Comparator 3000 Series
Introduction
The two discs containing the relevant colour scale are brought together
by a prism and viewed simultaneously alongside the sample making it
easier to achieve an optimum match.
When seen through the prism, the sample is viewed in the centre. The
colour standards are viewed either side of the sample, as in the following
diagram:
Colour Colour
Standard Sample Standard
1
The Comparator 3000 Series are available to measure the following
colour scales.:
Information
The colour standards for use with the Comparator 3000 Series consist of
flat chemically produced glass. Unlike plastics, Lovibond® permanently
coloured glass standards are guaranteed not to fade with continuous
exposure to daylight and will give reliable accurate results every time.
The standards are mounted into a disc which is easily positioned in the
Comparator 3000 Series.
The standards are mounted in a disc for use with the Lovibond®
Comparator 3000 Series.
2
“Colorimetric Chemical Analytical Methods” – CCAM
This CD, now in its 9th edition, is a comprehensive reference for the
chemical methods used for the range of Lovibond® discs.
The Individual methods and techniques are clearly laid out, enabling the
reader to see the principles of the various chemical methods as well as the
actual techniques for carrying out the tests.
Technical Specifications
Specification Description
Light Source 12V 20 Watt Tungsten Halogen
Optical System Fully Prismatic
Sample Requirements Sample, Glass Cells and Lovibond®
Glass Colour Standards
Instrument Housing Moulded Fire Retardant ABS
Power Pack:
Input Voltage 100-240V
Output voltage 12V DC
Power Consumption 2 amps 20 Watts (Max).
Operating Temperature Range -10°C to +70°C
Dimensions: 231w x 273l x 142h
Weight: 1.5 Kg
Technical Assistance Technical Services Department
The Tintometer Ltd
Lovibond House
Solar Way
Solstice Park
Amesbury. SP4 7SZ
Tel: +44 (0) 1980 446800
Fax: +44 (0) 1980 625412
E-mail: [email protected]
3
Parts Diagram
Lid
Top
Sample
Halogen chamber
Lamp
Disc
On/Off
switch
Disc Knobs
4
Operating Instructions for the Comparator 3000 Series
All scales except ASTM color
5
Press Once = 1 minute
illumination
Press twice = 2 minutes
illumination
Press three times = 3 minutes
illumination
Press and Hold = Off
Colour Colour
Standard Sample Standard
Note: The best colour match is obtained by viewing with one eye closed
Rotate the discs by turning the knobs on the front of the instrument until
the oil sample matches or falls between two Colour Standards. A reading
can then be taken from the disc.
FAC / EBC:- To change the disc pair: open the lid, pull the knobs and lift
out the discs. To insert the discs: pull the knobs, insert the discs and
rotate the knobs until they engage in the discs.
6
Operating Instructions for measuring ASTM colour scale on the
Comparator 3000 Series
1/2
1/2
Press Once = 1 minute illumination
Press twice = 2 minutes illumination
Press three times = 3 minutes illumination
Press and Hold = Off
Colour Colour
Standard Sample Standard
8
Reading the Petroleum Oil - ASTM Discs
NOTE: The best colour match is obtained by viewing with one eye closed.
Rotate the discs by turning the knobs on the front of the instrument until
the oil sample matches or falls between two Colour Standard. A reading
can then be taken from the disc.
If the sample is not clear, heat to 6°C above its cloud point and measure
the colour at that temperature. Petroleum waxes including Petrolatum,
should be heated to 11°C to 17°C above the congealing point. If the
sample is darker that 8 ASTM color, mix 15 volumes of sample with 85
volumes of solvent kerosene.
If diluted add “DIL” after the colour e.g L7.5 DIL ASTM Color
9
Maintenance
Lift the top clear of the unit. The lamp housing is located on the
base.
Remove the old lamp and replace with the new in the same
position, using the protective sleeve.
Should a spillage occur in the sample chamber, simply remove the cell or
glass tubes, remove the sample chamber and clean.
Precision fused glass cells and moulded glass tubes are made to the
highest quality standards. The cells perform an important dual function in
the Comparator System. Not only do they serve to contain the sample
being analysed, but the intensity of the colour sample is directly
proportional to the optical path length of the cell. For this reason it is
important to use cells with a path length matched to the colour standards
and the disk.
Description Order
Code
Gardner / FAC Test Tube AF228T 35 22 80
AF763 Reference Tube (ASTM D1500) 35 76 30
W680 2.5 mm fused glass cell 60 67 80
W680 5 mm 60 67 90
W680 10 mm 60 68 10
W680 15 mm 60 68 40
W680 20 mm 60 68 50
W680 25 mm 60 68 60
W680 33 mm 60 68 80
W680 40 mm 60 68 90
11
Optional Extras
Description Order
Code
Tungsten Halogen Lamp 12V 20W 12 23 40
4/30-1 Gardner Comparator 3000 disc 24 00 10
4/30-2 Gardner Comparator 3000 disc 24 00 20
4/32A FAC/A Comparator 3000 disc 24 02 10
4/32B FAC/B Comparator 3000 disc 24 02 20
4/32C FAC/C Comparator 3000 disc 24 02 30
4/32D FAC/D Comparator 3000 disc 24 02 40
EBC/1 Comparator 3000 disc 24 04 50
EBC/2 Comparator 3000 disc 24 04 60
EBC/3 Comparator 3000 disc 24 04 70
EBC/4 Comparator 3000 disc 24 04 80
ASTM D1500 Comparator 3000 disc 0.5 - 7.5 24 02 80
ASTM D1500 Comparator 3000 disc 1.0 – 8.0 24 02 90
Certificate of Conformity for new discs (order with discs) QA1
Conformance Filter Recertification (Certificate of QD2
Conformity)
Colour Disc Calibration (ISO 9001 Calibration Certificate) CAL2
12