ELECTRODES
Electrodes are used to pick up the electrical signals of the
body. They transfer the bioelectric event to the amplifier. The type
10
11
electrode metal, its area, electrolyte, charge density and frequency of
current. The electrode potential is measured with reference to
hydrogen electrode placed in electrolyte near metallic electrode. The
half cell potential development can be expressed by Nernst equation
as
RT C f
E= ln 1. 1
C f
nF 2 2
Where R – Gas Constant
T – Absolute Temperature
F – Faraday Constant
N – Valency of Ion
C1, C2 – Concentration of selected ion on two sides of
membrane
f1, f2 – Activity coefficients of ion on two sides of
membrane.
Purpose of Electrode paste:
The outer skin of the body is highly non- conductive and will
not establish a good electrical contact with an electrode. The skin
should be washed and rubbed to remove some of the outer cells. The
electrode paste decreases the impedance of contact and also reduces
the artifacts resulting from movement of electrode.
Electrode Material:
The electrode, electrode paste and body fluids can produce
a battery like action causing ions to accumulate on the electrodes.
This polarization of electrode can affect the signal transfer.
Types of Electrodes:
These are three types of electrodes
Micro electrodes (Intracellular Electrodes):
These are used to increase the bioelectric potential within a
single. It is divided into metallic and non metallic. The
microelectrodes should have smaller diameter and during insertion
of electrode into the cell, there will not be any damage to the cells.
They are formed by electrolytically electing the tip of a fine
tungsten or stainless steel wire to a fine point. This technique is
known as electro pointing. The metal microelectrodes are coated
almost to the micro tip with a material.
In standard leads, the potential are tapped from four locations of our
body. They are i) right arm ii) Left arm iii) Right Leg and iv) Left
leg.
Right leg electrode is acting as ground reference electrode.
Fig.1.20 shows the standard bipolar limp lead positions and the
corresponding wave patterns.
Lead I position – give voltage VI, the voltage drop from left arm(LA)
to right arm (RA)
Lead II position- gives voltage VII, voltage drop from left leg (LL) to
right arm (RA)
Lead III position – gives voltage VIII, voltage drop from left leg (LL)to
left arm (LA)
is provided by Rs.
Normal ECG
Here the PQ segment has prolonged conduction time i.e greater than
0.22second
Result: First degree AV block
Here QRS complex is widened i.e QRS interval is greater than 0.1
second
Result: Bundle block
PHONOCARDIOGRAPHY (PCG):
The graphic record of the heart sounds is called
―Phonogram. Since the sound is from the heart, it is called
phonocardiogram. The instrument used to measure the heart sounds
is called phonocardiograph.
1) Heart sounds
2) Murmurs
Heart sounds have a transient character and are of short duration.
Heart murmurs have a noisy characteristic and last for a longer time.
Heart sounds are due to the closing and opening of valves whereas
murmurs are due to turbulent flow of blood in the heart and large
vessels.
Heart sounds
Heart sounds are classified into four groups based on their
origin. They are
1) Valve closure sounds
2) Ventricular Filling sounds
3) Valve opening Sounds
4) Extra cardiac sounds
1) Valve Closure Sounds
The sounds occur at the beginning of systole (First heart
sound) and the beginning of diastole (Second heart sound). The first
heart sound is due to the closure of mitral and tricuspid valves. The
second heart sound is due to the closure of aortic and pulmonary
valves.
2) Ventricular filling sounds.
These sounds occur either at the period of rapid filling of the
ventricles (Third heart sounds) or during the terminal phase of
ventricular filling (ie) atrial contraction. These sounds are normally
in audible.
3) Valve opening sounds
They occur at the time of opening of atrio – ventricular
valves and semilunar valves.
4) Extra cardiac sounds
26
They occur in mid (or) late systole (or) early diastole. They
are caused by thickened pericardium which limites ventricular
distensibility.
Physical characteristics of sound
Heart sounds and murmurs are characterized by three
physical properties. They are
1) Frequency
2) Amplitude
3) Quality
1) Frequency: All heart sounds and murmurs are made up of
frequencies between 10 and 1000 Hz. They are divided into low,
medium and high- pitch frequencies
i) Low range: 10 – 60 Hz. It is represented by the third and
fourth heart sounds.
ii) Medium range: 60 – 150 Hz. It is represented by the first
and second heart sounds.
iii) High range: 150 – 1000 Hz. It is represented by snaps,
clicks and diastolic murmurs of aortic and pulmonary insufficiency.
1) Amplitude: Low frequency heart sounds have the biggest
amplitude while the high frequency murmurs have small amplitudes.
2) Quality: quality depends upon the overtones (or) harmonics
accompanying the fundamental frequency and applies to tones.
Origin of the heart sounds.
There are four separate heart sounds that occur during
the sequence of one complete cardiac cycle.
1) First heart sound: It is produced by a sudden closure of mitral
and tricuspid valves associated with myocardial contraction.
a) Timing: The low frequency vibrations occur approximately
0.05 sec after the onset of QRS complex of ECG.
b) Duration: It lasts for 0.1 to 0.12 sec.
c) Frequency : The first heart sound range from 30 – 50 Hz
d) Asculatory area: The first heart sound is best heard at the
apex of the mid pericardium.
2) Second heart sound: It is due to the closure of semi lunar valves
Medical Electronics 27
4) Placement of electrode
In EEG, electrodes are placed in standard positions on skill
in an arrangement called 10 – 20 system. The electrodes are arranged
as follows.
1) Draw a line on the skull from the nasion, the root of nose, to the
inion, ossification center on occipital lobe.
2) Draw a similar line from the left preauricular (ear) point to the right
preauricular point.
3) Mark the intersection of two lines as Cz which is the midpoint of
distance nasion and inion.
4) Mark points Fpz, Fz, Cz, Pz and Oz at 10, 20, 20, 20, and 10% of
total nasion – inion distance.
5) Mark points T3, C3, Cz, C4, and T4 at 10, 20, 20, 20 and 10%
of total distance between preauricular points.
6) Measure the distance between Fpz and Oz along the circle
passing through T3, and mark points as Fp1, F7, T3, T5, and O1 at
10, 20, 20, 20 and 10% of this distance.
7) Repeat this procedure on right side and mark the positions as Fp2,
F8, T4, T6, and O2.
8) Measure the distance between Fp1, and O1 along the circle
passing through C3 and mark point as F3, C3, and P3 at 25%
intervals.
9) Repeat this procedure on right side and mark as F4, C4 and P4.
10) Check that F7, F3, Fz,F4 and F8 are equidistant along transverse
circle passing through F7, Fz, and F8 check that T5, P3, Pz,P4, and T6,
a re equidistant along transverse circle passing through T5, Pz, &
T
6
Pg1 AND Pg2 are nasopharyngeal electrodes and A1 and
A2 are ear electrodes.
34
Fig 1.34: Placement of lectrodes on the Scalp for the EEG Recording
[Source: Leslie Cromwell - Biomedical instrumentation and
measurement]
Pg1 AND Pg2 are nasopharyngeal electrodes and A1 and
A2 are ear electrodes. The electrode systems are used to facilitate the
location of foci, (ie) cortical areas from which abnormal waves
spread.
5) Recording Setup
In EEG recording setup, there are and drive amplifier whose
gain can be increased by cascading several stages. The patient cable
consists of 21 electrodes and is connected to the eight channel
selector. The electrodes are attached to the channel selector in groups
of eight called a montage of electrodes. The interference is reduced
by employing differential amplifiers as preamplifiers.
Medical Electronics 35
Fig 1.36: Block Diagram for EMG Recording Setup [Source: Leslie
Cromwell - Biomedical instrumentation and measurement]
The needle electrode picks up the action potentials from
selected nerves or muscle. Further to record the action potentials from
a signal nerve, microelectrodes are used.
The amplitude of EMG signals depends upon the type and placement of electrodes used.
Surface electrode picks up many overlapping spikes and produces an average voltage from
various muscles.
The amplifier should have uniform frequency response, high CMRR and high input
impedance. The signal is also recorded in the tape recorder for further reference.
Determination of conduction velocities in motor nerves.
The measurement of conduction velocity is used to indicate the location and type of
nerve lesion.
The EMG electrode and stimulating electrode are placed at
two points on the skin, separated by a known distance l1.An electrical pulse is applied through the
stimulating electrode. The latency is now
measured as ‗ t 2‘seconds. The conduction velocity is
U = l1- l2 / t 1 –t2
The conduction velocity in peripheral nerves is normally 50 m/s. when it is below 40 m/s, there
is some disorder in nerve conduction.