EET6L Experiment 4 Pressure Sensor
EET6L Experiment 4 Pressure Sensor
ASSESSMENT: NOTE:
Max. Marks
Assessment For minor plagiarism:
Marks Awarded
Both student who copied and student
Performance 10 who allowed another student to copy
Data Collection & his/her work will be given Zero Mark.
20
Analysis For late submission:
Results & Discussion 20 1 Day – 10% deduction
Staff Signature
Prepared by
Engr. Rommel Aunario
September 2024
II. Introduction
In this experiment, you will have the opportunity to measure the dynamic response of the process
and test it in real time, by using the fill and drain valves
As you already performed the sensor calibration in the previous experiment, you will not need to
repeat the calibration procedure. In this standard calibration, we have 1 V for each 1 cm of the level.
VI Procedure
NOTE: The positioning of the valves can be seen on the DL 2314 board.
NOTE: You will notice that we have an ON/OFF switch in DC voltage sources. For the source to send the
present voltage signal, we have to activate the switch up.
NOTE:
1) Adjust the voltage on SETPOINT 1 at 8V. Start the pump by switching up the key.
2) Wait until the level of the water in the tank reaches 5 cm. of the tank scale.
5) Wait until the level is stabilized. If this will not happen adjust again the knob DRAIN VALVE
and/or the knob NEEDLE VALVE to maintain the level at a constant value of 5 cm.
6) Temporarily, adjust the voltage on SETPOINT1 back to 0 V and switch down the key.
7) Bring back the water level to 0 cm. through the RELIEF VALVE, without modifying the position of
the DRAIN VALVE and NEEDLE VALVE. After that, close the RELIEF VALVE.
8) Adjust the voltage on the SETPOINT1 at 8 V and switch up the key and start at the same time the
chronometer.
9) Measure the level at 20 seconds interval and write down the values in the measurement table.
10) Repeat the procedure to fill all the values in the measurement table.