Installation and Upgrade Instructions of Mindray LabXpert Software V2.0 - EN

The document provides detailed installation and upgrade instructions for Mindray labXpert Software, outlining the necessary environment configurations, software installation steps, and connection setups for both client and server. It also includes information on server interconnection modes, allowing multiple servers to connect and share data. The document is classified as confidential and is intended for use by authorized personnel only.

Uploaded by

Anar Isayev
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
369 views21 pages

Installation and Upgrade Instructions of Mindray LabXpert Software V2.0 - EN

The document provides detailed installation and upgrade instructions for Mindray labXpert Software, outlining the necessary environment configurations, software installation steps, and connection setups for both client and server. It also includes information on server interconnection modes, allowing multiple servers to connect and share data. The document is classified as confidential and is intended for use by authorized personnel only.

Uploaded by

Anar Isayev
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 21

QF502.01(2.

0)

File Type Technical Document Confidentiality Confidential

Document No. Z-110-007174-00 -01 Process Code 3214


Print
Color: Black and white
Requirements

Installation and Upgrade Instructions of


Mindray labXpert Software

Reference

Document No. Description

Issued by:(follow the two rules below:)


1. The right column must be filled in for fixture,
clamp, general, and reagent documents.
Hematology Product Manufacture Department
2. The column must be filled in for secret and top
secret documents to be issued. Documents with other
confidentiality can be delivered subject to this

Classified document, intellectual property of Mindray. Page of121


Page of 221

requirement.

MRSZ/R01N01.04.02
Confidentiality: Confidential Z-110-007174-00-01-2 .0

Revision Record
Vers Software Revised Effective
Description of Modification
ion version by Date

1. 0 / Mu 2021-01 - 13
New
Guanyun
Add Chapter 6 and 7: Server Interconnection
2. 0 / Mode and Central Server and Sub Server Liu Ping
Interconnection Mode

Classified document, intellectual property of Mindray. Page of 3


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Contents
1 Overview ............................................................................................................................................. 5
2 Software Installation .......................................................................................................................... 5
2.1 Environment Configuration of Software Installation ....................................................................... 5
2.1.1 Network Firewall Setting ................................................................................................... 5
2.1.2 User Account Control Settings .......................................................................................... 6
2.2 Installing the Client Software ........................................................................................................... 6
2.3 Installing on the Server................................................................................................................... 10
3 Running ............................................................................................................................................. 10
3.1 Running Client Software on the Client PC..................................................................................... 10
4 Connection Setup............................................................................................................................... 11
4.1 Logging in with the Administrator or Service Level Account ......................................................... 11
4.2 Establishing a Physical Connection between the labXpert Server and each Analyzer. ...................11
4.3 Instrument Connection Settings ...................................................................................................... 11
4.4 Connecting the Server PC to a Network accessible to LIS. ........................................................... 12
4.5 LIS Connection Settings................................................................................................................. 12
5 Update ............................................................................................................................................... 13
6 Server Interconnection Mode .......................................................................................................... 14
6.1 Overview ........................................................................................................................................ 14
6.2 Installation ...................................................................................................................................... 14
6.3 Interconnection Mode Settings ....................................................................................................... 14
6.3.1 Switching to Server Interconnection Mode ........................................................................ 14
6.3.2 Setting Server Interconnection ........................................................................................... 15
7 Central Server and Sub Server Interconnection Mode ................................................................. 15
7.1 Overview ........................................................................................................................................ 15
7.2 Installation ...................................................................................................................................... 16
7.2.1 Sub Server Installation ....................................................................................................... 16
7.2.2 Central Server Installation .................................................................................................. 16
7.3 Interconnection mode settings ........................................................................................................ 17
7.3.1 Sub Server Interconnection Mode Settings ........................................................................ 17
7.3.2 Central Server Interconnection Mode Settings ................................................................... 17
Appendix A Figures and Tables ....................................................................................................................... 21

Classified document, intellectual property of Mindray. Page of 4


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Installation and Upgrade Instructions of Mindray labXpert


Software
1 Overview
The labXpert refers to “Mindray labXpert Software”, which is based on the server-multiple client structure and
supports multiple clients to view server data at the same time.
LabXpert supports the connection to cell morphology analyzer (hereinafter referred to as the “analyzer”) and
LIS.
The following figure shows the logic structure of the connection among labXpert, analyzers and LIS.

Figure1 Logic Architecture of Connection


The client server, server, analyzer, and LIS use the network connection, and the arrows represent TCP .
2 Software Installation
2.1 Environment Configuration of Software Installation
2.1.1 Network Firewall Setting
Before the installation, turn off the PC firewall. For the WIN10 system, see the following settings. For operating
systems of other versions, refer to the WIN10 settings.

Classified document, intellectual property of Mindray. Page of 5


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Figure 2 Firewall Settings


2.1.2 User Account Control Settings
Enter the User Account Control screen, and set the notification to Never for WIN10 system; other systems can
refer to the WIN10 settings.

Figure 3 User Permission Control Settings Screen


2.2 Installing the Client Software

Classified document, intellectual property of Mindray. Page of 6


Confidentiality: Confidential Z-110-007174-00-01-2 .0
Install the client software on a Manual Validated PC or Smear Validated PC.
Decompress the client package ISO file using WinRAR or other file compressor, run the Setup.exe file, and
select a language:

Figure 4 Selecting Installation Language


The system detects and installs the essential run-time library automatically:

Figure 5 Installing the Run-time Library

Note: The computer will be restarted automatically after the run-time library is
installed. Save the related data before the installation.
After the run-time library is installed, application programs will be installed automatically:

Classified document, intellectual property of Mindray. Page of 7


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Figure 6 Auto Installation


Check the options as needed and click “Next”.

Figure 7 Selecting Custom Functions


Select the path to save the data file, and click “Next”:

Classified document, intellectual property of Mindray. Page of 8


Confidentiality: Confidential Z-110-007174-00-01-2 .0
Figure 8 Saving the Data File

Note: Try to save the data file on a non-system disk.


Select the location to install the program, and click “Next”:

Figure 9 Software Installation Path


After the setup, click “Next” to continue the installation:

Figure 10 Implementing Installation


After the auto installation, click “Finish” to complete the installation.

Classified document, intellectual property of Mindray. Page of 9


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Figure 11 Completing Installation


2.3 Installing on the Server
Install the server software on the server PC. For details about the installation procedure, see the client software
installation procedure.
3 Running
3.1 Running Client Software on the Client PC
Click the icon on the desktop, or choose “labXpert Client” folder > “labXpert Client” from the Start menu to
enter the login screen.

Figure 12 Login Screen


Click “Server” to set the IP address.

Classified document, intellectual property of Mindray. Page of 10


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Figure 13 Setting the Server IP Address


If the client software and server software are installed on the same PC, the default IP address 127.0.0.1 is used.
Otherwise, you need to enter the IP address of the server PC and log in again.
When you do not know the server IP address, check if the server software runs normally, and the network
connection is OK. If so, use the IP Auto Detection function to acquire the server IP address. When the client
software finds the server, it automatically acquires and enters the IP address to the IP address field.
After setting the server IP address, enter the password and click “Login”:

Figure 14 System Login


The default account ID and password are both “admin”. This account has the “administrator” authority. The
customer service account and password are consistent with other products.
4 Connection Setup
4.1 Logging in with the Administrator or Service Level Account
4.2 Establishing a Physical Connection between the labXpert Server and each Analyzer.
Note: If you want to connect it to cell morphology analyzer, connect the network cable to the router of the
analyzer.
4.3 Instrument Connection Settings
Use the administrator or customer service authority to enter the labXpert software settings screen: “System Setup
-> Instrument Connection Settings”, as shown in the following figure.

Classified document, intellectual property of Mindray. Page of 11


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Figure 15 Instrument Connection Settings Screen


1. Connect the cell morphology analyzer and click “Auto Detection” in the figure below. The software
automatically searches for the analyzer in the current network, as shown in Figure 17.
2.

Figure 16 Automatic Detection Screen


After the search, fill in the current instrument name in the name column, save the settings and exit.
2. Click “New Device” to add a cell morphology analyzer, save the settings and exit.
4.4 Connecting the Server PC to a Network accessible to LIS.
4.5 LIS Connection Settings
Enter the labXpert Setup screen with administrator or above permissions: System Setup -> Communication
Setup

Classified document, intellectual property of Mindray. Page of 12


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Figure 17 LIS Setup


Enter the IP address, port number, protocol type, and protocol version of the LIS.
After all connections are set, the corresponding indicators in the labXpert status bar will be on, as shown in the
following figure:

Figure 18 Connection Status


5 Update
1. Close the running server program.
a) Find the labXpert server icon in the lower right corner of the system, as shown in the figure below.

Figure 19 Server Status Icon


b) If the server status icon is not found, double-click the “labXpert server” icon as shown below.

Figure 20 Server Boot Icon


c) Right-click the server status icon and the menu as shown in Figure 21 will appear.

Figure 21 Context Menu of Server Status Icon


d) Click “Stop Service”. After the server icon changes to red, click “Exit”, as shown in the following
figure.

Figure 22 Stopping Service

Classified document, intellectual property of Mindray. Page of 13


Confidentiality: Confidential Z-110-007174-00-01-2 .0

6 Server Interconnection Mode


6.1 Overview
By setting the interconnection mode, the labXpert server can obtain historical samples of patients from other
servers for the calculation of rule delta and the display of historical results.
For each server, set up the connection with other servers. The overall connection is similar to the following
figure:

Server B

Server A

Server C
Figure 23 Server Interconnection
6.2 Installation
For details about the installation procedure, see the client software installation procedure.
6.3 Interconnection Mode Settings
1. Use the customer service tool to switch to server interconnection mode.
2. Set the interconnection server connection on the labXpert system.
6.3.1 Switching to Server Interconnection Mode
1、 Log in to the service auxiliary tool.
2、 Switch to the “Interconnection type setup” tab.
3、 Select the “server interconnection” option.
4、 Click “Save”. The server program will be restarted to enter the interconnection mode.

Classified document, intellectual property of Mindray. Page of 14


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Figure 24 Switch Server Type to Interconnection Server


6.3.2 Setting Server Interconnection
Suppose there are three servers that can be accessed through the network. The IP addresses are 192.168.1.1,
192.168.1.2, and 192.168.1.3 respectively.
1. Log in to one of the servers first. Suppose you log in to server 192.168.1.1.
2. Switch to “System Setup” -> “LabXpert Interconnection Setup”.
3. Click “New”, and enter the server name and IP address.
4. Switch the setup screen and save.
5. When the server is successfully connected, the connection status indicator is green.
6. Log in to another server and repeat the above steps to set interconnection.

Figure 25 Server Interconnection Settings


7 Central Server and Sub Server Interconnection Mode
7.1 Overview
A central server can connect to other laboratory servers as sub servers.
The user can log in to the client connected to the central server to review and view all recent sample results from
sub servers.
On the sub server, the history results of the same patient can be obtained from the central server and used for
rule delta calculation. View the history results in the history result tab.

Classified document, intellectual property of Mindray. Page of 15


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Sub server
A

Central Central Sub server


Client Server B

Sub server
C
Figure 26 Central Server connection
7.2 Installation
7.2.1 Sub Server Installation
For details about the installation procedure, see the client software installation procedure.
7.2.2 Central Server Installation
The central server stores sample results of sub servers intensively.
To ensure the response performance, there are several hard disks for PC configuration:
1. An SSD system disk for installing the Windows operating system;
2. An SSD hard disk for storing the central server database;
3. A large-capacity mechanical hard disk for storing sample pictures from the central server;

Suppose the drive letter for the operating system is drive C, the drive letter for the database SSD is drive D, and
the drive letter for the mechanical hard drive is drive E.
When setting up the path, you need to specify the installation disk for database and picture respectively.
Set disk D as the database path and backup path. Set disk E as the picture path.

Classified document, intellectual property of Mindray. Page of 16


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Figure 27 Central Server Installation Path Settings


After the path is set, the following installation steps are the same as the installation process of the
uninterconnected server.
7.3 Interconnection mode settings
7.3.1 Sub Server Interconnection Mode Settings
The sub server needs to switch to the interconnection mode. Perform the following operations:
1. Log in to the service auxiliary tool.
2. Switch to the “Interconnection type setup” tab.
3. Select “Sub Server”.
Click “Save”. The server program will be restarted to enter the sub server mode.

Figure 28 Switch Server Type to Sub Server


7.3.2 Central Server Interconnection Mode Settings

Classified document, intellectual property of Mindray. Page of 17


Confidentiality: Confidential Z-110-007174-00-01-2 .0
LabXpert Central Server Configuration Requirements
CPU Intel i5 -10500 or higher
Memory 64 G or above
Graphics card/
SSD 256 G(For installing the system)
SSD 1T(For storing database)
HDD 4T(For storing images)
System win10 Enterprise Edition 64 bit
The network interface supports dual Gigabit network adapters
7.3.2.1 Switching the Server Mode.
1、 Log in to the service auxiliary tool.
2、 Switch to the “Interconnection type setup” tab.
3、 Select Central Server.
Click “Save”. The server program will be restarted to enter the central server mode.

Figure 29 Switch Server Type to Central Server


7.3.2.2 Interconnection Settings
Perform this step after switching the sub-server mode and the central server mode.
Suppose the IP address of the central server is 192.168.1.10. There are three sub servers interconnected with
each other via the internet. IP addresses are 192.168.1.1, 192.168.1.2, and 192.168.1.3 respectively.
1. Log in to the central server.
2. Switch to “System Setup” -> “LabXpert Interconnection Setup”.
3. Click “New”, and enter the server name and IP address.
4. Switch the setup screen and save.
5. When the server is successfully connected, the connection status indicator is green.

Classified document, intellectual property of Mindray. Page of 18


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Figure 30 Central server “labXpert conn. Setup” screen


If the sub server has switched the interconnection mode to “sub server”, after successful connection, check the
sub server “LabXpert conn. Setup”, and you can see the central server. At this time, the sub server can obtain the
historical patient results from the central server. The sub server settings are as follows:

Figure 31 Sub server “labXpert conn. Setup” screen

Figure 32 Context Menu of Server Status Icon


a) Click “Stop Service”. After the server icon changes to red, click “Exit”, as shown in the following
figure.

Figure 33 Stopping Service


1. Close the running client program.
2. Decompress the new version installation package. (Select the corresponding installation package for
the client program and server program.)
3. Run setup.exe
4. The program automatically detects whether an earlier version is installed. If so, the program
automatically enters the update mode.
5. For details about the update procedure, see the installation procedure (the installation directory and

Classified document, intellectual property of Mindray. Page of 19


Confidentiality: Confidential Z-110-007174-00-01-2 .0
data path selections will be skipped).
6. After the installation, the upgrade program will automatically update
the user database to the new version. The update procedure may take a
long time. Do not shut down the PC during the update.
7. After upgrading the server program and each client program, start the server program, and then start
each client program to check whether the program runs normally and the data is complete.

Classified document, intellectual property of Mindray. Page of 20


Confidentiality: Confidential Z-110-007174-00-01-2 .0

Appendix A Figures and Tables


Figure1 Logic Architecture of Connection ................................................................................................... 5
Figure 2 Firewall Settings ............................................................................................................................ 6
Figure 3 User Permission Control Settings Screen ...................................................................................... 6
Figure 4 Selecting Installation Language ..................................................................................................... 7
Figure 5 Installing the Run-time Library...................................................................................................... 7
Figure 6 Auto Installation ............................................................................................................................. 8
Figure 7 Selecting Custom Functions .......................................................................................................... 8
Figure 8 Saving the Data File ....................................................................................................................... 9
Figure 9 Software Installation Path .............................................................................................................. 9
Figure 10 Implementing Installation ............................................................................................................ 9
Figure 11 Completing Installation .............................................................................................................. 10
Figure 12 Login Screen .............................................................................................................................. 10
Figure 13 Setting the Server IP Address ...................................................................................................... 11
Figure 14 System Login .............................................................................................................................. 11
Figure 15 Instrument Connection Settings Screen ..................................................................................... 12
Figure 16 Automatic Detection Screen....................................................................................................... 12
Figure 17 LIS Setup ................................................................................................................................... 13
Figure 18 Connection Status ...................................................................................................................... 13
Figure 19 Server Status Icon ...................................................................................................................... 13
Figure 20 Server Boot Icon ........................................................................................................................ 13
Figure 21 Context Menu of Server Status Icon .......................................................................................... 13
Figure 22 Stopping Service ........................................................................................................................ 13
Figure 23 Server Interconnection ............................................................................................................... 14
Figure 24 Switch Server Type to Interconnection Server ........................................................................... 15
Figure 25 Server Interconnection Settings ................................................................................................. 15
Figure 26 Central Server connection .......................................................................................................... 16
Figure 27 Central Server Installation Path Settings.................................................................................... 17
Figure 28 Switch Server Type to Sub Server ............................................................................................. 17
Figure 29 Switch Server Type to Central Server ........................................................................................ 18
Figure 30 Central server “labXpert conn. Setup” screen............................................................................ 19
Figure 31 Sub server “labXpert conn. Setup” screen ................................................................................. 19
Figure 32 Context Menu of Server Status Icon .......................................................................................... 19
Figure 33 Stopping Service ........................................................................................................................ 19