Official-Manual EVOSM7000
Official-Manual EVOSM7000
DISCLAIMER
TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL,
INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS
DOCUMENT, INCLUDING YOUR USE OF IT.
Trademarks
All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.
Cy is a registered trademark of GE Healthcare UK Limited. DRAQ5 is a registered trademark of Biostatus Limited. Windows is a
registered trademark of Microsoft Corporation. Kimwipes is a registered trademark of Kimberly-Clark Corporation. Hoechst is a
registered trademark of Hoechst GmbH. Swagelok is a registered trademark of Swagelok Company.
Audience This user guide is for laboratory staff operating, maintaining, and analyzing data
using the Invitrogen™ EVOS™ M7000 Imaging System.
User attention Two user attention words appear in this document. Each word implies a specific
words level of observation or action as described below.
Note: Provides information that may be of interest or help but is not critical to the
use of the product.
Safety alert words Three safety alert words appear in this document at points where you need to be
aware of relevant hazards. Each alert word—CAUTION, WARNING, DANGER—
implies a particular level of observation or action, as defined below:
Product description
EVOS™ M7000 The Invitrogen™ EVOS™ M7000 Imaging System (Cat. No. AMF7000) is a fully
Imaging System automated, digital, inverted multi-channel fluorescence and transmitted light
imaging system. The system is designed for a broad range of applications including,
but not limited to, multi-channel fluorescence imaging, multiple-position vessel
scanning, area scanning with montage or tile stitching, and time-lapse imaging.
EVOS™ M7000 The EVOS™ M7000 Imaging System is controlled by the integrated Invitrogen™
Software EVOS™ M7000 Software accessed by the touchscreen monitor or the computer
mouse and keyboard. The software comes pre-installed with the software shortcut
placed on the desktop and under StartAll Programs ThermoM7000.
Key features of the EVOS™ M7000 Software include:
• Capture: Allows control over every aspect of the system for image capture
through a simple user interface. All images acquired can be saved in TIFF,
PNG, C01, DIB, and JPEG formats, or compiled into a video sequence in AVI or
WMV formats.
• Automate: Allows the creation, saving, and running of user-defined routines to
automate image collection.
• Autofocus: Can be set up in five different modes to optimize speed and
accuracy.
• Tiled images and Image stitching: Allow the scanning of an area to acquire
multiple images to build tiled and stitched images. The Review tool allows
zooming in/out and panning of the composite image. The entire scan or only
regions of interest can be exported.
• Z-stacking: Captures a series of images along the z-axis that can be saved
individually or combined into a Z-stack projection with a greater depth of field
than any of the individual source images.
• Time lapse: Creates and runs time lapse movies based on user specifications.
• Review: Allows you to review captured images and analyze them with the
EVOS™ Analysis application that is included in the EVOS™ M7000 Software
package (page 6).
Note: For a detailed description of the EVOS™ M7000 Software controls, see
“Appendix C: Graphical user interface (GUI)” (page 126).
Note: For a detailed description of the EVOS™ Analysis application controls, see
“EVOS™ Analysis Application” in “Appendix C: Graphical user interface (GUI)”
(page 215).
Note: If you do not have your distributor information, contact Technical Support
(page 247).
IMPORTANT! Wiping the computer supplied with the EVOS™ M7000 Imaging
System (i.e., erasing the hard drive to remove all programs, files, and the
operating system) voids the product warranty. Do not install third party software
or update the Windows™ operating system.
Condenser
Automatic X-Y axis stage
Handholds
Rear view
Power switch
4-pin power input port (24 VDC, 5 A)
USB 3.1 Type B port
GUI layout The GUI of the system consists of the Viewing area on the left and a series of tabs
representing the main functions of the software (Capture, Automate, Review, and
Settings) on the right. Each tab contains the controls necessary to execute the
selected function. The Locations button, Area/Field View toggle, and the Zoom
slider are located above the Viewing area.
Locations: Opens the Locations controls, which allows you to select location for
automated scan protocols.
Area View/Field View toggle: Switches between Area and Field Views.
Zoom slider: Zooms in and out of the Viewing area.
Capture tab: Contains the controls for the manual capture of images.
Automate tab: Allows you to create and run automated scan protocols and time
lapse experiments.
Review tab: Allows you to review and annotate captured images.
Settings: Contains controls to select and adjust basic and advanced system
options and instrument functions.
Viewing area: Displays the sample in Area View or Field View modes.
Note: Click the View Help Content icon to open the Help window, which
provides EVOS™ M7000 Imaging System Help.
Position your pointer over the Tooltip icon without clicking it to view
additional information.
Note: For more information and detailed descriptions of software controls, see
“Appendix C: Graphical user interface (GUI)”, page 126.
IMPORTANT! Do not perform the initial installation of the EVOS™ M7000 Imaging
System yourself. A Thermo Fisher Scientific representative will contact you to
schedule the installation.
When the installation is scheduled:
1. Receive and inspect the system.
2. Move the crated instrument to the installation site.
IMPORTANT! Do not position the instrument so that it is difficult to turn off the
main power switch located on the back of the instrument base (page 9). In case of an
instrument malfunction, turn the main power switch to the OFF position and
disconnect the instrument from the wall outlet.
IMPORTANT! Do not lift the EVOS™ M7000 Imaging System by stage or condenser
arm. Lift the instrument by using the handholds in the base.
IMPORTANT! Do not subject the EVOS™ M7000 Imaging System to sudden impact
or excessive vibration. Handle the instrument with care to prevent damage.
Unpack and 1. Open the case and remove the monitor and accessories.
connect the 2. If a VGA cable is attached to the monitor, take it off.
monitor 3. Remove protective covering from monitor.
4. Plug the USB touch screen cable into the USB port on the monitor.
5. Plug power cord into monitor.
IMPORTANT! Wiping the computer supplied with the EVOS™ M7000 Imaging
System (i.e., erasing the hard drive to remove all programs, files, and the
operating system) voids the product warranty.
Unpack and connect 1. Open the box and remove the accessory box.
the instrument 2. Carefully lift the instrument out of the box, holding it by two of the four
handholds in the base (page 9).
IMPORTANT! Do not lift the EVOS™ M7000 Imaging System by stage or
condenser arm. Lift the instrument by using the handholds in the base.
3. Place the instrument on a flat, level surface that will be free from vibration
and leave enough room around it for the stage to move freely.
4. Remove the following from the accessory box (located in the instrument box):
• Power cable (for North America and as ordered)
• Power supply
• Display Port to Display Port cable
• Mini-Display Port adapter
• USB 3.1 type A to B cable
• White cardboard box (contains the light box and vessel holders)
• Dust cover
5. Confirm that the power switch is OFF (located on the back; page 9).
Note: At this point, everything should be plugged in and OFF. Save the packaging
for future shipping/storage of the instrument.
Remove shipping The EVOS™ M7000 Imaging System is equipped with three shipping restraints
restraints (X-Y stage, light cube, and camera shipping restraints), which prevent damage to
the instrument from shock and vibration during transport. The shipping restraints
must be removed before the EVOS™ M7000 Imaging System is powered on.
1. Unscrew the three thumb screws on the X-Y stage shipping restraint until they
spin freely. You do not need to remove them from the restraint altogether.
2. Gently pull the X-Y stage shipping restraint forward, away from the unit.
4. Unscrew the camera shipping restraint and pull it up to remove it from the
X-Y stage.
Note: Store the shipping restraints and the light cube tool in the accessories box
for future use. To reinstall the shipping restraints, page 118.
IMPORTANT! Do not subject the EVOS™ M7000 Imaging System to sudden impact
or excessive vibration. Handle the instrument with care to prevent damage.
Overview
Capture tab The basic functions of the EVOS™ M7000 Imaging System, such as viewing the
sample, setting optimal focus, and capturing and saving images are performed in
the Capture tab, which is the first screen after start-up.
Note: For a detailed description of the Capture tab controls, see “Capture tab” in
“Appendix C: Graphical user interface (GUI)” (page 139).
5. Click Done to complete your selection and close the dialog. The Vessel map
on the Capture tab displays your selected vessel.
By default, the first well on the sample vessel is selected and indicated in blue
on the Vessel map. The Viewing area displays the selected well.
6. To select another well, click on the desired well on the Vessel map.
2. Select the desired Light source from the available options for which you want
to adjust brightness and set focus. In the example below, the DAPI channel
has been selected.
Note: You can select only one light source at a time to adjust brightness and
set focus. However, you can display and capture multiple channels
simultaneously (see “Capture multiple channels simultaneously”, page 32).
Set brightness and 1. Click the Brightness and camera settings button to expand the controls for
camera options setting mode, camera, and phase options.
Note: Phase options are available only for the transmitted light channel; they
are not available for fluorescence channels. The phase contrast option does
not require phase contrast objectives. However, to obtain a phase image, you
must first install a phase contrast objective.
5. Click the Brightness and camera settings button again to collapse the
controls.
Note: The controls for image display settings are contextual; only the controls
for the selected channels will be available. In the example above, only the
controls for the selected DAPI, RFP, and transmitted light (brightfield)
channels are displayed.
3. Optional: To remove a channel from displaying in the Viewing Area, unselect
the corresponding Display checkbox. To display an available channel not
shown in the Viewing Area, re-select the checkbox.
4. Adjust the Brightness , Contrast , and Gamma settings for each of
the selected channels using the corresponding sliders.
5. Click the Image color display button to display the sample pseudo-
colored in the default emission color of the selected channels. By default, color
display option is on.
6. Click the Image display settings button again to collapse the controls.
Note: Adjustments made to Image display settings only affect how the image is
displayed in the Capture tab; they do not change how the image is captured .
The size the scale bar represents depends on the magnification of the selected
objective and it is displayed below the scale bar. After an image captured, the
size of the scale bar automatically changes as you zoom on a captured image
Note: The scale bar displayed in the Capture tab is not saved with your captured
images. However, you can add a scale bar to your saved images with the EVOS™
Analysis application (page 75).
Optional: Display 1. Click the Grid button to superimpose a grid over the Viewing area.
grid
Note: To include a grid when you save your captured images, check the Include
Grid option in the Save window and select the grid size (page 42). You can also
add a grid to your saved images with the EVOS™ Analysis application (page 75).
Note: If you run the autofocus procedure with the light off, the autofocused
image will be captured and stored in the image cache for that channel.
If you run the autofocus procedure with the light on (i.e., with the Light
button pressed), the instrument will simply find the optimal focus, but will
not capture the autofocused image.
Note: To learn more about the autofocus strategy, click Advanced focus settings
(the gear icon next to the AutoFocus button in the Capture tab), then click the
View help content icon to open the Help window. In the Help window, click
How do I set Software Autofocus?
2. Select the Objective and Light source (i.e., channel) you want to capture.
Note: To learn more about locked and unlocked Z-Offset states and how to set the
correct Z-Offsets, click Advanced focus settings (the gear icon next to the
AutoFocus button in the Capture tab), then click the View help content icon to
open the Help window. In the Help window, click How do I set Z-Offsets?
4. Focus on the sample manually using the Coarse focus and Fine focus sliders
or automatically by clicking the AutoFocus button.
You can also double-click on the handle of the focus slider to activate the text
box and enter the desired value for the focus position.
5. After you have found the optimal focus position in the first channel, click the
Advanced focus settings button, then uncheck the Lock Z-Offsets option.
6. Without changing the objective, select the next Light source (i.e., channel) you
want to capture.
7. If the light is off, click the Light button to turn it on, then focus on the sample
manually using the Coarse focus and Fine focus sliders or automatically by
clicking AutoFocus.
8. Repeat this procedure for every channel you want to capture.
9. When you have found the optimal focus position in all the additional channels
you want to capture, click the Advanced focus settings button, then check the
Lock Z-Offsets option.
When you have completed this procedure, the focus position in each selected
channel will be offset relative to each other.
5. Click the Light button again to exit the Live mode in Field View.
9. Click the Light source button to select the corresponding channel. If needed,
readjust brightness and focus.
10. Position the capture crosshair over the new field of interest, click to select the
new field, then click Capture.
In the example below, a third field located was captured in the DAPI channel
using the 4X objective.
11. To save your captured images, click Save…. For more information, see “Save”
(page 40).
3. Adjust brightness and focus for each of the selected channels as described
before.
4. Position the capture crosshair over the new region of interest, and click to
select the new field. Alternatively, navigate to the region of interest you want
to capture in the Live mode.
5. Click Capture Channels. The instrument captures an image
of the selected field in each of the selected channels using the current capture
settings.
A thumbnail of the captured image is displayed above the Light source button
for each channel in which it was captured.
The Viewing area displays a multicolor overlay of the images captured in the
selected channels at the location of their capture.
In the example below, the selected field was captured in the DAPI and RFP
channels using the 4X objective, and the Viewing area displays the merged
image captured in these channels.
6. To save your captured images, click Save…. For more information, see “Save”
(page 40).
Note: You can capture z-stack images in only one channel at a time; however,
you can save the z-stacks images captured individually in separate channels
into a single merged video (see “Save Z-Stack as a video”, page 38).
2. Position the capture crosshair over the region of interest, then click to select the
new field. Alternatively, navigate to the region of interest in the Live mode.
3. Adjust brightness and focus using the corresponding sliders on the Capture
tab. The focus position you set in this step is the Default Focus position.
4. Click Capture Z-Stack to open the Z-Stack Settings window. You can move
the Z-Stack Settings window anywhere on the screen.
12. The Area View displays the z-stack projection at the location of its capture.
The z-stack icon on the top left corner of the captured field identifies it as a
z-stack projection.
13. To view the individual planes that make up the z-stack image set,
double-click the field that shows the z-stack projection in Area View.
Alternatively, click the Field View icon. The Field View displays the z-planes
and the z-stack controls appear to the left of the Viewing area.
15. To view the the z-stack projection, click the Toggle Projection Image
button.
The Viewing area displays the z-stack projection, which is generated by
combining the most in-focus pixels from each z-plane. By default, the
projection image is on.
16. To view the z-planes in sequence as a movie, click the Start z-stack
playback button located below the z-stack slider.
The Viewing area displays the z-planes in sequence, starting at the
bottom slice (Z-plane: 1) and stopping at the top.
17. To play the z-stack movie in a loop, click the Repeat playback button.
The Viewing area displays the z-planes in a continuous loop, starting at
the bottom slice and looping back to beginning after displaying the top
slide.
• To change the playback speed, move the Playback speed slider up
(faster) or down (slower). The playback speed is expressed in
frames per second (FPS).
Note: During playback, z-planes that make up the z-stack are shown
sequentially in the channel in which they were captured. The projection image
is not displayed (i.e., the Toggle Projection Image is unselected).
19. To capture z-stack images in another channel, select the Light source (i.e.,
channel) you want to capture, then repeat the procedure as described.
20. To save the z-stack movie as a separate file, see “Save Z-Stack as a video”
(page 38).
21. To save the z-stack projection and individual z-planes for each field and each
channel as separate images, click Save… to open the Save dialog. For more
information, see “Save”, page 40.
2. Click Save as Video icon to save the z-stack images as a video sequence.
The Save Z-Stack Videos dialog opens.
3. Click Browse, navigate to the folder in which you want to save your z-stack
video, then click Select.
4. If desired, enter a new name for the z-stack video file in the File name textbox.
Note: When saving the z-stack movie as a separate video file, only the
z-planes that fall within the z-stack calipers are included in the movie.
To change the z-planes you want to include in the z-stack movie, click Cancel
and readjust the z-stack calipers as described on page 37.
Note: The Frame rate of your z-stack video sequence is set using the Playback
speed slider. To change the Frame rate for your z-stack movie, click Cancel
and readjust the Playback speed slider as described on page 37.
9. Click Save to save the z-stack images as a video sequence based on your
specifications.
2. Select the captured fields you want save. Available options are:
• Currently selected field: Saves images only from the currently selected field.
• All newly captured fields: Saves images that have been captured and
stored in the image cache, but not yet saved. This option is available only if
you have previously saved images from the same session.
• All captured fields: Saves images from all captured fields that are held in
the image cache. This is typically all the images captured in an imaging
session.
Note: If the software remains open while users access the instrument back to
back, there may be several sets of images stored in the image cache. To clear
these images from the cache, click the trash can icon (see “Delete image
options”, page 130).
Saving the same images in the cache again does not overwrite the existing
files in the destination directory; newly saved files simply have a
differentiator added to the name.
To create a new folder in which to save your captured images, navigate to the
desired location, click New, type in the name of the newly created folder, then
click Select.
4. If desired, type the prefix you want to use for your save images in the File
name prefix textbox.
Note: For an overview of the file naming convention for saved images, see
“File naming convention”, page 191.
Available Color options for displayed images are Grayscale (16-bit) and
Pseudocolor (24-bit RGB; 8-bit per RGB channel).
Available File format for displayed images are: TIFF, PNG, and JPEG.
If desired, check the Include Grid option and select the grid size.
Note: We recommend saving captured images in both Raw and Displayed image
formats to preserve the option of using the Raw images in downstream image
analysis and quantitation where 16-bit dynamic range is required, and the
Displayed images for instances where “prettier” images are required. For more
information on Raw and Displayed images, see “Raw vs. Displayed images”,
page 125.
7. Select the desired file types and save options by checking the corresponding
checkboxes. You can select multiple file types and options.
Available file type options are Raw images and Displayed images.
• Raw images:Saves each channel individually as 16-bit Raw images. Raw
image files contain the full dynamic range and metadata needed for
quantitative analysis and are the recommended format for image analysis.
- Single field, Z-Stack planes, and Tiled images can be saved as Raw
images with each channel saved individually.
- Available File format for Raw images are TIFF, PNG, C01, and DIB.
• Displayed images: Saves each channel individually or as a merged image
in a format that can be viewed in most image display applications.
- Single field and Tiled images can be saved as Displayed images from
individual channels or as a merged view.
However, Z-planes can be saved as Displayed images only from
individual channels.
- Available Color options for displayed images are Grayscale (16-bit) or
Pseudocolor (24-bit RGB; 8-bit per RGB channel).
- Available File format for displayed images are TIFF, PNG, and JPEG.
8. After you selected save options for your captured images, click Save.
Note: For more information on each save option available for Raw and Displayed
images, see the detailed descriptions in “Appendix C: Graphical user interface”
under Capture tabSave (page 154).
Note: For more information about Tiled and Merged images, see “Tiled, Stitched,
and Merged images”, page 125.
Overview
Automate tab The EVOS™ M7000 Imaging System allows you to create and recall scan protocols,
which are a series of automatically executed steps to capture multiple images over
an area and/or time period based on your specifications. For repeat experiments,
automated scan protocols can be saved, recalled, even edited.
The scan protocols are created using the Automate tab, which is organized into a
series of panels that contain the necessary controls that organized by functionality
(i.e., Load, Hardware, Scan Area, AutoFocus and Z Stacks, Time Lapse and
Incubator, Image Save Settings, Save, and Run).
You can access the Capture, Review, and Settings at any time by selecting the
corresponding tab.
Note: For a detailed description of Automate tab controls, see “Automate tab” in
“Appendix C: Graphical user interface (GUI)” (page 157).
Note: If you are using the EVOS™ Onstage Incubator, set it up for operation
as described in “Appendix D: EVOS™ Onstage Incubator”, page 228.
3. In the Automate tab, click Hardware Edit to open the Hardware panel.
4. Select the Objective you want to use for the scan protocol. You can select only
one objective at a time. In the example below, 10X objective is selected.
8. Click the Light button to turn on the excitation light for the currently selected
channel (i.e., Light source) and enter the Live mode.
9. While in the Live mode, adjust the brightness settings using the Brightness
controls as described in “Adjust brightness”, page 22.
10. If in Simple mode, adjust the brightness by moving the Brightness slider.
11. While still in the Live mode, select each of the remaining channels that will be
used for the automated scan and adjust the brightness settings.
12. When finished with the channel settings, click the Automate tab to return to
the Hardware panel.
13. Click Done to save hardware options, then proceed to “Select scan areas”,
page 47.
Note: For a detailed description of the controls available in the Scan Area
panel, see “Scan Area” in “Automate tab”, page 159.
To define the areas and fields you want to scan, you have two options:
• Scan Locations tool (page 161) allows you to manually assign
scan regions using the Locations tools (page 134).
To use the Scan Locations tool to select regions for automated
scan, go to Step 2 (page 48).
• Scan Pattern tool (page 166) allows you to define the capture
regions for the scan protocol as a pattern based on defined
parameters.
To use the Scan Pattern tool to select regions for automated scan,
go to Step 8 (page 51).
• If you have already defined locations using the Locations tools, they will
be listed in the Available Locations table.
In the following example, the first location in the list (an ellipse) is selected.
The Viewing are displays the selected location and the field group that best
fits the shape of the location.
5. To add a location to the scan protocol, click the desired location in the
Available Locations table to select, then click Add Selected.
To add multiple locations to the scan protocol simultaneously, Shift-click the
desired locations in the Available Locations table to select them, then click
Add Selected.
To add all the locations listed in the Available Locations table, click Add All.
The locations added to the scan protocol are displayed in the Scan Locations
table and removed from the Available Locations table.
You have the option to individually define scan locations for scan area you
want to capture or select the scan areas on the Vessel map to which the
selected scan locations are applied.
For instructions on how to define scan areas on the Vessel map, see page 165.
7. When finished, click Done to close the Vessel map and exit the area selection
mode.
Note: The region you define with the Field selection tools in Area View is
replaced by a set of capture fields that best fit the drawn shape or, in the
case of the Pattern tool, satisfy the criteria set for the region (see examples
on pages 166−167).
In a multi-well plate, multi-chamber slide, or multi-vessel holder, the fields
defined in a selected well, chamber, or vessel holder also apply to the other
selected wells, chambers, or vessel holders.
10. If needed, click the Delete field group button that is displayed when
you Position over the field group to delete that field group.
To delete all field groups from Area View, click the Clear field
selections button located at the bottom right corner of Area View.
12. When finished defining the scan areas, click Done to exit the selection mode.
16. Select Default Overlap for a faster scan or select More Overlap for a higher
image quality.
Stitching off
17. When finished with area selection, click Done to save your selections and
return to the Automate tab.
Note: To learn more about locked and unlocked Z-Offset states and how to set the
correct Z-Offsets, go to Automate tab, click the Edit button to open the AutoFocus
and Z-Stacks panel, then click the View Help Content icon to open the Help
window. In the Help window, click How do I set Z-Offsets?
For best results, optimize the Z-Offsets in the Capture tab, and then carry the
offsets over to the Automate tab.
1. In the Automate tabAutoFocus and Z Stacks panel, ensure that the Capture
Z-Stacks option is unchecked, then click the Capture tab.
Note: The Z-Offsets option is not available as one of the Focus Position
controls when the Capture Z-Stacks option has been selected. For more
information, see “Z-Stack Settings”, page 173.
2. In the Capture tab, click Advanced focus settings and verify that the
Lock Z-Offsets option is selected. By default, Lock Z-Offsets option is checked.
3. Select the Objective and Light source (i.e., channel) you want to capture.
4. Click the Light button to enter the Live mode (page 143).
6. After you have found the optimal focus position in the first channel, click
Advanced focus settings, then uncheck the Lock Z-Offsets option.
7. Without changing the objective, select the next Light source (i.e., channel) you
want to capture.
8. If the light is off, click the Light button to enter the Live mode, then focus on
the sample manually using the Coarse focus and Fine focus sliders.
Alternatively, click AutoFocus to automatically focus on the sample.
11. If the light is on, click the Light button to turn off the Light and exit the Live
mode.
Because the Z-Offsets are preserved, selecting the RFP channel for the default
focus position at 138.85 µm refocuses the instrument to 118.8 µm in the DAPI
channel (138.85 µm – 20.05 µm).
Note: If you need to adjust the Z-Offsets while in the Automate tab, make sure to
close out the Autofocus and Z Stacks panel by clicking the Done button before
navigating back to the Capture tab to make the adjustments.
Note: The options available on the AutoFocus and Z Stacks panel are
contextual; the controls that are displayed depend on the selections made in
this panel or elsewhere in the Automate tab. For more information, see
“AutoFocus and Z Stacks” in “Automate tab”, page 172.
3. Select Single Channel or All Channels to autofocus during the scan protocol.
Note: The effectiveness of the autofocus method depends on the plate type,
sample type, biomarker intensity, magnification, and channel. It may be
necessary to use different autofocus methods for different channels.
Note: For a single culture dish or flask, the scan area corresponds to the
entire dish or flask. For a multi-well plate or a multi-chamber slide, each scan
area corresponds to a single well or chamber.
Specifying fewer fields per area to autofocus reduces the overall scan times.
Frequent autofocusing may not be necessary for fields that are close together
or for samples with consistent z-locations on very flat vessel surfaces.
8. When finished with your selections, click Done to return to the Automate tab
and proceed to “Set Time Lapse options”, page 60.
Note: To learn more about the autofocus strategy, go to Automate tab, click the
Edit button to open the AutoFocus and Z-Stacks panel, then click the View Help
Content icon to open the Help window. In the Help window, click How do I
set Software Autofocus?
Note: The capture parameters defined in the Automate tab (z-stacks, exposures,
channels, etc.) are applied to every time point of a time lapse routine.
1. In the Automate tab Time Lapse and Incubator, click the Edit button to
open the Time Lapse and Incubator panel, then select Use Time Lapse.
2. Enter the Duration of the run into the corresponding Hours, Minutes, and
Seconds fields.
Note: The runs are performed in chronological order, beginning with Run 1
and ending with Run N. You can remove a run from the time lapse routine
by clicking the corresponding Delete Run button (page 184).
Note: Use Time Lapse option must be previously checked for the Use
Incubator option to be available.
2. Click Browse, navigate to the folder in which you want to save your images
captured during the automated scan protocol, then click Select.
4. If desired, type the prefix you want to use for your saved images in the Prefix
textbox.
Note: For file naming conventions for saved images, see “File naming
convention”, page 191.
5. Select File types to save by checking the corresponding checkboxes. You can
choose multiple file types for your images captured during the automated
scan protocol.
Available file type options are Raw images and Displayed images.
• Raw images: Saves images captured in different channels individually as
16-bit Raw images. Raw images contain minimally processed data from the
image sensor and it is the recommended format for image analysis and
quantitation.
Note: For more information about when to use tiled or stitched images, see
“Tiled, Stitched, and Merged images”, page 125.
6. Check the Include Grid option to superimpose a grid on the displayed images.
You have the following options for the grid size (in pixels):
Auto, 10 × 10, 50 × 50, 100 × 100, 200 × 200, and 500 × 500.
7. Select Tiled per location or Tile all locations together.
9. Select the desired file types and options by checking the corresponding
checkboxes. You can select multiple file types and options.
Available file type options are Raw images and Displayed images.
• Raw images: Saves each channel individually. This is the recommended
format for image analysis and quantitation.
- Single fields, Z-planes, and Tiled images can be saved as Raw
images.
- Available File format for Raw images are TIFF, PNG, C01, and DIB.
• Displayed images: Saves each channel individually or as a merged image
in a format that can be viewed in most image display applications.
- Single fields, Z-planes, and Tiled images from individual channels
can be saved as Displayed images.
However, you can only save Single fields and Tiled images as a
Displayed image in merged view; Z-planes can only be saved from
individual channels.
- Available Color options for Displayed images are Grayscale (16-bit) or
Pseudocolor (24-bit RGB; 8-bit per RGB channel).
- Available Brightness options for Displayed images are High and Full.
- Available File format for Displayed images are TIFF, PNG, and JPEG.
10. After you have configured Image Save Settings for your automated scan
protocol, click Done to return to the Automate tab.
At this point, we recommend saving the automated scan protocol for future recall
in repeat experiments. For convenience, you can also edit recalled scan protocols
to create new routines rather than creating them from the very beginning.
1. In the Automate tab, click Save to open the Save As dialog, then navigate to
the location in which you want to save your scan protocol.
2. In the File name text box, type in the name of your new routine.
The default name for new scan protocols is “New Protocol.scanprotocol”.
3. Click Save to save your scan protocol and close the Save As dialog.
4. To run your automated scan protocol, proceed to “Run newly created scan
protocol”, page 69.
The Scan Settings, Run Sequence Settings, and Run Settings specified in your
scan protocol are displayed below the Vessel map, and the Experiment
Progress bar tracks the progress of the run.
2. Click Pause to pause the automated scan protocol anytime during a run.
Click Resume to restart the scan protocol from the step where it was paused.
3. If you want to abort the scan protocol before it is completed, click Stop.
Click Save to save the images acquired during the aborted scan protocol in the
location specified in Image Save Settings, or click Discard to delete them.
Load and run a For repeat experiments, you can recall and run a saved scan protocol. If desired,
saved scan protocol you can also edit the settings of the recalled scan protocol.
1. On the Automate tab, click Load, then navigate to the folder containing the
saved scan protocol you want to recall.
2. Click the saved scan protocol of interest to select it, then click Open.
3. If you want to edit the recalled scan protocol, click Review to open the
Automate tab panel, adjust the settings you want to change, then click Done
to return to the Automate tab.
4. Click Run to run the recalled routine.
5. When the scan protocol is completed, click Done.
Note: Saved scan protocols retain the information about the scan areas as well as
the camera and lighting options set when creating the protocol. To change these
settings for a new experiment, click Review after loading the saved scan protocol
and make the desired changes.
Review tab The Review tab allows you to review saved images, including Z-Stack images and
images captured during automated scans, and scan and field metadata associated with
the images. It also allows you to launch the EVOS™ Analysis application to analyze
and annotate saved images (see “Analyze and annotate saved images”, page 75).
For the descriptions of the controls available in the Review tab, see page 192.
Note: For a detailed description of the Review tab controls, see “Review tab” in
“Appendix C: Graphical user interface (GUI)” (page 192).
The Vessel map shows the location of the imaged field in the original sample vessel.
Note: The Open in EVOS Analysis button becomes available in the Review
tab only after you have opened the image you wish to analyze in the Viewing
area.
Note: The Open in Celleste button is available only if you have the Celleste™
Image Analysis Software (available separately from Thermo Fisher Scientific)
installed on the computer running the EVOS™ M7000 Software. The button
becomes visible in the Review tab after you have opened the image you wish
to analyze.
Celleste™ Image The Celleste™ Image Analysis Software (available separately from Thermo Fisher
Analysis Software Scientific) is a full-feature image analysis suite designed for a range of biological
applications, from image adjustments and processing with manual and automatic
measurements over multiple channels, to segmentation and classification tools that
help you transform images into quantitative data in a streamlined and
customizable workflow.
For more information on the Celleste™ Image Analysis Software, go to
thermofisher.com/celleste. For instructions on how to analyze your images with the
Celleste™ Image Analysis Software, refer to the Celleste™ Image Analysis Software
user guide (Pub. No. MAN0018003), availabel for download at thermofisher.com.
Note: For a detailed description of the EVOS™ Analysis application controls, see
“EVOS™ Analysis Application” in “Appendix C: Graphical user interface (GUI)”
(page 215).
Launch the There are two ways to launch the EVOS™ Analysis application:
EVOS™ Analysis • Launch the application from the Review tab when there is a saved image file
application open in the Review tab (page 76).
• Launch the application directly from your desktop, then open the image you
wish to analyze using the application (page 77).
Note: The Open in EVOS Analysis button becomes available only after you
have opened the image you wish to analyze in the Viewing area.
2. Hover the pointer over the Viewing area to reveal Display Settings and
Analysis Tools, then click to open the desired tool.
3. To analyze another image, click on the image in the you wish to view it in the
Viewing area.
The contents of the selected folder are displayed in the File list/Image preview
panel.
3. To change the layout from File list to Image preview or to sort the files
by name, file type, or creation date, click the Display Settings button,
then make your selections:
• To toggle between file list and images preview formats, use the
Thumbnails and List buttons.
• To change the size of the image thumbnails, use the Zoom buttons.
• To sort the image files by Name, File Type, or Date Created, use the
Sort controls.
5. Hover the pointer over the Viewing area to reveal the buttons for Display
Settings and Analysis Tools.
6. Click a button to open the corresponding tool; click the button again to close it.
Note: The controls for image display settings are available only for captured
channels. In the example above, the controls are available for the TGBFP
(TagBFP), GFP, RFP, CY5, and TRANS (transmitted light or brightfield)
channels. The image that is displayed in the Viewing area is captured in the
TGBFP channel, as indicated by the check mark.
2. To change the grid size, click the Grid Settings button (arrow on the
Grid split button) to open the Grid Settings tool.
3. Select the Size for the grid. Available grid sizes depend on the magnification of
the selected objective.
4. Click the Grid Settings button again to save your settings and close the tool.
2. To move the scale bar, hover your pointer over the scale bar until a bounding
box appears, then click within the box and drag the scale bar to the desired
location within the Viewing area.
3. To adjust the length of the scale bar, hover your pointer over the scale bar until
a bounding box appears, then the click left or right side of the box and drag the
box to the desired length.
You can adjust the length by pre-fixed increments based on the objective
magnification.
4. To change scale bar settings, click the Scale Bar Settings button
(arrow on the Scale Bar split button) to open the Grid Settings tool.
5. Select Show End Bars to display the scale bar with the end bars.
6. Select the Color for the scale bar.
7. Click Scale Bar Settings button again to save your settings and close the tool.
2. Pixel Intensity histogram shows the Pixel count vs. Intensity data of the image
displayed in the Viewing area as well as the minimum, mean, and maximum
pixel intensities.
3. To move the histogram, click within the plot heading area and drag the plot to
the desired location.
4. To resize the histogram, click the grey triange at the lower right corner of the
plot, then drag the plot to the desired size.
5. Click the Histogram button again to close the Pixel Intensity histogram.
Alternatively, click the X on the plot to close the histogram.
3. If needed, change the Color and Thickness of the annotation to make it more
visible over the image.
5. To delete a selected annotation, click the X on the shape that appears when you
hover your pointer over it.
To delete all annotations, click Reset, then click OK in the dialog that opens.
6. Click the Measurement and Annotations button again to close the tools.
7. After you have added measurements and annotations to your image:
• Click Show Measurements and Annotations button (main part of the split
button) to turn the display on and off.
• Click Measurement and Annotations Controls button (the arrow on split
button) to display the controls to add new measurement and annotations
or to delete existing ones.
• Auto Count: Automatically counts the objects displayed in the Viewing area
based on your specifications (page 86). With Auto Count, you can count objects
only in a single fluorescence channel (nuclear stain channel).
• Manual Count: Allows you to tag objects in the Viewing area with up to six labels.
As you tag objects, the system keeps a running tally of the counts with percentages
for each label assigned (page 90). With Manual Count, you can count objects in
multiple channels simultaneously.
• Confluence: Allows you to select up to five reference objects for the target (i.e.,
cells) and one background reference in your image, then automatically
calculates the percentage confluence of your culture (page 92).
• Transfection Efficiency: Allows you to estimate the transfection efficiency of your
culture by calculating the ratio of fluorescence area (i.e., cells expressing the
fluorescence marker) to the total cell area in your culture (page 95).
• Batch Analysis: Allows you to save and apply the analysis parameters set in
Auto Count, Confluence, and Transfection Efficiency tools to other images that
you have collected and saved an image folder (page 100). Batch Analysis is not
available for Manual Count.
IMPORTANT! For analysis, only use 16-bit Raw image files, not Display or Merged
image files. Raw image files contain the full dynamic range and metadata needed
for quantitative analysis, whereas Display and Merged image files do not.
3. To identify the target objects to include in your count, click Target, then click
and drag to draw a circle (blue) around a nucleus.
Note: For best results, follow these guidelines when identifying target objects:
- When selecting objects, circle the entire object and include a slight border
around it.
- To include objects of lower intensity in your count, select dimmer objects
during identification.
- Circle only one object at a time to help define object size for segmentation.
5. To distinguish the target from background, click Backround, then click and
drag to draw a circle (orange) in a background area.
6. After you define the target and background areas, the software automatically
counts the objects based on your criteria.
9. To refine your count, select Intensity, Area, or Circularity, move the gate
handles to set the upper or the lower boundary for the selected parameter.
You can refine the count by a single parameter or by multiple parameters.
The software applies the selected boundaries and recalculates the count.
10. When finished with the count, save your count results (see “Save analysis
results”, page 96).
11. To save your count settings for Batch Analysis, see “Save current analysis
settings”, page 100.
2. Select the Channels to display in the Viewing area for manual count. You can
select multiple channels that contain captured images.
In this example, TGBFP (TagBFP) and RFP channels are selected for the manual
count.
3. Click in an Object Name field to enter a name for that label. You can use up to
six labels for the manual count.
As you tag the objects onscreen with the selected label, the system keeps a
running tally of the counts with percentages for each label assigned.
Guidelines for • We recommend that you visualize your cells using transmitted light and a
confluence phase objective with 4X to 10X magnification. Set the phase ring to Oly 4× (for
Olympus™ 4X phase objective) or 4×/10× (for EVOS™ phase objectives) using
measurements
the phase annuli selector (page 8).
• For analysis, only use 16-bit Raw image files, which contain the full dynamic
range and metadata needed for quantitative analysis.
• In the analysis tool, increasing the number of targets and background areas
improves accuracy. You can select up to 5 target areas and 5 background areas.
• The Confluence tool uses a texture and intensity-based algorithm. The
sensitivity slider adjusts the algorithm sensitivity to pixel intensity (higher
intensity = more pixels included). Decreasing the sensitivity reduces the
confluence value.
• Different cell types have different confluence “patterns”, and variability in
morphology and contrast can influence the absolute confluence measurement
between different cell types. However, within a given cell type, you can
optimize the reproducibility of your measurements. Reproducibility in
confluence measurements is more important than absolute percentages.
Measure 1. In the EVOS™ Analysis application, open the image captured in the transmitted
confluence light channel.
2. Move the pointer to the Viewing area to reveal Display Settings and Analysis
Tools, then click the Show Cell Count button.
7. To view the areas of the image counted as Target, select Show Mask. The areas
counted as Target are highlighted in the selected color.
Note: As you adjust sensitivity, observe the image with the Show Mask
option on. Ensure that the target areas are selected with minimal coverage of
the background areas. Note that decreasing the sensitivity reduces the
confluence value.
IMPORTANT! For analysis, only use 16-bit Raw image files, which contain the full
dynamic range and metadata needed for quantitative analysis.
Calculate 1. After you have completed the confluence measurement, click Transfection
transfection Efficiency to expand the controls for the Transfection Efficiency tool.
efficiency 2. Select the Fluorescence Channel for which you wish to measure the
transfection efficiency.
In this example, we want to calculate the percentage of cells that express GFP.
Therefore, the GFP channel is selected. The Fluorescence Channel and the
Transmitted Light Channel options are also checked, so that both channels are
displayed in the viewing area.
3. To view the pure fluorescence signal and to observe the various levels of
fluorescence marker (GFP) expression, uncheck the Transmitted Light Channel.
Note: Viewing the image only in the Fluorescence Channel and toggling the
Threshold Mask on and off will help you determine the best Threshold value
for your experiment.
As you adjust the Threshold, the software updates the calculation and displays
the new Transfection Efficiency value.
IMPORTANT! Images to be batch analyzed should all be of the same cell type and
have the same magnification and illumination settings. For consistent
measurements, do not mix different cell types, magnifications, or illumination
settings in the same folder when performing batch analysis.
IMPORTANT! For batch analysis, only use 16-bit Raw image files, which contain the
full dynamic range and metadata needed for quantitative analysis.
Save current 1. When finished with the Auto count, Confluence measurement, or Transfection
analysis settings Efficiency calculation, click Batch, then select Save Settings.
2. Enter the name for the settings, then click Save. The current analysis settings
are saved for reuse, which you can apply to other images that you have
collected and saved an image folder.
Note: You can also directly run Batch Analysis for images in an image folder
using previously saved Batch Analysis settings without first performing Auto
count, Confluence measurement, or Transfection Efficiency calculation (see
Step 4).
Note: Summary Data is always selected. After the analysis, summary data is
included in the analysis folder as a separate CSV file.
General care
• When cleaning optical elements, use only optical-grade materials to avoid
scratching soft lens coatings.
• Use the appropriate cleaning solutions for each component, as indicated in the
Decontamination Procedures below.
• If liquid spills on the instrument, turn off the power immediately and wipe
dry.
• Do not exchange objectives between instruments unless you know that the
components have been approved and recommended by Thermo Fisher
Scientific.
• After using, cover the instrument with the supplied dust cover.
Note: Always use the correct power supply. The power adaptor specifications
appear on the serial number label (front of LCD hinge) and in the Specifications.
Damage due to an incompatible power adaptor is not covered by warranty.
Note: To protect all optical components of the instrument, use the dust cover
when the instrument is not in use.
Decontamination procedures
In case hazardous material is spilled onto or into the components of the EVOS™
M7000 Imaging System, follow the decontamination procedure as described
below.
1. Turn power OFF.
2. Clean the LCD display.
a. Use a soft, dry, lint-free cloth to wipe off any dust from the screen.
b. Clean the LCD display with a non-alcohol-based cleaner made for flat-
panel displays.
IMPORTANT! Do not spray cleaning fluid directly onto the screen, as it may
drip into the display.
3. Lightly wipe working surfaces of the EVOS™ M7000 Imaging System (stage
top, objective turret, housing, etc.) with paper towels or Kimwipes™ tissues
dampened with 70% ethanol or 4,000 ppm hydrogen peroxide (H2O2).
2. Mount the vessel holder and the calibration slide as shown on the
screen, then click Calibrate XY.
As the instrument proceeds with the automated calibration procedure, the X-Y
stage moves between pre-set coordinates and the progress of the calibration is
shown on the screen.
4. Adjust the Brightness and Focus sliders to bring the image of the crosshairs into
focus, then click Done.
Calibrate vessel The following protocol describes the Vessel Calibration process for a generic
96-well plate as an example. The Vessel Calibration process follows a similar
workflow for the calibration of other vessel types and the software provides on-
screen instructions at each step.
1. Mount the vessel holder and the sample vessel on the X-Y stage.
2. Go to the Settings Vessels tab, then click the Vessel button to
open the Vessel Selection dialog.
3. Select the Vessel holder and Vessel type that corresponds to your
sample vessel, then click Done.
Note: If you need to restore the calibration values of a vessel to the default
settings, you can go to the first screen of the Vessel Calibration dialog and click
Restore Defaults, which replaces any existing calibration settings with the factory
default settings.
7. Drag the image or use the Jog control to bring the Top edge of the A1 well into
view. If necessary, use the arrow buttons on the edges of the image to jump to
an adhacent field.
9. When you are satisfied with the alignment of the green line with the top edge
of the A1 well, click Next.
10. Following on-screen instructions, repeat the alignment procedure for the left,
bottom, and right edges of the A1 well, clicking Next after each alignment.
11. Following on-screen instructions, repeat the alignment procedure for the right
and bottom edges of the H12 well, clicking Next after each alignment.
13. Review the images that show the alignment of the green line with the edges of
the A1 and H12 wells.
14. If satisfied with the results, click Save to update the existing vessel with the
new calibration values.
Alternatively, click Save As to save the vessel with the new calibration values
as new vessel with a name of your choice.
Change light cube 1. Remove the vessel holder from the X-Y stage.
2. On the Settings Filter Cubes tab, select the Position of the light cube you
want to change, then click Move to filter cube change position.
5. Use the light cube tool to tilt the light cube slightly toward you, then lift it out
gently. Unscrew and remove the light cube tool from the light cube.
6. Attach the tool to the new light cube, then lower the light cube into position.
Ensure that the electronic connections align properly (connections on the cube
facing the back of the microscope) and the light cube sits squarely in place.
7. Use the light cube tool to tighten the two slotted screws so that the screw
heads sit flush with the ridges on the light cube. Do not overtighten.
IMPORTANT! If the screws are not flush with the top of the light cube, they can
catch on the stage while moving and damage the system.
Change objective 1. Remove the objective you want to replace from the objective turret. You may
need to move the stage so that the objectives are accessible. Note the indicated
position (1–5) of the removed objective on the turret (red arrow).
2. Screw the new objective into the open position in the objective turret. Note the
part number of the objective and the turret position. In the following example,
a new objective is installed into the turret position 5.
3. Go to the Settings Objectives tab, and find objective in the Profiles list on
the left that matches the newly installed objective.
5. Optional: If desired, type in a label in the Display Name text box. This creates
a label that is displayed on the Objective button in the Capture tab.
In the following example, “LWD” is entered as the Display Name.
The Objective button in the Capture tab displays the label “LWD” below the
objective magnification.
Note: For best results, calibrate the newly installed objectives before using it in
your experiments (page 114).
Note: The pre-installed objectives supplied with the EVOS™ M7000 Imaging
System have been pre-calibrated. You do not need to calibrate them again unless
they are reinstalled after removal from the instrument.
Calibrate objective Calibration procedure involves two distinct steps and requires the use of the
EVOS™ M7000 calibration slide supplied with the EVOS™ M7000 Imaging System
(also available separately; Cat. No. AMEP4720).
First, the calibration crosshair on the screen is matched to the crosshair on the
calibration slide; then, the diameters of the reference circles on the calibration
slide are measured. Total time required objective calibration is about 5 minutes.
Note: For best parfocality and parcentration, calibrate the installed objectives one
after the other without removing the calibration slide.
The stage moves to the calibration target and the instrument automatically
finds the calibration crosshairs on the calibration slide (page 116).
Note: Depending on the type of objective being calibrated (e.g. Long Working
Distance vs. Coverslip Corrected objective), the calibration slide needs to be
mounted face up or face down. When mounting the slide, make sure to match the
orientation of the slide to the graphic shown on the calibration screen.
4. Manually align the green crosshairs over the black crosshairs or click Find
Crosshairs to automatically align once the black crosshairs are in view.
Note: If the circle on the screen is too large or too small, click Smaller Circle or
Bigger Circle to move to an appropriately sized circle. For best results, use the
largest circle possible.
6. Adjust the lighting and focus with the Brightness and Focus controls, then
click Find Circle to move the green calibration lines to the edges of the reference
circle. You can also manually move the lines by clicking and dragging them into
position.
7. When finished, click Next to complete the calibration.
8. Repeat the calibration process for each additional objective to be calibrated,
then click Done.
3. To install the X-Y Stage Shipping Restraint, attach to threaded holes in the
front right corner of stage, then tighten the screws in order from bottom to
top, 1-2-3.
If necessary, you can gently push on the side or front of the stage to move the
plates into alignment.
6. After securing all restraints, click Exit Application, then turn off the power to
the computer.
IMPORTANT! Do not power the instrument back on until the Shipping Restraints
have been removed.
Note: For additional technical support, contact your local EVOS™ distributor. If you do not have your
distributor information, visit thermofisher.com/evos or contact Technical Support (page 247).
Mechanical issues
Problem Possible solutions
Automatic stage does not move Remove shipping restraint.
Filter Cube Axis does not move Remove shipping restraint.
Camera Axis does not move Remove shipping restraint.
Vessel does not sit securely on moving Use the correct vessel holder for the application (visit
stage thermofisher.com/evos).
Technical specifications
Note: Technical specifications of the EVOS™ M7000 Imaging System are subject to change without notice.
For the latest product information, see the product page (thermofisher.com/evosm7000).
LED light cubes Each user-interchangable, auto-configured light cube contains an LED, collimating
optics, and filters. In addition to the channel dedicated to the transmitted light
from the condenser for brightfield contrast applications, the EVOS™ M7000
Imaging System can accommodate up to five fluorescent or specialty light cubes
for multiple-fluorescence research applications.
The table below lists some of the common fluorescent and specialty light cubes
available from Thermo Fisher Scientific. For a complete list of available light cubes
and to inquire about custom light cubes, go to thermofisher.com/evos or contact
Technical Support (page 247).
TagBFP TagBFP
CFP ECFP, Lucifer Yellow, Evans Blue
GFP GFP, Alexa Fluor™ 488, SYBR™ Green, FITC
YFP EYFP, acridine orange + DNA
RFP, Alexa Fluor™ 546, Alexa Fluor™ 555, Alexa Fluor™ 568,
RFP
Cy™3, MitoTracker™ Orange, Rhodamine Red, DsRed
Texas Red™, Alexa Fluor™ 568, Alexa Fluor™ 594, MitoTracker™
Texas Red
Red, mCherry, Cy™3.5
Cy5 Cy™5, Alexa Fluor™ 647, Alexa Fluor™ 660, DRAQ5™
Cy5.5 Cy™5.5, Alexa Fluor™ 660, Alexa Fluor™ 680, Alexa Fluor™ 700
Cy7 Cy™7, IRDye 800CW
CFP-YFP em CFP/YFP (for FRET applications)
AO Acridine orange + RNA, simultaneous green/red with FL color
White Refracted light applications
Tiled, Stitched, and If an object extends into multiple fields, a set of multiple images from the fields into
Merged images which the object extends may be required to show the object of interest in full.
Tiled images are images captured from multiple fields in an area and lined-up
close together into a tiled format, thus giving you a single image that includes
images from all the fields in the selected area.
When you use the Tiled image option without enabling Stitching, the images are
simply placed very close together with a barely noticeable seam to create an image
of all fields in that area. Tiling is quicker than stitching and is often sufficient.
If you notice that images from adjacent fields are not lining up correctly, you will
want to ensure that your instrument is fully calibrated and the camera is aligned.
Stitched images are images captured from multiple fields in an area and lined up
with a slight overlap into a tiled format. Stitching uses an algorithm to apply the
overlap, then removes the seam so you can visualize objects that cross fields.
Stitching option is available under Scan Area of Automate tab.
If you are analyzing images that cross fields (such as images of neurites), and the
images are slightly misaligned, you might want to consider Stitching the images.
Merged image is an image generated by combining the images captured in
different channels into a multicolor overlay. It is also referred to as a composite
image.
GUI layout The GUI of the EVOS™ M7000 Imaging System consists of the Viewing area on the left
and a series of tabs representing the main functions of the software (Capture,
Automate, Review, and Settings) on the right. Each tab contains the controls
necessary to execute the selected function. The Locations button, Area View/Field
View toggle, and the Zoom slider are located above the Viewing area, which can
contain additional controls depending on the view mode and the tab selected.
Locations: Opens the Location tools, which allow you to select locations or
designate capture fields for automated scan protocols (page 134).
Area View/Field View toggle: Switches between Area View and Field View.
Zoom slider: Zooms in and out of the Viewing area.
Capture tab: Contains the controls for the manual capture of images (page 139).
Automate tab: Used to create and run automated scan protocols (page 157).
Review tab: Allows you to review captured images (page 192).
Settings tab: Contains controls to select and adjust basic and advanced system
options and instrument functions (page 196).
Viewing area: Displays the sample in Area View or Field View mode
(page 127).
Note: Click on the View Help Content icon to open the Help window, which
provides EVOS™ M7000 Imaging System Help.
Position your pointer over the Tooltip icon without clicking it to view a
small Position box with additional information.
Area View In Area View, the Viewing area displays a graphical representation (i.e., map) of
the selected target area in its entirety. Depending on the vessel in use, the target
area may represent a single well of a multiwell plate, a chamber of chamber slide,
or an entire culture dish.
Area View allows you to select individual fields of view for manual image capture.
Note: If using a single culture dish, flask, or slide, the scan area corresponds to
the entire dish, flask, or slide. As such, the scan area displayed in Area View
represents the entire dish or flask selected.
In a multi-well plate, multi-chamber slide, or multiple dishes on multi-dish
holders, each scan area corresponds to a single well, chamber, or dish.
The following examples show the area map of a single well of a 96-well plate
(left) and a sample slide on a double-slide holder (right). The orange square in
each examle correspond to a single field view at 4X magnification and illustrate
the relative size of a single well in a 96-well plate and a sample slide.
Area map – Single well of a 96-well plate Area map – Sample slide
Field selection tool: Selects the target field you want to capture.
• The selection tool consists of an orange crosshair within a
broken orange rectangle. The crosshair indicates the position
of the objective lens over the target and the rectangle represents
the actual imaging area.
• The relative size of the field selection tool depends on the zoom level and
the selected objective.
Toggle field borders: Switches the display of the borders around captured
fields in Area View on and off. This option is not available in Field View.
• When field borders are turned on, all captured fields in Area View are
displayed with blue borders around them.
• Currently selected field, whether captured or not, is displayed with an
orange border, which cannot be hidden.
Image display settings: Opens the Image display settings window,
which allows you to adjust image display parameters (brightness,
constrast, gamma) for the Viewing Area. Clicking the button again
closes the window.
Note: Adjustments made to Image display settings only affect how the image
is displayed in the Viewing Area; they do not affect how the image is captured.
Note: The controls for image display settings are contextual; only the controls
for the visible channels will be available. In the example above, only the
controls for the GFP DAPI, and RFP channels are displayed.
Center on selected field: Centers the Viewing area on the selected field.
Delete image options: Opens the Delete image options window, which
contains the controls listed below. Clicking the button again closes the
window.
• Delete Selected Field: Deletes only the selected field, if the selected field
contains a captured image. If the selected field is empty, this option is not
available.
• Delete Area Fields: Deletes only the fields that were captured in the
current scan area. Images captured in other areas are not deleted.
• Delete Session Fields: Deletes all fields in all areas that were captured in
the current session.
Area and pixel information: Displays the coordinates of the pointer and pixel
intensity at the pointer location.
Toggle scale bar: Switches the display of the scale bar in the Viewing area on
and off.
Scale bar display is on Scale bar display is off
Toggle sample grid: Switches the display of the sample grid in the Viewing
area on and off.
Sample grid display is on Sample grid display is off
This information is displayed only if the Light is on for the selected channel, or
if the current field of view contains a captured image and image display is on
for the channel in which the image was captured.
• Area XY Location (Microns): Location of your pointer in relation to the
center of the selected well along the x-y axis, expressed in µm.
• Pixel XY Offset: Location of your pointer in relation to the top left corner of
the captured field, expressed in pixels along the x-y axis.
• Z Locations: Location of the captured plane along the z-axis, in µm.
• Pixel Intensities: Pixel intensity at the pointer location as reported by the
camera. This represents the number of photons detected by the camera
sensor.
Selected field of view: Displays the entire selected field of view.
• To select a location, click the desired location from the list. The X-Y axis
stage moves to align the objective with the selected (“current”) location.
In the example below, location number 3 (“Point_3”) is selected in the
Locations directory and the objective is centered on that location (as
indicated by the orange rectangle showing the field of view).
• To change the z-position at a selected location, exit the Locations tool and
adjust the focus using the focus sliders in the Capture tab, then Set Z for
that location as described.
• To remove a location from the list, select the location, then click the
Delete Location button.
Delete all: Deletes all locations in the Locations directory and removes them
from the Viewing area..
Done: Finalizes the location creation procedure and returns to the Capture tab.
The following images are example screens from the Vessel Selection dialog
for each of the vessel categories.
Note: The controls available in the Vessel Selection dialog are contextual; the
dropdown menus display only the options available for the selected vessel category.
Flasks Slides
Vessel map: Represents the vessel container (vessel + vessel holder) in use, as
determined by the selection made in the Vessel selection dropdown. The well
selected for imaging is indicated in blue on the Vessel map.
• To select a well to display in the Viewing area, click on the desired well,
flask, dish, or slide.
In the following examples, the first well of a 96-well plate and the first dish
on a 35-mm quad dish holder have been selected.
Thumbnail images: Display the most recently captured image stored in the
memory buffer for the specific channel.
• If unsaved, newly captured images will overwrite the previously captured
image in the channel.
• If no image has been captured for the channel, the corresponding
thumbnail will be blank.
• If z-stack was captured in the channel, the z-stack icon on the top left
corner of the thumbnail image for that channel identifies the image as
a z-stack projection.
• You can delete an image captured in a channel by clicking the Delete
channel image icon on the top right corner of the thumbnail image
for that channel.
In the example below, GFP channel displays the thumbnail for a z-stack
projection while the RFP channel displays the thumbnail for a field captured at
single focal plane. The rest of the channels do not have any images captured.
Light (Live mode): Turns on the excitation light for the currently selected light
cube and enters the instrument in the Live mode. The Live mode allows rapid
visualization of a large sample area illuminated with the selected light source
before image capture.
• The Vessel map displays the navigation crosshair, and the Jog
Control button appears next to the Vessel map.
• Jog Control moves the stage at an intermediate pace, allowing the quick
scanning of the sample in different parts of the sample vessel.
• Mode: Toggles between Simple and Actual modes for brightness controls.
- Simple mode allows you to control light intensity as a single
Brightness parameter (page 149).
- Actual mode allows you to adjust Light (i.e., LED intensity), Exposure,
and Gain parameters individually.
• Camera: Toggles between the Mono (monochrome)and Color cameras.
- Mono is used for image capture in fluorescence and transmitted light
(brightfield) channels. It uses a high-sensitivity 3.2 MP CMOS
monochrome camera with 2048 × 1536 pixel resolution.
- Color is used for image capture in the brightfield channel only. It uses
a high-sensitivity 3.2 MP CMOS color camera with 2048 × 1536 pixel
resolution.
• Phase: Selects between Phase options and Brightfield in the transmitted
light channel only. Phase contrast options change the phase annuli being
used to match the objective with a specific magnification. Available
options are:
- Small Ring: Used for objectives with low magnification (i.e.,
Olympus™ 4× PH)
- Medium Ring: Used for objectives with medium magnification (i.e.,
EVOS™ 4×/10× PH)
- Large Ring: Used for objectives with high magnification (i.e., EVOS™
20×/40× PH)
- Brightfield (phase contrast off)
Note: Phase options are available only for the transmitted light (brightfield)
channel; they are not available for fluorescence channels. The phase contrast
option does not require phase contrast objectives. However, to obtain a phase
image, you must first install a phase contrast objective.
• Autofocus method: Allows you to select the autofocus method best suited
for your experimental needs. Available options are:
- Fluorescence Optimized: The focal plane is derived from the highest
ratio between detailed, high-contrast objects against the background.
This option is recommended for fluorescence imaging.
- Transmitted Optimized: The optimal focal plane is derived through
statistics-based edge detection over 9 different regions to determine the
highest ratio of edge-to-background.
This option is recommended for transmitted-light imaging.
- Small Structure: This method computes the energy according to the
size of image features and can measure the presence or absence of small
image features.
Choose this method when capturing images of samples that have many
fine, hair-like structures (e.g., filaments or structural stains).
- Large Structure: This method is statistics-based and looks for large
changes in image content as parts of the image go in and out of focus.
Choose this method when your sample contains large structures (e.g.,
whole cell stains).
- Small Bright Objects: This method looks for changes in brightness at
the center of the cell and optimizes focus on cells that have a bright
center with a dark surround area.
Choose this method when capturing samples with localized staining
(e.g., nuclei).
Note: The effectiveness of the autofocus method depends on the plate type,
sample type, biomarker intensity, magnification, and channel. It may be
necessary to use different algorithms for different channels. This ensures the
best autofocus is used for your sample.
• Lock Z-Offsets: Allows you to lock the Z-Offsets, which specify the optimal
focus position in each channel relative to the focus position in other
channels.
- When locked, adjusting the Z-position in one channel changes the
Z-position in all channels, preserving the relative Z-positions of the
channels.
- When unlocked, adjusting the Z-position in one channel does not affect
the Z-position set for the other channels.
For an example, see “Note” on page 147.
For more information on configuring the autofocus settings for an
automated scan protocol, see page 176.
Note: Captured images are stored in the image cache for the selected channel.
If unsaved, newly captured images of the same field will overwrite the
previously captured image in that channel.
• The coarse focus slider bar represents the full focal range of the instrument
while the fine focus slider bar represents only a small fraction of the full
focal range.
• Coarse focus slider handle can be placed at any point within the focal
range of the instrument.
• Fine focus slider handle can be dragged as far as either end of the slider
bar, but it will snap back to the center position upon release.
• You can adjust the focus by moving the focus sliders in the desired
direction.
Alternatively, you can double-click on the handle of the focus slider to
activate the text box and enter the desired value for the focus position.
• If the instrument is in the Live mode, the live image changes in effective
real-time to reflect the changes made to the focus.
Record Video: Opens the Record Video controls, which allow
you to record capture a series of live images and compile them
into a video.
• Record Video controls are available only in the Live mode.
Note: The EVOS™ Onstage Incubator (Cat. No. AMC1000) is an optional accessory
that enables the incubation of cells directly on the automatic X-Y stage, allowing
the capture of images from the same sample over long periods of time and
recording of time lapse movies.
• Light: Turns on the light source in the selected channel to illuminate the
sample as you locate the top and bottom boundaries of the z-stack using
the focus sliders. Click the button again turns off the light.
• Set Top Position: Sets the current focal plane reached using the focus
sliders as the top position of the z-stack.
• Set Default Focus: Specifies the new default focus position for the z-stack.
• Set Bottom Position: Sets the current focal plane reached using the focus
sliders as the bottom position of the z-stack.
• Step Size: Describes the z-distance in µm between the successive focal
planes that are captured for the z-stack. Available methods to specify the
step size are:
- Multiple of Depth of Field: Enter the step size as multiple of depth of
field.
- Step Size: Enter the step size in µm in the corresponding text box.
• Number of planes: Specifies the number of focal planes that are captured
to generate the z-stack. Each captured image along z-axis represents a
single “optical section” of the sample.
Note: You can capture z-stack images in only one channel at a time; however,
you can save z-stack images of a field captured in different channels as a
single merged video.
• Select fields to save: Allows you to select the captured fields you want to
save. You can choose from the following options:
- Currently selected field: Saves images only from the current field.
- All newly captured fields: Saves images that have been captured and
stored in the image cache, but not yet saved. This option is available
only if you have previously saved images from the same session.
- All captured fields: Saves images from all captured fields that are held
in the image cache. This is typically all of the images that you have
captured during an imaging session.
• Location: Describes the location of the current folder or file.
• Select file types to save: Allows you to set I mage save options. You can
select one or more of the following options:
• Raw images: Saves images from
individual channels as 16-bit raw
images. This is the recommended
format for image analysis and
quantitation (see “Raw vs. Displayed
images”, page 125).
Available save options for raw images are:
- Single field, individual channels: Saves images captured in each
field and each channel individually.
- Z-stack planes, individual channels: Saves individual z-stack images
as well as the z-stack projection for each field and channel.
Available file formats for Raw images are:
- TIFF: No or low image compression with no loss in image quality.
- PNG: High image compression with no loss in image quality.
- C01: Thermo Fisher™ proprietary format. Only images captured by
monochrome cameras may be saved in this format.
- DIB: Color corrects across devices for true viewing. Only images
captured by monochrome cameras may be saved in this format.
Note: When viewing 16-bit images on most standard image viewers, they will
appear very dark or almost totally black, especially if a low exposure setting
was used. These images either need to be converted for display or opened in a
software application that allows viewing of 16-bit images such as the Thermo
Scientific™ Celleste™ Image Analysis Software (Cat. No. AMEP4816).
Note: 24-bit images (8-bit per RGB channel) are NOT recommended for image
analysis as not all channels will display in many image analysis applications.
Note: For more information about when to use tiled or stitched images, see
“Tiled, Stitched, and Merged images”, page 125.
Note: If you plan to analyze your images, save them in 16-bit TIFF Raw format.
The 16-bit dynamic range is required for analysis and there is additional
information in the file headers, including OME data, that can be useful.
Selected Vessel
Selected Camera
Select Objective
Select Channels
Phase Ring
Adjust Settings
Hardware controls Selected Vessel: Displays the selected vessel type. For more information, see
“Vessel” in “Capture tab controls”, page 140.
Selected Camera: Displays the selected camera. For more information, see
“Brightness and camera settings” in “Capture tab controls”, page 144.
Select Objective: Allows you select the objective for the scan protocol from the
currently installed objectives.
You can select only one objective at a time. In the example below, the 10X
objective is selected.
Select Channels: Allows you to select the channels you want to capture in the
scan protocol. The brightness value set in the Capture tab for each channel is
displayed below the corresponding checkbox.
You can select multiple channels by checking the corresponding channel boxes.
In the example below, the DAPI and GFP channels have been selected.
Phase Ring: Displays the phase option selected for the transmitted light
(brightfield) channel. For more information, see “Phase” in “Capture tab
controls”, page 144.
Adjust settings: Allows you to modify channel settings for the
scan protocol in the Capture tab (page 134).
In the capture tab, you can change Brightness and camera settings (page 144),
adjust Brightness parameters (page 149), and select a different Vessel
(page 140) and Objective (page 142).
Note: If using a single culture dish, flask, or slide, the scan area corresponds to the
entire dish, flask, or slide. As such, the scan area displayed in Area View represents
the entire dish or flask selected.
In a multi-well plate, multi-chamber slide, or multiple dishes on multi-dish holders,
each scan area corresponds to a single well, chamber, or dish.
You can specify multiple areas to scan, but the Area View displays only the current
(i.e., active) scan area.
• When using the Scan Locations tool (page 161), the scan areas are selected
using the Assign scan area button (page 164).
When using the Scan Pattern tool (page 166), the scan areas are selected
from the Vessel map using the Edit scan area button (page 168).
• When multiple scan areas have been defined, you can specify the order in
which they are captured using the Area Acquisition Order (page 169).
• You can specify the order in which the selected fields are captured using
the Field Acquisition Order (page 169).
In the following example, the first location in the list (an ellipse) is selected.
The Viewing are displays the selected location and the field group that best
fits the shape of the location.
• To add all of the locations listed in the Available Locations table, click
Add All.
• The locations added to the scan protocol are displayed in the Scan Locations
table and removed from the Available Locations table.
The Scan Locations table also provides information about the name and
shape of each location, and the number of fields it contains.
In the example below, two locations have been added to the Scan locations
list, and one location remains in the Available Locations table.
• When assigning scan areas, each selected scan location (i.e., highlighted in
Scan Locations table) is assigned to each selected scan area.
If no location is selected in the Scan Locations table, all scan locations listed
in the table are assigned to each selected scan area.
If the Vessel map contains no selected scan areas, the scan protocol uses
the default scan area associated with each scan location.
- Ctrl-click-drag to select
multiple blocks on the Vessel
map.
• As you define the parameters, the Area View displays a preview of the
selected fields.
• When you are finished defining the parameters for the capture pattern,
click Create.
Create Locations: Open the Locations tool (page 134), which
allows you to create new locations for the scan protocol.
Using the Locations tool, define the locations as described (page 135), then
navigate back to the Automate tabScan Locations.
Newly created locations will be listed in the Available Locations table.
• To define scan areas in the Vessel map, follow the same procedure as
described for the Assign scan area (page 165).
• If you want to view a larger version of the Vessel map, click the
Zoom button. The larger map displays the same content as the
smaller vessel map and behaves the same way.
• When in the edit mode (i.e., Edit scan area button is clicked), clicking on
an area in the larger map selects the corresponding area in the smaller map
and marks it for the scan protocol.
• Clicking the zoom button again closes the zoom window.
• When finished defining the scan areas, click Done to exit the selection
mode.
Toggle locations display: Switches the display of locations (defined with the
Location tool) in Area View on and off.
Toggle field borders: Switches the borders around captured fields in Area
View on and off. This option is not available in Field View.
• When field borders are turned on, all captured fields in Area View are
displayed with blue borders around them.
• Currently selected field, whether captured or not, is displayed with an
orange border, which cannot be hidden.
Note: Adjustments made to Image display settings only affect how the image
is displayed in the Viewing Area; they do not change how the image is
captured.
Note: The controls for image display settings are contextual; only the controls
for the visible channels will be available. In the example above, only the
controls for the GFP, DAPI, and RFP channels are displayed.
Capture Z-Stacks
Use AutoFocus
Channel
Focus Position
Z-Offsets
Note: The options available on this panel are contextual; the controls that are
displayed depend on the selections made in this panel or elsewhere in the
Automate tab. Objects and controls that are not available as a function in a
particular context are not shown in the panel.
Note: Click on the View Help Content icon to open the Help window, which
provides EVOS™ M7000 Imaging System Help.
AutoFocus and Z Capture Z-Stacks: Allows you to capture a series of images at various focal
Stacks controls planes along the z-axis within the sample.
Checking this option reveals the controls for Z-Stack Settings (page 173).
Use AutoFocus: Allows you configure the autofocus method and set the
autofocus frequency to use during automated scan.
Checking this option reveals the controls for AutoFocus Settings (page 176).
Channel: Indicates the channels selected for capture in scan protocol, and
allows you set the default focus position.
Depending on the context, Channel displays channel selection buttons for the
Default Focus Position and the Z-Offsets.
Focus Position: Indicates the active z-stage position (i.e., the position of the
focal plane along the z-axis) for the selected channel (page 178).
Z-Offsets: Allows you to specify the focus position in each channel relative to
the focus position in other channels when the fluorescent markers in different
channels are in different focal planes (page 178).
This option is not available when the scan protocol is set to capture images in a
single channel.
Note: When the Capture Z-Stacks option is checked, the Z-Offsets option is not
available as one of the Focus Position controls and the Z-Offsets that have been
previously set are not preserved.
Note: Click on the View Help Content icon to open the Help window, which
provides EVOS™ M7000 Imaging System Help.
Z-Stack Summary: Provides a summary of the Z-Stack Settings that were set up
in the Capture tab (Capture tab Capture Z-Stack, page 152).
The summary includes information on depth of field, step size, number of
planes, total range of the z-stack, top and bottom z-positions, default focus
position of the selected channel, and the z-stack focus position.
Note: Click on the View Help Content icon to open the Help window, which
provides EVOS™ M7000 Imaging System Help.
Note: The effectiveness of the autofocus method depends on the plate type,
sample type, biomarker intensity, magnification, and channel. It may be
necessary to use different autofocus methods for different channels. This
ensures that the best autofocus method is used for the relevant biology.
Start Focus Position: Indicates the currently active z-stage position (i.e.,
position of the current focal plane along the z-axis) for the selected channel.
• The Start Focus Position is used as starting point around which the
autofocus algorithm searches for optimal focus.
• You can only select a single channel for the Start Focus Position.
• When Z-Offsets option is checked, Start Focus Position is used as reference
point against which the focus position in other channels is set.
Z-Offsets: Allows you to specify the focus position in each channel relative to
the focus position in other channels. Setting the correct Z-Offsets is especially
important when the fluorescent markers in different channels are in different
focal planes.
AutoFocus Frequency: Allows you to specify which fields in a scan area are
used for autofocus during the automated scan protocol. Specifying fewer fields
per scan area for autofocus reduces the overall scan times.
Use Time Lapse (page 181) Image capture frequency (page 183)
Use Incubator (page 181) Incubator (page 184)
Run (page 182) Add run (page 184)
Duration (page 182) Autofocus Settings (page 185)
Delay Start (page 182)
Note: The options on this panel are contextual. Objects and controls that are not
available as a function in a specific context are not shown in the panel.
Note: Click on the View Help Content icon to open the Help window, which
provides EVOS™ M7000 Imaging System Help.
• A time lapse routine can be divided into individual “runs”, which allow
for different imaging frequency, duration, and incubator settings (see
“Run”, below).
• To add additional runs to your time lapse routine, click the Add run
button (page 184).
• The fields and locations to be captured during a time lapse experiment are
selected in the Scan Area panel (page 159).
• The autofocus options for the time lapse routine are set in the AutoFocus
and Z Stacks panel (page 172).
• The captured images are stored with time lapse information in their
headers and can be stitched together into a video.
Use Incubator: Enables the use the EVOS™ Onstage Incubator your time lapse
experiments. This option is available only when the EVOS™ Onstage Incubator
is connected to the EVOS™ M7000 Imaging System.
When checked, Use Incubator option reveals the controls for configuring the
incubator settings (see “Incubator”, page 184).
• The options available in this area are contextual and depend on the
selections made for Use Time Lapse and Use Incubator.
• Click the collapse/expand button to hide the controls for the run.
Click the button again to expand the controls.
• To add additional runs to your time lapse routine, click the Add run
button (page 184).
• If there is more than one run in the time lapse routine, each run has its own
set of controls that are similarly collapsible/expandable.
• Click the Delete Run button to delete a run and remove it from the
panel.
Duration: Allows you to specify the duration of the corresponding run in the
time lapse routine.
• Enter the desired values for the duration of the run into the corresponding
Hours, Minutes, and Seconds fields.
Delay Start: Allows you to postpone image capture for the corresponding run
of the time lapse routine and specify the duration of the delay.
• Enter the desired values for the delay into the corresponding Hours,
Minutes, and Seconds fields.
• When the Delay Start option is not selected, image capture starts
immediately after the time lapse experiment is initiated.
• By default the Delay Start option is deselected, with 0 hours entered for the
duration. After this feature has been used once, the most recent delay
period entered is used as the default value.
• The incubation parameters are set separately for each run in the time lapse
routine.
• Temperature: Sets the incubator temperature (ambient to 40°C).
• CO2: Sets the CO2 level (0% to 20%).
• Oxygen: Sets the oxygen level (0% to 20%).
• Use humidity: Enables the use of humidified atmosphere in the incubation
chamber.
• Shutdown: Allows you to select the shutdown method for the EVOS™
Onstage Incubator at the end of your time lapse routine.
- Turn off manually: The incubator remains on until the Use Incubator
option is manually deselected.
- Turn off at the end of experiment: Heat, humidity, and the flow of gas
are automatically turned off at the end of the experiment.
- Turn off after: Enter the time period in hours and minutes that must
elapse before the heat, humidity, and the flow of gas are automatically
turned off.
Add run: Adds a new run to the time lapse routine.
• Clicking the Add run button adds a new set of run controls
to the panel, which are identical to the existing run controls.
This allows you set the parameters for each run (delay start, duration,
capture frequency, and incubator settings) independently for each run.
• The runs are performed in chronological order, beginning with Run 1 and
ending with Run N.
• You can remove a run from the time lapse routine by clicking the
corresponding Delete Run button .
• First time point only: Performs the autofocus procedure only once at the
first time point of the run before proceeding with image capture.
• Every time point: Performs the autofocus procedure at every time point of
the run, ensuring that the sample is in focus for each image set captured.
• Autofocus fail options: Selecting Yes causes the instrument to skip the
autofocus procedure at locations where no optimal focal plane was found
at subsequent time points.
Selecting No ensures that the autofocus is run at each selected autofocus
location every time regardless of previous autofocus failure.
Note: The Autofocus Settings options available in this panel apply only to the
time lapse routine. Autofocus locations and method are determined
separately in the AutoFocus and Z Stacks panel (page 176).
Note: Click on the View Help Content icon to open the Help window, which
provides EVOS™ M7000 Imaging System Help.
Position your pointer over the Tooltip icon without clicking it for additional
information.
Note: For more information on the file naming convention when saving, see
“File naming convention”, page 191.
Note: If you plan to analyze your images, save them in 16-bit TIFF Raw format.
The 16-bit dynamic range is required for analysis, and there is additional
information in the file headers, including OME data, that can be useful.
Note: When viewed on most standard image viewers, 16-bit images will appear
very dark or almost totally black, especially if a low exposure setting was used.
These images will either need to be converted for display or opened in a
software application that allows viewing of 16-bit images.
Note: If you are analyzing images that cross fields and the images are slightly
misaligned, you might want to consider stitching the images (page 170).
Stitching applies an algorithm to apply an overlap, and then removes the seam
so you can visualize objects that cross fields. For more information about tiled
and stitched images, see “Tiled, Stitched, and Merged images”, page 125.
Note: For more information about when to use tiled or stitched images, see
“Tiled, Stitched, and Merged images”, page 125.
• The Vessel map shows the location of the Field that is displayed in the
Viewing area.
Note: You can also position your pointer over a scan or plate in the Browse panel or
over an image in the Image preview/Image list panel without clicking to display
the metadata associated with that scan, plate, or image.
Images/Metadata: Lists the images or scans saved within the selected folder (if an
image folder is selected in the Folders panel) or the scan metadata for the selected
scan (if a scan is selected in the Folders panel).
Note: Display options in the Review tab are visible only when an image is
displayed in the Viewing area.
General settings General: Allows you to set the Saturated Pixel, Jog Control, Scale Bar, Grid
controls Settings, and AutoFocus options (page 197).
Cameras: Allows you to perform Hot Pixel Correction for the monochrome
camera and to adjust White Balance for the color camera (page 199)
Objectives: Allows you to set up and calibrate objectives and assign objective
profiles (page 201).
Stage Calibration: Opens the Stage Calibration tool, which allows you to
calibrate the automatic X-Y axis stage for accurate instrument function
(page 204).
Vessels: Allows you to select sample vessels, calibrate the instrument for the
selected vessel, to import and export vessel information, and to import stage
insert information (page 205)
Incubator: Allows you to configure the gas inputs, calibrate the oxygen sensor,
and set the temperature offsets for the EVOS™ Onstage Incubator (page 208).
The Incubator panel is only visible when an EVOS™ Onstage Incubator is
connected to your EVOS™ M7000 Imaging System.
Service: Provides version information about instrument software and firmware
(page 210).
Camera Rotation: Allows you to calibrate the camera rotation for accurate
instrument function (page 212).
Filter Cubes: Allows you to add or remove EVOS™ LED light cubes and to
assign pseudocolors for installed light cubes (page 213).
Note: For samples that contain air bubbles or debris, we recommend that the
Quick AutoFocus option is disabled.
Cameras controls Hot Pixel Correction: Opens the Hot Pixel Correction
dialog, which allows you to search for hot pixels in the
monochrome camera and reset the brightness value of
all hot pixels to a base pixel intensity level.
Hot pixels are caused by electrical charges that leak into the sensor wells and
they appear brighter than the other pixels in the camera. Hot Pixel Correction
algorithm of the EVOS™ M7000 software stores the location and value of the
hot pixels in its memory for subsequent automatic hot pixel correction.
• Search for Hot Pixels: Prompts the software to search for hot pixels in the
image.
Click OK to save the location of hot pixels for subsequent correction.
• Reset Hot Pixels: Resets the hot pixel locations to the factory default.
• To automatically adjust the white balance, find a white area on the image,
click and drag to select that area on the field of view, then click Auto
Correct.
• To manually adjust the white balance, find a white area on the image, then
use the Red and Blue sliders.
Note: For specific instructions on how to change and calibrate objectives, see
“Change the objectives” (page 112) and “Calibrate the objectives” (page 114).
Details: Displays the detailed information about the active objective in the
turret.
Note: For specific instructions on how to calibrate the stage, see “Calibrate the
stage” (page 105).
Note: For specific instructions on how to calibrate the instrument for the selected
vessel, see “Calibrate vessel” (page 107).
• Select the appropriate Holder and Vessel type from the dropdown menus
available for your vessel category.
• Click Export to open the Save As dialog, which allows you to save the
vessel information as a *.vessel file at the desired destination directory.
Vessel map: Represents the vessel container (vessel + vessel holder) in use, as
determined by the selection made in the Vessel selection dropdown.
Note: For specific instructions on how to calibrate the instrument for the selected
vessel, see “Calibrate vessel” (page 107).
Note: The Incubator panel is only visible when an EVOS™ Onstage Incubator is
connected to your EVOS™ M7000 Imaging System.
Note: For specific instructions on how to set up the EVOS™ Onstage Incubator and
use it with the EVOS™ M7000 Imaging System, see “Appendix D: EVOS™ Onstage
Incubator” (page 224).
Software version (page 210) Copy Log Files to USB (page 211)
Firmware version (page 210) Move to Shipping Position (page 211)
Objective Turret (page 211) Cache (page 211)
Service controls Software version: Displays the installed EVOS™ M7000 software information.
• To calibrate the camera rotation, insert the EVOS™ Calibration Slide into the
instrument, adjust the Brightness, then Focus on the crosshairs of the
calibration slide.
When you have focused on the crosshairs, click Next to begin the calibration of
camera rotation.
Note: To insert or remove a light cube, first select the Position of the light cube you
want to change, then click Move to filter cube change position. For detailed
instructions, see “Change EVOS™ light cubes” (page 107).
Edit Pseudocolors: Contains the controls to edit the pseudocolor for installed
light cubes.
To select a new pseudocolor, move the Color slider to the desired position on
the spectrum or enter the corresponding wavelength into the Pseudocolor
wavelength text box.
Filter cube tabs: Each tab represents an installed light cube and contains the
controls to change the psudocolor assigned to that cube.
Current pseudocolor: Displays the pseudocolor assigned to the light cube.
Color spectrum: Represents the color spectrum available to choose as a
pseudocolor for the light cube.
Color slider: Allows you to select the desired pseudocolor.
Wavelength: Allows you to enter the wavelength that corresponds to the
pseudocolor you want to assign to the light cube.
Apply changes: Applies the selected pseudocolor to the light cube. To save the
changes, click Done after applying the changes.
Restore Default: Restores the factory default pseudocolor to the light cube.
Review tab controls Viewing area: Displays the image selected from the Image preview/Image list.
Image file name: File name of the image displayed in the Viewing area.
Zoom slider: Zooms in and out of the image. The zoom range is 100% to 1000%.
Folder: Displays the location of the current folder or image.
Search: Allows you to search by file name in the selected folder.
Layout: Allows you to toggle between grid or list view, increase or decrease the
display size of the folders or image files, and sort by name, file type, or date
created in ascending or descending order.
Image preview/Image list: Displays the preview or list of the files and
subfolders in the current folder.
Image properties: Displays the metadata for the selected image file.
Display and analysis tools: Allow you to change image display settings in the
Viewing area, annotate and analyze the captured images, and perform cell
count and cell culture analysis (confluence and transfection efficiency). See
page 217 for more information.
Settings: Opens the Settings tab, which allows you to select image format
options for saving TIFF files, view EVOS™ Analysis software version, and to
copy error logs to a storage device.
Refresh: Refreshes the list or grid of images in the current folder.
Display settings: Displays or hides the Layout controls.
Export: Allows you to export the currently selected folder or image to a storage
device.
Save: Saves the currently opened image.
Image Display Settings: Opens the Image display settings tool, which allows
you to adjust image display parameters (Brightness , Contrast , Gamma
Correction ) for the selected channels.
Measurements and
Annotations: Allows you to
draw regions of interest
(rectangle, ellipse, polygon,
line, or free-form) on the
captured image and
measure dimensions, area,
or perimeter of the drawn region.
Analysis tools: Allows you to toggle between Auto Count, Manual Count, or Cell
Culture (Confluence and Transfection Efficiency) tools for image analysis.
Channel: Selects the nuclear stain channel for Auto Count. You can select only a
single channel.
Select Target and Background: Allows you to select representative target objects
and background areas for Auto Count.
Split Cells: Allows you to split multiple objects that have been counted as one into
individual objects based on shape or pixel intensity to increase the count accuracy.
Refine: Selects intensity, area, or circularity by which to refine Auto Count results
using the Count Histogram.
Count Histogram: Allows you to set pixel intensity, area, or circularity thresholds to
refine the Auto Count results.
Object Count: Displays the object count based on the Auto Count parameters.
Object color: Selects the color by which the counted objects are identified.
Reset: Resets the count to 0 and clears the selected targets and background areas.
Batch Analysis: Allows you to save and apply the analysis parameters to other
images that you have collected and saved in an image folder (see page 100 for more
information).
Exit: Exists the Auto Count tool and displays the Review tab.
Save: Saves the analysis results as an image in the selected file format (see page 98
for more information).
Analysis tools: Allows you to toggle between Auto Count, Manual Count, or Cell
Culture (Confluence and Transfection Efficiency) tools for image analysis.
Channel: Selects the channels for Manual Count. You can select multiple channels.
Object #: Allows you to select the label (Object #) with which to tag objects in the
Viewing area. Left-click on the objects in the Viewing area to tag them with the
selected label; right-click to delete a tag. In this example, Object #2 is selected. You
can switch labels as desired. See page 90 for more information.
Object Name: Allows you assign a name for the Object #.
% and Count: Displays the label count and its percentage of the total object count
(total count of all labels).
Delete: Selects a label for deletion.
Total Count: Displays the total count of objects tagged with all labels.
Trash: Deletes the tags for the label and resets the label count to 0.
Reset: Resets all label counts and the total count to 0 and clears the Viewing area of
all tags.
Batch Analysis: Allows you to save and apply the analysis parameters to other
images that you have collected and saved in an image folder (see page 100 for more
information).
Exit: Exists the Manual Count tool and displays the Review tab.
Save: Saves the analysis results as an image in the selected file format (see page 98
for more information).
Analysis tools: Allows you to toggle between Auto Count, Manual Count, or Cell
Culture (Confluence and Transfection Efficiency) tools for image analysis.
Confluence: Expands or hides the controls for the Confluence tool.
Select Target and Background: Allows you to select representative cell and
background areas for the confluence measurement.
Sensitivity: Adjusts the algorithm sensitivity to pixel intensity (higher intensity =
more pixels included). Decreasing the sensitivity reduces the confluence value.
Show Mask: Indicates the areas included in the confluence measurement.
Mask Color: Selects the mask color.
Transfection Efficiency: Expands or hides the controls for the Transfection
Efficiency tool. The Transfection Efficiency tool is inactive until the confluence
measurement is completed.
% Confluence: Displays the percentage of the area covered by cells in the image,
based on the selected target and background areas and sensitivity.
Reset: Resets the Confluence and Transfection Efficiency measurements to 0 and
clears the selected targets and background areas.
Batch Analysis: Allows you to save and apply the cell culture analysis parameters to
other images that you have collected and saved in an image folder (see page 100).
Exit: Exists the Cell Culture tool and displays the Review tab.
Save: Saves the analysis results as an image in the selected file format (see page 98).
General: Allows you to configure TIFF file saving options. You can select to
save TIFF files in a Microsoft-compatible format (8-bit with reduced dynamic
range), uncompressed (16-bit Raw format with no loss in image quality), or
both.
Service: Display the EVOS™ Analysis Software version.
Copy Error Logs: Allows you to copy error logs to an external storage device
(USB drive).
Done: Saves your changes and returns to the Review tab.
EVOS™ Onstage The EVOS™ Onstage Incubator (Cat. No. AMC1000) is an optional accessory for the
Incubator EVOS™ FL Auto Imaging System that enables the incubation of cells on the
automatic X-Y stage, allowing the capture of images from the same sample over
long periods of time and recording of time lapse movies.
The EVOS™ Onstage Incubator consists of a Stagetop Environmental Chamber that
is placed on the automatic X-Y stage of the imaging system and a separate Control
Unit that supplies the power and gas (air or air-CO2 premix, CO2-only, and
nitrogen-only). The onstage incubator is controlled by the same software and user
interface that controls the EVOS™ M7000 Imaging System.
Physical Stagetop
characteristics Environmental Chamber Control Unit
Height: 25 cm (9.7 in) 37 cm (15 in)
Depth: 19 cm (7.6 in) 16 cm (6.3 in)
Width: 3.7 cm (1.5 in) 20 cm (7.9 in)
Weight: 1.5 kg (3.3 lb) 10 kg (22 lb)
Temperature range: Ambient to 40°C (± 0.1°C)
Humidity: >80% relative humidity (RH) at 37–40°C
CO2 range: 0% to 20%
O2 range: 0% to ambient
Operating power: 100–240 VAC, 1.8 A
Frequency: 50–60 Hz
Electrical input: 24 VDC, 5 A
Hardware Compatible vessels: Multi-well plates, 35-mm Petri dishes, T-25 flasks
Gas input ports: Air or air-CO2 premix, CO2-only, and N2-only (max. 50 psi input)
Stagetop environmental chamber accessories: Master Stage Plate, Vessel holder
for multi-well plates
Control unit
Hose heater connector
Heated hose
Sensor data cable with
6-pin connector
Environmental chamber
2. Secure the onstage incubator master plate to the X-Y stage using the four
thumb screws (indicated by red arrows).
2. Attach the vessel holder/adaptor to the incubator chamber using the four
thumb screws (indicated by red arrows).
Note: Place an empty “dummy” culture plate into the vessel holder/adaptor for the
initial warm up and equilibration to prevent build-up of condensation on the optical
components and the inside of the EVOS™ M7000 Imaging System.
4. Place the light shield with tinted plastic window on top of the heated glass lid.
Use of the light shield is required for fluorescence imaging applications.
5. If desired, place the light shield cover on the light shield for fluorescence
imaging applications. The light shield cover completely blocks any ambient
light from entering the environment chamber and improves image quality in
fluorescence imaging applications.
For the locations of the various input jacks and gas ports, refer to “Control unit
rear view”, page 226.
IMPORTANT! Do not position the control unit so that it is difficult to turn off the main
power switch. In case of an instrument malfunction, turn the main power switch to
the OFF position and disconnect the power cord from the wall outlet.
1. Plug power cord into the power input jack on the control unit and the wall
outlet.
2. Plug USB cable into the USB control cable jack on the control unit and the USB
port on the computer.
3. Connect each gas line to the appropriate gas tank via the PTC (push-to-click)
connectors threaded into the regulator. To do this, push the tubing into the
PTC connector until it clicks into place. Pull on tubing slightly to ensure a tight
connection; the tubing should not come out.
4. Attach the gas lines to the control unit via the PTC connectors for the
appropriate gas intake port.
• If using pre-mixed air, attach to Port 1: Air In
• If using compressed air and CO2, attach to Port 1: Air In and Port 3: CO2 In
• For oxygen displacement, attach to Port 1: Air In and Port 2: Nitrogen In
5. Plug the 6-pin sensor data cable to the environmental chamber and the
appropriate input jack on the control unit.
7. Place the water reservoir into the control unit with the fill holes to the front
and close the lid.
8. Attach the heated hose between the environmental chamber and the control unit
9. Plug the hose heater cable to the connector on the heated hose.
Note: Place an empty “dummy” culture plate into the vessel holder/adaptor for the
initial warm up and equilibration to prevent build-up of condensation on the optical
components and the inside of the EVOS™ M7000 Imaging System.
2. Select the appropriate options for the Gas Inputs that reflects your set-up for
the EVOS™ Onstage Incubator.
• For Port 1, you may select Air or Premix.
If you select Premix, manually enter the percentage of the CO2 and O2 to
reflect the specifics of your set-up.
• Port 2 is reserved for Nitrogen only.
• Port 3 is reserved for CO2 only.
3. Click Done once you have configured the gas connections for each port.
4. Turn on the regulators on the gas tanks. The meters on the regulators show
the tank fill on the right and gas flow on the left.
5. Set the flow on the regulators as follows. Do not exceed 50 psi of pressure.
• Air: 40–50 psi
• CO2: 40–50 psi
• Nitrogen: 40–50 psi
Symbol Description
Indicates the On position of the main power switch.
Symbol Description
Indicates that you should consult the manual for further information
and to proceed with appropriate caution.
Indicates the presence of an electrical shock hazard and to proceed
with appropriate caution.
Indicates the presence of a hot surface or other high-temperature
hazard and to proceed with appropriate caution.
Indicates the presence of a laser inside the instrument and to proceed
with appropriate caution.
Indicates the presence of moving parts and to proceed with
appropriate caution.
Indicates the presence of a biological hazard and to proceed with
appropriate caution.
Indicates the presence of an ultraviolet light and to proceed with
appropriate caution.
Environmental The following symbol applies to all Thermo Fisher Scientific electrical and
symbols on electronic products placed on the European market after August 13, 2005.
instruments
Symbol Description
Do not dispose of this product as unsorted municipal waste. Follow
local municipal waste ordinances for proper disposal provisions to
reduce the environmental impact of waste electrical and electronic
equipment (WEEE).
European Union customers:
Call your Customer Service representative for equipment pick-up and
recycling. See www.thermofisher.com for a list of customer service
offices in the European Union.
Hazard
English Français
Symbol
CAUTION! Hazardous chemicals. Read the ATTENTION! Produits chimiques dangereux.
Safety Data Sheets (SDSs) before handling. Lire les fiches techniques de sûreté de
matériels avant toute manipulation de
produits.
CAUTION! HAZARDOUS WASTE. Refer to ATTENTION! Déchets dangereux. Lire les
SDS(s) and local regulations for handling and fiches techniques de sûreté de matériels et la
disposal. régulation locale associées à la manipulation
et l’élimination des déchets.
DANGER! High voltage. DANGER! Haute tension.
WARNING! To reduce the chance of electrical AVERTISSEMENT! Pour éviter les risques
shock, do not remove covers that require tool d’électrocution, ne pas retirer les capots dont
access. No user-serviceable parts are inside. l’ouverture nécessite l’utilisation d’outils.
Refer servicing to Thermo Fisher Scientific L’instrument ne contient aucune pièce
qualified service personnel. réparable par l’utilisateur. Toute intervention
doit être effectuée par le personnel de service
qualifié venant de chez Thermo Fisher
Scientific.
DANGER! Class 3B visible and/or invisible DANGER! Rayonnement visible ou invisible
laser radiation present when open. Avoid d’un faisceau laser de Classe 3B en cas
exposure to beam. d’ouverture. Evitez toute exposition au
faisceau.
CAUTION! Moving parts. Crush/pinch ATTENTION! Pièces en mouvement, risque de
hazard. pincement et/ou d’écrasement.
Moving and lifting CAUTION! PHYSICAL INJURY HAZARD. The instrument is to be moved and
the instrument positioned only by the personnel or vendor specified in the applicable site
preparation guide. If you decide to lift or move the instrument after it has
been installed, do not attempt to lift or move the instrument without the
assistance of others, the use of appropriate moving equipment, and proper
lifting techniques. Improper lifting can cause painful and permanent back
injury. Depending on the weight, moving or lifting an instrument may
require two or more persons.
Moving and lifting WARNING! Do not attempt to lift or move the computer or the monitor
stand-alone without the assistance of others. Depending on the weight of the computer
computers and and/or the monitor, moving them may require two or more people.
monitors Things to consider before lifting the computer and/or the monitor:
• Make sure that you have a secure, comfortable grip on the computer or the
monitor when lifting.
• Make sure that the path from where the object is to where it is being moved is
clear of obstructions.
• Do not lift an object and twist your torso at the same time.
• Keep your spine in a good neutral position while lifting with your legs.
• Participants should coordinate lift and move intentions with each other before
lifting and carrying.
• Instead of lifting the object from the packing box, carefully tilt the box on its
side and hold it stationary while someone slides the contents out of the box.
Operating the Ensure that anyone who operates the instrument has:
instrument • Received instructions in both general safety practices for laboratories and
specific safety practices for the instrument.
• Read and understood all applicable Safety Data Sheets (SDSs). See “Safety
Data Sheets (SDS)”.
Gas cylinders
You must supply the required nitrogen, oxygen, and carbon dioxide gas cylinders
and accessories for the installation. This instrument requires pressurized house
lines, or one size 1-A gas cylinder that holds approximately 7.2 m3 (257 ft3) of gas
when full for each gas. Use only pre-purified gasses of 99.9% or greater purity.
CAUTION! Damage to the instrument and its products can result from
using impure gas, gases other than specified, or an inadequate amount of
gas.
WARNING! Gas cylinders are heavy and may topple over, potentially
causing personal injury and tank damage. Cylinders should be firmly
secured to a wall or work surface. Please contact your environmental
health and safety coordinator for guidance on the proper installation of a
gas cylinder.
Pressure regulator
You must supply a two-gauge regulator with a Compressed Gas Association
(CGA) 580-cylinder adaptor on the inlet side and a Swagelok™-type end-fitting
that accepts 6.35-mm (0.25-in.) o.d. tubing. The primary gauge (0 to 3000 psi;
0 to 25,000 kPa recommended) measures tank pressure, and the secondary gauge
(0 to 200 psi; 0 to 2000 kPa recommended) measures regulated pressure. The
secondary gauge must allow regulation to 50 psi. Compressed Gas Association
(CGA) 580-cylinder adaptor with a needle-type shutoff valve on the exit side. The
needle valves should have Swagelok™-type end-fittings ready for connection to
6.35-mm (0.25-in.) o.d. tubing.
Attaching the cylinder
Attach the pressurized gas cylinder firmly to a wall or gas cylinder cart by means
of approved straps or chains.
Ventilation requirements
Allow at least 50 cm (20 in) of clearance around the Instrument for ventilation.
Waste disposal If potentially hazardous waste is generated when you operate the instrument, you
must:
• Characterize (by analysis, if necessary) the waste generated by the particular
applications, reagents, and substrates used in your laboratory.
• Ensure the health and safety of all personnel in your laboratory.
• Ensure that the instrument waste is stored, transferred, transported, and
disposed of according to all local, state/provincial, and/or national
regulations.
Fuses WARNING! FIRE HAZARD. For continued protection against the risk of fire,
replace fuses only with fuses of the type and rating specified for the
instrument.
Overvoltage rating The EVOS™ M7000 Imaging System has an installation (overvoltage) category of
II, and is classified as portable equipment.
In the U.S.:
• U.S. Department of Health and Human Services guidelines published in
Biosafety in Microbiological and Biomedical Laboratories
(stock no. 017-040-00547-4;
www.cdc.gov/OD/ohs/biosfty/bmbl4/bmbl4toc.htm)
• Occupational Safety and Health Standards, Bloodborne Pathogens
(29 CFR§1910.1030;
www.access.gpo.gov/nara/cfr/waisidx_01/29cfr1910a_01.html)
• Your company’s/institution’s Biosafety Program protocols for working
with/handling potentially infectious materials.
• Additional information about biohazard guidelines is available at:
www.cdc.gov
In the EU:
• Check your local guidelines and legislation on biohazard and biosafety
precaution, and the best practices published in the World Health Organisation
(WHO) Laboratory Biosafety Manual, third edition
www.who.int/csr/resources/publications/biosafety/WHO_CDS_CSR_LYO_
2004_11/en/
U.S. and Canadian The CSA C/US Mark signifies that the product meets applicable U.S. and
safety standards Canadian standards, including those from CSA, CSA America, ANSI, ASME,
ASSE, ASTM, NSF and UL.
European safety The CE Mark symbolizes that the product conforms to all applicable European
and EMC standards Community provisions for which this marking is required. Operation of the
instrument is subject to the conditions described in this manual.
The protection provided by the instrument may be impaired if the instrument is
used in a manner not specified by Thermo Fisher Scientific.
Australian EMC The C-Tick Mark indicates conformity with Australian and New Zealand
standards standards for electromagnetic compatibility.
Obtaining support
Technical support For the latest services and support information for all locations, visit
www.thermofisher.com.
At the website, you can:
• Access worldwide telephone and fax numbers to contact Technical Support
and Sales facilities
• Search through frequently asked questions (FAQs)
• Submit a question directly to Technical Support (thermofisher.com/contact)
• Search for user documents, SDSs, vector maps and sequences, application
notes, formulations, handbooks, certificates of analysis, citations, and other
product support documents
• Obtain information about customer training
• Download software updates and patches
Limited product Life Technologies Corporation and/or its affiliate(s) warrant their products as set
warranty forth in the Life Technologies’ General Terms and Conditions of Sale found on
Life Technologies’ website at www.thermofisher.com/us/en/home/global/terms-
and-conditions.html. If you have any questions, please contact Life Technologies
at www.thermofisher.com/support.