method validation
method validation
2
Why Method Validation is Necessary?
Examples:
3
Method Development and Validation
Compliance to Standards:
FDA, GLP, ISO 17025, cGMP etc.
5
Outline
2 3
1
Method
Development and Method
Validation validation
Quality in selected
Analytical samples
Laboratory
6
METHOD VALIDATION
7
METHOD VALIDATION
8
METHOD VALIDATION
9
Method Validation
Method validation data submitted should
INCLUDE the following issues:
Accuracy
Precision
Limit of Detection (LOD)
Method Limit of Quantitation (LOQ)
Validation
Specificity
Linearity and Range
Ruggedness/Robustness
System Suitability
10
ACCURACY
Accuracy is the measure of exactness of an analytical
method, or the closeness of agreement between the
measured value and the value that is accepted as a
conventional true value or an accepted reference value.
12
12
METHOD VALIDATION : ACCURACY
The concentration should cover the range of concern and should
particularly include one concentration close to the quantitation limit.
Analyte recovery at different concentrations
“repeatability”
“reproducibility”
14
14
PRECISION (continued)
The acceptance criteria for precision depend very much on the type of
analysis:
15
PRECISION (continued)
Requirements:
According to International Conference on Harmonisation (ICH),
both repeatability and intermediate precision should be tested .
19
Evaluation of Precision
Calculate mean,
Standard deviation, S =
The relative standard deviation (RSD) is often times more
convenient. It is always expressed as percentage (%)
RSD (%) =
20
METHOD VALIDATION: PRECISION
Estimated precision data as a function the of analyte concentration
Analyte concentration versus precision within or between days
The lower the concentration of analyte, the higher the precision obtained 21
Accuracy and Precision
22
LOD and LOQ
Limit of detection (LOD) and limit of quantification
(LOQ) are two fundamental elements of method
validation that define the limitations of an analytical
method.
24
LOD
25
Methods in estimating LOD and LOQ:
Procedure:
• Run blank and the lowest concentration of analyte
• Run 10 replicates independently
• Measure the peak height values (noise magnitude was
measured by auto-integrator)
• Calculate LOD: S/N = 3, and
LOQ: S/N = 10
Analytical Method Development
LOQ
LOD
Signal to Noise = 3:1
Noise
Blank Determination
y a bx
y: instrument response
x: analyte concentration
a: y-intercept
b: slope
Linear Regression
3sa
LOD
b
10 sa
LOQ
b
32
LOQ (continued)
33
METHOD VALIDATION : LIMIT OF QUANTITATION
35
36
SPECIFICITY/SELECTIVITY
Example:
UV/Vis spectra of a drug molecule is used for
identification purposes, however, the UV spectra of the
drug is the same as the UV spectra of some drug
impurities. Thus, the UV/Vis method on its own is not
specific.
The chromatographic signal does not indicate any impurity in either peak.
Spectral evaluation, however, identifies the peak on the left as impure.
38
METHOD VALIDATION: LINEARITY AND RANGE
LINEARITY
RANGE
40
LINEARITY AND RANGE (continued)
42
ROBUSTNESS/RUGGEDNESS
Robustness can be defined as a measure of its
capability to remain unaffected by small, but
deliberate variations in method parameters and
provides an indication of its reliability during
normal usage.
- (involves internal factors to the method)
Example:
46