With VB10A CSTD - AXB7

The document provides installation instructions for the Ysio software system, specifically for the VB10A version. It includes preparation steps, software installation procedures, and guidelines for backing up data and configurations. Additionally, it contains disclaimers, copyright information, and a table of contents outlining various sections related to the installation process.

Uploaded by

Claudio
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
30 views55 pages

With VB10A CSTD - AXB7

The document provides installation instructions for the Ysio software system, specifically for the VB10A version. It includes preparation steps, software installation procedures, and guidelines for backing up data and configurations. Additionally, it contains disclaimers, copyright information, and a table of contents outlining various sections related to the installation process.

Uploaded by

Claudio
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 55

Ysio

Software
System
beginning with VB10A

Answers for life.


2 Document Version / Disclaimer / Copyright

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens office to order a current version or refer to our website
http://www.healthcare.siemens.com.

Disclaimer
Siemens provides this documentation “as is” without the assumption of any liability under
any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.

Copyright
“© Siemens, 2010” refers to the copyright of a Siemens entity such as:
Siemens Healthcare GmbH - Germany
Siemens Aktiengesellschaft - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany

Ysio AXB7-020.816.02.04.02 Page 2 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Table of Contents 3

1 Software Installation 4

1.1 Manufacturer’s Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4


1.1.1 How to Use these Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.2 Screenshots & Graphics in this Document . . . . . . . . . . . . . . . . . . 4
1.1.3 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.1 Reading Out the Service Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.2 Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.3 Saving Log Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2.4 Making a Note of the Customer Data . . . . . . . . . . . . . . . . . . . . . . 6
1.2.5 Uninstalling System Management . . . . . . . . . . . . . . . . . . . . . . . . 8
1.3 Software Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.3.1 Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.3.2 New Installation of Software, including Partitioning of the HD . . 13
1.3.3 New Installation of Software without Partitioning of the HD . . . . 13
1.3.4 Copying the Image Contents to the System Partition . . . . . . . . . . 13
1.4 Starting the Service Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
1.5 Reestablishing the System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
1.5.1 Restoring the Backup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
1.5.2 Restoring Test Images, optional only for the VB10B System SW. . 24
1.5.3 Image Rescue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
1.5.4 Reinstalling System Management . . . . . . . . . . . . . . . . . . . . . . . . . 25
1.5.5 Installing applicable RUH Packages . . . . . . . . . . . . . . . . . . . . . . . . 30
1.5.6 Reinstalling Service Online Help and Online User Manuals . . . . . 30
1.5.7 Installing the Service CDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
1.5.8 Final Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
1.5.9 Restoring the HIPAA Backup (optional) . . . . . . . . . . . . . . . . . . . . . 31
1.6 Copying Patient Studies from an Old Hard Disk to a New Hard Disk with an 32
Empty Image Partition (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.6.1 Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
1.6.2 Work Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
1.7 Component Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
1.7.1 XCU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
1.7.2 Touch User Interface (TUI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
1.7.3 Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
1.7.4 RAD Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
1.7.5 CAN Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

2 Changes to Previous Version 53

3 List of Hazard IDs 54

© Siemens, 2010 AXB7-020.816.02.04.02 Page 3 of 55 Ysio


11.13 CS PS AX R/F
4 1 Software Installation

1.1 Manufacturer’s Note

1.1.1 How to Use these Instructions


System knowledge (training, online Help), knowledge of Windows XP and patience
are required for the following work steps.
You will be successful only if you follow these instructions step by step and do not
skip any points.
n Please first read the entire paragraph in advance to have a better understanding
of the work steps.
n Go back to the beginning of the paragraph and follow the instructions step by
step.

1.1.2 Screenshots & Graphics in this Document


Screenshots and graphics used in this document should be viewed as examples only.
They cannot describe all possible input combinations; they have an informative character
and are used only for better orientation.
The actual parameters that must be configured can be found in the text and the tables, or
depend on the requirements and configuration (items included in the shipment) of the
individual system.

1.1.3 Introduction

The first section of this document, including ( Restoring the HIPAA Backup (option‐


al) / Page 31) is no longer applicable and is only for backup purposes!
SW reinstallation is performed using the COBR Tool. See the Ysio G2 (Installation / Soft‐
ware) CB-DOC on the CS Intranet, Print Number AX41-060.816.12.

This document describes installation of the system software for Ysio on a Celsius M460
PC. It is valid only for the SW versions mentioned on the cover sheet.
To install the software, you will need the corresponding SW installation DVD.

Ysio AXB7-020.816.02.04.02 Page 4 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 5

1.2 Preparations

1.2.1 Reading Out the Service Key


n Since the service key is not restored automatically, it must be read out manually prior
installing the software.
- Start the Service Software, see ( Fig. 11 Page 17).
- At the service software login, read out the service key and make a note of it in the
table below ( Tab. 1 Page 5).
Tab. 1 Service Key

14 digits 6 digits

1.2.2 Backup

1.2.2.1 HIPAA Backup


n If you have the HIPAA option installed, together with the network administrator, make
sure that an up-to-date backup of the HIPAA users exists.

1.2.2.2 System Backup


n Perform the following work steps to generate an up-to-date system backup on CD/
DVD.
- Insert a blank DVD into the upper drive of the FL-C PC.
- From the Service Software Main menu ( Fig. 16 Page 19), select <Backup & Re‐
store>.
 The Backup & Restore page appears.
- Select <Backup>.
- From the Select drive: list box, select <CD/DVD>.
- - Press the <Go> button to start the backup routine.
 The backup may take a couple of minutes.
 During the backup routine, status messages are displayed in the progress win‐
dow.
- After the backup has finished, remove the medium from the drive and close it.

1.2.2.3 Backup of Test Images

 Due to a software characteristic of the VB10B system SW, the Pattern folder, which re‐
sides on the D:\ partition, needs to be backed up prior to the SW installation!

Perform the following work steps:

© Siemens, 2010 AXB7-020.816.02.04.02 Page 5 of 55 Ysio


11.13 CS PS AX R/F
6 1 Software Installation

n Press the <Ctrl> and <Esc> keys simultaneously.


 The Windows Start menu appears.
n Select <Start> <Programs> <Windows Explorer>.
 The Windows Explorer will start.
n Plug in your USB stick.
n Navigate to D:\Studies\.
n Copy the Pattern folder to your USB stick.
n Close Windows Explorer and remove your USB stick.
n Burn the copied Pattern folder with your Service PC from your USB stick to a blank CD
you brought with you.
n Label the CD as follows:
Backup Test Images
D:\Studies\Pattern
Serial Number of the system
Current date

1.2.3 Saving Log Files


n Required only if the SW installation is performed to solve an error condition on the
specific system. In this case, a copy of the system log file should be saved for potential
error traceability.
For System SW VA2xx:
- Select <Diagnostic> <View FLC/OS Logs> <Save All> in the FL-C SSW.
For System SW VAB10:
- Select <File & Image Tools> <Logs> <Save All> in the FL-C SSW.
n Label the burned media as System Log Files and “date stamp”.

1.2.4 Making a Note of the Customer Data

1.2.4.1 Reading Out the Link Speed of the Primary Network Interface Card
n Select <System> <System Settings> <Control Panel>.
- Double-click on the <System> icon.
n Select the <Hardware> tab.
- Click on <Device Manager>.
- Double-click on <Network adapters>.
- Right mouse-click on the <Broadcom NetXtreme Gigabit> entry.
- Select <Properties>
 The Broadcom NetXtreme Gigabit Ethernet Properties window comes up.
-
n
Select the <Advanced> tab.

Ysio AXB7-020.816.02.04.02 Page 6 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 7

- Under Property: select <Speed & Duplex>.


- In the Value: list box, make a note of the entry for the link speed in
( Tab. 2 Page 7).
- Close the <Broadcom NetXtreme Gigabit Ethernet Properties> window with
<OK>.
- Close the <Device Manager> window.
- Close the System Properties window.
- Close the Control Panel window.
Tab. 2 Network Properties

Link Speed

1.2.4.2 Reading out the Configuration of the Paper Printer (Option)


An (optional) paper printer connection cannot be restored using Backup/Restore.
If a paper printer is configured, the settings must be backed up manually.
n Press the <Ctrl> and <Esc> keys simultaneously.
 The Windows Start menu appears.
n Select <Start> <Settings> <Printers and Faxes>.
 The Printers and Faxes window appears ( Fig. 1 Page 7).
n Check whether a printer is configured, e.g. HP Laser Jet 4050 Series PS.
Fig. 1: Printers and faxes

n If yes, right mouse-click on the printer icon and select <Properties> from the context
menu.
- On the General tab, read out the printer model and make a note of it in
( Tab. 3 Page 8).
- Click on the Ports tab.
- With the vertical scroll bar, scroll all the way down, read out the printer's TCP/IP ad‐
dress and make a note of it in ( Tab. 3 Page 8).

© Siemens, 2010 AXB7-020.816.02.04.02 Page 7 of 55 Ysio


11.13 CS PS AX R/F
8 1 Software Installation

- Select the Advanced tab and make a note of the Driver's name in
( Tab. 3 Page 8).
- On the Device Settings tab, read out the paper size used, i.e. A4, and make a note
of it in ( Tab. 3 Page 8).
Tab. 3 Paper Printer Configuration

Printer Model
TCP/IP address _____ . _____. _____ . _____
Driver name
Paper size

n Close the Paper Printer Properties window.


n Close the Printer and Faxes dialog.

1.2.5 Uninstalling System Management


n System Management needs to be uninstalled, e.g. prior every update of the System
SW or prior to when the host name needs to be changed for any reason.
- Select <Configuration> <System Management> in the FL-C SSW.
- Close the System Management install.log window with <Continue>.
- Activate the Deinstallation check box H and then press the <Go> button to trigger
the uninstallation.
 The uninstallation process is displayed in a new window.
 Uninstallation may take up to 10 minutes; do not interfere with the uninstalla‐
tion by pressing any other buttons!
 Please wait patiently until the <Continue> button is displayed at the lower right
corner in the browser window.

Fig. 2: System Management, uninstallation

 Sporadically while the uninstallation process is taking place, an opcle error pop-
up may appear. This error is not relevant for uninstallation. Close the pop-up
with <OK> to continue.

Ysio AXB7-020.816.02.04.02 Page 8 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 9

Fig. 3: opcle error pop-up

- Press <Continue> when uninstallation is complete and the System Management in‐
stall.log is displayed( 1/Fig. 25 Page 28).
- A reboot request is displayed on the following page, see ( Fig. 4 Page 9).
 Confirm the reboot request with <OK>.
 The FL-C will reboot.
Fig. 4: System Management uninstalled, reboot request

Uninstallation Successful
n Following the reboot, again select <Configuration> <System Management> in the
FL-C service mode.
n No checkboxes for Installed and Activated may be selected and the status
must be set to not installed.
n No template version may be displayed under Template update.
 System Management has been uninstalled successfully.
n Exit the service mode.
n Switch OFF the system.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 9 of 55 Ysio


11.13 CS PS AX R/F
10 1 Software Installation

n If support is necessary, contact the SRS Help Desk.

Hotline: +49 9191 18 8080-191

 e-mail: [email protected]
Service hours are 7.00 - 22.00 CEST

Ysio AXB7-020.816.02.04.02 Page 10 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 11

1.3 Software Installation

1.3.1 Preparations

For the following steps, it is important that the computer recognize the CD drive as the
bootable device.
 Therefore, you have to change the BIOS settings and enable the Boot menu.
The old BIOS settings must be re-established at the end of the installation!

1.3.1.1 Enabling the Boot Menu


n Switch on the system.
n Press the <F2> key when the message Press [F2] for setup is displayed in the
lower left corner of the screen.
n After approx. 2 sec., the FL-C asks for the BIOS password. See the corresponding entry
in the AX Password List, which is published in the CS Knowledge Base.
After entering the password, press the Return ↵ key.
 The Phoenix TrustedCore (tm) BIOS Setup Utility menu appears.
n In the Main menu, select <Boot Features> and confirm with the Return ↵ key.
n In the Boot Features menu, select <Boot Menu> and press the Return ↵ key.
n Select <Enabled> by using the down arrow key.
n Press the Return ↵ key.
n Press <Esc> two times to go back to the Phoenix TrustedCore (tm) BIOS
Setup Utility main menu.
n Select <Save Changes & Exit> and press Return ↵.
n Insert the Software Installation DVD into the CD drive and wait until the LED on the
drive stops blinking.
n Confirm the Setup Confirmation request with <Yes> (Return).

1.3.1.2 Booting the System DVD


n The PC will now start to reboot.
- Press the <F12> key several times while the FL-C reboots.
n In the Boot Menu that follows, select <SATA CD:> by using the arrow keys.
n Press Return ↵ to confirm.
 The PC will now reboot from the DVD drive and afterwards the Selection OF
THE SYSTEM window appears, ( Fig. 5 Page 12).
- Select <YSIO System>.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 11 of 55 Ysio


11.13 CS PS AX R/F
12 1 Software Installation

n After this, the FLUOROSPOT COMPACT Software Installation menu appears,


( Fig. 6 Page 12).

Fig. 5: Selection of the system Fig. 6: FLUOROSPOT Compact software installation

- Prepare hard disk + load application


This option is used if you want to install the entire application, the operating system and
an empty image partition. If you choose this option, you will lose your medical studies.
Continue at ( New Installation of Software, including Partitioning of the


HD / Page 13).
- Load application
This option is used if you need to install only the operating system and the application.
The second partition of the HD with the medical studies is not affected. The existing
medical studies are retained. Continue at ( New Installation of Software without Parti‐
tioning of the HD / Page 13).

Ysio AXB7-020.816.02.04.02 Page 12 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 13

1.3.2 New Installation of Software, including Partitioning of the HD


n Click on <Prepare hard disk + load application>.
The following screen appears with a warning about the consequences,
( Fig. 7 Page 13):

Fig. 7: Fig. 8:

n Click on <OK> for the next screen, ( Fig. 8 Page 13).


n Reenter the reboot password and click on <OK> to start the process.
After a few seconds, a new window comes up displaying the installation progress.
n Continue with ( Copying the Image Contents to the System Partition / Page 13).

1.3.3 New Installation of Software without Partitioning of the HD


The patient studies will not be deleted!
n Click on <Load application>.
n Continue with ( Copying the Image Contents to the System Partition / Page 13).

1.3.4 Copying the Image Contents to the System Partition

The SW installation and subsequent component download process may take up to 45 mi‐


nutes. During that time, no mouse / keyboard activity is necessary.
Do not remove the SW DVD unless the download process has ended and the FL-C re‐
mains in patient list.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 13 of 55 Ysio


11.13 CS PS AX R/F
14 1 Software Installation

n A new window appears and displays the installation progress. This will take approx.
15 minutes.

If the hour glass appears for more than 1 minute in the Ghost screen, disconnect the
 power connector on the back of the PC and reconnect it. Wait until the PC boots and
start again with ( Booting the System DVD / Page 11).

Fig. 9: Software installation progress window

If a blank screen with the note Clearing CAF file attribute... appears for


more than 1 minute, disconnect the power connector on the back of the PC and recon‐
nect it. Wait until the PC boots and start again with ( Booting the System
DVD / Page 11).

n At the end of the installation process, an automatic reboot is performed:


- Sysprep runs unattended.
- Product key definition runs unattended.
- Command windows will open and close automatically.


Patience is required!
Do not touch the keyboard or mouse. Do not remove the DVD from the drive.

n This reboot is followed by another reboot (certain pop-ups will appear). No user action
is required.
- The XP firewall is configured.
n Download of the XCU and the generator will be performed automatically. This may
take 30 - 40 minutes! Be patient, no interaction is necessary!

Ysio AXB7-020.816.02.04.02 Page 14 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 15

n When finished, the FL-C application starts and comes up with the FL-C patient list.

If a message box appears:


n FD in Table: Not calibrated

 n
or
Flat detector error ...
Confirm the message and ignore it.

 If a window named System Message appears in the lower right corner of the screen,
confirm it.

n Remove the Software Installation DVD from the drive.


n Switch OFF the system and wait until the system has shut down.

 If the PC does not shut down within 5 minutes, disconnect the power connector on the
back of the PC and reconnect it.

n Switch the system ON again.


n Press the <F2> key when the message Press [F2] for setup is displayed in the
lower left corner of the screen.

1.3.4.1 Disabling the Boot Menu


n After approx. 2 sec., the FL-C asks for the password. See the corresponding entry in
the AX Password List, which is published in the CS Knowledge Base.
After entering the password, press the Return ↵ key.
 The Phoenix TrustedCore (tm) BIOS Setup Utility menu appears.
n In the Main menu, select <Boot Features> and confirm with the Return ↵ key.
n In the Boot Features menu, select <Boot Menu> and press the Return ↵ key.
n Select <Disabled>.
n Press the Return ↵ key.
n Press <Esc> two times to go back to the Phoenix TrustedCore (tm) BIOS
Setup Utility main menu.
n Select <Save Changes & Exit> and press Return ↵.
n Confirm the Setup Confirmation request with <Yes> (Return).
 Then the PC reboots again and comes up with the FL-C patient list.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 15 of 55 Ysio


11.13 CS PS AX R/F
16 1 Software Installation

1.4 Starting the Service Software


English is set as the default layout for the keyboard. ( Fig. 10 Page 16) shows the
 English layout for the keyboard. This is intended to serve as an aid in finding keys if the
existing keyboard does not have the English layout.

Fig. 10:

The system must be booted and the application software started to start the service soft‐
ware.

If a message box appears:


n FD in Table: Not calibrated

 n
or
Flat detector error ...
Confirm the message and ignore it.

 If a window named System Message appears in the lower right corner of the screen,
confirm it.

n Proceed as follows to start the service software:


- In the patient list, click on the <Service login> icon ( 1/Fig. 11 Page 17) on the
Service tab.

Ysio AXB7-020.816.02.04.02 Page 16 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 17

Fig. 11: Service login

(1) Service login icon


n Two user messages may possibly appear, ( Fig. 12 Page 17) and
( Fig. 13 Page 17).
 Close them with <Cancel>.
Fig. 12: License file not detected Fig. 13: Service key not detected

n The service software will be started and the login screen will appear
( Fig. 14 Page 18).
- Mark the HSet as default checkbox, ( 1/Fig. 14 Page 18).

© Siemens, 2010 AXB7-020.816.02.04.02 Page 17 of 55 Ysio


11.13 CS PS AX R/F
18 1 Software Installation

- Enter the first 14 digits of the service key you noted in ( Tab. 1 Page 5) in the left
input box ( 2/Fig. 14 Page 18).
- Click in the right input box ( 3/Fig. 14 Page 18).
- Enter the last 6 digits of the service key you noted in ( Tab. 1 Page 5).
- Press the <OK> button to login.

Fig. 14: Service software, login

2 3

n If a pop-up appears, ( Fig. 15 Page 18), close it with <OK>.


Fig. 15:

Ysio AXB7-020.816.02.04.02 Page 18 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 19

n If a valid service password has been entered correctly, the Service Software Main
menu ( Fig. 16 Page 19) appears.
Fig. 16: Service software, Main menu

n For information only:

© Siemens, 2010 AXB7-020.816.02.04.02 Page 19 of 55 Ysio


11.13 CS PS AX R/F
20 1 Software Installation


The FL-C SSW ( Fig. 17 Page 20) is started by clicking on the <FL-C Service> button in
( Fig. 16 Page 19).

Fig. 17: FL-C Service Software, Main menu

Ysio AXB7-020.816.02.04.02 Page 20 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 21

1.5 Reestablishing the System Configuration

1.5.1 Restoring the Backup


Insert the backup medium (CD/DVD) into the drive and wait until the LED stops blinking.
n From the Service Software main menu, select <Backup/Restore>.
- The FL-C Backup / Restore window appears.
- Click on the <Restore> button.
 The Restore Files page is displayed ( Fig. 18 Page 22).
- From the Drives list box, select <CD / DVD>.
- Select the last backup that was made from the packages list.
- Press the <Go> button to start the restore routine.

 Patience is required. The restore process my take up to 20 minutes.

- Confirm all the safety inquiries that follow with <OK> or <Yes>.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 21 of 55 Ysio


11.13 CS PS AX R/F
22 1 Software Installation

- When Restore is finished, remove the CD from the drive and close the drive.

Fig. 18: SSW: Backup & Restore, selecting the restore package

Ysio AXB7-020.816.02.04.02 Page 22 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 23

n Once the restore has finished, the Distribute Restored Files dialog appears
( Fig. 19 Page 23).
Fig. 19: SSW: Backup & Restore, Distribute Restored Files

- In the action bar, press <Select All> to select all subcomponents.


 All checkboxes are marked, H.
- Press <Go> to copy the configuration files to their working directories.
- Wait until all configuration files are sent to the subcomponents.
 Each subcomponent states completed when distribution is completed.
Except wi-D: incomplete ...
Ignore the wi-D warning message.
- Close the Restore page with <Exit>.
- Confirm the request for Config Check and wait until it has ended.
 Config Check will run automatically.
- Run Config Check and / or Auto Conf if necessary / requested.
- Close the Config Check / Auto Conf page with <Exit>.
- Ignore the Reboot Required message.
n Press <Home> to return to the service software Main menu.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 23 of 55 Ysio


11.13 CS PS AX R/F
24 1 Software Installation

n Close the service software with <Logout>.


n An error message may appear:
- Confirm the pop-up with <OK>.

Fig. 20: Error pop-up

n Switch OFF the system and wait until the system has shut down.
n Switch the system ON again.

1.5.2 Restoring Test Images, optional only for the VB10B System SW

Due to a SW characteristic of the VB10B system SW in combination with Ysio systems,


 the <Import Test Images> button MAY NOT BE USED, because not all test images will be
recovered!

n If the Prepare hard disk + load application option was selected for the SW installa‐
tion procedure, the test images must be recovered as follows:

For backup purposes, the pattern folder is also published on the CS Intranet. Navigate

 to: For Service > AX Systems > Radiography > Ysio Systems > Software. Download
and unzip the pattern.exe folder and burn the extracted pattern folder on CD using
the Service PC.

Insert the CD which contains the backed up Pattern folder into the into the CD drive
and wait until the LED on the drive stops blinking.
- Press the <Ctrl> and <Esc> keys simultaneously.
 The Windows Start menu appears.
- Select <Start> <Programs> <Windows Explorer>.
 The Windows Explorer will start.
- Navigate to D:\Studies\.
- Delete the Pattern folder under D:\Studies.
- Press the <Ctrl> and <Esc> keys simultaneously.
 The Windows Start menu appears.
- Select <Start> <Programs> <Windows Explorer>.
 A second Windows Explorer will start.
- Navigate to the CD drive which contains the Pattern folder.
- Select the Pattern folder.
- Copy the Pattern folder to the clipboard by pressing the <Ctrl> and <C> keys simul‐
taneously.

Ysio AXB7-020.816.02.04.02 Page 24 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 25

- In the first Explorer window, paste the Pattern folder to D:\Studies by pressing the
<Ctrl> and <V> keys simultaneously.
- Close both Explorer windows.

 Do ! NOT ! press the <Import Test Images> button under <Diagnostic> <Import / Ex‐
port> in the FL-C service mode, otherwise test images will be deleted!

Check whether the number of test images is complete:


- In Patient List, select the Service tab card.
- On the Service tab card, click on the Check Test Images icon.
 The FL-C will change to the Postprocessing tab card.
- Double-click in the displayed image.
 The FL-C will change to the Documentation mode.
- 13 different test images are to be displayed.

1.5.3 Image Rescue


Start the SSW.
n From the FL-C SSW, select <System>
n Choose <Image Rescue> from the tree browser on the left.
n Press the <Image Rescue> button and answer the requester with <OK>.
 After a few seconds, the message Image save completed successfully
is displayed.
 The Patient Database is rebuilt.

1.5.4 Reinstalling System Management

Prerequisite for the following configuration steps of System Management is that the SRS
router is set up correctly (refer to the corresponding system Startup Instructions). Other‐

 wise the System Management installation will fail.


This document is N O T intended for initial SRS setup. For this, see the corresponding In‐
stallation and Startup Instructions (AX41-060.843.23)!

These instructions are not applicable for initial System Management installation!

 The following paragraph describes how to restart System Management if it was previ‐
ously in use.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 25 of 55 Ysio


11.13 CS PS AX R/F
26 1 Software Installation

1.5.4.1 Activating System Management


n Select <Configuration> <SRS Configuration> <System Management> in the FL-C
service software main menu.

Every time <System Management > is selected, the System Management install.log is
 updated and displayed in the browser window ( Fig. 21 Page 26). Press <Continue>
to proceed.

Fig. 21: FL-C SSW, Configuration, System Management install.log

n Press <Continue>.
n At the top of the page, the Location which was restored is displayed automatically
and appears dimmed.

Ysio AXB7-020.816.02.04.02 Page 26 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 27

n The IP address of the DICOM Connection that is configured under Network Settings
(Control Panel) is displayed under Network Adapter TCP/IP address
( Fig. 22 Page 27).

If Outside NAT is configured, the IP addresses for Event Mgmt. and SW Distribution Serv‐
 er are also shown in ( Fig. 22 Page 27) (the screen shot below illustrates an example
only!).

Fig. 22:

n To test the connectivity, press the <Ping> button. Since this is a very time-critical ac‐
tion, it is possible that a connection will not be immediately established. Because of
this, several attempts should be made.
The server must send back a message displayed in the status bar:
Fig. 23: Pinging the System Management server

n Starting the Installation / Registration Process


- With System Management not installed, the System Management page comes up
as shown in ( Fig. 26 Page 29).
- Activate the Installation & Registration checkbox.
 The Activation checkbox will be activated automatically.
- Simply start the installation / registration process by pressing <Go>.
 The installation / registration process is displayed in a new window and may
take up to 20 minutes and runs automatically.
 Sporadically while the installation / registration process is taking place, an opcle
error pop-up may appear ( Fig. 24 Page 27). This error is not relevant for in‐
stallation / registration. Close the pop-up with <OK>.

Fig. 24: opcle error pop-up

© Siemens, 2010 AXB7-020.816.02.04.02 Page 27 of 55 Ysio


11.13 CS PS AX R/F
28 1 Software Installation

Fig. 25: System Management installation / unnstallation

- Once the installation / registration process has finished, the System Management
install.log is displayed( Fig. 25 Page 28).
 Press <Continue>, ( 1/Fig. 25 Page 28).
-The status of the MNP is as shown in ( Fig. 27 Page 29).
Checking the Installation
n The following conditions must be met ( Tab. 4 Page 28):
- Select <Configuration> <SRS Configuration> <System Management> in the FL-C
service software.
Tab. 4 System Management successfully installed

Function Installation & Offline In‐ Uninstal‐ Current Status


Registration stallation lation
System Management H Q Q installed
Base
Activation Active
H
Template n.a. Q Template version,
e.g. 1.012


If the Installation / Registration fails, what is called a rollback to an Offline Installation
will be performed. For the status, see ( Fig. 29 Page 30).

Ysio AXB7-020.816.02.04.02 Page 28 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 29

Following a failed installation and subsequent rollback to an Offline Installation, an


additional attempt to install should be made. If this attempt also fails, contact the SRS
Help Desk.
Hotline: +49 9191 18 8080-191

 e-mail: [email protected]
Service hours are 7.00 - 22.00 CEST
For support in the U.S., please call Uptime Service Center
Hotline: 1-800-576-7336
and request SRS support

Different Status of System Management Installation


n The screen shots below represent the different status of the MNP:
- System Management not installed

Fig. 26: System Management not installed

- System Management installed and activated

Fig. 27: System Management installed and activated

- System Management installed and deactivated (e.g. for troubleshooting purposes)

Fig. 28: System Management installed and deactivated

© Siemens, 2010 AXB7-020.816.02.04.02 Page 29 of 55 Ysio


11.13 CS PS AX R/F
30 1 Software Installation

- Offline Installation

Fig. 29: System Management, offline installation

1.5.5 Installing applicable RUH Packages


n There are certain conditions under which a RUH package that was installed previously
is no longer installed. This might be the case after:

a) Restoring a recovery image which does not contain the RUH package(s)
b) Reinstallation of the system software
c) Replacing the PC (which leads to: a) restore of a recovery image)
n How to find out which RUH package(s) applies to the particular system:
- A technical info is published in the CS Knowledge Base (SKB0037721). This is a ref‐
erence table for all RuH packages that have been released for the RFU platform.
To make sure if a package applies to a particular system, contact your dispatcher/
USC/RSC.
- Install the RuH package per the appropriate UI.
n After this is done, create an up-to-date recovery image of the system partition.

1.5.6 Reinstalling Service Online Help and Online User Manuals


To reinstall the System Online Help, proceed as follows:
n Start the service software and select <Configuration> <Local Installation>.
- Insert the Ysio Online Help CS CD-ROM (Print No. AXB7-500.880.0X.0X.02) from
the update kit into the CD drive and wait until the LED stops blinking.
- Press the <Go> button.
 A command window will pop up. Enlarge it and follow the instructions on the
monitor
- Remove the CD from the drive.

To reinstall the Online User Manuals, perform the following steps:


- Insert the Online User Manuals CD from the update kit into the CD drive and wait
until the LED on the drive stops blinking.
- Press the <Local Installation> button and follow the specific instructions.

Ysio AXB7-020.816.02.04.02 Page 30 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 31

1.5.7 Installing the Service CDs


n If there are patches for the FL-C application or Microsoft hot fixes available, install
them according to the appropriate UI.

1.5.8 Final Work Steps


n If the link speed of the DICOM network interface card was previously set differently
from the default, ⇒ auto, then refer to ( Reading Out the Link Speed of the Primary
Network Interface Card / Page 6) to reestablish the desired setting.
n Check to make sure that all options for the site are available.
n Switch off the system.
n Wait until the FL-C has shut down (this may take up to 2 minutes).
n Switch on the system and wait until the FL-C application software is running.
n Check for correct functioning of the image flip with fluoroscopy.
This is the end of the software installation procedure. Congratulations!

1.5.9 Restoring the HIPAA Backup (optional)


n If the HIPAA option is installed, contact the customer network administrator to restore
the HIPAA backup and the user passwords.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 31 of 55 Ysio


11.13 CS PS AX R/F
32 1 Software Installation

1.6 Copying Patient Studies from an Old Hard Disk to a New


Hard Disk with an Empty Image Partition (optional)
This step becomes necessary when a PC is replaced and the customer wants to save pa‐
 tient studies from the old hard disk. Therefore they have to be transferred to the new
hard disk.

1.6.1 Prerequisites
n The step ’New installation of software, including the partitioning of the HD' must al‐
ready have been performed completely, including the restore of the site-specific da‐
ta.
n There are no images on the new hard disk.
n Switch off the system.
n Wait until the FL-C has shut down.
n System power has to be turned off at the main breaker.

1.6.2 Work Sequence

The following hardware configuration is expected:

 - New PC with new HD already installed


- SW installed on new HD

n If not already done, remove the left cover from the old PC now.
- Loosen the two green screws on the back of the PC housing and move the side pan‐
el slightly to the rear.
- Remove the side panel of the PC.
- Disconnect the power connector ( 1/Fig. 30 Page 33) and data cable
( 2/Fig. 30 Page 33) at the back of the hard drive.
- Slightly push in the 2 green ends of the mounting brackets and pull out the hard
disk.
n Insert the old hard disk ( 2/Fig. 31 Page 33) in the drive cage of the new PC, direct‐
ly below the first hard drive ( 1/Fig. 31 Page 33).
n Use one free connector of the SATA power connection cable and plug it into the pow‐
er socket of the secondary hard disk as shown in ( Fig. 32 Page 34).
n Unplug the SATA data cable of the first DVD-drive (upper one) and plug it into the da‐
ta socket the secondary hard disk as shown in ( Fig. 32 Page 34).
n Switch on the system at the generator and wait until the application is running.
n Enter the service mode.
n Start the FL-C SSW.
n Select <System>.
n Select <Image Rescue> from the tree browser on the left.

Ysio AXB7-020.816.02.04.02 Page 32 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 33

n Click on the <Save> button.


The Command window opens in the bottom right corner (below the menu). It is very
small. Increase the size by clicking on the corresponding NT symbol or:
- Click on the title bar and drag the window to the top left corner of the screen.
- Increase the size of the window (bottom right corner of the window) so that you
can see the full text (also use the scroll bars).

Fig. 30: Dismantle the old hard disk Fig. 31: Old hard disk in new PC

(1) Power plug


(2) Data cable
(1) Primary hard disk
(2) Secondary hard disk

© Siemens, 2010 AXB7-020.816.02.04.02 Page 33 of 55 Ysio


11.13 CS PS AX R/F
34 1 Software Installation

Fig. 32: Hard disk connected to new PC

(1) Primary hard disk


(2) Secondary hard disk

The following message is displayed:


Do you want to save the images?
Enter 1 for Yes, 0 for No
n Press <1> ↵ and Return. Copying of the images will start.

 Do not make any entries until copying is finished!

After a while, the following message appears:


Image saving completed successfully!
Press any key to continue...
n Press any key and close the System window with <OK>.
n Exit the service mode and switch off the system.
n When the FLC has switched off, remove the old HD from the PC and reinstall all cov‐
ers.
n Reconnect SATA data cable to the first (upper) DVD drive.
n Reinstall all covers.
n Reinstall the secondary hard disk in the old PC.
n Switch on the system and wait until the application is running.
n Start the FL-C SSW.
n Select <System>.
n Select <Image Rescue> from the tree browser on the left.

Ysio AXB7-020.816.02.04.02 Page 34 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 35

n Press the <Image rescue> button. Answer the requester with <OK>.
After a few seconds, the message Image save completed successfully is
shown.
 The Patient Database is rebuilt.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 35 of 55 Ysio


11.13 CS PS AX R/F
36 1 Software Installation

1.7 Component Download


This section describes the necessary work steps after one of the following components
has been replaced or has to be downloaded again:
-
n
XCU
- Touch User Interface (TUI)
- Polydoros F80
- RAD Controller
- CAN Components
Fig. 33: SSW: Configuration, Download

1.7.1 XCU
Start the service software.
n Select <Configuration> <Download>
 The PMS Download page appears ( Fig. 33 Page 36).
- Mark the Install checkbox H for XCU+.

Ysio AXB7-020.816.02.04.02 Page 36 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 37

- Start the download with <Go>.


n After the download is finished, leave the page with <Exit>.
n Press the <Home> button to return to SSW main menu.
n Select <Restore> from the tree browser on the left.
- Select <Restore> from the tree browser on the left.
 The following screen appears ( Fig. 34 Page 37).
Fig. 34: Distribute Files only

- Select <Distribute Files only>, ( 1/Fig. 34 Page 37).

© Siemens, 2010 AXB7-020.816.02.04.02 Page 37 of 55 Ysio


11.13 CS PS AX R/F
38 1 Software Installation

Fig. 35: XCU restore

In the next screen, ( 1/Fig. 35 Page 38), mark the checkbox H for X-Ray Control
Unit (XCU).
- Start the restore with <G0>.
- After the restore is finished, leave the page with <Exit>
n Press <Home> to return to the service software Main menu.

1.7.1.1 Work Steps after Replacing the XCU

After XCU replacement, a new service key and new license keys are necessary. (Li‐
cense.dat file)
The Host ID the System Part No. and Serial No. are required for a new service key and

 new license keys.


The Host ID is printed on a small sticker (labeled MAC address) on the body of the XCU. It
consists of the last 8 digits of the Serial Number, reading from right to left!
Please contact the Licensing Center to obtain a new service key / updated license.dat file

Perform the following sequence after the XCU has been replaced.
Download
n Switch ON the system and wait until the FL-C application software is running.
 An error message may be displayed on the screen: error 101/80.
 Confirm the error!
 The generator continues bootup sequence and finally displays 56 on the D600
board and 63 on the D611 board.

Ysio AXB7-020.816.02.04.02 Page 38 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 39

 The generator bootup time may take longer than usual.

n An error message may be displayed on the screen at the end of the generator bootup
phase: error 101/80.
 Confirm the error!
n Start the service software.
n In the Service Software log-in, highlight the checkbox H Set as default.
- Enter the first 14 digits of the N E W service key in the left input box.
- Click on the right input box.
- Enter the last 6 digits of the N E W service key.
- Press <OK> to log in.
n Select <Configuration> <Download>.
 The PMS Download page appears.
 The expected actual version of the XCU+ to be displayed is VB08B.


If the displayed actual version is lower than VB08B, download of the XCU will fail!
=> Order a new XCU.

- Highlight the Install checkbox H for XCU+.


- Start the download with <Go>.
n After the download is finished, leave the page with <Exit>.
n Press <Home> to return to SSW Main menu.
n Select <Backup/Restore> from the Service Software main menu.
- Click on <Restore> from the tree browser on the left.
 The Restore Files page is displayed ( Fig. 36 Page 40).
- Select <CD / DVD> from the Drives list box.
- Select the last backup that was made from the packages list.
- Press the <Go> button to start the restore routine.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 39 of 55 Ysio


11.13 CS PS AX R/F
40 1 Software Installation

- When Restore is finished, remove the DVD from the drive and close the drive.

Fig. 36: Backup & Restore, Select Package

n Then the Distribute Restored Files dialog appears ( Fig. 37 Page 41).
- Highlight the checkbox H for X-Ray Control Unit (XCU) ( 1/Fig. 37 Page 41).
- Start the restore with <G0>.
- After the restore is finished, leave the page with <Exit>.
- If prompted, confirm the request for Config Check and wait until it has ended.
 Config Check will run automatically.
- Run AutoConf, if requested.
- Close the AutoConf page with <Exit>.
- Press <Home> to return to the service software Main menu.

Ysio AXB7-020.816.02.04.02 Page 40 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 41

Fig. 37: XCU restore

Installing an Updated License.dat File


Insert or connect the medium (CD or USB stick) that contains the License.dat file.

1. Select <FL-C Service> <Configuration> <License> from the SSW Main menu.
2. Press <License Import>.
3. Press <Go> to import the file.
4. Confirm any pop-up windows that appear.
5. Exit FL-C service.
6. Switch the system OFF.
7. Wait until the FL-C has switched off.
8. Switch the system back on at the remote control console.
System Configuration:
1. Select <Configuration> <System config> from the service software main menu.
2. Select <AutoLic>.
3. Select <Save> and then exit the menu with <EXIT>.
Confirm the messages with <OK>.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 41 of 55 Ysio


11.13 CS PS AX R/F
42 1 Software Installation

4. A Config Check is started. Exit this window with <OK>.

1.7.2 Touch User Interface (TUI)


Start the service software.
n Select <Configuration> <Download>.
 The PMS Download page appears ( Fig. 33 Page 36).
- Highlight the Install check box H for Touch User Interface 1 or (2).
- Start the download with <Go>.
n After the download has finished, leave the page with <Exit>.
n Press <Home> to return to the service software Main menu.
n Select <Backup/Restore> from the Service Software main menu.
 The Backup / Restore window appears.
- Click on <Restore> from the tree browser on the left.
 The Restore Files page is displayed ( Fig. 38 Page 43).
- From the Select drive list box, select <CD / DVD>.
- Select the last backup that was made from the packages list.
- Press the <Go> button to start the restore routine.

Ysio AXB7-020.816.02.04.02 Page 42 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 43

- After the restore is finished, remove the DVD from the drive and close the tray.

Fig. 38: SSW, Restore

n The Distribute Restored Files dialog appears ( Fig. 39 Page 44).


- Mark the check box H for TUI1, ( 1/Fig. 39 Page 44).
- Press <Go> to copy the configuration files to their working directory.
 The selected sub-component states completed when distribution is completed.
- After the restore is completed, leave the page with <Exit>
- If prompted, confirm the request for Config Check and wait until it has ended.
 Config Check will run automatically.
- Run Auto Conf, if requested.
- Close the Auto Conf page with <Exit>.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 43 of 55 Ysio


11.13 CS PS AX R/F
44 1 Software Installation

- Press <Home> to return to the service software Main menu.

Fig. 39: TUI restore

1.7.2.1 Emergency Download of the TUI

 The TUI must be accessible from the rear, see the Replacement of Parts Instructions
(AXB7-020.841).

Perform the following work steps:


1. Switch ON the system.
2. Wait until the system has booted up and the FL-C remains in the patient list.
3. Press the Reset button ( 1/Fig. 40 Page 45) approx. 3 sec..
 The TUI will brighten.
4. Press the download button ( 2/Fig. 40 Page 45) and keep it pressed for approx. 30
sec. until the Dialog message box ( Fig. 41 Page 45) appears.
 After a wait period, the download process will start.
5. Once the download has finished, the following is displayed:

TUI_1

<< >>

Select

Ysio AXB7-020.816.02.04.02 Page 44 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 45

6. Choose <TUI_1> by clicking on the corresponding arrow button.


7. Confirm the setting with <Select>.
8. Subsequently perform a download of the TUI SW, ( Touch User Interface
(TUI) / Page 42).

Fig. 40: TUI component Fig. 41: Message box

2 1

(1) Reset button


(2) Download button

1.7.3 Generator
A download of the generator is necessary, e.g., if the MCB F80 board (piggyback board on
D600 board) was replaced.
Start the service software.
n Select <Configuration> <Download>
 The PMS Download page appears.
- Mark the Install checkbox H for Polydoros F80.
- Start the download with <Go>.
n After the download is finished, leave the page with <Exit>.
n Press <Home> to return to the service software Main menu.

 No restore of generator configuration files is necessary following a download of the gen‐


erator software.

1.7.4 RAD Controller


A download of the RAD Controller is necessary, e.g. if it was replaced or if there was a SW
update.

 Following a RAD controller download, its configuration must be restored.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 45 of 55 Ysio


11.13 CS PS AX R/F
46 1 Software Installation

 Prior downloading the RAD Controller / CAN Components, make sure an up-to-date sys‐
tem backup exists.

n In the service software Home menu, select <Configuration> <RAD Crtl. Download>
 The following page appears, ( Fig. 42 Page 46).
Fig. 42: Stand Download Manager

n Confirm the backup request with <OK>


n Prior to the download, a reboot of the RAD Controller is necessary to activate the
download mode.
- Trigger a reboot by pressing the <Reboot> button in the following page
( Fig. 43 Page 46).
Fig. 43: RAD Controller Download Manager, download preparation

- Confirm the Internet Explorer pop-up with <Yes>.

Ysio AXB7-020.816.02.04.02 Page 46 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 47

n After the reboot is done, next page appears ( Fig. 44 Page 47).
Fig. 44: Download section

n Press <RAD Controller Download> to start the download.


 The download will start automatically.
 Be patient, this will take a few minutes.
- After the download is completed, following page comes up ( Fig. 45 Page 47).
- A restart of the RAD Controller is required.
 Follow the instructions in the window below, ( Fig. 45 Page 47).
- Switch the system OFF, wait 5 seconds and switch it ON again (within 10 seconds)!
- Then press the <Confirm> button.
Fig. 45: RAD Controller Download Manager, reboot

© Siemens, 2010 AXB7-020.816.02.04.02 Page 47 of 55 Ysio


11.13 CS PS AX R/F
48 1 Software Installation

- Wait for completion. The message RAD Controller up again! Please per-
form a Config Check will be displayed.

Fig. 46: Download finished

n Perform a <Config Check> and / or <AutoConf>, if requested.


n Press <Home> to return to the service software Main menu.
n Select <Backup & Restore> from the SSW main menu.
- Select <Restore> from the tree browser on the left.
 The following screen appears ( Fig. 47 Page 48).
Fig. 47: Distributing files from the host to the components

- From the Select drive list box, select Local, ( 1/Fig. 47 Page 48).
- From the Select package list box, select Distribute files from Host (FL-C) to Com‐
ponents only, ( 2/Fig. 47 Page 48).
- Press <GO>.

Ysio AXB7-020.816.02.04.02 Page 48 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 49

- In the next screen, ( Fig. 48 Page 49), mark the checkboxes H for Stand Con‐
trol Unit (SCU), Collimator Control Unit and DCU2 (DAP Control Unit).
Fig. 48: Restore following a RAD controller download

- Start the restore with <GO>.


- After the restore is finished, leave the page with <Exit>.
- In the following window, confirm the request for the Config Check with <OK>,
( Fig. 49 Page 49).
Fig. 49: Confirming the Config check

© Siemens, 2010 AXB7-020.816.02.04.02 Page 49 of 55 Ysio


11.13 CS PS AX R/F
50 1 Software Installation

n After the Config Check is concluded, press <Exit> to return to the service software
Main menu.

1.7.5 CAN Components


CAN Components have to be downloaded if they have been replaced. Each CAN Compo‐
nent can be downloaded separately.
n In the service software Home menu, select <Configuration> <RAD Crtl. Download>
 The following page appears, ( Fig. 50 Page 50).
Fig. 50: Stand Download Manager

n Confirm the backup request with <OK>


n Prior to the download, a reboot is necessary to activate the download mode.
- Trigger a reboot by pressing the <Reboot> button in the next page.
 The RAD Controller will reboot.
- Then close the Internet Explorer pop-up with <Yes>.

Ysio AXB7-020.816.02.04.02 Page 50 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Software Installation 1 51

- After the reboot is done, the following page appears, ( Fig. 51 Page 51).
Fig. 51: Download section

- Press <Download CAN Components>.


n In the following dialog, the desired CAN Component (or all of them) can be selected
for download.
Fig. 52: Selection of the CAN Component

n Press <Start Download>.


 The download will start automatically.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 51 of 55 Ysio


11.13 CS PS AX R/F
52 1 Software Installation

 Depending on the installed controller HW for the Controller Table lift drive and for the
remote control, a version error may be displayed for this CAN component. Ignore it.

- Once a CAN component download is completed, the screen shown below appears.
( Fig. 53 Page 52)
 Follow the instructions in the window below.
- Switch OFF the system, wait 5 seconds and switch the system ON again (within 10
seconds)! Then press <Confirm>.
- Wait for completion.
The message RAD Controller up again! will be displayed.

Fig. 53: CAN download completed

n Press <Home> to return to the service software Main menu.


Fig. 54: Download finished

n In the service software Home menu, select <Configuration> and perform a Config
Check and Auto Conf, if requested.
n Afterwards, perform a Version Check.

Ysio AXB7-020.816.02.04.02 Page 52 of 55 © Siemens, 2010


11.13 CS PS AX R/F
Changes to Previous Version 2 53

Page 29: added Installing applicable RUH Packages.

© Siemens, 2010 AXB7-020.816.02.04.02 Page 53 of 55 Ysio


11.13 CS PS AX R/F
54 3 List of Hazard IDs

There are no Hazard IDs in this document.

Ysio AXB7-020.816.02.04.02 Page 54 of 55 © Siemens, 2010


11.13 CS PS AX R/F
All documents may only be used for
rendering services on Siemens
Healthcare Products. Any document
in electronic form may be printed
once. Copy and distribution of elec‐
tronic documents and hardcopies is
prohibited. Offenders will be liable
for damages. All other rights are re‐
served.

healthcare.siemens.com/services

Siemens Healthcare Headquarters