CbDF60a U0Sc

The ATEQ CDF60 User Manual provides essential information on the operation, installation, and maintenance of the ATEQ CDF60 leak/flow calibrator. It includes safety precautions, user interface details, and instructions for calibration and servicing. The manual emphasizes the importance of proper training for operators and outlines the support available through ATEQ's after-sales service.

Uploaded by

Mihai Pana
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
16 views52 pages

CbDF60a U0Sc

The ATEQ CDF60 User Manual provides essential information on the operation, installation, and maintenance of the ATEQ CDF60 leak/flow calibrator. It includes safety precautions, user interface details, and instructions for calibration and servicing. The manual emphasizes the importance of proper training for operators and outlines the support available through ATEQ's after-sales service.

Uploaded by

Mihai Pana
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 52

USER MANUAL

ATEQ CDF60
Version 1.0

Reference: UM-23900B-U
REVISION OF THE CDF60 USER MANUAL

Edition/ Date
Reference Chapters updated
Revision (week/year)

First edition UM-23900A-U 35/2007 -----


Recommendations for leak testing
instruments

Precautions for the test environment


• Keep the test area as clean as possible.

Precautions for the operators


• ATEQ recommends that the operators using the instruments should have a
suitable qualification and training with respect to the work bench requirements.

General precautions
• Read the user manual before using the instrument,
• all electrical connections to the instrument must be equipped with a safety
system (fuse, circuit breaker…) appropriate to its needs and complying with the
standards,
• to avoid electromagnetic interference, the cable connections to the instrument
should be less than two meters in length,
• it is essential that the electrical main is earthed,
• disconnect the electrical connections to the equipment before maintenance,
• cut the air supply for any kinds of operation on the pneumatic assembly,
• do not open the instrument when it is powered up,
• avoid water spillage near of the instrument,
• ATEQ is at your disposal for any further information concerning the use of the
instrument under maximum safety conditions.

We would like to bring to your attention that ATEQ will not be held responsible
for any accident connected to the improper use of the instrument, to the work
bench or to the lack of compliance with safety rules.
ATEQ Company is free from any responsibility for any adjustment of its
instrument which would not have been done by its own technicians.
The ATEQ cannot be held responsible if the instrument (program, mechanics or
electronics) has been modified without prior written consent.

0039/Ret-U
ATEQ, THE ASSURANCE OF A COMPETENT AFTER SALES
SERVICE
„ THE ATEQ AFTER SALES SERVICE IS :
• a team of qualified technicians,
• a permanent telephone assistance,
• agencies close to you for faster reaction,
• a stock of spare parts available immediately,
• a car fleet for rapid intervention,
• a commitment to quality ...
„ THE OVERHAUL
ATEQ carries out the overhaul of your instruments at interesting prices.
The overhaul corresponds to the maintenance of the instrument (checking, cleaning, replacing of used
parts) as part of preventive maintenance.
Preventive maintenance is the best way to guarantee reliability and efficiency. It allows the
maintenance of a group of instruments in good operational order and prevent eventual break-downs.
„ MAINTENANCE KITS
The ATEQ After Sales Service proposes, two kits destined for the preventive maintenance of the
pneumatic circuits of instruments.
„ CALIBRATION
This may be carried out on site or in our offices.
ATEQ is attached to the COFRAC and delivers a certificate following a calibration.
„ TRAINING COURSES
In the framework of partnership with our customers, ATEQ offers two types of training in order to
optimise the usage and knowledge of our instruments. They are aimed at different levels of technician:
• method / control training,
• maintenance / upkeep training.
„ A TARGETED TECHNICAL DOCUMENTATION
A number of technical documents are at your disposal to allow you to intervene rapidly in the event
minor breakdowns:
• problem sheets describing and offering solutions to the main pneumatic and electronic
problems,
• several maintenance manuals.
„ A QUALITY GUARANTEE
The instruments are guaranteed for parts and labour in our offices:
• 2 years for leak detection equipment,
• 1 year for electrical tests to norms instruments,
• 1 year for the accessories.
Our After Sales Service is capable of rapidly answering all your needs and queries.

We strongly recommend to send the instrument


back to ATEQ once a year for re-calibration

SAV-Uc/0650
PREFACE

Dear Customer,

You have just purchased an ATEQ instrument, we thank you for the trust you have
placed on our brand. This instrument has been designed to ensure a long and
unparalleled life expectancy, and we are convinced that it will give you complete
satisfaction during many long years of operation.

In order to maximise the life expectancy and reliability of your ATEQ instrument, we
recommend that you install this instrument on a secured workbench and advise you to
consult this manual in order to familiarise yourself with the functions and capabilities of
the instrument.

Our ATEQ After Sales Service centre can give you recommendations based on your
specific operation requirements.

ATEQ

0650/PREFd-U
Table of contents

TABLE OF CONTENTS

PREAMBLE 3
1. ATEQ CDF60 DEFINITION..................................................................................................................3
2. MEASUREMENTS CHARACTERISTICS ...........................................................................................4
2.1. Flow measurement ranges...........................................................................................................4
2.2. Atmospheric pressure and temperature indicators ......................................................................4

Chapter 1 INTRUMENT INSTALLATION


1. PRESENTATION OF THE ATEQ CDF60 ...........................................................................................5
2. ATEQ CDF60 INSTALLATION............................................................................................................6
2.1. Supply connector 24 V DC...........................................................................................................6
2.2. USB connector .............................................................................................................................6
2.3. RS232 connector .........................................................................................................................6
2.4. Pneumatic connector (first input) .................................................................................................6

Chapter 2 USER INTERFACES


1. KEYS....................................................................................................................................................7
1.1. “Start / Information” key................................................................................................................7
1.2. "Cancel" key .................................................................................................................................7
1.3. “HOLD” key ..................................................................................................................................7
1.4. "Enter" key....................................................................................................................................8
1.5. Navigation keys ............................................................................................................................8
2. OTHERS ITEMS ..................................................................................................................................9
2.1. LCD display ..................................................................................................................................9
2.2. Lights............................................................................................................................................9

Chapter 3 START UP ADJUST AND MEASURES


1. STARTING THE ATEQ CDF60 .........................................................................................................11
2. MEASURES DISPLAY ......................................................................................................................12
3. PARAMETERS ADJUST...................................................................................................................13
3.1. Filter ...........................................................................................................................................13
3.2. Flow unit .....................................................................................................................................14
3.3. Standard conditions (Std cond)..................................................................................................15
3.4. Gas.............................................................................................................................................16
3.5. Temperature unit........................................................................................................................17
3.6. Back light....................................................................................................................................18
3.7. Automatic power off (auto off) ....................................................................................................19
3.8. RS232 ........................................................................................................................................20

Chapter 4 CDF60 FUNCTIONS


1. MENUS STRUCTURE .......................................................................................................................21
2. SPECIALS CYCLES..........................................................................................................................22
2.1. Auto-zero....................................................................................................................................22
2.2. Measurements print (Print).........................................................................................................23
2.3. Sensors display ..........................................................................................................................24

UM-23900B-U User manual ATEQ CDF60 Page 1/40


Table of contents

Chapter 5 ACCESSORIES, SECURITY AND RECYCLING


1. ACCESSORIES .................................................................................................................................25
1.1. Power supply..............................................................................................................................25
1.2. Soft case ....................................................................................................................................25
1.3. USB wire ....................................................................................................................................25
1.4. Test tube ....................................................................................................................................26
1.5. Protection cover (option) ............................................................................................................26
1.6. Hard case (option)......................................................................................................................26
1.7. Needle valve (option) .................................................................................................................26
2. SAFETY INFORMATIONS ................................................................................................................27
3. RECYCLING ......................................................................................................................................28

Chapter 6 ERROR MESSAGES


1. ERROR MESSAGES .........................................................................................................................29

Chapter 7 PC INSTALLATION
1. PROGRAM AND DRIVERS INSTALLATION ...................................................................................27
1.1. USB connection..........................................................................................................................27
1.2. RS232 connection......................................................................................................................31
1.3. RS232 cable diagram.................................................................................................................31

Appendix ATEQ CDF60


1. TECHNICALS CHARACTERISTICS.................................................................................................37
2. PERSONNAL NOTES .......................................................................................................................38

Index 39

UM-23900B-U User manual ATEQ CDF60 Page 2/40


Preamble

PREAMBLE

1. ATEQ CDF60 DEFINITION

ATEQ CDF60
Pressure
sensor

Auto-zero
valve
the instrument
Connection to

to be tested

Security

Atmosphere
valve

Capillary

The ATEQ Leak/flow calibrator is a flow meter destined for checking leak instruments
and in particular those of ATEQ. It measures a loss of charge using a pressure sensor
placed on a calibrated flow tube. It can also be used for checking a leak flow or a
calibrated jet.

UM-23900B-U User manual ATEQ CDF60 Page 3/40


Preamble

2. MEASUREMENTS CHARACTERISTICS

2.1. FLOW MEASUREMENT RANGES

Flow Maximum
Standard Option (*) resolution
ranges
Accuracy in
conditions 4 ml/min 2% RV + 0,01 ml/min 1% RV + 0,01 ml/min 0,001 ml/min
defined as
ambient
No linearity +
Hysteresis +
Repeatability + T° drift 40 ml/min 2% RV + 0,1 ml/min 1% RV + 0,1 ml/min 0,01 ml/min
+ Long time drift (90
days)

Accuracy in
conditions 4 ml/min 2,5% RV + 0,01 ml/min 2% RV + 0,01 ml/min 0,001 ml/min
defined as
standard
(1013 hPa -
20°C)
No linearity +
Hysteresis + 40 ml/min 2,5% RV + 0,1 ml/min 2% RV + 0,1 ml/min 0,01 ml/min
Repeatability + T° drift
+ Long time drift (90
days)

RV = Read value.
(*) Instrument adjusted in our metrology laboratory with COFRAC agreement.

2.2. ATMOSPHERIC PRESSURE AND TEMPERATURE INDICATORS

Accuracy
Minimum Maximum
(20°C)

Atmospheric pressure 850 hPa 1050 hPa +/- 200 Pa

Temperature (internal) 0° C 50° C +/- 1,5%

Note: the zero offset linked to the temperature or the position can be take over by an
auto-zero. It is however advised to make an auto-zero before a measurement.

Recommendation of use: according to some use's conditions, the instrument can put
to two hours of time of heating before complete stabilization.

UM-23900B-U User manual ATEQ CDF60 Page 4/40


Chapter 1 – Instrument installation

Chapter 1
INTRUMENT INSTALLATION
Before the first use of the instrument, make a complete battery
charging, wait until the CHARGE light is off.

1. PRESENTATION OF THE ATEQ CDF60

Pneumatic
Connector

Display

Keyboard

Charging
battery
connector
USB and RS232
connectors

The ATEQ CDF60 is according a portable case.


The instrument has a pneumatic quick connector (Staublï kind) for the connection to the
flow to be measured.
The instrument has three electrics connectors:
¾ 1 connector for the battery charge, the instrument run with lithium-ion batteries
under 12 V DC voltage and it supplies with a specific supply transformer to
charge the batteries.
¾ 1 RS232 connector.
¾ 1 USB connector.

UM-23900B-U User manual ATEQ CDF60 Page 5/40


Chapter 1 – Instrument installation

2. ATEQ CDF60 INSTALLATION

2.1. SUPPLY CONNECTOR 24 V DC

Connector for the 24 V DC supply. For the battery charging. (Jack


connector).

0 V DC + 24 V DC

2.2. USB CONNECTOR

Allows the connection to a PC. (Type B USB connector).

2.3. RS232 CONNECTOR

Allows the connection to a printer with a RS232 connection. RJ45 type


connector. For further information see chapter 7, paragraph 1.2 "RS232
connection".

2.4. PNEUMATIC CONNECTOR (FIRST INPUT)

Allows the connection to the flow to be measured (quick connector


"Staublï" RBE03 female type).

UM-23900B-U User manual ATEQ CDF60 Page 6/40


Chapter 2 – User Interfaces

Chapter 2
USER INTERFACES

1. KEYS

1.1. “START / INFORMATION” KEY

KEY FUNCTION

Switched off instrument: this key switch on the instrument.


Switched on instrument: allows the displaying of:

¾ the program version, the flow range and the battery


level. Press and hold the key to display the battery
level.

¾ Switching off the instrument, the press and lengthy


hold this key is switching off the instrument (more than
3 seconds).

1.2. "CANCEL" KEY

KEY FUNCTION

Cancel he edition or the modification of the parameters.

Return to the previous menu.

1.3. “HOLD” KEY

KEY FUNCTION

To freeze or unfreeze the measurement display.

UM-23900B-U User manual ATEQ CDF60 Page 7/40


Chapter 2 – User Interfaces

1.4. "ENTER" KEY

KEY FUNCTION

ENTER: menus access, parameter edition, parameter


validation.

Menus access: its exists two modes:

1st mode, short press (< 1 second) : allows acceding to the


"special cycles" menu,

2nd mode, press and hold (> 3 seconds) allows acceding to


the "parameters" menu.

1.5. NAVIGATION KEYS

KEY FUNCTION

Scroll up or increase numeric values.

Swing from the single "Flow" or extended "Flow


/temperature / Atmospheric pressure" displays.

Scroll down or decrease numeric values.

Swing from the single "Flow" or extended "Flow


/temperature / Atmospheric pressure" displays.

UM-23900B-U User manual ATEQ CDF60 Page 8/40


Chapter 2 – User Interfaces

2. OTHERS ITEMS

2.1. LCD DISPLAY

Allows the measurements and parameters displaying.

2.2. LIGHTS

Three DEL indicate the instrument state.

Battery level: when this is on, the battery level becoming too low
LOW BAT for reliable operation, indicating that the battery must be
recharged before further use.

Charge light: this light is glowing red while the battery is being
CHARGE
charged. When charging is finished, the light turns off.

TX Not used.

UM-23900B-U User manual ATEQ CDF60 Page 9/40


Chapter 2 – User Interfaces

UM-23900B-U User manual ATEQ CDF60 Page 10/40


Chapter 3 – Start up, adjust and measures

Chapter 3
START UP ADJUST AND MEASURES

1. STARTING THE ATEQ CDF60

The CDF60 can run with the power supply or with the internal battery.

Check the pneumatics and electrics connections before the start up of the CDF60.

Switches on the instrument.

When this is on, the battery level becoming too low


for reliable operation, the "LOW BAT" light will light
on.

The instrument starts the auto-zero cycle.


AUTO ZERO

+/- 40.00 cm3/min


At the end of the auto-zero cycle, the instrument AIR STD.COND.

indicates the current flow measured and is ready


for measurement. 00.00
Note: it's important to make regularly an auto-zero cycle.

UM-23900B-U User manual ATEQ CDF60 Page 11/40


Chapter 3 – Start up, adjust and measures

2. MEASURES DISPLAY

The instrument displays continuously the current flow. There's no measurement start
cycle.
There are two display modes:
¾ The simple mode where only the current flow is displayed.
+/- 40.00 cm3/min
AIR STD.COND.

09.36
¾ The extended mode where all the measures are displayed: flow, internal
instrument temperature and atmospheric pressure.
+/- 40.00 cm3/min
AIR STD.COND.
FLOW : 09.36
TEMP. : 25.8°C
ATM.P : 1010.5 hPa

To switch from the simple mode to the extended mode, press the UP or the

DOWN keys.

Display description (extended mode):


Instrument
range

Selected gas Displayed if


+/- 40.00 cm3/min
AIR STD.COND. standard
Current flow conditions are
FLOW : 09.36
validated
Temperature TEMP. : 25.8°C
ATM.P : 1010.5 hPa
Atmospheric
pressure

Note: the data sending on the RS232 frame is different according to the display mode
selected, in simple mode, only the current flow is sent, in extended mode, the three
measures are sent.

UM-23900B-U User manual ATEQ CDF60 Page 12/40


Chapter 3 – Start up, adjust and measures

3. PARAMETERS ADJUST

It's existing three adjusted parameters.

3.1. FILTER

It can be adjusted from 0.1 to 9.9 seconds by 0.1 second of step. It allows making a
mean on the measurement time, making easier the measurement reading.

+/- 40.00 cm3/min


Enter into the parameters menu, by pressing and AIR STD.COND.
FLOW : 09.36
old the ENTER during 3 seconds. TEMP. : 25.8°C
ATM.P : 1010.5 hPa

The PARAMETER menu is displayed. By using the PARAMETER


>FILTER : 0.5 s
UP and DOWN key select FILTER FLOW UNIT.: cm3/min
STD.COND.: YES
parameter.

PARAMETER
Then press the ENTER key to modify the FILTER : 0.5 s <
FLOW UNIT.: cm3/min
parameter (the cursor moves on the right). STD.COND.: YES

PARAMETER
By using the UP and DOWN key FILTER : 1.0 s <
FLOW UNIT.: cm3/min
modify the parameter to the hoped value. STD.COND.: YES

PARAMETER
>FILTER : 1.0 s
Validate with the ENTER key. FLOW UNIT.: cm3/min
STD.COND.: YES

+/- 40.00 cm3/min


To return to the measurement mode, press AIR STD.COND.
FLOW : 09.36
CANCEL . TEMP. : 25.8°C
ATM.P : 1010.5 hPa

UM-23900B-U User manual ATEQ CDF60 Page 13/40


Chapter 3 – Start up, adjust and measures

3.2. FLOW UNIT

This parameter selects the flow unit displayed during the measurements.

+ /- 40 .00 c m 3/mi n
Enter into the parameters menu, by pressing and AI R ST D. CO ND.
FLOW : 09.36
old the ENTER during 3 seconds. TEMP. : 25.8°C
ATM.P : 1010.5 hPa

The PARAMETER menu is displayed. By using the PARAMETER


FILTER : 1.0 s
UP and DOWN key select >FLOW UNIT.: cm3/min
STD.COND.: YES
PRESS. UNIT parameter.

PARAMETER
Then press the ENTER key to modify the FILTER : 1.0 s
FLOW UNIT.: cm3/min<
parameter (the cursor moves on the right). STD.COND.: YES

UNIT
By using the UP and DOWN key >l/h
cm3/min
choose the unit to be displayed. cm3/s

PARAMETER
FILTER : 1.0 s
Validate with the ENTER key. FLOW UNIT.: cm3/h <
STD.COND.: YES

+ /- 2 40 0 cm 3 / h
To return to the measurement mode, press AI R S TD. CO ND .
FLOW : 0936
CANCEL . TEMP. : 25.8°C
ATM.P : 1010.5 hPa

UM-23900B-U User manual ATEQ CDF60 Page 14/40


Chapter 3 – Start up, adjust and measures

3.3. STANDARD CONDITIONS (STD COND)

This parameter allows displaying the measurements values corrected to the standards
conditions values, atmospheric pressure and temperature, to the user preferences.

+ /- 4 0. 0 0 c m 3 / mi n
Enter into the parameters menu, by pressing and AI R
FLOW : 09.36
old the ENTER during 3 seconds. TEMP. : 25.8°C
ATM.P : 1010.5 hPa

The PARAMETER menu is displayed. By using the PARAMETER


FILTER : 1.0 s
UP and DOWN key select FLOW UNIT.: cm3/min
>STD.COND.: NO
STAND. COND parameter.

PARAMETER
Then press the ENTER key to modify the FILTER : 1.0 s
FLOW UNIT.: cm3/min
parameter (the cursor moves on the right). STD.COND.: NO <

By using the UP and DOWN key PARAMETER


FILTER : 1.0 s
validate the function by YES. Validate with the FLOW UNIT.: cm3/min
ENTER key. STD.COND.: YES <

The standards temperature and atmospheric STANDARD CONDITION


pressure values are displayed, it's possible to >TEMP. : 20.0°C
modify following the user preferences. Default PRESS. : 1013.2hPa
values: 20° C and 1013.2 hPa.

To return to the measurement mode, press twice +/- 40.00 cm3/min


AIR STD.COND.
CANCEL . The measurements values FLOW : 09.36
TEMP. : 25.8°C
displayed are corrected following standards ATM.P : 1010.5 hPa
conditions.

UM-23900B-U User manual ATEQ CDF60 Page 15/40


Chapter 3 – Start up, adjust and measures

3.4. GAS

This parameter allows calculating the flow measurement for a other gas, default gas:
air.
The avalaible parameters are for the following gas:
¾ Helium (He),
¾ Nitrogen (N2),
¾ Other (the parameter had to be indicated, consult us).

+/- 40.00 cm3/min


Enter into the parameters menu, by pressing and AIR STD.COND.
FLOW : 09.36
old the ENTER during 3 seconds. TEMP. : 25.8°C
ATM.P : 1010.5 hPa

The PARAMETER menu is displayed. By using the PARAMETER


FLOW UNIT.: cm3/min
UP and DOWN key select GAZ STD.COND.: NO
>GAZ : AIR
parameter.

Then press the ENTER key to modify the


GAZ
parameter available the gas list is displayed. By >AIR
HELIUM
using the UP and DOWN key choose NITROGEN
the gas to be measured.

It's possible to configure a different gas than those GAZ


available, for that select the OTHER option, then HELIUM
configure the applied coefficient. For further NITROGEN
information, contact ATEQ company. >OTHER

GAZ
>AIR
Validate with the ENTER key. HELIUM
NITROGEN

+/- 40.00 cm3/min


To return to the measurement mode, press AIR STD.COND.
FLOW : 09.36
CANCEL . TEMP. : 25.8°C
ATM.P : 1010.5 hPa

UM-23900B-U User manual ATEQ CDF60 Page 16/40


Chapter 3 – Start up, adjust and measures

3.5. TEMPERATURE UNIT

This parameter selects the temperature unit displayed during the measurements.

+/- 40.00 cm3/min


Enter into the parameters menu, by pressing and AIR STD.COND.
FLOW : 09.36
old the ENTER during 3 seconds. TEMP. : 25.8°C
ATM.P : 1010.5 hPa

The PARAMETER menu is displayed. By using the PARAMETER


STD.COND. : YES
UP and DOWN key select GAZ : AIR
>TEMP.UNIT.: CELSIUS
TEMP. UNIT parameter.

PARAMETER
Then press the ENTER key to modify the FILTER : 0.5 s
PRESS.UNIT.: Pa
parameter (the cursor moves on the right). >TEMP.UNIT.: CELSIUS

UNIT
By using the UP and DOWN key CELSIUS
FAHRENHEIT
choose the unit to be displayed. >KELVIN

PARAMETER
STD.COND. : YES
Validate with the ENTER key. GAZ : AIR
>TEMP.UNIT.: KELVIN

+/- 40.00 cm3/min


To return to the measurement mode, press AIR STD.COND.
FLOW : 09.36
CANCEL . TEMP. : 25.8°C
ATM.P : 1010.5 hPa

UM-23900B-U User manual ATEQ CDF60 Page 17/40


Chapter 3 – Start up, adjust and measures

3.6. BACK LIGHT

The brightness of the backlighting can be set so as to suit the backlighting to the
ambient lighting or your personal preferences.
The screen backlighting is programmable and can be altered.
Backlighting of low brightness will save battery power. Specify a setting that suits the
way you use the instrument.

+/- 40.00 cm3/min


Enter into the parameters menu, by pressing and AIR STD.COND.
FLOW : 09.36
old the ENTER during 3 seconds. TEMP. : 25.8°C
ATM.P : 1010.5 hPa

The PARAMETER menu is displayed. By using the PARAMETER


PRESS.UNIT.: Pa
UP and DOWN key select >BACK LIGHT: 100 %
AUTO OFF: 05 min
BACK LIGHT parameter.

PARAMETER
Then press the ENTER key to modify the PRESS.UNIT.: Pa
BACK LIGHT: 100 % <
parameter (the cursor moves on the right). AUTO OFF: 05 min

PARAMETER
By using the UP and DOWN key
PRESS.UNIT.: Pa
BACK LIGHT: 050 % <
modify the backlight value between 0 % (switched
AUTO OFF: 07 min
off) and 100 % (maximum).

PARAMETER
PRESS.UNIT.: Pa
Validate with the ENTER key. >BACK LIGHT: 050 %
AUTO OFF: 07 min

+/- 40.00 cm3/min


To return to the measurement mode, press AIR STD.COND.
FLOW : 09.36
CANCEL . TEMP. : 25.8°C
ATM.P : 1010.5 hPa

UM-23900B-U User manual ATEQ CDF60 Page 18/40


Chapter 3 – Start up, adjust and measures

3.7. AUTOMATIC POWER OFF (AUTO OFF)

This feature conserve battery life by automatically turning the instrument off after a
user-defined period of time within there has been no key activity.

+/- 40.00 cm3/min


Enter into the parameters menu, by pressing and AIR STD.COND.
FLOW : 09.36
old the ENTER during 3 seconds. TEMP. : 25.8°C
ATM.P : 1010.5 hPa

The PARAMETER menu is displayed. By using the PARAMETER


GAZ : AIR
UP and DOWN key select TEMP.UNIT.: CELSIUS
>AUTO OFF: 05 min
AUTO OFF parameter.

PARAMETER
Then press the ENTER key to modify the GAZ : AIR
TEMP.UNIT.: CELSIUS
parameter (the cursor moves on the right). AUTO OFF: 05 min <

PARAMETER
By using the UP and DOWN key to GAZ : AIR
TEMP.UNIT.: CELSIUS
modify the switch off delay value. AUTO OFF: 07 min <

PARAMETER
GAZ : AIR
Validate with the ENTER key. TEMP.UNIT.: CELSIUS
>AUTO OFF: 07 min

+/- 40.00 cm3/min


To return to the measurement mode, press AIR STD.COND.
FLOW : 09.36
CANCEL . TEMP. : 25.8°C
ATM.P : 1010.5 hPa

It's possible to disable the AUTO OFF function and


leave continuously the instrument on, for that by
using the DOWN key go the minimum PARAMETER
GAZ : AIR
(inferior to 1 minute) then appears "DISABLE" TEMP.UNIT.: CELSIUS
>AUTO OFF: DISABLE
Validate with the ENTER key. The

instrument will be power on continuously.

UM-23900B-U User manual ATEQ CDF60 Page 19/40


Chapter 3 – Start up, adjust and measures

3.8. RS232

This parameter is configuring the RS232 port.

+/- 40.00 cm3/min


Enter into the parameters menu, by pressing and AIR STD.COND.
FLOW : 09.36
old the ENTER during 3 seconds. TEMP. : 25.8°C
ATM.P : 1010.5 hPa

The PARAMETER menu is displayed. By using the PARAMETER


TEMP.UNIT.: CELSIUS
UP and DOWN key select RS232 AUTO OFF: 05 min
>RS232
parameter.

RS232
Then press the ENTER key to enter into the >BAUDRATE : 57600
BITS COUNT : 8
RS232 links parameters configuration menu. PARITY : NONE

Select and adjust each parameter: BAUDRATE,


BITS COUNT, PARITY, HANDSHAKE by using
RS232
the UP and DOWN key and validate BAUDRATE : 19200 <
BITS COUNT : 8
PARITY : NONE
with the ENTER key.

+/- 40.00 cm3/min


To return to the measurement mode, press AIR STD.COND.
FLOW : 09.36
CANCEL . TEMP. : 25.8°C
ATM.P : 1010.5 hPa

UM-23900B-U User manual ATEQ CDF60 Page 20/40


Chapter 4 - CDF60 Functions

Chapter 4
CDF60 FUNCTIONS

1. MENUS STRUCTURE

Extended menu, access by press and hold > 3 seconds

PARAMETER > FILTER > 0,1 to 9,9 s


V
l/h > cm3/min > cm3/s
Flow unit > > cm3/h > in3/min >
in3/s > in3/h >
V
Standard conditions > Yes > Temperature
Atmospheric
V
pressure
V
Gas Air
V Helium
V Nitrogen
V Other
V
Temp. Unit > Celsius
V Fahrenheit
V Kelvin
V
Back light > 0 % > 100 %
V
Auto off > Disable > 99 min
V
2400 > 4800 > 9600
RS232 > Baud rate > > 19200 > 38400 >
57600 >
V
Bits count > 7
V 8
V
Parity > None > Even > Odd
V
Handshake > No / Yes

Simple menu, access by press < 1 second

SPECIAL CYCLE > Auto-zero


Print
Sensor display

UM-23900B-U User manual ATEQ CDF60 Page 21/40


Chapter 4 - CDF60 Functions

2. SPECIALS CYCLES

The specials cycles allows doing commands on the instrument.

2.1. AUTO-ZERO

This special cycle make a zero sensor according to the atmospheric pressure.

Enter into the parameters menu by short press on AIR


+/- 40.00 cm3/min
STD.COND.

the ENTER key.


17.61
Select the AUTO ZERO function by using the UP
SPECIAL CYCLE
and DOWN keys validate with the >AUTO ZERO
PRINT
SENSOR DISPLAY
ENTER key.

The instrument is doing its AUTO ZERO and return


to the measurement mode when the special cycle
is finished. AUTO ZERO

Note: this cycle allows taking over an offset linked to the temperature, position, etc…

UM-23900B-U User manual ATEQ CDF60 Page 22/40


Chapter 4 - CDF60 Functions

2.2. MEASUREMENTS PRINT (PRINT)

This special cycle send on the USB or RS232 port the measures displayed on the
screen.

Enter into the parameters menu by short press on AIR


+/- 40.00 cm3/min
STD.COND.

the ENTER key.


09.36
Select the PRINT function by using the UP SPECIAL CYCLE
AUTO ZERO
>PRINT
and DOWN keys. SENSOR DISPLAY

Frame format sent on the USB or RS232 port:


Simple mode display: +/- 40.00 cm3/min
AIR STD.COND.
FLOW : 17.61 cm3/min
17.61
Extended mode display: +/- 40.00 cm3/min
AIR STD.COND.
FLOW : 15.26 cm3/min FLOW : 15.26
TEMP. : 23.3oC TEMP. : 23.3°C
ATM.P : 1004.9 hPa
ATM.P : 1004.9 hPa

UM-23900B-U User manual ATEQ CDF60 Page 23/40


Chapter 4 - CDF60 Functions

2.3. SENSORS DISPLAY

This special cycle displays the raw values of the all sensors. This is to help the after
sales service diagnostic.

Enter into the parameters menu by short press on AIR


+/- 40.00 cm3/min
STD.COND.
FLOW : 15.26
the ENTER key. TEMP. : 23.3°C
ATM.P : 1004.9 hPa

Select the SENSOR DISPLAY function by using


SPECIAL CYCLE
the UP and DOWN keys validate with AUTO ZERO
PRINT
>SENSOR DISPLAY
the ENTER key.

U : 042727
FLOW : 040713
The instrument displays the sensors points. TEMP. : 016719
ATM.P : 037189

+/- 40.00 cm3/min


To return to the measurement mode, press AIR STD.COND.
FLOW : 15.26
CANCEL . TEMP. : 23.3°C
ATM.P : 1004.9 hPa

UM-23900B-U User manual ATEQ CDF60 Page 24/40


Chapter5 – Accessories, security and recycling

Chapter 5
ACCESSORIES, SECURITY AND RECYCLING

1. ACCESSORIES

1.1. POWER SUPPLY

The power supply of the CDP60


converts a network voltage (120 to 240
V AC) into a 24 V DC low voltage
supply. It has no power switch and
works as soon as it is plugged in.
It is protected against surges and short
circuits via a thermal fuse (Do not use
any other type of fuse).

Moreover, this supply can be plugged on all types of electrical plugs using different
removable pins.
The instrument is not designed to work during the battery charge.

1.2. SOFT CASE

The soft case is to put away the instrument


when out of use.

1.3. USB WIRE

To connect the instrument to a PC.

UM-23900B-U User manual ATEQ CDF60 Page 25/40


Chapter5 – Accessories, security and recycling

1.4. TEST TUBE

To make the pneumatic connection between


the instrument and the flow source to
control.

1.5. PROTECTION COVER (OPTION)

This cover in rubber material allows protect


the CDF60 and absorb impacts (option).

1.6. HARD CASE (OPTION)

Hard case to put away the CDP60


instrument and its accessories.

1.7. NEEDLE VALVE (OPTION)

Needle valves are used to set the leak.


Different needle valves are available,
contact the ATEQ company who
communicate you the best model adapted
to your application.

They must be handled with care especially when they are tightened.

UM-23900B-U User manual ATEQ CDF60 Page 26/40


Chapter5 – Accessories, security and recycling

2. SAFETY INFORMATIONS

Your device and its enhancements may contain small part. Keep them out of the reach of small
children.
Operating environment
Remember to follow any special regulations in force in any area, and always switch off your
device when its use is prohibited or when it may cause interference or danger.
Use the device only in its normal operating positions.

About Charging
Use only the charger supplied with your device. Use of another type of charger will result in
malfunction and/or danger.
Use a specified battery in the equipment.

About the Charger


Do not use the charger in a high moisture environment. Never touch the charger when your
hands or feet are wet.
Allow adequate ventilation around the charger when using it to operate the device or charge the
battery. Do not cover the charger with paper or other objects that will reduce cooling. Do not use
the charger while it is inside a carrying case.
Connect the charger to a proper power source. The voltage requirements are found on the
product case and/or packaging.
Do not use the charger if the cord becomes damaged.
Do not attempt to service the unit. There are no serviceable parts inside. Replace the unit if it is
damaged or exposed to excess moisture.

About the Battery


CAUTION: This unit contains an internal Lithium Ion battery, it is replaceable only by the
competent ATEQ personnel. The battery can burst or explode, releasing hazardous chemicals.
To reduce the risk of fire or burns, do not disassemble, crush, puncture, or dispose of the
battery or the instrument in fire or water, do not short-circuit or connect the contacts with a metal
object.
Use a specified charger approved by the ATEQ manufacturer.

Important instructions (for service personnel only)


CAUTION: Risk of explosion if battery is replaced by an incorrect type. Dispose of used
batteries according to the instructions.
Replace only with the same or equivalent type recommended by the manufacturer.
Use the battery only in the specified equipment.
The battery must be recycled or disposed of properly.

UM-23900B-U User manual ATEQ CDF60 Page 27/40


Chapter5 – Accessories, security and recycling

3. RECYCLING

Do not dispose of the rechargeable Lithium-Ion battery or the tool to


the dustbin.

These components must be collected and recycled.


The crossed-out wheeled dustbin means that within the EU the product
must be taken to separate collection at the product end-of life. This applies
to your tool but also to any enhancements marked with this symbol. Do not
dispose of these products as unsorted municipal waste. For further
information, please contact ATEQ.

UM-23900B-U User manual ATEQ CDF60 Page 28/40


Chapter 6 – Error messages

Chapter 6
ERROR MESSAGES

1. ERROR MESSAGES

The CDF60 can display the following error messages:

The applied flow is too high.


FLOW OUT OF RANGE
Reduce the flow and press on the Over Flow.........
Press [C] to return
CANCEL key. AUTO ZERO

The applied negative flow is too high.


FLOW OUT OF RANGE
Reduce the negative flow and press on the Under Flow........
Press [C] to return
CANCEL key. AUTO ZERO

The "LOW BAT" light flashes, the battery


level is too low for reliable operation.
When this light is on the instrument will be LOW BAT
quickly turned off automatically.
Charge the instrument.

UM-23900B-U User manual ATEQ CDF60 Page 29/40


Chapter 6 – Error messages

UM-23900B-U User manual ATEQ CDF60 Page 30/40


Chapter 7 – PC installation

Chapter 7
PC INSTALLATION

1. PROGRAM AND DRIVERS INSTALLATION

For this device, the connections for update can be carried out by two ways, USB or
RS232 following the user preferences.

1.1. USB CONNECTION

1.1.1. Windows© XP installation

1) Start the installation program:


USB_VIRTUAL_232_XP.EXE, the opposite
window appears.
2) Click on "Next >".

3) The "Serial emulation port" drivers will be


installed, at the end, the opposite window
appears.
4) Click on "Finish".

UM-23900B-U User manual ATEQ CDF60 Page 31/40


Chapter 7 – PC installation

5) Connect the USB wire on the USB port of


your PC and on the CDF60 device.
6) Switch on the CDF60 device.
7) When the CDF60 device is detected, its
installation begins. The opposite window
appears select "No, not this time" and click
on "Next >".

8) Select the "Install the software


automatically (recommended)" option and
click on "Next >".

9) The installation begins…

UM-23900B-U User manual ATEQ CDF60 Page 32/40


Chapter 7 – PC installation

10) When this window appears, click on


"Continue Anyway".

11) The drivers are installed.

12) The "ATEQ RS232 Emulation"


communication port is created; this port will
be used for each communication between the
CDF60 device and the PC through an USB
connector. Click on "Finish".

UM-23900B-U User manual ATEQ CDF60 Page 33/40


Chapter 7 – PC installation

1.1.2. Windows© Vista installation

1) Start the:
USB_VIRTUAL_232_VISTA.EXE, program
the opposite window appears.
2) Click on "Allow", to start the installation.

3) The opposite window appears.


4) Cliquer sur "Suivant >".

5) The "Serial emulation port" drivers will be


installed, at the end, the opposite window
appears.
6) Click on "Finish".

7) Connect the USB cable on the PC USB


port and on the CDF60 device.
8) Switch on the CDF60 device the drivers
will automatically installed.
9) The message at the right bottom corner of
the screen "ATEQ RS232 Emulation"
appears, to confirm the correct CDF60 device
driver's installation.

UM-23900B-U User manual ATEQ CDF60 Page 34/40


Chapter 7 – PC installation

1.2. RS232 CONNECTION

1) To connect the CDF60 to the RS232 port


(PC com1 or com2) it needs the RJ45 wire
and the SubD adapter.

2) Adjust on the CDF60 and in the PC the same RS parameters on each side. For
example:
¾ Baud rate: 19200.
¾ Bits count: 8.
¾ Parity: even.

1.3. RS232 CABLE DIAGRAM

1 5

6 9
SubD (9 pins) 8 4 1
RJ45 (8 pins)
Pin 1 Not used Pin 1 Not used
Pin 2 RXD Data reception Pin 2 Not used
Pin 3 TXD Data emission Pin 3 Not used
Pin 4 Not used Pin 4 Ground
Pin 5 Ground Pin 5 RXD Data reception
Pin 6 Not used Pin 6 TXD Data emission
Pin 7 RTS request to send Pin 7 CTS clear to send
Pin 8 Not used Pin 8 Not used
Pin 9 Not used

SubD9 RJ45

Gnd Gnd
7 2 3 5

7 6 5 4

Tx data Rx data
Rx data Tx data
RTS CTS

UM-23900B-U User manual ATEQ CDF60 Page 35/40


Chapter 7 – PC installation

UM-23900B-U User manual ATEQ CDF60 Page 36/40


Appendix

Appendix
ATEQ CDF60

1. TECHNICALS CHARACTERISTICS

CDF60

Dimensions
152 x 83 x 36
H x L x P (mm) :

Battery: Lithium ion, 12.6 V DC*

Autonomy: About 24 hours

Power supply: concentric Jack


Electrical connexions: Communication : USB and RJ45
(RS232)

Principal connector: quick


Pneumatics connexions: connector Staublï RBE03 female
type.

Display: LCD 4 lines 60 mm x 32 mm

Weight: About 470 g

Temperatures:

Use: + 0°C to + 50°C

Stock: -10°C to + 70 °C

* See the security and recycling instructions about this battery type.

UM-23900B-U User manual ATEQ CDF60 Page 37/40


Appendix

2. PERSONNAL NOTES

________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________

UM-23900B-U User manual ATEQ CDF60 Page 38/40


Index

Index
+ L
+ 24 V DC supply..................................6 Lights ....................................................9
A Low bat light .........................................9
Auto off ...............................................19 M
Auto power off ....................................19 Measurements characteristics ..............4
Auto-zero ............................................22 Measurements print ............................23
B Menus structure ..................................21
Backlight .............................................18 N
Battery ................................................27 Navigation keys ....................................8
Brightness...........................................18 Needle valves .....................................26
C P
C key ....................................................7 Parameters adjust...............................13
Caution ...............................................27 Power supply ................................25, 27
Characteristics....................................37 Presentation .........................................5
Charge light ..........................................9 Print ....................................................23
Charger ..............................................27 R
Charging .............................................27 Recycling ............................................28
Crossed-out wheeled dustbin .............28 RJ45 .....................................................6
D RS232.............................................6, 20
Definition...............................................3 RS232 frame.......................................23
Description de l'affichage....................12 RS232 installation...............................31
Display............................................9, 12 S
E Screen ..................................................9
Electric connector .................................6 Security...............................................27
ENTER key...........................................8 Sensor display ....................................24
Environment .......................................27 Simple menu.......................................21
Error messages ..................................29 Specials cycles ...................................23
Extended menu ..................................21 Standard condition..............................15
F Start up ...............................................11
Filter ...................................................13 Start up key ........................................11
First pneumatic connector ....................6 Std. Cond............................................15
Flow ranges ..........................................4 T
Flow unit .............................................14 Temperature unit ................................17
G TX light .................................................9
Gaz.....................................................16 U
H USB ......................................................6
Hold ......................................................7 USB frame ..........................................23
hold key ................................................7 USB installation ..................................27
I Uses ranges .........................................4
I key......................................................7 W
Installation ............................................6 Windows© Vista Installation ...............30
Windows© XP Installation...................27

UM-23900B-U User manual ATEQ CDF60 Page 39/40


Index

UM-23900B-U User manual ATEQ CDF60 Page 40/40


This document is the exclusive property of ATEQ.