R860 Data Output Guide - FINAL
R860 Data Output Guide - FINAL
CARESCAPE R860 ™
‘ ‘ Not Supported, ‘X’ Supported, ‘X1’ Supported by Unity Network ID v7A (refer to product specific documentation for more information)
Status data Alarm Status data bytes
FIELD COM1.3 COM1.5 COM1.7 FIELD COM1.3 COM1.5 COM1.7
set vent mode X1 X X Backup Buzzer POST X X X
set TV X1 X X Patient Connected X1 X X
set PEEP X1 X X No Battery Backup X X
set Tinsp X1 X X Temp High X X
set Pinsp X1 X X Primary Audio Failure X X X
set Psupp X1 X X Power Supply Fail X X
set Plimit X X X Exp Flow Sensor Error X X
set RR X1 X X Neo Flow Sensor Error X X
set mech RR X1 X X Neo Flow Sensor Reversed X X
set insp pause X X X Neo Flow Sensor Contaminated X X
set I:E (xxx.x:xxx.x) X1 X X Replace Neo Flow Sensor X X
set flow trigger X1 X X Neo Flow Sensor Off x X
set end flow X X X Negative Airway Pressure X1 X X
set bias flow X1 X X No Exp Flow Sensor X1 X X
set pressure rise time X X X No Neo Flow Sensor X X
set flow rise time X X Exp Flow Sensor Failure X X
set pressure trigger X X X O2 Sensor Failure X X X
set Pmax X X X Air Supply Pressure High X1 X X
set PSV rise time X X X Air Supply Pressure Low X X X
set flow X X O2 Supply Pressure High X1 X X
set vent assist control X X O2 Supply Pressure Low X X X
set pause time X X No Gas Supply Pressure X X X
set FiO2 X1 X X Air Supply Pressure Sensor Out of Range X X
set Phigh X X O2 Supply Pressure Sensor Out of Range x X
set Plow X X Pinsp Sensor Out of Range X X
set Thigh X X Pexp Sensor Out of Range X X
set Tlow X X Paux Sensor Out of Range X X
set Tsupp X X FiO2 Control Error X X
set high MVexp limit X1 X X Volume Delivery Error X X
set low MVexp limit X1 X X Air Temp Sensor Error X X
set high TVexp limit X1 X X O2 Temp Sensor Error X X
set low TVexp limit X1 X X Mixed Gas Temp Sensor Error X X
set high FiO2 limit X X X Total Flow Sensor Comm Fail X X
set low FiO2 limit X X X High Temperature X X X
set high RR limit X1 X X Module Not Compatible X X X
set low RR limit X X X Check Sample Gas Out X X X
set high EtCO2 limit X X X Replace D-Fend X X X
set low EtCO2 limit X X X Sample Line Blocked X X X
set high Ppeak limit X X X Check D-Fend X X X
set low Ppeak limit X X X Sample Flow Deviation X X
set high EtO2 limit X X MGAS Failure X X X
set low EtO2 limit X X No D-Lite Sensor X X
set high PEEPe limit X X Front Panel Comm Failure X X X
set low PEEPe limit X X Pressure Sensor Failure X X X
set high PEEPi limit X X Nebulizer Not Connected X X X
set high Paux limit X X On Battery X X X
set apnea delay time X1 X X Low Internal Battery - 20 min X X
set circuit leak limit X X Low Internal Battery - 10 min X X
set disconnect time X X Low Internal Battery - 5 min X X
set patient effort time X X Low Internal Battery - 1 min X X
set tube type X X X No Battery X X X
set tube size X X X Low Battery X1
set ARC level X X Battery Failure X
set patient type X X X Battery Charger Failure X
set BSA X X X Fan Fail X X X
set patient height X X X Relief Valve Failure X X
set patient weight X X X Backup Mode Active X X X
set weight units X X SBT Completed X X
set height units X X SBT Ended X X
set patient ID X X X Cannot Calculate FRC X X
set flow sensor type X X X Missed Scheduled FRC X x
set nebulizer time X X CPU Fan Speed Fail X X
set nebulizer volume X X X Case Fan Speed Fail X X
setting data status X X Ppeak High X1 X X
set system configuration X X Relief Valve Opened X X
set Pmin X X Ppeak Low X1 X X
set continuous nebuilzation X X Sustained Paw X1 X X
Breathing Circuit Occlusion X1 X X
Circuit Leak X1 X X
Status data byte Circuit Leak Off X X
Patient Connection Leak X X
FIELD COM1.3 COM1.5 COM1.7 Apnea X1 X X
new setting/10 sec data X X X Apnea Alarm Off X X
‘ ‘ Not Supported, ‘X’ Supported, ‘X1’ Supported by Unity Network ID v7A (refer to product specific documentation for more information)
Alarm Status data bytes (continued) System data
FIELD COM1.3 COM1.5 COM1.7 FIELD COM1.3 COM1.5 COM1.7
MVexp Low X1 X X current date (yyyymmdd) X X X
MVexp High X1 X X current time (hhmmss) X X X
MVexp Low Alarm Off X X system state X X
No Patient Effort X X
RR Low X1 X X
RR High X1 X X Procedures data
FiO2 Low X1 X X FIELD COM1.3 COM1.5 COM1.7
FiO2 High X1 X X
procedure status bytes X X
VTexp Low X1 X X
PEEPi X X
VTexp High X1 X X
PEEPi Volume X X
VT Not Achieved X1 X X
P0.1 X X X
EtO2 Low X X X
vital capacity X X
EtO2 High X X X
negative inspiratory force X X
EtCO2 Low X1 X X
Pplat X
EtCO2 High X1 X X
Static Compliance X
Patient Disconnected X1 X X
PEEPe High X X
PEEPe Low X X Procedures data bytes
PEEPi High X X
Paux High X X FIELD COM1.3 COM1.5 COM1.7
Plimit Reached X X X nebulizer on/off X X X
MGAS Warming Up - 2 min X X X pre-oxygenate on/off X X X
MGAS Warming Up - 5 min X X X suction on/off X X X
Alarm Light Fail X X post-oxygenate on/off X X X
Carrier Board Overheat X X intrinsic PEEP on/off X X
CPU Overheat X X inspiratory hold on/off X X
Touch screen fail X expiratory hold on/off X X
Service Cal Required X SBT on/off X
Standby On X oxygenation on/off X X
Alarms Silenced X X X P0.1 on/off X X
New Alarm/10 second data X X X negative inspiratory force on/off X X
vital capacity on/off X X
Waveform data
Data Transfer Configuration
FIELD COM1.3 COM1.5 COM1.7
1st waveform block X1 X X FIELD COM1.3 COM1.5 COM1.7
2nd waveform block X1 X X data to transfer X X
3rd waveform block X1 X X time period X X
4th waveform block X1 X X sample interval X X
5th waveform block X X X waveforms on/off X X
6th waveform block X X X patients X X
breath end/start index X1 X X transfer media X X
‘ ‘ Not Supported, ‘X’ Supported, ‘X1’ Supported by Unity Network ID v7A (refer to product specific documentation for more information)
Metabolics data FRC data
FIELD COM1.3 COM1.5 COM1.7 FIELD COM1.3 COM1.5 COM1.7
start date (yyyymmdd) X PEEP Inview measurement 3 time (hhmm) X
start time (hhmm) X PEEP Inview measurement 3 set PEEP X
end date (yyyymmdd) X PEEP Inview measurement 3 FRC volume X
end time (hhmm) X PEEP Inview measurement 3 PEEPi X
VO2 Average X PEEP Inview measurement 3 PEEPe X
VO2 Variance X PEEP Inview measurement 3 Cstat X
VCO2 Average X PEEP Inview measurement 3 Delta FRC X
VCO2 Variance X PEEP Inview measurement 3 Spiro
X
RQ X Dynamics Volume
EE X PEEP Inview measurement 3 Gain X
EE/kg X PEEP Inview measurement 4 date (yyyymmdd) X
EE/m2 X PEEP Inview measurement 4 time (hhmm) X
PEEP Inview measurement 4 set PEEP X
PEEP Inview measurement 4 FRC volume X
FRC data PEEP Inview measurement 4 PEEPi X
PEEP Inview measurement 4 PEEPe X
FIELD COM1.3 COM1.5 COM1.7
PEEP Inview measurement 4 Cstat X
procedure date (yyyymmdd) X PEEP Inview measurement 4 Delta FRC X
procedure time (hhmm) X PEEP Inview measurement 4 Spiro
FRC volume X X X X
Dynamics Volume
PEEPi X PEEP Inview measurement 4 Gain X
PEEPe X PEEP Inview measurement 5 date (yyyymmdd) X
Cstat X PEEP Inview measurement 5 time (hhmm) X
PEEP Inview measurement 1 date (yyyymmdd) X PEEP Inview measurement 5 set PEEP X
PEEP Inview measurement 1 time (hhmm) X PEEP Inview measurement 5 FRC volume X
PEEP Inview measurement 1 set PEEP X PEEP Inview measurement 5 PEEPi X
PEEP Inview measurement 1 FRC volume X PEEP Inview measurement 5 PEEPe X
PEEP Inview measurement 1 PEEPi X PEEP Inview measurement 5 Cstat X
PEEP Inview measurement 1 PEEPe X PEEP Inview measurement 5 Delta FRC X
PEEP Inview measurement 1 Cstat X PEEP Inview measurement 5 Spiro
PEEP Inview measurement 1 Delta FRC X X
Dynamics Volume
PEEP Inview measurement 1 Spiro PEEP Inview measurement 5 Gain X
X
Dynamics Volume FRC status byte X
PEEP Inview measurement 1 Gain X
PEEP Inview measurement 2 date (yyyymmdd) X
PEEP Inview measurement 2 time (hhmm) X
PEEP Inview measurement 2 set PEEP X FRC data byte
PEEP Inview measurement 2 FRC volume X FIELD COM1.3 COM1.5 COM1.7
PEEP Inview measurement 2 PEEPi X FRC Inview on/off X X
PEEP Inview measurement 2 PEEPe X PEEP Inview on/off X X
PEEP Inview measurement 2 Cstat X
PEEP Inview measurement 2 Delta FRC X
PEEP Inview measurement 2 Spiro
X
Dynamics Volume
PEEP Inview measurement 2 Gain X
PEEP Inview measurement 3 date (yyyymmdd) X
‘ ‘ Not Supported, ‘X’ Supported, ‘X1’ Supported by Unity Network ID v7A (refer to product specific documentation for more information)