EksigentSoftwareUserGuide (Eksigent, Part of AB Sciex)
EksigentSoftwareUserGuide (Eksigent, Part of AB Sciex)
D5031167 C
September 2012
This document is provided to customers who have purchased AB Sciex equipment to use in the
operation of such AB Sciex equipment. This document is copyright protected and any
reproduction of this document or any part of this document is strictly prohibited, except as
AB Sciex may authorize in writing.
Software that may be described in this document is furnished under a license agreement. It is
against the law to copy, modify, or distribute the software on any medium, except as specifically
allowed in the license agreement. Furthermore, the license agreement may prohibit the software
from being disassembled, reverse engineered, or decompiled for any purpose.
Portions of this document may make reference to other manufacturers and/or their products,
which may contain parts whose names are registered as trademarks and/or function as
trademarks of their respective owners. Any such use is intended only to designate those
manufacturers' products as supplied by AB Sciex for incorporation into its equipment and does
not imply any right and/or license to use or permit others to use such manufacturers' and/or their
product names as trademarks.
AB Sciex warranties are limited to those express warranties provided in connection with the sale
of its products, and are AB Sciex’s sole and exclusive representations, warranties, and
obligations with respect to its products, and AB Sciex makes no other warranty or any kind
whatsoever, expressed or implied, including without limitation, warranties of merchantability or
fitness for a particular purpose, whether arising from a statute or otherwise in law or from a
course of dealing or usage of trade, all of which are expressly disclaimed, and assumes no
responsibility or contingent liability, including indirect or consequential damages, for any use to
which the purchaser may put the equipment described herein, or for any adverse circumstances
arising therefrom.
The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective
owners. Eksigent is a division of AB Sciex, LLC.
AB SCIEX™ is being used under license.
Eksigent
5875 Arnold Road, Dublin, CA 94568.
© 2012 AB SCIEX.
Printed in Canada.
Chapter 1 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Install the Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Configure the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Import Instrument Calibration Settings and Configure Software Drivers . . . . . . .11
Chapter 2 Acquisition Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Components of the Acquisition Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Status Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Control Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Spectral Absorbance Region . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Main Plot Region . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Peak Table (UV Detector Systems only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Contour Plot Region (UV Detector Systems only) . . . . . . . . . . . . . . . . . . . . . .16
File Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Output Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
View Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
UV Detector Toolbox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
PeakPark ToolBox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Smoothing Toolbox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
System Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
System Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Hardware Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Instrument Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Notification Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Appearance Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Mobile Phases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Direct Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
User Authentication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
User Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Electronic Signatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Analysis Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Help Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Chapter 3 Run Manager. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Current Tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
Method Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
Run Manager Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
Run Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
About Run Table Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Edit a Menu Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Edit a Text Box or Edit Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Select Multiple Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Edit Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Move Selected Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
5. Click Next to select a desktop icon and a Quick Launch icon. Refer to Figure 1-3.
6. Click Next to open the Ready to Install dialog. Refer to Figure 1-4.
Note: This validation can be run at any time from the list of programs that is
accessed from the Programs menu. Refer to Figure 1-6
Note: If the software does not detect the instrument, a warning message is
shown. Check that the instrument is turned on and that the serial cable is
properly connected.
4. Select Instrument Configuration from the System menu to open the Instrument
Configuration dialog. Refer to Figure 1-8.
Note: When the Eksigent device is controlled by an external software application such
as the Analyst® software, the Run Manager is not available. This button becomes an LC
Method Editor button and enables the user to edit methods.
Status Area
The status area shows information about the current run, including the flow rate, run time, current
method, sample information, pump state, injection valve position, the location in the sequence,
and the file name used to save the data after the run is completed. Refer to Figure 2-2.
1
Figure 2-2 Status area
installed, Pc and Pd), and the pump power as a percentage of maximum pump
power for each channel.
Control Buttons
Run Manager
The Run Manager button opens the Run Manager dialog, where sample and method information
is entered to create sample sequences. For more information, refer to Chapter 3.
Pause Button
When an acquisition is in progress, the Pause button maintains the conditions of the device at
their current values indefinitely. Clicking this button during a run maintains the elution conditions
continuing to acquire data. After being clicked, the Pause button changes to Resume and is used
to continue the LC method.
Stop Button
The Stop button stops the LC method and then stops the sequence table at its current position.
For multi-channel devices, only the currently active channel is stopped. The device channel is
placed into the Standby state and the pumps are turned off.
The vertical dashed lines in the spectral absorbance region indicate the absorbance wavelength
range used to calculate absorbance for the chromatogram. These values are initially defined in
the LC Method Editor. Refer to About Run Table Cells on page 46. The wavelength range can be
modified graphically by using the pointer to drag the dashed lines to a new wavelength range. To
modify only the upper or lower wavelength setting, hold the Ctrl key while dragging the relevant
dashed line.
The axes limits for the spectral absorbance region are defined by the upper and lower
wavelength ranges (X-axis) of the spectrometer and the current chromatogram axis limits
(Y-axis).
After data is collected, moving the pointer over the contour plot region updates the spectral
absorbance region with the absorbance spectrum at the indicated acquisition time.
File Menu
Table 2-1 lists the commands available from the File menu in the Acquisition window.
Table 2-1 File Menu
Command Function
Open Opens a previously stored data file (.txt,.dat) in the main acquisition
window. Only a single file can viewed at one time. The Peak Viewer utility
should be used to view multiple files or to open files while an acquisition
is in progress. Refer to Peaks Menu on page 63.
Save As Saves a data file to the hard disk in a .txt, .xls, .dat, or .cdf (AIA/ANDI)
format. A bitmap image of the spectral absorbance contour can also be
independently exported using this command.
Output Options Opens the Output Options dialog, allowing access to the file output
parameters. Refer to Output Options on page 17.
Peak Viewer Opens the Open dialog, allowing users to select a file or list of files. When
a file is selected, then the Peak Viewer dialog opens to allow users to
view each file. Refer to Peak Preview on page 19.
Print Preview Opens the Report dialog, refer to Print Preview on page 19.
Print Prints a report summarizing the current data, refer to Print on page 21.
Autosave Instructs the system to automatically save the data file at the end of an
analysis.
Autoprint Prints a report summarizing the current data at the end of a run.
Exit Opens the Close Program dialog.
Output Options
The Output Options dialog (Figure 2-4) is used to specify the preferred data file formats, paths,
and naming conventions to be used when the AutoSave option is selected.
Output
Output folder: Used to indicate the default folder where Autosave data will be stored. Additional
sub-folders can be created with the output filename entry. If the folder does not exist, users are
prompted with a choice for creating that folder.
Output filename(s): Used to specify filename attributes for each data set after the run has
completed. Users can customize the file naming convention by using one or more tags. These
may include the method name, sample name, date, time, vial number, incremental counter, and
so on. The default filename format uses a dynamically created date subfolder and creates a
unique filename based on the time:
\<date>\ EK< channel>_<time>
Many other unique filenames can be generated from the various tags. Click Help for more
information on the available tags or use the list to insert new tags.
When the user creates a naming convention, an example filename is shown.
Prompt to save: Select to prompt users to name the data file at the completion of each run. This
is not recommended when there is more than one sample in the Run Manager. Tags that use
information from the Run Table are shown in the example filename as either blank or as the value
last used with that tag.
Autosave: Select to automatically generate a data file name and save the file in the folder
specified in the Output folder. If this command is not selected, the user is prompted to save the
data file after each sample run. The use of the Autosave option is recommended.
If a unique filename is not specified in the Output filename(s) edit box and Autosave is selected,
a unique filename is created if the current filename already exists in the specified folder.
Ownership Tags
Operator Name: The name of the person who is running the experiment or test that generated
the current dataset.
Dataset Origin: The name of the organization where the current dataset originated. Dataset
Origin information can include the organization's address, telephone number, e-mail nodes, and
the names of individual contributors, including the system operators, and any other appropriate
information.
Dataset Owner: The name of the owner of a proprietary dataset. The person or organization
named here is responsible for the accuracy of the information in this field. Data copyrights should
be indicated here.
Peak Preview
The Peak Preview function provides access to the Peak Viewer for loading and analyzing larger
sets of sample data. Refer to Peaks Menu on page 63.
Print Preview
The Print Preview function is used to preview the appearance of a report before it is printed. Print
previews are generated in the Report dialog (Figure 2-5). The Report window is shown with a
representation of the report for the currently loaded data.
Print
Click Print to print the report for the currently loaded data using the Print Preview or the Print
dialog. The report for the currently loaded data is sent to the default printer specified for that
computer. The most recently selected report format is used.
View Menu
Table 2-2 lists the commands available from the View menu in the Acquisition window.
Table 2-2 View Menu
Command Function
UV Detector Toolbox Opens the UV Detector dialog, which allows for manual control of the
detection settings. Refer to UV Detector Toolbox on page 21.
PeakPark ToolBox Opens the PeakPark toolbox, which allows for setting peak parking
options, refer to PeakPark ToolBox on page 23.
Smoothing Toolbox Opens the Smoothing window, which provides access to the post-
acquisition data spline values. Refer to PeakPark ToolBox on page 23.
System Logs Opens the System Logs sub-menu to access audit trails of system
messages, and warnings. Refer to System Logs on page 24.
UV Detector Toolbox
If the system includes a UV absorbance spectrometer, the UV Detector Toolbox can be used to
view the current UV detector settings and make adjustments as necessary.
Note: This window is typically used for diagnostic purposes, as each LC method
contains detector settings that are specific to that method. As a result, many of the
settings in this toolbox are not accessible during acquisition as they are being controlled
by the parameters specified in the LC method.
Baseline Region
The Baseline region is used to indicate the parameters for collection of the baseline. If the Real-
time correction option is selected, users can indicate the Wavelength and Bandwidth.
Reference/Background
Use the Reference/Background options to turn on or off the Automatic background and reference
acquisition. If this option is selected, the detector background and reference signal are acquired
prior to every LC method acquisition. When the option is not selected, the Acquire reference and
Acquire background buttons can be used to manually acquire those signals. In this case, each
LC method acquisition starts immediately and the current background and reference information
(in memory) are used.
Detector Sync.
Use the Detector Sync. options to determine how to synchronize the detector acquisition. The
Sync. with injection start option indicates that the detector should start acquiring data (time = 0)
when the injection valve moves to the inject position. The Sync. with gradient start option
indicates that the detector should start acquiring data (time = 0) when the injection has
completed and the gradient (flow profile) begins. This mode is useful when comparing retention
time reproducibility with varying injection volumes (binary gradient LC methods with metered
injections.)
PeakPark ToolBox
Eksigent LC devices can rapidly reduce the total column flow rate to allow for extended MS/MS
analysis at a particular point in the chromatogram. This procedure, which is commonly called
peak parking, is controlled using the PeakPark ToolBox (Figure 2-7). Adjustments to the
composition of the flow profile are paused during the parked period and resume after parking is
discontinued. Peak parking is only enabled during an acquisition.
• The Column flow rate reference method uses the measured column pressure at the
start of the PeakPark operation to estimate the current column flow resistance. This
value, with the real-time measured column pressure is used during the peak-park
transition to stabilize the column at the new flow rate. This method is typically the
fastest method to attain the target peak-park flow rate values.
• The Mobile Phases flow rate reference method uses the flow rates measured
directly from the individual mobile phase (A and B) flowmeters while the pump flow
rates are decreasing. This method is typically slower to lower the total flow rate
through the column, but more accurately maintains the mobile phase mixture
fraction.
Timing check boxes can be used to ignore external triggers or peak-park for a fixed duration after
the initial trigger event. This option is useful when noisy trigger signals are present (from the
mass spectrometer) and ensures that the device will stay in park mode for a configurable
acquisition period.
Smoothing Toolbox
Use the Smoothing Toolbox (Figure 2-8) to apply a cubic spline fit to the chromatogram after
acquisition is completed. The error band defines the looseness of fit in Y units, where a larger
number applies more smoothing. The Smoothing Toolbox can be used with systems with a
spectrometer.
System Logs
Use the System Logs sub-menu to show a variety of log files, which are audit trails that record all
pertinent system information, including ordinary usage statistics, system warning messages, and
experiment status logs. Log files are designed to prevent malicious tampering with their contents
to make sure of the long-term validity of the system audit trail. Each log file is archived on a
regular basis.
View a Log
• To view the contents of a log in the log window, click any available log in the left
column. The entries of that log are shown in the pane to the right.
• Alternatively, click Select Log to open the Open File dialog and then select a log
file.
Sort a Log
• To sort the log contents, click any column heading. The log entries are sorted based
on that column. Ascending sort is always used when the column is first clicked.
• To use descending sort, click the column heading again.
Log entries often appear in alternating lines of black and blue. The coloring scheme
is to help locate identical entries within a field. The color of the entries is based on
the column on which the sort is applied. For example, the log entries in Figure 2-10
are sorted based on date. Starting at the top, the color alternates when a different
value is detected. The first line is always black. On the second line, the date value is
different, so the color alternates to blue. On the third line, the date value is the same
as the second, so the color stays blue. On the fourth line, the date value changes
again, so the color alternates back to black. This pattern continues to the end of the
entries.
The Log Information field shows some basic information about the selected log,
including creation date, last modified date, and the file location of the log file.
• To exit the dialog, click OK.
Export a Log
The contents of any log can be exported to a PDF file or to an Excel spreadsheet (.xls).
• Select an Export format in the Log Information field and then click Export.
• If the Excel option is selected, a Windows Export to dialog opens. Type a file
name for the log and then click Save.
• If the PDF option is selected, a PDF preview window opens. Use the Portrait
or Landscape buttons to change the page orientation. Click Save to open a
Save As dialog to save the PDF file. Use the page navigation arrows to
navigate through multiple-page PDF documents. Use the magnifying glass
icon to switch between two best-fit views of the document. To print the PDF,
click the printer icon.
The Alerts log (Figure 2-10) contains a list of any warnings and messages automatically
generated by the system but which do not need immediate attention. The Alerts Log dialog opens
automatically when a system check discloses a potential issue and shows the issue as an Alert
event. The log includes information such as the Login name of the current user, the date of each
occurrence of the Alert event, and a detailed description of the event.
The System log (Figure 2-11) contains a list of user actions that affect the device, such as
calibration options.
The Channel01 log (Figure 2-13) contains a list of user actions that affect Channel 1 conditions,
such as flow rate. For multi-channel instruments, channel log information is available for each
additional channel.
System Menu
Table 2-3 lists the commands available from the System menu in the Acquisition window.
Table 2-3 System Menu
Command Function
Hardware Diagnostics Opens the Hardware Diagnostics dialog. Refer to Hardware
Diagnostics on page 28.
Instrument Opens the Instrument Configuration dialog. Refer to Instrument
Configuration Configuration on page 30.
Notification Options Opens the Notification Options dialog. Refer to Notification Options on
page 33.
Appearance Settings Opens the Appearance Settings dialog for customizing the data. Refer
to Appearance Settings on page 33.
Mobile Phases Opens the Mobile Phases dialog to describe the mobile-phases and
controlling the purge and flush functions. Refer to Mobile Phases on
page 36.
Direct Control Opens the Direct Control dialog, which allows for manual control of the
pumping system and injection valve. Refer to Direct Control on
page 38.
User Authentication Opens the User Logon dialog. Refer to Enable User Authentication on
page 129.
User Manager Opens the User Logon dialog, followed by the User Manager dialog.
Refer to User Manager on page 39.
Electronic Signatures Opens the Electronic Signatures dialog. Refer to Electronic Signatures
on page 40.
Hardware Diagnostics
Use the options on the Hardware Diagnostics dialog (Figure 2-14) to calibrate the LC hardware,
examine instrument usage statistics, and perform automated diagnostic routines.
.
Usage Information
• Total Sample Injections shows the total number of injections performed. This number
is useful for tracking instrument usage. Click CLR to clear the counter.
• Total Flowmeter Usage shows the total number of hours of flowmeter usage. This
number is useful for tracking instrument usage. Click CLR to clear the counter.
• Filter Usage shows the volume of fluid that has passed through the cell guard filter.
Click CLR to clear the field when the filter is changed.
Flow Region
The Flow region on the Calibration Values tab (Figure 2-15) shows fluid-independent and fluid-
dependent K values used for flow control. The actual K values may slightly change every few
seconds because they are calculated based on measured temperature.
Pressure Sensors
The Pressure Sensors region shows the pressure sensor calibration values.
PID Region
The PID region shows calibration values relating to the pump control parameters.
Instrument Configuration
System Tab
Use the Instrument Configuration dialog to set hardware configuration options. Usually these are
set during the initial installation, but they may be set any time the system configuration changes.
Advanced Tab
The Advanced Tab is shown in Figure 2-18.
• Use the Stop device if pump power remains at 100% option to set a time duration
that will stop the pumps and data collection if pump power remains at 100% for a
specified period of time. Pump power can remain at 100% for unusually long
durations if pump flow-control logic is attempting to compensate for leaks or
blockages in the flow path.
• Use the Stop device if Pc exceeds option to set an output column pressure sensor
(Pc) value that will stop data collection if the outlet pressure exceeds a specified
limit. High values of Pc can indicate a clogged column, post-column filter, or flow cell.
• Use the Stop device if Pc is below option to set an output column pressure sensor
value that will stop data collection if the outlet pressure is below the specified limit. A
low Pc value may indicate that there is a leak, the fitting is loose, and so on.
• Use the Rapid-Inject pressure limit option to protect the system from over-pressure
during a Rapid Inject process. This value is usually lower than the absolute column
pressure limit. The system flow rate is reduced as the system pressure approaches
this limit and is maintained just below the designated limit until the injection is
complete. After injection, the flow rate returns to the value specified in the analytical
method.
Notification Options
Use the Notification Options dialog to set e-mail notification options.
E-mail Settings
For notification of various events via e-mail, select the Notify by E-mail option and then indicate
the address and server. The Test button is used to verify that communication has been
established.
Notification Events
• LC device error allows the user to choose to receive an e-mail if the LC device
reports an error.
• Autosampler error allows the user to choose to receive an e-mail if the autosampler
device reports an error.
• RunTable complete allows the user to choose to receive an e-mail when the
instrument has completed acquiring all samples specified in the Run Manager
sequence.
Notification Frequency
Use the Notification Frequency region to select whether to receive e-mail notifications
immediately, hourly, daily, or not at all.
Appearance Settings
The Appearance Settings dialog (Figure 2-19) allows the user to determine the attributes of data
plotted in the main window.
A/D Channel
The Aux A/D (mV), scale X option allows the user to show the analog signal from an auxiliary
device and to select the color. The user can also designate a multiplier to be used to magnify the
analog signal plot. The offset value is a number added to the measured value.
Plot Axes
The Apply axes settings to all channels option allows the user to set the scale of the time axis to
vary automatically to fit the amount of data available to be shown.
More Options
Click More Options to open the Chart Options dialog. Select and order the columns for labeling
peaks on chromatograms, specifying plot colors, and so on that will be shown in the Main Plot
region.
Mobile Phases
Use the Mobile Phases dialog to configure the A and B operating fluids, document any additives,
and reset the usage counters. Accurate flow viscosity correction, flow rate measurement, and
flow stability depends upon correct identification of mobile phase fluid compositions in the Mobile
Phases dialog.
Direct Control
The Direct Control dialog (Figure 2-23) allows users to start the pumps at a specific flow rate.
Direct control is useful for operations such as system flushing, equilibrating, and tuning an
electrospray interface.
User Authentication
User authentication is used to set user access to the software. This allows applying varying
permissions to several users. For more information, refer to Chapter 9.
User Manager
The User Manager dialog is used to set different permissions for various users. It is described in
Chapter 10 and is only relevant if User Authentication is enabled.
Electronic Signatures
Electronic Signatures are used to view, apply, and revoke electronic signatures from data files.
Refer to Chapter 11.
Analysis Menu
Table 2-12 lists the commands available from the Analysis menu in the Acquisition window.
Table 2-12 Analysis Menu
Command Function
Settings Opens the Analysis Settings dialog, which contains peak detection
parameters. Refer to Chapter 7.
Choose Peak Table Prompts the user to select which peak analysis parameters to display
Columns in the peak table that is shown after a chromatogram is analyzed.
UV Detector Signal Analyzes the UV data trace when an analysis is performed.
AUX A/D Signal Analyzes the A/D data trace when an analysis is performed.
Automatic Automatically analyzes data at the conclusion of each sample data
collection.
Re-Analyze Now Analyzes the chromatogram using the current analysis method settings
and show the results on the screen.
Remove Results Removes the analysis results (peak table) from the screen.
Use the Analysis menu to adjust the parameters that control the analysis process. These
parameters are used by the peak-finding algorithm when identifying, placing, and integrating the
peaks found in the chromatogram. The Analysis menu can also be configured to automatically
analyze a chromatogram after acquisition is complete.
Use the Analysis Settings dialog to set peak integration parameters. These settings determine
which peaks the integration routines recognize during peak finding.
Help Menu
Table 2-13 lists the commands available from the Help menu in the Acquisition window.
Table 2-13 Help Menu
Command Function
Help Opens the interactive help system.
Check for Updates Opens the default Internet browser and selects a web page to show the
latest version of software.
Email Tech Support Opens the default e-mail client software and fills in the address and
system information to be sent to Eksigent Technical Support.
www.eksigent.com Accesses the Eksigent website with the default Internet browser.
About Opens installed software and firmware versions.
Current Tray
The Current Tray (lower left corner of Figure 3-1) shows an image of the configured autosampler
tray type and indicates active sample vials and their states.
Vial state:
• Green: Sample is currently running.
• Blue: Valid method associated with this vial.
• Blue/Checked: The sample finished successfully.
• Red: An error occurred while running this vial or the sample was stopped.
• Yellow: Vial is selected.
Autosampler
The Autosampler area indicates the currently selected (or active) tray and provides a text
description of the current autosampler action or status. The tray shown can be selected using the
arrow buttons next to the tray name in the Status area or by choosing a row in the table with the
desired tray.
For autosamplers with tray cooling, the tray temperature appears in the tray temperature area.
• The Pause button holds the autosampler in its current state or position. If an
autosampler method is running, the Pause button holds the method and stops the
autosampler in its current action.
• The Resume button allows the method to continue when the system has been
paused. This action is useful for changing sample trays while the autosampler is
running.
Method Definitions
Autosampler Methods: Opens the autosampler method editor that is used to create and edit
methods specifying the amount of sample to be drawn from the sample vial and the parameters
to be used for pre- or post-injection syringe cleaning. The editor shown depends on the type of
autosampler configured. CTC PAL Autosampler Control or NanoLC-AS1 Autosampler Control
editors are currently supported.
LC Methods: Opens a window for typing the parameters used for separating and detecting the
sample. This information includes flow rate, gradient profile, detection wavelength, and so on,
and is stored in a sample method. The method parameters available depend on the configuration
of the LC device. 1D LC methods and 2D methods are available.
Analysis Methods: Opens an Analysis Method Settings dialog that is used to set the peak
detection configuration parameters.
Run Table
The sample grid in the Run Table allows the user to link together an autosampler method, a
sample tray, a sample vial, an analysis method, sample method overrides, and sample
information such as injection amount, sample name, ID, type, and any other comments. Each
sample is represented by a single row.
Table 3-6 Run Manager Columns
Column Description
Seq. # The row number of the sample in sequence.
Run Indicates which samples are to be processed in the sequence.
(unchecked rows will be skipped).
Run Interval Time field for typing a sampling run interval. Time format is
hh:mm:ss. The specified time is the time between the start of each
acquisition. Blank values are treated as no interval.
Run Replicates Text field for specifying the number of replicates for each row in the
Run Table. Blank values are treated as a single run.
Autosampler Method Selects the autosampler method.
Autosampler Tray Selects the tray location of the sample vial or well.
Autosampler Vial Selects the vial to be sampled.
Autosampler Aspirate Specifies an override value for the aspirate volume in the
(µL) autosampler method. Leave blank to use the value specified in the
LC method.
When an acquisition is in progress, the color of the sample row indicates the state of its progress:
• Light green: The sample is preparing to run (equilibrating).
• Dark green: The sample is running.
• Red: The sample has stopped.
• Yellow: An error occurred.
• Gray: The sample has completed.
After acquisition has been completed, the table can be set to the pre-run state using the Reset
Table button. The sequence begins with the first sample row the next time it is started.
Cells in the table can be navigated with the arrow keys on the keyboard or by selecting cells with
the pointer.
Edit Cells
Use the following functions as required:
• In the Acquisition window, click Run Manager.
• After one or more rows are selected, right-click to open a menu to cut, copy,
paste, insert new rows, insert copied rows, insert series of rows, or delete
selected rows.
• Cut, Copy, and Delete operate on the selected rows. Paste puts copied or cut
rows into the currently active cell.
• Paste a single copied row into multiple rows by selecting the destination rows
prior to pasting.
Choose Columns
1. In the Acquisition window, click Run Manager.
2. Right-click the column labels at the top of the table and then click Columns.
Freeze Rows
Freeze rows in the Run Table so that they remain in the visible area while the rest of the table
area is scrolled.
1. In the Acquisition window, click Run Manager.
2. To freeze rows at the top, move the pointer over the border between the fixed
column label rows and the scrollable area until the pointer changes to a lock mouse
cursor.
3. Left-click and then drag the border down any number of rows to freeze. Any
movement or scrolling in the grid area does not affect the frozen rows.
Freeze Columns
Freeze columns in the Run Table so that they remain in the visible area while the rest of the table
area is scrolled.
1. In the Acquisition window, click Run Manager.
2. To freeze columns at the left of the grid, move the pointer over the border between
the fixed column and the scrollable area until the pointer changes to a lock mouse
cursor.
3. Drag the border across any number of columns to freeze. Any movement or scrolling
in the grid area will not affect the frozen rows.
Select a Vial
1. In the Acquisition window, click Run Manager.
2. Click a vial in the tray image to select the vial for the current sample row. If the active
row in the sample table is valid, the vial numbers are updated to reflect the vial
number selected in the graphic.
Add Vials
1. To add many new samples to the sample table, drag the cursor on the tray to select
multiple vials.
A Vial Selection dialog (Figure 3-6) opens to indicate how the sample table rows will
be filled. The new rows duplicate the information of the row number indicated in the
Current Table Row text box, but include the newly inserted vial numbers. The default
entry is the currently selected row in the table.
2. Do one of the following:
• Click Insert rows to add the newly selected vials into the table beginning with
the currently active cell.
• Click Overwrite existing rows to add the entries starting at the currently
active table cell and then fill down over the existing entries.
Tip! After all acquisitions are complete, to return the table to the initial
state, click Reset Table. The sequence begins with the first sample row the
next time it is started.
The Elapsed Time indicator shows the approximate time elapsed from when the Run
Table started. The Queued Time indicator shows the approximate time required to
run all of the active samples in the Run Table. This time is an approximation based
on the LC methods only and does not include additional time required for
autosampler methods.
Plot Dialog
The Plot dialog can be configured to show one or multiple data sets simultaneously. For systems
with spectrometers, the chromatogram (UV absorbance data) is shown. All other systems show
the AUX A/D signal (for use with external detectors).
Peak Table
When the utility starts, the peak table indicates data source information related to the loaded data
files and any peak information stored in the data file (resulting from automated analysis following
the acquisition).
The peak table shows the current peak analysis results for each data set and can be used to edit
(remove) peak information directly. A sample report, peak statistics, and peak trends are also
accessible from the peak table.
Analyze Peaks
By default the analysis parameters stored in the data files are used for the peak analysis. If the
data has not been previously analyzed or if there are no peaks detected, the table is empty
except for the data source information. For more information, refer to Peak Detection Tab on
page 82.
1. In the Acquisition window, click File > Peak Viewer.
2. Browse to files either in the Eksigent data or text format.
3. Click Open to load the data into the Peak Viewer.
Tip! The Peak Viewer can also be started directly from the installed
Program Group menu.
Note: The results in the data file are not permanently changed. However,
the new peak attributes can be saved.
5. Select the files in the table and then click File > Save Changes.
Sort Data
1. In the Acquisition window, click File > Peak Viewer.
2. Browse to files either in the Eksigent data or text format.
3. Click Open to load the data into the Peak Viewer.
4. Click the relevant column header labeled with the peak number.
The rows are sorted in ascending or descending order in that column each time the
column heading is clicked.
Tip! To sort the information by peak area, select Show Area and then
click the relevant peak number at the top of the table.
Select Data
Data in the table can be selected in various ways:
• Drag across a range of cells.
• Click the row label (first column) to select entire rows.
• Hold the Shift key and then click a range of cells or rows.
• Hold the Ctrl key and then click multiple cells or rows in any order.
Copy Data
Any range of information in the table can be copied to the Windows clipboard.
1. In the Acquisition window, click File > Peak Viewer.
2. Browse to files either in the Eksigent data or text format.
3. Click Open to load the data into the Peak Viewer.
4. Select the data, right-click, and then click Copy. The cell contents can be pasted into
any other application.
Delete Data
1. In the Acquisition window, click File > Peak Viewer.
2. Browse to files either in the Eksigent data or text format.
3. Click Open to load the data into the Peak Viewer.
4. Select the peak in the table and then press the Delete key.
Mark Values
Peak information in the table can be color-coded to easily identify trends in the data.
• To mark the data, type a value in the Mark Values field and then select a sign
(greater than, less than, equal) to mark all values in the table that meet the specified
criteria. A slight color gradient is used to differentiate the range of values.
Print a Report
• To print the peak table, sample report, or statistics, click File > Print.
Export a Report
• Click File > Export.
Statistics
Statistics are generated from all of the open files and for the peak attribute currently shown in the
peak table. The statistics table, shows displays the mean, median, relative standard deviation
(RSD), maximum, and minimum value generated for each peak number is at the top of the
Statistics tab. The chart below the table indicates the mean and the one and three sigma range
for the current peak attribute with the data source number on the abscissa.
Generate Charts
A chart can be generated for each individual peak. The statistics table and chart are useful for
examining the reproducibility of any of the peak attributes within a set of data files. The table and
chart are regenerated following any modifications to the peak attributes in the table or
chromatograms.
1. In the Acquisition window, click File > Peak Viewer.
2. Browse to files either in the Eksigent data or text format.
3. Click Open to load the data into the Peak Viewer.
4. Select the corresponding row in the statistics table.
In case of a numerical data source label, for example relative retention time, the
abscissa shows the actual numerical value.
Tip! All charts and table entries can be copied to the clipboard by right-
clicking and selecting Copy.
File Menu
Table 4-1 lists the commands available from the File menu.
Table 4-1 File Menu
Command Function
Open Used to load new data sets into the Peak Viewer. The existing peak
information and plots are removed and the new file data is used to
populate the peak viewer. No changes to the previously loaded data are
stored.
Add Files Used to add new data files to the current data set. The new data is added
to the end of the peak table (and plots windows).
Save Changes Used to save changes to the peak information for the open data files.
Modifications to the peak information are stored in the data files.
Only the selected files are updated. If no files are selected, a prompt to
save all files opens.
Export Used to export the current peak table, sample report, and statistics
information in various Windows file formats (Excel, tab-delimited text, and
so on).
Print Preview Used to preview the report before it is printed. The format of the report is
similar to the one generated in the Acquisition window. It contains file,
sample and method information, a plot of the chromatogram, and the set
of peak results is shown in the sample report.
Print Used to print the selected plots, current peak table, sample report,
statistics, and trends information on the default printer.
Print All Used to print all information for the data set, including the peak table, the
sample report, statistics, trends, and plots.
Monitoring Used to monitor samples in real time as they are completed from the Run
Table. When selected, the results of each sample are automatically
loaded into the Peak Viewer software, followed by an update of all charts,
trends, and reports. If not already open, it also loads the Run Manager
window, which allows users to queue and acquire a series of samples.
Monitor Folder As an alternative to monitoring the Run Table, this function observes a
specified folder for new files (in the text output format). After detecting a
new file, it is opened and analyzed automatically by the Peak Viewer
software, followed by an update of all charts and reports.
Skip First Sample Select this option to ignore data from the first sample acquired.
When selected, the first file after the user starts monitoring the Run Table
or a folder is skipped.
Exit Closes the Peak Viewer utility.
Edit Menu
Table 4-2 lists the commands available from the Edit menu.
Table 4-2 Edit Menu
Command Function
Copy Table Copies the selected section of the currently shown table to the clipboard.
Selection Sections of the peak table can be selected freely by dragging in the table
cells of interest. For the summary report and statistics table, click a row to
select the entire row.
Copy Plot Selection Copies the current selected plot to the clipboard. To select a plot, click the
desired plot or click anywhere in the peak table row for the data file.
Users can also select the plot, right-click, and then click Copy.
Select All Selects the entire table that is currently shown.
View Menu
Table 4-3 lists the commands available from the View menu.
Table 4-3 View Menu
Command Function
Full Screen Sets the plot area to show a single data set. The current active data file
(selected via plot window or peak table grid) is used to fill the entire plot
area. Full Screen is the default setting when the Peak Viewer utility starts.
Tile Rearranges the plot area to show multiple data sets in a tiled window
format. Use the right scroll-bar to scroll through the plots. Any data files
added to the data set are appended at the end. An example is shown in
Figure 4-6.
Peaks Menu
Table 4-4 lists the commands available from the Peaks menu.
Table 4-4 Peaks Menu
Command Function
Analysis Settings Provides access to the Analysis Method Editor. Refer to Chapter 7. Users
can also click Settings to open the Analysis Method Editor.
Import Analysis Imports analysis settings from an existing analysis method file and
Settings applies the settings to each selected file.
If no files (rows) in the table are selected, a prompt to apply the settings
to all files opens.
Find Peaks Analyzes the current data set. Each selected file is processed in
sequence using the analysis settings stored in the file.
If no files (rows) in the table are selected, a prompt to process all files
opens.
Plots Menu
Table 4-5 lists the commands available from the Plots menu.
Table 4-5 Plots Menu
Command Function
Plot Series Use this option to choose the data source for the plot data (that is, UV
data, A/D data, column pressure, and so on).
Overlay Chromatograms
1. To overlay chromatograms, in the peak table, select the data file rows using one of
the following functions:
• Hold the Shift key and then select multiple sequential rows.
• Hold the Ctrl key and then select multiple rows in any order.
• Drag an area to select rows in a range.
2. Click Plots > Overlay Selection.
The topmost data file in the peak table selection is used as the active chromatogram
and the plot window is updated with overlays of the data from the selected data files.
Refer to Figure 4-11.
Note: Depending on the configuration of the LC device, it may not be equipped with an
absorbance detector.
Create an LC Method
1. In the Acquisition window, click Run Manager.
2. In the Run Manager dialog, click LC Methods.
3. Type a name in the Name field.
4. Use the information in the following sections to create the LC method.
5. Click Save.
Edit an LC Method
1. In the Acquisition window, click Run Manager.
2. In the Run Manager dialog, click LC Methods.
3. Select a method in the Name field.
4. Use the information in the following sections to create the LC method.
5. Click Save.
Summary Tab
The Summary tab is used to type descriptive information about the column used for a particular
run (Figure 6-1). In addition, it summarizes the information from the other tabs in the LC Method
Settings dialog. The Sample Injection, Flow Profile, and Detection fields show information about
system conditions and are not editable.
Selected Method
Shows the name of the LC method. The user can select an existing method to view or edit or the
user can type a new name.
Method Identification
Used to type comments or to assign an identification number to the LC method.
Column Information
(Optional) Contains fields that can be used to describe the column used for a run.
• Manufacturer: The name of the manufacturer of the column used for the LC
method.
• Type: The description of column used in the current run. An example might be C18,
CN, and so on.
• Serial Number: The serial number of the column used with the method.
• particle size: The particle size of the column stationary phase.
• diameter: The column inner diameter.
• length: The length of the column used with the LC method.
Detection
For devices that include a spectrometer, the Detection region shows the details of detection data
collected during the run. Information includes data rate, wavelength range, and data file size.
Pre-Run
• Flush column for _ minutes using _% initial flowrate conditions: Select this
check box to equilibrate the column prior to the run for a defined duration and flow
rate using the method initial mobile phase composition. The specified duration is the
minimum time that the column equilibrates before acquisition begins. If the specified
duration elapses and the acquisition has not started (for example, column
temperature has not stabilized), the device equilibrates indefinitely.
• Stabilize column temperature at _ºC prior to injecting sample and beginning
Flow Profile: For systems with column heaters, select this check box to set the
temperature (ºC) for the column heater. Acquisition does not start until the column
reaches within 0.1ºC the specified temperature.
Sample Injection
Select a sample injection mode.
• None: The sample valve does not actuate during the acquisition.
• Standard: The sample valve switches to the Inject position when the acquisition
begins and then switches to the Load position when the acquisition ends. As a result,
the sample loop remains in the flow path during the acquisition.
• Metered: Offers the flexibility to vary the injection volume. The valve is switched to
the Inject position when the acquisition starts and the specified volume of sample is
metered to the column by the LC pumps at the specified conditions. After the
specified volume is injected, the sample valve switches to the Load position so the
sample loop is removed from the flow path during the acquisition.
• Rapid: Operates similarly to Metered injection mode option, but the LC pumps
steadily increase the flow rate during the injection (maintaining the initial mixture
fraction) to quickly meter a larger sample volume.
Post-Run
Select this check box to rinse the column after the run is complete for a defined duration and flow
rate using the method ending mobile phase composition.
Flow Mode
• Conserved flow: Select this option to maintain a constant total flow rate for the
entire duration of the acquisition while varying mobile phase composition.
• Independent flow: Select this option to specify the independent flow rates of each
mobile phase.
Profile Editor
Insert values at different points in time to create a mobile phase composition profile. Select A or B
from the option box or click the profile on the graphic to edit either the A pump or B pump flow
profile.
• Runtime: The total analysis time specified in minutes.
• Total flowrate: The combined flow rate of both A and B mobile phases. This value is
maintained constantly for the duration of the LC method run.
Note: Total flow rate is not shown for Independent Flow Mode.
• point: The point selector scrolls through the programmed points of the profile.
Programmable Events
For device firmware versions X.39 or greater, events are supported by the LC Method Editor
using the Gradient Profile tab.
Program an Event
• In the Gradient Profile tab, right-click in the chart area to show the Event menu, from
which one of 12 possible events may be selected (Figure 6-4). Up to 32 Events can
be inserted into the method. The event is inserted at the time located in the X axis
where the right-click was performed. The event may be moved by dragging the
cursor over the event flag icon.
Delete an Event
• To delete an event, right-click the event flag icon.
Note: Total flow rate is not shown for Independent Flow Mode.
Detector Tab
The Detector tab is available for instruments that include a spectrometer (Figure 6-6). This tab is
used to set up the spectral range that defines the main chromatogram, as well as the rate at
which spectra are acquired.
Data Acquisition
• Approximate acquisition rate in Hz: Specifies the data acquisition rate used to
collect the absorbance signal. For instruments equipped with an absorption
spectrometer, an entire absorption spectrum is measured at each time point.
• Average reference/background acq’ns: Specifies the number of data points
averaged together to calculate the spectrometer dark current and reference signals
prior to initiating acquisition. The default value is four.
Chromatogram
The Chromatogram region is used to indicate which mode of spectrometer acquisition (Signal
Type) to use. Absorbance is the default selection as it is typically used for HPLC detection.
However, for diagnostic purposes, the user can select either Transmission or Detector counts for
monitoring system performance.
Absorbance
Defines the parameters used to show the single wavelength chromatogram in the Acquisition
window. Two fields specify the spectral window wavelength range by defining a center
wavelength and a spectral band height.
• Wavelength: Corresponds to the center wavelength of the spectral window. It can
be set to within 1 nm of the specified detector wavelength range (for example, 201-
C379 nm) as long as the bandwidth setting keeps the spectral window within this
range.
• Bandwidth: Adjusted to set the wavelength range used to generate the
chromatogram. Wider bandwidth can be used to reduce baseline noise, but large
bandwidths can compromise detector response linearity. The bandwidth value is
limited by the acquisition wavelength range. For example, if the wavelength is set to
205 nm, the bandwidth can be no greater than 10 nm. The minimal bandwidth setting
is 1 nm.
Mode
Select between Single beam and Dual beam modes of operation.
• Single beam: The detector uses a detector reference signal acquired at the start of
the acquisition for the entire chromatogram.
• Dual beam: The reference signal is continuously measured with an independent
spectrometer. In both cases the reference signal uses the center wavelength and
bandwidth specified.
• Realtime baseline correction: Select the check box to enable real time baseline
correction. In this case a region of the spectrum, typically on the red (high
wavelength) end, is used to normalize the rest of the spectrum. This is useful for
removing a refractive index front that may occur on injection.
If the Realtime baseline correction check box is selected, then type values in the
Monitor baseline and Bandwidth fields. These fields together determine the spectral
window used to normalize the rest of the spectrum. Using typical values of 360 nm
for the monitor baseline and 20 nm for the bandwidth would have the effect of
dividing all other wavelengths in the spectrum by the average normalized intensity
between 350 to 370 nm. This normalization occurs at each time point in the run. The
spectral window defined by the monitor baseline and bandwidth must fit within the
specified range of the detector (200 to 380 nm).
Perform Analysis
1. In the Acquisition window, click Run Manager.
2. In the Run Manager dialog, click File > Peak Viewer.
3. In the Peak Viewer dialog, click open a data file.
4. In the Peak Viewer dialog, click Peaks > Analysis Settings.
The Analysis Method dialog (Figure 7-1) is divided into three tabs: Peak Detection,
Quantitation, and Options.
5. In the Analysis Method Settings dialog, open a method.
Insert a Table
Do one of the following:
• Click the arrow to the right of the calibration table to open the Insert New Table Entry
dialog (Figure 7-2).
• On the Analysis Method Settings dialog, click Advanced > Insert New Table
Entry.
Quantitation Tab
The Quantitation tab shown in Figure 7-3 consists of two parts: the calibration table and the
calibration curve window. The calibration table contains all the necessary information to
determine the concentration of unknown chemical compounds as well as sample specific
information.
Options Tab
The Options tab shown in Figure 7-4 contains the settings for peak identification and quantitation.
Refer to Chapter 9. This option is normally only used for UV systems.
Peak Results
Analyzing a chromatogram leads to a set of results for each identified peak. The results include
qualitative and quantitative properties that characterize the peak location and shape, the column
performance, and the amount of substance that created the peak. Peak results are shown in the
peak table in the Acquisition window. For more information, refer to Chapter 2.
The table columns can be customized using the Choose Peak Table Columns options on the
Analysis menu in the Acquisition window. The items that can be selected are shown in Table 7-1.
Table 7-1 Choose Peak Columns
Peak type Description
Peak Name The name of the peak as typed in the peak table.
Peak Number The peak number is assigned in ascending order according to the
retention time.
Retention Time (min) Position of the maximum peak absorbance, determined from a
quadratic curve fitted to peak data points around the absolute
maximum.
Peak Height (mAu) The maximum absorbance of a peak, determined from baseline-
corrected absorbance values.
Percent of Total Height Percentage from the sum of all peak heights.
(%)
*The available calculation methods are United States Pharmacopoeia (USP), German Pharmaco-
poeia (DAB), Japanese Pharmacopoeia (JP), and the method based on the Exponential Modified
Gaussian (EMG) model.
Capacity Factor
The capacity factor of a peak with retention time is calculated according to the formula in
Figure 7-5 where item 1 is the retention time of an unretained or reference peak.
Peak Asymmetry
Asymmetric peak shapes are characterized by different values for the front half width f and back,
or tail, half width b. As a quantitative measure of the peak asymmetry, the analysis module
provides values for the Asymmetry A
A=b/f
at 5, 10, and 50 percent of the peak height, and the Tailing Factor T which is calculated at 5
percent of the peak height:
T= (b+f)/2f
Figure 7-7 Asymmetric Peak with front width (f) and back width (b) indicated at 5, 10, and
50% of the peak height
Calculation Methods
The peak results include column performance parameters based on four different calculation
methods.
Figure 7-8 Tangent Method: The Base Width is Determined by the Intersection of the two
Tangents at the Inflection Points with the Peak Baseline
Resolution (R) between a peak of interest (peak 2) and the preceding peak (peak 1):
Where:
t1,2 = Retention time of peak 1 and peak 2
W1,2 = Base width of peak 1 and peak 2 determined by tangent method
Where:
t1,2 = Retention time of peak 1 (2)
W1(2), 50 = Width of peak 1(2) at 50% peak height
Where:
t1 ,2= Retention time of peak 1 and peak 2
W1 ,2 = Width of peak 1 and peak 2 at 50% peak height
Where:
t = Retention time of peak
W10% = Width at 10% peak height
f10% = Front width of peak at 10% peak height
b10% = Back width of peak at 10% peak height
Resolution (R) between a peak of interest (peak 2) and the preceding peak (peak 1):
With:
t1,2 = Retention time of peak 1 and 2
W1 ,2= Width of peak 1 and peak 2 at 10% peak height
Peak Detection
Detection Threshold
The Detection Threshold is applied to identify and locate peaks in a chromatogram. A non-zero
value is required by the software for the duration of the chromatogram. The threshold value is
specified as a percentage with respect to the highest value of the first derivative computed for the
entire chromatogram. A high value suppresses the detection of very slowly rising features and
small peaks, while a low value increases the number of small peaks that are reported. The
default threshold value is 1%.
Table 8-1 Event Parameters
Event Description
parameters
Start Time Start using the specified value for the Detection Threshold.
Stop Time Stop using the specified value for the Detection Threshold.
Parameter Detection threshold.
Shoulder Detection
Shoulders occur when two peaks are so close that no valley is formed in-between (for example,
Figure 8-1). The Shoulder Detection event enables the detection of shoulders attached to larger
peaks in a chromatogram. The shoulder detection algorithm is based on the second derivative.
The parameter indicates the sensitivity of the algorithm and is specified as a percentage with
respect to the highest value of the second derivative calculated for the entire chromatogram. A
higher value decreases shoulder sensitivity, while lower values increase the sensitivity to
shoulder peaks. A zero value indicates that shoulder detection is turned off. Shoulder detection
changes peak boundaries and the integrated areas of parent peaks.
Table 8-3 Event Parameters
Event Description
parameters
Start Time Start using Shoulder Detection.
Stop Time Stop using Shoulder Detection.
Parameter Shoulder Detection Sensitivity.
Minimum Area
The Minimum Area event sets a lower limit for the area that defines a peak. Features in a
chromatogram that lead to areas below this limit are not reported as peaks. A positive value is
required by the software. The area is specified in units of mAU. The default value is zero.
Table 8-6 Event Parameters
Event Description
parameters
Start Time Start using the specified value for the Minimum Peak Area.
Stop Time Stop using the specified value for the Minimum Peak Area.
Parameter Minimum Peak Area.
Integration Off
The Integration Off event disables peak detection and peak integration during the specified time
interval.
Table 8-7 Event Parameters
Event Description
parameters
Start Time Start disabling peak detection and integration.
Stop Time Stop disabling peak detection and integration.
Negative Peaks
The Negative Peaks event can be applied to integrate portions of the chromatogram that are
below the baseline (for example, Figure 8-3). As a consequence, features that appear as deep
valleys are reported as true peaks.
Table 8-8 Event Parameters
Event Description
parameters
Start Time Start detection of negative peaks.
Stop Time Stop detection of negative peaks.
Figure 8-4 Baseline with negative peaks enabled between 5.5 and 8 minutes
Peak Integration
Valley To Valley
The Valley To Valley event is applied to groups of fused peaks. By default, a common baseline is
drawn for all fused peaks, with each peak separated by a vertical drop line. During a Valley To
Valley event the baseline is constructed by connecting the valleys between individual peaks.
Table 8-9 Event Parameters
Event Description
parameters
Start Time Start using Valley To Valley.
Stop Time Stop using Valley To Valley.
Front Skim
The Front Skim event is applied to resolve one or more small child peaks or shoulder peaks
attached to the front of a larger parent peak. To construct the baseline of the child peak, the
following options are available:
• Tangent skim leads to a straight baseline for the child peaks.
• Gaussian skim leads to a Gaussian curve for the baseline of the child peaks.
• Exponential skim leads to an exponential curve for the baseline of the child peaks.
Table 8-10 Event Parameters
Event Description
parameters
Start Time Start of the first child peak.
Stop Time Stop of the parent peak.
Parameter Tangent, Gaussian, or Exponential.
Figure 8-7 Default integration of fused peaks using vertical drop lines
Back Skim
The Back Skim event is applied to resolve one or more small child peaks or shoulder peaks
attached to the back of a larger parent peak. To construct the baseline of the child peak the
following options are available:
• Tangent skim leads to a straight baseline for the child peaks.
• Gaussian skim leads to a Gaussian curve for the baseline of the child peaks.
• Exponential skim leads to an exponential curve for the baseline of the child peaks.
Table 8-11 Event Parameters
Event Description
parameters
Start Time Start of the parent peak.
Stop Time Stop of the last child peak.
Parameter Tangent, Gaussian, or Exponential.
Manual Baseline
The Manual Baseline event is applied to change a baseline of a peak without changing the
integration parameters. By default a baseline is constructed based on the peak boundaries as
they arise for a single isolated peak or in a group of fused peaks. A manual baseline is drawn
without changing the baseline of other peaks in the chromatogram. It is constructed by
connecting the data point at the specified start time with the data point at the specified stop time.
If the specified start time of the manual baseline is earlier than the start of the original baseline,
the original start time is maintained. Similarly, if the specified stop time of the manual baseline is
later than the original stop of the baseline, the original stop time is maintained.
Table 8-12 Event Parameters
Event Description
parameters
Start Time Start of Manual Baseline.
Stop Time Stop of Manual Baseline.
Figure 8-14 Forward Horizontal Baseline event between 4.5 and 7.2 minutes
Figure 8-16 Baseline after backward Horizontal Baseline Event between 4.3 and
5.75 minutes
Figure 8-18 Baseline after Lowest Point Horizontal Baseline event between 3.9 and
9.5 minutes
Reset Baseline
The Reset Baseline event resets the baseline at the data point that corresponds to the specified
start time of the event.
Table 8-16 Event Parameters
Event Description
parameters
Start Time Start of Reset Baseline.
Figure 8-24 Baseline after Force Peak Start event at 8.75 minutes
Figure 8-25 Baseline for peak in Figure 8-23 after Force Peak Stop event at 9 minutes
Split Peak
The Split Peak event divides a peak by inserting a perpendicular drop-line at the specified start
time of the event. The feature to the left and right of the drop line are reported as two separated
peaks. If the start time is not within the peak boundaries of a detected peak, the event is ignored.
Table 8-20 Event Parameters
Event Description
parameters
Start Time Indicate when to split a peak.
Merge Peak
The Merge Peak event reports the series of peaks between the start and stop of the event as one
peak. The retention time of the merged peaks is determined by the highest peak in the series. If
only one peak is detected between the start and stop time the event is ignored.
Table 8-21 Event Parameters
Event Description
parameters
Start Time Start of Merge Peak.
Stop Time Stop of Merge Peak.
Force Result
The Force Result event reports results for a peak in the specified time interval even if no peak
was detected. The event is useful for monitoring peaks that are expected to emerge within a
certain time interval during a series of runs.
Table 8-22 Event Parameters
Event Description
Parameters
Start Time Start of Force Result.
Stop Time Stop of Force Result.
Manual Peak
The Manual Peak event defines a feature in a chromatogram as a peak that was previously not
detected. The event is used to force integration without changing other integration parameters for
a specific region of the chromatogram.
Table 8-23 Event Parameters
Event Description
parameters
Start Time Start of Manual Peak.
Reference Peak
The Reference Peak event defines the peak which is used to calculate the capacity factor and
relative retention. If no peak is detected in the time window defined by the start and stop of the
event, the Capacity Factor and the Relative Retention is set to zero. If more than one peak is
detected the retention time of the first peak is used in the calculation. Only one Reference Peak
event can be typed into the integration timed event table.
Table 8-24 Event Parameters
Event Description
parameters
Start Time Start of the reference peak time window.
Stop Time Stop of the reference peak time window.
Incompatible Events
The analysis module restricts the simultaneous use of certain timed integration event
combinations. It is not possible to type an event in the integration timed event table if conflicts
arise with existing events. Conflicts are determined by checking for an overlap between two
events or, in the case of an event with only a start or stop time, if it is contained within a time
interval defined by another event.
Note: If an event is listed as conflicting with itself it cannot be applied a second time
within the range for which it already exists.
Note: If an event that is listed as not conflicting with itself is typed multiple times, the
parameter (for example, Detection Threshold) that was typed last is applied during the
overlapping time intervals.
Peak Identification
Peak Identification is an integral part of quantitation. A peak cannot be quantified unless it is
identified. The identification must have the following information in the calibration table:
Name and Retention Time: The user must supply a name for each retention time in the
calibration table for identification.
Retention Time Window: The retention time window accounts for the variability of retention
times by providing an upper and lower bound for the search. Two types are available to the user.
• Absolute: The upper and lower bounds are user defined.
• Relative: The window is calculated as a percentage.
Peak Identification Rules: More than one peak can be located within the defined retention time
window. The identified peak is chosen from one of five possible rules.
Rule 1—First Peak: The First Peak matching the criteria is found. Refer to Figure 9-1.
Figure 9-4 Third peak is found (the last peak in the window)
Rule 5—Best Match: Finds the peak with the most similar area relative to the indicated
specification in the calibration table.
Figure 9-5 Middle peak is chosen because its area is closest to 1100 mA2 (the peak area
in the calibration table was 1100 mA2)
Calibration
Calibration is based on the principle that a relationship exists between the amount of a chemical
compound and its measured detector response. The relationship (calibration curve) can be linear
or nonlinear and the calibration curves are stored in the calibration table.
Multiplication Factor: A correction factor that is used to convert units (for example, conversion
factor for grams to kilograms is 1000)
Dilution Factor: A correction factor that accounts for serial dilution
Types of Calibration
A calibration curve is characterized by the number of levels (for example, different amounts of
analyte in each sample). Each level can have of one or more replicates. The calibration curve is
generated from averaging the replicate amounts and response for each level.
• Single Level Calibration: A calibration that uses a single point. It assumes the
response is linear throughout the entire concentration range and the curve passes
through the origin. For an example of a single level calibration, refer to Figure 9-6.
Types of Fits
Linear Fit
A linear response follows the Beer-Lambert Law. The response of a chemical compound is
directly proportional to the amount of that compound. The following equation is used:
aX + b = Y
Where:
X = amount (ESTD), amount ratio (ISTD)
Y = response (ESTD), response ratio (ISTD)
a = slope
b = Y-intercept
Nonlinear Fits
If the response is nonlinear, the Quantitation module provides two polynomial fits: quadratic and
cubic.
Quadratic Fit
Quadratic: Minimum of three calibration levels
Y = aX2 + bX + c
Where:
X = amount (ESTD), amount ratio (ISTD)
Y = response (ESTD), response ratio (ISTD)
a = coefficient 1
b = coefficient 2
c = Y-intercept
Cubic Fit
Cubic: Minimum of four calibration levels
Y = aX3 + bX2 + cX + d
Where:
X = amount (ESTD), amount ratio (ISTD)
Y = response (ESTD), response ratio (ISTD)
a = coefficient 1
b = coefficient 2
c = coefficient 3
d = Y-intercept
Calibration Range
For quantitation with nonlinear fits, the Quantitation routine module will provide concentrations for
unknowns that are bound by the lower/upper limits of the amounts for calibration curves.
Note: Ideally, the equation generated should include the origin (for example, the Y-
intercept = 0, detector has zero response when the compound is not present). The user
has the option to force the fit through zero.
Correlation Coefficient
The correlation coefficient measures how well a polynomial fit matches the data. As the
coefficient approaches one, the curve is a better representation of the data.
0 ==> No fit
1 ==> Perfect fit
Weighting Factor
A weighting can be applied to each point in a calibration curve. The choice of weighting factor
determines the importance of a data point within a given data range. For example, a weight of 1/
amount places more emphasis on points with a smaller amount.
Scaling Factors
The quantitation module includes the ability to scale the X-axis of the calibration curves. For a list
of available scale factors, refer to Table 9-2.
Table 9-2 Scale Factors
Scale Factor
None
1/Amount
ln(Amount)
1/ln(Amount)
Sqrt(Amount)
Amount²
1/ResponseFactor
1/ResponseFactor²
ln(ResponseFactor)
1/ln(ResponseFactor)
Sqrt(ResponseFactor)
ResponseFactor²
Log(Amount)
1/Log(Amount)
Log(ResponseFactor)
1/Log(ResponseFactor)
Calibration Table
The calibration table contains all of the necessary information to perform Quantitation. The
following information is required to create the table:
• The retention time for each named peak.
• Amounts for each peak.
• Response for each peak.
Each row of the table corresponds to a given compound. For a description of each column, refer
to Table 9-3.
Note: The user can bypass calibration and perform peak identification by supplying the
retention time for a named peak.
the same conditions. Because the response is sensitive to changes in concentration, care must
be taken to make sure that the injection volume cannot vary.
The amount is calculated by solving the equation from its calibration curve using the response of
the unknown sample. Figure 9-8 shows a single-level calibration using the height as the
response.
Where:
a = the slope with units of Response/Amount.
The example equation is
Where:
a = the ratio:
b = the Y– intercept.
Rearranging the equation leads to:
User Authentication
By default, the user authentication features are not available when the software is installed. This
is an open system configuration, where every user who has access to the computer has full
access to the software. After user authentication has been enabled, all users are required to log
on to access the software.
The software automatically queries both the domain controller and the local computer for a list of
available users. These user accounts are defined on the primary domain controller or on the local
computer, but the privileges that these users have are defined in the software. Any Windows user
can access the software using the same accounts, assuming the right privileges have been
granted to the user.
If a valid system administrator user name and password combination is supplied for
the domain or the computer, the User Authentication dialog opens. Refer to
Figure 10-2.
Figure 10-3 User does not have software start and exit rights
User Manager
The specific system functions that are controlled by the User Manager dialog include:
• Adding and removing Windows users and groups.
• Granting system privileges to Windows users and groups.
• Defining electronic signature roles.
• Defining electronic signature reasons.
The action that occurs is dependent on whether User Authentication has been enabled and the
assigned user privileges.
• If User Authentication is enabled, and the user is logged on as a system
administrator or a user with User Administration privileges on software startup, the
user can access the User Manager dialog.
Tip! Click Cancel at any time in the Add Members dialog to exit without
making any changes
5. To add a member, select the member to be added in the Available Members region
and then click the right arrow. This member name appears in the Added Members
region.
6. To remove a member, select the member to remove in the Added Members region
and then click the left arrow. This member name appears in the Available Members
region. When you are satisfied with the membership arrangement, click OK to
confirm the changes and return to the User Manager dialog. The members shown
are the same as those that specified in the Added Members list.
Tip! Click Cancel at any time in the Electronic Signature Roles dialog to
exit without making any changes
3. To change the name of any of the roles, type a new role name in the text box for that
role.
4. To change the number of active roles, select a number between 1 and 5 in the
Number of Levels list. When reducing the number of roles, the highest signature
roles are deactivated first.
For example, choosing 3 makes the top two levels of electronic signature roles
unavailable, indicating that these two roles are not active, and only the bottom three
roles can be assigned to a user. If any users had previously been assigned to
Operations Manager or QA Manager, their role now defaults to Lab Manager, the
highest role available at this point.
5. Click OK.
General Tab
By default, all members automatically have permission to the Start/Exit Software feature when
they are added to the system. All other permissions on the General tab are not automatically
assigned. To give a member a particular permission on this tab, select the check box.
• Start/Exit Software: Gives a user the right to log in to the software when it is
opened. A user can also exit the software in the main program under File > Exit. A
user cannot log in to the software without this right.
• UV Detector Toolbox: Allows a user to open the UV Detector Toolbox from the
menu option View > UV Detector Toolbox in the Acquisition window. This menu
option is not available if a user does not have this right. For more information, refer to
UV Detector Toolbox on page 21.
• Peak Park Toolbox: Allows a user to open the PeakPark Toolbox from the menu
option View > PeakPark Toolbox in the Acquisition window. This menu option is not
available if a user does not have this right. For more information, refer to PeakPark
ToolBox on page 23.
• System Logs: Allows a user to view any of the available system logs from the menu
option View > System Logs in the Acquisition window. This menu option is not
available if a user does not have this right. For more information, refer to System
Logs on page 24.
• Analysis Settings: Allows a user to access the Analysis Settings dialog from the
menu option Analysis > Settings in the Acquisition window and Analysis Methods
in the Run Manager dialog. These controls are not available if a user does not have
this right. For more information, refer to Chapter 7.
• Notification Options: Allows a user to access the Notification Options dialog from
the menu option System > Notification Options in the main program. This menu
option is not available if a user does not have this right. For more information, refer to
Notification Options on page 33.
• Appearance: Allows a user to access the Appearance Settings dialog from the
menu option System > Appearance Settings in the Acquisition window. This
menu option is not available if a user does not have this right. For more information,
refer to Appearance Settings on page 33.
• Mobile Phases: Allows a user to access the Mobile Phases dialog from the menu
option System > Mobile Phases in the Acquisition window. This menu option is not
available if a user dos not have this right. For more information, refer to Mobile
Phases on page 36.
• Direct Control: Allows a user to access the Direct Control dialog from the menu
option System > Direct Control in the Acquisition window. This menu option is not
available if a user does not have this right. For more information, refer to Direct
Control on page 38.
• Check for Updates: Allows a user to check for software updates from the Eksigent
web site from the menu option Help > Check for Updates in the Acquisition
window. This menu option is not available if a user does not have this right.
To select or clear all permissions on this tab, click Check All or Uncheck All.
Runmanager Tab
This tab contains a slider on the left side that determines the user access level for the Run
Manager dialog. The access levels which range from least access to most access are: None,
Low, Medium, High, and Full. By default, all members have access level None in the Run
Manager dialog when they are added to the system.
None: Users cannot open the Run Manager dialog. The Run Manager button in the main
program is not available.
Low: Users can open the Run Manager dialog by clicking the Run Manager button in the main
program. The user can also exit the Run Manager dialog. In the File menu, the user can open,
export, and print pre-defined Run Tables.
Data Tab
The Data tab of the Member Permissions dialog (Figure 10-10) manages permissions to access
data in the main program and the application of electronic signatures. This tab also uses a slider
to determine the user level for data access. The access levels, from least access to most access,
are: None, Low, Medium Low, Medium, Medium High, High, and Full. By default, all members
have data access level None when they are added to the system.
Table 10-2 outlines the privileges granted to each data access level.
Table 10-2 Run Manager Permissions for Each Access Level
Permission None Low Medium Medium Medium High Full
Low High
Open Data
Toggle Auto-Print
Save Data
Toggle Auto-Save
Apply E-Signatures
Revoke E-
Signatures
All controls for data access are found in the File menu in the Acquisition window. The electronic
signatures are applied in the Electronic Signatures dialog.
• Open Data: Enables the Open command in the File menu in the Acquisition dialog.
• Toggle Auto-Print: Enables the Auto-Print command in the File menu in the main
Acquisition window.
• Save Data: Enables the items Save As and Output Options in the File menu of the
Acquisition window.
• Toggle Auto-Save: Enables the Autosave command in the File menu of the
Acquisition dialog.
• Apply E-Signatures: Allows a user to apply electronic signatures in the Electronic
Signatures dialog.
• Revoke E-Signatures: Allows a user to revoke electronic signatures in the
Electronic Signatures dialog.
The Electronic Signature Role of the user can also be set in the Data tab. The E-Signature Role
menu is only enabled when the user has a Data level of Apply E-Signatures or Revoke E-
Signatures. By default, all users have the lowest level of E-Signature role when they are added to
the system. Select a role appropriate for the user by using the menu. For more information, refer
to Chapter 11.
Diagnostics Tab
The Diagnostics tab of the Member Permissions dialog (Figure 10-11) is used to manage
permissions for the Hardware Diagnostics dialog. For more information, refer to Hardware
Diagnostics on page 28. Permissions are assigned by selecting the appropriate check box. By
default, all members have no permissions in the Hardware Diagnostics dialog when they are
added to the system.
Other Tab
The Other tab of the Member Permissions dialog (Figure 10-12) contains the remaining
permissions. Permissions are assigned by selecting the appropriate check box. By default, no
permissions on the Other tab are granted when members are added to the system.
6. To view a signature in detail, select the signature in the Signatures region and then
click View. The signature is shown in a separate Signature dialog (Figure 11-2). This
is useful if a signature contains long comments.
Explorer
When the folders are expanded, only data files and other folders, if any, are shown. If any data
files exist in a particular folder, the files are shown in the Data Files region. A data file with a blank
notepad icon indicates that the file contains no signatures, while a data file with a pencil-notepad
icon indicates that the file contains at least one signature.
Data Files
The Data Files region contains three columns: File Name, Creation Date, and Last Modified. The
File Name column contains the name of the data file, the Creation Date column contains the date
that the data file was created, and the Last Modified column contains the date that the data file
was last modified. Data files can be sorted in the Data Files region by any of these three
parameters.
Signatures
The Signatures region contains six columns that are the parts of an electronic signature saved to
file.
• Signer is the Windows user name of the user signing the document.
• Date is the date when the document was signed.
• Role is the signature role of the user signing the document.
• Reason is the reason for the signature.
• Revoke Date is the date a signature was revoked.
• Comments where optional comments can be typed.
For signatures that are not revoked, the revoke date field indicates active instead of a date. The
signatures can be sorted in the Signatures region by any of these parameters. To do so, click on
the column heading of the parameter.
If the user’s credentials have been established, then the user does not need to log
on. If the user’s credentials have not been established, then the logon prompt
(Figure 11-1) opens.
4. Type a valid user account that is either a system administrator or a user with the
Revoke E-Signatures right. If the user account is neither of these, the user cannot
revoke a signature.
When revoking signatures, remember that once a signature is revoked, it cannot be
activated. Signatures can only be revoked one at a time. There are no controls to
revoke multiple signatures simultaneously. Also users can only revoke their own
signatures. No user can revoke any other user signatures.
values, location 7
Calibration Disk button 12
A calibration functions 28
A/D channel, Appearance Settings 35 calibration table
Acquisition Window Quantitation tab 83
components 14 requirements for creating 124
overview 13 Calibration Values tab 29
acquisition, progress of 45 capacity factor, defined 86
Add Integration Event 16 Channel 1 log 27
adding users and groups 131 charts, generating 58
Advanced tab, Acquisition Window 32 Check For Leaks/Blockage 28
Alerts log 25 chromatograms
Amount Ratio 118 analyzing 40
Analysis methods defined 42 integration timed events 91
analysis module, overview 81 LC Method Editor 80
Analysis Settings Editor 54 overlaying 68
analyze peaks 53 Column Flow Rate Reference Method 24
ANDI files 19 column heaters 75
Appendable Peak Table Format 19 column pressure 14, 24, 33, 34
attributes, data 33 commands
Auto-Complete Command, Run Manager 48 Auto-Complete, Run Manager 48
Automatic background, turning on or off 22 configuring
automatic peak detection algorithm, described drivers 11
81 HPLC system 10
automatically purge 37 Plot window 53
autosampler methods system 10
defined 42 contour plot region 15
Run Table 44 Copy Plot Selection 62
Autosave correlation coefficient, described 122
File menu 17 creating
format 19 charts 58
using 17 LC methods 73
cubic spline fit, applying 24
B current mobile phase flow rates 14
Current Tray 41
Back Skim event 98
custom formula 58
Backward Horizontal Baseline 100
customize
C peak tables 16
calculation methods 86 D
calibration
data acquisition, LC Method Editor 79
curves, storage of 118
data files
curves, types 119
locating 141
described 118
revoking signatures 144
range 122
sample reports 56
terms 118
Integration Off 94
integration timed events, chromatograms 91 N
internal flowmeter pressures 14 Negative Peaks 94
Internal Standard Calculation, described 126 nonlinear fits 121
notification events 33
J notification frequency 33
Japanese Pharmacopoeia (JP) Calculation Notification Options 27
Method 89
O
L Options tab, peak identification 84
lamp, direct control 39 Other tab 140
LC Channel 45 Output D/A Range 32
LC methods Output Folder field 18
creating 73 output options 17
data acquisition 79 overlaid data, removing 60
defined 42 overlaying, chromatograms 68
editing 73 Ownership Tags 19
Mode 80
Run Manager 42 P
LC Wait for Contact Closure 43 Pause button 15
linear fit 121 PDF
log files 24 printing 20
Lowest Point Horizontal Baseline 101 report 20
peak asymmetry 87
M peak detection events, defined 91
Main Plot Region, modifying the plot scale 16 Peak Detection tab, overview 82
Make All Plots Identical 66 peak detection, disabling 94
Managing Electronic Signature Roles 132 peak identification
Managing User Permissions 134 described 113
Manual Baseline 98 Options tab 84
Manual Peak 106 peak integration events, defined 91
menus peak integration parameters, setting 40
File 42, 61 peak integration, disabling 94
Peak Viewer 61 peak label options 67
Run Manager 42 Peak Park Toolbox, overview 23
View menu 21, 62 Peak Preview
Merge Peak 105 File menu 61
Minimum Area 94 function 19
Minimum Peak Height 93 overview 19
Minimum Peak Width 93 Peak Preview function 19
mobile phases 37, 76 peak results
Mode, LC Method Editor 80 calculation methods 87
Monitor Folder 61 peak shapes, defined 87
monitoring peak trends 58 peak table
moving multiple cells 47 annotations 55
Multi-Level Calibration 120 customizing 16
multiple plots, showing 59 overview 53
Multiplication Factor 119 peak trends
generating 58