4 FLC PEX Editor
4 FLC PEX Editor
1
GUI Generator userinterface, t.i. integrated generator desk menue
HEI High energy image
HIS Hospital information system
H/V Horizontal/vertical
HW Hardware
II Image intensifier
IQ Image quality
IQC Image quality control
IQAP Image quality assurance program
LAT Laterality
LEI Low energy image
LIH Last image hold
LL Left lateral
LLD Left lateral decubitus
LLO Left lateral oblique
LUT Lookup table
M1I Material 1 image
M2I Material 2 image
mFD Mobile flatdetector
MS Microsoft
n. a. Not applicable
NLS Native language support
OFF Spatial Filter (German: Ortsfrequenz Filter)
OGP Organ Program
PA Posterior – anterior
PC Personal computer
PEX Program-Exam-Editor
RAD Radiography
REF Reference image
RIS Radiography information system
RM Roadmap
RL Right lateral
RLD Right lateral decubitus
RLO Right lateral oblique
SDM Selective measurement using dominants
SID Source image distance
STE Soft tissue enhancement
SUB Subtraction
SW Software
TET Total exposure time (for dual energy)
U10 Lithoskop Project
UI User interface
XCS X-ray communication system
XCU X-ray communication unit
2
Program, Film & Exam Editor (PEX)
The Program Film & Exam editor, also referred to as the PEX editor, allows the user to
define organ programs, define exams that consist of one or more organ programs, and
define parameter sets for filming layouts that can be used within exams. Additionally the
setup of Workflow parameters is done in the PEX editor. The output of the editor is a
database, referred sometimes as the PEX database, containing the definitions of the
Exams and Programs plus definitions for the layout of film sheets.
Hospital name
FLC serial no.
System type plus system serial no.
No of available targets (detectors)
XCU / generator configuration parameters
Tube type
Generator type
It is possible to import a PEX database into the PEX editor and merge this new
database with the current database.
It shall be possible to export the database for external modification or transfer to another
system or to save a copy of the current database (backup). The export function is
accessible in all acquisition modes of the PEX editor, including viewing mode. If the FLC
user does not have the CD option, the export function is also possible.
After activating the export function the user gets a list of all writeable devices connected
to the used PC (FLC or user PC with offline PEX editor) which can be used for
export/backup (e.g. local hard disk, CD/DVD writer, USB stick). The user selects the
destination device. If no medium is inserted in the selected drive (e.g. CD/DVD) an error
message is displayed: “Please insert medium into drive”. If there is not enough space
on the destination device an error message is displayed: “Not enough free space to
store database”. If the selected device is ready for writing, a button (“GO”) is enabled
and the user may start the database transfer. The PEX database is copied to the
destination device into a predefined folder (e.g. D:\PEX_backup\). The name of the
4
database copy is the original name extended by creation date and time. The extension
must have the format
“_YYYYMMDD-hhmmss” with the definitions:
• YYYY: year, 4 digits
• MM: month, 2 digits
• DD: day, 2 digits
• hh: hour, 2 digits, range 0 to23
• mm: minutes, 2 digits
• ss: seconds, 2 digits
With this name extension multiple backups are possible on all devices (multisession
CD/DVD). On local hard disk (e.g. drive C: and D:) the number of backups is limited to
10. At least this limitation is valid for the hard disk on the FLC. If the limit is reached, the
user is requested to select one of the stored backups, which is replaced by the new
database copy, or to cancel the export function.
The export function is located on the tab card “Configuration”.
An automatic backup is provided by the PEX. To avoid problems with free space on
storage media (external hard disk, USB stick), the automatic backup is provided only,
if the PEX runs in FLC environment (online). The automatic backup runs in the
background. It is triggered after each editing session, when the user closes the PEX
editor.
If the database was not changed, no backup will be written. The database backup is
stored on the local hard disk in a predefined folder (e.g. D:\PEX_auto_backup\). The
folder must differ from the one used for manual backup. Up to five backups can be
stored in this folder. If the folder contains five backups the next automatic backup will
overwrite the oldest version without informing the user. For automatic database backups
the same naming conventions are used as for manual backups. A restore of a database
stored via automatic backup can be done with the import function.
Test: Check that automatic backup is done in the background after closing PEX;
Check that no backup is done, if the database has not changed;
Check overwrite of the eldest backup version.
5
• Transfer the database to another system
• Modifying the database offline to avoid blocking the system during long editing
sessions.
The first one does not need most of the PEX configuration parameters, because they
are system dependent. But the second one needs these parameters, because they may
influence some Organ Program parameter settings. To satisfy both, the following is
implemented for:
• Export:
The database is stored on an external medium containing all configuration
parameters.
• Import:
The database is read from an external medium with the “old” configuration
parameters. Configuration parameter settings do not avoid a database import.
After successful import an automatic update of the configuration parameters
must be done (e.g. an update function is provided by the FLC host softwar).
When the parameter update has finished, the database has to be checked
against the new configuration parameter settings.
Transfer Database
The export/import function is used to transfer a PEX database from one system to
another.
Database transfers can only be done between systems of the same type. The following
system types are known:
• II-systems (e.g. Iconos MD, Artis U)
• II-systems with mobile flat detector (mFD) (e.g. Luminos TF)
• Iconos FD and RAD-systems
Depending on license settings some Exams/Organ Programs may become inactive
after transfer. Inactive Exam/Organ Programs cannot be selected by the customer.
6
PEX Data Categories
The PEX database supports the storage of different types of data (exam, organ
program and film layout data).
The following data types are supported:
User Data
• customer specific exams, organ programs and film layouts
SIEMENS Data
• SIEMENS default exams, organ programs and film layouts
• IQAP organ programs and exams
• Calibration organ programs and exams
• Service programs for maintenance
• Service programs for regulatory affaires (VA, RoeV §16)
7
PEX Password Protection
PEX Customer Password
If the PEX editor is invoked on the FLC or as an offline tool, the user is asked to enter a
password to run the editor or to select “viewing mode”. The system distinguishes two
kinds of passwords, one for customer and the other for SIEMENS IQ expert mode. In
customer mode, all customer specific organ programs, exams and film layouts are
visible and changeable. In SIEMENS IQ expert mode there is full access to all data
sets. If the user selects “viewing” mode instead of entering a password he can see all
data sets, but he cannot modify them.
The customer mode password is site specific. It can be set and modified by the user.
For usage of the PEX editor offline, inclusive the data modification option, the password
must be stored in the database itself.
In the service UI the customer mode password can be reset without knowledge of the
current password.
TEST: Check that all store functions are disabled in “viewing” mode. Check that all
store functions are enabled when the editor is started in customer mode. Check
that the password can be changed by the user. Check that the password can be
changed in service mode. Check that “SIEMENS” default data sets can only be
modified in SIEMENS IQ expert mode.
8
PEX IQ-Expert Password
There is an optional configuration item under configuration menu to run the PEX editor
tool in the IQ expert mode. Running the PEX editor tool in the IQ expert mode must be
additionally password protected. The password differs from the service UI / customer
PEX editor password. A modification of this password is not supported. The IQ expert
mode password is standardized for all databases and therefore hardcoded in the PEX
editor.
Test: Check that IQ expert password cannot be changed via change password dialog.
9
Structure of the PEX Editor
PEX General Layout
The main view of the PEX editor presents six sub menus containing
• the exam editor, to define and edit Exam datasets,
• the organ program editor, to define and edit OGP,
• the fluoro program editor, to define and edit FPG,
• the film layout editor, to define and edit film layouts
• the editor for parameter modules, to define and edit parameter subsets which are
referenced in other datasets, e.g. OGP
• the editor for configuration data, e.g. language configuration, generator type.
At least a “Close” button is provided at the bottom of the window.
Test: Check that a “Close“- button is available independent of the selected tab card.
10
Close PEX Editor
The user can close the PEX editor at any time. A warning message will pop up if there
are any unsaved data sets. If the user accepts (button “OK” in pop up window), all
unsaved changes he made since the editor was opened are discarded. Otherwise he
may cancel (button “Cancel” in pop up window) the close function. To save the changes
the user has to use the function buttons on the tab cards “Exam”, “Organ program”,
“Fluoro program”, “Film Layout” or “Param. Module.”
Test: Check that a warning message pops up if there are unsaved changes. Check
the functions “OK” and “Cancel” in the pop up window.
Exam Editor
Exam Editor General Layout
The sub menu Exam of the PEX editor presents the menu for exam programming. It
provides all functions to manage user defined exam datasets. The following actions are
possible:
Review the parameters of existing datasets
Create new data sets
Modification of datasets
Deleting of datasets
Remark: No password is requested when modified/new datasets are stored. The
user is requested to enter a password at startup of the PEX editor.
Remark: The datasets displayed in “Organ Program view” and “Examination view”
depend on the current user and the access right defined in Table 2, page 5.
11
Figure 3: Exam editor general layout
Remark: Also for Chinese and Japanese characters a language specific sort order
must be considered.
Test: Check the grouping of OGPs.
12
The exams in the examination view can be sorted by the following order:
• Name: alphabetic order of the exams
• Body Part: The exams are sorted by body part examined and the body parts are
displayed in alphabetic order. The value of the parameter “body part examined” of
the first Organ Program within an exam is used to define the rank for sorting.
• Acq. System: The exams are sorted by the acquisition system and the acquisition
systems are displayed in alphabetical order. The value of the parameter “acquisition
system” of the first Organ Program within an exam is used to define the rank for
sorting.
Remark: Also for Chinese and Japanese characters a language specific sort order
must be considered.
Test: Check the sorting of exams.
Remark: The same Organ Program can be linked multiple times to one exam.
Test: Define OGPs in exam. Define OGPs at different positions.
13
Remove Organ Program from Exam
The user has 2 possibilities to remove linked OGPs from an exam. The user selects an
exam to view the linked OGPs. Then he selects the Organ Program link to be removed
and presses the button “Remove OGP” or hits the “Del”- key on his keyboard. The
system removes the Organ Program link without any further confirmation. It is possible
to have exams without any linked Organ Program in the database during the current
editing session. When leaving the PEX’s exam editor the user is informed about these
empty exams. If the user presses the button “OK” the empty exams are deleted (not
valid for VF10). When he presses the button “Cancel” the PEX’s exam editor is not
closed and the user may continue the editing session. The PEX’s exam editor cannot be
closed as long as there are empty exams in the database (not valid for VF10).
Delete Exam
The button “Delete Exam” on tab card “Exam parameter” is used to delete entire exams.
The selected exam is deleted. After pressing the button “Delete Exam” a confirmation
window pops up to avoid unintended deletion of an exam. This button is active as long
as an exam is selected in “Examination view”. It becomes inactive if an Organ Program
link is selected within an exam.
14
• Camera: The user selects the camera he wants to use for filming/auto filming of
all the printer defined in user settings.
• Filming Layout: VF10: no selection is possible. Only two default film layouts are
defined, one for DFR and one for RAD
VF20: Two film layouts can be selected here. One for DFR and one for RAD
images. The selection depends on the value set for parameter “Camera”.
Default are the first layouts for the selected camera.
If “auto Scale” is selected, the scale factor will be automatically calculated from
the largest image, otherwise a fix factor is to select, which is valid for all
images.
• Add to Favorites: This parameter is used to set or reset the attribute “favorite” to
this exam. Favorite exams can be listed separately in the exam manager.
Default value is “not set”. An exam marked as “favorite” is displayed bold in the
examination view.
Store Exam
If the user has changed some parameters in the tab card “Exam parameter”, he has to
press the button “Store” to save the changes in the database. After selection of an exam
the button “Store” is inactive. It becomes active after the first parameter change and
inactive again after being pressed.
15
Figure 4: Exam editor Organ Program link
Test: Define an exam with more than 3 OGPs and set different values for this
parameter. Check if the parameter settings are handled correctly after successful
exposure.
For Organ Program with acquisition mode “Peri” the link parameter “automatic switch to
next Organ Program after acquisition” cannot be set. Organ Program for “Peri” must
stay active after acquisition.
Test: Check that link parameter “automatic switch to next Organ Program after
acquisition” cannot be set for “Peri” Organ Program.
Check if the parameter settings are handled correctly after successful exposure.
16
Default Exams
Up to 8 exams can be defined as default exams. The user can define an exam as
default exam by selecting tab card “Default Exams” and then double click an exam in
the“ examination view”. The selected exam is shown in the first free entry in the tab card
“Default Exams”. If no free entry is available the user is asked to delete one of the
former defined default exams. The same exam cannot be set twice as default exam. At
least 1 exam must be defined as default “Exam 1”.
Remark: For fluoro systems the default exam mechanism can be used to simulate
an Organ Program selection (on TUI) as known from the generator desk.
Therefore the user has to take care that only Organ Program with the
same "application type" setting are links to such an exam.
Test: Check defining and removing of defaults exams in all possible combinations.
Try to define more than 8 default exams. Try to remove all default exams.
17
Rearrange Default Exams
The buttons “move up” and “move down” are used to rearrange the default exams. This
is done by selecting an entry and pressing one of the two buttons, e.g. select “Exam 5”
and press button “move up”, the entries at “Exam 4” and “Exam 5” are exchanged. If
“Exam 1” is selected, button “move up” is not to use (dimmed). If “Exam 8” is selected or
the entry below the selection is empty, button “move down” is not to use (dimmed).
Empty entries cannot be moved.
Test: Check moving of default exam entries.
Check moving with respect to the restrictions defined above.
18
Figure 6: Exam editor Organ Program parameter
Test: Select different Organ Program links and exams and check the parameters/
parameter classes which are displayed.
19
If no Organ Program is selected when the organ program editor is started, the user gets
a layout as shown above. The only user action which is possible in this situation is to
select an acquisition mode. This may be one of the following:
20
After selection of the acquisition mode the corresponding editor layout is loaded or the
user is requested to select an application type and/or an organ program by name. If
acquisition mode and application type are set. The corresponding icon is displayed in
the field marked green in Figure 7: Program editor general layout.
21
Beside these parameters. For alternative OGP, if defined, the corresponding icons are also shown, see
Figure 7: Program editor general layout
If the user changes the acquisition mode, during a programming session, from one DR
mode to another DR mode (e.g. from “Single Shot” to “Series Native”) or from one DR
RAD mode to another DR RAD mode (e.g. “DR RAD” to “Tomography”), the displayed
parameter values does not change until another OGP is selected. Parameter values
which may conflict with the new acquisition mode change to their default values. Default
values are displayed for the parameters which are new in the selected acquisition
mode. The field “Program Name” will be empty until an OGP is selected or a new name
is typed in.
If no Organ Program and no Acquisition Mode is selected when the organ program
editor is started, then an empty layout (see above) is displayed.
In this case either an Organ Program or an Acquisition Mode can be selected.
The field “Program Name” will be empty until an Organ Program is selected or a new
name is typed in.
Test: Check that all acquisition modes are selectable.
Check that layout changes according to selected acquisition mode.
Check that icon corresponding to acquisition mode/application type is displayed.
Check that “reusable” parameter values do not change with the acquisition mode.
The following list of combinations of Acquisition Modes and Application Types is
22
supported for different systems. If an existing Organ Program will be selected, the
associated combinations of Acquisition Modes and Application Types will be selected
automatically.
Table 4: Available combinations of Acquisition Modes and Application Types in different
systems (II Systems) (shaded area = Future)
O = Option
A = alternate combinations of Acquisition Modes and Application Types (Operating
Mode)
1) only if Detector defect
23
systems (FD Systems) (shaded area = future)
24
O = Option
A = alternate combination of Acquisition Mode and Application Type (Operating Mode)
After selection of the combination of Acquisition Mode and Application Type or an organ
program by name the corresponding editor layout is loaded.
25
Select OGP
There are different ways to select an Organ Program for modification:
• Select an Organ Program in Organ Program view of the tab card “Exam” and
switch to tab card “Organ Program”. The selected Organ Program is loaded into
the editor;
• Double click an Organ Program in Organ Program view of the tab card “Exam”,
the PEX automatically switches to tab card “Organ Program” and the selected
Organ Program is loaded into the editor;
• Select an Organ Program from drop down list “Program Name” on tab card
“Organ Program”, the selected Organ Program is loaded into the editor. The
content of the drop down list depends on the grouping criterion defined on tab
card “Exam” for OPG view. This means, that all Organ Programs which are
accessible by the current user (customer or IQ-expert) are shown in the list.
The order of the Organ Program in the list is defined by the grouping criterion
currently set for Organ Program view (tab card “Exam”). The currently selected
Organ Program (if any) is highlighted in the list. If no Organ Program is selected
(neither on tab card “Organ Program” nor on tab card “Exam” in OPG view) the
top of the list is displayed, no Organ Program is highlighted.
Moving the cursor inside the field “Program Name”, by clicking the left mouse
button, it switches from selection to edit mode for program name.
• If an Organ Program is already selected on tab card “Organ Program”, the
buttons “previous” and “next” are used to switch to the previous or next Organ
Program of the currently active drop down list for “Program Name.”
The parameters “Body Part Examined”, “Acquisition Mode” and “Application Type”
cannot be used for Organ Program selection.
Store OGP
The button “Store” is used to save the parameter settings of the Organ Program
currently displayed under the name given in “Program Name”. The PEX editor must not
check any Organ Program parameters against each other. But the PEX user interface
has to ensure, that only valid parameter values can be selected/entered by the user. A
parameter value is regarded as valid,
if the current value is inside its range, and
if the conditions defined in the data tables are fulfilled.
For each parameter a default value must be defined. It is not possible to store
incomplete Organ Programs in the database.
26
storing is attempted.
Create OGP
The button “Store as new” is used to create a new Organ Program. The system checks
the Organ Program name. If it exists already in the database the user is requested to
enter a new name. The user has to press the button “Store as new” again to confirm the
changed Organ Program name. If the new name is accepted the system checks the
parameter settings as described above (see button “Store”). For languages is defined:
• online version: use language configured on FLC
• offline version: use language selected in PEX configuration
• general:
o customer defined names must not change with language
reconfiguration;
o SIEMENS default Organ Program must have a unique postfix, e.g. “_S”.
o This postfix cannot be set by the customer.
Test: Check the creation of new OGPs with valid and invalid parameter combinations
and with new and already existing program names;
Check usage of postfix.
Print OGP
The button “Print” is used to print the currently displayed Organ Program on the paper
printer. The format of the printout is a screenshot of the current layout of the program
editor. For online version “Print” is only possible with license “Paperprint”, For
offline version button “Print” is enabled, printing depends on the local computer settings.
Test: Make paper prints of all possible layouts of the program editor and compare
the printout with the screen display.
Delete OGP
The button “Delete” is used to delete the currently displayed OGP/dataset from the
database. After pressing button “Delete” a window pops up which must be confirmed by
the user before deletion. It is be possible to delete the last Organ Program of an
acquisition mode; in other words, there has to be at least one Organ Program for each
acquisition mode in the database. After deletion of an Organ Program all exams must
be checked and the links to this Organ Program must be removed. If Organ Program
deletion results in an empty exams, this exam will be handled as described above (see
remove OGP).
Test: Delete datasets, confirm and cancel the deletion. Check that it will not be
possible to delete the last Organ Program for the selected acquisition mode
27
Goto Previous / Next OGP
The buttons “previous” and “next” are used to switch to the previous or next Organ
Program according to the grouping selection defined on tab card “Exam” for “Organ
Program view”. If the first Organ Program is selected, the button “previous” becomes
inactive, if the last Organ Program is selected, the button “next”
becomes inactive.
Test: Change to previous/next Organ Program.
Check button states with first/last Organ Program selected.
Test: Define alternative OGPs, switch to the alternative OGP and check its alternative
OGP;
Check deletion of alternative OGP definitions.
For some Organ Programs alternate Organ Programs can be defined (only within the
Category RAD). The definition of alternate Organ Programs is limed to the combinations
of Acquisition Modes and Application Types of the Organ Program: Table 4 and Table 5
describe the combinations of Acquisition Modes and Application Types (marked with
“A”), which allows to define alternate Organ Programs.
Up to 3 alternative Organ Programs for the currently selected Organ Program with the
same or a different combinations of Acquisition Modes and Application Types can be
defined.
28
An alternate Program can be selected by choosing an Organ Program with the
dropdown list of alternate Organ Programs.
This dropdown list contains all Organ Programs for a preselected combinations of
Acquisition Modes and Application Types plus an empty entry to remove an already
selected alternate Organ Program.
A selection of an Organ Program as an alternate automatically links this Organ Program
to the current selected Organ Program as alternative Organ Program and vice versa. If
the “empty” Organ Program is selected, the link between the Organ Program and
alternative Organ Program is removed. (zu löschen)????
The displayed Organ Programs in the drop down list is filtered (optional) according the
combinations of Acquisition Modes and Application Types, the filter criteria are
displayed by two drop down button lists.
The drop down lists for combinations of Acquisition Modes and Application Types will
contain all optional combinations plus an empty entry to disable the filter criteria. If the
drop down list for the filter of combinations will be opened, the list shows all possible
alternate combinations of Acquisition Modes and Application Type (see Table 4 and
Table 5) for currently active Organ Program on the active system.
After selecting of an entry for the filter of the combinations of Acquisition Modes and
Application Types which differ to the currently combinations of Acquisition Modes and
Application Types for the alternate Organ Program, the current linked alternate Organ
Program will be removed, the content of the drop list of Organ Programs (if opened)
changes to the newly selected combination of Acquisition Mode and Application Type.
Test: Define alternative OGPs, switch to the alternative Organ Program and check
its alternative OGP;
Check deletion of alternative Organ Program definitions.
If the user selects the button for displaying the content of the alternate Organ Program,
the alternative Organ Program is loaded into the editor. If there are unsaved changes to
the Organ Program the user is asked to save the Organ Program before switching to
the alternative Organ Program. All fields in the editor are updated and the former Organ
Program is shown as alternative for the current one.
29
DFR Acquisition Mode
Acquisition Mode DR Single /DR Series
Selecting “Series Native” in the field acquisition mode activates the display of the organ
program data for single shot /series native in the program editor. The data of the first
organ program for the selected mode is displayed. The former acquisition mode “Single
Shot” is regarded as special case of acquisition mode “Series Native”. One editor layout
can be used for editing single shot and series native OGP. The user switches between
single shot and series acquisition by selecting / deselecting the checkbox “Single Shot”
in the parameter section of the organ program editor. If a single shot OGP is currently
selected and the user deselects the checkbox “Single Shot”, all parameters of the single
shot OGP which are also valid for series native OGP stay unchanged, other parameters
show there default values. The same is true, if the user switches from series native to
single shot.
If the checkbox “Single Shot” is activated the tab card “Organ program” will have the
following contents:
30
If the checkbox “Single Shot” is deactivated the tabcard “Organ program” will have the
following contents:
Figure 11: Organ Program Editor (Series Native, Variable Frame Rate)
31
If checkbox “Fixed Frame Rate” is activated by the user in a series native OGP, the
contents of the editor will change as shown in Figure 12 below:
Figure 12: Organ Program Editor (Series Native, Fixed Frame Rate)
32
Acquisition Mode DSA
Selecting “Series DSA” in the field acquisition mode activates the display of the organ
program data for series DSA in the program editor. The data of the first organ program
for the selected mode is displayed.
For acquisition mode “DR Series DSA” the tab card “Organ program” will have the
following contents:
33
Acquisition Mode Peri- /Phlebostepping
Selecting “Peristepping” in the field acquisition mode activates the display of the organ
program data for peristepping in the program editor. The data of the first organ program
for the selected mode is displayed.
For acquisition mode “DR Peri” the tab card “Organ program” will have the following
contents:
34
Figure 15: Organ Program Editor (DR Peristepping)
35
Acquisition Mode DR Scan
Selecting “DR Scanning” in the field acquisition mode activates the display of the organ
program data for DR Scanning in the program editor. The data of the first organ
program for the selected mode is displayed.
For acquisition mode “DR Scanning” the tab card “Organ program” will have the
following contents:
36
Figure 17: Organ Program Editor (DR Scanning)
37
RAD Acquisition mode
Acquisition Mode DR Rad
Selecting “DR RAD” in the field acquisition mode activates the display of the organ
program data for DR RAD in the program editor. The data of the first organ program for
the selected mode is displayed.
For acquisition mode “DR RAD” the tab card “Organ program” will have the following
contents:
Screenshot has to be updated according to „no change“ function for flip and rotate
38
Acquisition Mode DR Rad Tomography
Selecting “Tomography” in the field acquisition mode activates the display of the organ
program data for Tomography in the program editor. The data of the first organ program
for the selected mode is displayed.
For acquisition mode “Tomography” the tab card “Organ program” will have the
following contents:
Screenshot has to be updated according to „no change“ function for flip and rotate.
39
Acquisition Mode DR Rad Ortho
Selecting “Ortho” in the field acquisition mode activates the display of the organ
program data for Ortho in the program editor. The data of the first organ program for the
selected mode is displayed. An Ortho program is a set of 4 organ programs, “Ortho
Program 1” to “Ortho Program 4” (this may be more in the future). The user has to
define and store the Organ Program “Ortho Program 1” first, before he can program
“Ortho Program 2” to “Ortho Program 4”. The settings in “Ortho Program 1” are used as
defaults for the other Organ Programs of this Ortho program set. Restrictions on
parameter settings for “Ortho Program 2” to “Ortho Program 4” are defined in the data
section below. The buttons “Store”, “Store as new”, “Print”, “Delete”, “previous” and
“next” always work on a whole Ortho program set. Single Ortho Organ Program cannot
be stored or deleted. For Ortho programs the same parameter set as for acquisition
mode “DR RAD” is used. The user may switch between the Organ Program of an Ortho
program set by selecting the corresponding sub-tab card in the organ program editor.
Each sub-tab card shows a complete set of parameters.
For acquisition mode “Ortho” the tab card “Organ program” will have the following
contents. Only the main part is shown in the figure below:
Figure 20: Organ Program Editor (DR Ortho)
Screenshot has to be updated according to „no change“ function for flip and rotate
Test: Check the contents and the possible parameter settings
40
Acquisition Mode DR Dual Energy
For dual energy acquisition an Organ Program set is needed. The first two Organ
Programs in this set define the parameter settings for the high energy and low energy
exposures. In the third and fourth Organ Program only image processing parameters
are defined for calculating the soft tissue and bone image. These two images are
calculated from the two exposures taken with Organ Program 1 (high energy) and
Organ Program 2 (low energy).
A dual energy acquisition results in 4 images, two exposed and 2 calculated. For dual
energy programs the same parameters as for acquisition mode “DR RAD” are used.
The user may switch between the Organ Program of a dual energy program set by
selecting the corresponding sub-tab card in the organ program editor. Each sub-tab
card shows a complete set of parameters. In sub-tab card 3 and 4 only image
processing parameters can be set, the others are dimmed and not settable.
For acquisition mode “Dual Energy” the tab card “Organ program” will have the following
contents. Only the main part is shown in the figure below:
Figure 21: Organ Program Editor (Dual Energy)
Screenshot has to be updated according to „no change“ function for flip and rotate
Test: Check the contents and the possible parameter settings
41
Acquisition Mode DR Cassette free
Selecting “DRAD Cassette” in the field acquisition mode activates the display of the
organ program data for cassette in the program editor. The data of the first organ
program for the selected mode is displayed.
For acquisition mode “DRAD Cassette” the tab card “Organ program” will have the
following contents:
42
Fluoro Program Editor
Fluoro Program Layout
If at least one DFR Acquisition Mode is available (e.g. Acquisition Mode DR Series)
then Fluoro Program Sets are available in the PEX (e.g. available for all DFR systems,
not for RAD systems like NRF).
Selecting the tab card “Fluoro program” in the PEX activates the display of the fluoro
program editor. The data of the first fluoro program set is displayed if no fluoro program
set is selected in the Organ Program currently loaded in the tab card “Organ program”.
Otherwise the data of the fluoro program set of the Organ Program currently loaded in
the tab card “Organ program” is displayed. The fluoro program editor provides two
pages, one for the fluoro programs (automatic 1, automatic 2, automatic 3) and one for
the roadmap program.
For fluoro programs the tabcard “Fluoro program” will have the following contents:
43
Figure 24: Fluoro Program Editor (customer mode)
44
Roadmap Program Layout
If the DSA Acquisition Mode is available then the Roadmap is available in the PEX
(e.g. available for all DFR systems with DSA, not for RAD systems like NRF).
Selecting the button “Roadmap” in the fluoro editor activates the display of the roadmap
program data. The data of the roadmap program of the currently selected fluoro
program set is displayed.
The fluoro editor’s page for roadmap will have the following contents:
45
Figure 26: Fluoro Program Editor Roadmap (customer mode)
Test: Check the layout for the contents and the possible parameter settings in both user
modes (IQ expert and customer).
The buttons “Roadmap” on fluoro page and “Fluoro” on roadmap page are used to
toggle between these two pages. The button “Roadmap cannot be used without DSA
option.
46
General Functions
Store FPG
The button “Store” is used to save a complete fluoro program set in the database. The
set consists of a fluoro triplet (automatic 1, automatic 2, automatic 3) and a roadmap
program (only with DSA option installed). The dataset is stored under the name given in
the field “Fluoro Program Set Name”. The button can be used on both pages, but
always the complete set is stored. The user may also change the name of the fluoro
program set by typing a new name in the field “Fluoro Program Set Name” and pressing
the button “Store”.
Test: Check the functionality of the button. Press the button on both pages and
Check that all modifications done for fluoro and roadmap are stored.
Change the name of a fluoro program set and check that the new name is
displayed in the Organ Program referring this fluoro program set.
Create FPG
The button “Store as new” can only be used, if the PEX runs in IQ expert mode. This
function is used to create up to 10 fluoro program sets. The new dataset is stored under
the name given in the field “Fluoro Program Set Name”. The create request is rejected
by the PEX if the program set name already exists. If 10 datasets are stored the button
becomes inactive (dimmed). It becomes active again if at least one dataset is deleted
from the database.
Print FPG
The button “Print” is used to print the currently displayed fluoro program set on the
paper printer. The format of the printout is a screenshot of the currently displayed fluoro
program set. With DSA option installed two pages are printed, one with the fluoro page
and one with the roadmap page of the fluoro program editor, otherwise only the fluoro
page is printed.
Test: Make paper prints with and without DSA option and compare the printout with the
screen display.
Delete FPG
This function has been removed.
47
Film Layout Editor
Display the film layout tabcard
Display the current values of the following parameters:
o Destination (default: empty)
o Name (default: disabled and empty)
o Type (default: disabled and empty)
o Size (default: disabled and empty)
o Division (default: disabled and empty)
o Text size (default: medium)
o Enlarge patient name (default: no)
o Display the image text to be filmed
Test: Check that the PEX editor can be started and the film layout tab card can be
activated.
48
Destination
• The default destination will be displayed
• All available destinations can be selected in the combo box
• If a new destination will be selected the default DFR film layout and all
associated parameters (type, size and division) will be displayed
• The auto film destination (selected in auto film configuration – user settings) is
marked with an asterisk (*)
Test: Check that the default destination will be displayed after starting PEX editor.
Check that all available destinations can be selected.
Check that the auto film destination is marked with an asterisk.
Check that all associated parameters will be updated when selecting new
destination.
Name
• If a new destination will be selected the default DFR film layout will be displayed
and the combo boxes “Type”, ”Size”, “Division” and “Text size”, the check box
“Enlarge patient name” and the image text (VF20) configuration will be enabled
and all parameters associated to the actual selected film layout will be
displayed:
o Type
o Size
o Division
o Text size
o Enlarge patient name
o Image Text
• All available film layout names can be selected in the combo box
• If a new film layout will be selected all concerned parameters will be updated
• The default film layouts (DFR and RAD) of the current selected destination are
marked with an asterisk (*)
Test: Check that all available film layout names can be selected.
Check that all associated parameters will be updated when selecting a new
layout name.
Type
• “DFR layout” or “RAD layout” can be selected in the combo box
• If a layout type different from the layout type associated to the actual film layout
will be selected the combo boxes “Size” and “Division" will get empty and the
text configuration (valid for VF20) will get empty and adapted to the actual
layout type.
Test: Check that only DFR layout or RAD layout can be selected.
Check that all associated parameters will be updated when selecting a new
layout type.
49
Size
If a film layout with a non- modified layout type is selected, the size
associated to the actual film layout will be displayed. If not, the combo box
is empty and disabled.
All available sizes can be selected in the combo box.
If a size different from the size associated to the actual film layout will be
selected, and the new size does not support the actual division, the combo
box “Division" will get empty
Division
• If a film layout with a non-modified layout type and a size which supports the
actual division is selected, the division associated to the current film layout will
be displayed. If not, the combo box is empty and disabled.
• All available divisions can be selected in the combo box
Text Size
• The text sizes “standard”, “small” and “large” can be selected in the combo box.
The corresponding font sizes are defined in. Default setting for text size is
“standard”.
• Activating checkbox “enlarge patient name”, the patient name is printed larger
than the other text items.
• Deactivating checkbox “enlarge patient name”, the patient name is printed in
the same size as the other text items.
50
Group 1, upper left image corner:
Patient name
Patient ID
Date of birth
Ward
Request ID
Accession no.
51
Test: Check that all text associated to the current selected layout type can be
selected but no more.
For anonymous filming the checkbox “Anonymous print” is to select. With selection of
this checkbox the entries for patient demographics are automatically deselected and
displayed dimmed (selection disabled). Patient demographics are:
• Patient name
• Patient ID
• Date of birth
• Request ID
• Accession no.
Instead of patient name the string “Anonymous” is printed on film sheet. The other
patient demographics are not printed on film sheet, they are handled as unselected.
Figure 28: Film Layout Editor, Anonymous Print
Test: Check that patient demographics are not visible on film preview and film sheet
for anonymous print;
Check that patient demographics cannot be selected or anonymous print.
52
Set as default
• The button is always enabled
• If the button will be pressed the actual film layout will be set as the default
Film layout (of the actual type) of the current selected destination and the asterisk of
the corresponding layout name combo box will be updated.
Test: Check that the new layout gets the asterisk from the old one.
Create new
• The button is always enabled
• If the button is pressed and the current film layout has been modified a warning
message will be displayed: “All modifications will be lost! Continue? Yes/No”
• If “No” has been chosen the function will be aborted
• If the function has been activated all combo boxes and all text configuration
checkboxes will be cleared
• The new film layout can be stored if an unique name has been entered and all
Combo boxes show values
Save
• The button is disabled until
o a film layout has been modified and all combo boxes show values
o a film layout has been created new, a unique name has been entered and all
combo boxes show values
• With pressing this button the actual film layout will be overwritten or created
Test: Check that all values will be stored correct and no other layouts will be affected.
Remove
• The button is disabled until a film layout has been selected
• With pressing this button a confirmation dialog pops up: “Do you really want to
remove this film layout? Yes/No”
• If “No” has been chosen the function will be aborted
• If the function has been activated the current film layout will be removed
53
Parameter Module Editor
The tab card “Param. Module” has the contents shown in Figure 29. The following
parameter modules can be defined:
• Body Part Examined: the body parts examined which are selectable in the
Organ Program are defined here. The default values are part of the PEX default
database and cannot be deleted or modified by the customer. The body parts
examined are shown in alphabetic order, default values are on top and
customer defined values are listed below.
• Stand Position: stand positions can be stored only if the PEX runs online. Only
valid stand position (indicated in XCU telegram) can be stored. The user is
notified, if the current stand position is invalid.
• Spatial Frequency Parameter
• Gradation Parameter
• EXI Parameter: this module defines the upper and lower limit for controlling the
clinical EXI value.
Parameter types and ranges are defined in the data section of this document. The
“Gradation Parameter” modules cannot be changed in customer mode in VF10. The
buttons in each module have the following meaning:
• Button “delete”: The currently selected entry is deleted from the data base
without further user request. Parameter set which are still in use (referenced in
any OGP) cannot be deleted. A corresponding user message will be displayed.
• Button “replace”: After modification of the selected entry, this button is used to
store the changes. For the modules “Spatial Frequency Parameter” and
“Gradation Parameter” the parameter “Identifier” cannot be modified by
pressing this button.
• Button “new”: This button is used to create a new entry for a parameter module.
The editor checks, if name and identifier (if present) are not jet defined for the
module. In customer mode the parameter identifier is automatically set by PEX
(prefix “c” followed by the next free number). Deleted, former identifier values
are re-used. In IQ-expert mode the identifier must be set by the user.
54
Figure 29: Parameter Editor
Test: For all modules check the function buttons as described above.
Check that the same name cannot be used twice in a module.
Check that the same identifier cannot be used twice in a module.
Check that the identifier cannot be changed using the button “replace”.
55
Configuration Editor
Configuration Parameter
The PEX editor offers a configuration tab card to define the system on which the
database is used. Also the functions for database import and database export are
located on this tab card. The settings on this tab card are used by the editors (e.g.
Exam, OGP) for parameter checks and restrictions on parameter ranges (e.g. available
acquisition modes). If some acquisition modes are not available in the system (e.g. no
wall stand or hardware defect), the PEX editor should indicate this on the UI (e.g. Organ
Program or exams for wall stand are displayed dimmed). But these exams and Organ
Program are still accessible by the user. The parameters and functions on this tab card
are grouped by the definitions below. There is no need to have a description of the
entire system. Only those parameters, which have direct impact on the data sets
managed by the PEX, are listed on the tab card.
For ranges and values of the parameters see [2], [4] and [8]. The parameter groups are:
• System identification parameter
• X-ray system description parameter
• FLC description parameter
• Database import and export functions
Test: Check that the parameter and function groups described above are available on
Tab card “Configuration”
Test: Check that the defined parameters are visible on the tab card “Configuration”.
Check that the defined parameters are automatically set after system startup
(online mode).
Check that parameters are updated online if changed (e.g. hospital name)
56
X-ray system description parameter
These parameters describe the system configuration. The settings are used by the PEX
editor for parameter checks and range settings. If the PEX runs online the parameters
are set or updated by the FLC application after system startup. If the PEX runs offline
these parameters are set by the user, e.g. to define a new database for a specific
system. Parameter in this group are:
Test: Check that the defined parameters are visible on tab card “Configuration”.
Check that the defined parameters are automatically set after system startup (online mode).
57
Remark: The access for the customer to functions or editor pages are handled via
license options and not via system types, like Iconos MD. E.g. Organ
Program for DSA or Peri can only be defined in customer mode, if these
license options are set.
o License options:
The PEX holds a list of all possible PEX data relevant licenses. These
licenses are: DSA, Peristepping, TomoFL-C, ScantechnikFL-C,
DicomSend, Paperprinter (additional licenses are available for NRF,
e.g. Ortho, Tomo, Dual Energy).
At each system start-up the FLC updates this list by setting a flag for each
license which indicates that this license is currently active in the system. If
the PEX runs offline it is up to the user to mark licenses as active or
inactive by setting/resetting the flags on the configuration page. At startup
time in offline mode the PEX shows the last state of the license settings
stored in the database.
o HCC configuration:
At each system start-up the FLC updates the current HCC configuration in
the PEX database. This is done by setting the values for film format (size)
and division for each camera (destination) defined in the FLC service
settings. In offline mode it is up to the user to set these values correctly.
The default for these values is the last stored configuration in the PEX
database.
Remark: The Table shows the impact of the scale unit setting on the UI (Organ
Program editor). Scale unit “inch” is used on the UI only. Parameter
ranges must be adapted for display (input and output). Values stored in the
database are not in scale unit “inch”.
Test: Check that the defined parameters are visible on tab card “Configuration”.
Check that the defined parameters are automatically set after system startup
(online mode).
Check that parameters are updated online if changed (e.g. license options)
58
Offline Configuration
If the PEX editor runs offline, the parameters on tab card “Configuration” are read only
in viewing and customer mode. In IQ expert mode the parameters can be modified by
the user. Parameters set like hardcopy camera configuration or XCU parameter tables
can be loaded from files (e.g. text files) to the PEX database.
Test: Check the three automatic updates with different databases and system
configurations (e.g. add/remove any license, add/remove BWS).
Test: Check the manual update with different system configurations (e.g. add/remove
any license, add/remove BWS).
Check that the button can only be used in IQ expert mode.
59
PEX Language Support
The PEX must support all languages which will be supported by the FLC. It should use
the same language configuration mechanism as implemented in the FLC application. All
UI tags and tool tips are implemented using this language file mechanism. For the
naming of the data entries (like exam, Organ Program or FPG) the following is defined:
60
Fluoro / Organ Program Data
Organ Program Data
Organ Program Parameter and Ranges for DFR
Programs (all DFR Acquisition Modes)
Simple dependencies shall be considered for the input of organ program entries:
Facts:
• Max mAs: 800
• Min ms = 20
• Max ms = 500 with FD
• Max ms = 5000 with cassette
Calculations:
The maximum mAs value also depends on the used focus and the tube load. It can be
calculated by multiplying the maximum current with the maximum exposure time and
the tube load.
The maximum current can be calculated by dividing the kW (40kW with small focus, 80
kW with large focus) by the used kV but it never exceeds 800 mA.
The maximum exposure time is 500 ms with the detector and 5000 ms with cassette.
Examples:
1. Detector organ program with 80 kV, large focus (80 kW) and 80% tube load
• Maximum current = 80 kW/80 kV = 1000 mA
But the maximum current is limited to 800 mA, so for further calculation 800 mA
has to be used.
• Maximum exposure time for detector is 500 ms
• Maximum mAs = 800 mA * 500 ms * 80% = 320 mAs
2. Detector organ program with 100 kV, small focus (40 kW) and 100 % tube load
• Maximum current = 40 kW/100 kV = 400 mA
• Maximum exposure time for detector is 500 ms
• Maximum mAs = 400 mA * 500 ms * 100% = 200 mAs
3. Tomo organ program with 80 kV, large focus (80 kW) and 80% tube load and 2500
ms tomo time.
• Maximum current = 80 kW/80 kV = 1000 mA
But the maximum current is limited to 800 mA, so for further calculation 800 mA
has to be used.
• Maximum exposure time is 2500 ms
• Maximum mAs = 800 mA * 2500 ms * 80% = 1600 mAs
But the maximum mAs value is limited to 800 mAs
Minimum mAs = 1 mA * 2500 ms = 2,5 mAs
61
The dose settings in the OGPs for all DR modes are specified in uGy/p.
Test: Check that the selectable values for dose are given in uGy/p.
The parameter “autowindow” cannot be set in Organ Program for DSA it is always
“OFF”.
Test: Check that this parameter cannot be set in Organ Program for DSA.
activates the display of the data of the current selected Organ Program. The parameters
and ranges of these Acquisition Modes are shown in the table Table 10.
62
Table 10: DFR- Acquisition Mode Organ Program Parameters (FD and II Systems)
63
64
1) The parameter ranges for these Acquisition Modes are configured via init data. The
configuration updates the ranges to the different functionality supported by the used
subsystem. E.g. the data of available grids for different Acquisition Modes in a
dedicated system is sent to the FLC, then the editor supports the range of data setup.
Remark: The relationship between kV, mAs and ms does not depend on generator
power. Only values from the EP-tables defined in the XCU are selectable.
65
Table 11: Dependencies of Organ Parameters for Exposure Data from Fluoro FD Systems
In an Organ Program only frame rates shown in one line of the table can be used.
Remark: Requirement was updated due to wrong description in the old text
(Category C)
Remark: Future version: 3f/s and 6f/s
Remark: This position is only used for post processing flips because Organ Program
flips will not be displayed as symbol on DFR images
Test: Check all available values, selections and ranges.
If the user defines more than one framerate (“Acquisition Rate”) in an series OGP,
the maximum value for “Scene length” is defined by the highest framerate.
Test: Check that the maximum scene length depends on the highest framerate
selected. The highest framerate shall be set for framerate 1, 2 and 3.
For DR Scan Organ Program parameters the following restrictions are defined:
• VF10: The frame rate is not changeable, fix 2 f/s;
VF20: The frame rate is changeable, allowed values: 2 f/s and 3 f/s.
Remark: 3/fs depends on availability of this conditioning mode (current
restriction in VF10/Iconos FD Phase 1, but planned for VF20/Iconos Phase 2)
• The collimation height is not changeable; default value is defined in service
mode. (Restriction for VF10: 6cm)
• The collimation width is changeable; default value “no change”.
• The scan speed is 4 or 6 cm/s
• Max pulse width is limited to 50 ms, default: 30 ms
• The following combinations of “Scan mode” and “Scan direction” are allowed:
- Spine: .......................................... “feet to head”
- Leg: ............................................. “head to feed”
- Colon:.......................................... both
Test: Check that these restrictions are implemented.
66
Selection of measuring fields for DFR modes
For all DFR modes (e.g. single, series, DSA) the user selects a measuring field
combination by a drop down list. The drop down list contains all possible measuring
field combinations for the selected acquisition mode. The currently selected measuring
field combination is marked in the drop down list. After selection of an measuring field
combination, the drop down list is closed and the selected measuring field is displayed
in the editor.
Remark: Requirement was updated due to wrong description in the old text
(Categorie C)
Test: Change the SDM selection and check if the selected SDM is displayed in the
editor.
For DFR acquisition modes (incl. fluoro) in the FD systems the following measuring field
combinations are offered in a dropdown list:
Figure 30: Elements of the dropdown list for measuring field selection of DFR modes (FD systems)
Remark: Requirement was updated due to wrong description in the old text
(Categorie C)
Remark: Pedeatric not valid for VF10/20
Test: Check that all shown SDM combinations are displayed and can be selected.
For the DFR acquisition modes (incl. fluoro) in II systems the following measuring field
combinations is offered in in a dropdown list:
Figure 31: Elements of the dropdown list for measuring field selection of DFR modes (II systems)
Test: Check that all shown SDM combinations are displayed and can be selected.
67
Organ Program Parameter for Radiography (all
Rad Acquisition Modes)
Selecting the one of the following Acquisition Modes
• DR Rad (FD) Table
• DR Rad (FD) Table lateral
• DR Rad (FD) Wallstand
• DR Rad (FD) Vertical
• DR Rad (FD) free
• DR Rad Tomo (FD)
• DR Rad Ortho (FD)
• DR Dual Energy (FD)
• Rad Cassette Table
• Rad Cassette Wallstand
• Rad Cassette Tomo
• Rad Cassette free
activates the display of the data of the current selected Organ Program. The parameters
and ranges of these Acquisition Modes are shown in table 13.
68
Table 13: Rad Acquisition Mode Organ Program Parameters (mFD / FDp and II Systems)
69
70
71
1) The parameter ranges for these Acquisition Modes are configured via init data. The
configuration will date the ranges to the different functionality supported by the used
subsystem. E.g. the data of available grids for different Acquisition Modes in a
dedicated system are sent to the FLC, then the editor supports the range of data
setup.
2) The relation of Patient, Viewing Position and Display Indicator is described in the
tables in the chapter appendix for the patient position “undefined” and the 16
“defined” patient positions describes the “anatomic correct view” image operation
(green line) in relation to the patient position. Additionally all other image flip and
rotation combination are listed (column PEX editor).
Remark: The relationship between kV, mAs and ms does not depend on tube or
generator. Only values from the EP-tables defined in XCU are selectable.
Remark: OPEN ISSUE for parameter 18: 100cm has not been confirmed by IC;
115cm was the goal. To be checked during IT
72
Selection of measuring fields for Rad modes
The user selects a measuring field combination by a drop down list. The drop down list
contains all possible measuring field combinations for the selected application type. The
currently selected measuring field combination is marked in the drop down list. After
selection of a measuring field combination the drop down list is closed and the selected
measuring field is displayed in the editor.
Test: Change the SDM selection and check if the selected SDM is displayed in the
editor.
For application type of different systems the following measurement field combinations
is offered in the dropdown list.
Remark: Requirement was updated due to wrong description in the old text (Category C)
73
Tomography for Rad modes
For the Acquisition Mode DR Rad Tomo (FD) the Organ Program parameters are the
parameters for DRAD table and the parameter combination for tomo-time and tomo-
angle. Tomo-height cannot be set via Organ Program for Iconos FD systems. For NRF
systems the Tomo-height is adjustable in a range from 3mm to 250 mm. One of the
following parameter combinations can be set for time and angle:
Test: Check that all 6 parameter combinations are available and can be set.
74
Patient Position for Rad modes (FD)
The display indicator is a hint for the user, how the image will be displayed at the
monitor. The display indicators shown in the table are examples and only valid if no
additional flip or rotation is selected. With selection of flip or rotation the display indicator
must change accordingly.
If no patient position is defined, the display indicators may look like described in [6].
These display indicators correspond to the settings for flip and rotate.
75
The following workflow is an example how the setting of these parameters can be done:
• Create new OGP
• Select 1-point technique
• Define parameter:
o kV
o Dominant
o Exposure correction
o Dose
• Select 2-point technique
• Define parameter
o mAs
• Select 3-point technique
• Define parameter
o ms
• Select 1-point technique
• Store Organ Program (Organ Program is stored as 1-point technique)
The values defined for “Dominant”, “Exposure correction” and “Dose” are used for
1-point technique and 3-point technique with AEC.
Test: Define an Organ Program for multi technique support (e.g. with workflow
defined above).
Check that the parameters for all techniques are stored in the database.
Select this Organ Program in examination mode, do acquisition for all
techniques;
Check that the acquisition is done with the parameters defined in Organ
Program for each technique.
76
Table 18: Organ Program Parameter Restrictions for Rad Acquisition Mode DR Rad Ortho (FD) / (FDp)
77
Test: Check restrictions for parameter settings in the Ortho step Organ Program.
78
Table 19: Organ Program Parameter Restrictions for Rad Acquisition Mode Dual Energy OGP
79
Test: Check settings for parameter settings in dual energy Organ Program.
Check restrictions for parameter settings in dual energy Organ Program.
80
Table 20: Fluoro Program Parameters II + FD Systems
81
Remark: RM- Mask Image is stored after 2 sec with transition from phase 1 to phase 2
for all framerates and all fluoro modes.
Remark: Fluoro Harmonization (DDO) is possible up to 30 images/sec
Remark: with m = 0 the dose/pulse is constant
with m = 1 the dose rate remains constant
The FPG parameters dose per pulse, rate index, fluoro curve, copper filter and frame
rate can only be changed if the editor runs in SIEMENS IQ expert mode.
Test: Check that these parameters and can only be changed in SIEMENS IQ expert
mode
82
Gradation Parameter Module
The following parameter can be defined in the Gradation parameter module:
7) System will be removed, as the generation of Organ Programs is not related to the
system (last 3 columns)
Remark: The parameter “Identifier” is only settable by the user, if the PEX runs in
IQ expert mode. If the PEX runs in customer mode, this parameter is
automatically set by the PEX, if button “Store as new” is pressed. In this
case, the prefix is always ‘c’ and the next free number is used to complete
the identifier.
83
Spatial Frequency Parameter Module
The following parameter can be defined in the Spatial Frequency parameter module:
8) System is removed, as the generation of Organ Programs is not related to the system
(last 3 columns)
Remark: The parameter “Identifier” is only settable by the user, if the PEX runs in
IQ expert mode. If the PEX runs in customer mode, this parameter is
automatically set by the PEX, if button “Store as new” is pressed. In this
case, the prefix is always ‘c’ and the next free number is used to complete
the identifier.
84
EXI Parameter Module
The following parameter can be defined in the EXI parameter module:
9) System is removed, as the generation of Organ Programs is not related to the system
(last 3 columns)
Remark: The parameter “Identifier” is only settable by the user, if the PEX runs in
IQ expert mode. If the PEX runs in customer mode this parameter is
automatically set by the PEX, if button “Store as new” is pressed. In this
case, the prefix is always ‘c’ and the next free number is used to complete
the identifier.
10) System is removed, as the generation of Organ Programs is not related to the
system (last 3 columns)
85
Remark: No change for VF10, table was moved from requirement.
Migration of Databases
The system provides means to migrate the data content of current Aristos systems into
the new PEX data structure. This can be done manually (not preferred) or by
implementing a tool to transform the Aristos Organ Program into the PEX Organ
Program data structure.
The tool may be implemented as a simple command line tool without any Windows®
like UI. The resulting PEX database has to be checked by an IQ expert before it can be
used at any system.
Remark: This tool is totally independent from the PEX software and it is not part
of the PEX software delivery. It must not be tested within SSIT or SIT. It
is only a helpful tool for Organ Program authors to get started with the
PEX based on the data currently defined for Aristos systems.
86