Chapter Two-Instrument Types and Performance Characterstics
Chapter Two-Instrument Types and Performance Characterstics
for
Mechanical Engineering students
CHAPTER–TWO
An example of a passive instrument is the pressure-measuring device shown in Figure 2.1. The pressure of the fluid is translated
into a movement of a pointer against a scale. The energy expended in moving the pointer is derived entirely from the change in
pressure measured: there are no other energy inputs to the system.
An example of an active instrument is a float-type petrol tank level indicator. Here, the
change in petrol level moves a potentiometer arm, and the output signal consists of a
proportion of the external voltage source applied across the two ends of the potentiometer.
The energy in the output signal comes from the external power source: the primary
transducer float system is merely modulating the value of the voltage from this external
power source. Fig. 2.1 Passive pressure gauge.
In active instruments, the external power source is usually in electrical form, but in some cases, it can be other forms of energy
such as a pneumatic or hydraulic one.
One very important difference between active and passive instruments is the level of measurement resolution that can be
obtained. With the simple pressure gauge shown, the amount of movement made by the pointer for a particular pressure change
is closely defined by the nature of the instrument. Whilst it is possible to increase measurement resolution by making the
pointer longer, such that the pointer tip moves through a longer arc, the scope for such improvement is clearly restricted by the
practical limit of how long the pointer can conveniently be. In an active instrument, however, adjustment of the magnitude of
the external energy input allows much greater control over measurement resolution. Whilst the scope for improving
measurement resolution is much greater incidentally, it is not infinite because of limitations placed on the magnitude of the
external energy input, in consideration of heating effects and for safety reasons.
Deflection-type Null-type
Analogue and digital instruments
An analogue instrument gives an output that varies continuously as the quantity being measured changes. The output can have
an infinite number of values within the range that the instrument is designed to measure. The deflection-type of pressure gauge
described earlier in this chapter (Figure 2.1) is a good example of an analogue instrument. As the input value changes, the
pointer moves with a smooth continuous motion. Whilst the pointer can therefore be in an infinite number of positions within
its range of movement, the number of different positions that the eye can discriminate between is strictly limited, this
discrimination being dependent upon how large the scale is and how finely it is divided.
A digital instrument has an output that varies in discrete steps and so can only have a finite number of values. The rev counter
sketched in Figure 2.4 is an example of a digital instrument.
The final way in which instruments can be divided is between those that merely give an audio or visual indication of the
magnitude of the physical quantity measured and those that give an output in the form of a measurement signal whose magnitude
is proportional to the measured quantity.
The class of indicating instruments normally includes all null-type instruments and most passive ones. Indicators can also be
further divided into those that have an analogue output and those that have a digital display. A common analogue indicator
is the liquid-in-glass thermometer. Another common indicating device, which exists in both analogue and digital forms, is the
bathroom scale.
Instruments that have a signal-type output are commonly used as part of automatic control systems. In other circumstances,
they can also be found in measurement systems where the output measurement signal is recorded in some way for later use.
Smart and non-smart instruments
Smart instruments
Can provide additional data, such as diagnostics, in addition to their primary purpose. They can detect faults in
installation or problems with the application. Smart instruments can also be connected to site networks using
protocols like Ethernet/IP.
Non-smart instruments
Also known as dumb instruments, they can't perceive any process information other than input from their direct
sensors.
2.2 Static characteristics of instruments
2.2.1 Accuracy and inaccuracy (measurement uncertainty)
2.2.5 Linearity
It is normally desirable that the output reading of an
instrument is linearly proportional to the quantity being
measured. The Xs marked on Figure 2.6 show a plot of the
typical output readings of an instrument when a sequence of
input quantities are applied to it. Normal procedure is to draw
a good fit straight line through the Xs, as shown in Figure 2.6.
The non-linearity is then defined as the maximum deviation of
any of the output readings marked X from this straight line.
Non-linearity is usually expressed as a percentage of full-scale
reading.
2.2.6 Sensitivity of measurement
The sensitivity of measurement is a measure of the change in instrument output that occurs when the quantity being
measured changes by a given amount. Thus, sensitivity is the ratio:
The sensitivity of measurement is therefore the slope of the straight line drawn
on Figure 2.6. If, for example, a pressure of 2 bar produces a deflection of 10
degrees in a pressure transducer, the sensitivity of the instrument is 5
degrees/bar (assuming that the deflection is zero with zero pressure applied).
Example 2.1: The following resistance values of a platinum resistance thermometer were measured at a range of
temperatures. Determine the measurement sensitivity of the instrument in ohms/°C.
Zero drift or bias describes the effect where the zero reading of an instrument is modified by a change in ambient
conditions. This causes a constant error that exists over the full range of measurement of the instrument. Zero drift is
normally removable by calibration.
Sensitivity drift (also known as scale factor drift) defines the amount by which an instrument’s sensitivity of
measurement varies as ambient conditions change. It is quantified by sensitivity drift coefficients that define how
much drift there is for a unit change in each environmental parameter that the instrument characteristics are sensitive
to.
Example 2.2: A spring balance is calibrated in an environment at a temperature of 20°C and has the following
deflection/load characteristic.
2.2.10 Hysteresis effects
In any linear, time-invariant measuring system, the following general relation can be written between input and
output for time (t) > 0:
Where 𝑞𝑖 is the measured quantity, 𝑞𝑜 is the output reading and 𝑎𝑜 . . . 𝑎𝑛 , 𝑏𝑜 . . . 𝑏𝑚 are constants.
If we limit consideration to that of step changes in the measured quantity only, then equation (2.1) reduces to:
Any instrument that behaves according to equation (2.3) is said to be of zero order type. A potentiometer, which
measures motion, is a good example of such an instrument, where the output voltage changes instantaneously as the
slider is displaced along the potentiometer track.
It is convenient to re-express the variables 𝑎𝑜 , 𝑎1 , 𝑎2 and 𝑏𝑜 in equation (2.8) in terms of three parameters 𝐾 (static
sensitivity), 𝜔 (undamped natural frequency) and 𝜉 (damping ratio), where: