5 hydrogens out of every one
million align parallel but against the direction of the magnetic field. This is a tiny
number and yet it is the behaviour of these protons which eventually form an MRI
image.
True
False
True
False
False
False
Vectors
A vector is a force in a particular direction. Different vectors with different magnitudes
and directions or angles can be added (or subtracted) to give an overall vector. In MRI
physics we deal with two vectors primarily; the Mz vector, and the Mxy vector.
Regarding MRI safety, which of the following is true?
True
False
True
False
True
False
True
False
True
False
• The magnetic field. The magnetic field is always on. This is because it is
easier/better for the machine to remain on permanently than to repeatedly have
to supercool the coil to the extent needed in order for it to function correctly.
The immediate concerns regarding the static magnetic field relate to risk from
ferromagnetic objects entering the room becoming projectiles. Equipment used
within the MRI room must be made from MR-safe, non-ferrous materials.
Austenitic stainless steel (a type of stainless steel) is MR-safe, as is aluminium.
Cardiac pacemakers are of significant concern due to the risk, not just of lead
movement, but also of abnormal current induction in the leads leading to
arrhythmias. Some types of pacemaker (or neural stimulators) are MR
conditional (an update concept of MR-safe, with limits on the strength of the
magnetic field they have been tested to), although should still be switched off
temporarily for the duration of the scan.
• The gradient magnetic fields. These are made to apply in various different
directions for varying amounts of time during the scan. Notable side effects
include peripheral nerve stimulation giving the sensation of tingling, but also a
metallic taste in the mouth, vertigo and nausea. Involuntary muscular
contractions and cardiac arrythmias have also been reported. These effects are
all more prevalent when the background magnetic field strength is more
significant (for example in a 3T magnet as opposed to a 1.5T one). Obviously the
shorter the duration of the variations occurring, the lower the risk of side effects
occurring. The risk of the gradient magnetic fields to the developing foetus are
theoretical however it is to guard against the effects of this aspect of the scan
that scans are usually ideally not performed on or by patients/staff in their first
trimester of pregnancy (where declared), though most follow-up studies do not
show noticeable effects for the foetus.
The second aspect of the gradient magnetic fields which causes a problem is the
noise incurred when the gradients are applied. Having an MRI scan is a very
noisy experience. The noise comes from switching on and off the gradient
magnetic fields. Patients are given ear defenders to guard against hearing
damage, and for their own comfort.
a. Short TR
True
False
True
False
True
False
d. Reduced SNR
True
False
e. Use of gadolinium
True
False
Scan time
Scanning time is an important parameter. Longer scans are associated with
significantly more patient discomfort which in turn leads to motion artefact and in
many cases, early termination of the scan without the necessary sequences having
been acquired. The following all reduce scan time but many of these must be traded
off to preserve image quality:
• Short TR – this causes a reduction in SNR and causes the images to be more
heavily T1 weighted
• Reduced number of phase encoding steps – i.e. reduced phase matrix. This
reduces the resolution
• Reduced number of excitations. This reduces the SNR and increases the effect
from motion artefact if there is any.
a. Improving SNR
True
False
b. Gadolinium
True
False
c. SPIOs
True
False
True
False
True
False
True
False
c. Using a larger matrix
True
False
True
False
e. Using gadolinium
True
False
Contrast
Contrast between different tissues within an MRI image depends on the comparison of
the SNRs within those tissues. Contrast does therefore depend, to an extent, on the
SNR. Other techniques can help improve the contrast between structures as well:
• Fat suppression – This is a useful technique for some specific purposes. Fat
suppression is performed with post-gadolinium images such that the high signal
on the T1 images used to show the contrast medium, can be assured not to
have come from fat. STIR images are fat supressed.
• Contrast media – such as gadolinium or SPIOs help promote image contrast by
returning more signal from structures which contain the contrast.
• Magnetisation transfer constant – this involves including in the sequence used,
some RF pulses delivered at specific off-resonant frequencies. These are
targeted at protons bound within big molecules (instead of in water molecules),
and they help reduce the signal from the big molecules so that the signal from
the water molecules is better appreciated
Spatial resolution
The spatial resolution of an MRI image relates to the size of the pixels being used. This
itself depends on the FOV (field of view) which is selected on the scout images, and on
the matrix selected on the computer. Smaller pixels allow smaller structures to be
distinguished hence smaller pixels improve spatial resolution. The following all improve
spatial resolution:
• Smaller pixels
• Smaller FOV
• Larger matrix
• Use of a local coil (e.g. a head coil, as opposed to the large coil contained within
the gantry)
• Thinner slices (reduce the partial volume effect).
True
False
b. Each horizontal line represents data from each frequency encoded gradient
applied
True
False
True
False
True
False
e. In a normal spin-echo sequence, many lines of k-space can be acquired during
each TR
True
False
Only one line of k-space can be acquired in a TR during a normal spin-echo sequence.
More can be acquired during a turbo spin echo sequence though.
Spatial encoding
This is the process of encoding the returning signal such that we can work out where in
the body the signal has come from. This has to be done in all three directions (x, y and
z).
K-space**
K-space is the theoretical graph which has to be filled in for each slice before the image
can be produced. It is filled in line-by-line by successive repetitions of each sequence.
For example, in a spin-echo sequence, the first TR might be used to collect the bottom
line along k-space, then the phase encoding gradient is changed for the next TR such
that the next line up can be read and added. Only once the entirety of the k-space
graph has been collected for that slice can a Fourier transform be applied to the whole
thing, the result of which is the diagnostic image for that slice.
Image © Medical Exam Prep
It is important to note that each area in k-space does not correspond to individual
pixels in the final image. Each point in k-space contains frequency and phase encoded
information which relate to the picture as a whole.
The centre of k-space contains spatial frequencies relating to basic image contrast.
The periphery of k-space contains spatial frequencies relating to the detail within the
image, the edges or boundaries between two tissue types.
Image © Medical Exam Prep
**The letter 'k' itself comes from the universal term given to the concept of a
'wavenumber'. A wavenumber is 1 divided by the wavelength i.e. it is the number of
wavelengths per unit distance.
Which of the following regarding the radiofrequency pulses used in MRI scanning is
true?
False
False
d. After a 180o pulse there are more spins pointing antiparallel than parallel to M z
True
False
e. Both 90o and 180o pulses are emitted at the same frequency; the Larmor
frequency
True
False
• Some of the spin-up protons gain enough energy to 'flip' to become spin-down
protons. This causes a change in the Mz vector. The Mz vector can increase,
decrease or disappear completely depending on the strength and duration of
the RF pulse.
• The protons in the tissue become are forced to precess in phase with one
another. This means that their hitherto random distribution (like clocks in
different time zones) in the Mxy plane becomes a single magnetic vector in that
plane, spinning around like clocks in synch. Their synchronisation is responsible
for a net magnetic vector being produced in the Mxy plane.
90o pulse
A 90o pulse is one which tips the net magnetisation (by 90o (ie into the xy plane), and
also causes re-phasing such that the spins are all precessing in phase with one another.
180o pulse
A 180o pulse is one which tips the net magnetisation by 180o (ie into the opposite
direction to the Mz direction), and also causes re-phasing such that the spins are all
precessing in phase with one another. Prior to a 180o pulse there is a tiny excess in
protons pointing parallel to the direction of the magnetic field compared to pointing
anti-parallel. After the 180othis excess is reversed; there are slightly more protons
pointing antiparallel.
Which of the following statements are true regarding phase encoding?
True
False
True
False
True
False
d. Phase encoding must be repeated as many times as there are pixels in that
direction
True
False
e. Within k-space, each horizontal line represents signal from a different phase
encoding gradient
True
False
patial encoding
This is the process of encoding the returning signal such that we can work out where in
the body the signal has come from. This has to be done in all three directions (x, y and
z).
Phase encoding - this process encodes one of the other two axes, typically the y-axis
(front to back). After the slice has been selected, a gradient is applied in the y-direction
which alters the phases of the spins from the front of the patient to the back. This is
akin to time-zones around the world. Time is still passing with the same speed but the
clock hands are pointing to different numbers. The gradient applied can be a strong
gradient (where the phase shifts a lot from front to back), or a shallow gradient (where
the phase shifts only a small amount from front to back), or anything in between. The
sequence is repeated with many different gradient strengths, one after another and
each time, when the signal is read out, comparing the phase of the waves between
successive gradient strengths allows us (a computer) to deduce where along the y-axis
those spins gave signal. The phase shift is proportional to the position of the spins
generating the signal, along the y-axis. The resultant data is added line by line to
something called k-space which is a theoretical graph, and the whole of k-space must
be filled in before an image can be produces from it. To collect enough data there must
be as many repetitions, all with different strength phase gradients applied as there are
number of pixels from front to back.
Each line added to k-space represents the signal collected under different strengths of
phase-encoding gradient. There might be a strong gradient where the magnitude of
the phase shift changes significantly over the width of the slice, or a mild gradient
where the magnitude of the phase shift changes only slightly over the width of the
slice. The line added to k-space does not itself relate to position along the y-axis, only
that the phase shifts undergone by the different frequencies within that signal are
proportional to position along the y-axis. Phase encoding is complicated!
Image © Medical Exam Prep
k-space**
k-space is the theoretical graph which has to be filled in for each slice before the image
can be produced. It is filled in line-by-line by successive repetitions of each sequence.
For example, in a spin-echo sequence, the first TR might be used to collect the bottom
line along k-space, then the phase encoding gradient is changed for the next TR such
that the next line up can be read and added. Only once the entirety of the k-space
graph has been collected for that slice can a Fourier transform be applied to the whole
thing, the result of which is the diagnostic image for that slice.
It is important to note that each area in k-space does not correspond to individual
pixels in the final image. Each point in k-space contains frequency and phase encoded
information which relate to the picture as a whole.
**The letter 'k' itself comes from the universal term given to the concept of a
'wavenumber'. A wavenumber is 1 divided by the wavelength i.e. it is the number of
wavelengths per unit distance.
True
False
True
False
True
False
True
False
True
False
Motion artefact
Frequency encoding takes place very quickly, within a fraction of a millisecond and as
such is not as severely affected by motion as phase encoding, which takes place over a
matter of hundreds-thousands of milliseconds. This means that on the resultant image,
the effects of patient motion are only really noticeable in the phase-encoding direction.
For structures which pulsate (e.g. the aorta), there can be multiple images of the same
structure all lined up on top of one another along the phase encoding gradient. This is
known as ghosting.
Magnetic susceptibility
The metal of a joint replacement distorts the magnetic field around the joint and alters
the signal returning from that area. The affected area is usually black with anatomical
distortion at the point where the signal becomes recognisable.
Its paramagnetic properties cause significant spin dephasing in any spins in its vicinity.
It reduces both T1 and T2 however it has a far greater effect on T1 than it does on T2,
hence all IV contrast sequences are performed as T1 weighted sequences.
Gd-DTPA does not cross the normal blood-brain barrier so can be used to discern
when this has been pathologically breached.
Contrast – endogenous
Arterial spin labelling – this is a way of providing vascular contrast without using an
exogenous substance. It relies on the different T1 re-growth times of blood as
compared to surrounding tissue. Two images are acquired, one where the signal from
blood is nulled by spin inversion, and a second control image. When the two images
are combined, an image of perfusion can be obtained.
Time of flight imaging – This technique relies on the blood in the slice downstream
from the slice of interest is targeted and the spins excited by RF pulses. As the blood in
this upstream slice flows into the target slice downstream, the spins from the blood
show bright against the slice which hasn’t received these initial RF pulses.
True
False
b. It is a qualitative test
True
False
True
False
d. NAA is usually pathological if low
True
False
True
False
True
False
True
False
True
False
d. Using gadolinium
True
False
True
False
Field of view
The physics behind changing the field of view is difficult to understand but in general,
smaller field of view:
• Higher resolution
• Higher gradient strengths needed
• The returning signal strength is lower
Rectangular fields of view can be used, for example in spine imaging. The number of
phase encoding steps is halved but the difference between each phase encoding step
is doubled such that the remaining steps cover the same overall range. There are two
problems which arise due to this technique however. Firstly, the signal to noise ratio is
reduced and secondly aliasing can occur.
Aliasing, or wrap-around artefact
This occurs when the body part being imaged is bigger than the field of view selected.
The signal coming from the part of the body outside the field of view is misinterpreted
as coming from the other side of the image. Although this can theoretically occur in
both the phase encoding and the frequency encoding gradients, in practice the phase
encoding gradient is only applied to the field of view. Aliasing occurs in the frequency
encoding gradient if the returning signal is not sampled frequently enough. To avoid
this artefact the signal must be sampled at at least twice the wavelength of the highest
frequency. This (sampling twice per wavelength) is known as the Nyquist criteria. If the
signal is under-sampled then pixels from outside of the field of view are accidentally
mapped onto areas within the field of view. The classic example is the nose of a patient
appearing as an occipital mass.
True
False
True
False
True
False
True
False
e. Using SPIOs
True
False
Regarding contrast agents used in MRI scanning, which of the following is true?
True
False
True
False
True
False
True
False
True
False
Contrast agents
The composition of gadolinium (Gd) ions includes seven unpaired electrons. These
unpaired electrons make them strongly paramagnetic. Unfortunately, gadolinium is
very toxic so must be attached or chelated to other compounds which bind the
gadolinium tightly, allowing it first to dissolve in water and then to be excreted by the
body shortly following administration.
Its paramagnetic properties cause significant spin dephasing in any spins in its vicinity.
It reduces both T1 and T2 however it has a far greater effect on T1 than it does on T2,
hence all IV contrast sequences are performed as T1 weighted sequences. The
reduction of the spin-lattice relaxation time (T1) caused by the gadolinium means that
tissues containing the contrast return a higher signal on T1 weighted sequences.
Gd-DTPA does not cross the normal blood-brain barrier so can be used to discern
when this has been pathologically breached.
Direction of Relative
Property Typical material
polarisation strength
Gadolinium, Manganese,
Paramagnetic Same +1
Copper
• Diamagnetic materials have paired electrons and have a negative (albeit weak)
susceptibility to surrounding magnetic fields. They align against the direction of
polarisation but revert to a random distribution if the magnetic field is removed.
• Ferromagnetic materials have at least one unpaired electron. They are strongly
attracted to align in the same direction as the magnetic field and, once the
magnetic field is removed, remain that way aligned afterwards, retaining also
magnetisation of their own.
• Paramagnetic materials are weakly attracted to align in the same direction as
the magnetic field. They also have at least one unpaired electron.
Which of the following statements are true regarding k-space?
True
False
True
False
True
False
True
False
e. K-space must be filled in for both the phase encoding gradient and the
frequency encoding gradient, then the two are averaged
True
False
K-space is filled in once per slice with data from each repetition of the sequence
occurring under each new phase encoding gradient.
Spatial encoding
This is the process of encoding the returning signal such that we can work out where in
the body the signal has come from. This has to be done in all three directions (x, y and
z).
K-space**
K-space is the theoretical graph which has to be filled in for each slice before the image
can be produced. It is filled in line-by-line by successive repetitions of each sequence.
For example, in a spin-echo sequence, the first TR might be used to collect the bottom
line along k-space, then the phase encoding gradient is changed for the next TR such
that the next line up can be read and added. Only once the entirety of the k-space
graph has been collected for that slice can a Fourier transform be applied to the whole
thing, the result of which is the diagnostic image for that slice.
It is important to note that each area in k-space does not correspond to individual
pixels in the final image. Each point in k-space contains frequency and phase encoded
information which relate to the picture as a whole.
The centre of k-space contains spatial frequencies relating to basic image contrast.
The periphery of k-space contains spatial frequencies relating to the detail within the
image, the edges or boundaries between two tissue types.
Image © Medical Exam Prep
**The letter 'k' itself comes from the universal term given to the concept of a
'wavenumber'. A wavenumber is 1 divided by the wavelength i.e. it is the number of
wavelengths per unit distance.
True
False
True
False
False
True
False
True
False
True
False
True
False
True
False
True
False
True
False
Sequences
To produce a signal from the tissue of interest, different strength pulses are used for
different durations, in different orders. Each set of pulses is known as a 'sequence'. The
sequences are designed for different purposes for example:
• T1 weighted sequence - fat, blood, melanin, proteinaceous fluid and gadolinium
are all high in signal on a T1 weighted sequence. This sequence can be used with
or without fat suppression. Once the signal from fat has been supressed,
gadolinium based IV contrast can also be given and imaged using this sequence.
• T2 weighted sequence - fluid and fat are both high in signal
• PD sequence - proton density sequence, similar to T2, used a lot for MSK
imaging and neuroradiology.
• Diffusion weighted sequences - These are concerned with the movement of
water molecules within tissues. Anisotropy (An-eye-sot-rop-y) is a property which
describes how non-linear the diffusion of water through a tissue is. In highly
organised/structured tissues such as cardiac muscle fibres, or neuronal
networks, there is a much stronger anisotropic diffusion of water through the
tissue along the lines of the muscle fibres or neurones. Certain pathologies (in
brain imaging: infarction, demyelination, pus, and some cancers) can disrupt this
anisotropy and 'restrict diffusion'. DW sequences must always be coupled with
an 'ADC' map (Apparent Diffusion Coefficient). An ADC map is produced by
eliminating the effect of T2 from the DW sequence; DW sequences are
inherently very heavily T2 weighted. If something is high in signal intensity on
both sequences, this is due to its T2 effects (T2-shine through), and not its
diffusion properties.
• Flow-related sequences - there are a few different varieties of these. Some use
injected contrast, and some rely on the movement of blood around the field of
view.
Which of the following regarding the 90o radiofrequency pulses used in MRI scanning is
true?
a. The RF pulses are set to emit the frequency of the Larmor frequency of water
hydrogen nuclei
True
False
True
False
c. Protons rephase during the duration of the RF pulse
True
False
False
False
• Some of the spin-up protons gain enough energy to 'flip' to become spin-down
protons. This causes a change in the Mz vector. The Mz vector can increase,
decrease or disappear completely depending on the strength and duration of
the RF pulse.
• The protons in the tissue become are forced to precess in phase with one
another. This means that their hitherto random distribution (like clocks in
different time zones) in the Mxy plane becomes a single magnetic vector in that
plane, spinning around like clocks in synch. Their synchronisation is responsible
for a net magnetic vector being produced in the Mxy plane.
Which of the following statements are true or false regarding MRI sequences?
True
False
True
False
True
False
True
False
True
False
Sequences
To produce a signal from the tissue of interest, different strength pulses are used for
different durations, in different orders. Each set of pulses is known as a 'sequence'. The
sequences are designed for different purposes for example:
Gradient-echo sequence: Useful if the sequence needs to have a short TR. This
sequence suffers from magnetic inhomogeneities because there is no 180 o pulse to
'refocus' the spins which have suffered from this effect. Consequently, any T2 effect
seen is referred to as T2* (T2 star). Local susceptibility artefact can be used to
diagnostic advantage in many circumstances though. In cerebral imaging, if there are
microhaemorrhages, the presence of excess iron causes spin dephasing leading to
'blooming' artefact.
Spin-echo Gradient-echo
In the field of MRI, regarding the concept of spin relaxation, which of the following is
true?
True
False
True
False
True
False
True
False
• Some of the spin-up protons gain enough energy to 'flip' to become spin-down
protons. This causes a change in the Mz vector. The Mz vector can increase,
decrease or disappear completely depending on the strength and duration of
the RF pulse.
• The protons in the tissue become are forced to precess in phase with one
another. This means that their hitherto random distribution (like clocks in
different time zones) in the Mxy plane becomes a single magnetic vector in that
plane, spinning around like clocks in synch. Their synchronisation is responsible
for a net magnetic vector being produced in the Mxy plane.
They undergo 'relaxation' which reverts them (as a population) to their original states.
This is either spin-lattice relaxation or spin-spin relaxation.
Spin-spin relaxation (T2 decay, or transverse relaxation) - T2 decay occurs due to spins
dephasing from one another. This occurs due to field inhomogeneities. The
electromagnet produces a very strong magnetic field but no field is 'perfect' i.e. the
strength of the field is not completely uniform throughout. There exist miniscule
fluctuations in the strength of the field which are called 'inhomogeneities'. These
inhomogeneities affect the tiny magnetic fields produced around each spinning proton
to different extents and as the protons move around within the water of the tissues,
the different amounts of energy they have is shared out (hence spin to spin). T2 decay
times are therefore faster in tissues with very little water in them (like bone or air). The
protons in these cannot move around as much, and therefore cannot share their
energies around; they dephase faster. The higher the water content of a tissue, the
longer its T2 decay time is; hence fluid shows as high signal on T2 weighted sequences.
T2 decay is only relevant to the Mxy plane. T1 is also relevant to the M xy plane but also
describes the signal gain in the Mz direction too.
Spleen 400 60
Kidney 550 60
Liver 400 40
Fat 250 80
Which of the following statements are true regarding proton density weighted
sequences?
True
False
True
False
c. This is one of the only sequences where bone returns high signal
True
False
d. Has a long TR
True
False
e. Is used a lot for neuroradiology
True
False
Sequence TR (ms) TE (ms)
T1w (both short) 500 15
T2w (both long) 4000 90
FLAIR (both very long) 9000 115
Proton density (long TR,
2000 15
short TE)
Sequences
To produce a signal from the tissue of interest, different strength pulses are used for
different durations, in different orders. Each set of pulses is known as a 'sequence'. The
sequences are designed for different purposes for example:
Regarding MRI safety considerations during pregnancy, which of the following is true?
True
False
True
False
True
False
True
False
True
False
Gadolinium in pregnancy
Gadolinium also crosses the placenta and is excreted by the foetus into the amniotic
fluid. The chelated form of gadolinium in itself poses no risk to the foetus however the
free form of gadolinium is teratogenic and the longer the foetus is exposed to the
chelated gadolinium administered for the scan, the greater the risk of each molecule of
gadolinium becoming dissociated from its chelate. For this reason therefore,
gadolinium should only be administered in a patient known to be pregnant, if there is
likely to be significant diagnostic benefit from the resultant images.
In the field of MRI, regarding the concept of spin relaxation, which of the following is
true?
a. Spin-lattice interactions cause longitudinal relaxation
True
False
True
False
True
False
d. Both T1 recovery and T2 decay are concerned with regrowth of the M z vector
True
False
True
False
• Some of the spin-up protons gain enough energy to 'flip' to become spin-down
protons. This causes a change in the Mz vector. The Mz vector can increase,
decrease or disappear completely depending on the strength and duration of
the RF pulse.
• The protons in the tissue become are forced to precess in phase with one
another. This means that their hitherto random distribution (like clocks in
different time zones) in the Mxy plane becomes a single magnetic vector in that
plane, spinning around like clocks in synch. Their synchronisation is responsible
for a net magnetic vector being produced in the Mxy plane.
They undergo 'relaxation' which reverts them (as a population) to their original states.
This is either spin-lattice relaxation or spin-spin relaxation.
Spin-spin relaxation (T2 decay, and T2* decay, or transverse relaxation) - T2 decay
occurs due to spins dephasing from one another causing a reduction in the height of
the Mz vector. T2* is when this occurs due to the effects of field inhomogeneities, in
addition to the effects of normal molecule interactions. The electromagnet produces a
very strong magnetic field but no field is 'perfect' i.e. the strength of the field is not
completely uniform throughout. There exist miniscule fluctuations in the strength of
the field which are called 'inhomogeneities'. These inhomogeneities affect the tiny
magnetic fields produced around each spinning proton to different extents and as the
protons move around within the water of the tissues, the different amounts of energy
they have is shared out (hence spin to spin). T2 decay times are therefore faster in
tissues with very little water in them (like bone or air). The protons in these cannot
move around as much, and therefore cannot share their energies around; they
dephase faster. The higher the water content of a tissue, the longer its T2 decay time is;
hence fluid shows as high signal on T2 weighted sequences.
Image © Medical Exam Prep
T2 decay is only relevant to the Mxy plane. T1 is also relevant to the M xy plane but also
describes the signal gain in the Mz direction too.
Spleen 400 60
Kidney 550 60
Liver 400 40
Fat 250 80
True
False
True
False
True
False
True
False
True
False
In the field of MRI, regarding quenching the electromagnet, which of the following is
correct?
True
False
True
False
d. The valves in the door to an MRI room are to equalise the air pressure in the
case of a quench
True
False
True
False
Each atom has a nucleus and each nucleus contains a specific number of protons
(plus/minus neutrons) depending on which element it is. Nuclei of elements which have
an odd number of protons in their nuclei (1,3,5,7 etc.) possess a property called nuclear
magnetic resonance. Hydrogen nuclei which consist only of one single proton are
present within water in almost everything we are looking to image with MRI and so for
the most part, it is the behaviour hydrogen nuclei with which we are most interested in
seeking to understand how MRI works. Of all the atoms in a human body around 63%
(by number) are hydrogen.
Image © Medical Exam Prep
Each proton has a property called 'spin'; they spin like the earth spins on its north-
south axis. The axis is also called the dipole. The direction of the dipole is represented
by an arrow. Normally within a substance there is no overall direction but when that
substance is placed inside a magnetic field the dipoles all align along the field-lines of
the magnetic field pointing either in the same direction or the opposite direction. If the
dipole points along the direction of the field it is described either as 'spin-up' or
parallel. If it points against it, it is termed 'spin-down' or antiparallel.
Image © Medical Exam Prep
The electromagnet of an MRI machine is very powerful. Due to the high currents
passing through the coil the apparatus would overheat if it was not supercooled.
Cooling is achieved by bathing the coil in liquid helium at temperatures of -270oC or
4Kelvin. Normally helium is a gas. At 5Kelvin it forms a liquid.
Magnetic fields are measured in either 'Gauss' or 'Tesla' where 1 Gauss is the strength
of the earth's own magnetic field, 1Tesla is equivalent to 10,000Gauss and a typical
diagnostic MRI scanner is 1.5 or 3Tesla in strength.
Precession
'Precession' is the term used to describe the spinning action of a proton which is, still
spinning on its axis, but moving so that the ends of the axis seem to trace a circle.
Individually, these spinning precessing protons generate minute magnetic fields
themselves. Under the influence of a strong external magnetic field, these minute
magnetic fields line up in parallel with the field (either with it or against it). The result is
that the 'net magnetisation' is zero. This is not to say that in the context of a static
magnetic field the protons stop precessing; they don't. Only that they precess in such a
way that their tiny magnetic fields, when summed together create a net magnetisation
of zero. To create signal, radiofrequency (RF) pulses are applied to the patient which
cause the axes to tip-over. When the RF pulses are switched off, the precession forms a
spiral path to return their original positions.
Image © Medical Exam Prep
Terminology
Bo - This is the direction of the magnetic field.
Mz - When an RF pulse is removed, the z-component of the spins increases or 'grows'.
Phrases such as 're-growth of the Mz vector' are used.
Mxy - When an RF pulse is removed, the spins lose strength in the M xy plane. This is
described as 'decay of the Mxy plane'.
The Larmor frequency
The frequency (in MHz) with which protons precess is dependent on a constant called
the gyromagnetic ratio, which is different for the protons in different nuclei.
The Larmor frequency (ω) = gyromagnetic ratio (γ) x magnetic field strengh (B o)
Historically DWI and EPI imaging was achieved on gradient echo sequences but these
days different manufacturers use different sequences, either gradient or spin echo
sequences to produce diffusion weighting.
Noise in MRI
Noise in MRI comes from a variety of different sources.
SNR in MRI
There are a number of different parameters which can improve SNR in MRI:
True
False
True
False
c. Slice selection is achieved by causing the magnetic field to vary from one end of
the scanner to the other
True
False
d. Slice select relies on varying the Larmor frequency from one end of the patient
to another
True
False
e. The Larmor frequency is always at least 42.6MHz or higher (for a 1.5T magnet)
True
False
Spatial encoding
This is the process of encoding the returning signal such that we can work out where in
the body the signal has come from. This has to be done in all three directions (x, y and
z).
Slice select - this process encodes which slice we are imaging, i.e. the z-axis. The
electromagnet strength is varied linearly from one end of the patient to the other. For
example, if we are imaging the brain then at the vertex the magnetic field is made to be
very slightly stronger and at the base of the skull it is made to be very slightly weaker.
Or the variation can be across the whole body. Variation across shorter lengths allows
for thinner slices to be selected. The strength of the magnetic field determines the
precession frequency of the spins within that slice so to 'select the slice' to image, the
RF pulses are applied at the Larmor frequency according to the slice needed. The
magnetic field can be made to vary through the use of additional coils which act
against the background of the existing magnetic field.
Which of the following are true regarding frequency encoding during an MRI scan?
True
False
b. The strength of the magnetic field is made to vary from left to right
True
False
c. The frequency encoding gradient is applied at the same time as slice selection
True
False
d. A Fourier transform is used to show the different frequencies as amplitudes
True
False
True
False
To produce these sequences, different combinations of radiofrequency pulses are
used.
Gradient-echo sequence: Useful if the sequence needs to have a short TR. This
sequence suffers from magnetic inhomogeneities because there is no 180 o pulse to
'refocus' the spins which have suffered from this effect. Consequently, any T2 effect
seen is referred to as T2* (T2 star). Local susceptibility artefact can be used to
diagnostic advantage in many circumstances though. In cerebral imaging, if there are
microhaemorrhages, the presence of excess iron causes spin dephasing leading to
'blooming' artefact.
Spin-echo Gradient-echo
Sequences
To produce a signal from the tissue of interest, different strength pulses are used for
different durations, in different orders. Each set of pulses is known as a 'sequence'. The
sequences are designed for different purposes for example:
Inversion recovery
This sequence is a little bit like a spin-echo sequence but with an extra gradient bolted
onto the front. To begin with, a 180o inversion sequence is used which flips the
Mz vector upside down. Different tissues revert at slightly different rates, with the
Mz vector gradually growing. Within each tissue type, there becomes a point in time
(specific to that tissue type) where the Mz vector is zero.
Image © Medical Exam Prep
By choosing the tissue type (e.g. water for FLAIR, or fat for STIR imaging), the normal
spin-echo sequence can be made to begin at exactly the point at which the Mz signal
from that tissue type is ‘nulled’. This means that that tissue type then does not
participate in the remainder of the sequence; signal comes only from other tissues.
This is referred to as that tissue type being ‘nulled’.
Image © Medical Exam Prep
FLAIR stands for Fluid Attenuation Inversion Recovery. FLAIR images show fluid as
black.
STIR stands for Short T1 Inversion Recovery. (Fat has a short T1). STIR images cause the
signal from fat to be supressed.
• FLAIR stands for Fluid Attenuation Inversion Recovery. FLAIR images show fluid
as black.
• STIR stands for Short T1 Inversion Recovery. (Fat has a short T1). STIR images
cause the signal from fat to be supressed.
• PD sequence - proton density sequence, similar to T2, used a lot for MSK
imaging and neuroradiology.
• Diffusion weighted sequences - These are concerned with the movement of
water molecules within tissues. Anisotropy (An-eye-sot-rop-y) is a property which
describes how non-linear the diffusion of water through a tissue is. In highly
organised/structured tissues such as cardiac muscle fibres, or neuronal
networks, there is a much stronger anisotropic diffusion of water through the
tissue along the lines of the muscle fibres or neurones. Certain pathologies (in
brain imaging: infarction, demyelination, pus, and some cancers) can disrupt this
anisotropy and 'restrict diffusion'. DW sequences must always be coupled with
an 'ADC' map (Apparent Diffusion Coefficient). An ADC map is produced by
eliminating the effect of T2 from the DW sequence; DW sequences are
inherently very heavily T2 weighted. If something is high in signal intensity on
both sequences, this is due to its T2 effects (T2-shine through), and not its
diffusion properties.
Contrast agents
The composition of gadolinium (Gd) ions includes seven unpaired electrons. These
unpaired electrons make them strongly paramagnetic. Unfortunately, gadolinium is
very toxic so must be attached or chelated to other compounds which bind the
gadolinium tightly, allowing it first to dissolve in water and then to be excreted by the
body shortly following administration.
Its paramagnetic properties cause significant spin dephasing in any spins in its vicinity.
It reduces both T1 and T2 however it has a far greater effect on T1 than it does on T2,
hence all IV contrast sequences are performed as T1 weighted sequences.
Gd-DTPA does not cross the normal blood-brain barrier so can be used to discern
when this has been pathologically breached.
Direction of Relative
Property Typical material
polarisation strength
• Diamagnetic materials have paired electrons and have a negative (albeit weak)
susceptibility to surrounding magnetic fields. They align against the direction of
polarisation but revert to a random distribution if the magnetic field is removed.
• Ferromagnetic materials have at least one unpaired electron. They are strongly
attracted to align in the same direction as the magnetic field and, once the
magnetic field is removed, remain that way aligned afterwards, retaining also
magnetisation of their own.
• Paramagnetic materials are weakly attracted to align in the same direction as
the magnetic field. They also have at least one unpaired electron.
In the field of MRI, regarding the concept of spin relaxation, which of the following is
true?
True
False
True
False
c. T2* decay is caused by magnetic field inhomogeneities
True
False
d. Both T1 recovery and T2 decay are concerned with regrowth of the M z vector
True
False
e. Transverse relaxation has a greater effect in more rigid tissues
True
False
• Some of the spin-up protons gain enough energy to 'flip' to become spin-down
protons. This causes a change in the Mz vector. The Mz vector can increase,
decrease or disappear completely depending on the strength and duration of
the RF pulse.
• The protons in the tissue become are forced to precess in phase with one
another. This means that their hitherto random distribution (like clocks in
different time zones) in the Mxy plane becomes a single magnetic vector in that
plane, spinning around like clocks in synch. Their synchronisation is responsible
for a net magnetic vector being produced in the Mxy plane.
They undergo 'relaxation' which reverts them (as a population) to their original states.
This is either spin-lattice relaxation or spin-spin relaxation.
Spin-spin relaxation (T2 decay, and T2* decay, or transverse relaxation) - T2 decay
occurs due to spins dephasing from one another causing a reduction in the height of
the Mz vector. T2* is when this occurs due to the effects of field inhomogeneities, in
addition to the effects of normal molecule interactions. The electromagnet produces a
very strong magnetic field but no field is 'perfect' i.e. the strength of the field is not
completely uniform throughout. There exist miniscule fluctuations in the strength of
the field which are called 'inhomogeneities'. These inhomogeneities affect the tiny
magnetic fields produced around each spinning proton to different extents and as the
protons move around within the water of the tissues, the different amounts of energy
they have is shared out (hence spin to spin). T2 decay times are therefore faster in
tissues with very little water in them (like bone or air). The protons in these cannot
move around as much, and therefore cannot share their energies around; they
dephase faster. The higher the water content of a tissue, the longer its T2 decay time is;
hence fluid shows as high signal on T2 weighted sequences.
T2 decay is only relevant to the Mxy plane. T1 is also relevant to the M xy plane but also
describes the signal gain in the Mz direction too.
Spleen 400 60
Kidney 550 60
Liver 400 40
Fat 250 80
True
False
True
False
False
False
Vectors
A vector is a force in a particular direction. Different vectors with different magnitudes
and directions or angles can be added (or subtracted) to give an overall vector. In MRI
physics we deal with two vectors primarily; the Mz vector, and the Mxy vector.
True
False
True
False
d. Patients with a cardiac pacemaker cannot have MRI scans
True
False
True
False
• The magnetic field. The magnetic field is always on. This is because it is
easier/better for the machine to remain on permanently than to repeatedly have
to supercool the coil to the extent needed in order for it to function correctly.
The immediate concerns regarding the static magnetic field relate to risk from
ferromagnetic objects entering the room becoming projectiles. Equipment used
within the MRI room must be made from MR-safe, non-ferrous materials.
Austenitic stainless steel (a type of stainless steel) is MR-safe, as is aluminium.
Cardiac pacemakers are of significant concern due to the risk, not just of lead
movement, but also of abnormal current induction in the leads leading to
arrhythmias. Some types of pacemaker (or neural stimulators) are MR
conditional (an update concept of MR-safe, with limits on the strength of the
magnetic field they have been tested to), although should still be switched off
temporarily for the duration of the scan.
• The gradient magnetic fields. These are made to apply in various different
directions for varying amounts of time during the scan. Notable side effects
include peripheral nerve stimulation giving the sensation of tingling, but also a
metallic taste in the mouth, vertigo and nausea. Involuntary muscular
contractions and cardiac arrythmias have also been reported. These effects are
all more prevalent when the background magnetic field strength is more
significant (for example in a 3T magnet as opposed to a 1.5T one). Obviously the
shorter the duration of the variations occurring, the lower the risk of side effects
occurring. The risk of the gradient magnetic fields to the developing foetus are
theoretical however it is to guard against the effects of this aspect of the scan
that scans are usually ideally not performed on or by patients/staff in their first
trimester of pregnancy (where declared), though most follow-up studies do not
show noticeable effects for the foetus.
The second aspect of the gradient magnetic fields which causes a problem is the
noise incurred when the gradients are applied. Having an MRI scan is a very
noisy experience. The noise comes from switching on and off the gradient
magnetic fields. Patients are given ear defenders to guard against hearing
damage, and for their own comfort.
True
False
True
False
True
False
True
False
e. Gadolinium has super-paramagnetic properties
True
False
Motion artefact
Frequency encoding takes place very quickly, within a fraction of a millisecond and as
such is not as severely affected by motion as phase encoding, which takes place over a
matter of hundreds-thousands of milliseconds. This means that on the resultant image,
the effects of patient motion are only really noticeable in the phase-encoding direction.
For structures which pulsate (e.g. the aorta), there can be multiple images of the same
structure all lined up on top of one another along the phase encoding gradient. This is
known as ghosting.
Magnetic susceptibility
The metal of a joint replacement distorts the magnetic field around the joint and alters
the signal returning from that area. The affected area is usually black with anatomical
distortion at the point where the signal becomes recognisable.
Its paramagnetic properties cause significant spin dephasing in any spins in its vicinity.
It reduces both T1 and T2 however it has a far greater effect on T1 than it does on T2,
hence all IV contrast sequences are performed as T1 weighted sequences.
Gd-DTPA does not cross the normal blood-brain barrier so can be used to discern
when this has been pathologically breached.
Contrast – endogenous
Arterial spin labelling – this is a way of providing vascular contrast without using an
exogenous substance. It relies on the different T1 re-growth times of blood as
compared to surrounding tissue. Two images are acquired, one where the signal from
blood is nulled by spin inversion, and a second control image. When the two images
are combined, an image of perfusion can be obtained.
Time of flight imaging – This technique relies on the blood in the slice downstream
from the slice of interest is targeted and the spins excited by RF pulses. As the blood in
this upstream slice flows into the target slice downstream, the spins from the blood
show bright against the slice which hasn’t received these initial RF pulses.
True
False
b. Gradient magnetic fields can cause arrhythmias in patients who do not have
pacemakers
True
False
c. The use of multiple RF pulses during sequences creates a lot of acoustic noise
True
False
d. The patient’s weight is needed to calculate the SAR
True
False
e. Burns can occur at the point of contacts between patients and wires
True
False
• The magnetic field. The magnetic field is always on. This is because it is
easier/better for the machine to remain on permanently than to repeatedly have
to supercool the coil to the extent needed in order for it to function correctly.
The immediate concerns regarding the static magnetic field relate to risk from
ferromagnetic objects entering the room becoming projectiles. Equipment used
within the MRI room must be made from MR-safe, non-ferrous materials.
Austenitic stainless steel (a type of stainless steel) is MR-safe, as is aluminium.
Cardiac pacemakers are of significant concern due to the risk, not just of lead
movement, but also of abnormal current induction in the leads leading to
arrhythmias. Some types of pacemaker (or neural stimulators) are MR
conditional (an update concept of MR-safe, with limits on the strength of the
magnetic field they have been tested to), although should still be switched off
temporarily for the duration of the scan.
• The gradient magnetic fields. These are made to apply in various different
directions for varying amounts of time during the scan. Notable side effects
include peripheral nerve stimulation giving the sensation of tingling, but also a
metallic taste in the mouth, vertigo and nausea. Involuntary muscular
contractions and cardiac arrythmias have also been reported. These effects are
all more prevalent when the background magnetic field strength is more
significant (for example in a 3T magnet as opposed to a 1.5T one). Obviously the
shorter the duration of the variations occurring, the lower the risk of side effects
occurring. The risk of the gradient magnetic fields to the developing foetus are
theoretical however it is to guard against the effects of this aspect of the scan
that scans are usually ideally not performed on or by patients/staff in their first
trimester of pregnancy (where declared), though most follow-up studies do not
show noticeable effects for the foetus.
The second aspect of the gradient magnetic fields which causes a problem is the
noise incurred when the gradients are applied. Having an MRI scan is a very
noisy experience. The noise comes from switching on and off the gradient
magnetic fields. Patients are given ear defenders to guard against hearing
damage, and for their own comfort.
• RF pulses. The risk from the RF pulses is of heating. In a whole-body sense, there
are limits applied to scans which curb the SAR (specific absorption ratio) allowed
to be given to a patient during the scan. The ratio limits depend on the weight of
the patient, hence patients are weighed prior to scanning in order to calculate
the limit allowed. Furthermore, there is a risk of contact-point heating. This can
occur if there is metal in contact with the patient (non-ferrous), and for this
reason patients are usually advised to change into a patient gown prior to the
scan. Contact-point heating can also occur in ‘closed-loop’ scenarios. Where a
closed-loop exists, a small current can run within it during the application of an
RF pulse. This includes, for example, if the patient has crossed their arms; the
loop of the arms and body would count as a loop. At the point of skin contact
there is inefficiency in the conduction of this current which manifests as heat
and can lead to burns (even second/third degree burns in rare cases). For this
reason, care is taken to ensure, not just that the patient has not crossed their
arms or legs during the scan, but also that where monitoring leads are
necessary, they do not cross over the patient but rather run straight down the
bore of the scanner. It is also the reason why it is not possible for caregivers (e.g.
a parent) to hold the hand of their child during paediatric scans, or other care-
related scenarios.
• The fringe magnetic field. The strength of the magnetic field diminishes
exponentially away from the bore. In the fringes of the magnetic field there is far
less of a projectile risk however there is still a risk to electrical items such as
watches, phones and items containing chips such as cards, especially contactless
cards. For this reason, public access to the area around the magnet where the
magnetic field strength is still above 0.5mT is restricted even if it is not within the
confines of a designated ‘inner controlled area’. Hence there will be areas
outside the inner controlled area where public access is restricted; because the
field strength is still greater than 0.5mT. The inner controlled area must contain
all parts of the field in excess of 3mT.
Regarding aliasing, the following reduce the likelihood of aliasing occurring?
True
False
True
False
True
False
d. Using gadolinium
True
False
True
False
Field of view
The physics behind changing the field of view is difficult to understand but in general,
smaller field of view:
• Higher resolution
• Higher gradient strengths needed
• The returning signal strength is lower
Rectangular fields of view can be used, for example in spine imaging. The number of
phase encoding steps is halved but the difference between each phase encoding step
is doubled such that the remaining steps cover the same overall range. There are two
problems which arise due to this technique however. Firstly, the signal to noise ratio is
reduced and secondly aliasing can occur.