Lab 2
Lab 2
Proprietary Notice:
This document and the information contained in it is proprietary and confidential information of Philips and
may not be reproduced, copied in whole or in part, adapted, modified, disclosed to others, or disseminated
without the prior written permission of the Philips Legal Department. Use of this document and the
information contained in it is strictly reserved for current Philips personnel and Philips customers who have a
current and valid license from Philips for use by the customer’s designated in-house service employee on
equipment located at the customer’s designated site. Use of this document by unauthorized persons is strictly
prohibited. Report violation of these requirements to the Philips Legal Department. This document must be
returned to Philips when the user is no longer licensed and in any event upon Philips’ first written request.
CSIP Level 2:
This document or digital media and the information contained in it is restricted proprietary and confidential
information that CANNOT BE DISTRIBUTED OR LICENSED AND IS ONLY FOR THE USE OF PHILIPS PERSONNEL
AND LICENSED REPRESENTATIVES. IT IS NOT TO BE DISTRIBUTED, SHARED, COPIED, TRANSFERRED OR
OTHERWISE PROVIDED TO ANYONE ELSE, ELECTRONICALLY OR OTHERWISE.
Page 1 of 18
Objectives
When you have successfully completed this lab, you will be able to:
• Explain the user interface layout
• Power up the system
• Warm up the X-ray tube
• Perform air calibrations
• Conduct a surview scan
• Plan a study and conduct Axial scans
• Conduct a Helix (Spiral) scan
• Verify operation of the ACS/DOM function
• Find, view and delete images from the home page
• Manipulate the images
• Perform a post-reconstruction
• Power down the system
Prerequisite Lessons
• System Introduction Lecture
• Safety Lecture
Tools/Reference Materials
• Access Dual/CT, MX16, Incisive CT Instruction for use manual
Time
2 hours
Introduction
The purpose of this module is to make the student familiar with the basic operation of the Access CT/DUAL, MX16
and Incisive CT systems. Refer to the correct CT system “Instructions for Use” throughout this entire lab.
The student will properly apply power up the system, warm up the x-ray tube, perform air calibrations, and
remove power from the system.
The basic system functionality is verified by performing scans to created surview, axial and helical images. This lab
allows the student to become familiar with the methods of image selection and manipulation on the CT systems.
During each of the following labs the students will properly power up and power down the system with a
functional scan at the start and end of lab to ensure the scanner is working.
Page 2 of 18
Procedure Preview
In this lab exercise, the learner will perform:
• Power up and down the system
• Basic functional scan
• Image manipulation
Special Instructions/Precautions
<CAUTION:
Damage to the equipment and personnel can occur if you are not careful. If you do not know the outcome of a
motion before initiating, DON’T DO IT! Please seek help from the instructor if unsure.>
Instructions
Tube Warm Up/ Short Tube Conditioning
For Access and MX16 warming up the tube before performing a scan is called Warm Up (Dunlee Tube). Short tube
conditioning is the name used by Incisive CT (MRC tube)
Tube Warm Up
Question:
What is the tube heat after warm up?
Can you change this heat percentage when logged on as user CT? (Non-Service Mode)
Page 3 of 18
Air Calibration
Question:
Can you perform specific calibrations for one protocol only?
If yes, how?
Surview Scan
The purpose of this step is to learn how to make a surview scan and plan a study using the
surview image.
1. Ask your instructor for a “Mystery Box” and put it on the patient table. Activate the laser
indicator.
2. Position the mystery box just before the inner laser.
3. Make a surview scan of the mystery box. The position of the patient table can now be read
from the gantry control panels.
Question:
What is the position of the patient table according to the gantry control panel?
Do not manually change the table position during this part of the lab.
Page 4 of 18
Question:
Why do we have to indicate how the patient is positioned on the table?
5. For Access and MX16 you need to press the arrow to go to the next screen for selecting a
protocol. For Incisive CT the protocol selection is done on the same screen. Select the “Brain
Exam (Ax)”
6. First perform the surview scan. On the surview image there will appear a blue box / window
which can be used to plan the scan.
Questions
What is the position of the X-ray tube during a 90 surview?
Look at the image, what do you think is inside the black box?
What is the meaning of the increment (mm) parameter and how much will the couch move
after the first axial scan?
Page 5 of 18
Planning a Scan
Question
How many images will be made to cover the complete “Mystery Box”?
What will change if you change the size of the blue box / window in vertical direction?
3. Change the default mAs value to 46 and execute the plan by pressing the green “GO”
button.
Questions:
How many images are made?
Did you discover any other items inside the “Mystery Box”?
4. We will now plan a second series of axial scans on the mystery box. The same surview will
be used. On Access/MX16 press the “Plan Scan Button” then the “Insert Protocol” Button
and then select the “Exam protocols” tab. On Incisive CT press the Insert Exam Card Icon.
5. Select the “Brain Exam. (Ax)” protocol, change the slice thickness to 5mm and increment to
10.
Question
What’s the meaning of the different colors used for the different protocols?
Page 6 of 18
How many different slice thicknesses can be selected in this protocol?
What is the couch position for the last axial scan (calculate)?
Question:
What is the scan position after the scan series?
Page 7 of 18
Helical (Spiral) Scan
The purpose of this step is to learn how to make a helical scan and to understand the meaning
of the different parameters used.
Make use of the Instructions for Use manual to make an abdomen helical scan.
3. Select the “Abdomen Protocol Group” and the Abdomen Exam Protocol.
Question:
How many procedures are in this examination?
Question:
What is the difference between a helical and an axial scan?
8. Adjust the blue box so that the complete mystery box is covered.
Question:
In this protocol what is the default setting for the pitch? What is the meaning of pitch?
Page 8 of 18
How many images will be made?
Question:
Did the number of images change, and why?
11. Change the “Collimation” to 2*5, “Thickness” to 3mm. Under ‘Increment’ select
‘Continuous’.
Question:
Did the number of images change and why?
12. Change the thickness to 4mm and change the “increment” to 4mm.
Question:
13. Leave the increment parameter on 3mm. If you change the length to 300mm the scan time
will also change.
14. Check your helical plan and see if it still covers the mystery box. Execute the scan as planned
and wait until all images are reconstructed. Later we will use this data for post
reconstructions and volume reconstructions. After all images are reconstructed, end the
study.
Page 9 of 18
Find, View and Delete image
1. After closing the helical study, you will see the Patient register page. On Access/MX16 click
the “Home” button on Incisive CT click Completed.
The following screen is used to find, retrieve and delete patient images and raw data on Access
and MX16.
Access/MX16
Page 10 of 18
Incisive CT
1
3
The right side of screen shows a window which is split up into three parts.
The first session allows you to see and choose different data storage devices available on the
system (1).
The second session contains the patient examinations (2).
The third session shows the series that were made during the exam (3). There are three tabs:
Series, Images and Reports. Selecting a series and then clicking on the image tab will show all the
images in that series.
Preview of the images can be seen in the lower right corner.
Explore the directory page and perform the following operations:
Questions:
Name different possibilities to store patient image?
Page 11 of 18
When copy Patient data to a CDR, which data (Images, Raw, or Both) will be copied to the CD?
2. Select an image series (not a surview) and press the ‘viewer’ button under the ‘Data
options’. For Incisive UI, select the image series and press the “review” button.
This action will load the series to the viewer and will show the review screen. The viewer
can be used to diagnose the images.
3. Look at the image and try to adjust the Window Width (WW) and the Window Level (WL).
Question:
How did you change the WW and the WL?
4. On the lower left of the screen there is a ‘Windowing’ section. Click on the ‘Modified>>’
button and select the ‘lung’ window.
Question:
How can you apply the ‘Lung’ Window on all the images?
5. Locate the Image Parameters button and view the Image information. For Incisive UI,
navigate to the Information list located at the top icon bar.
6. Exercise other options in the viewer.
7. Exit the review window.
Page 12 of 18
ACS & DOM Functionality
ACS (Automatic Current Selection) is a volume-based function that helps clinicians get
consistent noise characteristics for different sized patients.
C-DOM (Combined Dose Modulation) is a prospective scanning mode in which the X-ray tube
current is modulated according to patient’s size in both the rotation and Z directions, this
reduces streaking artifacts, and improves image noise consistency along the Z direction.
This exercise demonstrates how to verify the modulations functions are working properly.
1. Position the non-linear phantom on the table top with use of the phantom holder.
2. Start a new study and select a Lung protocol. (surview and helical)
3. Perform the surview scan and make sure to capture the whole phantom + 5 cm overlap on
both sides.
4. Plan the scan and make sure of the whole phantom + overlap will be scanned and duplicate
this series.
5. Deselect dose modulation for the first series and check if for the second series the dose
modulation is activated.
6. Execute both scans. When the images are previewed pay attention to the mA values. The
first series should show a stable mA value through the slices. In the second series as the
phantom size differs the mA value will change.
Page 13 of 18
Archiving Images
Archiving images is done from the home page of the application software. For Incisive UI you
need to select the “completed” button.
The Dicom images, the raw data and the calibration tables are stored on the D and E drive of the
host computer. There is no automatic deletion of the Dicom images from the host computer.
Manual deletion of the Dicom images should only be done after they are archived. Archiving can
be done to a PACS system, CD-R, DVD or any other (correctly configured) device.
For this exercise we will archive patient data to a CD-R.
Question:
What is the function of the “lock” button and can you lock individual series?
What is the function of the “Modify” or “Edit” button and which patient data can still be
modified?
Will the original data still be available after you modified the patient data?
Page 14 of 18
Is it possible to delete only the raw data and keep the images? How can you do this?
3. Use the “copy to” button to copy the images you made to CD-R.
Question:
Can you hide the patient information (anonymize) from the images being written to CD?
5. Copy the raw data to a CD, DVD or any external storage device.
Page 15 of 18
Post Reconstruction
The purpose of this exercise is to learn how you can execute a post reconstruction from
available RAW data.
1. Navigate to the IRS data storage device. Before you can make a post reconstruction you
must have RAW data available on the host computer.
For Incisive UI you need to select the raw data tab in the completed window
Question:
How can you check if there is raw data available?
2. Select the Raw data for ‘training Spiral’ study. Click on the Recon button on left side of the
screen under ‘Recon’ section.
3. Change the following parameters:
Slices Thickness: 5mm
Increment: 5mm
FOV: 250mm
Metal Artifact Reduction: ON
4. Perform the reconstruction. After the reconstruction is finished, click ‘Exit’. This will bring
you back to the home page.
Page 16 of 18
Changing Users/ Logging Out
When handing over the system to the customer, it is very important to leave the service mode.
This will prevent the user to access windows operating system.
Leaving the system incorrect can be a system security and instability treat.
From being logged in as SERVICE, Log out from the HOME screen. Notice the system logs out
and leaves you at the service desktop, with multiple shortcuts to different tools, i.e. my
computer, backup/restore…
We DO NOT want to return the system to the customer in this state, with access to the desktop
icons. To leave the system at the correct point:
1. Launch the host application from the Icon on the desktop.
2. Log in as: philips_service, pwd: service_only
3. Navigate to > Service and select the “Switch User” button
4. Now login as user: CT
5. Navigate to the Home screen and logout from the application.
Notice on the desktop; only the START button in the lower left corner and a system Icon is
available.
After every service call, leave the system in this state before you turn it over to the customer.
Page 17 of 18
Additional Information
<For further information about the topic, read/do….>
Page 18 of 18