URIT-910C Engineer Training Procedure

The document outlines the operation, structure, installation, and maintenance procedures for the URIT-910C Plus Electrolyte Analyzer. It details the system's measurement parameters, hardware components, and step-by-step instructions for setup and calibration. Additionally, it provides guidelines for quality control, data management, and troubleshooting to ensure optimal performance of the analyzer.

Uploaded by

BaBuiDinh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
553 views85 pages

URIT-910C Engineer Training Procedure

The document outlines the operation, structure, installation, and maintenance procedures for the URIT-910C Plus Electrolyte Analyzer. It details the system's measurement parameters, hardware components, and step-by-step instructions for setup and calibration. Additionally, it provides guidelines for quality control, data management, and troubleshooting to ensure optimal performance of the analyzer.

Uploaded by

BaBuiDinh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 85

Engineer Training

Procedure
URIT-910C Plus
Electrolyte Analyzer
URIT-910C Plus Electrolyte Analyzer

1- Main Idea For Operation.


2- System Structure.
3- System Installation.
4- Operating System.
5- Hardware Structure.
1- Main Idea For Operation.
► Ion-selective Electrode (ISE), Also Known As A Specific Ion
Electrode (SIE), Is A Transducer (Or Sensor) That Converts The
Activity Of A Specific Ion Dissolved In A Solution Into An Electrical
Potential, Which Can Be Measured By A Voltmeter Or Ph Meter..
1- Main Idea For Operation.
► One Electrode Is Only Sensitive To One Type Of Ions. For Example,
The Na+ Electrode Is Only Sensitive To Na+ Ion, Not Sensitive To Other
Ions
2- System Structure.
A) Parameters :
1- URTI-910 Plus:
► Measurement Parameters: K+, Na+, Cl-, Ica2+, Ph, TCO2
► Calculation Parameters: Nca2+, Tca2+, AG
2- URTI-910C Plus:
► Measurement Parameters: K+, Na+, Cl-, Ica2+, Ph
► Calculation Parameters: Nca2+, Tca2+
Where :
1- Tca2+ Means Total Ca2+ In Plasma.
2- Ica2+ Means Ionic Ca2+ (Ica2+occupies 45-50% Of Tca2+)
3- Nca2+ Means Standard Ionic Ca2+(when Ph=7.4, Ica2+= Nca2+ )
4- TCO2 Means Total Carbon Dioxide In Plasma.
5- Ag Means Anion Gap, Which (AG= [ Na+]-[cl-+hco3-])
B) Main View :
Thermal Printer

Liquid Crystal
Display
Probe Cover

Start And Stop


Buttons

Sample Plate Reagent Kits

Unit Cover Bottom Cover


B) Main View :
► Liquid Crystal Display: Display The Menu And Test Results.

► Buttons: Button Above For Quick Test, Button Below For Stop Or
Back.

► Cover: Describe The Flow Cell And The Internal Flow Path.

► Reagent Kits: Contains Calibration Solutions.

► Bottom Cover: Connect Sample Plate And The Main Engine.


► Sample Plate: Automatically Sampling.
► Probe Cover: Avoid Human Touch In The Operation.
► Printer: Print The Test Results By Build-in Thermal Printer.
C) Flow View (Without Co2 Module (ISE Method) ) :

Main Peristaltic Pump


Flow Cell
Induction light
coupling

Sample Distribution Silicone adapter of reagent bag


injection needle valve
C) Flow View (With Co2 Module(Pressure Method)) :
Main Peristaltic Pump
Induction light Flow Cell
coupling

Silicone adapter of reagent bag Reaction Prestaltic Pump

Sample Distribution
injection needle valve

Pressure
sensor

Pipe pressed reaction


Solenoid valve valve vessels
A Set Of Electrodes
2- System Installation. Automatic Sampling Disc

Calibration
Solution C

Two Rolls
Of Printing

Refill Solutions

One Sample Disc One Power Cord

One Stylus, Two Pump Pipes, And Respectively One


Manual And Packing List
Auto Sampler Installation.
1. Take Out The Protective Foam For
Instrument Base Plate On The Instrument
Side From The Packing Box.

2. Take Out The Instrument Base Plate On


The Instrument Side From The Packing
Box.
3. Place The Instrument Base Plate On
A Horizontal And Firm Desktop Free Of
Vibration.

4. Take Out The Main Machine, Pass The


Cable Of The Automatic Sampling Disc
From Below Through The Square Hole To
Above The Large Hole.
5. Take Out The Main Machine, Correctly
And Steadily Place It On The Base Plate By
Aligning Its Four Feet With The Four Holes
Of The Base Plate.

6. Separately Align The Four Food Pads Of


The Main Machine With The Four Holes Of
Base Plate.
7. Pass The Cable Of The Sampling Disc
Through The Large Hole, Take Out The
Automatic Sampling Disc, And Insert The
Cable Plug.

8. Align The Notch Of The Automatic


Sampling Disc With The Syringe
Mouthpiece In Place.
9. Take Out The Sample Disc, Align It With
The Hole Direction, And Gently Press It In
Place.

10. After Leveling The Sample Disc, Slightly


Rotate It To Align “W” With The Syringe
Position.
Reagent Pack Installation.
► The Reagent System Of The Instrument Includes:
► Standard Solution A.
► Standard Solution B .
► For URIT-910 Plus:
► Tco2standard Solution.
► Tco2reaction Solution.
 All Reagents Are Inside The Kit.

 High Safety Features ,All Reagent &


Waste Cycle Inside The Kit ,No Need
To Dump Waste Everyday.

 ID Card To Close Reagent ,Protect


Distributor Interest .
Reagent Pack Installation.
1. Take Out The Reagent Bag, Take
Off The Silica Gel Plug, And Keep The
Plug For Future Plugging After End
Of Use Of The Kit, To Prevent The
Biohazard.

2. Push The Reagent Bag Horizontally


Until The Position Line Is Aligned With
The Machine Body.
Electrodes Installation.

1. Open The Door, And Tear Down The


Fixing Tape On The Electrode Holder.

2. Take Out All Electrodes, Add The


Internal Filling Solution For Reference
Electrode, Check The Situation Of The
Internal Filling Solution For The
Ordinary Electrode, Flip Off The
Bubbles, Wipe Clean The Housing, And
Install The Electrodes According To The
Corresponding Positions.
1.Check If Internal Filling Solution Reach
1/2~3/4 Level .If Not ,Replace With Fresh
Solution .
2.Remove Air Bubble In The Solution By
Knocking Gently .
3. Clean Liquids Outside

1. Remove 2 caps in the red cycle


2. Soak electrode whole body in normal
temperature distilled water for 15-30
minutes, caps no need to soak
3. Clear liquid ,close the cap with
4.Fill reference solution (about 1/2~3/4
full)
5. Lean To Left About 45
Degree ,And Knock The Electrode
Gently To Remove Air Bubble In The
Lower Chamber .
6.Clean The Liquid On The Outside
Printer Installation.
1. Chamber Switch
2. SEL Button
3. LF Button
4. Chamber Cover

1. Put The Thermo sensitive Printing Paper In


The Paper Bin (The Thermo sensitive Face Of
The Printing Paper Shall Be Downward,
Otherwise, Characters Cannot Be Printed Out),
Then Close The Paper Bin Cover. Pay Attention
To The Paper Direction When Installing The
Thermo sensitive Printing Paper.
4- Operating System.
1- Check Whether The Supply Voltage Is 100V-240V~, If The Voltage Is
Not Stable, UPS Or Stabilizer Are Recommended To Be Used.
2-Connect The Instrument And The Socket With Power Line, Avoid
Looseness.
3-Power Socket Must Be
Grounded Well.

4-Turn The Power ON.

5- System Self Check.


6- After self check all items
Must give (Ok) , if not please
Check Errors then try again.

7- Reagent Pack Massage


Then Perform ISE Calibration if we
choose (Back) or replace with
another pack first before
calibration if we choose (OK).
Main Menu:
1. Cal Menu :
► On the left column is “Auto Calibration” option. Operator can
change the calibrating time by clicking the text-box “ ”on the right side
as illustrated :
► ISE Automatic Calibration Function : It must be turn “on”, the default
calibration time is “23:59”.
► TCO2 Automatic Calibration Function: It is optional to turn on or turn
off, the factory setting is turning “on”. (It is for URIT-910 Plus only).
► ISE Cal: The instrument will calibrate with Standard Solution A
and Standard Solution B successively. When the calibration is finished,
the instrument will display the evaluation on the results

► The Calibration Value Must Be Sure:


A-a'≤0.5mv b-b'≤0.5mv
► TCO2 System Calibration :
Click the check-box “” in front on the “TCO2 Cal.” and then click
“Start”, the instrument will conduct the “TCO2 System Calibration”
procedure. The instrument will calibrate with TCO2 Standard Solution
and Reaction Solution successively.

► Calibration Voltage Range 900-1300 Is ,If We Have


Two OK In Three Times ,It Will Pass .
2. Sample Menu :

• S.Type (Sample Type): Serum, Whole Blood, Urine.


• Test Item: ISE, ISE+T, TCO2.
• Start No.: 01,02,...,31.
• End No.: ≥First Cup, No.≤31.
• Recycles.: 01,02,…,15.
► Note:
This Device Allows The Insertion Of Emergency Cup :
1- Put The Sample Cup Into The Emergency Cup Position And Then
Press The Button Of "Emergency Insertion" .
3-QC Menu :

► point Standardization: Absorb Right Amount (More Than 300


Microlitres) Sample To The Sample Cup, Placed On The
Sample Hole Of The Sample Tray, To Ⅰpoint Standardization,
Execution Steps:
► Click On The "QC" Of The Main Menu "Ⅰpoint Sta." Enter The
Concentration Of Target Value Of The Sample Set The Position
Of Sample Placement Click On The "Test".
► Ⅱ point Standardization: Absorb Right Amount (More Than
300 Microlitres) Sample To The Sample Cup, Placed On The
Sample Hole Of The Sample Tray, To Ⅱ point Standardization ,
Execution Steps:
► Click On The "QC" Of The Main Menu Ⅱ point Standardization
Enter The Concentration Of Target Value Of The Sample Set The
Position Of Sample Placement Click On The "Test".
► TCO2 standardization: Absorb Right Amount (More Than 300
Microlitres) Sample To The Sample Cup, Placed On The Sample Hole
Of The Sample Tray, To TCO2 standardization, Execution Steps:
► Click On The "QC" Of The Main Menu TCO2 standardization Enter
The Concentration Of Target Value Of The Sample Set The Position
Of Sample Placement Click On The "Test".
►Stat Factor:
► It will illustrate Slope and Intercept For Each Electrodes.
►QC Setting :
► On The “QC Item”, Click “QC Setting”, The Instrument Will
Enter The “QC Setting” Interface.
1- Click The Text Box On The Right Side Of "QC Lot." And It Will Pop
Up Numeric Keypad. Operator Should Input The Corresponding Batch
Number Of QC Solution. Such As 1202001.

2- Click The Drop-down List On The Right Side Of "Qc Name" And
Select A Qc Solution Is Used. Such As Low QC Solution: Q1.

3- Click Numeric Keys On The Left Column Of The Screen And It Will
Pop Up A Numeric Keypad. Operator Should Input The Target Value
Of Potassium (K), Sodium (Na), Chlorine (Cl), Calcium (Ica), Ph, 2
Times The Standard Deviation (2SD) In Accordance With The
Instruction. Click "Save" To Save The Settings.

Note: Q1, Q2, Q3 Are ISE Qc Solutions.


TCO2 QC Solution Setup :

1- Click The Text Box On The Right Side Of "QC Lot." And It Will Pop
Up Numeric Keypad. Operator Should Input The Corresponding
Batch Number Of QC Solution. Such As 1202001.

2- Click The Drop-down List On The Right Side Of "Qc Name" And
Select A Qc Solution Is Used. Such As Low QC Solution: C1.

3- Click Numeric Keys On The Left Column Of The Screen And It Will
Pop Up A Numeric Keypad. Operator Should Input The Target Value
Of TCO2 QC Solutions And 2 Times The Standard Deviation (2SD) In
Accordance With The Instruction.

Note: C1, C2, C3 Are Tco2 Qc Solutions


►QC Test :

1- Click On The Check Box On The Left Side Of "QC Name" And
Select QC Solutions To Test . Such As The Middle QC Solution: Q2.
(Note: Click “CLEAR” Can Remove The Selected QC Solution)

2- Click On The Drop-down List On The Right Side Of "Recycles" To


Select Test Times, Select 3 For Example.

3- Click "Start Test" To Start The Test.


►QC Data :
1- Click On The Drop-down List On
The Right Side Of “QC Name”,
Select A QC Solution.
2- Click On The Drop-down List On
The Right Side Of “QC Item”, Select
A Test Item.
3- Click the text box on the right
side of the screen, it will pop up
numeric keypad to set up the
date of inquiry.

4- Click “OK” to start the inquiry procedure.


4- Maintain :

► Flushing Pipeline
- “Flushing Pipeline”, the
instrument will suck standard
solution A to flush the whole
pipe system.
► Electrode Protein Cleaning:
- The Instrument Will Suck The Protein Clean Solution To Flush The
Pipes.
Note: After Electrode Protein Cleaning, It Needs ISE System
Calibration, Till The Calibration Result Is Stable Can It Measure
Samples
► System Maintenance :
- Enter the password “123” if you want
to set the protein removal interval in
the manufacturer maintenance.
- Users can set the number of days
or times of instrument automatic
protein removal as required by
the hospital, and the instrument
default is separately 7 days and
100 times of test.
5- Service:

Test Data:
- According to operator needs , we can
choose first condition of search then we
press Start to review old data and easy
get it direct.
Cal Data:
- According to operator needs , we can
choose first condition of search
between (ISE & TCO2) then we press
on it to review calibration data and
easy get it direct.
Slope :
Electrode K+ Na+ Cl- Ca2+ pH
Reference range 40~ 40~ 40~ 20~ 40~
of 70 70 70 40 70
slope/(mv/dec)
Replace Bag :
- Firstly we connect New Bag
then press Replace Bag .

- It will take several minutes to


replace the reagent bag. The
instrument states will display “OK”
when each pipeline is filled with
reagent
Transfer :
- Select "Open" The Serial Port, With
19200 Baud Rate For The Instrument,
You Can Send The Sample Data By
Means Of Connecting The RS - 232 Port
In The Back Cover Of The Instrument .
Bluetooth Can Transmit The Sample
Data As Well
Stop Use :
- If Operator Needs To Turn Off For A Long Period Of Time, You
Can Click On The "Stop Use" Option In The Second Column Of The
"Assistant", And Then Operate Follow The Step Tips Instrument
6- Setup :

Normal Range
- Operator can set the range according
to actual situation and the existing
data of the instrument can be used as
a reference. Touch Save after
completing the setting to quit
Date & Time
- Operator Can Touch the
corresponding digit to modify it
when the instrument time is
incorrect.

Printer Option
- Operator Can Touch the
corresponding Item to print it or not.
Set Channel :

-Operator Can Touch the


corresponding digit to Choose between
run item or not.

- When Certain Items Do Not Pass The


Electrode Calibration, Such Items Can
Be Disabled, Thus To Use The
Instrument.
5- Hardware Structure (Electrode Refill Solution).
5- Hardware Structure (Reference Refill Solution).
5- Hardware Structure (Reference Refill Solution).
5- Hardware Structure (Cleaning and maintenance of sampling
needle nozzle ).
5- Hardware Structure (Cleaning and maintenance of Reaction
Vessels ).
5- Hardware Structure (Cleaning and maintenance of Reaction
Vessels ).
5- Hardware Structure (Disassembly and assembly of Reaction
Vessels ).
5- Hardware Structure (Cleaning and maintenance of Distribution
Valve).
5- Hardware Structure (Disassembly and assembly of Distribution
Valve ).
5- Hardware Structure (Disassembly and assembly of Automatic
sampling disc).
- Start The Machine By Pressing
And Holding The Quick Sample
Test Key, To Enter The Backdoor
Debugging Program. .

Plate Adjustment
- Reset Adjustment Refers To
The Position Where The Syringe
Is At “W” Hole Position During
Reposition Of The Sampling
Disc. Increase The Digit If The
Position Deviates To Left And
Vice Versa.
- Positioning Adjustment Refers To The Position Where The
Syringe Points Towards The Sample Cup During Sampling.
Increase The Digit If The Position Deviates To Left And Vice Versa
Assembly, Disassembly And Maintenance Of Main And Auxiliary
Peristaltic Pump
Assembly, Disassembly And Maintenance of Flow Cell :
Assembly, Disassembly And Maintenance of Pressure Sensor :
Assembly, Disassembly And Maintenance of Pipe Pressed Valve:
Assembly, Disassembly And Maintenance of Display Screen:
Assembly, Disassembly And Maintenance of Elevator:
Assembly, Disassembly And Maintenance of Induction Light Opt
coupler:
Solution Level Adjustment :

1. Adjust Clockwise To VR1 Until A Click Sound Is VR1


Heard, At This Time The Voltage Is Adjusted To
The Maximum And No Further Adjustment Is
Required.
2. Boot Up The Instrument, The Multimeter Shows 20V Gear DC Voltage, The Pen
Connection Diagram Corresponds To Two Test Points; Adjust VR2 Clockwise To
Range From 1.4V To 1.6V When There Is No Liquid In The Pipelines Of Liquid Level
Optical Coupler. The Voltage Is Larger Than 2.3V When There Is Liquid. After
Adjustment, Calibrate And Observe Whether Or Not The Liquid Is Located
Normally.
Reaction Cup Rotational Speed Adjustment :

1. The DC Voltage Of Multimeter


Is 20V Gear, The Red Pen
Connects TP5 Round Hole And
The Black Pen Contacts The Earth,
The Ground Voltage At TP5 Test
Point Shall Be 12V±0.5v.
TP5
2. If the voltage is less than
12V±0.5V, then TCO2 calibration VR3
procedure shall be executed and
observe whether or not the voltage
remains at about 12V when the
stirring bar in the reaction cup
rotates; if not, adjust VR3 clockwise
to turn it down and vice versa.
Fault Fault Reason Resolving
Code
1) Check The Pipe, It Is OK Or Not.
2) Replace The Electrode Silver Stick.
3) Replace The Filling Solution.
E01 Potential Mill volt <10 Or >120 4) Replace The Electrode Contact.
5) Replace The Electrode.
6) If All Electrodes Are Drift, There May Be Some Faults In
The Reference Electrode.
1) Check The Pipe, It Is OK Or Not.
2) Execute The Protein Removal Operation.
3) Replace A Reagent Bag.
4) Check The Electrode, If There May Be Leakage,
A And A’ Or B And B’ Mill volt Bubble, Dirt, Silver Stick Coating Stripping And Non
E02 D-value >0.4 Filling Solution In It Or Not.
Ph D-value:>2.0 5) Check The External Power, The Ground And Stabilivolt
Device Are Connected Well Or Not.
6) If All Electrodes Are Drift, There May Be Some Faults In
The Reference Electrode.

Electrode Slope K:<40 Or >70 1) Check The Pipe, It Is OK Or Not.


Na:<40 Or >70 2) Replace A Reagent Bag.
E03 Cal:<35 Or >70 3) Execute The Protein Removal Operation.
Ca:<20 Or >40 4) Replace The Filling Solution.
Ph:<40 Or >70 5) Replace The Electrode.
E04 No Standard Solution A 1) If The Reagent Is Used Up Or Not.
2) Check The Reagent, If It Is Pumped Up, Check The
E05 No Standard Solution B Induction Light Coupling.
3) Check The Electrode Assembly, It Is Or Not.(Holes Are
E06 No Standard Solution A And B Aligned, Electrode Is Even.)
4) Check The Pipes, If There Are Blocks And Leakages In
Them Or Not.
E07 No TCO2 No Standard Solution 5) Check The Pump Tube, If There May Be Conglutinated
And Cracked.
URIT Medical Electronic Group Co., Ltd.

Head Office
Add.: No.4 East Alley, Jiuhua Road, Guilin, Guangxi 541001, P.R. China
Tel: +86-773-2288586 Fax: +86-773-2288560
Shenzhen Office
Add.: Room EJ, 24th Floor, New Baohui Building, Nanhai Road, Nanshan District, Shenzhen, P.R. China
Tel: +86-755-26050853 Fax: +86-755-86036150
E-mail: [email protected], [email protected] Http://www.urit.com

You might also like