Research Journal of Pharmaceutical, Biological and Chemical Sciences
Research Journal of Pharmaceutical, Biological and Chemical Sciences
1
Department of Natural Resources, Isfahan University of Technology, Isfahan, 84156-83111 Iran
2
Biotechnology Research Center, School of Pharmacy, Mashhad University of Medical Sciences, Mashhad, Iran.
ABSTRACT
*Corresponding author
INTRODUCTION
Discharge of municipal and industrial wastewaters as well as agricultural and industrial drainage
containing organic materials in the water resource cause the reduction of dissolved oxygen (DO) [1, 2].
Biochemical oxygen demand (BOD) is an empirical parameter that refers to the amount of required oxygen for
microorganisms to oxidize organic materials in wastewater, effluent, and polluted water samples [3]. Dilution
and manometric methods are the two most conventional methods for measuring the BOD. In these methods,
BOD is calculated through the changes in the amount of DO in water samples or by pressure drop, which is the
result of oxygen consumption in a bottle containing the sample [4].
◦
In the dilution method, samples are incubated n-days (5 days as BOD5 in the standard test) at 20 C in
a dark bottle and the initial and final amount of oxygen is determined using an amperometric sensor or
iodometric titration [5, 6]. The BOD is calculated by the following equation:
Where C1 is the initial oxygen concentration of sample, C2 is the oxygen concentration after n-day of
incubation, C3 is the initial oxygen concentration of blank, C4 is the blank oxygen concentration after n-day
incubation, Ve is the initial sample volume, and Vt is the final sample volume after dilution.
Dilution method is a standard test method for BOD. However, it has some disadvantages such as
necessity for aeration of some samples (for example, the samples containing hydrogen peroxide). Dilution is a
critical step in this method and the accurate estimation of dilution order is very important. Moreover, DO
sensors always show some uncertainties in the samples with high electrical conductivity.
In the manometric method, the pressure drop due to oxygen consumption in a bottle containing
sample is measured by a manometer. The produced carbon dioxide is absorbed into a potassium hydroxide
solution. BOD is calculated through following equation:
In spite of having a lot of advantages, manometric measurements need some requirements that must
be met. One of the disadvantages of this method is the impact of temperature on the gas pressure. A little
change in temperature can lead to a change in pressure difference and consequently, the BOD amount. On the
other hand, for instruments based on measuring pressure reduction, to equalize temperature of sample and
incubator, reading of initial pressure is performed 20 minutes after inserting the sample bottle into the
incubator. This could be an origin of error in measurements. Also, entering a little amount of air into the bottle
sample during incubation period causes all measurements not to be valid.
Generally, the nitrification processes could consume dissolved oxygen and cause the interferences in
the BOD measurement [7, 8]. Moreover, the BOD is affected by oxidation of inorganic materials such as
sulfides and ferrous iron [5, 9, 10]. Because of the importance of BOD measurement in the environmental
monitoring, many attempts have been made to develop new rapid methods for determination of BOD [6, 11,
12]. However, most of these methods cannot be regarded as standard methods and suffer from some
limitations [13, 14]. In order to eliminate the noted interferences, a method basically identical to that of
manometric one is presented, in which the BOD is determined by indirect measuring of carbon dioxide.
Turbidity produced by converting carbon dioxide to carbonate in the presence of barium chloride is
determined and related to the BOD amount. One of the most important advantages of this method, unlike the
two mentioned standard methods, is that the oxygen consumed in nitrification and other oxidant process such
as mineral oxidant bacteria has no quota in calculations. This is because the products of inorganic biooxidation
do not enter to the absorber solution and consequently not any precipitation with barium ion. The method is
performed only with a turbidity meter or a spectrometer that can be found in most laboratories and does not
need DO sensors or expensive sensitive manometers.
Produced carbonate amount is in proportion to the amount of consumed oxygen for oxidizing
biodegradable organic materials.
All chemicals used in the experiments were in the analytical reagent grade and were purchased from
Merck Co. The solutions were prepared in distilled water. The biochemical oxygen demand measurements
were performed by a commercial BOD meter (model Aqualytic Al 606). Turbidity determination was performed
by a commercial WTW turbidity meter (model Turb 550). A commercial Metrohm pH meter (model 827) was
used for all pH measurements.
Phosphate buffer solution: 21.75 g K2HPO4 , 8.5 g KH2PO4, 33.4 g Na2HPO4 , and 1.7 g NH4Cl was dissolved in
water and then diluted to a final volume of 1000 mL.
Calcium chloride solution: 27.5 g CaCl2 was dissolved in water and then diluted to a final volume of 1000 mL.
Magnesium sulfate solution: 22.5 g MgSO4.7H2O was dissolved in water and then diluted to a final volume of
1000 mL.
Iron chloride solution: 0.25 g FeCl3.6H2O was dissolved in water and then diluted to a final volume of 1000 mL.
Potassium hydroxide (absorber): 40 g potassium hydroxide rinsed by deionized water to remove absorbed
carbon dioxide and diluted to a final volume of 100 ml after dissolution and stored in a closed bottle away
from air contact.
Experimental procedure
1000 ml of water sample was placed in a 2 L Becker, and then 1ml of each nutrient solution was
added to the Becker. The samples pH was adjusted at 6.5 to 7.5. Then 224 ml of the prepared sample was
placed in an opaque glass bottle containing a magnet. Five droplets of a 40% KOH solution were placed into
the container mounted on the cap to absorb carbon dioxide produced in the period of incubation and the
◦
bottle cap was tightly fastened so that no air could diffuse into it. The bottle was incubated at 20 C for 5 days
in an incubator while the sample was being agitated on a stirrer to accelerate carbon dioxide and oxygen
transfer (Figure 1).
After five days, potassium hydroxide solution was moved to a 25 ml balloon and diluted to volume.
The carbonate content of absorption solution was determined by turbidity meter using barium chloride as a
suspension reagent based on calibration curve.
Calculations
BOD value is obtained from the concentration of carbonate according to the following equation:
Where C carbonate is carbonate concentration (mg/L) in balloon after absorber dilution which
obtained from calibration curve, Vballoon is balloon volume used to absorber dilution, Vsample is incubated
sample volume (ml).
To obtain the optimum pH range of the method, the synthetic seeded solutions of glucose with four
different pH levels were prepared and the BOD5 was analyzed by manometric and turbidimetric methods
(Table. 1).
Table 1: Optimization of pH for BOD5 (mg O2/L) measurement
Manometric
54 48 47
(Mean BOD5)
Carbonate
turbidimetric 56, 47, 42, 61, 53 (14*) 39, 53, 40, 55, 51 (16*) 43, 50, 41, 52, 45 (10*)
(BOD5)
Carbonate
turbidimetric 51.8 47.6 46.2
(Mean BOD5)
* RSD%
As shown in Table 1, the optimum pH for the measurements is between 6.5 to 7.5 and in this range
the mean value of BOD, obtained by carbonate turbidimetric method, is very close to that of manometric.
However, at pH=8.0, a negative significant deviation was observed. This could be due to the solubility of the
carbon dioxide in aqueous solution of sample at high pH, which results in the formation of soluble forms of
carbonate and bicarbonate ion. The relative standard deviations (%) of the Results of proposed method are
shown in Table 1. The RSD of this method is satisfactory and comparable to other methods. It has been
reported the RSD between the respirometric test, iodometric method, and oxygen sensor were 1–35% [15,
16]. These deviations can be explained due to the heterogeneity of the samples.
To investigate the linear response range of the method, glucose solutions were prepared in the range
of 1-1000 mg O2/L. To prevent the shortage of oxygen in the period of incubation, a lower volume of sample
having higher BOD was chosen. All sample bottles were 500 ml and measurements were performed in pH= 7.
Table 2: BOD5(mg O2/L) Results for dynamic range of proposed method compare to manometric standard method
As shown in Table 2, the mean value of BOD, obtained by carbonate turbidimetric method, is very
close to that of manometric in BOD lower than 50 mg O2/L. However, at BOD>50, a negative significant
deviation in relative error was observed. Therefore, using this method, a linear range was obtained between
generally in the range of 1–50 mg O2/L.
To assess the feasibility of the application of the proposed method for the analysis of environmental
samples, three different matrix wastewater samples were analyzed by both manometric and turbidimetric
methods. Nutrient solutions were added and the pH of water and waste water samples was adjusted by the
addition of diluted sulfuric acid or sodium hydroxide and buffer solutions before incubation. Table 3 shows
that the results of presented method are comparable with standard manometric method and it is suitable for
the determination of BOD in urban and industrial water and waste water samples.
One of the advantages of the turbidimetric method, as compared with the conventional manometric
and dilution methods, is that the nitrifying bacteria and oxidation of sulfide and iron have no influence on the
experimental results. The metabolites products from microorganisms’ activities include nitrate, sulfate and
some minerals which are soluble in the water, are not absorbed by potassium hydroxide solution at the top of
the sample bottle [8]. Therefore, contrary to the manometric or dilution method, no inhibitor reagents are
needed to add to the samples. Also, the results of BOD are not affected by the errors from sensitive apparatus
of DO meter and manometer. On the other hand, the required equipment for turbidimetric method is usually
found in every water and wastewater laboratory.
Table 3: BOD determination by both manometric and turbidimetric methods in real samples
CONCLUSION
In this study, we described a new, simple and inexpensive method for determination of BOD in
environmental samples. Optimum pH range for measurement, like other conventional methods, was between
6.5 -7.5. The samples with BOD in the range of 1-50 mg O2/L can be directly analyzed and those with higher
concentrations will be analyzed after an appropriate dilution. The results and RSD of presented method are
comparable with standard manometric method and it can be applied for the BOD analysis of real water
effluents such as urban and industrial wastewater samples.
ACKNOWLEDGEMENT
The authors would like to acknowledge Isfahan University of Technology Research Council for
providing facilities and laboratory supplements.
REFERENCES
[1] Luo P, He B, Takara K, Razafindrabe BH, Nover D, Yamashiki Y. J Environ monitoring 2011;13(10):2819-
2829.
[2] Dhall P, Siddiqi TO, Ahmad A, Kumar R, Kumar A. The Scientific World Journal 2012; 2012:105712.
[3] Dhall P, Kumar R, Kumar A. The Scientific World Journal 2012; 2012:861903.
[4] Kamal D, Khan AN, Rahman MA, Ahamed F. Pakistan J Biol Sci 2007; 10(5):710-717.
[5] Radojevic M, Bashkin VN. Practical Environmental Analysis. RSC publication, Cambridge, 1999.
[6] Hudson N, Baker A, Ward D, Reynolds DM, Brundson C, Marquet CC, Browning S. Sci total Environ
2008; 391:149-158.
[7] Hur J, Cho J. Sensors 2012; 12(1):972-986.
[8] Su JJ, Liu BY, Chang YC. Lett Appl Microbiol 2001; 33(6):440-444.
[9] Eye JD, Ritchie CC. Wat Pollut Control Federation 1966; 38(9):1430-1440.
[10] Jacome A, Molina J, Novoa R, Suarez J, Ferreiro S. Wat Sci Technol 2014; 69(1):217-221.
[11] Wang J, Zhang Y, Wang Y, Xu R, Sun Z, Jie Z. Biosensors Bioelectronics 2010; 25:1705-1709.
[12] Namour PH, N. JR, Namour PH, TrAC. Trends anal Chem 2010; 29(8):848-857.
[13] Galleguillos M, Vasel L. J Environ Technol 2011; 32(11-12):1259-1267.
[14] Liu C, Zhao H, Ma Z, An T, Liu C, Zhao L, Yong D, Jia J, Li X, Dong S. Environ Sci Technol
2014;48(3):1762-1768.
[15] Maximova N, Dahl O. Chem Soc Rev 2007; 36(8):1323-1349.
[16] Roppola K, Kuokkanen T, Rämö J, Prokkola H, Heiska E. J Autom Methods Manag Chem
2007;2007:39761-39765.