Sonoline G20..
Sonoline G20..
Sonoline G20..
Ultrasound Division
Revision Data
Page 1 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 2 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
2.0 Scope 5
3.0 Definitions 6
5.0 AE Specifications 19
Page 3 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
8.0 Configuration 53
9.0 Security 58
Page 4 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
1.0 Purpose
This document describes the conformance to the ACR-NEMA DICOM 3.0
Standard by the SONOLINE G20 ultrasound system software version 2.0
from Siemens Medical Solutions USA, Inc. Ultrasound Division. It shall
establish the conformance specifications for this system only, and does not
apply to other products offered by Siemens Medical Solutions USA, Inc., or
its affiliates.
The SONOLINE G20 system is a device that generates ultrasound images
that can be sent using DICOM standard protocols and definitions to other
DICOM compliant devices that support SOP classes as defined in Table 2
on page 19 in this document.
2.0 Scope
The DICOM standard provides a well-defined set of structures and proto-
cols that allow inter-operability to a wide variety of medical imaging devices.
When configured with the DICOM option, the SONOLINE G20 system pro-
vides support for essential services related to ultrasound scanning and con-
nectivity to DICOM compliant devices. The SONOLINE G20 system prod-
ucts will not support all features supported by the DICOM standard. This
document clearly states the DICOM services and data classes that are sup-
ported by the applications included with the SONOLINE G20. The intent of
this document is to allow users and other vendors who also conform to the
DICOM standard to exchange information within the specific context of
those elements of the DICOM standard that the SONOLINE G20 system
supports.
This document is written with respect to the adopted portions of the DICOM
standard, Revision 3. The following sections of this document follow the
1
outline specified in the DICOM Standard NEMA publication PS3.2.
Page 5 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
3.0 Definitions
The following table provides a list of terms, their acronyms (if applicable),
and their descriptions.
Table 1 Terms, Acronyms, and Descriptions.
American College of ACR- The American College of Radiology (ACR) and the
Radiology - National NEMA National Electrical Manufacturers Association
Electrical Manufacturer's (NEMA) formed a joint committee to develop a
Association standard for Digital Imaging and Communications
in Medicine (DICOM).
DICOM Message Service DIMSE Defines an Application Service Element (both the
Element service and protocol) used by peer DICOM
Application Entities for the purpose of exchanging
medical images and related information.
Digital Imaging and DICOM A well-defined set of structures and protocols that
Communications in 3.0 allow inter-operability to a wide variety of medical
Medicine, Version 3.0 imaging devices.
Picture Archiving and PACS A DICOM server that accepts medical images from
Communications Systems another DICOM system and stores the images for
later retrieval.
Protocol Data Unit PDU The PDUs are message formats exchanged
between peer entities within a layer. A PDU shall
consist of protocol control information and user
data.
Page 6 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Real-World Activity RWA That which exists in the real world which pertains
to specific area of information processing within
the area of interest of the DICOM Standard. Such
a Real-World Activity may be represented by one
or more computer information metaphors called
SOP Classes.
Service Class Provider SCP The role played by a DICOM Application Entity
(DIMSE-Service-User) which performs operations
and invokes notifications on a specific Association.
Service Class User SCU The role played by a DICOM Application Entity
(DIMSE-Service-User) which invokes notifications
and performs operations on a specific Association.
Service-Object Pairs SOP The union of a specific set of DIMSE Services and
one related Information Object Definition which
completely defines a precise context for
communication.
Unique identifier UID A series of digits and periods (.) used to uniquely
identify an object such as an Ultrasound image in
DICOM.
Page 7 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 8 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
SONOLINE G20
B/W Print AE B/W Print
SCU SCP
Modality Worklist
Worklist SCU SCP
MPPS MPPS
SCU SCP
Storage Storage
Commit Commit
SCU SCP
Page 9 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
4.1.1 Verification
Verification is a part of the DICOM configuration located on the ‘DICOM’
page of the System Presets. Verification can be used to send a DICOM Ver-
ification request to a remote Application Entity (AE) and listen for a
response.
When used as a diagnostic tool, Verification returns the following messages
to the user:
If the verification succeeds: “DICOM - Successfully contacted
system”
C-ECHO-RQ
Page 10 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
C-STORE-RQ
C-STORE-RSP
N-ACTION-RQ
DICOM N-ACTION-RSP DICOM
G20 STORAGE STORAGE
STORE COMMITMENT N-EVENT-RQ COMMITMENT
SCU N-EVENT-RSP SCP
Page 11 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
N-SET-RQ
N-ACTION-RQ
N-CREATE-RQ
N-DELETE-RQ
G20 DICOM DICOM
PRINT PRINT SCU PRINT SCP
N-SET-RSP
N-ACTION-RSP
N-CREATE-RSP
N-DELETE-RSP
Figure 5 Print Model.
Page 12 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
wildcard was appended at the end of each field. Patient ID, Requested Pro-
cedure ID and Accession number will be exact match only. If no matches
are found, a message will be presented to the operator indicating so. If
more that one matching patient is found, a pick list of patient procedures will
be presented to the user to select from. Each of the fields will be sortable in
ascending and descending order.
The pick list of patient procedures will be limited to a number of preset
entries. If more than this number of matching records are found in the query,
the search will terminate and the user will be notified. The search list criteria
will contain:
Patient name
Patient ID
Accession number
Exam start date/time range
Requested Procedure ID
US/All modalities
Scheduled station AE title
Scheduled Performing Physician’s Name
C-FIND-RQ
G20
New Patient
C-FIND-RSP
Once a Worklist query is initiated, a “Retrieving worklist, please wait ...” dia-
log will be presented to the user. The user will only have one option, “Can-
cel,” which will abort the query operation.
The following data fields in Modality Worklist Screen are initially populated
from the New Patient Screen, if filled in, and can be used for query:
Page 13 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
The following data fields will be populated on the worklist screen for each
return:
The first procedure step is highlighted initially. The user can highlight a dif-
ferent procedure step, if desired, and select the highlighted patient proce-
dure step or cancel the operation. Selection of a procedure step from the list
will cause demographic information for the patient to be loaded in to the
patient data fields. Additional information from the procedure step will be
used to populate attributes in any images and MPPS messages generated
during the procedure step.
If the Worklist Server provides one or more of the attributes in the following
table with the procedure step, the data provided will be displayed to the
user.
Page 14 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
The following data fields will be populated on the patient data screen:
N-CREATE-RQ
N-SET-RQ
G20 DICOM DICOM
New Modality Performed Modality Performed
Procedure Step Procedure Step
Patient SCU N-CREATE-RSP SCP
N-SET-RSP
Page 15 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 16 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 17 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
b. The series is closed before all images are stored successfully, all previ-
ous stores have succeeded and the last image stores successfully.
c. The series is closed before all images are stored successfully, at least
one store has succeeded, at least one store has failed and the last store
with non-zero retry count fails or succeeds.
d. A series has been partially committed as in c. Later, due to "Retry Job”
button press on the Store Status UI screen the store jobs are retried.
Another Storage Commit is sent when at least one store has succeeded
and the last store with non-zero retry count fails or succeeds.
MPPS
The MPPS (if enabled) command is sent in the following situations:
a. N-CREATE command is sent whenever a new procedure step is
selected. The state of the MPPS command is set to “In-Progress”.
b. N-SET command is sent when the Procedure Step is closed by the user
pressing either the Completed or Discontinued button on the Close Proce-
dure dialog. The state of the MPPS command is set, according to the state
(Completed or Discontinued) set by the user.
Page 18 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
5.0 AE Specifications
The following specifications apply to the SONOLINE G20 AE as depicted in
Figure 1.
Verification 1.2.840.10008.1.1
Ultrasound Image Storage (Retired) 1.2.840.10008.5.1.4.1.1.6
Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1
Ultrasound Multi-Frame Image Storage (Retired) 1.2.840.10008.5.1.4.1.1.3
Ultrasound Multi-Frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1
Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7
Storage Commitment - Push Model 1.2.840.10008.1.20.1
Basic Grayscale Print Management Meta SOP Class 1.2.840.10008.5.1.1.9
Basic Grayscale Image Box SOP Class 1.2.840.10008.5.1.1.4
Printer SOP Class 1.2.840.10008.5.1.1.16
Modality Worklist Information Model C- FIND 1.2.840.10008.5.1.4.31
Modality Performed Procedure Step 1.2.840.10008.3.1.2.3.3
5.1.1.1 General
The SONOLINE G20 system utilizes TCP/IP. The Maximum Length PDU
negotiation is included in all association establishment requests. The max-
imum length PDU offered for an association initiated by the SONOLINE
G20 is:
Maximum PDU Offered: 28672
The SONOLINE G20 initiates each C-Store Request one at a time, one for
each transfer request being processed.
Image format on the SONOLINE G20 can be set to one of “Automatic”, “Old
Ultrasound” or “Secondary Capture”.
Page 19 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 20 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
The SONOLINE G20 facilitates users to store images as they are being cre-
ated or later in review mode.
Queueing images during acquisition:
User can select one or more closed studies and queue them up for Storage.
The DICOM Store button is available in Review screen for this operation. All
images are stored.
Transfer of images to the storage server:
Page 21 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
The SONOLINE G20 system always acts as an SCU for store and is the cli-
ent in a client-server model.
Page 22 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
This is accomplished using the DIMSE C-STORE Service. The SCU issues
a service request with a SOP instance that meets the requirements of the
desired ultrasound or secondary capture IOD.
The following table denotes the attributes included in the Ultrasound Image
Object as implemented on the SONOLINE G20. Attributes not listed are not
used.
Table 6 Ultrasound Image and Ultrasound Retired Image IOD Attributes
Page 23 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
(b)
Series Time (0008,0031) Time the series started.
(b)
Series (0008,103E) Populated with Scheduled
Description Procedure Step Description if a
value was provided by Modality
Worklist.
Page 24 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
(b)
Protocol Name (0018,1030) The exam type of the most recent
image stored in a particular series.
If no images are stored for a series
then the value is set to
“Ultrasound”.
(b)
Request (0040,0275) Populated with Scheduled
Attributes Procedure Step Sequence
Sequence (0040,0100) if provided by Modality
Worklist.
(b)
Performed (0040,0260) Populated with Scheduled Protocol
Procedure Protocol Code Sequence (0040,0008) if
Code Sequence provided by Modality Worklist.
Page 25 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
(b)
Comments on (0040,0280) Populated with Comments on the
the Performed Scheduled Procedure Step
Procedure Step (0040,0400) if provided by Modality
Worklist.
(c)
> Region Data (0018,6014) B-Mode, M-Mode = 0001H (Tissue)
Type
Page 26 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
(c)
> Region Flags (0018,6016) 1st Bit (LSB) = 1 (All images
acquired are transparent)
2nd Bit = 1 (All images acquired are
automatically scaled)
The value of the 3rd bit is undefined
4th Bit is Reserved and value is
always 0
(c)
> Region Location (0018,6018)
Min X0
(c)
> Physical Delta (0018,602C)
X
(c)
> Physical Delta (0018,602E)
Y
(b)
The Attribute is only provided if the procedure step is queried
from the MWL server.
(c)
Region Calibration is provided only for 2D (B-Mode) and M-
Mode Regions. Region Calibration is not supported on Ultra-
sound Retired, Secondary Capture images and Screen Cap-
tures.
Page 27 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Table 7 Ultrasound MultiFrame and Ultrasound MultiFrame Retired Image IOD Attributes
Page 28 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
(b)
Series Time (0008,0031) Time the series started.
(b)
Series (0008,103E) Populated with Scheduled
Description Procedure Step Description if a
value was provided by Modality
Worklist.
(b)
Protocol Name (0018,1030) The exam type of the most recent
image stored in a particular series.
If no images are stored for a series
then the value is set to “Ultrasound”.
(b)
Request (0040,0275) Populated with Scheduled
Attributes Procedure Step Sequence
Sequence (0040,0100) if provided by Modality
Worklist.
(b)
> Scheduled (0040,0007) Populated with Scheduled
Procedure Step Procedure Step Description
Description (0040,0007) from Scheduled
Procedure Step Sequence
(0040,0100) if provided by Modality
Worklist.
(b)
> Scheduled (0040,0008) Populated with Scheduled Protocol
Protocol Code Code Sequence (0040,0008) from
Sequence Scheduled Procedure Step
Sequence (0040,0100) if provided
by Modality Worklist.
Page 29 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
(b)
> Requested (0040,1001) Populated with Requested
Procedure ID Procedure ID (0040,1001) from
Scheduled Procedure Step
Sequence (0040,0100) if provided
by Modality Worklist.
(b)
Performed (0040,0244) Date the Performed Procedure Step
Procedure Step was started.
Start Date
(b)
Performed (0040,0245) Time the Performed Procedure Step
Procedure Step was started.
Start Time
(b)
Performed (0040,0253) Populated with Scheduled
Procedure Step ID Procedure Step ID (0040,0009) if
provided by Modality Worklist.
(b)
Performed (0040,0254) Populated with Scheduled
Procedure Step Procedure Step Description
Description (0040,0007) if provided by Modality
Worklist.
(b)
Performed (0040,0260) Populated with Scheduled Protocol
Procedure Protocol Code Sequence (0040,0008) if
Code Sequence provided by Modality Worklist.
(b)
Comments on (0040,0280) Populated with Comments on the
the Performed Scheduled Procedure Step
Procedure Step (0040,0400) if provided by Modality
Worklist.
Page 30 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 31 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
(c)
> Physical Units X (0018,6024) B-Mode = 0003H (cm)
direction M-Mode = 0004H (seconds)
(c)
> Physical Units Y (0018,6026) B-Mode = 0003H (cm)
direction M-Mode = 0003H (cm)
(b)
The Attribute is only provided if the procedure step is queried
from the MWL server.
(c)
Region Calibration is provided only for 2D (B-Mode) and M-
Mode Regions. Region Calibration is not supported on Ultra-
sound Retired images and Screen Captures.
Page 32 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
(b)
Series Time (0008,0031) Time the series started.
(b)
Series (0008,103E) Populated with Scheduled
Description Procedure Step Description if a
value was provided by Modality
Worklist.
(b)
Protocol Name (0018,1030) The exam type of the most recent
image stored in a particular series.
If no images are stored for a series
then the value is set to “Ultrasound”.
Page 33 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
(b)
Request (0040,0275) Populated with Scheduled
Attributes Procedure Step Sequence
Sequence (0040,0100) if provided by Modality
Worklist.
(b)
> Scheduled (0040,0007) Populated with Scheduled
Procedure Step Procedure Step Description
Description (0040,0007) from Scheduled
Procedure Step Sequence
(0040,0100) if provided by Modality
Worklist.
(b)
Performed (0040,0260) Populated with Scheduled Protocol
Procedure Protocol Code Sequence (0040,0008) if
Code Sequence provided by Modality Worklist.
(b)
Comments on (0040,0280) Populated with Comments on the
the Performed Scheduled Procedure Step
Procedure Step (0040,0400) if provided by Modality
Worklist.
Page 34 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
(b)
The Attribute is only provided if the procedure step is queried
from the MWL server.
Error Handling
The following table indicates the response status codes that are handled by
the SONOLINE G20 AE, which an SCP may return following the SCU’s C-
STORE-RSP command.
Page 35 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 36 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
The SONOLINE G20 system will act as an SCU in the ‘Push Model” Stor-
age Commitment SOP Class.
SOP Specific Conformance to Storage Commitment SOP Class
The Storage Commitment AE provides conformance to the following
DICOM Service SOP Class as an SCU at a standard level of conformance.
Page 37 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
The following table provides the list of attributes supported by the AE in the
implementation of Storage Commitment SOP Class:
Table 12 Storage Commitment Request Attributes in N-ACTION REQUEST
The SONOLINE G20 facilitates printing of images as they are being created
or later in review mode.
Preparing image pages during acquisition
The “Print/Store 1” and “Print/Store 2” keys on the control panel can be con-
figured for Print (DICOM B/W Print). When the user presses the configured
key on the control panel, the image is acquired, stored on the hard disk and
placed in a page under the printer layout (DICOM B/W Printer Layout).
Preparing image pages in Review mode
The user can select either individual images from open or closed studies, or
one or more entire closed studies and queue them up for print. A DICOM B/
W Printer button is available in Review screen for this operation. When a
study is selected for print, all single-frame images belonging to the study
will be printed.
Transfer of pages to the Printer
Pages may be immediately transferred or delayed until the end of study
using the transfer configuration. The SONOLINE G20 supports two config-
urations: “Print At End of Exam” and “Print When Page Is Full”.
If the configuration is set to “Print At End of Exam”, all pages queued to des-
tination devices will be transferred as a batch when the user selects “Close
Study” or “New Patient”. If the configuration is set to “Print When Page Is
Full”, a page is transferred to destination devices immediately after it is full.
Page 38 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
For both “Print At End of Exam”, and “Print When Page Is Full” settings,
image transfer will be delayed if the SONOLINE G20 is busy performing
another DICOM Print Command.
Associated Real World Activities
An association is established when the user initiates a “B/W Print” opera-
tion. Individual images or entire exams can be transferred to the selected
DICOM Print device. The association is closed when no pages are avail-
able to be printed for five seconds. An association may also be opened after
a network outage or when the system is powered-on if images are queued
to be printed.
Proposed Presentation Context to a Grayscale Print Server
Table 13 Grayscale Print Presentation Context.
Page 39 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 40 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Table 17 Attributes set for the Basic Film Box SOP Class.
Page 41 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
N-Set M The SCP for each potential image of the film box creates an
image box instance. Only those instances, which actually
contain images, will be updated with the N-SET message.
Table 19 Attributes set for the Basic Grayscale Image Box SOP Class.
Error Handling
The SONOLINE G20 Print AE supports the following error codes and
reports failures to the user.
Table 22 Supported Error Codes for Printer Classes.
If the print operation is not successful, the image(s) are spooled on the
SONOLINE G20 hard drive. A user-configured number of additional
attempts are made to print the image(s). If these attempts fail, the user must
select the job and press “Retry Job” on the Print Status page to complete
the print operation.
Page 42 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
The SONOLINE G20 will always act as an SCU and as the client in a client-
server model.
SOP Specific Conformance to Modality Worklist Service SOP Classes
The Worklist AE provides conformance to the following DICOM Service
SOP Classes as an SCU all at a standard extended level of conformance:
Table 24 Supported SOP Classes
The following table provides the list of attributes requested in the Modality
Worklist Query.
Table 25 Modality Worklist Information Model Attributes
Page 43 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 44 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Only the IHE (refer to IHE Rev 5.5) Simple and Abandoned Cases for the
relationship between Scheduled Procedure Steps and Performed Proce-
dure Steps are supported. Both cases specify that a 1-to-1 relationship
must exist between Scheduled Procedure Step and Performed Procedure
Step. In the Simple Case the Performed Procedure Step is completed suc-
cessfully. In the Abandoned Case the Performed Procedure Step is aban-
doned before being completed.
A list of scheduled procedures and procedure steps will be accessible from
the Worklist and Procedure screens. The Performed Procedure Step User
Interface allows the operator to set the status of the performed procedure
step. The system shall establish an association for N-CREATE and N-SET,
if another N-CREATE or N-SET is available within 5 seconds, it will be sent
using the same association.
Starting a Performed Procedure Step
When the user depresses the ‘OK’ button on the New Patient Screen a per-
formed procedure SOP Class instance will be created using the N-CREATE
DIMSE service for the selected scheduled procedure.
Ending a Performed Procedure Step
When the user selects ‘Completed’ or ‘Discontinued’ from the MPPS User
Interface, the performed procedure step will be closed using the N-SET
DIMSE service.
New Patient Request
If the ‘New Patient’ button is selected and there are opened performed pro-
cedure steps, the user shall be prompted for a closure status for the opened
procedure step by the MPPS User Interface. Any opened procedure steps
must be closed before any ‘new patient’ data can be entered.
System Shutdown
If the user requests ‘System Shutdown’ and there is an open performed pro-
cedure step, the user will be prompted for a closure status for the open pro-
cedure step. All procedure steps should be closed before the system can
be shutdown. Failure to close a procedure step will result in the procedure
step being marked as Discontinued.
Error Handling
If the MPPS operation is not successful, the MPPS command is spooled on
the SONOLINE G20 hard drive. A user-configured number of additional
attempts are made to complete the MPPS Commands. If these attempts
fail, the user must select this job and press “Retry Job” on the Store Status
page to complete the MPPS operation.
Page 45 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
The SONOLINE G20 system will always act as an SCU and be the client in
a client – server model.
SOP Specific Conformance to Modality Performed Procedure Step
SOP Classes
The Modality Performed Procedure Step AE provides a conforming imple-
mentation of the following DICOM Service SOP Class as an SCU at a stan-
dard extended level of conformance.
Table 27 Supported SOP Class
The following tables provide the list of attributes supported by the AE in the
implementation of MPPS SOP Class including N-CREATE, N-SET and
Final State attributes. The SONOLINE G20 sends N-SET only at final state.
Table 28 Modality Performed Procedure Step Attributes in N-CREATE
Page 46 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 47 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 48 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Service SOP Class Name SOP Class UID Transfer Syntax Name Transfer Syntax UID List
Page 49 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Table 32 US Image Attributes Used (Refer to Table 6 on page 23 for additional attributes used)
File Preamble No Tag or Length fields All bytes are set to 00H
DICOM Prefix No Tag or Length fields Set to DICOM Prefix “DICM”
Group length (0002,0000)
File Meta Information Version (0002,0001) Always set to 0001H
Media Storage SOP Class UID (0002,0002) Always Ultrasound Image
1.2.840.10008.5.1.4.1.1.6.1
Media Storage SOP Instance UID (0002,0003)
Transfer Syntax UID (0002,0010) Always Explicit VR Little Endian
1.2.840.10008.1.2.1
Implementation Class UID (0002,0012) Always set to 1.3.12.2.1107.5.5.5
Implementation Version Name (0002,0013) Always set to MergeCOM3_330
The following table denotes the attributes included in the Ultrasound Multi-
Frame Image Object as implemented on the SONOLINE G20 in addition to
the attributes listed in Table 7 on page 28.
Table 33 USMF Image Attributes Used (Refer to Table 7 on page 28 for additional attributes used)
Page 50 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 51 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 52 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
8.0 Configuration
The SONOLINE G20 Networking and DICOM parameters can be config-
ured through the SONOLINE G20 System Presets Menu screens. The fol-
lowing configuration is supported:
General system
Network (local and remote)
DICOM Store
DICOM Print
DICOM Modality Worklist
DICOM Storage Commitment
DICOM Modality Performed Procedure Step
8.2.1 Local
The SONOLINE G20 local network parameters are configurable. The fol-
lowing network parameters can be configured for a SONOLINE G20
device:
Host Name
IP address
Subnet IP mask
Page 53 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Default Gateway
DICOM Application Entity Title
8.2.2 Remote
Multiple DICOM service class providers can be configured through the sys-
tem presets. The following network parameters can be configured for each
remote device:
DICOM Device Application Entity Title
IP address
Port Number
The “Store Image Format” setting controls the Referenced SOP Class UID
Page 54 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Page 55 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Parameter Description
or Landscape:
Display Format You must supply the number of rows and columns of
images on the printed sheet.
For example, a 6 on 1 print with Landscape mode should
have 3 columns and 2 rows:
Page 56 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
Parameter Description
Page 57 of 58 REV 03
Siemens Medical Solutions USA, Inc. Ultrasound Division
9.0 Security
Page 58 of 58 REV 03