0% found this document useful (0 votes)
2K views29 pages

Manual DTL150

The document provides a comprehensive tutorial for the DTL150 DC 150W electronic load, detailing its technical specifications, wiring methods, and operational modes. It includes instructions for calibration, setting protection parameters, and managing discharge limits, along with interface descriptions and product dimensions. Key features include overvoltage, overcurrent, and overpower protections, as well as a user-friendly interface for managing various load settings.

Uploaded by

pfpa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
2K views29 pages

Manual DTL150

The document provides a comprehensive tutorial for the DTL150 DC 150W electronic load, detailing its technical specifications, wiring methods, and operational modes. It includes instructions for calibration, setting protection parameters, and managing discharge limits, along with interface descriptions and product dimensions. Key features include overvoltage, overcurrent, and overpower protections, as well as a user-friendly interface for managing various load settings.

Uploaded by

pfpa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 29

DTL150

DC 150W 2.4 "Color Screen CNC Electronic


Load Usage Tutorial

Important reminder


If U≥120V then P≤60W

01
产品展示
Product exhibition

4P terminal DC power interface


and NTC port


2.4 TFT Set button
主机介绍

02
技术参数
Technical
parameter

Load type: DC CNC electronic load


Working voltage: DC12V equipped with 12V 1A power adapter
Load voltage: DC0~150V
Load current: DC0-20A
Load power: 0-150W (maximum load capacity of 60W above 120V)
Wiring method: A+V+V - A - four wire system, downward compatible with two wire system
Upper limit of electricity metering: 999.999Wh
Capacity measurement upper limit: 999.999Ah
Accumulated timing range: ddd: hh: mm: ss (day: hour: minute: second)
Working mode: four modes CC - constant current mode CV - constant voltage mode CP -
constant power mode CR - constant resistance value mode
Overvoltage protection: Supports maximum 150V overvoltage protection
Overcurrent protection: Supports up to 20A overcurrent protection
Overpower protection: Supports a maximum of 150W over power protection
Voltage limited discharge: also known as low voltage protection, supports, defaults to no
voltage limited discharge, and can be set according to your needs
Limited time discharge: supports up to 99 hours and 59 minutes of limited time
discharge
Radiator: Alloy integrated silent (20 ± 3dBA) Radiator internal temperature ≥ 40 ℃
Automatic operation
External temperature measurement: Connect the NTC 3950 10K 1% temperature sensor
to measure and display the temperature between 0 and 100 ℃
Chinese English interface switching: supported
Accumulated data reset: supported
Voltage and current zero return without load: supported
Voltage open calibration: supported
Current open calibration: supported
Screen brightness adjustment: supported
Screen hibernation standby setting: supported
Computer software online: not supported
Restore factory settings: Supported
Product size: 165 × ninety-five × 80(mm)
Product packaging size: 195 × one hundred and five × 135mm
Product packaging weight: 625g, net weight of host 400g

03
界面介绍
Interface
The interface introduction is for clear display, made by a computer, and may differ from the actual
introduction
display effect. The final display effect is based on the actual product!

1.1Introduction to the top menu bar

2.1Constant current working mode and settable working current (default


working mode)

2.2 Constant voltage working mode and settable working voltage (requires
manual switching of working mode)

2.3 Constant power working mode and settable working power


(requires manual switching of working mode)
2.4 Constant resistance working mode and settable working resistance
(requires manual switching of working mode)

3.1 Load operation status - Off not running (default state)

3.2Load operation status - ON in operation (manual triggering required)

4.1Real time parameters voltage, current, power

4.2 Accumulated parameter capacity, energy

4.2 Real time parameter resistance value

5.1Voltage, current, and power over limit protection (can be set independently)

OVP-Overvoltage protection

OPP-Overcurrent protection
OPP-Overpower protection

6.1Stop discharge voltage (low voltage protection) can be set independently

7.1Limited time discharge --: -- Unlimited time discharge (default state) can be
set independently

7.2Limited time discharge 99:59 (up to 99 hours and 59 minutes) requires


independent setting

8.1 Stop running interface END | End (normal end interface)

8.2 Voltage over limit protection stop over limit warning+preset protection
voltage

8.3 Current over limit protection stop over limit warning+preset protection
current
8.4 Power over limit protection stop over limit warning+preset
protection power

9.1 Menu

04
使用方法
Wiring
method

Click to view the corresponding tutorial

01 02 03
Voltage and current
calibration RESET No return to zero
with no load
Calibrated

04 05 06

Data Clear Adjusting display Set working mode


brightness

07 08 09

Set overrun Stop voltage Limited time


protection setting discharge setting

10 11

Discharge Start load


parameter settings

05
使用演示
Application
examples

Usage demonstration: Using a 4-wire wiring method and CC


constant current mode, set a discharge current of 19.5A to test
the automotive power battery
Voltage and current calibration method
Calibration essential equipment

Prepare a power supply that can set DC 30V3A constant current and constant voltage, and set

the power supply to 30V 3A CCCV mode

Set the power supply to 30V 3A CCCV mode

Connecting the power supply to the load

Connect the power supply to the load using a four wire or two wire system, and

demonstrate the use of two wire wiring.

Enter the system menu

Long press the ON/OFF button for 3 seconds to enter the system menu
Voltage calibration

Click the SET key to switch the cursor to the voltage calibration option

Click the button "+" to confirm the calibration voltage, and a "OK"

prompt will pop up on the screen indicating that the calibration is

complete,

After voltage calibration, the menu displays 30.000V

Current calibration

Click the SET key to switch the cursor to the current calibration option

Click the button "ON/OFF" to start the load, and the current calibration item displays

real-time current

Click the button "+" to confirm the calibration current, and a "OK"
prompt will pop up on the screen indicating that the calibration is

complete,

After voltage calibration, the menu displays 03.000A

Calibration completed, click the SET key to switch the cursor to the exit menu, and
click the "+" button to exit the system menu

—warn—

If there is an error in operation,

the factory settings can be restored!

点击图标查看如何恢复出厂设置
RESET
Enter the system menu
Long press the ON/OFF button for 3 seconds to enter the system menu

Click the SET key to switch the cursor to 08. Reset

Click the "+" button to confirm the restoration of factory settings, and a "OK" prompt
will pop up on the screen indicating that the restoration is complete
Zeroing without load
Empty parameter is not zero
When the load is not started or not connected to the load power supply, real-time
readings of voltage, current, and power indicate no load return to zero

Enter the system menu


Long press the ON/OFF button for 3 seconds to enter the system menu

Click the SET key to switch the cursor to 03


Click the button "+" to confirm resetting, and a "OK" prompt will pop up on the
screen indicating that the recovery is complete

When there is interference in the usage environment, a single reset cannot be made

and multiple operations are required

If multiple operations still fail to return to zero, it can only be counted as a system

error

The interface display after returning to the main interface after clearing the empty

load (the figure shows the display of connecting to a 30V power supply)
Accumulated data reset
Clear historical cumulative data when starting a new cumulative cycle
Including three cumulative data: cumulative operating time, energy, and capacity

Enter the system menu


Long press the ON/OFF button for 3 seconds to enter the system menu

Click the SET key to switch the cursor to 02.Clear cumulative


Click the button "+" to confirm the reset, and a "OK" prompt will pop up on the

screen indicating that the reset is complete

The interface display after returning to the main interface after clearing the empty

load (the accumulated data has been cleared)


Adjusting screen brightness
Enter the system menu
Long press the ON/OFF button for 3 seconds to enter the system menu

Click the SET key to switch the cursor to 06.Display brightness: 8

Click the button "+" or "-" to adjust the brightness


Mode set
Set working mode (default CC constant current mode)
Press and hold the "SET" button for 2 seconds on the main interface to enter the
auxiliary parameter setting mode. Working mode CC flashing

Click "+" to switch to CV constant pressure mode

Click "+" to switch to CP constant power mode


Click "+" to switch to CR constant resistance mode

Switch the working mode according to your own needs, and after switching the working mode,
click "SET" to switch to other items
Or continuously click "SET" to exit the setup
Overlimit protection settings
Press and hold the "SET" button for 2 seconds on the main interface to enter the
auxiliary parameter setting mode. Working mode CC flashing

Click "SET" to switch to the over limit protection setting

OVP voltage over limit protection flashing


Click the "+" or "-" keys to adjust to the target voltage
After completing the adjustment, click the "SET" button to switch to the current
overload protection setting
OCP current over limit protection flashing
Click the "+" or "-" keys to adjust to the target current

After completing the adjustment, click the "SET" button to switch to the power
over limit protection setting

OPP power over limit protection flashing


Click the "+" or "-" keys to adjust to the target power

Set the overrun protection parameters according to your own needs, and click "SET" to
switch to other options after completion
Stop voltage (low voltage protection) setting
Press and hold the "SET" button for 2 seconds on the main interface to enter the
auxiliary parameter setting mode. Working mode CC flashing

Click "SET" to switch to the stop voltage setting

Stop voltage flashing


Click the "+" or "-" keys to adjust to the target voltage

Set the stop voltage (low voltage protection) according to your own needs, and click "SET" to
switch to other options after completion
Limited time discharge setting
Press and hold the "SET" button for 2 seconds on the main interface to enter the
auxiliary parameter setting mode. Working mode CC flashing

Click "SET" to switch to the limited time discharge setting


Default --: -- is unlimited discharge
Flashing for 1 (minute) period of limited time discharge
Click the "+" or "-" keys to adjust to the target time, with a maximum setting of 59
minutes and automatic rounding for 1 hour at the end of 60 minutes

After setting the minute period, click "SET" to switch to the limited time discharge
2 (hour period) setting
Flashing for 2 (hour) period of limited time discharge
Click the "+" or "-" keys to adjust to the target time, with a maximum setting of 99
hours

After setting the discharge time according to your own needs, click "SET" to exit the setting
Discharge parameter settings
Taking CC mode as an example for demonstration, the setting method for other working modes is the same
Click the "SET" button on the main interface to switch and set the cursor position

After the cursor reaches the target mantissa, click the "+" or "-" keys to adjust to
the target current

After adjusting to the target parameters, the settings are completed. Press "ON/OFF" to start
the load or make other settings according to the needs
Operating load
Taking CC mode as an example for demonstration, the starting method for other working modes is the same

Turn on the power and complete the parameter settings

Click the "ON/OFF" button to activate the load status indicator "ON"

When it is necessary to stop the load, click the "ON/OFF" button to stop the load
status indicator "OFF"