LCD3 Manual EN
LCD3 Manual EN
LCD3
DIGITAL LOAD
LIMITER
SYMBOLS
Here are the symbols used in the manual to draw the reader's attention:
Further information.
WARRANTY
24 months from the date of the delivery note. Warranty covers only failures of defective components (due to construction
defects or defects in materials) and includes replacement or repair of the components and related labor costs.
Warranty is automatically forfeited in the event of:
- tampering, deletion, removal of the identification label and/or serial number of the product
- misuse, transformation, alteration, repair of products not carried out by Laumas personnel
Laumas provides a 1-year warranty from the date of the delivery note on defects in material or manufacture of the battery.
The cell cable must be individually led to its panel input (from one side or the other of the panel) and
not share a conduit with other cables; connect it directly to the instrument terminal block without
breaking its route with support terminal blocks.
Use “RC” filters on the instrument-driven solenoid valve and remote control switch coils.
Avoid installing the instrument in a panel containing inverters; if inevitable, use special filters for the
inverters and separate them with sheet metal partitions.
The panel installer must provide electric protections for the instruments (fuses, door lock switch, etc.).
It is advisable to leave the equipment always switched on to prevent the formation of condensation.
The instrument is available in two versions and the manual is valid for both:
- LCD3 Rev 1: 2 changeover contact relays and 1 NO relay 18÷50 VDC/VAC; 2 A.
- LCD3 Rev 2: 3 changeover contact relays 30 VDC/250 VAC; 2 A.
The specific characteristics of only one of the two versions will be suitably indicated.
-1-
MAIN SPECIFICATIONS OF THE INSTRUMENT
VERSION IDENTIFICATION
To define the version of the instrument, check the number of contacts on the terminal block:
-2-
TECHNICAL SPECIFICATIONS
-3-
ELECTRICAL CONNECTIONS
- The cell cable must not be routed with other cables (e.g. outputs connected to remote control
switches or power supply cables), but must follow its own path.
- Any cable extension connections must be carefully shielded, observing the color code and using
the cable of the type supplied by the manufacturer. Extension connections must be made by
welding, either through back-up terminal blocks or via the junction box supplied separately.
- The cell cable must have no more conductors than those used. In the case of an unused multi-
conductor cable, connect the remaining wires to the negative pole of the cell power supply (N. 1).
- The cell cable shield must be connected to terminal 1 (- cell power supply) or to ground.
- For connections with 4-wire cells, jumper terminal 1 with 4 and 2 with 3.
The cell power supply voltage is 4 VDC and is protected against temporary short-circuiting.
The load cell cable must be connected to terminals 1...6 of the relevant terminal block.
REV 1
LCD display Status LED
instrument
Status LED
relay
Programming
keys
Load cell
terminal block
Alarm relay
-4-
REV 2
LCD display Status LED
instrument
Status LED
Programming relay
keys
Load cell
terminal block
Alarm relay
-5-
LED FUNCTION AND KEYS
LED Function
L1 instrument status led
L2 output 1 relay active/energized
L3 output 2 relay active/energized
L4 output 3 relay active/energized
In normal operating conditions, the function of the 4 keys is indicated on the bottom line of the
display every 3 seconds
KEYS Awaiting commands
When no programming procedure is in progress, the function of
the keys is as follows:
VIS: Switching the data displayed on the bottom line
(see section VIEWING ON THE DISPLAY).
>T<: semi-automatic tare control (see section SEMI-AUTOMATIC
TARE (NET/GROSS)).
>0<: command for zero-setting the gross weight within the
programmed threshold in the relevant parameter [1306].(see
section WEIGHT ZERO-SETTING).
PRG: Pressed and held for 2 seconds, it takes you to the set-up
menu (see section SET-UP MENU).
Parameters menu
When navigating a parameter menu, the first line on the left shows
the unique numeric identifier that identifies the function and on the
right the current set value.
The composition of the keys is as follows:
<-: switches to the previous parameter display.
->: switches to the next parameter display.
ESC: exits the menu and returns to the upper level or exits the
procedure.
SEL: you enter parameter editing/programming.
Numeric parameter programming
During programming, the digit being formed flashes:
(+): increments the flashing digit.
(-): decreases the flashing digit.
->: moves to the next digit on the right.
->: press and hold the key for 2 seconds to zero-set the value.
ENT: confirms and saves the displayed value.
Parameter programming with default values
During programming, the current value flashes:
<-: selects the previous value.
->: selects the next value.
ENT: confirms and saves the displayed value.
-6-
VIEWING ON THE DISPLAY
When a programming procedure is not in progress, the display shows the current gross weight, with
the indication of stability (S).
The following information is displayed on the bottom line, which can be switched with the VIS key:
In alarm conditions, by pressing the VIS key, the values on the lower line are displayed for
5 seconds.
-7-
ERROR MESSAGES ON THE DISPLAY
When an alarm occurs, the alarm message describing the cause is shown on the bottom line of the
display, the related relay of alarm 1 is activated, and the other relays are deactivated at the same time.
In the case of multiple alarms occurring, the highest priority message is displayed and it is possible to
scroll the list of alarms by repeatedly pressing the VIS key. Alarms are listed below in order of priority
(highest to lowest):
-8-
SET-UP MENU
Access to the set-up menu can be protected by a programmable password (default: not set).
To set the password or to disable it, go to the set-up menu by pressing the PRG key for two seconds, enter
the current password if prompted, and go to FUNCTIONAL PARS. GENERAL PASSWORD [1002].
Permissible values range 0÷9999. If the value is 0 the password is disabled.
Functional pars.
CALIBR. PASSWORD [1001] Calibration parameter access password
GENERAL PASSWORD [1002] Password to access the set-up menu
SAFETY FUNCTIONS [1003] SAFETY/OVERLOAD function activation selection
STAND-BY [1004] Stand-by activation time
SERIAL NUMBER [1006] Instrument identification code
Weighing const.
L.CELLS CAPACITY [1102] Total load cell rated capacity
L.C.SENS. [mV/V] [1103] Load cell average rated output
PRE-LOADED TARE [1105] Theoretical zero
DIVISION VALUE [1106] Weight division value
Weight calibrat.
ACQUIRE ZERO Zero calibration
WEIGHT CALIBR. Calibration of full scale with sample weight
LINEARIZATION Weight linearization procedure
THEORETICAL CAL Theoretical calibration calculation
Parameters
FILTER FACTOR [1301] Filter to attenuate weight fluctuations
STABILITY FACTOR [1302] Weight stabilization factor
INITIAL AUTOZERO [1304] Initial automatic zero-setting limit when instrument is switched on
ZERO-TRACKING [1305] Zero tracking function selection, expressed in divisions per second
ZERO BAND [1306] Semi-automatic zero command acceptance band
Input/output
FUNCTION INPUT 1 [1401] Input 1 function selection: Semi-automatic tare, Motor for
SAFETY function, weight data transmission.
FUNCTION INPUT 2 [1402] Input 2 function selection
SETPOINT OUT 1 [1404] Programming setpoint corresponding to output 1
DELAY OUT1 [1405] Output 1 intervention delay time in tenths of a second
POLARITY SET 1 [1406] Programming output 1 intervention polarity
HYSTERESIS SET 1 [1407] Setpoint 1 hysteresis
OUT 1 LOAD SEL. [1408] Load selection compared to setpoint 1
-9-
SETPOINT OUT 2 [1410] Programming setpoint corresponding to output 2
DELAY OUT2 [1411] Output 2 intervention delay time in tenths of a second
POLARITY SET 2
Programming output 2 intervention polarity
[1412]
HYSTERESIS SET 2
Setpoint 2 hysteresis
[1413]
OUT 2 LOAD SEL.
Load selection compared to setpoint 2
[1414]
SETPOINT OUT 3 [1416] Programming setpoint corresponding to output 3
DELAY OUT3 [1417] Output 3 intervention delay time in tenths of a second
POLARITY SET 3
Programming output 3 intervention polarity
[1418]
HYSTERESIS SET 3
Setpoint 3 hysteresis
[1419]
OUT 3 LOAD SEL.
Load selection compared to setpoint 3
[1420]
Serial ports
COM1 BAUD RATE [1501] Default baud rate: 115200
COM1 FRAME FORM. [1502] Default format: n-8-1
COM1 PROTOCOL [1503] Protocol types
COM1 ADDRESS [1504] Instrument address
COM2 BAUD RATE [1505] Default baud rate: 115200
COM2 FRAME FORM. [1506] Default format: n-8-1
COM2 PROTOCOL [1507] Protocol types
COM2 ADDRESS [1508] Instrument address
COM3 BAUD RATE [1510] Default baud rate: 115200
COM3 FRAME FORM. [1511] Default format: n-8-1
COM3 PROTOCOL [1512] Protocol types
COM3 ADDRESS [1513] Instrument address
SUM UNITS NUMBER [1509] Number of units connected in sum function
RF CHANNEL SEL [1514] RF channel number
- 10 -
Overload (if OVERLOAD on)
SETPOINT 1 [1802] Overload 1 threshold setting
SETPOINT 2 [1804] Overload 2 threshold setting
SETPOINT 3 [1806] Overload 3 threshold setting
HYSTERESIS [%] [1807] Overload hysteresis value setting
COUNTER 1 [1809] Overload 1 threshold counter display
COUNTER 2 [1811] Overload 2 threshold counter display
COUNTER 3 [1813] Overload 3 threshold counter display
COUNTER 4 [1815] Full scale threshold counter display
PEAK [1817] Maximum peak value display
RESET Counter reset procedure
WARNING: All changed settings are permanently applied and saved on exiting the set-up
menu by repeatedly pressing the “ESC” key. If you switch off the instrument without exiting
the set-up menu, the changes will be discarded.
- 11 -
INSTRUMENT COMMISSIONING
Switch on the microprocessor and wait five minutes for all the components to have a stable
temperature. Check that the instrument indication is positive and that it increases when a weight force
is applied on the container. If the indication is negative or the instrument displays an error message,
check the connection of the load cells and their correct positioning (direction of load).
FUNCTIONAL PARAMETERS
FUNCTIONAL PARS. CALIBR. PASSWORD [1001], password to access the weight calibration
procedure. If the value is 0 the password is disabled (default. 2792).
FUNCTIONAL PARS. GENERAL PASSWORD [1002], general password to access the set-up
menu. If the value is 0 the password is disabled (default. 0).
FUNCTIONAL PARS. STAND-BY [1004], you can program a time (0–999 sec.) after which,
without using the keypad, the display backlighting turns off and the keypad is disabled. Press and hold
a key for 3 seconds to exit this condition.
If the value is 0 the function is disabled.
- 12 -
INSTRUMENT CALIBRATION VERIFICATION
If you need to check the instrument calibration, for example, when you first use it, proceed as directed.
Make sure the system is unloaded and the instrument indicates zero. Apply a significant quantity of product
to the system (at least 50% of the maximum quantity to be weighed) and check that it is correctly indicated.
- If the difference is not significant but is in the order of 1–2 %, correct it (see section KEYPAD
CORRECTION OF THE INDICATED WEIGHT).
- If there is a significant difference (greater than 1–2%) make sure that this is not due to mechanical
issues and check the electrical connections and direction of load of the load cells again. Check
that the currently set theoretical calibration meets the application characteristics by referring to the
section THEORETICAL CALIBRATION.
- If necessary, make a real calibration by sample weight, referring to the section REAL
CALIBRATION (WITH SAMPLE WEIGHTS).
If the instrument indicates a different value to the weight of the product that is applied to the system,
it is possible to correct the weight by means of a simple calibration with a sample weight:
Referring to the section REAL CALIBRATION (WITH SAMPLE WEIGHTS), adjust the parameter
WEIGHT CALIBRAT. WEIGHT CALIBR., by applying a sample weight of at least 50% of the
maximum quantity to be weighed.
- 13 -
PROGRAMMING OF SYSTEM PARAMETERS
INSTRUMENT CALIBRATION
WARNING: The calibration procedures described in this section may require accessing the “WEIGHT
CALIBRAT”menu, which can be password protected: CALIBR. PASSWORD [1001]. (see section
FUNCTIONAL PARAMETERS).
THEORETICAL CALIBRATION
In the submenu “WEIGHING CONST.” you can find the items for theoretical calibration without the
need for a sample weight, of course we always recommend checking the calibration performed with
a known quantity of product.
WEIGHING CONST. L.CELLS CAPACITY [1102] (kg, integer values only), load cell full scale.
Example: If there are 3 load cells with a nominal capacity of 300 kg set 900 for this parameter.
WEIGHING CONST. L.C.SENS. [mV/V] [1103] (mV/V), load cell rated output (default 2.0000 mV/V).
Example: Set the average load cell rated output to 4 decimal places (for load cells with 2 mV/V rated
output set 2.0000; whereas, for 3 load cells with different rated output, for example 1.9800 mV,
1.9600 mV, 1.9900 mV, set 1.9766 mV).
WEIGHING CONST. DIVISION VALUE [1106] (divisions), instrument resolution with decimal point
if required.
WEIGHING CONST. PRE-LOADED TARE [1105] Setting theoretical zero, prior to calibration.
To confirm the entered settings and activate the theoretical calibration, go to the “WEIGHT
CALIBRATION” submenu and select THEORETICAL CAL. The “DONE” message will confirm that
the set calibration has been activated.
- 14 -
REAL CALIBRATION (WITH SAMPLE WEIGHTS)
The “WEIGHT CALIBRAT.” submenu contains the items for calibrating with a sample weight.
Calibration with a sample weight is not an alternative to theoretical calibration, but requires the
theoretical calibration to have been performed correctly, in particular the full scale value
WEIGHING CONST. L.CELLS CAPACITY [1102] overall will be considered as the
maximum weight limit that can be entered during the calibration procedure with a sample weight.
Proceed with the system in an idle condition and without any load to calibrate 0 using the item:
WEIGHT CALIBRAT. ACQUIRE ZERO, “Zero calibration (zero signal acquisition)” and confirm.
Afterwards, for the load, apply a sample weight of at least 50% of the maximum quantity to be weighed.
Select:
WEIGHT CALIBRAT. WEIGHT CALIBR., “Calibration of full scale with sample weight”.
The device display will prompt you to confirm the load by pressing “ENTER” and to indicate the
corresponding value of the representative quantity of the sample.
For greater weighing accuracy, 5-point linearization can be performed using different sample weights.
Select:
WEIGHT CALIBRAT. LINEARIZATION, “Weight linearization procedure”.
During the linearization procedure, the device display will prompt you to confirm the weight and to
indicate the representative sample quantity for at most 5 times. A smaller number of points can be
entered by interrupting the procedure with the “ESC” key.
- 15 -
WEIGHING PARAMETERS
PARAMETERS FILTER FACTOR [1301], you can set a value between 0 and 9 (default: 4).
The digital weight filter can be increased or decreased. In the first case, the display weight response
is slowed down, which allows absorbing oscillations or vibrations of the weighing system. In the
second case, a faster response of changes in weight is obtained.
PARAMETERS STABILITY FACTOR [1302], you can set a value between 0 and 4 (default: 2).
The stable weight condition is indicated by the relevant symbol ‘(S)’ on the LCD screen, this condition
can be determined more or less accurately and more or less quickly. Maximum accuracy is achieved
by setting the stability value to 0.
WEIGHT ZERO-SETTING
You can zero-set the weight to correct small zero displacements of the instrument in the following ways:
- From the condition of awaiting commands, by pressing the >0< key. Refer to section LED
FUNCTION AND KEYS.
The gross weight zero-setting command is not executed under the following conditions:
▫ Unstable weight (weight does not stabilize within the preset time set by PARAMETERS
STABILITY FACTOR [1302]).
▫ Gross weight greater (positive or negative) than the automatic zero-setting threshold set in
PARAMETERS ZERO BAND [div] [1306] (max 200 div).
- By activating zero tracking at power-on via PARAMETERS INITIAL AUTOZERO [1304]. When
the instrument is switched on, if the displayed weight is less than the value programmed in this
item, the display will be reset to zero. Setting the value 0 disables the function. To perform zero-
setting, the weight must be stable: if the weight does not stabilize within 3 seconds of switching
on, the automatic zero-setting function will not be performed.
- By activating zero tracking for small variations via PARAMETERS ZERO-TRACKING [1305].
You can set a value between NO/0.5d/1d/2d (Divisions per second). When the scales are unladen,
the instrument automatically corrects small and slow weight changes from the scales zero, the
corrections made are canceled when the instrument is switched off, the maximum correction limit
is the automatic zero-setting threshold. Setting the value 0 disables the function.
- 16 -
SEMI-AUTOMATIC TARE (NET/GROSS)
- From the condition of awaiting commands (see section LED FUNCTION AND KEYS) proceed as
follows:
Short press >T<: semi-automatic tare activation.
Long press >T< (2 seconds): cancellation of the set tare.
- By closing one of the available inputs by programming the relevant parameter: INPUT / OUTPUT
FUNCTION INPUT (1-2) in semi-automatic tare mode.
Once the semi-automatic tare is activated, the device will continue to show the gross weight from the
main screen. To display the net weight, the data displayed must be switched via the VIS button until
the net weight is displayed (see section DISPLAYS AND ERROR MESSAGES ON THE DISPLAY).
INPUT PROGRAMMING
To activate an input, the contact with the common terminal must be closed. Use the shortest possible
connection cable.
The behavior when the contact is closed is programmable using the following parameters:
INPUT / OUTPUT FUNCTION INPUT 1 [1401], Function selection of input 1. Selectable values:
Semi-automatic tare (Tare), Motor for SAFETY function (Motor), weight data transmission (Txd).
INPUT / OUTPUT FUNCTION INPUT 2 [1402], Function selection of input 2. Selectable values:
Semi-automatic tare (Tare), Motor for SAFETY function (Motor), weight data transmission (Txd).
For more information on the weight data transmission mode TXD see section SINGLE
TRANSMISSION PROTOCOL.
- 17 -
OUTPUTS PROGRAMMING
ELECTRICAL CONNECTIONS
When the programmed setpoint value is exceeded, the setpoint relay switches off. This behavior is
indicated by the corresponding status LED switching off.
When an alarm occurs, the alarm relay (1) is energized, and all other relays are de-energized at the
same time.
REV 1
- 18 -
REV 2
- 19 -
PARAMETERS
Through the parameters of the INPUT / OUTPUT menu it is possible to configure the outputs and the
behavior of the instrument with respect to the 3 setpoints available. If the setpoint value is = 0, the
setpoint output is never activated.
INPUT / OUTPUT SETPOINT OUT 1 [1404], setpoint corresponding to output 1, value: 0÷<capacity>.
INPUT / OUTPUT DELAY OUT1 [1405], output 1 intervention delay time in tenths of a second,
value: 0÷99.9 sec.
INPUT / OUTPUT HYSTERESIS SET1 [1407], hysteresis setpoint 1, value to be subtracted from
the setpoint to obtain the contact switching threshold for decreasing weight. Value: 0÷9999.
INPUT / OUTPUT OUT 1 LOAD SEL. [1408], load selection compared to setpoint 1. See table
Output load selection.
INPUT / OUTPUT SETPOINT OUT2 [1410], setpoint corresponding to output 2, value: 0÷<capacity>.
INPUT / OUTPUT DELAY OUT2 [1411], output 2 intervention delay time in tenths of a second,
value: 0÷99.9 sec.
INPUT / OUTPUT HYSTERESIS SET2 [1413], hysteresis setpoint 2, value to be subtracted from
the setpoint to obtain the contact switching threshold for decreasing weight. Value: 0÷9999.
INPUT / OUTPUT OUT 2 LOAD SEL. [1414], load selection compared to setpoint 2. See table
Output load selection.
INPUT / OUTPUT SETPOINT OUT3 [1416], setpoint corresponding to output 3, value: 0÷<capacity>.
INPUT / OUTPUT DELAY OUT3 [1417], output 3 intervention delay time in tenths of a second,
value: 0÷99.9 sec.
INPUT / OUTPUT HYSTERESIS SET3 [1419], hysteresis setpoint 3, value to be subtracted from
the setpoint to obtain the contact switching threshold for decreasing weight. Value: 0÷9999.
INPUT / OUTPUT OUT 3 LOAD SEL. [1420], load selection compared to setpoint 3. See table
Output load selection.
- 20 -
Output load selection
The setpoint is compared to the gross weight of
SINGLE
the instrument.
The setpoint is compared to the total load value
TOTAL of the weighing system. In this case the
sum/difference protocol must be active.
The setpoint is compared to the value of the
difference of the loads in the weighing system.
- In the case of 2 instruments, it indicates the
difference between load A and load B.
- In cases where the number of instruments
DIFFER.
exceeds 2, it indicates the difference
between the minimum load and the
maximum load of the weighing system. In
this case the sum/difference protocol must
be active.
The setpoint is compared with the total partial
load value (A+B in the case of instrument
PAIR address A or B, C+D in the case of instrument
address C or D). In this case the sum/difference
protocol must be active.
The sum/difference protocol (TOTAL, DIFFER, PAIR) can be activated for multi-instrument
configurations. (see section CONFIGURATION OF MULTI-INSTRUMENT WITH
MULTIPLE WEIGHINGS).
- 21 -
SERIAL PORTS PROGRAMMING
Through the RS485 serial port it is possible to connect several instruments to each other and activate
the sum/difference protocol to evaluate the load detected by the different devices and to activate the
setpoints available accordingly. (see section CONFIGURATION OF MULTI-INSTRUMENT WITH
MULTIPLE WEIGHINGS).
The various ports and instrument behavior can be configured using the respective parameters:
SERIAL PORTS COM1 BAUD RATE [1501], set the desired baud rate (1200 2400 4800 9600
19200 38400 57600 115200).
SERIAL PORTS COM1 FRAME FORM. [1502], set the desired format (n-8-1, n-8-2, E-7-2, E-8-
1, o-7-2, o-8-1).
SERIAL PORTS COM1 ADDRESS [1504], different value depending on the protocol
set (see section SERIAL COMMUNICATION PROTOCOLS).
SERIAL PORTS COM2 BAUD RATE [1505], set the desired baud rate (1200 2400 4800 9600
19200 38400 57600 115200).
SERIAL PORTS COM2 FRAME FORM. [1506], set the desired format (n-8-1, n-8-2, E-7-2, E-8-
1, o-7-2, o-8-1).
- 22 -
SERIAL PORTS COM2 ADDRESS [1508], different value depending on the protocol
set (see section SERIAL COMMUNICATION PROTOCOLS).
SERIAL PORTS COM3 BAUD RATE [1510], set the desired baud rate (1200 2400 4800 9600
19200 38400 57600 115200).
SERIAL PORTS COM3 FRAME FORM. [1511], set the desired format (n-8-1, n-8-2, E-7-2, E-8-
1, o-7-2, o-8-1).
SERIAL PORTS COM3 ADDRESS [1513], different value depending on the protocol
set (see section SERIAL COMMUNICATION PROTOCOLS).
SERIAL PORTS SUM UNITS NUMBER [1509], set the desired format (0-4).
SERIAL PORTS RF CHANNEL SEL [1514], set the desired value (0-7). This value must be the
same on all the devices that you want to connect. Only with the RF module installed.
Serial protocols
PC Slave protocol with bidirectional communication (see
SLAVE
section SLAVE PROTOCOL).
Continuous transmission repeater protocol
REPEATER
(see section REPEATER/REPEATER C PROTOCOL).
Continuous transmission repeater C. protocol
REPEAT.C.
(see section REPEATER/REPEATER C PROTOCOL).
Continuous weight transmission and instrument
WEIGHT status protocol (see section CONTINUOUS
WEIGHT TRANSMISSION PROTOCOL).
Activation sum/difference protocol (ONLY COM 2-3)
SUM / DIF.
(see section SUMMING PROTOCOL).
Deactivation (see section SINGLE TRANSMISSION
NONE
PROTOCOL).
- 23 -
LIFE COUNTERS CONFIGURATION
One of the life control functions of the system can be enabled. The first one, called ‘Safety’, counts
the number of lifts detected and saves the system operating time. The second one, called ‘Overload’,
allows configuring a maximum of 3 progressive weight thresholds and records the threshold count as
well as the maximum peak reached. The counts can be reset with the relevant command.
The two modes of operation are mutually exclusive and can be activated or deactivated via the menu
item FUNCTIONAL PARS. SAFETY FUNCTIONS [1003].
The related configuration menus are shown or hidden depending on the setting of this parameter.
SAFETY COUNTERS OPERAT. COUNTER [1702], lifting operation counter (motor input
active). Increased value with variable frequency, according to the formula (CS / FS) ^ 3.
CF = Load raised.
FS = Rated capacity of the lifting system.
Value: 0÷999999
SAFETY COUNTERS OPERATIONAL TIME [1706], System operating time (motor input active).
Value expressed in the format “HHHH:MM:SS” for total hours less than or equal to 9999, or
“HHHHHH:MM” for total hours greater than 9999.
SAFETY COUNTERS SYSTEM CAPACITY [1708], rated capacity of the lifting system (password
protected 2792). Value: 0÷<capacity>.
SAFETY COUNTERS RESET COUNTERS, counter reset procedure (password protected 1687).
OVERLOAD SETPOINT 1 [1802], setting threshold of overload 1. The threshold values must be
progressive, otherwise the setting will not be accepted. Set this parameter to 0 to override overload
threshold control. Value: 0÷<capacity>.
OVERLOAD SETPOINT 2 [1804], setting threshold of overload 2. The threshold values must be
progressive, otherwise the setting will not be accepted. Set this parameter to 0 to override overload
threshold control. Value: 0÷<portata>.
OVERLOAD SETPOINT 3 [1806], setting threshold of overload 3. The threshold values must be
progressive, otherwise the setting will not be accepted. Set this parameter to 0 to override overload
threshold control. Value: 0÷<portata>.
OVERLOAD HYSTERESIS [%] [1807], setting the overload hysteresis value. Value expressed as
a percentage of the capacity value. This parameter is used to determine the end threshold of the
loading procedure.
The loading procedure is determined by exceeding at least one overload threshold. The procedure
ends when the gross weight decreases to the threshold determined by the hysteresis value
(SETPOINT 1 - hysteresis). Value: 0÷100.
- 24 -
OVERLOAD COUNTER 1 [1809], Overload 1 threshold counter display. Only the counter for the
maximum threshold reached during the loading procedure is increased. Value: 0÷999999.
OVERLOAD COUNTER 2 [1811], Overload 2 threshold counter display. Only the counter for the
maximum threshold reached during the loading procedure is increased. Value: 0÷999999.
OVERLOAD COUNTER 3 [1813], Overload 3 threshold counter display. Only the counter for the
maximum threshold reached during the loading procedure is increased. Value: 0÷999999.
OVERLOAD COUNTER 4 [1815], Overload 4 threshold counter display. Only the counter for the
maximum threshold reached during the loading procedure is increased. Value: 0÷999999.
OVERLOAD PEAK [1817], displaying the maximum peak value reached during the loading
procedure. Value: 0÷999999.
- 25 -
CONFIGURATION OF MULTI-INSTRUMENT WITH MULTIPLE WEIGHINGS
Available configurations:
- A+B = 2 Instruments
- A+B+C = 3 Instruments
- A+B+C+D = 4 Instruments
With reference to the section SERIAL PORTS PROGRAMMING and to the section OUTPUTS
PROGRAMMING, acting on each instrument, configuration involves the following steps:
- Connect terminal 12 to the corresponding terminal 12 of each instrument, and terminal 13 to the
corresponding terminal 13 of each instrument.
- Make sure you have set the same number of decimal places, correct accordingly. WEIGHING
CONST. DIVISION VALUE [1106].
- Enable the Sum-Difference protocol: SERIAL PORTS COM2 PROTOCOL [1507], value SUM / DIF.
- Differentiating each instrument, set its address according to the solution that is being created:
SERIAL PORTS COM2 ADDRESS [1508]: A, B, C, D.
- Enter the total number of devices required using the parameter SERIAL PORTS SUM UNITS
NUMBER [1509], value: 2 – A + B, 3 – A+B+C, 4 – A+B+C+D.
- Complete the serial port configuration by suitably setting the baud rate (SERIAL PORTS COM2
BAUD RATE [1505]) and frame format (SERIAL PORTS COM2 FRAME FORM. [1506]).
Once the communication between the various instruments has been set, it will be possible to view the
sum / difference value on the display of each instrument (see section VIEWING ON THE DISPLAY).
Finally, configure the desired setpoint behavior for each instrument, in particular act on the parameter
INPUT / OUTPUT OUT (1-3) LOAD SEL., in order to determine the behavior of the output as a
function of the weight values considered.
- 26 -
RF CONNECTION (OPTIONAL)
The multi-device connection can also be made via an on-board RF module (optional). If the module
is installed, its configuration items are shown in the SERIAL PORTS menu.
With reference to the section SERIAL PORTS PROGRAMMING and the section OUTPUTS
PROGRAMMING, acting on each instrument, configuration involves the following steps:
- Set the same RF-CHANNEL on each instrument that you want to connect using the
parameterSERIAL PORTS RF CHANNEL SEL [1514].
- Make sure you have set the same number of decimal places, correct accordingly. WEIGHING
CONST. DIVISION VALUE [1106].
- Enable the Sum-Difference protocol: SERIAL PORTS COM3 PROTOCOL [1512], value
SUM/DIF.
- Differentiating each instrument, set its address according to the solution that is being created:
SERIAL PORTS COM3 ADDRESS [1513]: A, B, C, D.
- Enter the total number of devices required using the parameter SERIAL PORTS SUM UNITS
NUMBER [1509], value: 2 – A + B, 3 – A+B+C, 4 – A+B+C+D.
- Complete the serial port configuration by suitably setting the baud rate (SERIAL PORTS COM3
BAUD RATE [1505]) and frame format (SERIAL PORTS COM3 FRAME FORM. [1506]).
Once the communication between the various instruments has been set, it will be possible to view the
sum / difference value on the display of each instrument (see section VIEWING ON THE DISPLAY).
Finally, configure the desired setpoint behavior for each instrument, in particular act on the parameter
INPUT / OUTPUT OUT (1-3) LOAD SEL., in order to determine the behavior of the output as a
function of the weight values considered.
- 27 -
SERIAL COMMUNICATION PROTOCOLS
SLAVE PROTOCOL
Communication is always initiated by the PC. Transmit the described commands as ASCII strings. On
each generic command received correctly and, with an address corresponding to that of the device,
an acknowledgment of receipt is sent:
- In the case of a successful command: & & a a ! \ ck ck CR
- In case of a command received but with an error: & & a a ? \ ck ck CR
where:
- & & = 2 c. of string start (38 ASCII)
- a a = 2 c. of instrument numeric address
- ! = 1 c. of ok (33 ASCII)
- ? = 1 c. of ko / error (63 ASCII)
- \ = 1 c. of separation (92 ASCII)
- ck ck = 2 c. of check-sum (see section CALCULATION OF CHECK-SUM)
- CR = 1 c. of string end (13 ASCII)
If the command requires the device to return additional information such as a weight value, the specific
answer will be described in the relevant section.
On each request received via serial protocol during a fault condition among the listed ones, the device
will respond with a warning concerning the fault:
- Load cell nominal capacity exceeded = & a a s s O-L s t \ ck ck
- Incorrect load cell connection = & a a s s O-F s t \ ck ck
where:
- a a = 2 c. of instrument numeric address
- s = 1 c. separator (32 ASCII, space)
- t = type of data required by the command received
where:
- $ = 1 character of string start (36 ASCII)
- aa = instrument address
- xxxxxx = 6 setpoint characters
- y = A (setpoint 1)
- y = B (setpoint 2)
- y = C (setpoint 3)
- ck ck = 2 check-sum characters (see section CALCULATION OF CHECK-SUM)
The setpoints set in this way are volatile, save them if you want to keep them after the device has
been restarted.
- 28 -
SAVING SETPOINT TO E2prom MEMORY
The PC-programmed setpoints are saved in RAM volatile memory and lost when the instrument is
switched off. To save them as final, send the following command:
where:
- j = a (setpoint 1)
- j = b (setpoint 2)
- j = c (setpoint 3)
- j = t (gross weight)
- j = n (net weight)
- j = p (peak)
The first character from the left of the weight takes the value " -" (minus sign – 45 ASCII) in the case
of a negative weight, e.g. -00082.
In the case of a request for zero-setting above the maximum resettable value, the response is: & a a # CR
The zero-setting will also be maintained after the instrument is switched off
- 29 -
SWITCHING FROM NET WEIGHT TO GROSS WEIGHT
where:
- a a = instrument address;
- x = number of decimals;
- y = division value.
In the case of a fixed zero, the field “ y ” takes the following values:
- '6' for division value = 10;
- '7' for division value = 20;
- '8' for division value = 50;
- '9' for division value = 100.
- 30 -
REPEATER/REPEATER C PROTOCOL
where:
- & = character of string start (38 ASCII)
- N = a net reference character (78 ASCII)
- x x x x x x = 6 ASCII characters of net or peak weight right justified (4857 ASCII).
- L = a gross weight reference character (76 ASCII)
- y y y y y y = 6 ASCII characters of gross or peak weight right justified (4857 ASCII)
- \ = 1 c. of separation (92 ASCII)
- ck ck = 2 check-sum characters. See section CALCULATION OF CHECK-SUM.
- CR = 1 c. of string end (13 ASCII)
The first character from the left of the net or gross weight takes the value " -" (minus sign – 45 ASCII)
in the case of a negative weight.
If the protocol “Repeater C.” is selected, the net weight, gross weight or peak will include the decimal
places separated by a point.
The total number of characters available includes any “.” or “-” symbols.
If the maximum number of characters is exceeded:
- in case of positive weight the value “^^^^^^“ is transmitted
- in case of negative weight the value “_ _ _ _ _ _ “ is transmitted
- 31 -
CONTINUOUS WEIGHT TRANSMISSION PROTOCOL
This protocol is used for continuous transmission, usually to the repeater panel.
Protocol compatible with RIPLED5100 repeater.
Baud rate 1200/115200, string transmitted 10 times per second.
where:
- STX = start of text (02 ASCII)
- <STATUS> = character coded as per table below (bit = 1 if condition = True)
- <WEIGHT> = field composed of 8 ASCII characters with the weight value justified right.
- Under conditions of weight above maximum capacity, it takes the value: “^^^^^^^^^“.
- In conditions with a negative weight greater than 9 divisions, it takes the value: “_ _ _ _ _ _ _ _“.
- EXT = End of text (03 ASCII).
- <CK> = 2 check-sum characters (see section CALCULATION OF CHECK-SUM).
- EOT = End of transmission (04 ASCII).
- 32 -
SUMMING PROTOCOL
String transmitted in case of activation of an input set to “Txd” operation. This transmission is
performed on each serial port that is set to “None” or “Slave” operation.
- <NET>: fields composed of 8 ASCII characters with the weight value justified right.
- Under overweight conditions, the field takes the value: “^^^^^^^^”.
- Under underweight conditions, the field takes the value: “_ _ _ _ _ _ _ _”.
- Under weight read error conditions, the field takes the value: “ O-L ”.
- ETX: end of text; data string end character (value 03 ASCII).
- <CK>: see section CALCULATION OF CHECK-SUM.
- EOT = end of transmission (04 ASCII).
- 33 -
CALCULATION OF CHECK-SUM
The check-sum consists of 2 ASCII characters (ck ck) obtained by following this procedure:
- For the calculation, consider only the significant characters in the string between the delimiter
characters, excluding the delimiter characters. For example, all the characters between
STX<…>ETX.
- Calculate the EXCLUSIVE OR or (XOR) of the ASCII character codes:
- the binary operation XOR gives the result 1 when the number of 1 of the operands is odd, otherwise
it gives the result 0.
Example:
character code Decimal ASCII code Hexadecimal ASCII code Binary ASCII
0 48 30 00110000
1 49 31 00110001
t 116 74 01110100
XOR = 117 75 01110101
- The result of the XOR operation expressed in hexadecimal notation is composed of 2 hexadecimal
digits (i.e. numbers 0 to 9 or letters A to F).
- The check sum consists of the 2 characters representing the result of the XOR operation in
hexadecimal notation (in our example the character "7" and the character "5").
Examples of calculated check sum: $01t75<CR> (address 1 weight required, slave protocol).
- 34 -
DECLARATION OF CONFORMITY - EU
Email: [email protected] Fabbricante metrico Prot. N. 7340 Parma - R.E.A. PR N. 169833 - Reg. Imprese PR N.19393 - Registro Nazionale Pile
Web: http://www.laumas.com N. IT09060P00000982 - Registro A.E.E. N. IT08020000002494 - N. Mecc. PR 008385 - Cap. Sociale € 100.000 int. vers.
SISTEMA QUALITÀ CERTIFICATO UNI EN ISO 9001 - SISTEMA GESTIONE AMBIENTALE ISO 14001 - MODULO D: GARANZIA DELLA QUALITÀ DEL PROCESSO DI PRODUZIONE
Dichiarazione di Dichiariamo che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito
I
conformità citate.
Declaration of
GB We hereby declare that the product to which this declaration refers conforms with the following standards.
conformity
Declaración de Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las
E
conformidad siguientes normas
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden
D Konformitäts-erklärung
Normen übereinstimmt.
Déclaration de Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est
F
conformité conforme aux normes citées ci-après.
Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými
CZ Prohlášení o shode
normami.
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde
NL Conformiteit-verklaring
normen overeenstemt.
Declaração de Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às
P
conformidade normas seguintes.
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi
PL Deklaracja zgodności
normami.
Заявление о Мы заявляем, что продукт, к которому относится данная декларация, соответствует
RUS
соответствии перечисленным ниже нормам.
- 35 -
DECLARATION OF CONFORMITY - UKCA
Email: [email protected] Fabbricante metrico Prot. N. 7340 Parma - R.E.A. PR N. 169833 - Reg. Imprese PR N.19393 - Registro Nazionale Pile
Web: http://www.laumas.com N. IT09060P00000982 - Registro A.E.E. N. IT08020000002494 - N. Mecc. PR 008385 - Cap. Sociale € 100.000 int. vers.
SISTEMA QUALITÀ CERTIFICATO UNI EN ISO 9001 - SISTEMA GESTIONE AMBIENTALE ISO 14001 - MODULO D: GARANZIA DELLA QUALITÀ DEL PROCESSO DI PRODUZIONE
Dichiarazione di Dichiariamo che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito
I
conformità citate.
Declaration of
GB We hereby declare that the product to which this declaration refers conforms with the following standards.
conformity
Declaración de Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las
E
conformidad siguientes normas
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden
D Konformitäts-erklärung
Normen übereinstimmt.
Déclaration de Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est
F
conformité conforme aux normes citées ci-après.
Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými
CZ Prohlášení o shode
normami.
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde
NL Conformiteit-verklaring
normen overeenstemt.
Declaração de Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às
P
conformidade normas seguintes.
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi
PL Deklaracja zgodności
normami.
Заявление о Мы заявляем, что продукт, к которому относится данная декларация, соответствует
RUS
соответствии перечисленным ниже нормам.
- 36 -
On our website www.laumas.com there are videos on the guidelines for correct installation of
weighing systems and video tutorials on configuring our transmitters and weight indicators.
All Laumas product manuals are available online. You can download the manuals in PDF format
from www.laumas.com by consulting the Products section or the Download Area.
Registration is required.