R860 Data Output Guide - FINAL

The document is a quick reference guide for the CARESCAPE R860, detailing setup and measured data across different versions (COM1.3, COM1.5, COM1.7). It includes information on alarm status, system data, and procedures data, indicating which features are supported by each version. Additionally, it outlines data transfer configurations and calculations related to respiratory measurements.

Uploaded by

emrekeskiner
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
21 views4 pages

R860 Data Output Guide - FINAL

The document is a quick reference guide for the CARESCAPE R860, detailing setup and measured data across different versions (COM1.3, COM1.5, COM1.7). It includes information on alarm status, system data, and procedures data, indicating which features are supported by each version. Additionally, it outlines data transfer configurations and calculations related to respiratory measurements.

Uploaded by

emrekeskiner
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Data Output Quick Reference Guide

CARESCAPE R860 ™

Datex-Ohmeda Com 1.3, 1.5, 1.7

Setup data Measured data (continued)


FIELD COM1.3 COM1.5 COM1.7 FIELD COM1.3 COM1.5 COM1.7
soft rev X1 X X VCO2/m2 X X
language X X X VO2/kg X X X
alarm loudness X X X VCO2/kg X X
TVexp/wt X X
altitude setting X X X
MVexp/wt X X
altitude unit X X dynamic PEEPi X1 X X
model number X1 X X pressure units X X X
serial number X X X flow units X X X
CO2 units X X X
gas supply units X X
Measured data EE units X X
FIELD COM1.3 COM1.5 COM1.7 O2 supply pressure X X X
Ppeak X1 X X air supply pressure X X X
Dynamic PEEPe X1 X X Paw source X X X
Pmean X1 X X flow source X X X
Pplat X1 X X volume source X X
Leak X X O2 source X X X
MVexp X1 X X ambient pressure X X X
TVexp X X X RRCO2 X1 X X
RR total X1 X X Tvinsp X1 X X
FiO2 X1 X X Mvinsp X1 X X
MVexp spont X1 X X I:E (xxx.x:xxx.x) X X X
TVexp spont X X X meas data status X X X
RR spont X1 X X inspiratory time X X X
RSBI X X expiratory time X X X
MVexp mech X X
TVexp mech X X
RR mech X X
Measured data byte
Paux Peak X X X FIELD COM1.3 COM1.5 COM1.7
Paux Mean X X X new breath/10sec data X X X
Paux Min X X X mechanical breath X X X
etCO2 X1 X X ctrl triggered breath X X X
FiCO2 X1 PSV triggered breath X X X
etO2 X X X
FiO2 (MGAS) X1 X X
FiO2 - EtO2 X X X CARESCAPE R860 Electrical Interface
dynamic compliance X1 X X RS-232C signal standards
Raw X1 X X 9 pin female D connector – The connectors are located on the back of the
static compliance X X X display unit and are labeled Port 5 and Port 6
static Pplat X X
static PEEPe X X X
Data Communications Equipment configuration (DCE)
static PEEPi X X Pin 2 – receive data
PEEPe+I X1 X X Pin 3 – transmit data
EE X X X Pin 5 – signal ground
RQ X X X Pin 7 – do not connect
VO2 X X X
Baud Rate: 19.2K
VCO2 X X X
VO2/m2 X X X Byte format: Start bit + 7 data bits + parity bit + stop bit
Parity: odd
NOTE: Requires use of serial isolation cable, GE part number 2070435-001

‘ ‘ Not Supported, ‘X’ Supported, ‘X1’ Supported by Unity Network ID v7A (refer to product specific documentation for more information)
Status data Alarm Status data bytes
FIELD COM1.3 COM1.5 COM1.7 FIELD COM1.3 COM1.5 COM1.7
set vent mode X1 X X Backup Buzzer POST X X X
set TV X1 X X Patient Connected X1 X X
set PEEP X1 X X No Battery Backup X X
set Tinsp X1 X X Temp High X X
set Pinsp X1 X X Primary Audio Failure X X X
set Psupp X1 X X Power Supply Fail X X
set Plimit X X X Exp Flow Sensor Error X X
set RR X1 X X Neo Flow Sensor Error X X
set mech RR X1 X X Neo Flow Sensor Reversed X X
set insp pause X X X Neo Flow Sensor Contaminated X X
set I:E (xxx.x:xxx.x) X1 X X Replace Neo Flow Sensor X X
set flow trigger X1 X X Neo Flow Sensor Off x X
set end flow X X X Negative Airway Pressure X1 X X
set bias flow X1 X X No Exp Flow Sensor X1 X X
set pressure rise time X X X No Neo Flow Sensor X X
set flow rise time X X Exp Flow Sensor Failure X X
set pressure trigger X X X O2 Sensor Failure X X X
set Pmax X X X Air Supply Pressure High X1 X X
set PSV rise time X X X Air Supply Pressure Low X X X
set flow X X O2 Supply Pressure High X1 X X
set vent assist control X X O2 Supply Pressure Low X X X
set pause time X X No Gas Supply Pressure X X X
set FiO2 X1 X X Air Supply Pressure Sensor Out of Range X X
set Phigh X X O2 Supply Pressure Sensor Out of Range x X
set Plow X X Pinsp Sensor Out of Range X X
set Thigh X X Pexp Sensor Out of Range X X
set Tlow X X Paux Sensor Out of Range X X
set Tsupp X X FiO2 Control Error X X
set high MVexp limit X1 X X Volume Delivery Error X X
set low MVexp limit X1 X X Air Temp Sensor Error X X
set high TVexp limit X1 X X O2 Temp Sensor Error X X
set low TVexp limit X1 X X Mixed Gas Temp Sensor Error X X
set high FiO2 limit X X X Total Flow Sensor Comm Fail X X
set low FiO2 limit X X X High Temperature X X X
set high RR limit X1 X X Module Not Compatible X X X
set low RR limit X X X Check Sample Gas Out X X X
set high EtCO2 limit X X X Replace D-Fend X X X
set low EtCO2 limit X X X Sample Line Blocked X X X
set high Ppeak limit X X X Check D-Fend X X X
set low Ppeak limit X X X Sample Flow Deviation X X
set high EtO2 limit X X MGAS Failure X X X
set low EtO2 limit X X No D-Lite Sensor X X
set high PEEPe limit X X Front Panel Comm Failure X X X
set low PEEPe limit X X Pressure Sensor Failure X X X
set high PEEPi limit X X Nebulizer Not Connected X X X
set high Paux limit X X On Battery X X X
set apnea delay time X1 X X Low Internal Battery - 20 min X X
set circuit leak limit X X Low Internal Battery - 10 min X X
set disconnect time X X Low Internal Battery - 5 min X X
set patient effort time X X Low Internal Battery - 1 min X X
set tube type X X X No Battery X X X
set tube size X X X Low Battery X1
set ARC level X X Battery Failure X
set patient type X X X Battery Charger Failure X
set BSA X X X Fan Fail X X X
set patient height X X X Relief Valve Failure X X
set patient weight X X X Backup Mode Active X X X
set weight units X X SBT Completed X X
set height units X X SBT Ended X X
set patient ID X X X Cannot Calculate FRC X X
set flow sensor type X X X Missed Scheduled FRC X x
set nebulizer time X X CPU Fan Speed Fail X X
set nebulizer volume X X X Case Fan Speed Fail X X
setting data status X X Ppeak High X1 X X
set system configuration X X Relief Valve Opened X X
set Pmin X X Ppeak Low X1 X X
set continuous nebuilzation X X Sustained Paw X1 X X
Breathing Circuit Occlusion X1 X X
Circuit Leak X1 X X
Status data byte Circuit Leak Off X X
Patient Connection Leak X X
FIELD COM1.3 COM1.5 COM1.7 Apnea X1 X X
new setting/10 sec data X X X Apnea Alarm Off X X

‘ ‘ Not Supported, ‘X’ Supported, ‘X1’ Supported by Unity Network ID v7A (refer to product specific documentation for more information)
Alarm Status data bytes (continued) System data
FIELD COM1.3 COM1.5 COM1.7 FIELD COM1.3 COM1.5 COM1.7
MVexp Low X1 X X current date (yyyymmdd) X X X
MVexp High X1 X X current time (hhmmss) X X X
MVexp Low Alarm Off X X system state X X
No Patient Effort X X
RR Low X1 X X
RR High X1 X X Procedures data
FiO2 Low X1 X X FIELD COM1.3 COM1.5 COM1.7
FiO2 High X1 X X
procedure status bytes X X
VTexp Low X1 X X
PEEPi X X
VTexp High X1 X X
PEEPi Volume X X
VT Not Achieved X1 X X
P0.1 X X X
EtO2 Low X X X
vital capacity X X
EtO2 High X X X
negative inspiratory force X X
EtCO2 Low X1 X X
Pplat X
EtCO2 High X1 X X
Static Compliance X
Patient Disconnected X1 X X
PEEPe High X X
PEEPe Low X X Procedures data bytes
PEEPi High X X
Paux High X X FIELD COM1.3 COM1.5 COM1.7
Plimit Reached X X X nebulizer on/off X X X
MGAS Warming Up - 2 min X X X pre-oxygenate on/off X X X
MGAS Warming Up - 5 min X X X suction on/off X X X
Alarm Light Fail X X post-oxygenate on/off X X X
Carrier Board Overheat X X intrinsic PEEP on/off X X
CPU Overheat X X inspiratory hold on/off X X
Touch screen fail X expiratory hold on/off X X
Service Cal Required X SBT on/off X
Standby On X oxygenation on/off X X
Alarms Silenced X X X P0.1 on/off X X
New Alarm/10 second data X X X negative inspiratory force on/off X X
vital capacity on/off X X

Waveform data
Data Transfer Configuration
FIELD COM1.3 COM1.5 COM1.7
1st waveform block X1 X X FIELD COM1.3 COM1.5 COM1.7
2nd waveform block X1 X X data to transfer X X
3rd waveform block X1 X X time period X X
4th waveform block X1 X X sample interval X X
5th waveform block X X X waveforms on/off X X
6th waveform block X X X patients X X
breath end/start index X1 X X transfer media X X

Spiro Dynamics data Snapshot Notification


FIELD COM1.3 COM1.5 COM1.7 FIELD COM1.3 COM1.5 COM1.7
pressure waveform X X X X
volume wavform X X
breath start/end index X X
sequence number X X Calculations data
FIELD COM1.3 COM1.5 COM1.7
sample date (yyyymmdd) X
Dynostatic Curve data sample time (hhmm) X
FIELD COM1.3 COM1.5 COM1.7 set hb X
Paux curve data X X set SaO2 X
volume curve data X X set SpO2 X
low compliance X X X set SvO2 X
mid compliance X X X set PaO2 X
high compliance X X X set PvO2 X
sequence number X X set PaCO2 X
calculated Pa/FiO2 X
calculated AaDO2 X
Checkout data calculated CO X
calculated Vd/Vt X
FIELD COM1.3 COM1.5 COM1.7
calculated Vd X
last checkout date stamp (yyyymmdd) X X X
calculated VA X
last checkout time stamp (hhmm) X X X
calculated PAO2 X
breathing circuit leak X X
calculated PaO2/PAO2 X
breathing circuit compliance X X
calculated OI X
breathing circuit resistance X X
calculated SpO2/FiO2 X
checkout verdict X X

‘ ‘ Not Supported, ‘X’ Supported, ‘X1’ Supported by Unity Network ID v7A (refer to product specific documentation for more information)
Metabolics data FRC data
FIELD COM1.3 COM1.5 COM1.7 FIELD COM1.3 COM1.5 COM1.7
start date (yyyymmdd) X PEEP Inview measurement 3 time (hhmm) X
start time (hhmm) X PEEP Inview measurement 3 set PEEP X
end date (yyyymmdd) X PEEP Inview measurement 3 FRC volume X
end time (hhmm) X PEEP Inview measurement 3 PEEPi X
VO2 Average X PEEP Inview measurement 3 PEEPe X
VO2 Variance X PEEP Inview measurement 3 Cstat X
VCO2 Average X PEEP Inview measurement 3 Delta FRC X
VCO2 Variance X PEEP Inview measurement 3 Spiro
X
RQ X Dynamics Volume
EE X PEEP Inview measurement 3 Gain X
EE/kg X PEEP Inview measurement 4 date (yyyymmdd) X
EE/m2 X PEEP Inview measurement 4 time (hhmm) X
PEEP Inview measurement 4 set PEEP X
PEEP Inview measurement 4 FRC volume X
FRC data PEEP Inview measurement 4 PEEPi X
PEEP Inview measurement 4 PEEPe X
FIELD COM1.3 COM1.5 COM1.7
PEEP Inview measurement 4 Cstat X
procedure date (yyyymmdd) X PEEP Inview measurement 4 Delta FRC X
procedure time (hhmm) X PEEP Inview measurement 4 Spiro
FRC volume X X X X
Dynamics Volume
PEEPi X PEEP Inview measurement 4 Gain X
PEEPe X PEEP Inview measurement 5 date (yyyymmdd) X
Cstat X PEEP Inview measurement 5 time (hhmm) X
PEEP Inview measurement 1 date (yyyymmdd) X PEEP Inview measurement 5 set PEEP X
PEEP Inview measurement 1 time (hhmm) X PEEP Inview measurement 5 FRC volume X
PEEP Inview measurement 1 set PEEP X PEEP Inview measurement 5 PEEPi X
PEEP Inview measurement 1 FRC volume X PEEP Inview measurement 5 PEEPe X
PEEP Inview measurement 1 PEEPi X PEEP Inview measurement 5 Cstat X
PEEP Inview measurement 1 PEEPe X PEEP Inview measurement 5 Delta FRC X
PEEP Inview measurement 1 Cstat X PEEP Inview measurement 5 Spiro
PEEP Inview measurement 1 Delta FRC X X
Dynamics Volume
PEEP Inview measurement 1 Spiro PEEP Inview measurement 5 Gain X
X
Dynamics Volume FRC status byte X
PEEP Inview measurement 1 Gain X
PEEP Inview measurement 2 date (yyyymmdd) X
PEEP Inview measurement 2 time (hhmm) X
PEEP Inview measurement 2 set PEEP X FRC data byte
PEEP Inview measurement 2 FRC volume X FIELD COM1.3 COM1.5 COM1.7
PEEP Inview measurement 2 PEEPi X FRC Inview on/off X X
PEEP Inview measurement 2 PEEPe X PEEP Inview on/off X X
PEEP Inview measurement 2 Cstat X
PEEP Inview measurement 2 Delta FRC X
PEEP Inview measurement 2 Spiro
X
Dynamics Volume
PEEP Inview measurement 2 Gain X
PEEP Inview measurement 3 date (yyyymmdd) X

‘ ‘ Not Supported, ‘X’ Supported, ‘X1’ Supported by Unity Network ID v7A (refer to product specific documentation for more information)

Information is subject to change without notice. This guide provides only


general information related to the data output capabilities of this GE
product. For complete specifications, along with communication protocol
information, please visit GE’s Open Interface Commitment website:
http://www3.gehealthcare.com/en/Products/Interfacing_Commitment

GE Healthcare
P.O. Box 7550 © 2014 General Electric Company – All rights reserved.

Madison, WI 53707-7550 General Electric Company reserves the right to make


U.S.A. changes in specifications and features shown herein,
or discontinue the product described at any time without
notice or obligation. Contact your GE Representative