CALIBRATION
CALIBRATION
EN
CONTENTS
1. REQUIRED INSTRUMENTS
• Support plate
• Dentition phantom
• Aluminium filter
• Laser reference sheet
Insert the phantom, making sure that the pins coincide with
the holes highlighted in the following images.
Enter the serial number and the part number of the column
in the suitable fields (96600231 in case of floor-mounted co-
lumn, 96600230 in case of wall-mounted column). Click on
OK.
PERFORM THIS OPERATION ONLY IF THE DEVICE REMAINS UNUSED FOR MORE THAN
THREE MONTHS
3 4 5
Repetitions
2
Shot Parameters
kV 70
mA 4
3
kV mA Time (ms) Repetitions
a 60 4 1000 2
b 70 4 1000 2
5 4 c 80 4 1000 2
d 85 4 1000 2
e 85 8 1000 2
f 85 12 1000 2
g 85 12 5000 1
h 68 15 5000 1
3D ONLY
3. 2D CALIBRATION
• Rotation
• Horizontal alignment
• Vertical alignment
Click on “Acquire”.
4. 3D CALIBRATION
• DIAMETER 10
•
• Select the 10x10 field, enter the values -3 and -7
inside the “Left” and “Right” fields, then press “Move”.
10
If the resulting image is like the following one
• 10x7
• 10x6
• 10x10 HiRes
• 10x7 HiRes
• 10x6 HiRes
• DIAMETER 8
Select the 8x10 field, enter the -17 and -18 values in
the “Left” and “Right” fields and press “move”. As soon
as the machine is ready, perform a new acquisition (8
mA and 20 ms).
• 7x7
• 8x6
• 8x10 HiRes
• 8x7 HiRes
• 8x6 HiRes
• DIAMETER 6
Select the 6x7 field, enter the -31 and -28 values in
the “Left” and “Right” fields and press “move”. As soon
as the machine is ready, perform a new acquisition (8
mA and 20 ms).
• 6x6
• 6x7 HiRes
• 6x6 HiRes
then “Start”.
Press “Next”
Press “Next”,
Press CTRL, click the image with the left mouse but-
ton and move it (keeping both keys pressed). Once
the desired position has been reached, release mou-
se button and wait for the Software to warn you to
release also the CTRL button. Once finished, press
the rays button when prompted; the system acquires
a new scout to evaluate the new position.