ALGO 3i Newborn Hearing Screener
A. Power-On slider switch
B. Display
C. Navigation buttons (keys)
D. Earphone (ATA) cable connection
E. Sensor (Preamp/PCA) cable
connection
F. Infrared (IR) port: printing/data
export
G. DC power port: battery charger
connection
H. SoftClip™: device stand/bassinet
clip
Selecting A Baby
Indications for use:
• Term, healthy newborns • Ready for discharge, or scheduled to go
• NICU graduates, 34 weeks gestational home within a few days
age and under 6 months of age • Sleeping or in a relaxed state (usually
• Newborns with normal outer ear after a recent feeding)
anatomy and no obvious deformities of • In an open crib
the head or neck
Exclusion criteria:
• Infants not between 34 weeks • Infants exhibiting cranio-facial head
gestational and 6 months of age abnormalities, including structural
• Infants on ventilators or in incubators abnormalities of the outer, middle, or
• Infants on CNS (central nervous inner ear
system) stimulants • Infants with known or suspected
• Infants receiving ototoxic medications neurologic conditions
• Infants with compromised skin or • Infants in an agitated state
jaundice
Toggle key: This large key enables up/down/left/right movements, and is primarily used to
scroll through the alphanumeric keypad to input data and passwords.
OK key: This green key controls the “YES”, “OK”, and “CONFIRM” functions, which
allow a command to be executed.
X key: This red key controls the “NO”, “STOP”, and “PAUSE” functions, which allow a
command to be stopped, screening to be paused, or allows the user to return directly to the
MainMenu.
? key: This key controls the Help system and will access the section of the ALGO 3i
screener’s help system corresponding to the area of the interface where the user is located.
The user can scroll through the Help system to access different information as desired.
Up/Down
Arrow Keys
Toggle Key
OK Key
X Key ? Key
The display will always indicate to the user which keys are used to select, scroll, advance to
the next window, or go back to the previous window.
Examples:
Whenever the ALGO 3i screener presents you with options, they will be framed by blue bars
showing an up and down arrow key. This indicates you may scroll through the options by
pressing either arrow key. Press the arrow key once to scroll to the next option. Press and
hold the arrow key to scroll rapidly through the available options.
Scroll until the black select bar highlights the desired option. The select bar will indicate what
to do to activate your option (in most cases, clicking on the OK key) and the Message Box,
located on the bottom of the window, will update to provide further information about your
choice.
A few options within the ALGO 3i screener are multi-choice options, meaning the user may
select more than one at a time.
Multi-choice options are indicated by a toggle key symbol . See illustration below.
In this case, the black select bar indicates use of the toggle keys to select or deselect the
desired option set. Use the OK key to save your option set.
Data is entered into the ALGO 3i screener by using the up/down toggle key to scroll through
a keypad containing the numbers 0–9, the Roman character alphabet A–Z, a blank space, ‘+’, ‘-’,
Note: Only if the alpha-numeric function is enabled) will the capitalized Roman character
alphabet sequence appear with the number sequence in the Infant Identifier and User
Identifier fields. (See Chapter VIII. Service Manual: Device Set-Up: User Preferences)
Blank Spaces
There is a blank space in the keypad between A and 9 (or between 9 and 0 if alphanumeric
entry is disabled). This allows users to insert blank spaces if needed in the data fields or to
remove characters.
Note: The infant identifier data field may not begin with a blank space.
The black select bar will indicate which data field is active, and a cursor symbol will
indicate the current position within the data field. To begin scrolling through the keypad,
click on the up or down toggle keys (press and hold key for rapid scrolling). When you have
located the correct number or letter, simply click on the right or left toggle keys to set it and
move one space right or left. To advance to a different data field, click one of the arrow keys.
Example: Enter infant identifier 6C, then move to the Last Name field
1. Enter the Infant Data window. The cursor will be set on the first space of the 12-digit
infant identifier data field.
2. Click the down toggle key to scroll through the numbers 9…8…7…6.
3. Click on the right toggle key to set the first space to the number 6 and advance to the
second space.
4. Click the up toggle key to scroll through the letters A…B…C.
Page 1 of
Infant Data
Window
5. Click the down arrow key to set the second space to the letter C and move to the Last
Name data field.
Page 2 of
Infant Data
Window
Step 1. Power on: Turn on the ALGO 3i screener by sliding the On/Off switch upward, and
releasing it.
Start-up/Warnings: The ALGO 3i screener will briefly display a Start-Up window
while it performs a self-check. Warnings may appear which may require your
acknowledgement. The MainMenu will appear next.
Step 2. Observe MainMenu: The MainMenu shows each of the major functions of the
ALGO 3i screener, and is indicated by this icon:
a. SpeedScreen, if enabled (SpeedScreen mode)
b. Enter Data/ScreenBaby (ScreenBaby mode)
c. DataManagement
d. Device Set-Up
a. There are three data fields in the first Infant Data Window. Infant Identifier, Last
Name, and Frist Name. There are another three data fields in the second window.
Date of Birth, User Identifier, and Comment.
Step 4. Prepare baby and attach supplies: Prepare the infant for screening by ensuring that
no exclusion criteria are present, and that the infant is in a quiet state. Collect
screening supplies: Natus single-use earphones (pair) and single-use sensor (triplet)
for the infant.
a. Attach sensor cable clips to sensors (clip onto the purple section of sensor). See
Figure 1 below.
Figure 1
Figure 2
or
Figure 3 Figure 4
e. Attach earphones to baby by rolling from behind the ear to the front. Hold
earphones in place for a few seconds to warm and activate earphone adhesive.
The earphone tab should be directed towards the back of the infant’s head. The
earphone cable may emerge either “up” or “down” from the earphone, see Figure
4 above.
The RED transducer is placed on the infant’s RIGHT ear. The BLUE transducer
is placed on the LEFT ear. See Figure 5 above.
Step 5. Select screening parameters: Select either 35 dBnHL or 40 dBnHL screening (if 40
dB option is enabled). Select Simultaneous (not available for 40 dB), Sequential,
Right-only, or Left-only screening. Use the up/down arrow keys to scroll and select
the parameters then click on the OK key to proceed.
Step 7. Monitor screening in progress: Monitor impedances and the Message window
during the screening. The ALGO 3i screener will alert you to any conditions which
require your attention during the screening process.The Screening in Progress
window displays the following information:
a. Infant data
b. Sweep Counts
c. Elapsed screening time
d. Current Date and Time of screening
e. Impedance monitor
f. Screening Progress bar
g. Screening Result
h. Ambient and Myogenic noise bars
i. Message Box
j. “X” to Pause Screening
To cancel screening, click X again. The ALGO 3i screener will ask for confirmation
that you want to exit the screening. If a result for at least one ear has been obtained,
the ALGO 3i screener will save and confirm a screening record. If neither ear has
obtained a result, no screening record will be saved.
There are a few conditions under which the ALGO 3i screener will automatically
terminate the screening process:
• Cable(s) disconnected
• Battery charger connected
• Screening time limit exceeded
• Battery too low to complete screening
Step 8. Observe screening complete: At the completion of the screen, the ALGO 3i screener
will save a screening record to memory, and will display the infant data, including a
summary of the screening results.
Step 9. Remove/discard supplies: Remove from the baby and discard the Natus screening
supplies.
Step 10. Return to MainMenu/Disposition screening record: Return to the MainMenu or
go to DataManagement for printing, exporting, or deleting the screening record.
a. From the Screening Complete window, you may click X to return to the
MainMenu, then either power off the device or screen another infant.
b. From the Screening Complete window, you may instead click OK, which will
bring you to the DataManagement Window. You may immediately print or
export the screening record just saved. See the Chapter V: ALGO 3i Screening
Step 11. Power off: Turn off the device by sliding and holding the ON/OFF switch for 3
seconds. The ALGO 3i screener will count down from 5 to 1 before powering off.
Step 12. Clean Preamp/PCA and ATA cables before and/or after use: Clean patient
cables after use. Residual sensor adhesive may corrode Preamplifier/PCA cable clips.
Note: There is no time constraint for entering infant data once you are in the
Infant Data window. The screen may be completed while you are entering infant
information but the ALGO 3i device will wait until you click OK before matching
this infant information to the screening results and save it as a screening record.
e. If you do not wish to enter infant data for the screening record, proceed with
screening.
At the completion of the screening, the device will save the screening record and will
display the screening results. The infant data fields will be blank.