CbDF60a U0Sc
CbDF60a U0Sc
ATEQ CDF60
Version 1.0
Reference: UM-23900B-U
REVISION OF THE CDF60 USER MANUAL
Edition/ Date
Reference Chapters updated
Revision (week/year)
General precautions
• Read the user manual before using the instrument,
• all electrical connections to the instrument must be equipped with a safety
system (fuse, circuit breaker…) appropriate to its needs and complying with the
standards,
• to avoid electromagnetic interference, the cable connections to the instrument
should be less than two meters in length,
• it is essential that the electrical main is earthed,
• disconnect the electrical connections to the equipment before maintenance,
• cut the air supply for any kinds of operation on the pneumatic assembly,
• do not open the instrument when it is powered up,
• avoid water spillage near of the instrument,
• ATEQ is at your disposal for any further information concerning the use of the
instrument under maximum safety conditions.
We would like to bring to your attention that ATEQ will not be held responsible
for any accident connected to the improper use of the instrument, to the work
bench or to the lack of compliance with safety rules.
ATEQ Company is free from any responsibility for any adjustment of its
instrument which would not have been done by its own technicians.
The ATEQ cannot be held responsible if the instrument (program, mechanics or
electronics) has been modified without prior written consent.
0039/Ret-U
ATEQ, THE ASSURANCE OF A COMPETENT AFTER SALES
SERVICE
THE ATEQ AFTER SALES SERVICE IS :
• a team of qualified technicians,
• a permanent telephone assistance,
• agencies close to you for faster reaction,
• a stock of spare parts available immediately,
• a car fleet for rapid intervention,
• a commitment to quality ...
THE OVERHAUL
ATEQ carries out the overhaul of your instruments at interesting prices.
The overhaul corresponds to the maintenance of the instrument (checking, cleaning, replacing of used
parts) as part of preventive maintenance.
Preventive maintenance is the best way to guarantee reliability and efficiency. It allows the
maintenance of a group of instruments in good operational order and prevent eventual break-downs.
MAINTENANCE KITS
The ATEQ After Sales Service proposes, two kits destined for the preventive maintenance of the
pneumatic circuits of instruments.
CALIBRATION
This may be carried out on site or in our offices.
ATEQ is attached to the COFRAC and delivers a certificate following a calibration.
TRAINING COURSES
In the framework of partnership with our customers, ATEQ offers two types of training in order to
optimise the usage and knowledge of our instruments. They are aimed at different levels of technician:
• method / control training,
• maintenance / upkeep training.
A TARGETED TECHNICAL DOCUMENTATION
A number of technical documents are at your disposal to allow you to intervene rapidly in the event
minor breakdowns:
• problem sheets describing and offering solutions to the main pneumatic and electronic
problems,
• several maintenance manuals.
A QUALITY GUARANTEE
The instruments are guaranteed for parts and labour in our offices:
• 2 years for leak detection equipment,
• 1 year for electrical tests to norms instruments,
• 1 year for the accessories.
Our After Sales Service is capable of rapidly answering all your needs and queries.
SAV-Uc/0650
PREFACE
Dear Customer,
You have just purchased an ATEQ instrument, we thank you for the trust you have
placed on our brand. This instrument has been designed to ensure a long and
unparalleled life expectancy, and we are convinced that it will give you complete
satisfaction during many long years of operation.
In order to maximise the life expectancy and reliability of your ATEQ instrument, we
recommend that you install this instrument on a secured workbench and advise you to
consult this manual in order to familiarise yourself with the functions and capabilities of
the instrument.
Our ATEQ After Sales Service centre can give you recommendations based on your
specific operation requirements.
ATEQ
0650/PREFd-U
Table of contents
TABLE OF CONTENTS
PREAMBLE 3
1. ATEQ CDF60 DEFINITION..................................................................................................................3
2. MEASUREMENTS CHARACTERISTICS ...........................................................................................4
2.1. Flow measurement ranges...........................................................................................................4
2.2. Atmospheric pressure and temperature indicators ......................................................................4
Chapter 7 PC INSTALLATION
1. PROGRAM AND DRIVERS INSTALLATION ...................................................................................27
1.1. USB connection..........................................................................................................................27
1.2. RS232 connection......................................................................................................................31
1.3. RS232 cable diagram.................................................................................................................31
Index 39
PREAMBLE
ATEQ CDF60
Pressure
sensor
Auto-zero
valve
the instrument
Connection to
to be tested
Security
Atmosphere
valve
Capillary
The ATEQ Leak/flow calibrator is a flow meter destined for checking leak instruments
and in particular those of ATEQ. It measures a loss of charge using a pressure sensor
placed on a calibrated flow tube. It can also be used for checking a leak flow or a
calibrated jet.
2. MEASUREMENTS CHARACTERISTICS
Flow Maximum
Standard Option (*) resolution
ranges
Accuracy in
conditions 4 ml/min 2% RV + 0,01 ml/min 1% RV + 0,01 ml/min 0,001 ml/min
defined as
ambient
No linearity +
Hysteresis +
Repeatability + T° drift 40 ml/min 2% RV + 0,1 ml/min 1% RV + 0,1 ml/min 0,01 ml/min
+ Long time drift (90
days)
Accuracy in
conditions 4 ml/min 2,5% RV + 0,01 ml/min 2% RV + 0,01 ml/min 0,001 ml/min
defined as
standard
(1013 hPa -
20°C)
No linearity +
Hysteresis + 40 ml/min 2,5% RV + 0,1 ml/min 2% RV + 0,1 ml/min 0,01 ml/min
Repeatability + T° drift
+ Long time drift (90
days)
RV = Read value.
(*) Instrument adjusted in our metrology laboratory with COFRAC agreement.
Accuracy
Minimum Maximum
(20°C)
Note: the zero offset linked to the temperature or the position can be take over by an
auto-zero. It is however advised to make an auto-zero before a measurement.
Recommendation of use: according to some use's conditions, the instrument can put
to two hours of time of heating before complete stabilization.
Chapter 1
INTRUMENT INSTALLATION
Before the first use of the instrument, make a complete battery
charging, wait until the CHARGE light is off.
Pneumatic
Connector
Display
Keyboard
Charging
battery
connector
USB and RS232
connectors
0 V DC + 24 V DC
Chapter 2
USER INTERFACES
1. KEYS
KEY FUNCTION
KEY FUNCTION
KEY FUNCTION
KEY FUNCTION
KEY FUNCTION
2. OTHERS ITEMS
2.2. LIGHTS
Battery level: when this is on, the battery level becoming too low
LOW BAT for reliable operation, indicating that the battery must be
recharged before further use.
Charge light: this light is glowing red while the battery is being
CHARGE
charged. When charging is finished, the light turns off.
TX Not used.
Chapter 3
START UP ADJUST AND MEASURES
The CDF60 can run with the power supply or with the internal battery.
Check the pneumatics and electrics connections before the start up of the CDF60.
2. MEASURES DISPLAY
The instrument displays continuously the current flow. There's no measurement start
cycle.
There are two display modes:
¾ The simple mode where only the current flow is displayed.
+/- 40.00 cm3/min
AIR STD.COND.
09.36
¾ The extended mode where all the measures are displayed: flow, internal
instrument temperature and atmospheric pressure.
+/- 40.00 cm3/min
AIR STD.COND.
FLOW : 09.36
TEMP. : 25.8°C
ATM.P : 1010.5 hPa
To switch from the simple mode to the extended mode, press the UP or the
DOWN keys.
Note: the data sending on the RS232 frame is different according to the display mode
selected, in simple mode, only the current flow is sent, in extended mode, the three
measures are sent.
3. PARAMETERS ADJUST
3.1. FILTER
It can be adjusted from 0.1 to 9.9 seconds by 0.1 second of step. It allows making a
mean on the measurement time, making easier the measurement reading.
PARAMETER
Then press the ENTER key to modify the FILTER : 0.5 s <
FLOW UNIT.: cm3/min
parameter (the cursor moves on the right). STD.COND.: YES
PARAMETER
By using the UP and DOWN key FILTER : 1.0 s <
FLOW UNIT.: cm3/min
modify the parameter to the hoped value. STD.COND.: YES
PARAMETER
>FILTER : 1.0 s
Validate with the ENTER key. FLOW UNIT.: cm3/min
STD.COND.: YES
This parameter selects the flow unit displayed during the measurements.
+ /- 40 .00 c m 3/mi n
Enter into the parameters menu, by pressing and AI R ST D. CO ND.
FLOW : 09.36
old the ENTER during 3 seconds. TEMP. : 25.8°C
ATM.P : 1010.5 hPa
PARAMETER
Then press the ENTER key to modify the FILTER : 1.0 s
FLOW UNIT.: cm3/min<
parameter (the cursor moves on the right). STD.COND.: YES
UNIT
By using the UP and DOWN key >l/h
cm3/min
choose the unit to be displayed. cm3/s
PARAMETER
FILTER : 1.0 s
Validate with the ENTER key. FLOW UNIT.: cm3/h <
STD.COND.: YES
+ /- 2 40 0 cm 3 / h
To return to the measurement mode, press AI R S TD. CO ND .
FLOW : 0936
CANCEL . TEMP. : 25.8°C
ATM.P : 1010.5 hPa
This parameter allows displaying the measurements values corrected to the standards
conditions values, atmospheric pressure and temperature, to the user preferences.
+ /- 4 0. 0 0 c m 3 / mi n
Enter into the parameters menu, by pressing and AI R
FLOW : 09.36
old the ENTER during 3 seconds. TEMP. : 25.8°C
ATM.P : 1010.5 hPa
PARAMETER
Then press the ENTER key to modify the FILTER : 1.0 s
FLOW UNIT.: cm3/min
parameter (the cursor moves on the right). STD.COND.: NO <
3.4. GAS
This parameter allows calculating the flow measurement for a other gas, default gas:
air.
The avalaible parameters are for the following gas:
¾ Helium (He),
¾ Nitrogen (N2),
¾ Other (the parameter had to be indicated, consult us).
GAZ
>AIR
Validate with the ENTER key. HELIUM
NITROGEN
This parameter selects the temperature unit displayed during the measurements.
PARAMETER
Then press the ENTER key to modify the FILTER : 0.5 s
PRESS.UNIT.: Pa
parameter (the cursor moves on the right). >TEMP.UNIT.: CELSIUS
UNIT
By using the UP and DOWN key CELSIUS
FAHRENHEIT
choose the unit to be displayed. >KELVIN
PARAMETER
STD.COND. : YES
Validate with the ENTER key. GAZ : AIR
>TEMP.UNIT.: KELVIN
The brightness of the backlighting can be set so as to suit the backlighting to the
ambient lighting or your personal preferences.
The screen backlighting is programmable and can be altered.
Backlighting of low brightness will save battery power. Specify a setting that suits the
way you use the instrument.
PARAMETER
Then press the ENTER key to modify the PRESS.UNIT.: Pa
BACK LIGHT: 100 % <
parameter (the cursor moves on the right). AUTO OFF: 05 min
PARAMETER
By using the UP and DOWN key
PRESS.UNIT.: Pa
BACK LIGHT: 050 % <
modify the backlight value between 0 % (switched
AUTO OFF: 07 min
off) and 100 % (maximum).
PARAMETER
PRESS.UNIT.: Pa
Validate with the ENTER key. >BACK LIGHT: 050 %
AUTO OFF: 07 min
This feature conserve battery life by automatically turning the instrument off after a
user-defined period of time within there has been no key activity.
PARAMETER
Then press the ENTER key to modify the GAZ : AIR
TEMP.UNIT.: CELSIUS
parameter (the cursor moves on the right). AUTO OFF: 05 min <
PARAMETER
By using the UP and DOWN key to GAZ : AIR
TEMP.UNIT.: CELSIUS
modify the switch off delay value. AUTO OFF: 07 min <
PARAMETER
GAZ : AIR
Validate with the ENTER key. TEMP.UNIT.: CELSIUS
>AUTO OFF: 07 min
3.8. RS232
RS232
Then press the ENTER key to enter into the >BAUDRATE : 57600
BITS COUNT : 8
RS232 links parameters configuration menu. PARITY : NONE
Chapter 4
CDF60 FUNCTIONS
1. MENUS STRUCTURE
2. SPECIALS CYCLES
2.1. AUTO-ZERO
This special cycle make a zero sensor according to the atmospheric pressure.
Note: this cycle allows taking over an offset linked to the temperature, position, etc…
This special cycle send on the USB or RS232 port the measures displayed on the
screen.
This special cycle displays the raw values of the all sensors. This is to help the after
sales service diagnostic.
U : 042727
FLOW : 040713
The instrument displays the sensors points. TEMP. : 016719
ATM.P : 037189
Chapter 5
ACCESSORIES, SECURITY AND RECYCLING
1. ACCESSORIES
Moreover, this supply can be plugged on all types of electrical plugs using different
removable pins.
The instrument is not designed to work during the battery charge.
They must be handled with care especially when they are tightened.
2. SAFETY INFORMATIONS
Your device and its enhancements may contain small part. Keep them out of the reach of small
children.
Operating environment
Remember to follow any special regulations in force in any area, and always switch off your
device when its use is prohibited or when it may cause interference or danger.
Use the device only in its normal operating positions.
About Charging
Use only the charger supplied with your device. Use of another type of charger will result in
malfunction and/or danger.
Use a specified battery in the equipment.
3. RECYCLING
Chapter 6
ERROR MESSAGES
1. ERROR MESSAGES
Chapter 7
PC INSTALLATION
For this device, the connections for update can be carried out by two ways, USB or
RS232 following the user preferences.
1) Start the:
USB_VIRTUAL_232_VISTA.EXE, program
the opposite window appears.
2) Click on "Allow", to start the installation.
2) Adjust on the CDF60 and in the PC the same RS parameters on each side. For
example:
¾ Baud rate: 19200.
¾ Bits count: 8.
¾ Parity: even.
1 5
6 9
SubD (9 pins) 8 4 1
RJ45 (8 pins)
Pin 1 Not used Pin 1 Not used
Pin 2 RXD Data reception Pin 2 Not used
Pin 3 TXD Data emission Pin 3 Not used
Pin 4 Not used Pin 4 Ground
Pin 5 Ground Pin 5 RXD Data reception
Pin 6 Not used Pin 6 TXD Data emission
Pin 7 RTS request to send Pin 7 CTS clear to send
Pin 8 Not used Pin 8 Not used
Pin 9 Not used
SubD9 RJ45
Gnd Gnd
7 2 3 5
7 6 5 4
Tx data Rx data
Rx data Tx data
RTS CTS
Appendix
ATEQ CDF60
1. TECHNICALS CHARACTERISTICS
CDF60
Dimensions
152 x 83 x 36
H x L x P (mm) :
Temperatures:
Stock: -10°C to + 70 °C
* See the security and recycling instructions about this battery type.
2. PERSONNAL NOTES
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
Index
+ L
+ 24 V DC supply..................................6 Lights ....................................................9
A Low bat light .........................................9
Auto off ...............................................19 M
Auto power off ....................................19 Measurements characteristics ..............4
Auto-zero ............................................22 Measurements print ............................23
B Menus structure ..................................21
Backlight .............................................18 N
Battery ................................................27 Navigation keys ....................................8
Brightness...........................................18 Needle valves .....................................26
C P
C key ....................................................7 Parameters adjust...............................13
Caution ...............................................27 Power supply ................................25, 27
Characteristics....................................37 Presentation .........................................5
Charge light ..........................................9 Print ....................................................23
Charger ..............................................27 R
Charging .............................................27 Recycling ............................................28
Crossed-out wheeled dustbin .............28 RJ45 .....................................................6
D RS232.............................................6, 20
Definition...............................................3 RS232 frame.......................................23
Description de l'affichage....................12 RS232 installation...............................31
Display............................................9, 12 S
E Screen ..................................................9
Electric connector .................................6 Security...............................................27
ENTER key...........................................8 Sensor display ....................................24
Environment .......................................27 Simple menu.......................................21
Error messages ..................................29 Specials cycles ...................................23
Extended menu ..................................21 Standard condition..............................15
F Start up ...............................................11
Filter ...................................................13 Start up key ........................................11
First pneumatic connector ....................6 Std. Cond............................................15
Flow ranges ..........................................4 T
Flow unit .............................................14 Temperature unit ................................17
G TX light .................................................9
Gaz.....................................................16 U
H USB ......................................................6
Hold ......................................................7 USB frame ..........................................23
hold key ................................................7 USB installation ..................................27
I Uses ranges .........................................4
I key......................................................7 W
Installation ............................................6 Windows© Vista Installation ...............30
Windows© XP Installation...................27