Amx3 Operator Manual
Amx3 Operator Manual
II
PRODUCT
mmm
: v ' .
OPERATING MANUAL
OM A0654F
11-481
AMX-3
X-RAY UNIT
G E N E R A L S ELECTRIC
Operating (^Manual OM A0654F
11-681
AMX-3
X-RAY UNIT
11-681 O M A0654F
FOREWORD
TABLE 1
Bask: Unit, Control Mobile Base, 46-217900G2 Next to exposure control handswitch
Support Column, 115V, 50/60 Hz bracket on front of unit
X-ray Tube Unit 46-155750G6 Caltiode trim cap housing, rear trim
cover
1
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UNIT
11-681 O M A0654F
• A n audible tone and/or red X - R A Y light indicates If electrolyte gets o n the skin, w a s h the affected areas
p r e s e n c e of radiation. If either indicator is o n at any with large quantities of water; then neutralize skin with
time other t h a n d u r i n g the e x p e c t e d X - R A Y e x p o s u r e vinegar, l e m o n juice, or 3 % acetic acid.
Interval, immediately release the handswitch
R O T O R and/or X - R A Y button a n d turn off m a i n circuit
breaker. Do not e n e r g i z e t h e unit a g a i n until it h a s If electrolyte gets into the e y e s , flush with water and
b e e n c h e c k e d by a qualified X-ray service represent- GET IMMEDIATE MEDICAL ATTENTION.
ative.
11-681 OM A0654F
SECTION 1 — DRIVING
A. MAIN POWER
SWITCH
B. DIRECTION
SELECTOR
SWITCH
33631-D01
E DRIVE ASSIST C. BRAKE HANDLE D. SPEED CONTROL
FIGURE 2
3
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UNIT
11-681 O M A0654F
33630-001
PIGURE 3
4
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION A M X - 3 X - R A Y UNIT
11-681 OM A0654F
LATCH
LATCH
BAR
3421B-001
FIGURE 5
FIGURE 4
5
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UNIT
11-681 OM A0654F
2-2 R E L E A S E A N D A D J U S T M E N T O F H O R I Z O N -
TAL TELESCOPING ARM
VERTICAL
ARM LOCK
33637-D01 33628-D01
FIGURE 6
6
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UNIT
11-681 OM A0654F
YOKE ROTATION
LOCK
TUBE UNIT
ROTATION LOCK
33639-001
X-RAY
TUBE
ALUMINUM
FILTERS
ROTATIONAL
FIELD LIGHT DETENT PLUNGER
BUTTON KNOB
BLADE CONTROL
CALIBRATED
KNOBS
SCALE DRUM
TRACK FOR 33641-DOl
EXTENSION
CYLINDER MEASURING 33663-D01
7
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UNIT
11-681 O M A0654F
D e p r e s s t h e field light b u t t o n o n t h e h a n d s w i t c h cylinder into the tracks f r o m the front of the collimator
W I T H O U T d e p r e s s i n g t h e rotor s w i t c h . until the cylinder latch s n a p s into the slot o n the right
track. Press t h e latch to release the cylinder for r e m o v a l .
N O T E : T h e field light will r e m a i n O N as long as t h e T h e a d d e d weight of the e x t e n s i o n cylinder will affect
h a n d s w i t c h field light b u t t o n is d e p r e s s e d a n d will the c o u n t e r b a l a n c e of the vertical c o l u m n . Lock the
turn off u p o n release of the field light b u t t o n , or d e - vertical a r m w h e n installing or r e m o v i n g the extension
pression of t h e rotor b u t t o n . cylinder a n d apply extra caution w h e n positioning the
unit w i t h t h e e x t e n s i o n cylinder installed.
Position the collimator a n d t u b e unit u s i n g the illumi- A d j u s t t h e size of t h e field b y positioning t h e collimator
n a t e d field a n d cross hairs s h a d o w . T h e cross hairs blades u s i n g the Blade C o n t r o l K n o b s (see Figure 7 ) .
indk;ate the field center. R e f e r to S e c t i o n 2-3 for t u b e Field sizes f r o m 0" x 0 " up to 17" x 17" (43 c m x 43 c m ) at
positioning a d j u s t m e n t s . a 4 0 " ( 1 0 2 c m ) S I D can be o b t a i n e d . T w o pointers on t h e
c a l i b r a t e d scale d r u m indicate field d i m e n s i o n s for a
U s e t a p e m e a s u r e o n tx)ttom of collimator (see Fig- s e l e c t e d S I D . T h e right Blade Control K n o b controls the
u r e 7 ) , calibrated in i n c h e s a n d c e n t i m e t e r s , to m e a s u r e t r a n s v e r s e d i m e n s i o n i n d i c a t e d by the upper pointer.
s o u r c e - t o - i m a g e (receptor) d i s t a n c e (S(D). M e a s u r e t h e T h e left k n o b controls the longitudinal d i m e n s i o n indi-
con^ect S I D b y h a v i n g the b o t t o m of t h e collimator h o u s - c a t e d by t h e lower pointer.
i n g e v e n with the m a r k i n g s o n the t a p e w h i c h c o r r e s -
p o n d to t h e s o u r c e - t o - i m a g e (receptor) d i s t a n c e . T h e Dial n u m b e r s o n t h e Calibrated Scale D r u m (see Fig-
t a p e t a b o n t h e e n d of t h e t a p e m u s t b e oriented parallel u r e 7) are for 4 0 , 4 8 , 6 0 , a n d 70-tnch (102^ 122, 152,
w i t h the tape for a c c u r a t e m e a s u r e m e n t s . 178 c m ) SID'S a n d s t a n d a r d film sizes of 5 to 17 inches
(13 c m to 4 3 c m ) .
N O T E : T h e a c c e s s o r y E x t e n s i o n Cylinder m o u n t i n g
t r a c k s (see Figure 7) limit m i n i m u m source-to-skin d i s - S e l e c t the scale for o n e of t h e four SID's by turning t h e
t a n c e t o 30 c e n t i m e t e r s . T h e s e tracks accept a n o p - S c a l e Selection K n o b until d e s i r e d detent position is
tional e x t e n s i o n cylinder. T o u s e , slide the e x t e n s i o n reached.
COLUMATOR ROTATIONAL
DETENT PLUNGER
33642-D01
FIGURE 9
8
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UNIT
11-681 O M A0654F
D e p r e s s t h e Rotational Detent Plunger K n o b (see Fig- N O T E : Rotational Detent Plunger wit! not operate u n -
u r e 7) to rotate the collimator a b o u t its vertical axis. less knob is first rotated fully counterclockwise.
9
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y ^ N ) T
11-681 O M A0654F
M a k e sure m a i n p o w e r circuit breaker is o n and battery s w i t c h c a n n o t be withdrawn while the unit is switched
c a b l e is not p l u g g e d in. Referring to Figure 1p, p e r f o r m " O N " . T a k e the key out of the socket a n d retain in
ttie following s t e p s . p o s s e s s i o n of d e s i g n a t e d operator w h e n unit is not
in u s e .
Rotate the X-ray G e n e r a t o r Keyswitch to the O N 2. Select d e s i r e d m A s value. All exposures with the
position for e x p o s u r e s . T h e " O N " light should g l o w at A M X - 3 are m a d e at 100 milliamperes. m A s is select-
this time. C h e c k S t a t u s P a n e l for battery c o n d i t i o n . able in 26 s t e p s f r o m 0.64 m A s to 3 2 0 m A s with e a c h
T h e red R e c h a r g e I m m e d i a t e l y light should not be step in 2 5 % i n c r e m e n t s . Exposure time is deter-
o n . If it is, the battery m u s t be r e c h a r g e d before m i n e d by dividing the selected m A s value by 100.
m a k i n g e x p o s u r e s . W h e n not m a k i n g e x p o s u r e s , For e x a m p l e , if 50 m A s is selected, Ihe exposure
rotate this switch to C h a r g e w h e n c h a r g i n g the bat- time w o u l d be .5 s e c o n d s .
teries, or O F F w h e n unit is not being u s e d for X-ray or
charging, 3. Select d e s i r e d k V p v a l u e . kVp is selectable in 24
steps f r o m 50 k V p to 125 k V p with 2 kVp steps f r o m
A captive key is p r o v i d e d for this switch to assure that 5 0 kVp to 76 k V p a n d 5 kVp steps f r o m 80 k V p to
only a u t h o r i z e d p e r s o n n e l operate the unit. T h e key- 125 k V p .
FIGURE 10
10
G ^ E R A L ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UNIT
11-681 OM A0654F
D e p r e s s the handswitch Rotor button (see Figure 11) to N O T E : T h e rotor button should be released immedi-
start rotation of tube a n o d e . T h e r e will be a 2 . 5 - s e c o n d ately after e x p o s u r e to m a x i m i z e battery charge life.
delay before the g r e e n R e a d y indicator g l o w s .
N O T E : A l t h o u g h e x p o s u r e s can be m a d e in rapid suc-
After R e a d y Indicator g l o w s , d e p r e s s t h e X-ray button c e s s i o n , for stated technique factor accuracy, sufficient
o n the right side of the h a n d s w i t c h (Figure 11) to initiate time s h o u l d be allowed for the Wait light to go out before
X-ray e x p o s u r e . A n a u d i b l e tone signal will s o u n d a n d a g a i n d e p r e s s i n g the rotor switch, approximately
the red X-ray O n Indicator will g l o w during the time that 2 5 s e c o n d s following prior release of the rotor switch.
X-rays are e m i t t e d .
C h e c k the Status P a n e l before each exposure. If the
If the Rotor a n d X-ray buttons a r e d e p r e s s e d s i m u l t a - R e c h a r g e Immediately indicator is o n , the batteries
neously, the X-ray e x p o s u r e will o c c u r after the Z.5-sec- m u s t l>e r e c h a r g e d as described in Section 7. Failure to
o n d delay a n d as the R e a d y Indicator c o m e s o n . r e c h a r g e batteries at this time m a y result in failure to
meet s t a t e d technique accuracies a n d possible d a m a g e
W A R N I N G : I F E I T H E R T H E X - R A Y " O N " INDICA- to the battery. If t h e R e c h a r g e S u g g e s t e d indicator is o n ,
TOR G L O W S OR T H E A U D I B L E SIGNAL CON- c h a r g i n g of the batteries is not mandatory and additional
T I N U E S F O R M O R E T H A N 3.2 S E C O N D S , R E L E A S E e x p o s u r e s a r e permitted. However, the number of re-
T H E R O T O R AND X - R A Y B U T T O N S AND T U R N T H E m a i n i n g e x p o s u r e s available will d e p e n d upon condition
MAIN P O W E R C I R C U I T B R E A K E R O F F . IN T H I S of batteries, t e c h n i q u e s selected, and distance traveled
C A S E , DO NOT A T T E M P T A N O T H E R E X P O S U R E between exposures.
U N T I L T H E UNIT H A S B E E N C H E C K E D B Y A Q U A L -
IFIED S E R V I C E R E P R E S E N T A T I V E , Turn the X-ray Generator keylock switch to O F F w h e n
completed.
N O T E : T h e Ready indicator stays o n if the rotor switch
is held after an e x p o s u r e , but the rotor s w i t c h m u s t be T h e unit m a y be driven to a new location. Secure the unit
released a n d d e p r e s s e d a g a i n b e f o r e another e x p o s u r e for transport as d e s c r i b e d in Section 6. Only the Main
can be m a d e . P o w e r C i r c u i t B r e a k e r s h o u l d b e O N while driving.
I
FIELD X-RAY
f
JGH1
3363Z-D01
FIGURE 11
11
GENERAt ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y GNIT
11-681 OMA0654F
YELLOW GREEN
AMBER
FIGURE 12
12
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UNIT
11-681 OM A0654F
Turn keylock switch to " O F F " . L o o s e n vertical a r m lock a n d slowly lower arm until the
latch a n d latch bar e n g a g e . W h e n driving the A M X - 3
Return collimator to 0° position {refer to Figure 9). f r o m r o o m to r o o m , the telescoping a r m is s e c u r e d by
m e a n s of the latch a n d latch bar. The telescoping a r m
R e m o v e e x t e n s i o n cylinder, if u s e d . m u s t be in the position s h o w n in Figure 13. The latch at
the b o t t o m of the telescoping a r m automatically latches
Return tube unit to 0" position (refer to Figure 8). to t h e talch'bar w h e n they a r e j o i n e d . T o release, rotate
t h e latch clockwise a n d raise the telescoping a r m (refer
Return yoke to 0° position a n d s e c u r e (refer to Figure 7 ) . to Figure 4 ) .
Rotate c o l u m n until a r m is In s t o r a g e position (refer to M a k e sure charger cable is fully retracted (refer to Fig-
Figure 13). ure 14).
33629-DOf
33640-D01
FIGURE 13 FIGURE 14
13
GENERAL ELECTWC MEDICAL SYSTEMS OmSlON AMX-3 X - R A Y ONIT
11-681 OM A0654F
14
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UNIT
11-681 OM A0654F
26-1/2
{.67 m)
FRONT VIEW
SIDE VIEW
34217-D10
15
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y L?NIT
11-681 O M A0654F
P o w e r r e q u i r e m e n t s for r e c h a r g i n g the battery: 110 For radiation output, t h e coefficient of variation is less
to 1 2 5 V A C , 5 0 / 6 0 Hz, 6.0 a m p s . t h a n 0.05 for successive e x p o s u r e s having constant
t e c h n i q u e factors.
kVp
X-ray G e n e r a t o r . R a t i n g s a n d Duty C y c l e
50 kVp fo 74 kVp ± ( 9 % + 1 kV}
M a x i m u m Ratings
76 kVp to 125 kVp ± ( 1 2 % 4- 2 kV)
Ail r a d i o g r a p h i c e x p o s u r e s w i t h t h i s unit a r e at
100 milliamperes. k V p a c c u r a c i e s subject to additional ± ( 5 % + 2 kV)
a c c u r a c y of i n s t m m e n t a t i o n u s e d for calibration and
kVp (Maximum) mAs (Maximum) measurement.
All ratings a n d duty c y c l e s are subject to limitations Conditions required to a c h i e v e the a b o v e accuracies:
of the X-ray t u b e rating a n d apply at altitudes up to
8 0 0 0 f e e l (2.438 k i l o m e t e r s ) , a v e r a g e relative h u m i d - 1. T h e rotor h a n d s w i t c h must have b e e n released
ity not e x c e e d i n g 8 0 % , a n d ambient t e m p e r a t u r e after t h e c o m p l e t i o n of the previous e x p o s u r e for a
b e t w e e n e O T (IB-e^C) a n d 100"F ( 3 7 . 8 X ) . sufficient time (approximately 25 seconds) tor the
16
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UNIT
11-681 OM A0654F
M i n i m u m inherent filtration of the tube unit is 1.0 m m DIAGNOSTIC SOURCE ASSEMBLY MINIMUM
a l u m i n u m equivalent, m e a s u r e d at 130 k V p . INHERENT FILTRATION
17
GENERAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UJ^IT
O M A0654F
11-681
\
HRT-B F
1.200 (MODEL NO, 46-155750)
9-54 50-60
\
<
lU
X 1,000 MAX, COOLING RATE =
20,000 HU/MIN.
^ 900
< d
lU 800
CO (/>
o
3
I
X
<
700
O ^ 500
<
DC ^ 400
o 300
(/)
o 200
z
CO 100
3
1 2 3 4 5
TIME IN MINUTES
6 7 0
^179^0.0
o 0 50 100 150
TIME IN MINUTES
200 250 300 350
»)6HDIO
X
18
GEf^RAL ELECTRIC MEDICAL SYSTEMS DIVISION AMX-3 X - R A Y UNIT
11-681 O M A0654F
SECTION 9 — SERVICING
19