Installation and Upgrade Instructions of Mindray LabXpert Software V2.0 - EN
Installation and Upgrade Instructions of Mindray LabXpert Software V2.0 - EN
0)
Reference
requirement.
MRSZ/R01N01.04.02
Confidentiality: Confidential Z-110-007174-00-01-2 .0
Revision Record
Vers Software Revised Effective
Description of Modification
ion version by Date
1. 0 / Mu 2021-01 - 13
New
Guanyun
Add Chapter 6 and 7: Server Interconnection
2. 0 / Mode and Central Server and Sub Server Liu Ping
Interconnection Mode
Contents
1 Overview ............................................................................................................................................. 5
2 Software Installation .......................................................................................................................... 5
2.1 Environment Configuration of Software Installation ....................................................................... 5
2.1.1 Network Firewall Setting ................................................................................................... 5
2.1.2 User Account Control Settings .......................................................................................... 6
2.2 Installing the Client Software ........................................................................................................... 6
2.3 Installing on the Server................................................................................................................... 10
3 Running ............................................................................................................................................. 10
3.1 Running Client Software on the Client PC..................................................................................... 10
4 Connection Setup............................................................................................................................... 11
4.1 Logging in with the Administrator or Service Level Account ......................................................... 11
4.2 Establishing a Physical Connection between the labXpert Server and each Analyzer. ...................11
4.3 Instrument Connection Settings ...................................................................................................... 11
4.4 Connecting the Server PC to a Network accessible to LIS. ........................................................... 12
4.5 LIS Connection Settings................................................................................................................. 12
5 Update ............................................................................................................................................... 13
6 Server Interconnection Mode .......................................................................................................... 14
6.1 Overview ........................................................................................................................................ 14
6.2 Installation ...................................................................................................................................... 14
6.3 Interconnection Mode Settings ....................................................................................................... 14
6.3.1 Switching to Server Interconnection Mode ........................................................................ 14
6.3.2 Setting Server Interconnection ........................................................................................... 15
7 Central Server and Sub Server Interconnection Mode ................................................................. 15
7.1 Overview ........................................................................................................................................ 15
7.2 Installation ...................................................................................................................................... 16
7.2.1 Sub Server Installation ....................................................................................................... 16
7.2.2 Central Server Installation .................................................................................................. 16
7.3 Interconnection mode settings ........................................................................................................ 17
7.3.1 Sub Server Interconnection Mode Settings ........................................................................ 17
7.3.2 Central Server Interconnection Mode Settings ................................................................... 17
Appendix A Figures and Tables ....................................................................................................................... 21
Note: The computer will be restarted automatically after the run-time library is
installed. Save the related data before the installation.
After the run-time library is installed, application programs will be installed automatically:
Server B
Server A
Server C
Figure 23 Server Interconnection
6.2 Installation
For details about the installation procedure, see the client software installation procedure.
6.3 Interconnection Mode Settings
1. Use the customer service tool to switch to server interconnection mode.
2. Set the interconnection server connection on the labXpert system.
6.3.1 Switching to Server Interconnection Mode
1、 Log in to the service auxiliary tool.
2、 Switch to the “Interconnection type setup” tab.
3、 Select the “server interconnection” option.
4、 Click “Save”. The server program will be restarted to enter the interconnection mode.
Sub server
A
Sub server
C
Figure 26 Central Server connection
7.2 Installation
7.2.1 Sub Server Installation
For details about the installation procedure, see the client software installation procedure.
7.2.2 Central Server Installation
The central server stores sample results of sub servers intensively.
To ensure the response performance, there are several hard disks for PC configuration:
1. An SSD system disk for installing the Windows operating system;
2. An SSD hard disk for storing the central server database;
3. A large-capacity mechanical hard disk for storing sample pictures from the central server;
Suppose the drive letter for the operating system is drive C, the drive letter for the database SSD is drive D, and
the drive letter for the mechanical hard drive is drive E.
When setting up the path, you need to specify the installation disk for database and picture respectively.
Set disk D as the database path and backup path. Set disk E as the picture path.