PMManager User Manual
PMManager User Manual
___________________________________________________________________________________
PMManager User Manual 1
Measurement modes .......................................................................................................................... 28
TTL Output........................................................................................................................................... 30
Analog Output ..................................................................................................................................... 31
Functions ............................................................................................................................................. 32
Average .................................................................................................................................................. 33
Offset...................................................................................................................................................... 35
Scale Factor ............................................................................................................................................ 35
dBm and dB offset .................................................................................................................................. 36
Normalize ............................................................................................................................................... 36
Density ................................................................................................................................................... 37
Pass/Fail Limits ....................................................................................................................................... 38
Graphic Setup ...................................................................................................................................... 39
Line Graph .............................................................................................................................................. 39
Histogram Graph .................................................................................................................................... 41
Pulse Chart ............................................................................................................................................. 42
Needle Graph ......................................................................................................................................... 43
Position................................................................................................................................................... 44
Stability................................................................................................................................................... 46
Apply to All ............................................................................................................................................. 49
Chapter 5 - Measuring with Thermopile Sensors ................................................................................... 50
About Thermopile Sensors .................................................................................................................. 50
Configure Measurement Settings ....................................................................................................... 50
Configure Power Settings.................................................................................................................... 51
Laser Wavelengths .............................................................................................................................. 51
LP1 Sensors ......................................................................................................................................... 51
Power Ranges ...................................................................................................................................... 51
Auto Range .......................................................................................................................................... 52
Average the Measurements................................................................................................................ 52
Configure Energy Settings ................................................................................................................... 52
Instrument is Ready ............................................................................................................................ 52
Energy Thresholds ............................................................................................................................... 52
Configure Pulsed Power Settings ........................................................................................................ 53
___________________________________________________________________________________
PMManager User Manual 2
Pulsed Power.......................................................................................................................................... 53
Pulse Width ............................................................................................................................................ 53
Optimize the Readings ........................................................................................................................ 54
Apply an Offset....................................................................................................................................... 54
Zero the Instrument ............................................................................................................................... 55
Specify Line Frequency........................................................................................................................... 56
Calibration Factors .............................................................................................................................. 57
Adjust Power Calibration Factors........................................................................................................... 57
Configure the Response Factor .............................................................................................................. 58
Adjust Energy Calibration Factors .......................................................................................................... 61
PEPS Sensors ....................................................................................................................................... 63
Configure Track w/Power Settings ..................................................................................................... 63
Additional Graphical Display Options ................................................................................................. 63
Chapter 6 - Measuring with Photodiode Sensors ................................................................................... 64
About Photodiode Sensors ................................................................................................................. 64
Configure Measurement Settings ....................................................................................................... 64
Configure Laser Wavelengths ................................................................................................................ 65
Select the Range..................................................................................................................................... 67
Filter Settings ......................................................................................................................................... 67
Photodiode Exposure ............................................................................................................................. 68
Low Frequency Power ............................................................................................................................ 69
Average the Measurements ................................................................................................................... 71
Optimize the Readings ........................................................................................................................ 71
Apply an Offset....................................................................................................................................... 71
Zero the Instrument ............................................................................................................................... 71
Specify Line Frequency........................................................................................................................... 71
Adjust Calibration Factors ...................................................................................................................... 71
Additional Graphical Display Options ................................................................................................. 72
Chapter 7 - Measuring with Pyroelectric and PD Energy Sensors .......................................................... 73
About Pyroelectric and PD Energy Sensors ........................................................................................ 73
Configure Measurement Settings ....................................................................................................... 74
Configure Laser Wavelengths ............................................................................................................. 75
Broadband (BB) Sensors ......................................................................................................................... 75
___________________________________________________________________________________
PMManager User Manual 3
Select the Range..................................................................................................................................... 75
Add a Diffuser......................................................................................................................................... 76
Configure the Pulse Width ..................................................................................................................... 76
Average the Measurements ................................................................................................................... 77
Set the Thresholds.................................................................................................................................. 77
Optimize the Readings ........................................................................................................................ 77
Zero the Instrument ............................................................................................................................... 77
Measure the Total Energy Exposure ................................................................................................... 78
Adjust Calibration Factors ................................................................................................................... 80
Additional Graphical Display Options ................................................................................................. 81
Chapter 8 - Triggers and Missing Pulses ................................................................................................. 82
About the External Trigger .................................................................................................................. 82
Hardware Considerations ................................................................................................................... 82
Configure the External Trigger Settings .............................................................................................. 83
Configure the External Trigger Window Time .................................................................................... 83
Choosing the Window Time Setting ....................................................................................................... 83
External Trigger Input Modes ............................................................................................................. 84
Rising Edge Mode ................................................................................................................................... 85
Enable or Disable a Channel to Work with External Trigger .................................................................. 85
External Trigger Output Modes .......................................................................................................... 86
Enable an Active Channel for Output Control........................................................................................ 86
Chapter 9 - Multiple Channels ................................................................................................................ 87
Connect Multiple Sensors ................................................................................................................... 87
Select Sensors ..................................................................................................................................... 88
Select Device Options ......................................................................................................................... 89
Two Channel Displays ......................................................................................................................... 89
Merge and Split ...................................................................................................................................... 90
Split......................................................................................................................................................... 90
Merge ..................................................................................................................................................... 91
Multiple Sessions ................................................................................................................................... 92
Track w/Power ....................................................................................................................................... 94
View the List of Active Sensors .............................................................................................................. 95
___________________________________________________________________________________
PMManager User Manual 4
Chapter 10 - Math Channels ................................................................................................................... 96
Add and Close a Math Channel ........................................................................................................... 96
Adjust Default Operations .................................................................................................................. 96
Create a User Defined Formula........................................................................................................... 97
Chapter 11 - Log Files .............................................................................................................................. 99
Default Log Files .................................................................................................................................. 99
Configure Log File Settings ................................................................................................................ 100
Log One Screen of Data Only ............................................................................................................... 100
Configure Log Duration ........................................................................................................................ 101
Configure the Number of Measurements ............................................................................................ 102
Begin and End a Log ............................................................................................................................. 103
Add Notes to a Log File ........................................................................................................................ 104
Standard Format Log Files .................................................................................................................... 105
Excel Friendly Format Log Files ............................................................................................................ 105
Select the Log File Format .................................................................................................................... 106
Chapter 12 - View Log Files ................................................................................................................... 107
Upload Files ....................................................................................................................................... 107
Access the Log Viewer....................................................................................................................... 107
Log Viewer Window .......................................................................................................................... 108
Zoom In, Zoom Out .............................................................................................................................. 108
View Log Files in NotePad ................................................................................................................. 109
Log Files Entries .................................................................................................................................... 110
Open Log Files in Excel ...................................................................................................................... 111
Chapter 13 - Preferences and Features ................................................................................................ 115
Save Sensor Settings ......................................................................................................................... 115
Preferences ....................................................................................................................................... 115
StartUp ................................................................................................................................................. 116
Communication .................................................................................................................................... 117
Logging ................................................................................................................................................. 117
Advanced.............................................................................................................................................. 118
Export and Print a Screen Capture .................................................................................................... 118
Diagnostics ........................................................................................................................................ 120
___________________________________________________________________________________
PMManager User Manual 5
Upgrade Internal Software ............................................................................................................... 121
Upgrade Sensor Software ................................................................................................................. 121
Access Help ....................................................................................................................................... 122
Appendix A - Device Specifications ....................................................................................................... 123
1938-R / 2938-R Specifications ......................................................................................................... 123
841-PE-USB / 844-PE-USB Specifications .......................................................................................... 126
843-R-USB Specifications .................................................................................................................. 127
1919-R Specifications ........................................................................................................................ 129
Appendix B – Calibration, Traceability, and Recalibration ................................................................... 131
Newport Thermopile Sensors ........................................................................................................... 131
Surface Absorbers ............................................................................................................................. 131
Volume Absorbers............................................................................................................................. 131
Factory Calibration of Thermopile Sensors....................................................................................... 132
Linearity and Accuracy of Thermopile Sensors ................................................................................. 133
Recalibration from a Known Source of Laser Energy ........................................................................ 134
Newport Photodiode Sensors ........................................................................................................... 135
Factory Calibration of Photodiode Sensors.......................................................................................... 135
Linearity and Accuracy of Photodiode Sensors .................................................................................... 135
User Recalibration ............................................................................................................................. 135
Newport Pyroelectric PD Energy Sensors ......................................................................................... 136
Metallic................................................................................................................................................. 136
Broadband ............................................................................................................................................ 137
Calibration ......................................................................................................................................... 137
Accuracy of Calibration ........................................................................................................................ 138
Recalibration from a Known Source of Laser Energy ........................................................................... 138
Periodic Recalibration .......................................................................................................................... 140
Appendix C – Safety and Compliance ................................................................................................... 142
CE Compliance................................................................................................................................... 142
___________________________________________________________________________________
PMManager User Manual 6
About this Manual
This manual describes how to operate the Newport PMManager application with any of the following
Newport measuring devices:
PC Interfaces (without a GUI)
• 841-PE-USB / 844-PE-USB – compact sensor to USB interface
Power/Energy Meters
• 1938-R, 2938-R, for data-logging
• 1919-R, for data-logging
• 843-R-USB, for data-logging
Definition of Terms
The Newport devices are referred to in this manual as device or instrument. The connection between the
smart sensor detector, referred to as a sensor, the 841-PE-USB / 844-PE-USB.
1938-R, 2938-R, 1919-R or 843-R-USB devices, and the PMManager software running on your PC, is
referred to as a channel. Using the PMManager system, you can view each channel’s laser power and
energy information. You can work with multiple channels depending on the type of device you have and the
number of free USB ports you have on your PC.
PMManager Features
PMManager converts a PC into a laser power/energy multi-channel analysis workstation.
The PMManager software features include:
• Extensive graphic display of data:
- Line Plot, Histogram, Pulse Chart, Simulated Analog Needle, Position, and Stability Graph
- Multiple data sets on separate graphs on the same screen
• Advanced measurement processing
- Power/energy density, scale factor, normalize against a reference
- Multi-channel comparisons
- User defined mathematical equations: channels A/B, etc.
• Connect additional devices during active measurements
• Data logging for future review
- Displayed graphically or saved in text format
- Exported to an Excel spreadsheet
• Printing of graphs and data
• Interfaces and supports data logging with Newport’s devices
1938-R, 2938-R
The 19389-R & 2938-R package contains the following items:
• 1938-R / 2938-R device
• USB cable
• Power cable
1919-R
The 1919-R package contains the following items:
• 1919-R device
• USB cable
• 12v power supply
843-R-USB
The 843-R-USB package contains the following items:
• 843-R-USB device
• USB cable
• 12v power supply
841-PE-USB / 844-PE-USB
The 841-PE-USB / 844-PE-USB package contains the following items:
• 841-PE-USB / 844-PE-USB device
• USB cable
• Analog output connector 2.5 mm mono jack
Item Requirement
RAM 4GB
Operating System Windows 7 / 8 / 10 (32-bit or 64-bit)
Display 1024 X 768 resolution (17” recommended)
Install PMManager
Install PMManager software
Download the PMManager from the product page in the web site (https://www.newport.com/g/optical-
power-meters) and following the Installation Wizard steps.
3. Connect the power cable between the electrical outlet and the 110-220V input located on the device.
4. Open PMManager - double-click the PMManager shortcut icon, located on the desktop.
• The device is connected and ready for use.
Connect a 1919-R
After completing the software installation, connect the 1919-R device.
❖ If this is the first time you are connecting the 1919-R to your PC, ensure Windows is running in
Administrator mode.
1. Connect the sensor to the sensor input located on the device.
2. Connect the USB cable between the USB port located on the device, and your PC.
3. Connect the 12v power cable between the electrical outlet and the 12VDC input located on the device.
4. Open PMManager - double-click the PMManager shortcut icon, located on the desktop.
• The device is connected and ready for use
❖ In order for the 843-R-USB device to communicate with PMManager it must have its USB capability
activated. To do so, contact your Newport distributor to acquire a unique USB Activation Code.
841-PE-USB
844-PE-USB
3. Open PMManager - double-click the PMManager shortcut icon, located on the desktop.
• The device is connected and ready for use.
• 1938-R / 2938-R
• 1919-R
• 843-R-USB
• 841-PE-USB / 844-PE-USB
Multiple Sensors
PMManager application supports the use of multiple sensors.
Channels
Each connection between a sensor, device, and PMManager is referred to as a channel.
You can view channels as either separate graphs or combined graphs.
Multiple Channels
In PMManager you can simultaneously work with multiple channels from one or more devices.
The devices can be connected to many different kinds of sensors, including thermal, photodiode,
pyroelectric, or a combination of different sensor types.
Example:
• Use a combination of single-channel devices (e.g., 1938-R, 843-R-USB, 841-PE-USB / 844-PE-USB)
working with one channel per device.
• Use a multi-channel device (e.g., 2938-R) working with multiple channels.
• Use a combination of single-channel and multi-channel devices (a & b above).
In addition to being able to connect PMManager to numerous channels simultaneously, you can also
specify which of the connected channels you wish to view and use multiple Newport devices to create
additional channel connections. For more information, see Multiple Channels.
This section describes the following:
• Start and end a PMManager session
• PMManager Window
• Resize PMManager Window Areas
Select a Sensor
PMManager enables you to work with multiple channels from one device or various devices
simultaneously. The devices may be connected to many types of sensors including thermopile, photodiode,
pyroelectric or a combination of different sensor types. In addition to being able to connect the application
to numerous channels at once, you can specify which of the connected channels you wish to view.
For more information about multiple sensors, refer to Multiple Channels.
Select a device
1. Click Select Device(s) in the window title bar.
• PMManager begins to search for attached devices
• In addition to devices that connect via USB, there are devices that require a wireless or an ethernet
connection. When using such devices, open the menu adjacent to the relevant icon.
• When the search is completed, a separate window opens, enabling you to select device(s) that are
currently connected
PMManager Window
PMManager enables you to work with multiple channels from one device or various devices
simultaneously.
The PMManager window contains the following areas for all sensors:
• Channel Area
• Numeric Display Area
• Graph Configuration Area
• Statistics Area
• Logging Area
• Title Bar Controls
The specific items displayed in these areas depend on which channel is active – Math or sensor
(Thermopile, Photodiode, Pyroelectric, or PD energy).
❖ See PMManager Window Features for details and examples.
Channel Controls
When a sensor is connected, only the current reading and properties of the device and sensor are shown in
the channel area. Its settings, math, and function dialogs are closed.
Icons are available to disconnect the device and to connect other devices.
Channel display with all controls closed:
Select Opens Math
Device(s) Channel
Disconnects
Device
Time Frame Value selected determines time period for the X-axis.
Time is divided into three sections HH:MM:SS.
To change a section value, click in the section (numbers shown
with dark background), and click the up/down arrow to the desired
value.
Click the down arrow to open the list of graphs.
Power and Energy Graphs • The Line graph plots the values by connecting a line between
points. The Y-axis is the value and the X-axis the time.
• The Histogram presents a statistical analysis of all the
measurements.
• The Pulse Chart graph displays readings as bars with lengths
proportional to the measurement.
• The Needle graph displays value(s) as an analog meter.
Option to include previous displays as a gray persistence.
• Apply to all’: Applies the current graph mode of the selected
channel to the other channels, where applicable.
Track Graphs These graph types are for sensors in the Track w/ Power measurement
mode.
• The Position graph shows the location of the laser beam as an
offset from the center of the graph.
• The Stability graph shows how stable the beam retains its position
by counting the number of times the center of the beam remains
within defined x-y cells on the graph.
Apply to all': Applies the current graph mode of the selected channel
to the other channels, where applicable.
A y-axis range is set from the Range control in the Device Channel. For
the graph modes Line and Pulse, different range values can be
Y-axis manually entered on the graph display itself, by clicking and entering a
new value on top range value at the top of the Y-axis.Measurements
that are outside the immediate display area can be moved into view
by the y-slide control. See Line Graph.
❖ For more information on configuring graph display settings, refer to PMManager Window Features.
Statistic Description
Min Displays the minimum measurement taken.
Total Pulses
(Thermopile sensors in
Energy Mode and
Displays the total number of measurements taken.
pyroelectric and PD energy
sensors in power or energy
modes)
Frequency
(Pyroelectric and PD energy Displays the frequency at which the laser is firing.
sensors only)
The logging area is used to configure log file settings and to start and stop the log.
For information on working with log files, refer to Log Files.
Control Description
Select Device(s) Command to select and connect devices (see Select a Sensor)
This area can be expanded to fill the entire screen by clicking the icon of the display area.
To contract to the default size, click the icon in the expanded view.
• These actions hide and show the left control panel.
- To fill the display area, click the hide/reveal triangle located in the lower corner of the numerical
area. While in this expanded mode, …you can hide the statistics area so that only the measurement
area is shown. To do this, click ‘X’ at at the end of the horizontal line below the measurement value
- To return to the default size, click the hide/reveal triangle in corner of the numerical area.
- To invert the color of the measurement area at any viewing option, click the Invert Color icon,
located in the lower left corner.
Channel
A channel is opened for each sensor that is connected to PMManager.
There are two sets of configuration fields for each channel: Settings
and Functions. Each channel is assigned a letter character to identify
the channel for graphic configurations and displays. The title or
banner of the channel displays the sensor and the current value.
Holding the mouse indicator over the banner displays full details of
the sensor including sensor and devices serial numbers.
(Refer to Channel Controls)
The channel measurement frame displays the current value measured by the sensor, adjusted by enabled
(if any) Function parameters.
Measurement modes
The measurement modes and parameters, displayed in the Settings area of the PMManager Window, are
sensor dependent - meaning, the modes and parameters differ according to the type of sensor deployed.
For example, a photodiode sensor has different measurement modes and related parameters when
compared to those of a pyroelectric sensor.
• Each time a mode or any other sensor parameter is modified, it is automatically saved for the next time
the sensor is used
The table below displays the measurement modes available for different sensor types.
TTL Output
This feature is available only with the 1938-R / 2938-R
Each 1938-R / 2938-R channel is equipped with a digital output to signal to the outside world the
measurement’s status. This is beneficial in situations such as interlocking to shut down an out-of-range
laser.
Select a TTL Output mode
Navigate to TTL Output (Options > Advanced Functions > TTL Output), and choose from one of the four TTL
Output modes:
• Disable (Low)
• On (High)
• Signal On Error
• Pass/Fail Limits
Disable (Low)
TTL Output mode is set to output a 0. This is the default state of TTL Output. It will not be left floating.
On (High)
This mode is intended for manually setting the TTL signal to 1 for controlling equipment connected to the
device.
Signal On Error
In this state the TTL Output is set to High (1), when the measurement process reports one of the followings
errors:
Analog Output
Analog output is useful for driving chart recorders and feeding measurements to analog input equipment.
PMMnager enables you to remotely configure the analog output settings in the 841-PE-USB and the 844-
PE-USB only.
For meters with analogue output, such as the 1919-R, the analog output can be set locally on the meter.
Configure the analog output voltage range
5. Navigate to Analog Output (Options > Advanced Functions > Analog Output).
6. Select the relevant voltage setting.
The analog output is driven through an impedance of 100 ohm. For best accuracy, it is recommended to limit the
external load to 100K (or larger). A smaller load (down to 1K) is possible but may result in loss of accuracy.
• For thermal and photodiode sensors in Power mode, the analog output is updated 15 times per
second with the latest power measurement
• For thermal sensors in single shot energy mode, the analog output is held until the next pulse is
measured
Functions
Each sensor channel comes with a Functions section. The parameters are the same for each sensor, but
their values are related to the sensor and the current readings of the settings.
Functions apply to power and energy readings. For a sensor in Track w/Power measurement mode, the
functions will apply to the power reading, not to the position and size readings.
Every time you click the current value becomes the offset value and shows in the adjacent
field. The offset value remains until another value is selected.
The difference between the “current” value and the Offset is displayed in the measurement
frame.
If you click the button while the offset is in effect, the new offset value is the “current” value,
not the displaced difference.
3. Slide the toggle to off , to remove the Offset value, but maintain the current value.
Scale Factor
The Scale Factor function multiplies the “current” value with the positive or negative value you enter in the
adjacent field once you select the check box.
1. Open the Function dialog.
If this is the first use of Offset, the value is 1.000 (grey).
2. Slide the toggle to ON, to enable the value field.
Normalize
The Normalize function uses the selected “current” value and divides all subsequent values by the selected
value and displays the result in Channel Measurement frame.
1. Open the Function dialog.
2. Verify that dBm function is off.
3. To begin normalization, click the adjacent button.
4. The current value becomes the normalizing value and is displayed in the text field.
You can also enter the normalization value manually.
5. To remove normalization, keeping current value, slide the toggle.
Graph selection starts by clicking the adjacent list arrow to a list to select the graph type.
The icon picture updates to show your selection.
Multi-channel displaying would follow the same use of the controls, only then you can show the channels
on separate graphs or merged into one.
When merged, the channel you click defines the axis definitions.
Line Graph
Configure a line graph
1. In the Graph option list, select Line.
2. In the Time Period section, select the time period to represent the width of
the X-axis.
Since the X-axis always has 10 major divisions, then in this case,
each one is 1‑second.
After 10 seconds has elapsed, the X-axis will begin sliding to right and after
each second another division is added to the right and one division is
removed on the left.
3. Observe the value in the Channel A Measurement frame as well as the Min and Max displayed in the
Statistics frame. This information will guide you in selecting Y-axis limits.
4. Open the list box to select the units for the Y-axis.
5. Select a unit based on the information gathered from the previous steps.
With the slide controls in this position, the values are 0 to 1,000 of the unit that you
selected.
❖ When options from Functions are used, the unit selection will change accordingly.
Refer to Y-axis with Functions for details.
6. Adjust the slide controls to give you the definition that you wish to see in the display.
Selecting one of these functions does not change the units of the Y-axis
Offset and Scale Factor
scale but may require adjusting the limits.
7. The line graph also has a pause button to enable temporarily ‘freezing’ the presently displayed graph.
3. Open the list box to select the units for the X-axis.
In this example, the Y-axis was modified to display 5.000W at the top because the measurement
values were less. Initially, the value 30.000 was changed and after clicking in the bottom field
(0.0), the chart show was displayed.
Needle Modes
The Needle graph has two modes:
• Persistence
In this mode, previous measurements remain and are displayed in gray.
Also, the Min and Max values are displayed.
Click Clear to clear the display of previous measurements, including Min and Max.
Graphic area
The default size for the area is the position tracking area available on the PEPS sensor being used, with zero
in the center. The measuring dimensions can be altered by using one of the Zooming options.
Each axis is divided into 10 units creating 100 cells in which the counting is made for the beam center
position. In the Stability example, each cell is 0.05mm by 0.05mm.
Each cell is given a color that represents the number of readings for that cell, see Legend. As the stability
measurement proceeds, the cells change colors.
Autoscale
Selecting this option in Graph Setup scales the X- and Y-axes so that the display occupies the maximum area
of the graph.
If you then de-select the option, these axis settings remain. One of the Zooming options must be used to
change the display.
Legend
The legend is a series of vertical, colored squares where each square’s number is the number of counts for
that x-y cell in the graph. The value zero, which means no hits in that cell, is the background color of the
display. The cell with the greatest number of counts is made white. The values of the squares in between
are given values divided between zero and the highest values. As the stability measurement proceeds, you
will see the values of the legend’s squares change.
Data area
Laboratory System. An X, Y coordinate system defined in the space of the laser sensor. The zero point is the
sensor center, X is horizontal and Y is vertical.
Beam Axis System. An X, Y coordinate system where the X direction is the direction of the maximum
amplitude of movement of the laser beam, and the Y direction is perpendicular to the X direction. The zero
point is the average position. This coordinate system moves relative to the Laboratory System as the laser
beam moves.
Shows duration of the counting. Counting begins when the Graph option
Elapsed Time
Stability is selected or when the Reset button is clicked.
- Mouse wheel
Rolling the mouse wheel away from you zooms in, and towards you zooms out. The display may be
scrolled in any direction by pressing the mouse wheel (or middle mouse button) and dragging.
Pulsed Power mode is only available with 1919-R and 841-PE-USB / 844-PE-USB devices
Laser Wavelengths
Thermopile sensors have different absorption at different wavelengths. To compensate for these
differences, each sensor has been calibrated by a laser at several wavelengths. By specifying the correct
laser wavelength, the correction factor for that wavelength is automatically introduced.
Specify the laser wavelength
In the Laser list, specify a laser wavelength.
LP1 Sensors
For the LP1 family of sensors, the absorption of the detector coating varies somewhat according to
wavelength. The correction curve for the absorber is stored in the sensor EEROM.
This correction curve ensures the power reading is correct at all laser wavelengths.
Power Ranges
Thermopile sensors cover a wide range of powers, from microwatt to 1000s of watts, depending on the
type of sensor. To provide accuracy at each end of the range, the electronics of the device must be
configured to work in a range that is most suited to your needs.
Specify the power range
In the Range list, specify a power range.
❖ When the range of the expected readings is unknown, or if highly varied readings are expected, select
AUTO (see Auto Range)
Auto Range
When the range is set to Auto, the instrument configures itself in the lowest range possible that is higher
than the latest readings.
• If the readings exceed 100% of the present range, the instrument reconfigures itself for the next higher
range
• If the readings fall below 9% of the present range, the instrument reconfigures itself for the next lower
range after a short delay
The delay prevents an infinite range-changing loop when readings are close to the end of the scale.
Instrument is Ready
When the instrument is ready to measure a new pulse, the word READY flashes in the numeric display area.
The next pulse automatically clears the screen, and the new value is then displayed.
If a further pulse is fired before the word READY is displayed, the reading may be inaccurate or may not be
displayed.
Energy Thresholds
If the instrument is used in a noisy environment, or where high levels of background thermal radiation are
present, the instrument may trigger spuriously on the noise or background radiation. It would then fail to
measure the intended pulse. Since some degree of noise or background radiation is inevitable, the
instrument is designed not to respond to pulses below a preset minimum level.
Pulsed Power
Thermopile sensors models are capable of measuring pulsed power in order to display instantaneous power
of a laser pulse. Power can be calculated from energy if the length of the pulse is known.
ⓘ The accuracy of measurement is dependent on setting the pulse width correctly
Pulse Width
A key element in measuring pulsed lasers is to define the pulse width ( or pulse length ) of the laser pulses.
Specify the pulse width
Specify the pulse width to the shortest time that is longer than the expected pulse width.
1. In the Pulse Width list, select Modify.
2. In the Modify Pulse Width dialog box, specify the pulse width duration in the seconds field.
• Specify a shorter pulse width setting for higher frequency lasers
Apply an Offset
When there is thermal background in the environment, thermopile sensors may show a non-zero power
reading, even when there is no laser. For example, the display reads 0.1 Watts when the laser is blocked,
and reads 20.5 Watts when laser power is applied.
In this case, the true power is 20.5 - 0.1 = 20.4 Watts.
To subtract the background, apply an offset while the laser is blocked. The display will now read zero, and
the 0.1 Watt background will be subtracted from all subsequent readings.
The laser power reading will be 20.4 Watts.
❖ If you suspect the instrument has a permanent zero offset, disconnect the sensor while the sensor is in
power measurement mode. If the instrument still displays a similar reading even when the sensor is not
connected, zero the instrument.
When the zeroing is complete, a dialog opens with a message describing if the zeroing was successfully
completed or failed. Click OK to dismiss the message.
Zero the device with a sensor attached
1. Start up the instrument with a sensor attached.
2. From the Settings option Cogwheel, Open the Advanced Functions menu and select Zeroing. The
Zeroing Instrument dialog opens.
3. In the Zeroing Instrument dialog, click Start to begin zeroing the instrument.
4. When the zeroing is complete, click Save to save the zeroing.
Calibration Factors
PMManager enables you to adjust power calibration factors, configure the response factor, and adjust
energy calibration factors.
The Original area displays the original Measurement and Calibration Factor. The Current wavelength is
displayed beneath the Original area.
2. Use the scroll bar to adjust the Calibration Factor in order to attain an accurate Calibration Factor
reading.
3. Click Save to save the adjustment for the active channel.
2. Select the wavelength you wish to remove. You can only remove one wavelength at a time.
• When the range of the expected readings is not known, or highly varying readings are expected:
In the Range list, select AUTO.
❖ AUTO instructs the instrument to configure itself in the lowest range possible that is higher than the
latest readings.
- If the readings exceed 100% of the present range, the instrument reconfigures itself for the
next higher range.
- If the readings fall below 9% of the present range, the instrument reconfigures itself for the
next lower range after a short delay. The delay prevents an infinite range-changing loop when
readings are close to the end of the scale.
Filter Settings
Most photodiode sensors are equipped with an optional filter that allows them to measure up to 300mW or
more without saturating the detector. The exact maximum power is reached when the reading reaches full
❖ Make sure to physically insert or remove the filter, according to the filter setting selected.
Failure to do so will result in erroneous readings.
❖ For sensors with built-in filter state detection, the Filter list displays the current active setting only
Photodiode Exposure
❖ Photodiode exposure is currently only supported on 1919-R devices.
• For Photodiode sensors, measuring total exposure is based on summing photodiode power
measurements over time
• In exposure mode, the devices measure 15 times per second, updating the exposure displayed on the
screen 5 times per second
Measure total exposure
1. In the Measuring list, select Exposure.
2. In the Stop Mode list, select either Manual or Timeout.
3. Click Start.
The accumulated laser power exposure and elapsed time are displayed.
• Click Stop, to stop measuring the exposure before the chosen period (Manual or Timeout).
• Click Reset, to reset the reading to zero before another reading
4. In the Modify Pulse Frequency dialog, manually enter the required pulse frequency,
OR
Use the scroll bar or arrows to modify the pulse frequency.
• If the frequency setting is too low, this may result in periodic positive power spikes above the
average power level being measured
• If the frequency is too high, the spikes will be negative, below the average power level
5. Click OK to save the modified pulse frequency, and close the dialog.
❖ Due to slight errors in the nominal internal clock frequency of the measurement device, or of the laser
pulse frequency, the exact frequency setting required for perfect measurements may not be exactly as
expected.
Apply an Offset
Newport’s unique dual-detector sensors detect and subtract 98% of background light. The residual
background signal can be removed using the Offset feature.
For example, the display reads 0.1 µW when the laser is blocked, and reads 20.5 µW when laser power is
applied. In this case, the true power is 20.5 - 0.1 = 20.4 µW. To subtract the background, apply an offset
while the laser is blocked. The display will now read zero, and the 0.1 µW background will be subtracted
from all subsequent readings. The laser power reading will be 20.4 µW.
To apply an offset, see Offset
❖ If you suspect the instrument has a permanent zero offset, disconnect the sensor while the sensor is in
power measurement mode. If the instrument still displays a similar reading even when the sensor is not
connected, zero the instrument.
2. Use the Calibration Factor scroll bar to attain an accurate reading in the Calibration Factor field in the
Adjusted area,
OR
Enter the factor into the text box above the scroll bar.
❖ Adjusting the Calibration Factor changes the calibration of all wavelengths by the same factor.
3. Click Save to save the adjustment for the active channel.
A test slide is provided with each pyroelectric sensor, which contains the same
⚠
coating as the pyroelectric detector. (You can obtain additional slides from your
supplier.)
WARNING Use this slide to test the damage threshold of your laser pulses. If the laser pulses
damage the slide, either enlarge the beam or lower the laser energy until no damage
is detected.
⚠ While measuring pulsing lasers, an erroneous energy reading will result if the energy
WARNING range is not set up correctly.
❖ When the diffuser setting is changed, a message is displayed, reminding you you to either insert or
remove the diffuser
❖ This setting is not enabled for sensors that are not equipped with a diffuser - the sensors are configured
to measure in Diffuser Out mode.
Metallic sensor
1. The 1938-R device includes one TRIG IN BNC connector as an External Trigger Input the channel, as
well as one TTL BNC connectors for signal output from the channel – see TTL Output.
2. The 2938-R device includes one TRIG IN BNC connector as an External Trigger Input for both
channels combined, as well as two TTL BNC connectors for signal output from each channel
individually – see TTL Output.
Hardware Considerations
The External Trigger input characteristics are as follows:
• Can be driven by any digital signal, TTL or CMOS, 3.3v or 5v logic
• Can tolerate input voltages between 10v and -10v
• A voltage is considered high when it is 2.4v or greater
• A voltage is considered low when it is 0.6v or smaller
• Minimum pulse width 200ns (>1us preferable)
Falling Edge • The device is sensitive to a trigger on the FALLING EDGE of the input
• The trigger is valid for a pulse arriving during a window of time before or
after the active (falling) edge (see Figure 8‑2)
• The inactive (rising) edge of the signal is ignored
• Missing Pulses are recorded when an External Trigger edge is received, but
no pulse arrives within the Window Time before or after the active edge
• Pulses are ignored if they arrive outside the Window Time, before or after
the active edge
• Multiple pulses within the Window Time are counted as one pulse
High Level • Pulses are recorded only when the input signal is at a HIGH LEVEL
• Any pulse arriving while the signal is high is counted. Any pulse arriving
while the signal is low is ignored
• No Missing Pulses are recorded in this mode
Low Level • Pulses are recorded only when the input signal is at a LOW LEVEL
• Any pulse arriving while the signal is low is counted. Any pulse arriving while
the signal is high is ignored
• No Missing Pulses are recorded in this mode
❖ When the External Trigger is not in use, ensure that the Ext. Trigger list is set to Off, otherwise the
application may not function correctly
Multiple USB devices can be attached to the PC by using more than one USB port.
Most PCs are equipped with between 2 and 6 USB ports. However, the USB standard specifies that up to
127 devices can be attached to one USB controller. The standard relies on a device called a hub. Hubs
provide ports to allow multiple devices to be attached.
• 841-PE-USB / 844-PE-USB devices are bus-powered and therefore require a hub that can provide power
to downstream high-power devices
• 1938-R, 2938-R, 1919-R and 843-R-USB are self-powered and do not require power from the USB port
Split
By default, when starting PMManager with multiple channels, the channels are
displayed (split) as Line graphs in the Display area.
The split feature is available individually in all graph modes. Each graph mode has its own Y settings, but
shares the same X setting with the other display.
A channel is selected for focus by clicking its colored horizontal line in its Channel Control window.
Create a Math Channel for a selected device channel by clicking the relevant icon (located above the
opened device Channels), see Math Channels.
This display method may have a purpose when certain time-related analyses are required because each
X-axis can have its own definition. Each channel is given the same default color.
However, the method does not allow performing Math Channel operations between two channels. Only
operations on the same channel are available. See Math Channels. Moreover, the option to
Capture Screen, outputs only the selected channel and not both channels.
The above image shows the PEPS sensor displayed in Track w/Power mode as the orange Position graph
and power from another sensor as the Line graph in blue.
Since the focus is the orange display, the Statistics window displays the results of power measured with the
PEPS sensor.
The Graph Setup also corresponds to display in focus. In either case, Merge is disabled.
The option, Apply to All, is unavailable in this display configuration.
Click the expand/reveal triangle to minimize the list when multiple devices are connected.
Logging Menu
Log Control
Log Settings
Begin Log
• The Log Control icons change to display a Stop and Pause icon.
• The time left before logging stops indicates the time remaining in the time range that was selected for
the measurement.
• Hover the mouse over the button to display the logging type.
❖ No changes can be made to the settings in the Measurement
Parameters or Graph Configuration Areas throughout the log process
duration.
Upload Files
The process for uploading files differs depending upon the device.
Upload files from 1938-R, 2938-R and 1919-R devices
To upload files from the devices, simply remove the USB flash drive from the device and attach it to a
computer.
The devices store the files in a directory called Log Files. The log files in PMManager-friendly format can be
opened directly from the USB flash drive similar to any other file that PMManager stores on the computer.
❖ For details of the default location for log files, see Default Log Files.
2. Select the required file and click Open. The log file opens in the PMManager application’s Log Viewer.
In this example, a log file having a Multi prefix is selected denotes multiple channels were logged.
Numeric prefixes indicate a one-channel log file.
The first paragraph identifies the PMManager, graph mode, logging time, and any note that was added by
the Add Notes to a Log File option.
The next paragraphs contain the information that describe the devices, sensors, channel association,
parameters per channel when the log measurement was made, followed by statistics for each channel.
The next section is the data that was measured. Using the First Pulse Arrived value, 11:05:23.980000 as a
basis, the times in the first column are the incremental changes in the X-axis.
After new readings on both channels have been logged, the math channel is evaluated.
A log file for Track w/Power is similar in form to other log files, but with added information. The log file
includes power, position, and size data, whichever graph is shown. When Stability graph is shown, the final
data from the Data area are included.
Preferences
1. Click the Preferences icon in the title bar, and select Preferences.
The Preferences dialog opens consisting of four-tabs:
• StartUp
• Communication
• Logging
• Advanced
Option Description
Restart devices on application
When PMManager begins, all connected devices are restarted
startup
Show Select Device dialog When selected, always show Select Device(s) window when
PMManager starts.
Open all sensors When selected, connects to all devices without showing Select
Device(s) window
Open last configuration When selected, uses previous PMManager configuration
settings.
Select configuration file When selected, enables the buttons Browse and Save current
configuration.
• Use browse to load an existing config file
• To create a new config file, enter the file name in the text
field, and click Save current config
Communication
There are three communication options.
Click to select which found device(s) are to be used in the current session.
Logging
Navigate to the Logging tab to specify the logging preferences
Option Description
Flush file every Set PMManager to zero log file every 60 (default), 10, and 3
seconds
Log file format Set log format as either Excel-type (default) or standard
Open Log Viewer automatically Have Log View automatically open when log is finished
when log finishes
Advanced
Navigate to the Advanced tab to specify how PMManager handles firmware downgrade and software
graphics rendering.
Option Description
Allow firmware downgrade Allows loading a firmware version that is lower than the installed
firmware in the device.
See, Upgrade Internal Software
Use software graphics rendering Use of this option if PMManager displays any graphic anomalies
1. Click the camera icon. A dialog opens overlaying the screen that includes the captured screen and
option buttons.
1. Click the . A dialog opens overlaying the screen that includes the captured screen and option
buttons.
2. Click to print directly to the default printer. The Default printer is always selected.
3. To print to a different printer, de-select the option and click Print. The standard Windows dialog opens
to select a different printer.
❖ When opened, the Functions dialog is adjacent to the channel dialog. You can select the
Functions dialog and move it anywhere on the screen. However, the dialog closes when you click
anywhere else on the screen so that the dialog loses focus. Thus, regardless of its position, the
Function dialog is not included in the Capture Screen option.
Diagnostics
The device’s internal software must be upgraded when a newer version of the internal software is released;
or a fault is detected with the device.
PMManager provides diagnostics tools that check device status and upgrade a device’s internal software.
Access Help
Access the PMManager Help
1. Do one of the following:
• Select Options Cogwheel -> Help
• The PMManager Help module appears.
Electrical accuracy (up to 1.5mA) ±0.25% ± 20pA new; ±0.5% ±50pA after 1 year
Electrical accuracy (up to 1.5mA) ±0.25% ± 20pA new; ±0.5% ±50pA after 1 year
A to D resolution 12 bits
Electrical accuracy (up to 1.5mA) ±0.25% ± 20pA new; ±0.5% ±50pA after 1 year
General Specifications
Thermopile (919P, PEPS series)
Detector compatibility Photodiode (918D, 818, 819C/D series, 818-MSCOPE)
Pyroelectric (919E, LIAD series)
Analog output User selected 1v, 2v, 5v, or 10v full-scale; 0.03% resolution. 100 ohms
impedance
Polarity of analog output Inner connector positive and outer connector negative (shield)
Analog output accuracy ±0.2% (of reading) ±0.3% of full scale volts
Mass 2.8 kg
LCD Backlight LED’s. Operates from charger or battery. Backlight level is user adjustable.
Loudspeaker Yes
Communications Specifications
Electrical input noise level 500nV or 1.5pA + 0.0015% of input range @3Hz
General Specifications
User selected 1v, 2v, 5v, or 10v full-scale; 0.03% resolution. 100 ohms
Analog output
impedance
Polarity of analog output Inner connector positive and outer connector negative (shield)
Dimensions (m) 841: 117mm x 68mm x 32mm / 844: 105mm x 80mm x 29mm
Mass 0.43 Kg
USB specifications Full speed, bus-powered, high power device, one interrupt IN endpoint
Electrical input noise level 500nV or 1.5pA + 0.0015% of input range @3Hz
General Specifications
Polarity of analog output Inner connector positive and outer connector negative (shield)
Analog output accuracy ±0.2% (of reading) ±0.3% of full scale volts
Mass 0.47 Kg
Display 320 x 240 pixel TFT LCD; Active area ~70 x ~52 mm
DC 12-16v, 1W
Charger input Charge time approx. 5 hours
Automatically stops charging when battery is full
Electrical input noise level 500nV or 1.5pA + 0.0015% of input range @3Hz
General Specifications
Polarity of analog output Inner connector positive and outer connector negative (shield)
Analog output accuracy ±0.2% (of reading) ±0.3% of full scale volts
Mass 0.47 Kg
Display 320 x 240 pixel TFT LCD; Active area ~70 x ~52 mm
DC 12-16v, 1W
Charger input Charge time approx. 5 hours
Automatically stops charging when battery is full
Max points stored onboard per file Limited only by USB capacity
Surface Absorbers
BB (Broadband) Absorber
A special refractory coating is used to provide high absorption from the UV through to the IR on standard,
high power density, broadband monitor sensors. This coating withstands very high power densities (up to
20 kW/cm²) without changing calibration. The absorption ability of this coating is above 90% for most of its
range.
EX (Excimer) Absorber
The EX Absorber provides high absorption in the UV, and can withstand both the pulse energies and the
average power of excimer lasers. These discs also have excellent absorption for 10.6m and other
wavelengths. Consequently, they can also be used for other types of lasers.
LP1 (long pulse) Absorber
This absorber has a very high damage threshold for long pulse (ms) or continuous lasers and is therefore
offered for use with high power and energy lasers. It is calibrated for the spectral range
250-2200nm with some sensors also being calibrated for 2940nm. It is not suitable for long wavelength CO2
operation.
LP (Long Pulse) Absorber
The LP Absorber has a particularly high damage threshold for long pulse (ms), or continuous lasers and is
therefore offered for use with high power sensors. It is calibrated for use with the YAG laser at 1.064m, or
the CO2 laser at 10.6m, and absorbs about 90% at these wavelengths.
Volume Absorbers
P (Pulse) Type Absorber
The models with the P suffix, are used with pulsed lasers, and have a special absorbing glass with an
absorbance of 95 +2% over the operating range. Since the surface is glass, its reflectivity does not change
even if damaged or melted locally.
HE/HE1 (High Energy) Absorber
The HE and HE1 Absorbers have a particularly high damage threshold for pulsed and repetitively pulsed
lasers, of both the short and long pulse variety. The HE and HE1 Absorbers are useful where the highest
pulse energies and average powers are used.
The Original area displays the original Measurement and Calibration Factor fields.
The Current wavelength is displayed beneath the Original area.
2. Adjust the Calibration Factor until the Adjusted measurement equals the energy applied to the
sensor.
3. Click Save to save the adjustment for the active channel.
User Recalibration
Adjust Calibration Factors of Photodiode Sensors
Adjust the power calibration factor
1. Apply the known laser power.
The Original area displays the original Measurement and Calibration Factor.
The Current wavelength is displayed beneath the Original area.
❖ A different wavelength can be selected in the Measurement Parameters Area.
3. Adjust the Calibration Factor until the Adjusted measurement equals the power applied to the sensor.
4. Click Save to save the adjustment for the active channel.
❖ The relative readings at different wavelengths are fixed by the wavelength calibration curve stored in
the sensor EEROM. When changing the calibration at one wavelength, the calibration at all other
wavelengths changes proportionately.
Metallic
Metallic sensors do not have a BB suffix in the sensor name. They have a partially reflective multi-layer
metallic coating that absorbs approximately 50% of the laser pulse (see image below).
The metallic coating permits very high repetition rates, up to 5000Hz, as well as relatively high damage
threshold.
Calibration
The sensitivity of the various Newport pyroelectric sensors varies from one to another, as well as with
wavelengths. Therefore, Newport pyroelectric detectors are individually calibrated against NIST traceable
standards. In addition, the calibration is corrected in the device for different wavelengths.
Newport pyroelectric detectors are calibrated using a 1.06µm repetitively pulsed laser referenced to a NIST
traceable thermal power meter. The average energy is set to the average power of the standard power
meter, divided by the laser frequency. The metallic PE25 and PE50 sensors are also calibrated with an
excimer laser at 248nm to correct the rather large absorption variations in that spectral region with those
sensors.
Accuracy of Calibration
Since the instruments are calibrated against NIST standards, the accuracy is generally 3% at the energy level
and wavelength at which the calibration has been performed. This accuracy has been verified by checking
the scatter of the results when several instruments are calibrated against the same standard. The maximum
error in measurement will be less than the sum of the specified accuracy, linearity, and inaccuracy due to
errors in the wavelength curve. The non-linearity is approximately 2%.
For error due to wavelength, see table below.
In addition to the above errors, the reading of a pyroelectric sensor changes with frequency. The device has
a built-in correction for this error. For frequencies above 50% of maximum frequency, inaccuracies in this
correction can increase the total error up to 3%.
The maximum error in measurement will be less, and in general will be considerably less, than the sum of
the above mentioned errors.
See http://www.Newportopt.com/laser-measurement-instruments/laser-power-energy-
meters/tutorial/calibration-procedure for a complete analysis of Newport calibration accuracy.
Coating Type
Wavelength Error
Broadband Metallic
190 - 350nm ±2% ±2%
400 - 800nm ±2% ±2%
1064nm 0 0
2 - 3µm ±2% ±2%
10.6µm ±5% ±15%
Broadband sensor
- The Original area displays the original Measurement and Calibration Factor
- The Current wavelength is displayed beneath the Original area
❖ A different wavelength can be selected in the Measurement Parameters area.
3. Adjust the Calibration Factor until the Adjusted measurement equals the power applied to the sensor.
4. Click Save to save the adjustment for the active channel.
❖ For metallic sensors, when the calibration is changed at one laser wavelength, the overall calibration of
all other wavelengths changes proportionately.
For broadband sensors, there is an option to adjust the calibration factor for all wavelengths or only for
a selected wavelength.