Tomosynthesis
Tomosynthesis
4:30 pm Adjourn
About the Instructor
PRE‐00023
Breast Tomosynthesis
Hologic Selenia Dimensions DBT
• A 3D screening modality that
preserves the very high
resolution of 2D FFDM • 2D or 3D imaging
– 2D only
• Multiple images of the breast are – 3D only
acquired at different angles
during a sweep of the x-ray tube
• Combo mode
• Allows radiologists to see around – 3D image
overlapping structures – Return to 0 degrees for 2D image
– Single compression for both images
How Does Hologic’s Tomosynthesis Work? Potential Benefits of 3D Imaging
• Tube moves in a 15 arc
• Better imaging
– Improved lesion margin visibility
•15 low dose images are acquired
– Precise lesion localization
• 1 image at each degree
• Four second sweep – Identification and location of multi-focal cancers
• Total dose ≈ one 2D mammogram
• Obtaining a 2D exam along with the 3D exam will allow direct Low rate of Cancer Detection =
comparison of current 2D images with prior 2D images Low Recall Rate
2D plus 3D
2D
AUC diff 0.072, p‐value 0.0001
February 2011
PRE‐00023
LCC LCC
A 2D Mammography Image
2D 2D
Hologic – Proprietary and Confidential Hologic – Proprietary and Confidential
2D 3D
2D 3D
2D
The Difference is Clear
Hologic – Proprietary and Confidential Hologic – Proprietary and Confidential
Recall Reduction
Superimposed Tissue Examples
2D 3D 3D Click image to
Hologic – Proprietary and Confidential
run 3D in cine
Slice 30
26
22
14
18
Recall Reduction – Superimposed Tissue (Case 2)
Click image to
run 3D in cine
Hologic – Proprietary and Confidential Hologic – Proprietary and Confidential
2D 3D
Click image to Click image to
Hologic – Proprietary and Confidential
run 3D in cine Hologic – Proprietary and Confidential
run 3D in cine
Tubulolobular Adenocarcinoma Region of Interest
Click image to 2D 3D
run 3D in cine
Click image to
run 3D in cine
Hologic – Proprietary and Confidential Hologic – Proprietary and Confidential
Screening vs DBT in a
57 y/o woman Micropapillary type ductal
carcinoma in situ in 65 y/o woman
39
Movie 3
Park J M et al. Radiographics 2007;27:S231-S240 Movie 1 Park J M et al. Radiographics 2007;27:S231-S240
Metastasis from endometriod
carcinoma in 59 y/o woman
DBT
Artifacts
3 primary masses Mass not seen
(see arrows) on FFDM
On basis of US
appearance Mass
Dx as Cyst
44
Park J M et al. Radiographics 2007;27:S231-S240 Movie 7 Wu, T. et al. Med Phys 33 (7), Jul 2006
Small Calcification Artifact Needle Artifact
45 46
Wu, T. et al. Med Phys 33 (7), Jul 2006 Wu, T. et al. Med Phys 33 (7), Jul 2006
Questions?
Talk 2
Technical Approach to The technology behind
Tomosynthesis tomosynthesis
• DBT
Digital Breast Tomosynthesis Imaging
– Complex imaging technique
• Uses computer algorithms and digital images to
create a 3D image
14 March 2015 – Indianapolis, IN • Underlying technologies
– Digital detectors
Jerry A. Thomas, MS, FAAPM, DABR, CHP, DABSNM – X-ray unit
Diagnostic Medical Physicist
Via Christi Hospitals Wichita, Wichita, KS
– Reconstruction algorithms
– Image Display 2
• Solid state
3 4
X-ray tube travel Ends of Tube Travel
+ 7.5 degrees - 7.5 degrees
Motion
Continuous
Step-and-shoot
Angle of Travel
15°-- ± 7.5°
40° -- ± 20°
# of Projections
15 to 25
5
Park J M et al. Radiographics 2007;27:S231-S240 Park J M et al. Radiographics 2007;27:S231-S240
7 8
Park J M et al. Radiographics 2007;27:S231-S240 Park J M et al. Radiographics 2007;27:S231-S240
Basic Principle of Basic Principle of
Slice Recon Projection 1 Slice Recon Shift Right
Projection 2 No Shift
Basic Principle of
Slice Recon Shift Left
Shift Right
13 14
22
Blurring - Ripple
Ghost Artifact - Distortion
• Blurring occurs along the sweep direction
– Results from imaging structures that exist out-side • Caused by high-contrast structures outside
the section plane the section plane whose long axis is parallel
– Most prominent for high-contrast objects that are to the sweep direction
perpendicular to the sweep direction – Called “parasitic streak” in conventional linear
tomography
• Ripple
– Due to limited number of projections
– Caused by a high-contrast source far outside imaging
plane
Poor Spatial Resolution DBT Reconstruction Algorithms
• In-section resolution • Filtered back projection (FBP)
– Depends on the resolution of the detector • Maximum likelihood (ML)
– Independent of acquisition parameters • Simultaneous algebraic reconstruction
technique (SART)
• Depth resolution • Gaussian frequency blending (GFB)
– Affected by sweep angle
• Voting strategy
• Section thickness decreases and depth resolution
increases as sweep angle increases.
26
27 28
Simultaneous algebraic
Gaussian frequency blending (GFB)
reconstruction technique (SART)
• A refined version of the iterative method of • Combines matrix inversion tomosynthesis
algebraic reconstruction. with FBP
• More effective than FBP at making tissue • Improved reconstruction of micro
conspicuous calcifications
• Better Calcification enhancement
• Better mass margin enhancement
• FBP superior in noise reduction and CNR
29 30
31
MIP of slices showing MIP of slices showing speculated
Calcifications lesions
45
46
Questions?
47
Break Time
10 minute break
Talk 3
Digital Detector Technology Acknowledgement
• Original talk content developed by
Digital Breast Tomosynthesis Imaging
Dr. J. Ed Barns, Executive Director MTMI
FCM
Fuji Mammo CR
• Pixel size 50 micron
• Use dual sided reading to DQE
• Imaging Plate has thicker phosphor layer
• Results in 40-50% increase in NEQ (DQE)
• Post processing enhances image
(e.g. filtering to recognize and enhance calcifications)
FUJI FCR 5000: MTF FUJI FCR 5000: DQE
From Centre for Evidence-based Purchasing Report 04094, October 2004 From Centre for Evidence-based Purchasing Report 04094, October 2004
Combination
Processing
enhancement of
selective edge-enhancement microcalcifications
Calcification
Position
Detection Calcification
Original Processing
Image Edge
Information
Detection of Center
of Calcification Image
Konica CR Mammography
CR Mammography Systems
• Use CsBr plate
• Columnar crystal structure which collimates the for AGFA, FUJI, KODAK, KONICA
light
• Scan with 43.75 u laser “ All measurements were equivalent to
or exceeded the performance of a typical
• Scan slower than regular plate
modern film/screen system”
• Use smoothing algorithm to reduce noise
• Motor accuracy is limiting parameter for resolution
Photo D
Mammo Panel with Bonded Scan & Data Modules
TFT Data Data
Modules Modules
Scintillator
Drivers
Photo D/TFT
9M$ - 30units
33%
100%
Data line
Gate Line
Photodiode
IP CP
Cstorage
HV
Biasing
Electrode
Scan
Electric field
Line blocking
layer
selenium
Data blocking
Line layer
Selenium
Structure
Pixel thickness
TFT Collection
Active Matrix Electrode
Array
Assuming perfect material, no trapping of charge in selenium facilitates rapid imaging
Principle of Blocking Layers High Efficiency of a-Se
Metal
electrode
n-layer •Selenium provides almost 100% charge collection efficiency due to
electrostatic field effects in the photoconductor
•Almost complete absorption of the x-rays is achieved within ~200µm layers
selenium
•Selenium allows high fill factor for small pixel size required in
mammography
p-layer
TFT
Substrate
N-layer: one way charge valve. Conducts electrons but traps holes
P-layer: one way charge valve. Conducts holes but traps electrons
Pixel electrodes
Data line
Hologic DQE Curves
Gate Line
a-Selenium Plate
Ise Cse
Cstorage
Z. Jing, Hologic
a-Selenium Detector
PHOTON COUNTING
Mammography stand
• Five programmable stops for individual projection
settings Ergonomic design – Adjustment buttons in
several places
• Free positioning
• Field of View - 24 x 26 cm
• Time to display image < 20 seconds
• Scan time 4-15 seconds
Detector Technology - Photon Counting Sectra Silicon Detector Geometry
• Dead volume layer on top
• 20 parallel silicon microstrips X-rays
• Detector several cm long
• 50 micron pixels
• Pre-collimator defines pixel in scan • Detection rate 5 x 106 x-rays
direction /sec/pixel
• Sample data every 50 microns of scan Dead area
motion • Slit-scanning, 5 sec. scan time
• Tungsten target tube operated at about • 3 mm deep – 95% absorption
30kVp
to 20 kev
• 5 second scan time – varies with breast Active Detector 8 degrees
size Volume • Counts individual photons
-digitized upon detection
Dead area
-suppresses noise
Energy Result:
X-ray photon
- Number of X-rays
From Centre for Evidence-based Purchasing Report 06045 August 2006 From Centre for Evidence-based Purchasing Report 06045 August 2006
Sectra DQE Dose Dependence
Other Detectors – not discussed
• Xcounter
– Gas filled Krypton Detector Not FDA Approved
– Have developed a 3D prototype
• Fuji Aspire HD
– Dual aSe detector
– Readout – optical switching
• DASALA
– CMOS detector coupled to CsI
– 39 micro pixel Not FDA Approved
– Partnering with Giotto
From Centre for Evidence-based Purchasing Report 06045 August 2006 – 30 frames/second readout
Acknowledgements
Summary Thanks for input from:
• GE Healthcare
• Properties of a detector useful for DBT • Fujifilm Medical Systems
– Fast readout • Hologic/LORAD
• Planmed
– Low noise
• Sectra Imtec AB
– Low lag • Siemens Medical Solutions
– High DQE • XCounter
• Tao Wu, Ph.D.
– Niquist frequency: 5 lp/mm or greater • Ed Barns, PhD
• Mats Danielsson, Ph.D.
Questions?
Talk 4
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
INTERFACE
1 2
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
Agenda
System Description
• Go over the features of the Selenia
Dimensions DBT system
• Walk through the Selenia Dimensions user
interface
3 4
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
7 8
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SSYYSSTTEEMM DDEESSCCRRI PI PTTI O
I ONN AANNDD UUSSEERR I N
I NTTEERRFFAACCEE SYSTEM DESCRIPTION AND USER INTERFACE
X-Ray Generation
System Components
• Conv: 50 μm Rh; 50 μm Ag
X-Ray Filtration • Tomo: 0.7 mm Al
9 10
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
Digital Detector
Detector Field-Of-View
• Electronic cropping to
– 18 cm x 24 cm or
– 24 cm x 29 cm
• Smaller FOV dynamically moves to left,
TFT based amorphous Retracting HTC Grid center and right side of the full detector
selenium detector (aSe) • Automatic grid control • Detector cropping depends on position of
• Similar technology to • Grid retracts for
Selenia detector (70 μm • tomo the attached shifting paddle
pixel pitch) • Conventional mags
• Optimized for fast image • Grid in place for
acquisition (4 fps) conventional contact
• Active temperature control imaging
11 12
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
13 14
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
15 16
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
17 18
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
19 20
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
21 22
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
23 24
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
physicist
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SSYYSSTTEEMM DDEESSCCRRI PI PTTI O
I ONN AANNDD UUSSEERR I N
I NTTEERRFFAACCEE
User Interface
Quality Control
34
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SSYYSSSYSTEM
TTEEMM DDEEDESCRIPTION
SSCCRRI PI PTTI O
I ONN AAND
ANNDD UUSSEERRINTERFACE
USER IN
I NTTEERRFFAACCEE SYSTEM DESCRIPTION AND USER INTERFACE
User Interface
Image Acquisition
47
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE
49 50
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SSYYSSTTEEMM DDEESSCCRRI PI PTTI O
I ONN AANNDD UUSSEERR I N
I NTTEERRFFAACCEE
User Interface
Other Tools
58
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SSYYSSTTEEMM DDEESSCCRRI PI PTTI O
I ONN AANNDD UUSSEERR I N
I NTTEERRFFAACCEE
User Interface
Test Patterns and QC Review
70
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
SYSTEM DESCRIPTION AND USER INTERFACE SYSTEM DESCRIPTION AND USER INTERFACE
Thank you!
78
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
Lunch TIME
79
© 2011 Hologic, Inc. All rights reserved. PRE00027 rev 03/2011
QUALITY CONTROL PROCEDURES
Quality Control
Quality Control Procedures Procedures
14 March 2015 – Indianapolis, IN
Selenia® Dimensions® Digital Breast Tomosynthesis Systems
Conventional Only
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
2D Tomosynthesis
Detector a-Se, 24x29 cm a-Se, 24x29 cm a-Se, 24x29 cm
Pixel 70 μm 70 μm 70 μm 140 μm
Filter 30 μm 58 μm 50 μm 700 μm
thickness
mA (Max) 100 100 200
SID (cm) 66 66 70
Grid Contact mode: HTC Contact mode: HTC Contact mode: HTC Contact mode: no
Mag mode: no grid Mag mode: no grid Mag: no grid grid
Alternative Standard 9
Technologist Tests –
Tomosynthesis Option
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
2. Collimation Assessment
Equipment
• X-ray recording media (e.g. screen-film cassettes, CR
plate, self-developing film)
• Screening paddles
2. Collimation Assessment
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
• 2D
o 24 × 29 cm screening paddle
o 18 × 24 cm screening paddle, LEFT position
o 18 × 24 cm screening paddle, CENTER position
o 18 × 24 cm screening paddle, RIGHT position
• Tomosynthesis
o 18 × 24 cm screening paddle, CENTER position, using
“Zero-Degree Tomo” 18 × 24 cm, Left 18 × 24 cm, Center 18 × 24 cm, Right
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
Note:
• Put 4.0 cm BR-12 or PMMA under
the paddle
• Apply 10-12 lb compression force
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
2. Collimation Assessment
(1) X-Ray Field to Light Field
• 2D
o 24 × 29 cm screening paddle
(2) X-Ray Field to Image Receptor
• 2D
o 24 × 29 cm screening paddle
o 18 × 24 cm screening paddle, LEFT position
3. Artifact Evaluation
o 18 × 24 cm screening paddle, CENTER position
o 18 × 24 cm screening paddle, RIGHT position
• Tomosynthesis
o 18 × 24 cm screening paddle, CENTER position, “Zero-Degree Tomo”
(3) Compression Paddle to Image Receptor
• 2D
o 24 × 29 cm screening paddle
o 18 × 24 cm screening paddle, CENTER position
o
© 2011 Hologic, Inc. All rights reserved.
Small Breast screening paddle (when available) PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
3. Artifact Evaluation
“Actual Pixel” button to
get into 1:1 display mode
(1 detector pixel to 1
monitor pixel)
Image acquisition
Image review
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
• If non-invasive kVp meter is used to measure kVp, it must • At 28 kVp, the coefficient of variation
be calibrated to the target-filter combination and the kVp of the kVp must be ≤ 0.02
range used.
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
5. Beam Quality Assessment — Half-Value • Protect the detector by covering the detector with a layer of
Layer Measurement ≥0.5 mm lead or lead equivalent
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
Image Acquisition
• One 2D acquisition
• One Tomosynthesis acquisition
“Magnification” button
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
Performance Criteria:
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
Magnification stand
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
SFS
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
© 2011 Hologic, Inc. All rights reserved. Example: Data sheet for 2D AEC test
PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
LFS
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
10. Phantom Image Quality Evaluation 10. Phantom Image Quality Evaluation – 2D
• Image Acquisition • Click the “Actual Pixel” button to
o Use ACR phantom bring the image to full resolution
o Use “ACR Phantom View” provided by the system • Score the phantom following
o Acquire 2D and tomosynthesis phantom images using AEC 1999 ACR Mammography Quality
Control Manual
• 2D Phantom Score:
- 5 fibers, 4 specs, 4 masses
- Due to phantom variation, a
score of (4.5, 4.0, 3.5) is
acceptable if SNR and high
contrast resolution tests pass
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
• Performance Criteria
SNR ≥ 40
CNR within ±15% of the reference value set by physicist
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
13. Detector Ghosting (Troubleshooting Use Only) 13. Detector Ghosting (Troubleshooting Use Only)
• Create Ghosting Effect
1. Create Ghosting Effect
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
13. Detector Ghosting (Troubleshooting Use Only) 13. Detector Ghosting (Troubleshooting Use Only)
• Create Ghosting Effect • Measure Ghosting Factor
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES QUALITY CONTROL PROCEDURES
13. Detector Ghosting (Troubleshooting Use Only) 13. Detector Ghosting (Troubleshooting Use Only)
• Measure Ghosting Factor • Measure Ghosting Factor
3 2 1
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
ROI1 in background
without ghosting Ghosting Factor = Thank you!
meanregion2 - meanregion3
ROI2 in Al filter
without ghosting meanregion1 - meanregion2
ROI3 in Al filter
with ghosting
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011 © 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
QUALITY CONTROL PROCEDURES
10 minute break
Cokes & Soft Drinks
are on back table
Talk 6
© 2011 Hologic, Inc. All rights reserved. PRE00028 rev 03/2011
Quality Control of
Softcopy Display
Tomosynthesis Displays
• Monitor design/types
Digital Breast Tomosynthesis Imaging
• Flat Panels
• Monitor performance characteristics
• DICOM Part 14 Grayscale Standard Display
14 March 2015 – Indianapolis, IN Function
• Flat Panel Calibration
Jerry A. Thomas, MS, FAAPM, DABR, CHP, DABSNM
Diagnostic Medical Physicist
Via Christi Hospitals Wichita, Wichita, KS
Flat Panels
1280 x 1024 2048 x 1536 (3 MP) 1200 x 1600 (2 MP) 1200 x 1600 2048 x 2560 (5 MP)
1024 x 1280 (1 MP)
55 fL / 55 fL / 35 fL / 75 fL / 235 fL / 175 fL /
200 cd/m2 200 cd/m2 120 cd/m2 255 cd/m2 800 cd/m2 600 cd/m2
3
Each Sub-Pixel is Independent
Liquid Crystal (LC) Light Valve and controlled by a TFT
Not a Liquid … and not a Crystal
AM-LCD Structure
Graphic boards
Backlight
8
Monitor Performance Characteristics Contrast Ratio
• Luminance Uniformity • Anti-reflective coating • Depends on
– Fixed at time of – Maximum luminance
– Will add color to monitor
manufacture
– Should be < 18% (center to • Glass color – Glass Transmission
edges) – Ambient Light
– Best monitors – 8% – 12% • Monitor gamma
– White areas of the
• Contrast Ratio ~ 2.2 image
– Lmax/Lmin • Focal Spot Size (CRT)
• Phosphor Type (CRT)
– P45 • Input impedance
– P104 – 50 or 75 ohms
Phosphors - CRT
Glass Envelope Construction - CRT
• P45
– One component phosphor
– Low aging
• 27% total transmission
– Short decay time after 2 ms • Ambient Light
– No color variation over
time
attenuation (93%)
• Anti-reflective/anti-
• P104 static coating
– Dual component phosphor
– High efficiency • Same refractive index
– Can be run at luminance throughout
levels > 100 fL
11 12
11 12
Resolution vs Luminance - CRT Monitor Lifetime – CRT and FP
• Can last up to
50,000 hr
• Phosphor and
glass decrease
• Cut-off control
• Control of
Gamma function
13 14
14
Hands-off
Conf Test
Sampling
Points
Regulating the
Gamma
Function
Stabilizing the
Black Level
15 16
Grayscale Standard Display GSDF
Function
• A methodology for measuring the • Neither
characteristic curve of a display system – A performance standard
– For conformance testing – Image display standard
– For altering display system to match GSDF
• Does not define
• Key elements of GSDF – Luminance
– Target: 10% of the monitor area – Luminance range
– Optical density range
– Surround: 20% of Lmax
– Presentation of picture element values
• Icons on screen
• Veiling glare
Monitor Gamma vs Perceptual Monitor Perceptual Linearization
Linearization and
JND Determination
• Background is set
• Line Pair target in center 10% of viewing area
• Increase target until it is just visible
• Increase background to target level
– Repeat previous two steps
• Example follows:
Grayscale Standard Display Function Grayscale Standard Display Function
DICOM Part 14 DICOM Part 14
Perceptually Linearized Response Curve
white 1000
100
Different change
in absolute luminance 10
Luminance
1
0 200 400 600 800 1000
GSDF defines
.1
transformation of
Digital Driving
.01
Levels (DDL) JND Index
to Luminance.
34
35
The First Piece:
Monitor Calibration
• Setting the display black and white levels
Application of an LUT
Display Natural Gamma Gamma Correction LUT Conformance Validation
Conformance Validation
Automated Conformance Testing
• Manual Conformance Testing
• Barco
• DICOM Part 14 Grayscale Standard Display Function – Medi-Cal
• Monitor Luminance Uniformity – Medi-Cal QA Web
Raw Image
• Produces:
•An image with diagnostic content (pizzazz)
Sigmoid
Acquisition image Presentation
Dicom GSDF
LUT
+
• But if the GSDF transformation
resulted in making big skips near
Processing Processing
+
black...
Perceptually Linearized
Display
Raw Image
Monitor Calibration
Image Presentation
65 66
Video LUT with 256 Calibration values 256 Cal values - Showing JND Shifts
67 68
Optigrayscale – TRUE 256 JNDs Optigrayscale 256 JNDs Annotated
69 70
72
71
Conformance Testing
QC/QA of Digital Mammography and
Brightness Uniformity QC
Workstation
Image quality
Luminance uniformity
Ambient light
Control of Specular reflections
DICOM Part 14 – Gray Scale
Performance Standard
Measure at System Level and Application Level
QC using SMPTE Pattern
73 74
Barco QC Program
• QC Tests for the Radiologic Technologist
– Display Screen Cleanliness – Daily
– Viewing Conditions Check for the Display
• Daily when clinical image review is planned
– Image Quality Check – Weekly
• Equipment positioning
– View boxes and Softcopy displays should be parallel to
one another … not at an angle or back to back
80
Coronis 5MP QC Testing Nio 5MP QC Testing
81 82
83 84
Pattern Evaluation Frequency
Do you remember this?
Physician
- Daily prior to each reading session
Technologist
- Daily on acquisition workstation
- Weekly on Dx monitors in work area
Medical Physicist
- Monthly – Department QC review
- Annually – detailed eval/comparison with prior years data
85 86
First Children’s
program to run 5
days a week
87 88
SMPTE Test Pattern Modified SMPTE Test Pattern
Evaluation
0 / 5% Black
Center Targets
95/100% White
For measuring
Density Step Ramp Lmax and Lmin
(0 to 100%)
Veiling Glare will
Increase black level
High Contrast Bars
measurement
Center
Corners Pattern Should NOT
be used for this
Low Contrast Bars measurement
Center
Corners
Letters
90
91 92
92
AAPM Briggs Test Pattern Circle
• Measures:
– Geometrical distortion
– Linearity
– Warping
93 94
94
END
Printer Overview
• High resolution Laser Printers
– Kodak 8600 – No longer manufactured Still used
WIDELY
Hard Copy Device QC –
–
–
Kodak 8610 – No longer manufactured
Kodak 8900
Kodak 6800
– Agfa LR5200 – No longer manufactured or supported
– AGFA DS4500M
– Fuji DryPix 7000
– Fuji DryPix 5000
– Fuji DryPix 4000
– Fuji FM-DP-L
– Konica-Minolta DryPro 793
– Quality Control
• Acceptance Testing
• NEMA “standard” XR 23-2006
Printers Validated by FFDM Common Mammo Laser
Manufacturers Characteristics
VALIDATED Printers
• AGFA 4500M
In service No longer
Manufactured
• Specifications
• • AGFA 5500 – Pixel matrix: 5025 x 6200
• • AGFA 5503 • KODAK 8600
• • AGFA AXYS
• KODAK 8610 – Pixel size: 38.75 to 50 um – newer system down to 25 um
• FUJI FM-DP L
• • KODAK 8900 – 12 bit (4096 gray levels) – newer systems 14 bit
• • KODAK 6800 No Longer Supported by
• • CARESTREAM DryView 5850 Printer Manufacturer
– Resolution: 508 to 655 dpi – newer system up to 1016 dpi
• • KONICA 793 • AGFA LR5200 – Image area: 19.5 x 24 cm up to 35 x 43 cm
• • KONICA 873
• • FUJI DRYPIX 4000 (see note 1)
– Throughput: 40 to 200 films/hr
• • FUJI DRYPIX 5000 (see note 1) – Dmax: Film dependent – normally < 3.6 – now ~ 4.0
• • FUJI DRYPIX 7000 (see note 1)
• • SONY UP-DF750
– Dmin: Film dependent – normally ~ .25
97
98
101 102
114
116
SMPTE in LP QC SMPTE in LP QC
• Medium Density – (5th step
– Medium Density – (5th step down from max white)
down from max white) – Density Difference
– Density Difference • (5th step minus 2nd step)
• (5th step minus 2nd step)
60% - 90% – Lower Density – (1st step
above max black) (10% step)
Talk 7
126
Accreditation of Digital Accreditation of
Breast Tomosynthesis FFDM SYSTEMS
Digital Breast Tomosynthesis Imaging
•FDA Requirements – 2D FFDM
3 4
FDA Regulations FDA Regulations
• Provide a satisfactory FFDM equipment • Follow the manufacturer's guidelines for
evaluation (including an evaluation of the quality assurance and quality control
Soft Copy Display system) performed by a tests as described by the manufacturer's
qualified medical physicist within 6 manual.
months prior to the date of your application.
5 6
7 8
FDA Facility / FFDM Statistics FDA Facility / FFDM Statistics
• As of 01 August 2007 • As of 01 August 2008
– Certified facilities, as of October 1, 2006: 8,832 – Certified facilities, as of October 1, 2007: 8,837
– Total certified facilities: 8,825 – Total certified facilities: 8,835
– Total accredited units 13,402 – Total accredited units 13,465
– Certified facilities with FFDM units: 2,090 (24%) – Certified facilities with FFDM units: 3,515 (40%)
– Accredited FFDM units: 3,077 (23%) – Accredited FFDM units: 5,356 (40%)
• Total annual mammography procedures reported, as of August 2007: 34,808,145 • Total annual mammography procedures reported, as of August 2008: 36,154,201
9 10
• Total annual mammography procedures reported, as of August 2009: 36,983,579 • Total annual mammography procedures reported, as of August 2010: 38,823,849
11 12
FDA Facility / FFDM Statistics FDA Facility / FFDM Statistics
• As of 01 August 2011 • As of 01 August 2012
– Certified facilities, as of October 1, 2010: 8,652 – Certified facilities, as of October 1, 2011: 8,620
– Total certified facilities: 8,620 – Total certified facilities: 8,626
– Total accredited units 12,270 – Total accredited units 12,414
– Certified facilities with FFDM units: 6,766 (78%) – Certified facilities with FFDM units: 7,388 (86%)
– Accredited FFDM units: 9,760 (79%) – Accredited FFDM units: 10,786 (87%)
• Total annual mammography procedures reported, as of August 2011: 38,844,222 • Total annual mammography procedures reported, as of August 2012: 39,351,056
13 14
• Total annual mammography procedures reported, as of August 2013: 38,233,396 • Total annual mammography procedures reported, as of 01 Feb 13: 38,747,608
15 16
FDA Approved DBT & FFDM Systems
FDA Facility / FFDM Statistics 19 Systems
approved by
13 Systems
FDA since
approved in
• Changes from 1 Aug 2006 to 1 Jun 2014 2000
2011
How longiswill
FFDM the it take for DBTOF
STANDARD to CARE
becomefor
the 5 Systems No
17
STANDARD of Imaging
Mammographic CARE? longer
manufactured
23 24
FDA Certification Extension FDA Certification Extension
• For FFDM systems for which there is not an approved
Accreditation Body FDA will provide a Certification
• Processing time for Certification Extension: Extension of Accredited Sites
– 14 days
• System is exempt from Accreditation requirement
• FDA Contact: Go to FDA MQSAWeb Site
• Facility must submit a request to FDA – in Writing – to
obtain Certification Extension
25 26
29 30
3. Hologic Selenia Dimensions DBT Digital 4. Identification of Printer for Hard Copy
Image Receptor Identification Interpretation
a. Receptor Manufacturer (mandatory even for facilities performing only soft copy
b. Receptor Model interpretation)
c. Year of Manufacture
d. Serial Number (if applicable)
a. Printer Manufacturer
b. Printer Model
c. Year of Manufacture
d. Serial Number
31 32
5. Final Interpretation Review Monitor
6. Phantom Identification
Identification
(if soft copy display is available) a. Phantom Manufacturer
b. Phantom Model
a. Monitor Manufacturer
b. Monitor Model
c. Year of Manufacture
d. Serial Number
33 34
7. Personnel Qualifications – Pre 11 Feb 2011 7. Personnel Qualifications – Pre 11 Feb 2011
• Interpreting Physicians • Radiologic Technologists
– PERSONNEL QUALIFICATIONS: INTERPRETING PHYSICIANS WHO – PERSONNEL QUALIFICATIONS: RADIOLOGIC TECHNOLOGISTS
ARE QUALIFIED TO INTERPRET DBT MAMMOGRAMS WHO ARE QUALIFIED TO PERFORM DBT MAMMOGRAMS
– (1) meet all the requirements of CFR 900.12(a)(1) "Mammography Quality – (1) meet all the requirements of 21 CFR 900.12(a)(2) "Mammography Quality
Standards; Final Rule" that became effective on April 28, 1999 AND Standards; Final Rule" that became effective on April 28, 1999 ; AND
– (2) began interpreting DBT mammograms prior to February 11, 2011 – (2) began performing DBT mammography examinations prior to February 11,
2011.
35 36
7. Personnel Qualifications – Pre 11 Feb 2011 7. Personnel Qualifications – Post 11 Feb 2011
• Medical Physicists • Interpreting Physicians
– PERSONNEL QUALIFICATIONS: MEDICAL PHYSICISTS WHO ARE – PERSONNEL QUALIFICATIONS: INTERPRETING PHYSICIANS WHO
QUALIFIED TO PERFORM DBT SURVEYS ARE QUALIFIED TO INTERPRET DBT MAMMOGRAMS
– (1) meet all the requirements of 21 CFR 900.12(a)(3) "Mammography Quality – (1) meet all the requirements of 21 CFR 900.12(a)(1) "Mammography Quality
Standards; Final Rule" that became effective on April 28, 1999 ; AND Standards; Final Rule" that became effective on April 28, 1999 *;
– (2) began performing equipment evaluations and/or surveys of DBT – (2) began interpreting DBT mammograms after February 11, 2011; AND
mammography units prior to February 11, 2011.
– (3) have 8 hours of initial training in DBT Mammography*.
• Note: Full Field Digital Mammography training can not be used as a substitute for DBT
training.
37 38
7. Personnel Qualifications – Post 11 Feb 2011 7. Personnel Qualifications – Post 11 Feb 2011
• Radiologic Technologists • Medical Physicists
– PERSONNEL QUALIFICATIONS: RADIOLOGIC TECHNOLOGISTS – PERSONNEL QUALIFICATIONS: MEDICAL PHYSICISTS WHO ARE
WHO ARE QUALIFIED TO PERFORM DBT MAMMOGRAMS QUALIFIED TO PERFORM DBT SURVEYS
– (1) meet all the requirements of 21 CFR 900.12(a)(2) "Mammography Quality – (1) meet all the requirements of 21 CFR 900.12(a)(3) "Mammography Quality
Standards; Final Rule" that became effective on April 28, 1999 ; Standards; Final Rule" that became effective on April 28, 1999 ;
– (2) began performing DBT mammography examinations after February 11, – (2) began performing equipment evaluations and/or surveys of DBT
2011; AND mammography units after February 11, 2011; AND
– (3) have 8 hours of initial training in DBT Mammography. – (3) have 8 hours of initial training in DBT Mammography.
• Note: Full Field Digital Mammography training can not be used as a substitute for DBT • Note: Full Field Digital Mammography training can not be used as a substitute for DBT
training. training.
39 40
8. Report of Mammography Equipment 8. Report of Mammography Equipment
Evaluation Evaluation
Report of Mammography Equipment Evaluation (MEE) (must have been conducted in b. The results of quality control tests as required under the following sections of the MQSA
accordance with 900.12(e)(10) within the 6 months prior to the request for use final regulations 21CFR 900.12(e):
approval)
(4)(iii) Compression Device Performance (5)(vii) X-Ray Field/Light field/Image
a. Statement that equipment performance, as required under the following sections of the receptor/Compression paddle alignment
(5)(i) Automatic Exposure Control
MQSA final regulation 21 CFR 900.12(b), is met:
Performance (5)(ix) System Artifacts
(1) Prohibited Equipment (6) Magnification (if applicable to the DBT system used) (5)(x) Radiation Output
(2) Specifically Designed for (7) Focal Spot Selection (5)(ii) Kilovoltage Peak Accuracy and (5)(xi) Decompression
Mammography Reproducibility (or alternative standards allowed for these
(8) Compression
(5)(iii) Focal Spot Condition (Resolution) requirements)
(3) Motion of Tube-Image Receptor (9) Technique Factor Selection and Display
Assembly (5)(iv) Beam Quality and Half-Value Layer (6) Quality Control Tests – Other
(10) Automatic Exposure Control Modalities
(4)(iii) Removable Grid (if applicable to the (5)(v) Breast Entrance Air Kerma and AEC
(14) Lighting (Facilities must perform all DBT
DBT system used) Reproducibility (if applicable to the
(if hard copy display is used for image DBT system used) manufacturer recommended quality control
(5) Beam Limitation and Light Fields evaluation) tests including the medical physicist’s tests
(5)(vi) Dosimetry for Soft Copy Display system)
(15) Film Masking Devices
(if hard copy display is used for image
evaluation) 41 42
43 44
10. Signature
Lead Interpreting Physician Attestation to Staff Personnel Qualifications
45
Advanced Applications in
Advanced
Digital Imaging Applications in
Digital Breast Tomosynthesis Imaging Digital Imaging
Fast Imaging •
•
Tele-Radiology
Dual-Energy Subtraction
• Digital Subtraction Angiography
Low Noise •
•
Low-Dose Fluoroscopic Positioning
Stereo / Computer Aided Localization
• Multi-modality Imaging
• Temporal Subtraction
• Image Pasting
• ...
Computer-Aided Detection CAD Goals
Applications
• Automated detection
• Automated classification • Improve diagnosis
Current ImageCheckerTM • Standardize interpretation
Performance
• Flagged 85% of • Facilitate interpretation
overlooked cancers • Reduce reading load
• Flagged 83% of
diagnosed cancers • Improve positive predictive value
Increase the sensitivity for • Facilitate Image “triage” in Telemammography
cancer detection when
combined with the radiologist • Educational tool
M Calcification
a
s
Cluster Detection
s
D
e
t
e
c
t
i
o
n
Calcification
Artificial Neural Network
Original image Cluster
Classification
Classification Mammogram Enhancement
Malignant
Enhanced
18
19
DSA Leg DSA Leg
No Contrast Subtracted
Image
• GE Senobright
• FDA approved for sale in the US
Neuro Angiography System
• Change filter and algorithms only
• No hardware change required
• Haven’t found uptake kinetics clinically helpful
• Dose 20% higher-1.5 mGy/view
30 60 180 300
Physical examination
0
Iodine injection
Non palpable lesion
Images courtesy of Dr Diekmann Technique :
Charité – Berlin, Germany Mo/Cu, 45kV,
100mAs Mammography
- 1 stellate opacity
- 1 round opacity
Conventional Mammograms
Mammography
0 120 120 + Time (s)
Small mass
Iodine injection only visible on CC view
• Low confidence of
presence
Low Energy Spectrum Rh/Rh 28 kVp High Energy Spectrum Rh/Cu 44 kVp
• Global
classification
BIRADS 3
Fluence(10 6ph./cm 2)
10 14
12
8
Rh 10
6 8 Iodine
4 Cu 6 K-edge
4
2
2
0 0
0 10 20 30 40 10 15 20 25 30 35 40 45
Energy (keV) Energy (keV)
2nd
Non-overlapping spectra better tissue cancellation threshold
Concentrated around K-edge improved I-contrast
30
29
34