National Lab Low Temperature Upright Freezer
National Lab Low Temperature Upright Freezer
ProfiLine Pegasus
User Manual
Low Temperature Upright Freezer
ProfiLine Pegasus
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
printed in Germany
User Manual
ProfiLine Pegasus
Contents
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
printed in Germany
User Manual
ProfiLine Pegasus
1. Packing:
The packing material is entirely recyclable.
Please follow the local regulations for scrapping.
2. Appliance:
The freezer is made of recyclable material. Dispose of the appliance should be
according with the local environmental regulations.
Please deliver the appliance to an authorised centre for recovery of refrigerant.
Information:
This appliance contains a mixture of different refrigerant. For more information,
please refer to the rating plate affixed on the freezer.
All personnel have to be qualified by the operator under consideration of the special
requirements for an operation down to -40 °C / -86 °C. There is a risk of frost-bite if
the contents or parts of the inner container are touched!
In the event of breakdown or irregularities contact professional instructors or repair
technicians immediately.
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
1 printed in Germany
User Manual
ProfiLine Pegasus
2. The freezer should be located in a cool, dry place out of direct sunlight. The
freezer gives off a great deal of heat when in operation. If the ambient
temperature is too high there is a risk that the freezer will not operate
correctly. It is recommended that the room be ventilated so that the room
temperature does not exceed +25 °C. Class N.
3. Place the freezer on a solid and flat substrate. This will eliminate any
vibration and irritating noise. The freezer should be placed with at least
10 cm free to the sides, at least 15 cm free at the back free to the ceiling.
Inner doors:
NOTE: The inner doors should be used at all times. Make sure to close them tightly.
The manufacturer is not responsible for damage to the freezer caused by failure to
do so.
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
2 printed in Germany
User Manual
ProfiLine Pegasus
Electrical Connection
The machine should be given extra protection in accordance with Electricity Supply
Regulations in order to protect the user against dangerous electric shocks in the
event of faults.
Data regarding voltage and absorbed power / current is given on the rating plate
placed on the rear side of the freezer.
Warning:
If the socket and plug are not of the same type either the socket or the plug must be
replaced by a qualified electrician.
Under no circumstances use extension leads or adaptors.
1. Place the freezer on a flat floor that can bear the weight of the freezer
4. The freezer should be placed in a dry location. Do not allow water to come
into contact with the cabinet. Water entering electrical parts may cause short-
circuiting.
Before commissioning
Clean the freezer both inside and out using a damp, wrung out cloth. Dry with a dry
cloth.
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
3 printed in Germany
User Manual
ProfiLine Pegasus
Start procedure
• Put the plug in the socket.
• Turn the unit on by using the on/off switch (password “0000” if necessary)
• It will take about 3-5 minutes before the compressor starts
Start the freezer. Set the digital thermometer at the temperature required. It will now
take about 6-10 hours before the set temperature is reached.
The freezer should be cooled down empty.
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
4 printed in Germany
User Manual
ProfiLine Pegasus
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
5 printed in Germany
User Manual
ProfiLine Pegasus
Setting icons
Custom Settings
Custom Settings (page 7)
Advanced Settings
Advanced Settings (page 8)
Service Settings
Service Settings
Status
Status (page 10)
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
6 printed in Germany
User Manual
ProfiLine Pegasus
Custom Settings
Enter password
The menu Custom Settings is protected with a password.
(Default password is “0000”)
Setpoints
The temperature for the unit will be set.
Alarm Settings
The below menu will be available.
Alarm delay
The setting of the time from an alarm will occur on the unit
and until it will be shown on the display.
Power failure
The alarm displays that there is no power to the unit.
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
7 printed in Germany
User Manual
ProfiLine Pegasus
Advanced Settings
Enter password
The menu Advanced Settings is protected with a password.
(Default password is “0000”)
Calibration
The setting for offset of the temperature in the display.
Automatic defrost
The setting of the time where a defrost automatic will be performed on the unit.
Manual defrost
The manual defrost can be started immediately.
Hysteresis
The setting of the different from the set point where the compressor will start and stop.
Service
See page 9.
Password protection
Set password protection on switching ON or OFF the unit.
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
8 printed in Germany
User Manual
ProfiLine Pegasus
Service
Under this menu, there is information about the unit, which is important for the service
on the unit.
Compressor hours
Displays how many hours the compressor has been running.
Fan hours
Displays how many hours the fan has been running.
Probe
Displays the temperature for the sensors installed in the unit.
Probe/Compressor
Displays the temperature of the sensor mounted on the compressor.
(Not all models).
Probe evaporator
Displays the temperature of the sensor, which is located on the
condenser.
Calibration Settings
Probe 1
Calibration of probe 1.
Probe 2
Calibration of probe 2.
Probe 3
Calibration of probe 3.
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
9 printed in Germany
User Manual
ProfiLine Pegasus
Status
Current alarms
Displays the current alarms with date and time.
Graph
Temperature curve
Use arrow up, will show the graph for the actual day. Using the left arrow, will show up to 10 days left.
Press ESC to return to Main screen.
Change password
It is possible to change the password for Customer Settings, Advanced Settings and
Advanced Service Settings.
Reset password
It is possible to reset the password for Customer Settings, Advanced Settings and
Advanced Service Settings. Contact the manufacturer to get the password for reset
password.
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
10 printed in Germany
User Manual
ProfiLine Pegasus
Download Data
USB connected
Place the USB drive in the connection on the display and press the „right button“ for
download data to the USB drive.
Read data
This symbol will be shown in the display when the data is loading the
USB drive.
Transfer completed
This symbol will be shown in the display when the loading is
finished and the USB drive can be removed from the display.
These two files can be open in Excel: “data00” and
“param00”
USB connected
Turn off the display before start.
Place the USB drive in the connection on the display and press the “left button” for upload data from the
USB drive.
Service options
Choose “Load Program” to upload new software or parameters to the unit.
“Clear SC memory” is only used for units with Card lock mounted and is described in the manual for the
Card lock.
Data upload
This symbol will be shown in the display when the data is loading from the
USB drive.
Transfer completed
This symbol will be shown in the display when the loading
is finished and the USB drive can be removed from the
display.
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
11 printed in Germany
User Manual
ProfiLine Pegasus
Alarms on display
To reset an alarm, press “Enter” on the main screen, and press “OK” for each
alarm.
Alarm icons
Probe/eprom failure
The alarm displays that the probe is not working correctly.
Power failure
The Power failure alarm indicates that the unit has experienced a power failure. The display goes black,
just for saving power from the battery. The battery icon in the display will start flashing, and an alarm bib
will start sounding. Press few seconds on the On/Off button, the display will start light up, and show there
is a Power failure and the actually temperature in the unit.
The display will go black again after few seconds, just to save power from the Battery. When the Power
returns, the display will start lightning up and show the normally set up again, and the compressor will
start up after 1 minutes, and return into normally operating.
Compressor alarm
The Compressor failure alarm indicates that a compressor on unit is malfunctioning. When Compressor
alarm occurs, the controller switch off the specific Compressor relay, for protecting the Compressor to
prevent too high current.
Low battery
The alarm displays that the power on the backup battery is to low.
Auto defrost
The Auto defrost status indicates that an automatic defrost cycle is
running.
Manual defrost
The Manual defrost status indicates that a manually activated defrost
cycle is running.
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
12 printed in Germany
User Manual
ProfiLine Pegasus
Default settings
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
13 printed in Germany
User Manual
ProfiLine Pegasus
Power failure
The ProfiLine Pegasus is equipped with battery back up for display and contact for
remote monitoring only. Back up time is minimum 48 hours. The freezer must be
connected voltage in minimum 72 hours before maximum battery is available.
Please note the lifetime of the battery is estimated to be 3 years, but should be
checked by authorized personnel every 6 months.
The condenser is placed on top of the cabinet behind the control panel.
Thermostat with digital display: once the freezer is cooled down, the set
temperature and the temperature shown will be the same.
If the freezer and its electrical parts are reconstructed, the guarantee and safety
regulations shall cease to apply.
There is no guarantee for the contents of the freezer. There is a 1-year parts
warranty for functional parts related to the freezer.
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
14 printed in Germany
User Manual
ProfiLine Pegasus
Ice will form on the inner doors of the freezer. This ice should be removed
with a scraper at regular intervals. (Remember gloves).
Ice on the inside of the cabinet should be removed as required. Take the
contents out of the freezer. Pull out the plug and leave the door open until the
ice can be removed. (Remember gloves).
Dry the inner cabinet and shelves with a cloth until it is not damp any longer.
For external maintenance use stainless steel polish.
Ice formed on the inside of the door should be removed with a wooden/plastic
scraper. This ice will build up because of the low temperatures inside the
cabinet, and if it is not removed it will cause the door to open.
Cleansing agents containing chlorine or compounds of chlorine as well as
other corrosive means may not be used as they might corrode the stainless
panels of the cabinet and the evaporator system.
All parts of the refrigerating circuit are suitable for continuous operation and
require no maintenance. Only the condenser should be regularly checked and
cleaned. This depends on the degree of contamination of the site (at least
once a year).
Note:
Unplug the appliance before carrying out maintenance.
Spare parts:
When ordering spare parts that you can easily fit yourself, please give unit type
number and the function of the part(s) concerned
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
15 printed in Germany
User Manual
ProfiLine Pegasus
Wiring Diagram
PLPE 0186, PLPE 06xx und PLPE 12xx
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
16 printed in Germany
User Manual
ProfiLine Pegasus
Reg. BPLPEEN-G214-160801 All rights and changes reserved © 2016 Copyright by National Lab
17 printed in Germany
National Lab GmbH
Kälte- und Temperiertechnik
Grambeker Weg 157 • D-23879 Mölln
Tel: +49 4542/8491-10 • Fax: +49 4542/8491-11 • e-mail: [email protected]
EU-Konformitätserklärung
EC-Declaration of Conformity
Das (die) nachfolgend bezeichnete(n) Gerät(e) erfüllt (erfüllen) die Anforderungen der unten
genannten EG-Richtlinien, harmonisierten EN, sowie die nationalen Normen und Richtlinien.
The machine(s) designated in the following complies (comply) with the requirements of the EC directives
and harmonized ENs given below as well as with the national standards and recommendations.
Bei einer nicht mit uns abgestimmten Änderung des (der) Gerät(e)s verliert diese Erklärung ihre
Gültigkeit.
This declaration becomes invalidated if any modification is made to the applicance(s) without our prior
authorization.
Mölln, 04.01.2016
...................................................
Reinhard E. H. Schuldt,
Geschäftsführer Managing Director