CALIBRATION

The document is a calibration manual for the Hyperion X5 2D/3D OPT E.L. system, detailing required instruments, preliminary operations, and step-by-step procedures for both 2D and 3D calibration processes. It includes instructions for X-ray alignment checks, sensor calibration, mechanical centering, and laser calibration, as well as daily checks and data backup procedures. The manual is intended for users to ensure accurate calibration and optimal performance of the imaging system.

Uploaded by

Bernard Koffi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
4 views36 pages

CALIBRATION

The document is a calibration manual for the Hyperion X5 2D/3D OPT E.L. system, detailing required instruments, preliminary operations, and step-by-step procedures for both 2D and 3D calibration processes. It includes instructions for X-ray alignment checks, sensor calibration, mechanical centering, and laser calibration, as well as daily checks and data backup procedures. The manual is intended for users to ensure accurate calibration and optimal performance of the imaging system.

Uploaded by

Bernard Koffi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 36

hyperion X5 2D/3D

EN

2D/3D OPT E.L. CALIBRATION MANUAL

Code Rev Date


9707122197071221 01 10.2017

EN 2D/3D OPT E.L. CALIBRATION MANUAL 1


hyperion X5 2D/3D

2 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

CONTENTS

1. REQUIRED INSTRUMENTS ................................................................................................................... 5


2. PRELIMINARY OPERATIONSPRELIMINARY OPERATIONS ............................................................... 7
2.1 TUBE HEAD WARM-UP.................................................................................................................... 9
3. 2D CALIBRATION.................................................................................................................................. 12
3.1 X-RAY ALIGNMENT CHECK........................................................................................................... 12
3.2 SENSOR CALIBRATION CHECK.................................................................................................... 17
3.3 VERIFYING THE MECHANICAL CENTERING .............................................................................. 18
3.4 LASER CALIBRATION .................................................................................................................... 20
3.5 CALIBRATION DATA BACKUP ....................................................................................................... 22
4. 3D CALIBRATION.................................................................................................................................. 23
4.1 RAY BEAM CALIBRATION.............................................................................................................. 23
4.2 DAILY CHECK AND BLANK ACQUISITION.................................................................................... 29
4.3 TIN SCAN........................................................................................................................................ 31
4.4 QA PHANTOM ACQUISITION (OPTIONAL)................................................................................... 34
4.5 CALIBRATION DATA BACKUP........................................................................................................ 36

EN 2D/3D OPT E.L. CALIBRATION MANUAL 3


hyperion X5 2D/3D

Page left intentionally blank

4 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

1. REQUIRED INSTRUMENTS

Calibration kit, including:

• Support plate
• Dentition phantom
• Aluminium filter
• Laser reference sheet

• Allen wrench set


• Screwdriver set

When you will be required to assemble the support plate and


the dentition phantom, remove the chinrest from the machi-
ne and insert the highlighted pins into the holes where the
chinrest was assembled.

EN 2D/3D OPT E.L. CALIBRATION MANUAL 5


hyperion X5 2D/3D

Insert the phantom, making sure that the pins coincide with
the holes highlighted in the following images.

6 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

2. PRELIMINARY OPERATIONSPRELIMINARY OPERATIONS

Open the acquisition server software, click on “Settings”,


then on “Change Mode” and confirm when prompted.

Select “Service mode” and enter the password “cefla”.

Open the Acquisistion server software and click on “Calibra-


tion”.

EN 2D/3D OPT E.L. CALIBRATION MANUAL 7


hyperion X5 2D/3D

Enter the serial number and the part number of the column
in the suitable fields (96600231 in case of floor-mounted co-
lumn, 96600230 in case of wall-mounted column). Click on
OK.

8 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

2.1 TUBE HEAD WARM-UP

PERFORM THIS OPERATION ONLY IF THE DEVICE REMAINS UNUSED FOR MORE THAN
THREE MONTHS

EN 2D/3D OPT E.L. CALIBRATION MANUAL 9


hyperion X5 2D/3D

3 4 5

Repetitions
2

10 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

Shot Parameters

kV 70

mA 4

Time exposure (ms) 1000

3
kV mA Time (ms) Repetitions

a 60 4 1000 2

b 70 4 1000 2

5 4 c 80 4 1000 2

d 85 4 1000 2

e 85 8 1000 2

f 85 12 1000 2

g 85 12 5000 1

h 68 15 5000 1

3D ONLY

Perform Daily Check (see paragraph 4.2)

EN 2D/3D OPT E.L. CALIBRATION MANUAL 11


hyperion X5 2D/3D

3. 2D CALIBRATION

3.1 X-RAY ALIGNMENT CHECK

The aim is to check that the X-ray beam is inside the


red demarcation lines;

click on “Start procedure”

12 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

Remove any object between source and sensor and


press “Start”

Press the button when prompted.

Upper part Lower part


Check that the entire length of the X-ray beam is cen-
tred inside the red lines (compare with the following
images), in particular concerning rotation, horizontal
alignment and vertical alignment.

If adjustments are necessary, open the generator co-


ver and operate on the collimator as follows:

EN 2D/3D OPT E.L. CALIBRATION MANUAL 13


hyperion X5 2D/3D

• Rotation

If the X-ray beam is turned, open the monobloc cover


and slightly loosen the following screws (A, B,C):

Correct the rotation manually; tighten the screws, se-


lect “Image not centred” and press “ok”.

Repeat the procedure until a perfectly vertical image


is obtained.

14 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

• Horizontal alignment

Slightly loosen the following screws (D, E):

Correct the horizontal alignment manually; tighten the


screws, select “Image not centred” and press “ok”.

Repeat the procedure until a horizontally-centred ima-


ge is obtained.

EN 2D/3D OPT E.L. CALIBRATION MANUAL 15


hyperion X5 2D/3D

• Vertical alignment

Slightly loosen the following screws (F, G, H, I).

Correct the vertical alignment manually; tighten the


screws, select “Image not centred” and press “ok”.

Repeat the procedure until a vertically-centred image


is obtained.

When the image is perfectly centred, select “Image


centred” and press OK; the software will automatically
check the proper alignment even for the Bitewing and
Children panoramic positions.

At the end of the procedure, press Ok to confirm and


save the data.

16 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

3.2 SENSOR CALIBRATION CHECK

Press “Start procedure” and make sure that any object


between source and sensor has been removed.

The procedure is automatic; just press the X-ray but-


ton when required; at the end, save the data.

EN 2D/3D OPT E.L. CALIBRATION MANUAL 17


hyperion X5 2D/3D

3.3 VERIFYING THE MECHANICAL CENTERING

Remove bite, chinrest and aluminium filter and posi-


tion the calibration phantom as shown in the figure.

Click on “Acquire”.

The calibration tool will acquire a complete panora-


mic image and will display it; make sure that the fo-
cus is centred with respect to the series of circles in
the middle of the image, i.e. the number of off-focus
circles must be the same both above and below the
median horizontal line.

18 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

To correct the positioning of the vertical line, adjust


the highlighted grub screws and screws (placed in the
lower part of the kinematic mechanism – remove the
accessory drawer to see them).

If the image obtained is similar to the following one,


adjust the left-hand screw and grub screw; slightly lo-
osen screw B and tighten grub screw A (some strength
is necessary) to modify the position of the chinrest.

Vice versa, if the image is similar to the following one,


adjust the right-hand screw and grub screw.

At the end of each adjustment, press “Retry” to acqui-


re a new image and assess it; if necessary, repeat the
operation until an image similar to the following one
is obtained.

To change the image focus position, modify the value


contained in the “offset Y” field, confirm by means of
the arrow and press “Retry” in the dialog box in the
foreground.

At the end of the procedure, when the focus is centred


and the line is perfectly vertical, press “OK” to confirm
and save the data.

EN 2D/3D OPT E.L. CALIBRATION MANUAL 19


hyperion X5 2D/3D

3.4 LASER CALIBRATION

Position the laser reference template on the support


(as shown in the figure)

Press “Start calibration”

20 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

The machine will turn automatically until it has rea-


ched the right position; assess the alignment betwe-
en the three laser beams (midsagittal, canine and
Frankfurt plane) and the references on the template.

If one or more lasers are not aligned, use a gripper to


loosen the springs of the corresponding supports and
adjust them manually; the Frankfurt plane laser and
the canine laser are inside the generator block.

The midsagittal laser is under the mirror.

At the end, press “Exit”.

EN 2D/3D OPT E.L. CALIBRATION MANUAL 21


hyperion X5 2D/3D

3.5 CALIBRATION DATA BACKUP

Select “Device data” and “Tubehead calibration data”.

Select the folder where the calibration data copy is


going to be saved and confirm.

22 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

4. 3D CALIBRATION

4.1 RAY BEAM CALIBRATION


From iRYS main menu, press Ctrl+Alt+Shift+End keys
together to set software in service mode (“Service Le-
vel” will be displayed at the bottom left, under RYS)

From the “Tools” menu, select “Scanner tests”,

press “Reset” to set device in 3D mode.

Then, from the “Tools” menu, select “Beam limiter


test”,

EN 2D/3D OPT E.L. CALIBRATION MANUAL 23


hyperion X5 2D/3D

tick “Enable manual tuning”.

During this procedure, collimator horizontal position


must be adjusted for the different available types of
view.

24 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

The aim of the procedure is to make the area hit by the


rays (in grey) match with the green box delimiting the
view field. In addition, the shadow area at the edges
must be within the red lines (see next image).

Before continuing, make sure that the scan area is


free from any object.

• DIAMETER 10


• Select the 10x10 field, enter the values -3 and -7
inside the “Left” and “Right” fields, then press “Move”.

Enter the 8 and 10 values in the mA and time fields,


then press “Acquisition” to perform a ray acquisition.

10
If the resulting image is like the following one

EN 2D/3D OPT E.L. CALIBRATION MANUAL 25


hyperion X5 2D/3D

slightly “close” left and right collimators (using the ar-


rows under the “left” and “right” fields, press “move”
and acquire a new image; repeat this procedure until
an image like the following oneis acquired

where you can distinguish the black colour on the side


edges. Slightly “open” collimators to get a uniform
grey image again.

Once finished, press “Save” to save the new setting.

Then, select and save also for the following view


fields:

• 10x7
• 10x6
• 10x10 HiRes
• 10x7 HiRes
• 10x6 HiRes

26 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

• DIAMETER 8

Select the 8x10 field, enter the -17 and -18 values in
the “Left” and “Right” fields and press “move”. As soon
as the machine is ready, perform a new acquisition (8
mA and 20 ms).

Evaluate the image quality and, if necessary, change


collimator position until obtaining an image like the fol-
lowing one (...)

Once finished, press “Save” to save the new setting.

Then, select and save also for the following view


fields:

• 7x7
• 8x6
• 8x10 HiRes
• 8x7 HiRes
• 8x6 HiRes

• DIAMETER 6

Select the 6x7 field, enter the -31 and -28 values in
the “Left” and “Right” fields and press “move”. As soon
as the machine is ready, perform a new acquisition (8
mA and 20 ms).

Evaluate the image quality and, if necessary, chan-


ge collimator position until obtaining an image like the
following one

EN 2D/3D OPT E.L. CALIBRATION MANUAL 27


hyperion X5 2D/3D

Once finished, press “Save” to save the new setting.

Then, select and save also for the following view


fields:

• 6x6
• 6x7 HiRes
• 6x6 HiRes

After having calibrated all view fields, press “close”.

The software will ask you whether you want to cancel


blanks; press yes.

28 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

4.2 DAILY CHECK AND BLANK ACQUISITION

From RYS main menu, select “Scan”, “Daily Check”,

then “Start”.

EN 2D/3D OPT E.L. CALIBRATION MANUAL 29


hyperion X5 2D/3D

As indicated by the software, some automatic expo-


sures will be performed; before continuing, make sure
that nobody is present in the room and that the scan
area is free.

The procedure, fully automatic, lasts approx. five mi-


nutes. Once finished, the blank acquisition window
will be displayed.

In this case as well, some automatic exposures will be


performed; once procedure is completed, make sure
that the final image is uniform. Then press “Close”.

30 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

4.3 TIN SCAN

From RYS main menu, select “Scan” and “Mattarel-


lum acquisition”

Position Tin on octagonal plate, as shown in the figu-


re, then press “XRay Flash”

Press rays button when prompted; the system acqui-


res the first scout to check cylinder positioning. Make
sure that the image is similar to the next one, and that
there are no error messages about a wrong Tin posi-
tioning.

EN 2D/3D OPT E.L. CALIBRATION MANUAL 31


hyperion X5 2D/3D

Press “Next”

The device acquires the second evaluation scout ima-


ge; make an evaluation like the previous one.

Press “Next”,

then press “Start”

32 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

Press the rays button when prompted; the system


makes a complete tin acquisition, then it will automati-
cally open the “Tin Report” with the results.

Make sure that none of the reported values has aste-


risks; should this not be the case, check tin position
and repeat the procedure.

EN 2D/3D OPT E.L. CALIBRATION MANUAL 33


hyperion X5 2D/3D

4.4 QA PHANTOM ACQUISITION (OPTIONAL)

From RYS main menu, select “scan” and “QA Phan-


tom scan”

Position phantom on the octagonal plate, as shown in


the figure (positioning is not that important, as it has a
central symmetry).

From scout acquisition window, click on “X-ray Flash”


and press the rays button when prompted.

34 2D/3D OPT E.L. CALIBRATION MANUAL EN


hyperion X5 2D/3D

Evaluate phantom position; if necessary, correct the


side position to reach a position similar to the one
shown in the figure.

Press CTRL, click the image with the left mouse but-
ton and move it (keeping both keys pressed). Once
the desired position has been reached, release mou-
se button and wait for the Software to warn you to
release also the CTRL button. Once finished, press
the rays button when prompted; the system acquires
a new scout to evaluate the new position.

To confirm the correct position, press “Next”,

then press “X-ray Flash”.

The front scout is acquired, no corrections are nee-


ded.

Press “Next”, then press “Start”.

Press the rays button when prompted. The system

EN 2D/3D OPT E.L. CALIBRATION MANUAL 35


hyperion X5 2D/3D

performs a complete acquisition of the QA phantom;


once finished, when system prompts you to continue
the analysis, press “Yes”.

The calculations are made automatically; once fini-


shed, a report like the one shown in the figure will be
displayed:

If no value has asterisks, close the window; should


this not be the case, check QA phantom positioning
and repeat the procedure.

4.5 CALIBRATION DATA BACKUP

From RYS main page, select “Setting Backup”

Select the folder for the Backup and press OK

36 2D/3D OPT E.L. CALIBRATION MANUAL EN