PURE SPECT User Manual
PURE SPECT User Manual
User Manual
www.PureImagingPhantoms.com
Durham, N 405
2
3
Contents
Introduction ............................................................................................................................5
Use of the Benchmark, Standard and Elliptical SPECT Phantoms .......................................6
Use of cylinder without inserts ..............................................................................................6
Placement of Solid Spheres ...................................................................................................8
How much activity do I use? (also see Appendix I) ..............................................................8
How is the phantom mounted in the scanner? .......................................................................9
Slice thickness: thin vs. thick .................................................................................................9
Draining the phantom ............................................................................................................10
What if the phantom leaks?....................................................................................................10
Proper care .............................................................................................................................11
Appendix I: Filling and securing Phantom ............................................................................13
Appendix II: Example protocol and phantom specifications .................................................14
Sensitivity measurements.......................................................................................................15
Line spread measurements .....................................................................................................16
Image statistical (count density) noise ...................................................................................19
Image contrast - example using sphere ..................................................................................21
Appendix III: Sample images ................................................................................................23
SPECT phantom ..................................................................................................................24
Deluxe cold rods ..................................................................................................................25
Standard cold rods ...............................................................................................................26
Benchmark cold rods ...........................................................................................................27
Deluxe cold rods .................................................................................................................28
Bibliography ..........................................................................................................................29
General ................................................................................................................................29
Physics .................................................................................................................................29
Image artifacts and compensation techniques .....................................................................30
Warranty ................................................................................................................................31
4
SPECT Phantom User Manual
Introduction
It is possible from a single scan of the SPECT Phantom to evaluate many system
performance characteristics, including:
System single-slice volume sensitivity
System total volume sensitivity
Effect of regional variations in intrinsic system response using uniform cylindrical
portion
Accuracy of attenuation compensation algorithms
Variation in spatial resolution within the field-of-view using the multi-sized rod insert
Lesion detectability using the multi-sized spheres
Effect of finite spatial resolution and Compton scattering on image quality
Image contrast, %-rms noise and signal-to-noise (S/N) measurements
Line spread measurements in air and in water, without having to remove the cover plate
It is our philosophy that the reported variability of SPECT results reinforces the need
for standardized, routine quality control in ECT, even to a greater degree than in
5
SPECT Phantom User Manual
6
SPECT Phantom User Manual
be introduced into the cylinder through one of the openings (“filler-holes”) in the cover plate.
After screwing both caps (plugs) into the cover plate, agitate the cylinder to completely mix
the radionuclide and the water solution. Remove both caps, to permit air to escape, and
finish filling the cylinder with water through one of the openings. A 30 or 60 cc disposable
syringe, with a “catheter” tip, is useful for this purpose. Repeatedly withdrawing into and
expelling from the syringe will afford sufficient mixing of the extra water and radionuclide.
Be careful of contamination during final filling of the cylinder. Replace both caps and set the
phantom aside for a few minutes to allow for more uniform distribution of the radionuclide
with the extra water that was added. Be sure to observe the usual precautions when handling
radioactive substances within the laboratory: wear gloves and protective clothing, use
shielding whenever possible, and avoid unnecessary exposure to yourself and others by using
time, shielding and distance.
The above procedure describes one technique of preparing the cylinder for scanning.
An alternative approach is to pre-mix the solution in another separate container which has a
larger volume than the cylinder. You may then fill the SPECT phantom from the second
container. This method may be faster for some individuals and allows immediate use of the
phantom, since there is no waiting required for the radionuclide to completely mix. A large
(30 or 60 cc) catheter tipped syringe can be used to complete the filling once the cylinder is
nearly full and you have secured the cover plate in place.
A third method is to use a magnetic stirrer to mix the radionuclide once it has been
introduced into the water. This method can be executed by filling the cylinder with “cold”
(non-radioactive) water and sealing, except for one of the caps, leaving just enough air to be
replaced by the radionuclide. Cap the phantom and place on the magnetic stirrer. A small
bar, placed in the bottom of the phantom before filling with water, will then quickly mix the
contents. This method may not work with the cold rod insert, unless the bar is placed on top
of the insert and the phantom turned upside down and placed on the stirrer. In using this
technique keep in mind that the bar might be visible in some of the reconstructed image
slices.
We encourage that you experiment to become aware of the best method of filling the
phantom for your particular application. As you gain experience filling the phantom this task
will become more efficient and seem routine. One hint with regard to water temperature: Do
not fill the phantom with cold water and tightly seal it since this water must expand as it
warms to room temperature. This expansion can damage the phantom, especially if it is
sealed extremely tightly. Prior to filling the phantom, we have found it useful to fill a
separate large container with several liters of cold tap water and allow it to set for a few
hours until it reaches room temperature. In general it is not possible to pick the "ideal" water
temperature each time, so quite often due to the increase in pressure that is associated with
small amounts of expansion, a bit of spray may result when a filler cap is removed. To
protect yourself and nearby objects from this spray, always loosen filler caps with a towel
surrounding the cap. Note that allowing the water to equilibrate to room temperature before
sealing the phantom, or simply using warmer, room temperature water to start with can help
alleviate most of these temperature related issues.
7
SPECT Phantom User Manual
8
SPECT Phantom User Manual
9
SPECT Phantom User Manual
collimator dents. Thicker slices may be used with the Rod Insert and the Line Sources,
provided the phantom is mounted such that the long axis of the phantom is parallel to the
axis-of-rotation. Obtaining the scan with a thin slice thickness allows for later slice addition,
which is often useful for viewing the rod insert. For this method to work well, you must very
accurately align the phantom (in this case, the long axis of the rods) with the axis-of-rotation.
How do I know if the phantom is accurately aligned with the axis-of-rotation?
One may position the phantom and view a persistence scope or computer monitor
image to help determine proper alignment. Quickly acquired planar test images might work
better on some systems. Make sure that the top/bottom and the sides of the phantom are
“square” with the sides of the image matrix. You will need to perform this for the camera
looking up or down at the cylinder (like an anterior or posterior view), and then with the
camera looking from the “side” (like a lateral view). Repeat this another time or two until
you are satisfied with the positioning. This alignment is most critical for situations in
which you want to add many, perhaps all, of the slices containing the cold rods. Alignment
with the axis-of-rotation is not very critical for imaging the other inserts such as the spheres
or uniform activity. Line sources should be reasonably well aligned so that measurements of
resolution (FWHM or FWTM) provide you with accurate values.
10
SPECT Phantom User Manual
Proper Care
1. Periodically clean the phantom using only a mild (dishwashing or other) soap
solution.
2. If the phantom is not to be used for an extended period of time, remove liquid, clean
and dry. Remove caps and leave the inside of the phantom open to air for ventilation.
3. Periodically inspect phantom for damage. If damage has occurred, repair or replace
phantom prior to further use.
4. The O-rings require lubrication and should have a liberal amount of silicone-based
grease applied every month or so depending on how often the phantom is used.
Silicone grease may be purchased in the plumbing department of a home
improvement store such as Home Depot or Lowes.
5. Handle the phantom and inserts with care. Each component is precision- and custom-
made. Avoid dropping or striking since acrylics are relatively fragile..
** WARNING ** Avoid over-tightening caps or other plastic or nylon parts. Threads
can be damaged or screws broken. Usually, hand tightening of any threaded part is adequate.
Avoid using metal screws on the cover plate. With the flangeless model never, never use
longer replacement cover plate screws than the ones that come with the original phantom.
Longer screws, particularly longer metal screws, will “bottom out” before the cover plate is
seated and potentially will crack the thin acrylic cylinder. This would void the warranty.
Also, with the flangeless model, be careful not to confuse the threaded screw holes in the
cover plate that can be used to easily remove the cover plate from the cylinder with the un-
threaded holes that are use to secure the cover plate to the cylinder. Failure to follow this
precaution may damage the screws and/or the phantom. Data Spectrum Corporation can
supply replacement components. The screws can also be lubricated with silicone-based
grease. Silicone grease can be obtained in the plumbing department of home improvement
stores such as Home Depot™ and Lowes™.
Avoid strong detergents or “glass cleaners” containing ammonia, alcohols,
solvents, chlorinated hydrocarbons, petroleum products, extreme hot or cold water, and
exposure to extreme heat and cold. Never, under any circumstances, fill the phantom
with alcohols, solvents, chlorinated hydrocarbons, or petroleum products.
A maximum of 5 parts per million of chlorine can be safely dissolved in the water
used to fill the phantom to retard biological growth. Swimming pool chlorine additives may
11
SPECT Phantom User Manual
be used, but it is important that the chlorine concentration be measured accurately with an
appropriate test kit, such as those used to test swimming pool or spa water.
We wish you the best of success with the use of your emission computed tomography
system, and hope that your Data Spectrum SPECT Phantom can play an important role in
acceptance testing and routine quality control procedures. We would be most interested in
your comments, questions, suggestions and results. Thank you for considering Data
Spectrum Corporation. For your convenience we have included in the following pages, a few
representative quantitative calculations that can be performed using the SPECT Phantom.
12
SPECT Phantom User Manual
* Activity in SPECT Phantom required for obtaining approximately 20 million counts within an 8 cm-thick
slice containing the cold rods. A single camera SPECT system has been assumed. Note that a “high”
resolution collimator (or “LEAP”, “LEUR” etc.) from a given manufacturer does not necessarily correspond
to the “high” resolution collimator of another manufacturer, either in sensitivity, high energy penetration or
resolution.
13
SPECT Phantom User Manual
II. COLLIMATION: Use of your High or Ultra High Resolution Collimator (LEHR or
LEUR) is recommended.
14
SPECT Phantom User Manual
SENSITIVITY MEASUREMENTS
Sensitivity can be measured using any of several methods, although each method may
provide slightly different results. Sensitivity measurements can be made with an isolated
point source in air, distributed (“flood”) sources or cylindrical phantoms. One widely used
method, described below, uses a water-filled, cylindrical phantom filled with a low
concentration of radioactivity. As one may expect, a number of variables should first be
considered before acquiring the data for the sensitivity measurements.
One debate revolves around the question of whether sensitivity should be stated in
terms of single slice sensitivity or total volume sensitivity. For scanners that acquire only
one slice at a time, the numbers are the same. Total volume sensitivity can be considered as
the sensitivity for a single slice multiplied by the total number of slices for a multi-slice,
camera based single photon or multiple ring positron ECT system.
Sensitivity may be expressed in terms of (cnts•sec-1)•(Ci•ml-1)-1. In order to
determine these measurements, first fill a 21 cm diameter cylinder with a measured volume
of water. No inserts (rods, spheres, etc.) should be in the phantom. Add—and mix well in
this water—a known amount of radioactivity of the radionuclide chosen for the
measurements. For example, 2.49 mCi (at scan time) was mixed into 7155 mL of water,
resulting in 0.348 Ci•ml-1. The SPECT data acquisition was for a total of 349 seconds,
using a low energy, high resolution collimator. The radionuclide was Tc-99m and the energy
discriminator was centered at 140 keV with a 20% (FWHM) wide window. The slice
thickness was 2.5 cm with a total of 11 simultaneous slices in one acquisition. This
particular study was performed with a dual-head SPECT system, a fact that must be taken
into account. Note that the amount of radioactivity was kept low mainly to eliminate the
effects of system dead-time on the sensitivity measurements.
When the projection data were reconstructed, the number of counts-per-slice for one
detector in the original data were as listed below:
Note that the counts in the lower and higher slices are significantly lower in number
than those in the middle of the phantom (slices 4 through 7). The reason is because this
particular phantom did not span the entire length of the field-of-view along the axis-of-
rotation. This must also be taken into account. If we average the counts for the center slices
-1 -1
(5 through 7), we shall arrive at an average of 165,521 (counts•slice •detector ). Dividing
165,521 counts by 349 seconds (the scan time), and dividing this quotient by 0.348 Ci•ml-1
(the concentration you prepared) gives us the sensitivity for one slice (pixel thickness equal
to 2.5 cm) for one detector, or approximately 1363(cnts•sec-1)•(Ci•ml-1)-1. Recall that the
total volume sensitivity is the product of the single slice sensitivity (1,363) times the number
15
SPECT Phantom User Manual
of slices (11) times the number of detectors (2). Thus the total volume sensitivity of this
ECT system, using the collimators and energy window settings previously described, is equal
-1 -1 -1
to 29,986 (cnts•sec )•(Ci•ml ) . The single slice sensitivity for both detectors, for a pixel
-1 -1 -1
slice thickness of 2.5 cm, is equal to 2,726 (cnts•sec )•(Ci•ml ) . Sensitivity measurements
are required at low counting rates so that deadtime effects are avoided or negligible.
Sensitivity measurements are a critical part of routine quality control. It is a good
idea to perform this type of measurement immediately after initial installation, after software
upgrades and after hardware modifications. Remember that collimators, window-widths,
dead-time, gamma ray energy, slice thickness, field-of-view and other factors can directly
influence the resulting values.
16
SPECT Phantom User Manual
size; although, remember that if the inside diameter of the tube is increased too much, the
resolution measurements could be affected due to “tilting” of the tube. As the glass capillary
tubes are very fragile, it is advisable to practice the procedures of cutting the tubes and
mounting them a few times, prior to actually using source-filled tubes. Of course, the
phantom is to be filled either with “air” for line source measurements made in air, or with
water for measurements made in water.
After the line sources have been mounted, the phantom is positioned so that the line
sources are parallel to the axis-of-rotation, and then the line sources are scanned with the
SPECT system. Of course, conventional measurements may be made by simply taking
planar views. Measurements are made by generating numerical (or graphical) profiles
(displayed either linearly or on a log scale) through the reconstructed images of the line
sources. Since the line source response function of most SPECT systems is not precisely
circularly symmetric (largely due to geometric collimation variations as a function of the
distance of the source from the collimator surface), it is recommended that more than one
profile be obtained. Typically a radial profile (i.e., consisting of a line drawn through both
the axis-of-rotation and the center of the line source image), and a tangential profile (i.e.,
consisting of a line which passes through the center of the line source image and is tangent to
an imaginary circle whose center is at the axis-of-rotation, and whose circumference passes
through the center of the line source image) are used. These profiles may be a few pixels
wide in order to encompass the central region of the image of the line source. With the
numerical data values for the pixels comprising these profiles, either modulation transfer
functions (MTFs) may be computed, or, more simply, the full-width-at-half-maximum
(FWHM) and full-width-at-tenth-maximum (FWTM) of the line source image may be
estimated. The latter calculations are illustrated in the sample profile presented below.
These data correspond to a SPECT system using
-1
3.2 mm•pixel linear samples, 180 angular views (2 degree angular sampling), continuous
rotation of the gantry, a 15 cm radius-of-rotation (measured from the front surface of the
-1
collimator) and a Von Hann filter with a cut-off frequency equal to 1.6 cycles•cm .
17
SPECT Phantom User Manual
17748 13807
times1pixel
17748 6054
. ymax 2761.4
01
FWTM(pixels) = 5 pixels
3704 2761
+ times1pixel
3704 508
6054 2761
+ times1pixel
6054 1133
Note that the data values were simply linearly interpolated to obtain the fractional
pixel widths. A more precise interpolation method (e.g., five-point) may also be used.
Another approach would be to perform a "least-squares-fit" of the data values to an assumed
Gaussian function, and then use that function to analytically calculate the FWHM and
FWTM. Note that since the linear sampling is only 3.2 mm•pixel-1, the measured response
function will be a convolution of this sampling with the true response. Thus aliasing can
18
SPECT Phantom User Manual
occur and make the measurements somewhat inaccurate. Unfortunately, this effect will be
increased if one uses 6.4mm sampling (i.e., 64×64 reconstructions). Fortunately, most
SPECT camera manufacturers are supplying 128×128 capability. Ideally, even finer
sampling (e.g., 256×256 or 512×5l2) would be desirable for these measurements. Sometimes
a “zoom-mode” spatial resolution is a complicated function of intrinsic and extrinsic camera
performance, sensitivity variations (non-uniformities), spatial non-linearity’s and distortions,
collimation characteristics, centering, gain, asymmetric image size, radius-of-rotation, linear
and angular sampling intervals, reconstruction filters, attenuation compensation algorithm,
amount of scattering material present, etc. Thus, if one would like to use the above described
procedure to compare the SPECT system’s performance over a period of time, one must be
careful to ensure that none of these parameters have changed.
N
1
Yavg
N
y
i 1
i
where N is the number of pixels (picture elements) within a region of interest (ROI)
near the center of the reconstructed image of the uniform cylindrical portion of the
SPECT phantom, and yi is the number of counts per pixel in the ith pixel. It is assumed
for the computation that the SPECT image was reconstructed using an appropriate (e.g.,
Chang) attenuation compensation algorithm.
y
N
1 2
S dev Yavg , or
N 1
i
i 1
19
SPECT Phantom User Manual
N
1
S dev
N 1
( y
i 1
2
i ) N Y avg
2
% rms S dev / Yavg 100
IV. Example Data Values:
The following pixel values - expressed here as real fractions - were measured within a
5×5 pixel (each pixel equals 3.175 mm) ROI located near the center of the reconstructed
image of the uniform cylindrical portion of the SPECT phantom (using a 128×128 image
matrix and a generalized Von Hann filter having a cutoff frequency equal to 1.6 cycles per
cm):
Using the equations and definitions presented above, the measured parameters are:
Sdev 2.8
2.8
Fractional root-mean-square (rms) noise 0.086 , or
32.5
%rms 0.086100 8.6%
It should be noted that for a pixel “signal” value (i.e., the difference between that
pixel value and the average background Yavg) to be statistically significant, this signal value
should be at least 3 to 4 times greater than the measured standard deviation (Sdev). Also note
that the derivation presented above corresponds to noise related to regions of the size of a
“pixel” however, it is possible to extend this methodology to reflect the noise level found in
larger regions (i.e., such as 1 to 2 cm, etc.), although care, of course must be exercised.
Finally, remember that the noise spectral density of reconstructed images is not “white,” or
Poisson, but is actually structured due to finite angular sampling and other characteristics of
the reconstruction process.
20
SPECT Phantom User Manual
sphere bkg
Counts Pixel
C im age 1
= Signal cnts/background cnts.
bkg
For example, the following pixel values (expressed here simply as normalized
integers) were measured within the reconstructed image of a large sphere using a small ROI
(3×4 pixels):
Note that the mean counts per pixel within this ROI or the counts per pixel is simply 15618.
A background ROI was selected adjacent to the sphere of interest, and the following pixel
values were measured within an ROI having the dimensions of 4×4 pixels:
21
SPECT Phantom User Manual
Note that the estimated mean counts per pixel within the selected background region
of interest is equal to 25507 counts per pixel. The sphere’s image contrast is thus equal to:
C im age
15618 25507 = Signal cnts/Normal background cnts.
25507
0.39 or 39%
It has been pointed out in scientific literature that two conditions must be satisfied in
order that a lesion located within a relatively uniform background be reliably detected. First,
the signal (i.e., the numerator in the contrast equation) must be greater than 3 or 4 times the
standard deviation, that is, the value of Sdev as defined in the last section. This is at least
valid for small “resolution-sized” lesions. For larger lesion diameters this criterion may be
relaxed somewhat. Furthermore, even if this first condition is satisfied the lesion's image
contrast must still be greater than the observer’s visual contrast threshold defined as the
minimum contrast of an object of the shape of a small sphere that an observer can reliably
detect on a “screen or film”.
This latter value has been estimated to be equal to an image contrast of about 12%, or
equal to an optical density of about 0.09 when using a typical nuclear medicine film. SPECT
images of small lesions that satisfy both of the above conditions may or may not be detected.
If the SPECT system were perfect, then it would be possible to detect the small lesions.
However, if there are systematic errors such as non-uniformities, etc., then the small lesions
may not be detected.
It can finally be noted that this derivation can be extended to determine an estimate of
signal-to-noise (S/N) ratios for arbitrarily sized defects having arbitrary radionuclide uptake
ratios compared to surrounding background activity. However, extreme care must be taken
in appropriately determining the regions used for the noise power - as it will now depend, not
only on the various reconstruction parameters, but also on the size of the simulated lesions.
These calculations are beyond the scope of this manual. Needless to say, it is also imperative
that the slice selected to perform these quantitative measurements on should be precisely
centered through the diameter of the sphere, since partial volume effects will effect the
measured results.
22
SPECT Phantom User Manual
23
SPECT Phantom User Manual
SPECT Phantom
RODS SPHERES
UNIFORM LINE
SOURCES
The SPECT image of the Cold Rods had an effective slice thickness
of approximately 8 cm, while the effective slice thickness for the Spheres was approximately
12 mm. The radius-of-rotation was equal to approximately 13 to 14 cm. Over 30 million
counts were used to reconstruct the image of the Cold Rods. Approximately two to three
million counts were acquired for the slice containing the Spheres. The SPECT system was
reasonably well-calibrated, although as indicated by the image of the Uniform Section, the
value used for the attenuation coefficient was slightly too great, resulting in an apparent
increase in activity towards the central region of the Phantom.
24
SPECT Phantom User Manual
EFFECT OF COLLIMATION
Data Spectrum's Deluxe Phantom (Model ECT/DLX/P) was used. Each SPECT image
shown has an effective slice thickness equal to approximately 8 cm. Radius-of-rotation
(ROR) was equal to approximately 13 to 14 cm. Approximately a total of 20 to 30 million
events were acquired to reconstruct each section shown. The SPECT system was
“reasonably” well-calibrated. As a result of its relatively large loss of geometric resolution
as a function of distance from the collimator surface, the Low Energy, All Purpose collimator
may not be appropriate for many SPECT imaging situations, except perhaps for very low
count density studies. Although the images obtained with the Ultra-High Resolution
Collimator are good, it has a very low relative sensitivity, as compared with the other
collimators illustrated.
25
SPECT Phantom User Manual
ROR
14cm
ROR
22cm
EFFECT OF RADIUS-OF-ROTATION
Data Spectrum’s Standard Phantom (Model ECT/STD/P)) was used. The SPECT system was
moderately well-calibrated. The fall-off with distance for both collimators are illustrated.
SPECT system (not clinical) performance evaluation is best performed with the collimator
positioned as close as possible to the phantom (i.e., using the 13 to 14 cm radius-of-rotation,
or ROR). The image quality of the Low Energy All-Purpose Collimator degrades more
rapidly as the radius-of-rotation is increased.
26
SPECT Phantom User Manual
27
SPECT Phantom User Manual
28