Different Measuring Instruments
Different Measuring Instruments
MODULE: 1
SYLLABUS
Instrument
The figure shows half wave rectifier based instrument. Pure sinusoidal signal
(whose rms value is to be measured) having peak value V p is applied at the input
terminals. The half wave instrument consists of an ideal diode, a multiplier
resistance RS and a PMMC meter having resistance Rm. The series connected
multiplier resistor limits the current flow through the PMMC instrument. When
the input sinusoidal signal is positive, diode conducts and current will flow
through the meter. Diode is reverse biased for negative input of sinusoidal signal
and does not conduct. The resulting current waveform is a discontinuous series of
positive half cycles. The meter deflection is proportional to the average current
which 0.318 times peak current is. But the actual current (or voltage) to be
indicated in ac measurements is normally the rms quantity which is 1.57(form
factor of half-wave rectifier is 1.57) times the average value. Since there are direct
relationships between rms, peak and average values, the meter scale can be
calibrated to indicate rms voltage. The average current of half wave rectifier is
given by Iav= =0.318xIp Where Ip is the peak value of current.
.
or Iav=
The AC sensitivity of half wave rectifier is Sac=0.45 ⁄ I where Ifsd is the full scale
deflection current and the average value of voltage Vdc=0.45xVrms
Rectifier type of instrument is used only for pure sinusoidal waveform. The
voltage vs. current characteristic of a diode is nonlinear at low value of current. To
keep the meter scale linear the diode should operate in the linear region that is
for relatively high value of current. The lower part of the ac scale of a low range
rectifier type voltmeter is often crowded, separate low voltage calibrated scale is
provided for this purpose. The effect of low current can be countered by
connecting a shunt resistor Rsh parallel with the meter. The following drawbacks
exist in the circuit.
a) In the negative half cycle of the input , reverse current flows in the circuit ,
thereby reducing the overall average value of current. The meter therefore
gives low reading.
b) In the negative half cycle, the diode may breakdown due to a large reverse
voltage if it is not selected for appropriate rating.
To overcome the problems discussed above, in practical half wave rectifier
voltmeter, another diode D2 is used which bypasses the negative half cycle
current . The figure shows a practical half wave voltmeter circuit. In the positive
half cycle diode D1 is forward biased and conducts. Thus average current
corresponding to input voltage passes through the meter. In the negative half
cycle of input, diode D2 is forward biased and diode D1 is reversed biased .Then
current passes through diode D2 and multiplier resistance Rs to the input. Thus
no current will flow through the meter.
To overcome the effect of nonlinearity and increase average current through the
diode D1, a shunt resistance is connected in parallel with the meter. If the value
of meter resistance is equal to the shunt resistance that is R sh=Rm , the average
current through the diodeD1 becomes double that is 2Iav .Then the value of
multiplying resistor becomes
The series connected multiplier resistor limits the current flow through the PMMC
instrument. The meter resistance of PMMC is denoted by R m in the circuit. When
the input sinusoidal signal is positive, diodes D3 and D4 conduct causing current
to flow through the meter from top to bottom. When the sinusoidal input goes
negative , diodes D1 and D2 conduct and current again flow through the meter
from top to bottom i.e from positive to negative terminal of the meter. The
resulting current waveform is a continuous series of positive half cycles. The
meter deflection is proportional to the average current which is 0.636 times peak
current Ip. But the actual current(or voltage) to be indicated in AC measurements
is normally the rms quantity which is 0.707 of the peak value or 1.11( form factor
of full wave rectifier is 1.11) times the average value. Since there are direct
relationships between rms ,peak and average values , the meter scale can be
calibrated to indicated rms voltage. The average current Iav of full wave rectifier is
given by
Iav= =0.636Ip
or Iav=0.9xVrms/(Rs+Rm)
The AC sensitivity of fullwave rectifier is Sac=0.9 ⁄ I where Ifsd is the full scale
deflection average current and the average value of voltage becomes Vav=0.9xVrms
Rectifier type of instrument is used only for pure sinusoidal waveform and the
instrument has a linear scale for higher value of current. At low value of forward
current , the ac scale of a low range voltmeter is often crowded, separate low
voltage calibrated scale is provided for this purpose. The effect can be countered
by using a shunt resistance.
The indication of true rms voltmeter is produced by sensing the heating power
which is proportional to the square of the rms value of the input ac signal to
be measured.
The unknown ac input voltage is amplified and applied to the heating element
of the measuring thermocouple. The application of heat produces an output
voltage that upsets the balance of the bridge. The unbalance bridge is amplified
by the dc amplifier and fed back to the heating element of the balancing
thermocouple, so that the voltage outputs of both thermocouples are the same.
The true rms value is measured independently of the waveform of the ac signal,
provided that the peak excursions of the waveform do not exceed the dynamic
range of the ac amplifier.
This instrument can be used for indicating the AC signal of wide frequency
range. True rms voltmeter has high accuracy and sensitivity with low power
consumption. Most true rms instrument can measure voltage of 100µV to
300V within a frequency range of 10Hz to 10MHz .
Limitation
In this time interval between the start and stop pulse, the gate
opens and the oscillator circuit drives the counter. The magnitude
of the count indicates the magnitude of the input voltage and the
counted value is displayed . Therefore the voltage is converted into
time and the time count represents the magnitude of voltage.
The ramp circuit is easy to design and its cost is low. The single
ramp required excellent characteristics regarding linearity of the
ramp and time measurement. Large errors are possible when noise
is superimposed on the input signal. Input filters are required with
this type of converter.
1.3 DUAL SLOPE INTEGRATING TYPE DVM
e0=- ∫ 𝑒 𝑑𝑡 = - 𝑡 ……(1)
e0 = ∫ −𝑒 𝑑𝑡 = - 𝑡 …..(2)
therefore 𝑡 = 𝑡 or ei=ert2/t1
ei=ern2T/(n1T)= ern2/n1…..(3)
The dual slope technique has excellent noise rejection because the
integration process averages out the noise, provided the time period
of noise is greater than the integrating period t1 and has good
accuracy and resolution, also an accuracy of + 0.05% in 100ms is
available.
1.4 Q METER-APPLICATION AND ERROR
BASIC Q –METER CIRCUIT
𝑋 = 𝑋
𝐸 = IX =IX
𝐸 = 𝐼𝑅
𝐼 =Circuit current
𝑄= = =
therefore Ec=EQ
The Q reading on the output meter must be multiplied by the index setting of the
"Multiply Q by" to obtain actual Q value.
Measurements method:-
There are three methods for connecting unknown components to the test
terminals of a Q meter: direct, series and parallel. The type of component and its
size determine the method of connection.
Direct connection
Here E is a voltage source with a small internal resistance. The voltage E across
the shunt is measured with a meter marked "Multiply Q by". The voltage across
the variable capacitor, corresponding to E ( in figure 2 below,) is measured with
an electronic voltmeter whose scale is calibrated directly in Q values.
Figure 2
The effective Q(actual value) of the reactive element will be somewhat greater
than the indicated Q. The difference can generally be neglected, except in certain
cases where the resistance of the coil is relatively small in comparison with the
value of the insertion resistor.
The inductance L of the coil can be calculated from the known values of frequency
(f) and resonating capacitance (C)
EC=EQ
Series connection
Low impedance components, such as low value resistors small coils and large
capacitors are measured in series with the measuring circuit.
Figure 3
𝑋 =𝑋 or, = 𝜔𝐿 …..(1)
ω
Q1 = XL/R = XC1/R
Q = = …….(2)
For the second measurement, the react nce of the unknown can be expressed in
terms of the new value of the tuning capacitor(C ) and the in-circuit value of the
inductor(L).
𝑋 =𝑋 -X or, 𝑋 = - ……(3)
ω
So that X = …..(4)
ω
R1 = and R =
Also,
R =R −R = −
ω ω
So that,
R = …..(5)
ω
If the unknown parameter is purely resistive, the setting of the tuning capacitor
would not have changed in the measuring process, and C1 = C2.The equation for
resistance reduces to
R = = …..(6)
𝐿 = ……(7)
𝑄 =
and
( )( )
𝑄 = ……(8)
And,
𝐶 = …..(9)
Parallel connection.
High impedance components such as high value resistors certain inductors and
small capacitors are measured by connecting them in parallel with the measuring
circuit. Figure 4 shows the connections. Before the unknown is connected the
circuit is resonated by using a suitable work coil to establish reference values for
Q & C .Let the value of quality factor for resonating condition is Q 1 and the
corresponding value of resonating capacitor is C1 . When the component under
test is connected in series to the circuit the capacitor is readjusted for resonance
and a new value for the tuning capacitance (C2) is obtained and a change in the
value of circuit Q(𝝙Q) from Q1 and to Q2..
Figure 4
𝜔𝐿 = ….(1)
and, 𝑄 = = ….(2)
When the unknown impedance is not connected into the circuit and the capacitor
is tuned for resonance, the reactance of the working coil(XL) equals the parallel
reactances of the tuning capacitor(XC2) and the unknown (X P).Therefore
( )
X =
𝐿 = ….(4)
( )
If the unknown is capacitive,XP =1/𝞈CP and eq (3) yields the value of the unknown
capacitor:
C =C −C ….(5)
R T = Q 2 XL
Or by substitution of Eq. (1 ),
Q2
RT = Q2 XC1 = ….(6)
ωC1
The resistance (RP) of the unknown impedance is most easily found by computing
the conductances in the circuit of Fig 4 . Let,
Then,
GT = GP + GL or, GP = GT − GL ….(7)
From Eq 6
1 ωC1
GT = =
RT Q2
Therefore ,
1 ωC1 R
= −
RP Q2 R2 ω 2 L2
ωC1 1 1
= - ( )( )
Q2 R 1 ω2 L2 /R2
ωC1 1
= − 2
Q2 1
The Q of the unknown is then found by using Eqs. (3) and (8) so that
RP (C1 C2 )(Q1 Q2 ) (C1 C2 )(Q1 Q2 )
QP = = = ….(9)
XP C1 (Q1 Q2 ) C1 ΔQ
Sources of error:-
Probably the most important factor affecting measurement accuracy and the
most often overlooked, is the distributed capacitance for self- capacitance of the
measuring circuit. The presence of distributed capacitance in a coil modifies the
actual or effective Q and inductance of the coil. At the frequency at which the self
capacitance and inductance of the coil resonant ,the circuit exhibits a purely
resistive impedance. This characteristic may be used for measuring the
distributed capacitance.
One simple method of finding the distributed capacitance (Cd ) of a coil involves
making two measurements at different frequencies. The coil under test is
connected directly to the test terminals of the Q meter, as shown in the circuit of
figure. The tuning capacitor is set to a high value, preferably to its maximum
position, and the circuit is resonated by adjusting the oscillator frequency.
Resonance is indicated by maximum deflection on the “Circuit Q” meter. The
values of the tuning capacitors (C1) and the oscillator frequency (f1) and noted.
The frequency is then increased to twice its original value (f2 = 2f1 ) and the circuit
is returned by adjusting the resonating capacitor ( C2).
𝑓 = …….(10)
√
At the initial resonance condition ,the capacitance of the circuit equals C 1 +Cd ,
𝑓 = ……..(11)
√ ( )
After the oscillator and the tuning capacitor are adjusted, the capacitance of the
circuit id C2 +Cd , and the resonant frequency equals
1
f2 = … … . . (12)
2π L(C2 Cd )