Manual For Service Training (Interal Archive)
Manual For Service Training (Interal Archive)
1
CONTENTS
2
4.8 Coarse Collimator…………………………………………………………………..36
4.9 Optical Plummet…………………………………………………………………….37
5. AXIS BLOCKING………………………………………………………………38
6. SOFTWARE DOWNLOADING………………………………………………39
7. TOOLS & CHEMICALS………………………………………………….....40
3
1. ANGLE MEASUREMENT PART
P.S
1 grid 1 division
If the signals are still not right, you may also try below method
(not recommended if the signal is not too serious):
First loosen the small group of screws, and adjust the big group of screws
on the horizontal axis
Loosened: decrease the signal value
Tightened: increase the signal value
F
zero-point voltage
D
E
zero-point signal
C
(Note:in theodolite,
we only test the 3rd and 4th
test points)
1.Press “CH2 MENU” to switch off the CH2 signal. Just use CH1 probe.
2. Firstly, on the CSY-2 machine, put theCH1 probe to “Voset”, and its plier to
“GND”. And connect the white connecting cable of CSY-2 to the analog.
Rotating the telescope of instrument, we can see zero-point signal wave (see
the picture above).
Distance between the horizontal line and the peak point L
should be between 1 grid and 2 grids. If not,
L adjust adjusting screw on the analog W2
properly.
3. Put the probe at test point E on main board (see P4), while rotating the
EDM, and we get the wave of zero-point signal; then put the probe at point F,
we get the zero-point voltage line. So it is with point C and D.
4. Adjust the TR1 or TR2 (see P4) to make the zero-point voltage line in the
position as described above. (for theodolite, 2 divisions above peak line; for
total station, 3 divisions above peak line).
5.Press “MEASURE” on oscilloscope, and check the test points on main
board (see picture on P5) just to see the “Pk-Pk” values and “Mean” values of
each channel of the four analogs (2 for Hr., 2 for V.), are all up to the
requirement. If not, readjust the analogs.
Note:
If the zero-voltage line is too high, E08 will appear;
If the zero-voltage line is too low, OSET will fail, “OSET” will appear on the
screen, or even it will sometimes cause incorrect reading.
(400,500,600 Series)
Series 400,500,600 use angle measurement system of absolute encoder and CCD
sensor, which is different from Series 300 and theodolite that uses rasters.
With absolute encoding circle the machine can be turned on without initialization,
measuring results are more accurate.
Working principle: obtaining the code information from specific position on the circle
by photoelectric detector. Multiplexer will conclude if horizontal disk or vertical diskis
working, then all information will be transmitted to microprocessor CPU to calculate.
CPU will emit instructions to guide the operation then transmit the result to display.
Absolute Encoder Circle code strip
Inspection
1. On the oscilloscope, shift to State “ X-Y, 1.00V, 25μs “(See picture below);
2. Put the probe at the test points C and D (for vertical disk), or A and B (for
horizontal disk);
(Caution: the test points’ locations vary on different series of main board, see
P8 );
3. Cover main board with a black cloth, to prevent light interfering to the
signal.
4. Rotate telescope gently at the meantime.
1V
2.8~3.2V
Peaks
Arch Symmetry
0
upper CCD,Vert. D
lower CCD,Vert. C
the 3rd leg
Series 400
Series 500/600
Adjustment
Fixing screw
Adapt the CCD to a proper position
until the arch wave is up to standard;
② If the arch wave is not good
enough, adjust the four fixing screws
or the two fixing screws on diode, to
make the diode at center of the
bracket.
③ Clean the glass surface of the
CCD and encoder disk with chemical Adjust the position of CCD
solution (alcohol : ether=3:7). Do it
carefully without any scratching.
④ Rotate disk and check the signal
to insure the waves from all part of
the disk should be in good condition.
⑤ Finally, restore those screws and
add some varnish on to fix them.
(Adjustments of both the horizontal
CCDs and vertical CCDs are the
same, except for the different
locations of test points , A,B for
horizontal and C,D for vertical ).
Error 08:
a. The diameter of the Lisharing figure is too big. See P1
b. Adjust the zero-point voltage.(zero-point voltage is too high)See P5 ~ P6
c. Main board is damaged, replace for a good one.
Additional Errors for Theodolite
E20—— I angle is not well set. Press “0SET”+“V%”+“0SET” to make a force
setting.
E21—— Check the compensator (See P32 ~ P33).
Battery
Bad welding in the wires or joints on battery cover
3. EDM PART
Procedures:
1. Enter the test mode: for Series 300, POWER+F1 (5 times)+F4 (5 times).
for Series 400/500, Power→Config→“7”→”4”/”3”→press F1(5 times) and F4
(5 times); for Series 600, Power→Para→System→State.
2. By the implication on the screen, choose State 3, and go to Inside Path.
3. In such state, the signal values should be in accordance with the following
criterion:
TAPE 0
≈2.5V When rotating EDM, SIG value in the test mode should TAPE 1
be within 2.2—2.7V, otherwise, re-adjust the high voltage.
TAPE 3 1.6—2.5V
Inspections:
Inside path, no signal value:
Check emitting board, receiving board, and intermediate frequency board
Inside path, signals exist, but:
Tape 0,1,3 lower than required, problem with intermediate frequency board
Tape 0,1 signals ok, but signal of Tape 3 not good, check the receiving board
If signals of both inside and outside path are right, but hi-voltage is not
adjustable, it is problem with fibers.
After entering test mode, the path changes automatically between inside and
outside. Check the slip ring.
If there is big change in SIG under inside or outside path, the problem is with
slip ring or light reducing sheet.
Notice:
When adjusting the holding screw, better not overturn it (no more than two
rounds), or else it will cause light leaking.
Two ways of diagnosing light leaking:
① Check through the EDM lenses in Inside Path state of the test mode to see
whether there is a red light point. If so, light leaking exists.
② Cover the EDM with a prism. Shake the prism and see whether the SIG
value in the Inside path of test mode changes or not. If it changes, light
leaking exists.
Solution to light leaking:
Readjust the holding screw back to a proper position. If the high voltage is not
within correct range, try to replace light-reducing sheet, see Method 2.
If the high voltage still fails after practice the above two methods, there is
probably something wrong with the intermediate frequency board. Replace for
a new one.
3.3 Circuit Boards of the EDM
3.3.1 Emitting board
①the fourth leg,Power leg, Voltage ≈6.0V,it is connected with main board, slip ring,
and emitting board. If the voltage here is not around 6.0V, then we have to check
with multimeter to see whether it has problem with EDM main board or slip ring.
②Voltage supplied to
emitting tube ≈5V
EDM main board damaged
CH1 500mV
M 500us
CH1 2.00V
M 100us
fixing screws
test point
Check the signal wave while
rotating EDM
See Picture
optical coupler
Standard of wave:
The widths of each wave column CH12.00us M 1.00us
should be approximately the same
If not,
① Loosen the two fixing screws;
② Move the coupler board up or
down to a proper position, and
make the edge of the circle sheet
go across about 1/2 of the
coupler foot. (See Figure below).
③ Check the signal wave. If the
wave is not desirable, repeat the
Step ②.
coupling signal wave
dimmer sheet
Beams Alignment
1) Inspection
A. Put the total station in front of a collimator and adjust it to infinity distance
until you can see the reticle of collimator clearly; (note: there are two
positions of distance in the collimator, the near distance point and the infinity
distance point. the scale of the infinity distance point is a bit smaller.)
B. Adjust an electronic theodolite to infinity distance, too;
C. Make the object lens of the two instruments be opposite, and we use the
telescope of the theodolite to observe the EDM. In order to make image clear,
we’d better put a white paper as background;
D. Turn the total station on, enter distance measurement mode, and focus it
until reticle of total station and red point can be clearly viewed.
E. Observe if the red point and center point of reticle of total station are
overlapped, if overlapped they are ok, otherwise it needs to be adjusted.
The checking procedures for receiving fiber is almost same with checking
emitting fiber, but you need to add an outside light source at the head of
receiving fiber.(See below)Remove screw A, take off the tube base, and add
an outside light source in the small hole.
1
3
4 2
Adjustment:
A. Adjustment of Emitting Fiber
1)Make sure that the machine is in distance measurement mode.
2 ) Loosen the fixing screw 1 on emitting fiber (above figure), move the fiber
higher or lower, make the facula be clearest, then fix the screw 1.
3)Loosen the fixing screws 2 on emitting fiber board, adjust the position of the
bracket of emitting tube, make the center of facula be on the center of reticle
of total stations. Finally fix the screw and add some varnish on them.
B. Adjustment of Receiving Fiber.
1)Add an outside light source on receiving fiber.
2 ) Loosen screw 3, move the fiber to find facula in view field, adjust the
position of fiber, make the facula be clearest, then fix the screw.
3)Loosen screw 4, adjust the bracket of receiving tube, make the facula be in
the center of reticle, finally fix the screws and add some varnish on them.
4)Remove outside light source, install the original receiving board.
GND
CLOCK
DATA
POWER
(Power Leg)
(on emitting board)
Volt.≈6.0V
emitting
board volt. Check CZ9
tube
Emitti
board
≈3.9V P17
Bad
Fib
ken
bro
test point P17
er
abnormal
Bad
ng
no not good
Check frequency
Check board&Emitting
receiving board
board Method: Check
the voltages of
two leads of the
e for a
frequen
suitable
reducin
signals
Adjust
g sheet
Replac
holding
Check
voltage
light-
screw
P14
on
cy
emitting light:
Yes: fiber broken
No: emitting tube broken
or Emitting board bad
Inspection Procedures:
At the moment of powering on instrument, listen if there is voice from motor in
EDM (and voice from motor of light-path transition system).
1. If no voice, add a DC 3V to motor to see if it has been broken. If motors are
good, test followings:
1) Inspect the Power leg on emitting board to connect with slip ring part (see
P22,) (under distance measuring state)to see if it is about 6.0V.
2) Check if slip ring is in shorted circuit, or weak welded.
3) Check the connection with main board and relevant ports.
4) If above is ok, go to next step. Turn the instrument on, under distance
measurement state, look into objective lens, to see if there is red facula, if
not, try to rotate motor of light-path transition system. If still no facula, it
needs to check emitting fiber or emitting board.
5) Inspect whether emitting fiber is broken.
6) Test voltage of the ports on emitting board.
If above are ok, we need to inspect intermediate frequency board to find
error points. See below:
a) Use a multimeter to test the voltage of CZ9 point, the voltage is the
output voltage of emitting board. The value should be ≈2.5V (300series)
or ≈1.6V (400,500,600series), if not, there is problem with emitting board.
b) Use a multimeter to test the voltage of CZ10 point, the point is the testing
point of receiving board. The value should be around 3.9v, if not, inspect small
board and receiving fiber.
c) If above points are ok please inspect the hi-voltage of CZ4 point. The value
should be lower than the recommended voltage for 10~15V
(please find the recommended voltage on the back of EDM cover). If the hi-
voltage is out of the normal range, we need to adjust it as below:
①If there is no big difference between testing voltage and recommended
voltage, we can adjust hi-voltage controlling screw.
Turn on the instrument, enter adjustment state, enter state 3.
(for 300, to change light path, press F1. To change measuring tape, press F4.
0, 1 are low frequency, 3 is high frequency, for 400/500/600, see manuals).
②In adjustment state 3, under inside light path, low frequency state, loosen the
fixing screw, then adjust hi-voltage control screw.:
When inside light path signal become lesser, restart the total station, enter test
mode again. Now high-voltage controlling value and high voltage value will be
larger.
When inside light path signal become larger, restart the total station, enter test
mode again. Now high-control value and high voltage value will be lower.
(High-voltage control screw can be turned in 2 rounds, if hi-voltage can not be
well adjusted by this screw, it needs to change light-reducing piece in inside
light path).
To change light-reducing piece:
a. Take zero-fixing board down, turn inside light path out, take off its light-
reducing piece;
b. Change for a suitable light-reducing piece, paste on with varnish, fix zero-
fixing board on the bracket. Turn on the total station, check the high-voltage. If
still not ok, repeat the adjustment steps of high-voltage.
(When high-voltage is lower than recommended voltage, use a darker light-
reducing piece; When high-voltage is higher than suggested voltage, use a
lighter light-reducing piece.)
7) After fixing above, if the problem still appears, the middle-frequency board
should be changed.
8) If above are ok, we enter Adjustment State 2 to check Optical Coupling
board.
.9) Checking Optical Coupling Wave.
Set oscilloscope to be 0.1V state, place probe on the middle point on the
optical coupling board to watch signal of this point (meanwhile rotating the light
reducing motor gear). (See P19)
Gear value Checking: Under adjustment state 2, put a prism onto Objective
lens, the correct gear value should be about 11.
If there is big error, adjust it as below:
Loosen 2 fixing screws, move the optical coupling board upward or downward,
make the Gear value be correct.
10) When above is ok, enter adjustment state 3.
a) Check the CZ5 point on intermediate frequency board. In inside light path,
set oscilloscope to be 0.1V state, 10db attenuation, put probe on CZ5.
Observe the signal. Under low frequency, the signal should be 2.2~2.7V.
Under high frequency, the signal should be 1.6~2.5V. Middle-frequency
signals should be sine wave(See P17).
11) Optical and electronic crosstalk: Under outside light path, set oscilloscope
to be 5mV, 10db attenuation, and place probe on the CZ6 point. Under low
frequency, Vp-p<100mV; under high frequency Vp-p is a little higher.;
12) Use a multimeter to test CZ7 point on middle-frequency board. The point is
High Voltage controlling point, normal value should be 1—3V.
13) Check the stability of inside light path(State 3, inside light path).
①Under inside light path, low-frequency signal is about 2.5v, turn EDM a round,
the signal should change between 2.2—2.7V, if it is not in this range, adjust
high-voltage controlling screw.
②Under inside light path, hi frequency signal is about 1.7v, turn EDM a round,
the signal should change between 1.6—2.5V.
If it is not stable, adjust it as above.
③Under inside light path, high frequency, put a prism onto objective lens,
shake the prism. If signals voltage change, it needs to adjust high-voltage
controlling screw. If no changes, it is good.
④After fixing all of screws, then check again high-voltage, high-voltage control,
intermediate frequency signals.
2. Exaggerated reading
①Without rotating EDM (telescope), reading is correct; while EDM
(telescope) being rotated, exaggerated reading occurs, it is the problem with
slip ring;
②Without rotating EDM, exaggerated reading appears, then something
wrong with intermediate frequency board, receiving board or EDM main
board.
③Prism system gets dirty or moldy, clean it or change it. After being re-
assembled, we have to readjust all the calibrations.
10. Inside path signal is zero, and the voltage of high-voltage controlling is 0
Communication between EDM and main board is bad.
Slip ring fault: short circuit or open circuit
Solution: replace slip ring
11. High-voltage too high, and inside path signal 4.9V, distance measuring
fails
Initiative response of high-voltage is too slow, and high-voltage controlling
transformer is bad.
Solution: replace intermediate-frequency board
4. CALIBRATIONS
Adjustment:
Take off reticle cover, and adjust the left and right screw properly.
(Note: Don’t adjust the screws beyond 1′; otherwise it will affect the beam
alignment)
For example, the “a” value on the reversed position is 180º00'40" (in which
40" is just the difference). Adjust the horizontal clamp and tangent screw to
make the horizontal angle value to half of the difference as 180º00'20". Then
sight at the horizontal collimator, and see how much the deviation is from
Point 10. Next we take off the reticle cover on the eyepiece and adjust the left
and right adjusting screws to make the vertical reticle line back to the Point
10. Finish.
Inspection:
First Position Second Position
1. On first position (main board
side is on your left when
sighting), sight at the center of
crosshair in horizontal
collimator.
2. Sight in the lower collimator,
note down the corresponding
grid value A on the horizontal
reticle crosshair.
3. On second position (main Hor. collimator Hor. collimator
board is on your right when
sighting), do as the same with A B
above two procedures, and
note down the other
corresponding grid value B.
4. The difference between A and
B, |A-B|, is just the high-low
difference value, which should
be less than 9″ (0.3 grid on
the scale).
(Note: on the scale of reticle,
1 grid=30″) Lower collimator Lower collimator
Adjustment:
Take off the center cover below EDM.
Tighten the left or right screw under center cover as per which side is higher.
If the above procedure is not satisfactory, we have to disassemble the main
body and scratch the side which is higher than the other.
Collimator
B C
Leveling Screws A, B, C
500 Series
Setting Coefficient of Horizontal Compensator
1. Level the instrument precisely. Do the Vertical Index adjustment (high-low
difference, see P31).Then go to compensator mode (press “Esc”, “CONFIG”,
“7”, “1”) and adjust the compensator value within the range of ±30 (take the
battery back cover off the instrument and tap two sides of the compensator).
2. Initially, in compensator mode (electronic bubble mode) we input the
coefficient as 4.00, and remain the temperature as 0.20 (temperature
coefficient). Press [ENT] twice, then [SFT] + [.], take a look at "Y" on the
display, note down the value θ (0).
3. Quit to Angle Measuring. Sight the instrument at a certain point E(point 10
for example) in the horizontal collimator, then 0set the horizontal angle. After
that, sight the instrument in the lower collimator, and note down the point F on
the vertical cross hair. Press [SFT] + [.].
4. Enter the compensator mode (electronic bubble mode) again, and turn the
leveling hand wheel B and C to make instrument tilt left-side until the number
in Y of electronic bubble reach +1'40". (For right-side it should be -1'40")
5. Rotate the telescope and turn the horizontal tangent screw. Sight the EDM
at the point E on the vertical cross hair in the horizontal collimator. Then press
[ESC]. Fix the horizontal tangent screw. And sight the telescope vertical cross
hair in the lower collimator. 0set the horizontal angle. Next, sight the
instrument at point F on the vertical cross hair in the lower collimator. Here,
note down the horizontal angle θ(h) and vertical angle θ(v).
6. Finally, we are able to calculate the Compensator's coefficient with the
formula:
K1= 4×(1'40"-θ0)×0.577/θH 0< θH <90º
or
K1=4× (1'40"-θ0)×0.577/(360º-θH ) 270º<θH<360º
Note: θ(v) is the angle between horizontal collimator and the lower collimator.
Usually, we take tgθ(v)=0.577, for the angle of the two collimator scopes we
used is always 30°. So this angle value depends on the collimators you used.
7. Level the instrument again. And follow the steps above. Tilt the instrument
right-side to -1'40", so that we attain the other coefficient K2.
8. Lastly, the coefficient of horizontal compensator is calculated as:
K= (K1+K2)/2
9. At the end, we go to the compensator mode, press “OK” and input the K
value in place of the initial coefficient 4.00.
600 Series
Setting Coefficient of Horizontal Compensator
1. Level the instrument precisely in all directions. Do the Vertical Index
adjustment. Then go to the compensator mode (PARA→System→HADJ) and
adjust the compensator value within the range of ±30(take the battery back
cover off the instrument and tap two sides of the compensator). Initially, we
input the coefficient as 4.00(press OK and input 4.00), and keep the
temperature coefficient as 0.20(the lower row). Note down the value θ0 at Y of
the horizontal angle in the electronic bubble mode(F2→F6→F5). Quit to Angle
Measuring.
2. Sight the instrument at a certain point E (10, for example) on horizontal line
of the scale line in the upper collimator, then 0set the horizontal angle. After
that, sight the instrument in the lower collimator and remember the
corresponding point F on the horizontal scale line.
3.Enter the electronic bubble mode (F2→F6→F5), turn the screw B and C to
make instrument tilt left-side until the value at Y of the electronic bubble
screen reaches +1'40".
4. Rotate the telescope and turn the horizontal tangent screw, and sight at the
point E on the horizontal scale line in the upper collimator. Fix the horizontal
tangent screw. Then sight the telescope at the horizontal scale line in the
lower collimator. 0set the horizontal angle. Next sight the instrument at point F
on the scale line in the lower collimator. Here, note down the horizontal angle
θH in the screen.
Then, we can calculate the compensator's coefficient as follows:
K1= 4×(1'40"-θ0)×0.577/θH 0< θH <90º or
K1=4× (1'40"-θ0)×0.577/(360º-θH ) 270º<θH<360º
5. Level the instrument again and follow the steps above. Tilt the instrument
right side to -1'40" in the same way. We can get the other coefficient K2.
Lastly, the coefficient of horizontal compensator is calculated as:
K= (K1+K2)/2.
6. At the end, we go to the compensator mode (PARA->System->HADJ),
press “OK” and input the K value in place of the initial coefficient 4.00.
Checking the Accuracy of the Coefficient
1. Level the instrument precisely in all directions.
2. Sight the instrument at a reference point N (10 for example)on the
horizontal scale line of the lower collimator. Lock the horizontal fixing
screw(e.g. point 10). Turn the EDM and sight at the upper collimator. Here,
0set the horizontal angle value, remember the corresponding position point M.
2. Enter the compensator mode. Turn the hand wheel B and C to make the
instrument tilt left around +2' (see "Y" value on the display), then sight the
EDM at point M again. 0set the horizontal angle value. Turn the EDM and
sight at point N in the lower collimator, and then look at the horizontal value
"HAR", which is just the compensating error of compensator. This error should
be within
3. Level the instrument; tilt it on the other side (right side), by turning the
leveling hand wheels B and C as the steps above. And the compensating
error of the horizontal compensator on the right side can be attained. And we
should make sure that both errors should be within the range of 6". Otherwise,
repeat the steps of calculating of the compensator coefficient and re-input the
coefficient until the compensating errors are within the range of 6".
4.4 I Angle
(Vertical Angle Index)
Adjustment:
1. Go to the item “V ANGLE 0 POIN”
(for Series 300, Press F1 +
Power, and then F1; for Series
400/500,
Power→Esc→Config→“2” → “1”
for 600 Series, Power→ADJ→V
ANGLE 0 POINT). Then 0set;
2. On first position (main board side
is on your left when sighting), aim at
a point (10, for example) on the
vertical crosshair of the scale in
horizontal collimator, and press
“Enter”;
3. Reverse the instrument to second position (main board side is on your right
when sighting), and sight at the same point, and press ENTER;
At this moment, if the I Angle error is too much deviation, “ERROR” will be
displayed. At this time press keys as follows for force alignment:
Series 300, press F3→F2→F2;
Series 400, press F1→F2→F2;
Series 500, press F1→F2→F2;
Series 600, press F1→F2→F2.
Theodolite, press “0SET”+“V%”+“0SET”
4. Power off to save the force setting.
5. AXIS BLOCKING
Adjustment
Note: Once the blocking occurs, do not force turning the instrument.
Otherwise it will cause a greater damage to the axis.
1.Vertical Axis Blocking:
axis sheath
1). Disassemble the instrument.
2). Keep the axis vertical, and separate
the axis shaft and the axis sheath, then axis shaft
remove those balls bearing carefully.
3). Clean the oil on the surface of both
axis shafting and axis sheath.
4). Add some green grease (Cr2O5) on the grinding stick. And then grind the
axis sheath with the stick uniformly. Be aware that you should grind in one
direction, clockwise or counter-clockwise avoiding damage to the axis.
5). Grinding finished, clean the grease on the surface with a clean cloth.
6). Add some oil to the place grinded.
7). Clean the surface of the encoder disk with chemical solution (ethanol :
aether=3:7).
8). Mount the balls bearing back. And reassemble the vertical axis.
If the tolerance between axis shafting and axis sheath still feel tight, repeat
the procedures above.
6. SOFTWARE DOWNLOADING
Series 300 jumper
CSY-2 Box
Oscilloscop
e
Multimeter
Tweezer
Extender
Screwdrivers