UnivFrag
UnivFrag
UnivFrag
Universal Fragmatome ®
Service Manual
ALCON SURGICAL
15800 Alton Parkway
Irvine, California 92618-3818
U.S.A.
Telephone: 714/753-1393
800/832-7827
FAX: 714/753-6614
ii 906-8400-501
Universal Fragmatome®
TABLE OF CONTENTS
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
System Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
System Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
The Fragmatome® System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
The Coagulation System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
The Gas Fluid Exchange System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
The Gas Forced Infusion System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3
Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3
The Surgical System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
Surgical System Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
The GFX and GFI® System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
GFX and GFI® System Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
Display Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
Mode Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
System Status and Messages Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
Primary Parameters Section. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
Multi-Function Key Menu Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
Memory Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
Multi-Function Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
Adjust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
Sub-mode Arrow (----->) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
Audio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
Mode Keys and Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.8
Fragmentation (Frag) Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.8
Coagulation (Coag) Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.9
Footswitch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.10
Handpiece Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.10
Single Use Disposable Paks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.11
Universal Fragmatome® Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.12
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.12
906-8400-501 iii
Universal Fragmatome®
iv 906-8400-501
Universal Fragmatome®
906-8400-501 v
Universal Fragmatome®
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
EMI Shield Gaskets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
Voltage Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2
Universal Fragmatome® Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3
Pressure Transducer Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3
Coagulation Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.4
U/S Driver Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5
Remove and Replace Fiber Optic Lamps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5
Clean Light Source Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5
Troubleshooting Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.6
vi 906-8400-501
Universal Fragmatome®
Explanation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1
906-8400-501 vii
Universal Fragmatome®
LIST OF FIGURES
LIST OF TABLES
viii 906-8400-501
Universal Fragmatome®
The data presented in this manual was current at the time of publication; accordingly, the data in this
manual may not reflect exactly your system configuration.
Alcon Surgical makes no warranties, expressed or implied, that the information contained in this service
manual is complete or accurate. If this manual is used to perform service on the equipment by other than
trained personnel, the user assumes all risks in the use of this manual.
Pay close attention to warnings and cautions in this manual. A WARNING! statement is written to protect
individuals from bodily injury. A caution statement, with the CAUTION heading centered above the text,
is written to protect the instrument from damage.
Caution: U.S. federal law restricts this device to sale by or on the order of a physician.
Alcon Surgical
Technical Services Group
PO BOX 19587
Irvine, CA, USA 92623-9587
All rights reserved. No part of this manual may be reproduced, transmitted, or stored in a retrieval system,
in any form or by any means; photocopying, electronic, mechanical, recording, or otherwise; without prior
written permission from Alcon Surgical.
906-8400-501 ix
Universal Fragmatome®
x 906-8400-501
Universal Fragmatome®
SECTION ONE
GENERAL INFORMATION
The Alcon Surgical Universal Fragmatome® is an ophthalmic The Fragmatome® is designed to promote operator safety
surgical tool for use in posterior segment surgeries. Control of when used properly.
the system is achieved via front panel controls and the • The operator must follow basic safety practices to ensure
surgeon’s footswitch, which is connected to the Universal that no bodily injury or damage to the system occurs.
Microtome Vitrectomy System (UMVS). When connected to a • The user of this system must read this manual carefully and
UMVS, the Fragmatome® has the capability to perform a pars become familiar with all Warnings and Cautions that
plana lensectomy using the Fragmatome® handpiece, and appear within.
coagulate blood vessels using coagulation brushes and • The Fragmatome® is portable, with two recessed hand-
forceps. The Fragmatome® provides gas fluid exchange to holds on the side panels to help grip the system during
infuse an air bubble into the eye, and gas forced infusion to transportation. Care should be exercised when lifting and
maintain accurate control of intraocular pressure; transporting this instrument.
communications need not be established with the UMVS for
these air pump functions. Portable Equipment Note: Prior to moving instrument,
disconnect all cords, cables, accessories, and tubing.
The Universal Fragmatome® offers the following features:
• Ultrasonic fragmentation.
• Bipolar coagulation. SYSTEM FEATURES
• Gas fluid exchange and gas forced infusion: GFX and GFI ®.
• Portable, table-top design. System Self-Test
• Microprocessor control. The Fragmatome® electronics provide an automatic system
• Programmable memories (3). self-test, initiated upon system power-on. Alcon Universal
• Footpedal control of: Fragmatome is flashed on and off in the middle of the display
Power activation in Fragmentation (Frag) mode. while the following components and functions are tested:
Output voltage activation in Coagulation (Coag) mode. • Central Processing Unit (CPU)
• Self-tuning (continuous) piezo-electric ultrasonic handpiece. • Read Only Memory (ROM)
• Audible tone feedback for all functions. • Random Access Memory (RAM)
• Back-lit front panel. • Electronically Erasable Programmable Read Only Memory
(EEPROM)
• Analog to Digital Converter (ADC)
SYSTEM COMPONENTS • 85 Volt functionality
• CPU watchdog timer functionality
Following are the basic system components shipped along • Display pixels
with the electronic console:
After successful completion of the self-test, the system
• Operator’s manual defaults to the Frag mode. If the UMVS is not connected
• Dust cover during power-up CHECK UMVS CONNECTION is displayed.
• Power cord
• Fragmatome® handpiece If the system does not pass the self-test, the system enters a
• Fragmatome® Accessory Pak safe, non-operational mode, and an error message may be
• Universal Fragmatome® to UMVS interconnect cable printed on the display screen identifying the failure. The
exception to this is the pixel test when the user must observe
the display screen during power-up to identify any unlit pixels,
and GFX and GFI® when the system enters a non functional
loop. If there are problems, the local service engineer should
be contacted for assistance.
906-8400-501 1.1
Universal Fragmatome®
Up/Down
Parameter
Ultrasonic Alphanumeric Adjustment
Tuning/Priming Key Display Screen Keys
Infusion Pressure Pressure Frag Mode
Adjust Knob Display Key
Ultrasonic
Air Infusion Infusion Valve Elapsed Time Multi Function Coag Mode
On/Off Key LED Reset Key Keys (4) Key
Air Infusion
23 Tune Standby
Key
On/Off mmHg
Reset
Figure 1-1. FRONT PANEL - The Universal Fragmatome® front panel contains all of the controls, connectors, and
displays required by the operator to perform lens fragmentation surgery. Handpiece activation is
performed through the UMVS footswitch.
1.2 906-8400-501
Universal Fragmatome®
SURGICAL
ALCON LABORATORIES, INC.
Universal
FRAGMATOME®
TYPE B BF
! CAUTION! GEFAHR:
~ 100-120/220-240 V 50-60 Hz 6.3 A MAX
EXPLOSIONSGEFAR. NICHT IN DER BRENNBARER ANÄSTHETIKA VERWENDEN.
POWER
(W)
25
VORSICHT: EINE ZUVERLÄSSIGE ERDUNG KANN NUR ERREIGHT WERDEN, WENN DAS
GERÄT AN EINE KRANKENHAUSSICHERE STECKDOSE ANGESCHLOSSEN WIRD.
CAUTION: GROUNDING RELIABILITY CAN ONLY BE ACHIEVED WHEN EQUIPMENT IS
CONNECTED TO AN EQUIVALENT RECEPTACLE MARKED HOSPITAL GRADE.
0123
VORSICHT: VERBRENNUNGS - UND BRANDGEFAHR - NICHT IN DER NÄHE VON LEITFÄHIGEN
MATERIALIEN VERWENDEN. ELEKTRODENKABEL AUSWECHELN, SOLLTEN
SCHÄDEN SICHTBAR SEIN.
CAUTION: RISK OF BURNS AND FIRE - DO NOT USE NEAR CONDUCTIVE MATERIALS.
RENEW ELECTRODE CABLES UPON EVIDENCE OF DETERIORATION.
T6.3A250V
REPLACE FUSE
AS MARKED.
SERIAL IN /OUT
Figure 1-2. REAR PANEL - The Universal Fragmatome® rear panel contains a power switch, a multi-pin cable
connector, and a variety of warning and identification labels.
906-8400-501 1.3
Universal Fragmatome®
Refer to the UMVS Operator's Manual for the Standard • Fragmatome® Handpiece Connector
Fragmentation System Setup Procedure. The fragmentation handpiece is plugged into this multi-pin
electronic connector.
• Audio Speaker
A speaker behind the front panel grill provides audio
feedback to the operator.
1.4 906-8400-501
Universal Fragmatome®
The GFX and GFI® System GFX and GFI® System Controls
For gas fluid exchange an air pump inside the Fragmatome® is The GFX and GFI® system provides accurate, consistent air
used to provide air pressure for infusion of an air bubble into pressure (0-99 mmHg) for use in gas forced infusion or during
the eye, while the UMVS aspiration system simultaneously gas fluid exchange procedures. The system is controlled from
performs passive displacement of the existing fluid. This the front panel with one knob and four keys. One LED and two
procedure is done following core surgery to provide a display windows provide operator interface (see Figure 1-1).
temporary tamponade for the retina. GFX or GFI® tubing is connected to the air infusion port.
Gas forced infusion is intended to maintain accurate control of • Infusion Pressure Adjust Knob - This rotational knob sets
intraocular pressure by providing regulated air pressure infusion pressure, allowing the user to either increase or
directly into the infusion bottle. This helps to regulate the fluid decrease the amount of pressure delivered.
flow rate.
• Air Infusion On/Off Key - When pressed, this key opens
The UMVS does not need to have power applied in order for and closes the valve that controls the output air pressure.
the Fragmatome's air pump to function. If the Fragmatome® is
powered-up without establishing communications with the • Pressure Display (mmHg) - The requested air pressure is
UMVS, CHECK UMVS CONNECTION is displayed. shown in this display window. When the infusion valve
LED is illuminated, this display represents the actual
A GFX and GFI® power-on self-test is automatically delivered pressure.
performed each time system power is turned ON, and each
time the front panel Air Pump On/Off key is turned ON. Two • Infusion Valve LED - This green LED, located to the right
tones are emitted upon successful completion of the self-test. of the pressure display, illuminates whenever the infusion
Tests performed are listed in the Troubleshooting section of valve is open.
this manual. The timer and pressure displays become
illuminated at the completion of the GFI® self-test, with the • Infusion Timer (MIN : SEC) - Air infusion elapsed time is
timer display showing 00:00 and the pressure display showing shown in this display window. This display automatically
the current pressure. accrues time that delivered infusion pressure is greater than
60 mmHg. It can also be used to manually keep track of
Infusion pressure is controlled with the Infusion Pressure any other timing function when the delivered infusion
Adjust Knob: turning the knob clockwise raises the pressure, pressure is less than 60 mmHg.
while turning the knob counterclockwise lowers the pressure.
A green LED on the front panel illuminates when air pressure • Timer Start/Stop Key - The timer start/stop key is used to
is available from the front panel air infusion port. If the start or stop the infusion timer when the delivered pressure
displayed pressure is UU, available pressure is between 100 is less than 60 mmHg.
and 125 mmHg.
• Timer Reset Key - The timer reset key is used to reset the
The timer display is under manual or automatic control, infusion elapsed time to zero when the delivered pressure
depending upon the output air pressure sensed by the system. is less than 60 mmHg.
When the pressure goes higher than 60 mmHg, the timer
operates under automatic control, with the timer automatically • Air Pump On/Off Key - The gas forced infusion pump
starting from 00:00 (without regard to time accrued, or if the On/Off key is used to either turn the air pump ON or OFF.
timer was previously ON or OFF). The timer continues to
increment each second until the pressure drops below • Air Infusion Port - This port is the tubing connection for
60 mmHg, when it automatically stops at the time accrued. A the gas forced infusion.
tone will sound every integer minute to indicate the high
pressure condition.
WARNING!
Shutting down the Fragmatome® system shuts down
GFI® pressure as well.
906-8400-501 1.5
Universal Fragmatome®
Universal Fragmatome®
®
SURGICAL
Mode Section
System Status
and Primary
Tune Messages Parameters
Section Section
Figure 1-3. DISPLAY SCREEN - The display screen is located in the upper right corner of the front panel. The display contains
eight lines of data, with 42 characters in each line.
1.6 906-8400-501
Universal Fragmatome®
906-8400-501 1.7
Universal Fragmatome®
MODE KEYS AND FUNCTIONS The Fragmatome® presets to aspiration only when entering
Frag mode. To activate fragmentation, the operator must press
The Fragmatome® has three modes of operation: Continuous sideways on the UMVS footswitch’s right momentary switch;
Fragmentation (FRAG), Pulsed Fragmentation (FRAG Pls), to turn fragmentation off, press the left momentary switch (see
and Coagulation (COAG). Each mode and its functions are Figure 1-6). This can be performed with the footpedal in any
described in detail starting with the Continuous Fragmentation position, including the resting position.
mode (see Figures 1-4 & 1-5, and Table 1-4).
Pressing the UMVS footpedal activates aspiration and
Fragmentation (Frag) Mode ultrasonic power to the Fragmatome® handpiece. Linear
The Frag mode offers the surgeon lens fragmentation aspiration activates at minimum vacuum and increases to
capability. The UMVS provides microprocessor control of the maximum vacuum at the bottom of the footpedal's stroke.
fragmentation functions through its footswitch. All Fragmentation activates at the power Limit set by the operator.
fragmentation settings, controls, and alarms are resident in the
Fragmatome®.
Universal Fragmatome®
®
SURGICAL
Frag
Frag HP Not tne
Panel Time 0.0 min
Tune Limit 70 %
Figure 1-4. MULTI-FUNCTION KEYS - Four multi-function keys are used to adjust and store operating parameters in one of
three programmable memories.
MODE KEYS - Two designated mode keys, and one multi-function key, are used to select modes of operation.
COAGULATION
Figure 1-5. MODE FLOW CHART - This flow chart shows the modes of operation, and after the power-on self-test, the paths
taken to access them.
1.8 906-8400-501
Universal Fragmatome®
Pulsed Fragmentation The voltage Limit is adjusted by pressing the Up/Down keys to
The fragmentation function can also be operated in a pulsed the right of the display screen. Pressing the footpedal activates
mode. To enter Frag Pulse mode the operator must press the voltage to the coagulation handpiece. Volts appears on the
FRAG Pulse multi-function key. display screen, and a continuous audible tone is generated
while coagulation voltage is delivered to handpiece.
FRAG Pls
Frag HP Not tne NOTE: For proper operation in Coag mode, the ATFS-601
Panel Time 0.0 min footswitch must have its detent knob in the active position.
Limit 70 %
Pulses 10 p/sec
–––––––––––––––––––––––––––––––––––––––––– WARNING!
FRAG Memory Adjust –––> DM Do not use the coagulation function on patients with
pacemakers. If electrosurgery is used on patients with
implanted cardiac pacemakers or pacemaker electrodes,
be aware that irreparable damage to the pacemaker and
Refer to the UMVS Operator's Manual for the Standard its function may occur and lead to ventricular fibrillation.
Fragmentation System Setup Procedure.
CAUTION
FOOTPEDAL POSITION
Continuous Press sideways and release STANDBY LINEAR ASPIRATION & FRAGMENTATION
right momentary switch
Pulse Press sideways and release STANDBY LINEAR ASPIRATION & PULSED FRAGMENTATION
right momentary switch
Table 1-4. OPERATION BY MODE AND FOOTSWITCH POSITION - The footpedal initiates fragmentation power or coagulation
voltage to the handpieces. Aspiration is always enabled in fragmentation mode; momentary switches are used
to enable and remove fragmentation capability.
906-8400-501 1.9
Universal Fragmatome®
Momentary switches are located on the right and left sides of • Coagulation Handpieces (see Figures 1-8 & 1-9)
the footswitch. To enable the fragmentation function you must
press sideways on the right switch with the side of your foot. Single-use Bipolar Coagulation Brushes are available in a
To remove fragmentation capability, press the left switch. wide variety of configurations: straight, curved, 20-gauge,
23-gauge, tapered, and widestroke. All single-use bipolar
HANDPIECE DESCRIPTIONS accessories are available with and without cables. Also
Alcon handpieces are available for each operating mode of the available are reusable and single-use bipolar cables.
Fragmatome®. Alcon's High Performance Fragmatome®
handpiece is required for fragmentation procedures, and a Bipolar Coagulation Forceps are lightweight and
variety of brushes and forceps are offered for use in the Coag ergonomically designed to reduce hand fatigue, as well as
mode. Following are general descriptions of each (see the to provide precise control and safety. They are available in
Accessories and Parts section of this manual, or consult your high-conductive, non-stick alloy, titanium, or single-use
Alcon representative for a complete selection of all configurations. They are also available with a wide variety
handpieces). of tip styles.
• High Performance Fragmatome® Handpiece (see Figure 1-7) Coagulation Cables (see Figure 1-10) are available in
disposable and reusable configurations, and in two
The HP Fragmatome® handpiece provides superior lens different styles.
fragmentation and is virtually clog free. The handpiece can
perform simultaneous fragmentation & aspiration, or See the Accessories and Parts section for a complete listing
of available products.
Detent knob
Figure 1-6. ATFS-601 FOOTSWITCH - For precise operator control of Universal Fragmatome® functions, the footswitch
contains a footpedal, a detent knob, and two momentary switches.
1.10 906-8400-501
Universal Fragmatome®
Component Quantity
Fragmatome® vacuum tubing 1
3-way stopcock 1
HP Fragmatome® needle 1
Cleaning stylette 1
Needle wrench 1
Directions for Use (DFU) 1
Figure 1-8. SINGLE USE BIPOLAR BRUSH - One example Vitrectomy Surgery Set (8065807101)
of the many styles of bipolar brushes is shown here. This pak is required to use the Fragmatome® in combination
with the UMVS for fragmentation.
Component Quantity
MicroSmooth® probe with tubing 1
Fluid collection cassette 1
Extrusion extension tubing set 1
20 cc syringe 1
3-way stopcock 1
Component Quantity
Infusion line with filter 1
®
GFI® Tubing Set for Pressurized Infusion (8065808001)
This pak contains a tubing assembly to be connected between
the Fragmatome® system's air infusion port, the fluid bottle,
not approved for European Union and an infusion cannula. The assembly is made up of filtered
tubing, a drip chamber, and a 3-way stopcock valve.
Component Quantity
GFI® tubing set 1
Directions for Use (DFU) 1
906-8400-501 1.11
Universal Fragmatome®
The Universal Fragmatome® is controlled by a factory- The Fragmatome® is designed, certified, listed, and labeled to
programmed Intel 8085 8-bit microprocessor running at a meet the requirements of IEC 601-1 and UL 544 Electrical
6.0 MHz clock rate. The microprocessor is configured with Standards for Medical Electrical Equipment. It is classified as
memory space in three separate blocks: 52 K of EPROM, 4 K Class 1 (grounded), Type B equipment.
of EEPROM, and 8 K of RAM. The microprocessor receives
information from the front panel controls and the footswitch. Local, State, or Federal law may restrict the use of this
equipment to, or under direction of, a licensed physician.
The user communicates with the system by way of the front
panel controls, footswitch connected to a Universal
Microtome Vitrectomy System, and front panel display screen.
The alphanumeric display screen presents its data with 8 lines
of 42 characters each, utilizing a 256 X 64 dot matrix vacuum
fluorescent module.
ELECTRICAL POWER:
The Universal Fragmatome® is Class I equipment with protective Ultrasonic Drive: 2.7 mil stroke ±0.5 mil at 42.9 KHz
earth conductor. Power supply protection is through two fuses,
Type F6.3A 250 V. Ultrasonic Pulse: 1 to 15 pulses per second
100 VAC nominal:
120 VAC nominal: Air Infusion Pressure: 0 +5/–0 mmHg to 99 mmHg delivery
220-240 VAC nominal: IEC601-1, IEC601-1-1, IEC601-2-2 pressure range
Grounding: Class I
Applied part: Type B for gas forced infusion and
ultrasonic fragmentation.
Type BF for bipolar coagulation.
Ingress of water: Ordinary
Mode of operation: Continuous
Anesthetics: Not suitable for use in presence of
flammable anesthetics.
Table 1-5. UNIVERSAL FRAGMATOME® SYSTEM SPECIFICATIONS - This table is a quick reference point to identify basic
system specifications, system requirements, and performance figures.
1.12 906-8400-501
Universal Fragmatome®
SECTION TWO
THEORY OF OPERATION
This discussion is intended to give an overview of system The Fragmatome® system only operates when connected to a
operation. It is broken down into five sections: Ultrasonic Universal Microtome Vitrectomy System (UMVS). When an
Control, Frag Communications, Coagulation Control, Front interface communications cable is connected between the two
Panel Interface, and Gas Forced Infusion. This text is written systems, and power is ON for both units, the Universal
in conjunction with the System functional block diagram (see Fragmatome® defaults to Fragmentation mode. Coagulation
Figure 2-1). Most of the signals discussed here travel through can also be accessed at this time.
the Backplane PCB, although, some of them travel through
interconnect cables. The system block diagram is a system The information between the two systems flows in serial
overview, and is not a cabling diagram. format. The information is forwarded to the Front Panel
Control PCB to chip U10, a USART. This chip converts the
serial data to parallel data for interpretation by the CPU.
ULTRASONIC (U/S) CONTROL
The Front Panel Control PCB receives signals FRAGEN and
The Ultrasonic output to the Fragmatome® handpiece is FRAGPU from the UMVS, which are status signals from the
controlled by the U/S Driver PCB and CPU PCB. UMVS footswitch. They identify when the footswitch requests
frag power. They are received by an opto-isolator chip which
The system identifies the installed handpiece through a forwards the request to the CPU as BSTATMVS*.
voltage divider network formed between a resistor installed in
the handpiece and circuitry on the U/S Driver PCB. This The Front Panel Control PCB also receives serial data from
handpiece identification value is sent back to the CPU as the UMVS as ANO (anode) and CAT (cathode) into an optic
USCAL/I. isolator. These signals are forwarded to the USART as RXD
(received data). The USART converts this data to parallel
The U/S Driver PCB is enabled by the USEN signal from the format for CPU interpolation. Data sent from the USART to
CPU. It also utilizes SU* and SD* to adjust the output up and the UMVS is accomplished by the TXD (transmit data) and
down. Signals CS0 and CS1U are a two-bit code from the TXDPU (transmit pull up).
CPU to select between the three Digital to Analog Converter
(DAC) outputs. These outputs adjust frequency span CPU and USART know when to transmit and receive data by
(FREQSPAN), used to coarse adjust the frequency during signals RXINT (receive interrupt) and TXINT (transmit
tuning; frequency select (FREQSEL), used to fine tune the interrupt). These signals are from the USART to the CPU
handpiece; and AMPLITUDE, used to adjust the output interrupt multiplexer, and interrupt the software routine to
power. handle new data. The signals are received by the UMVS opto
isolator as RXLED and RXD, then sent on to the CPU in the
The U/S output is monitored by the ADC on the CPU PCB. UMVS as just RXD.
The U/S Driver sends signals HPV, a representation of the
handpiece voltage; HPI a voltage representation of the The CPU in the UMVS is able to receive and transmit directed
handpiece current; 85VSTAT, a level of the 85V; and serial data without the use of a USART chip.
USDAC, the output of the DAC for adjusting amplitude.
These voltages are compared and analyzed by the CPU in
order to provide a precise ultrasonic output.
906-8400-501 2.1
Universal Fragmatome®
Front
Right SL0-SL2, RL0-RL4 FRAGEN*, FRAGPU, ANO, CAT
Knob
Position
GFI
Control
Tone information Pressure Info
PCB
COAGEN*, PU, PD
Figure 2-1. SYSTEM BLOCK DIAGRAM – This block diagram shows the most significant signals used by the Universal
Fragmatome® to perform its functions. When connected to a UMVS, the system drives the ultrasonic and
coagulation handpieces. Independent of the UMVS the system performs gas forced infusion.
2.2 906-8400-501
Universal Fragmatome®
The UMVS does not need to have power applied in order for
the GFI® system to function, nor does the interconnect cable
need to be connected. If the Universal Fragmatome is powered
up without establishing communications with the UMVS,
906-8400-501 2.3
Universal Fragmatome®
BACKPLANE PCB
The Backplane PCB is the large board mounted on standoffs The following table lists the cable connectors on the
at the bottom of the card cage, providing electronic paths for Backplane PCB, and their associated system interfaces:
the assemblies within the Universal Fragmatome® to
communicate. Four S-100 PCB connectors (A1J1, A1J2, Connector Interface
A1J3, & A1J4) interface the main function PCBs to the A1J5 GFI® Control PCB.
system. These connectors are not interchangeable; each A1J6 Ultrasonic handpiece identification.
specific PCB must be placed in its own specific connector: A1J7 Front panel vacuum fluorescent module
display communications.
Connector PCB A1J8 Front panel vacuum fluorescent module
A1J1 Front Panel PCB. display power.
A1J2 U/S Driver PCB. A1J9 Front right panel overlay.
A1J3 Coagulation PCB. A1J11 Rear panel communications port
A1J4 CPU PCB. (Fragmatome® to MVS).
A1J12 Audio speaker.
CAUTION A1J13 Power supply.
A1J15 Standby/85V enable in power supply.
Applying power to the system with a PCB in the wrong A1J16 Cooling fan.
location can cause serious damage to the system.
FRAGMATOME® COAG
HANDPIECE HANDPIECE
CONNECTOR CONNECTORS
W2
W6
W8
A7P2 A6P1 A5P2 A4P1 A4P2
BACKPLANE PCB
W1
P7 P5
PRESSURE REGULATOR J7 J5
POTENTIOMETER W14 W3
GFI DISPLAY PCB
POWER SUPPLY
W15
Figure 2-2. BACKPLANE PCB – The Backplane provides electronic paths for system communications. The board contains
four slots for printed circuit board insertion, and ten connectors are used for electronic cables.
2.4 906-8400-501
Universal Fragmatome®
Ready - RDY
This input signal is always true (ready function not used).
CONTINUED ON PAGE 2.7
906-8400-501 2.5
Universal Fragmatome®
CLK Bi-directional
CLK RD* Data Buffer
RST 5.5 U19
RD* WR* RD* BAD0-BAD7
RST 7.5
WR* RST BUFFEN Misc
U18
TIMER* Microprocessor SOD FPC*
Programmable ALE
U25
Clock INTR INTA*
U26
SID A8-A14
System RST* FPC* RD*
Clock Output Data
U7 AD0-AD7 RD* FPINT1*
Buffer
WR* ALE*
ALE* U17 WR*
LAD0-LAD7 EPROM U9
HRST RST,RST*
RAM U8 HRST*
RD*
WR* EEPROM U10
ADCCLK EPROM*
RAM*
EEPROM*
2.5V Ref
5.12V
AGND
LAD0-LAD3
USCAL/I
HPV
ERRTIMER (XDCR2)
ADC RD* SU*,SD*
KEYTONE (XDCR1)
Data/Address WR* USEN
HPI U6
Multiplexer HRST SO0
USDAC
U16 WDOG* CS1U
UPRIM ALE U18
USINST* WDEN1
COAGPWR
COAGINST* GNDFRST*
85VSTAT
RSTSTAT PU,PD
ADCCLK
COAGSTAT* COAGEN*
RD* PROG0-PROG2 MSKINT*
ADCSTART
WR* U15 ADC_EOC USERR VSOLEN*
ADCOE*
ADC* BSTATMVS* PSOLEN*
PIA's VENT1EN*
U11,U12 GNDFRSTR*
RD* EPROM*
ADC_EOC Data Buffers REQ 85V
RAM
U2,U3
2.5VREF Power FAULT* EEPROM* U20-U23
5.12V Fault Memory ADC*
Circuitry GNDF Mapping
+15V AD3-AD5 Decoder TIMER*
U28
R19-R24 KEY*
-15V FPC*
RB U14
INTR USWDT
TXINT Interrupt
Multiplexer SID DACWR*
RXINT
U15 PIA1*
MSKINT*
PIA2*
INTA*
CONTREQ*
MUXA
MUXB ENB
MUXC
CONTREQ*
MUXD
ENB
DIRBUFF1
CLK
U1,U5 DIRBUFF2
WR* I/O Control BUFFEN
Register
U4 MUXA
MUXB
MUXC
RST* MUXD
Figure 2-3. CPU PCB – The CPU PCB provides instructions of control to the rest of the system. User requests are
accepted from the front panel. The CPU processes the requests, then issues the appropriate commands. The
responses to the commands are monitored to ensure that the instructions are followed.
2.6 906-8400-501
Universal Fragmatome®
CPU
ADDRESS REGISTER & DECODING ANALOG-TO-DIGITAL CONVERTER
The Address Register & Decoder block provides three major The analog-to-digital converter is an ADC 0816 A/D converter
functions to the system: System Memory-I/O Mapping, Data/ chip. The chip accepts 16 analog inputs. Four address lines
Address Multiplexing, and I/O Control Decoding. select any one of the sixteen inputs for conversion. The
ADC0816 is powered by the 5V supply, and a 5.12 voltage is
System Memory-I/O Mapping generated on the CPU PCB, used as the reference voltage for
PLD U14 is programmed as the System Memory-I/O Map the A/D to ensure consistency in conversions.
Decoder. The eight most significant address bits (A8-A15) are
used by U14 to decode the individual function block Analog Inputs
Since the microprocessor is a digital device, and the real world
addresses. This memory maps the system. This can be used for
is analog, a device is used to convert analog signals into
troubleshooting, meaning that if the CPU locks up, the time-
digital signals which the processor can understand and work
out timer can be disabled, and U14 points to the area causing
with. U6 is a 16-input multiplexed analog-to-digital (A/D)
the lockup. Mapping is accomplished as follows: converter with an 8-bit word for each input. An 8-bit word
provides 256 steps of resolution; therefore, for a reference
Interrupt Traps . . . . . . . . . . . . . . . . . 0000-003F (H) voltage of 5.12 VDC, each step equals 20 millivolts.
EPROM - U9 . . . . . . . . . . . . . . . . . . 0040-BFFF (H)
EEPROM-U10 . . . . . . . . . . . . . . . . . C000-CFFF (H) U/S Handpiece Calibration Resistor - USCAL/I
RAM-U8 . . . . . . . . . . . . . . . . . . . . . . D000-EFFF (H) This signal is used to provide a precision voltage from the
PIA1-U12 . . . . . . . . . . . . . . . . . . . . . F000-F003 (H) calibration resistor which indicates to the processor the
PIA2-U11 . . . . . . . . . . . . . . . . . . . . . F100-F103 (H) maximum DAC position for U/S output. After recognition,
ADC-U6 . . . . . . . . . . . . . . . . . . . . . . F200-F20F (H) this channel is used to monitor the current feedback to indicate
Timer-U7 . . . . . . . . . . . . . . . . . . . . . F300-F303 (H) over-current conditions.
Control Register-U4 . . . . . . . . . . . . . F400 (H)
FPC PCB Interface1-U19 . . . . . . . . . F500 (H) U/S Handpiece Feedback Voltage - HPV
Keyboard-FPC . . . . . . . . . . . . . . . . . F570-F571 (H) Indicates actual voltage to the handpiece. This signal is used to
Sound Chip-FPC . . . . . . . . . . . . . . . . F5B0-F5B1 (H) generate the actual percentage of output to the handpiece. The
Communications-FPC . . . . . . . . . . . F5D0-F5B1 (H) voltage is a linear equivalent of the input voltage to the DAC
Display-FPC . . . . . . . . . . . . . . . . . . . F5E0-F5E1 (H) used to set the output power limit. The actual voltage can be
US Watchdog-U/S DRIVER . . . . . . FC00 (H) up to 20% above the U/S Power set point if set at 10-90, and
US DAC Write-U/S DRIVER . . . . . FD00 (H) up to 10% above the set point if set at 100. A failure occurs
Memory Mapped I/O . . . . . . . . . . . . FFFF (H) when the actual voltage is greater than the set point tolerance.
906-8400-501 2.7
Universal Fragmatome®
CPU
5 Volt Reference - 5V ADC Control Lines
This signal indicates the level of the 5 V power supply. Used There are nine control lines that control the function of the
for comparison purposes. ADC0816 so that interfacing with the system can be
accomplished when requested, and accomplished within the
2.5 Volt Reference - 2.5VREF timing constraints of the system.
Indicates the value of the precision 2.5 V reference. Used to
evaluate ADC performance by comparing this reading to 2.5 Data Select Lines - LAD0-LAD3
volts. These four lines determine which of the sixteen inputs is to be
converted.
Analog Ground - AGND
This signal indicates analog ground value. Used as a reference Address Latch Enable & Start - ADCSTART
for comparison monitoring. The address latch enable and the start lines are tied together.
When the address decoder toggles these lines, a conversion
U/S Output - USDAC event is started. The selected channel to be converted is
This voltage can be used to compare the input value of the selected, sampled, and the information latched.
DAC to the output value of the DAC for verifying the DAC
conversion performance. End of Conversion - ADC_EOC
When the conversion is completed ADC_EOC goes high,
Coagulation Power Output Feedback - UPRIM signaling that valid data is present.
This signal indicates magnitude of coagulation output energy
and is monitored to assure that output energy does not exceed Output Enable - ADCOE*
that which is requested. The processor activates this input which puts the 8-bit output
of the ADC0816 on the data bus to be processed. When the
Coagulation Compliance - COAGPWR processor has read the data, the output enable line is de-
This signal indicates coagulation output and is monitored to activated, restoring the output lines and the bus to their high-
determine whether or not there is output energy to the impedance state.
coagulator handpiece.
Clock - ADCCLK
85 Volt Status - 85VSTAT The ADC 0816 clock input is 1/4 the frequency of the
This signal indicates whether or not the 85 volt source is processor clock output.
present, and is generated on the U/S Driver PCB through a
voltage divider. During power-up this signal is used to verify Reference Voltage - 5.12V
the operation of the 85 V switching circuitry located in the The 5.12 V provides the reference voltage used to assure
power supply. consistent conversions, and is generated from a 2.5 V
precision source.
2.8 906-8400-501
Universal Fragmatome®
CPU
DATA I/O CONTROL Coagulation Enable - COAGEN*
This control signal has a dual role of enabling/disabling
Most of the data I/O is handled by programmable peripheral coagulation output and resetting the coagulation watchdog
devices (PIA). The function of the PIA’s is to interface timer (when pulsed).
peripheral equipment to the processor. The configuration of
the PIA’s is programmed by system software per factory U/S PCB Installed - USINST*
mandate (PFM) so that normally no external logic is necessary This control signal indicates that the U/S Driver PCB has been
to interface to peripheral devices. The three data ports on the installed, and therefore the Frag mode is a valid selection. If
PIA’s are normally configured either as input or output ports this PCB is not installed, then the Frag mode is not an
with their own directional data buffer IC’s. available option to the user.
U/S DAC Directional Control Lines - SU*, SD* Watchdog Timer Reset Status - RSTSTAT
These control signals indicate the counter direction for pulsing This control signal indicates the status of the watchdog timer.
up or down and clearing the counter or putting it in standby. This signal is monitored during power-up when the watchdog
timer functionality is being verified.
U/S Enable Control Line - USEN
This control signal enables the U/S output, and simultaneously Coagulation Fault Status - COAGSTAT*
enables the 85 V required to drive the U/S handpiece. This signal indicates that an over voltage condition exists on
the Coagulation PCB.
U/S Frequency/Voltage DAC Control Line - CS0
This control signal, along with CS1U, is used to select the CONTINUED ON PAGE 2.10
required DAC on the U/S Driver PCB before DAC positioning
information is sent.
906-8400-501 2.9
Universal Fragmatome®
CPU
Programmable Peripheral Interface Device #2 (PIA2) 85V Request REQ85V
Ports A, B, & C on PIA #2 (U11) utilize data buffers U2, U3, This control signal enables the 85 V generation circuitry in the
& U20 respectively. The direction of data on these buffers is power supply, provided no fault conditions exist. This signal is
fixed and is not under software control. U2 is strictly an input used upon power-up to verify the circuitry operation which is
buffer, while U3 and U20 are strictly output buffers. The ports utilized in the Frag and Coag modes.
are initialized at power-up to be either input or output ports
with default states assigned to the corresponding control U/S Driver PCB Watchdog - WDOG*
signal. Like PIA#1, output signal levels are established by This control signal allows the processor to monitor the U/S
resistor pull-ups/downs (as required) until PIA initialization is watchdog timer and disable handpiece power in the event of a
complete and BUFFEN = TRUE. time out.
2.10 906-8400-501
Universal Fragmatome®
CPU
FRONT PANEL PCB INTERFACE & BUFFERING MEMORY
This Functional block consists of connector A4J1, data buffer There are three types of memory used on the CPU PCB:
U17, bi-directional data buffer U19, and logic gates U18. All RAM, EPROM, and EEPROM. U8 is a 64K static RAM
switch, display, and control communications between the device which provides 8K bytes of volatile storage for system
Front Panel PCB and the processor goes through this interface. scratch pad calculation memory. EPROM U9 is a 512K device
that provides 64K bytes of program memory. EEPROM U10
Front Panel Interface Bi-directional Data Buffer is a 64K device that provides 8K bytes of user programmable
Bi-directional data buffer U19 interfaces the Address/Data bus memory. The EEPROM is used when selecting the default
to and from the Front Panel PCB. Logic gates U18 set the language option, or when setting new default limits for the
enable bit. selected mode of operation.
Multiplexed Address/Data Bus - AD0-AD7 RAM is enabled by address decoder U14, and the data send/
This is a dual function bus. During the second and third timed receive is controlled by the processor RD* and WR* control
event (clock cycles) this bus is used for data transfer. During signals.
the first timed event this is an 8-bit microprocessor address
bus for the lower 8-bits of I/O addressing as described in EPROM is also enabled by address decoder U14, or on request
Address Decoding. from the control register U4 ENB line in response to a
software interrupt. Data send is controlled by the RD* signal
Read - RD* from the processor.
The RD* command from the processor determines the
direction of data flow to/from the Front Panel PCB. EEPROM is addressed the same way as EPROM. The
exception is that U14 selects the EEPROM, and the device
Front Panel Control Select - FPC* also has a write function controlled by the WR* signal from
This address decoder output signal enables data interchanges the processor. During the write function U10 sends a RDY/
between the CPU and Front Panel PCB. BSY* (RB) interrupt to the processor via the interrupt
multiplexer telling the processor to hold all activity frozen on
Output Data Buffer the data bus until U10 has completed writing data into
Output buffer U17 is constantly enabled to interface the memory.
processor control signals to the Front Panel PCB.
Address Bus - A8-A15
Address Latch Enable - ALE* This 8-bit microprocessor output bus addresses the high order
This signal clocks the above mentioned octal D flip-flop. memory and I/O functions of the system as described in
Address Decoding.
Front panel Select - FPINT1*
This signal enables the Front Panel PCB, and is simply a Multiplexed Address/Data Bus - AD0-AD7
buffered form of FPC*. This is a dual function bus. During the second and third timed
event (clock cycles) this bus is used for data transfer. During
Read Control - RD* the first timed event this is an 8-bit microprocessor address
This signal provides the appropriate control to synchronize bus for the lower 8-bits of I/O addressing as described in
communication between the processor and the Front Panel Address Decoding.
PCB.
Read Control - RD*
Write Control - WR* The READ output signal is negative true when the
This signal provides the appropriate control to synchronize microprocessor wants to accept data from the I/O devices or
communication between the processor and the Front Panel memory.
PCB.
Write Control - WR*
Hardware Reset - RST*, RST The WRITE output signal is negative true when the
Both senses of this signal provide for asynchronous clear or microprocessor wants to send data to the I/O devices or
reset of the directed circuitry on the Front Panel PCB. memory.
906-8400-501 2.11
Universal Fragmatome®
CPU
when processor address information is present on the Address/ INTERRUPT MULTIPLEXER
Data bus. Octal tri-state D flip-flop U16 uses ALE to latch the
lower eight address bits onto the true address bus (LAD0- PLD U15 is programmed to function as the software interrupt
LAD7). This clears up the Address/Data Bus allowing the encoder to the processor. The system interrupts are divided
processor to send or receive data from these lines during the into two groups: vectored interrupts, and polled interrupts.
next clock cycles.
Vectored interrupts are defined as those that warrant
RAM Select - RAM* immediate action by the software. Most of these are associated
This signal from the memory decoder selects and enables this with system hazards or otherwise require immediate action.
device for operation. When one of these interrupts is present, the interrupt request
signal INTR is issued to the processor. The processor
EPROM Select - EPROM* acknowledges the request with INTA (interrupt acknowledge).
This signal from the memory decoder selects and enables this INTA outputs the appropriate interrupt vector address onto
device for operation. AD3-AD5 for automatic interrupt service branching by the
8085 µP.
EEPROM Select - EEPROM*
This signal from the memory decoder selects and enables this Polled interrupts are monitored once each main control loop.
device for operation. Address/Data bits AD0-AD3 are used to poll the requested
interrupt and are sent to the interrupt multiplexer through
EEPROM Ready/Busy* - RB register control U4 as MUXA-MUXD. The polled interrupt
This signal is used to indicate the status of the EEPROM. If state is monitored through the Serial Input Data line (SID). If
RB* (busy) is indicated, the EEPROM is still in the write the SID state for the polled interrupt indicates an error
cycle. RB (ready) is issued when writing is complete. condition, the software responds accordingly.
2.12 906-8400-501
Universal Fragmatome®
CPU
EEPROM Ready/Busy - RB WATCHDOG TIMER CONTROL
This is a polled interrupt that indicates the status of the
EEPROM. A low level indicates the EEPROM is busy (still in The Watchdog Timer Control circuitry is comprised of clock
the write cycle). oscilllator U31 (and misc), µP monitor U24, and PLD U27.
Communications Transmit - TXINT Because of the requirement for WDOG timeout assurance
This is a vectored interrupt used to indicate when information before 500 mS, but not before µP loop cycle time of
can be sent through the communications port. approximately 300 mS, the targeted WDOG period of
400 ±50 mS requires a precise frequency source, which is
Communication Received - RXINT provided by U31.
This is a vectored interrupt used to indicate when information
sent through the communications port has been received. The watchdog timer function and the 5V failure detection are
provided by U24. PLD U14 is used to combine the signals
Ground Fault Interrupt Mask - MSKINT* logically to generate the required outputs to disable or reset
This control signal allows the masking of the Ground Fault the system. U14 also provides the ability to disable the
Interrupt when it occurs so that the processor can take the watchdog circuitry from the rest of the PCB functions, while
appropriate action without being continuously interrupted. still providing an output that can be monitored to indicate
watchdog status. This enables the processor to test the circuits
Interrupt Request - INTR functionality upon initialization, and also for field service
This general interrupt signal has the lowest priority of the five testability.
available interrupts. This interrupt signal is sent to the
processor indicating that a system error has occurred that Although the WDOG is initially disabled to provide ample
requires action. INTR is filtered for noise. time to test µP resources such as ROM, RAM, TIMERS, etc;
once it is enabled it cannot be disabled again until the next
Interrupt Acknowledge - INTA* power OFF/ON cycle. The watchdog alarm asserts HRST until
This signal is issued by the µP during the instruction cycle the next POR cycle.
after an INTR is accepted, to enable the 3-bit latched output
error code from the interrupt multiplexer onto the data bus. One peculiarity of WDOG timer U24 is that once WDOG
timeout has occurred, WDOG strobes issued during the
Interrupt Selection - MUXA-MUXD WDPOR pulse do not have the desired effect of resetting the
This 4-bit code is used to select any polled or vectored WDOG. Thus, the µP must verify WDI has reset the timer by
interrupt. monitoring WDOG via RSTSTATA.
906-8400-501 2.13
Universal Fragmatome®
CPU
Watchdog Enable (latched) - WDEN4 85V ENABLE CIRCUITRY
This signal is a latched form of WDEN3, which once set,
enables the WDOG until the next POR, PBRST, or WDTR1. The 85V Enable Circuitry is comprised of PLD U13, LED
DS1, and transistor Q5. Although originally designed to be
Hardware reset - HRST, HRST* present only when requested by the functions that require
Both active high/low forms of this signal are provided to force 85 volts, which include ultrasound and coagulation, this
the CPU hardware to the initialize state. HRST latches on circuitry now enables 85 volts when requested by the µP, and
WDOG timeout until POR, PBRST, or WDTR1 clear turns OFF only in the event of fault conditions. Typically, the
WDEN4. µP sets up for 85V enable at power-up, to ensure sufficient
U/S Driver PCB warm up time.
Reset Status - RSTSTATA
This signal is used both by the µP and during manufacturing Ultrasound Output Enable - USEN
test to monitor status of various WDOG/Reset signals. This signal enables the U/S Driver PCB. 85 volts is disabled if
USDIS is true during USEN.
CPU Reset #2 - CRST2
This input indicates when the µP is being reset, either by Ultrasound Disable - USDIS
external hardware (via CRST1), or by a µP-initiated reset This signal disables the 85 V in the event of a U/S fault
instruction. condition during USEN.
2.14 906-8400-501
Universal Fragmatome®
CPU
SYSTEM CLOCKS handpiece. In Frag mode, it controls the pulsing of the
handpiece by turning the U/S power ON and OFF at the
There are two clocks on the CPU PCB: system clock U26, and specified rate to produce the pulsing action.
programmable clock U7. These clocks provide the
synchronized timing required by the system in order to Restart Interrupt 7.5 - RST 7.5
function properly. This software interrupt controls the operation of the scheduler.
The programmable clock is set up to generate this interrupt
System Clock every 25 mS to control and check the task queue. The task
The system clock takes the CLK signal from the processor queue control includes checking for a null entry on the queue,
(1/2 the crystal frequency) and divides it into the individual entering a task on the queue, removing a task from the queue,
frequencies required by the various function blocks of the and clearing the queue.
system.
CONTINUED ON PAGE 2.16
Clock - CLK
This is the main system clock output from the processor. It is
used to synchronize all timed events within the system. The
clock timing is 1/2 the crystal frequency.
Clock Frequency Q2 - Q2
Q2 provides the required timing to the ADC.
Clock Frequency Q7 - Q7
Q7 provides the required timing to the programmable clock.
Programmable Clock
The programmable timer has three independent counter/clocks
under software control. A binary count is placed on the data
bus and is read into the addressed counter. The counter then
counts down from the specified number to zero, and sets a
counter-out bit. The counter-out bits are received by the
processor as software interrupts.
906-8400-501 2.15
Universal Fragmatome®
CPU
POWER FAULT DETECTION CIRCUITRY ±5.12 VOLT REFERENCE SUPPLIES
The Power Fault Detection Circuitry verifies the integrity of The ±5.12 V reference supplies generated on the CPU PCB
the ±15 V and ±5.12 V analog supplies. Quad comparator U28 originate from the 2.5 V output of precision low voltage
forms a hardwired OR gate of the inputs. If the voltages reference supply Q1. U29, U30, Q3, Q4, and associated
measure out of the defined tolerance, an error exists and the discrete components comprise the amplifier circuitry required
output state changes. to generate the ±5.12 voltages.
2.16 906-8400-501
Universal Fragmatome®
The Front Panel PCB contains the circuitry to perform the Address Register
following functions: U4 is a tri-state octal D-type flip-flop and is used to separate
• Bi-directional communication between the CPU and Front the functions of data and address bussing. U4 latches the
Panel PCBs. information on the address bus when requested by the
• Address Register. processor. This clears up the address/data bus, allowing the
• Fragmatome Interface Communications. processor to send or receive data during the next clock cycles.
• Keyboard Encoding.
• Display Interface. Data Bus - DB0-DB7
• Audio Generation and Feedback. This bus information provides the required data to enable the
• CPU To GFI® Interface device requested by the processor.
906-8400-501 2.17
Universal Fragmatome®
906-8400-501
Figure 2-4. FRONT PANEL PCB – The Front Panel PCB is the interface between the Front Panel Overlay Switches/
VENT1EN* U5
Quad PMPVLV
Driver BISTATUS
STATUS* to
Inverting U19 GFI Control PCB
PSOLEN* Buffer
VSOLEN* +24V
U12 Q5,Q6
24VRTN
from
CPU AD0-AD7 Bi-Directional DB0-DB7
Data Buffer
Bi-Directional
RD* U2 Buffer BDB0-BDB7
RD*1 DIR
U13
LA4* OE
RD* RD*1
WR* Data Buffer WR*1
ALE*
U1 FPINT1*B LA4* LA4*
FPINT1* BRD*
ALE*1 LA0 LA0
IRQ* Address BWR*
LA5* Data Buffer to
Register RST BCS*
U4 VFM Display
LA7* U18
RD*1 BC*/D
LA6*
WR*1 BRST
Bit Rate 2.4576 MHz 2.4576 MHz
JP1-JP4 Generator TXINT
2.18
Universal Fragmatome®
Front Panel PCB
Fragmatome® Interface Communication Reset Control - RST1
The serial communications functional block consists of a This control signal forces the USART into the idle mode
programmable bit rate generator (U14), USART (U10), and where it will remain until initialized again with a new set of
opto-isolators (U15 & U17). control words.
The programmable bit generator provides the necessary clock Data Control Bit - LA0*
signals for digital data transmission. The four rate select inputs This control signal, along with the RD* and WR* inputs,
are externally set for 9600 baud based on the crystal frequency informs USART to accept or provide either a data character, a
supplied of 2.4576 MHz. The crystal frequency is buffered control word, or status information via the data bus.
through this device and used to supply the necessary
frequency to the audio circuitry. Receive Data - RXD
A composite serial data stream is received by the receiver
The USART is used as a peripheral, and is programmed by the control logic on this pin.
8085 processor to communicate in serial data transmission
technique. The USART receives serial data streams and Transmit Data - TXD
converts them into parallel data characters for the processor. The transmit control logic outputs the composite serial data
The USART signals the processor when it has completely stream on this pin.
received or transmitted a character and requires service.
The opto-isolator chips provide the necessary electrical Transmit Interrupt - TXINT
isolation required for receiving data while maintaining the This interrupt signals the processor that the transmitter is
required TTL compatibility to interface with the rest of the ready to accept a data character.
circuitry.
Receive Interrupt - RXINT
Baud Rate Selection - JP1-JP4 This interrupt signals the processor when the receiver buffer is
External configuration selects baud rate of the bit rate ready with an assembled character for input to the processor.
generator.
Transmitter Clock - TXCLK*
Crystal Frequency - Y1 The transmitter clock controls the serial transmission rate.
Input frequency from external crystal which determines
multiplying factor for baud rate generation. Receiver Clock - RXCLK*
The receiver clock determines the rate at which the incoming
Clock Output - 2.4576 MHZ character is received.
Buffered crystal frequency output.
Anode - FRAGPU
Bit Rate - Z Return path for opto-isolator (footswitch integrity).
Selected baud rate output.
Cathode - FRAGEN*
Clock2 - 307.2 KHZ Input of the opto-isolator for receiving incoming footswitch
Divided clock frequency equal to 1/8 of crystal frequency. integrity information.
906-8400-501 2.19
Universal Fragmatome®
Front Panel PCB
Keyboard Encoding Display Interface
Keyboard encoding and multiplexing for all of the front panel The Universal Fragmatome® system contains an OEM
overlay switches is performed through the 8279 (U16), a display. The vacuum fluorescent dot matrix display (VFM
programmable keyboard interface device (the front panel display) appears just like a dot matrix printer to the processor.
overlay switches act as the keyboard). U16 is programmed by The display interfaces to the processor via connector A1J9 on
the CPU via data bus DB0-DB7. The data bus is bi-directional the Backplane PCB to the FPC PCB. U13 and U18 make up
to receive commands from, and transmit data to, the processor. the display interface circuitry. Data buffer U13, enabled by
The direction of data flow is a function of the RD*1, WR*1, LA4* from the address register, interfaces the data bus to/from
and chip enable (LA7*) control inputs. the processor. The RD* signal controls the direction of data
flow. Buffer U18 interfaces the control data to the display.
The front panel overlay switches are configured in a keyboard
matrix with the select input lines controlled by the scan line Data Bus - DB0-DB7
outputs of U16. The return line inputs of U16 are connected to 8-bit bi-directional data bus to the display.
the scan lines through the front panel overlay key switches via
connector A1J9 on the Backplane PCB. Switch activity within Read Control - RD*1
U16 is stored in a FIFO RAM. When U16 receives This control signal determines the direction of data flow
information from the switches it issues an interrupt request to between the display and the processor. This control signal
the processor. The processor responds with a read command to enables the display data buffer to send data from the external
bi-directional data buffer U2, and a read command to the bus.
keyboard encoder. Keyboard information is passed through
the encoder and data buffer back to the processor. Enable Control - LA4*
This control signal enables the output of bi-directional data
Bi-directional Data Bus - DB0-DB7 buffer U13. This signal enables the display when requested by
All data and commands between the processor and the 8279 the processor.
are transmitted on these lines.
Write Control - WR*1
Read Control - RD*1 This control signal enables the display data buffer to receive
This control signal enables the data buffer to send data to the data from the external bus.
external bus.
Data Control - LA0*
Write Control - WR*1 This control signal indicates whether the inputs and outputs
This control signal enables the data buffer to receive data from are interpreted as commands or data.
the external bus.
2.20 906-8400-501
Universal Fragmatome®
Front Panel PCB
Data Bus DB0-DB7 CPU to GFI® Interface
This is an 8-bit bi-directional data bus. These signals provide feedback to the GFI® Control PCB to
provide system operational mode information, burn-in mode
Read Control - RD*1 operational information, and operational mode tone
Status of chip is read. acknowledgement.
Synchronization Signal - SH
This signal is used to control the synchronization of
information sent from the device to be converted.
Serial Output - MO
This is the serial audio output signal to the D/A.
906-8400-501 2.21
Universal Fragmatome®
2.22 906-8400-501
Universal Fragmatome®
+10.24V
+5.12V U1 FreqSpan
FREQSPAN Amplifier
U10
DACWR*
FREQSEL VCO
U2
Watchdog
Timer WDOG CS00, CS10
USWDT Feedback Summing
H/PI Amplifier Amplifier
U6 DAC
D0-D7 H/PV
U5/14 U5/7
U9
SU* PLD
Counter USDIS
SD* Control
CS0
from CPU U8
CS1U
USEN
GNDFRST*
RST
USENB
GNDSENS1
USSIGNAL
Lock
Amplifier T1
LOCK VCA Power U/SHV
U5/1 VCAOUT USOUT
Amplifier
AMPLITUDE U3 USCAL/I
+5V Q1,Q3,Q4 U/SHVRTN
HNDPC
Recognize USCAL
USCALR
Circuitry
U1
Output Switch U4
GND Fault
Detection
Circuitry GNDF
GNDFRSTR U7
Figure 2-5. U/S DRIVER PCB – The U/S Driver PCB uses control commands from the CPU PCB to configure itself to
deliver the required drive signal directly to the handpiece. The processor monitors the output to ensure the
commands issued are being properly followed.
906-8400-501 2.23
Universal Fragmatome®
U/S Driver PCB
Voltage Controlled Oscillator and Feedback Amplifier Output Switch
VCO U2 has three summed inputs. Once the handpiece has been The output signal from the VCA is switched by U4, a high
recognized by the processor, the optimum frequency is selected speed analog switch. The open/closed function of this switch
and the output of DAC 1 (U9) becomes the selected input, is set by the USENB from the PLD. A second portion of the
setting the VCO to approximately 45 KHz. When the handpiece switch selects the monitor function of the processor. Initially
is locked in and running, the output of the feedback amplifier set to monitor the handpiece calibration resistor, once the
circuit fine tunes the VCO for maximum handpiece efficiency. handpiece has been recognized, this portion of the switch then
selects the current feedback loop to monitor the fine tuning of
Handpiece Voltage - H/PV the handpiece.
This signal represents the handpiece feedback voltage from
the primary side of the transformer. The actual voltage can be
VCA Output - USOUT
up to 20% above the U/S Power set point if set at 10-90, and
This signal is switched to the power amplifier when USENB is
up to 10% above the set point if set at 100. A failure occurs
asserted.
when the actual voltage is greater than the set point tolerance.
2.24 906-8400-501
Universal Fragmatome®
U/S Driver PCB
Ground Fault Detection Watchdog Timer Control
The output of the handpiece has potentially hazardous Watchdog timer circuit U6 is designed to time-out in
voltages. For this reason the integrity of the handpiece 400 mS ±84 mS unless reset by a rising edge on the watchdog
grounding is monitored. The high voltage return signal timer input. The watchdog output is hardwired to the
(U/SHVRTN) is coupled to the input of flip-flop U7 and PLD processor for status monitoring, and it is logically coupled to
U8. If the U/SHVRTN opens, the clamped high voltage will the USENB signal. If the watchdog timer times-out, analog
set the flip-flops. The processor will acknowledge the fault, switch U4 is not allowed to be asserted, which opens the
disable the output by resetting the DAC counter and opening output and closes the ID path creating an over-current
the output switch, and reset the circuitry by issuing the condition. This condition de-tunes the handpiece and clears
GNDFRST* and GNDFRSTR* signals. The flip-flop in PLD the DAC counters.
U8 provides the redundant circuitry necessary for single fault
conditions. Watchdog Strobe - USWDT
This signal is the input to the watchdog timer and must
Handpiece Ground Sense -U/SHVRTN transition within 400 mS or a time-out condition will exist.
This signal represents the handpiece return voltage, and is at
ground potential unless all of the grounds are removed. Watchdog Output - WDOG*
This signal is sent to PLD U8 where it is gated with the USEN
Ground Fault Redundant Reset - GNDFRSTR* and ground fault latch output to create the USENB signal
This signal is sent by the processor to reset flip-flop U7 as which drives analog switch U4.
soon as the ground fault condition is removed.
Handpiece Recognition
U1 and its associated passive components comprise the
handpiece recognition circuitry. This circuitry is used by the
processor to identify the handpiece. Since each handpiece has
its unique operating characteristics, a precision resistor is
installed during manufacturing to identify the handpiece to the
system. The resistor completes a voltage divider circuit to
identify to the processor the maximum setting for the output
power.
906-8400-501 2.25
Universal Fragmatome®
2.26 906-8400-501
Universal Fragmatome®
PU,PD,PU*,PD*
U2
CAUTEN*
One-Shot U2 Level EN
WATCHDOG Converter
U9
TIMER
U1
COAGEN*
PU,PD
EN*
U6 COAGPWR
4-bit
Counters U-PS
D0-D7
U3,U4 DAC U-DAC
U5
EN U6 U-CON
Q1,Q2
L1
Oscillator
U8 PWM
One-Shot Switcher I-PS
U7
EN U9 U10,Q4
CAUTEN
EN U11 EN* SD
U6
U-PS
85V
Control COAGSTAT*
Circuitry
HP CAUT1
Control
U8,U9,U10 Circuitry CAUT2
Q4
Output Level
DRIVE Converter UPRIM
Circuitry (to CPU)
U12,U13
Figure 2-6. COAGULATION PCB – The Coagulation PCB generates coagulation output, with the output being monitored
and adjusted by the CPU PCB. The CPU generates a COAGEN* signal that enables the PCB, and utilizes PU
and PD for power adjustments up and down, respectively.
906-8400-501 2.27
Universal Fragmatome®
2.28 906-8400-501
Universal Fragmatome®
GAS FORCED INFUSION SYSTEM The signal derived by transducer XD2 is digitized by the ADC
(U13, channel 1). The signal derived by transducer XD1 is
The Gas Forced Infusion (GFI®) system housed in the also digitized by the ADC (channel 2). The ADC output, an 8-
Universal Fragmatome® is designed to provide accurate, bit word, is the address of EPROM U14 (see Table 2-1). The
consistent pressure (0–99 mmHg) for use in gas forced 256 possible addresses (0000H to 00FFH) represent a
infusion or during fluid/gas exchange procedures. The GFI® numerical value of the pressure detected (in mmHg) by
system consists of the GFI® Control PCB (see Figure 2-7), transducer XD1 with the valve energized and transducer XD2
GFI® Display PCB, GFI® air pump, GFI® pressure regulator, with the valve shut. The programmable gate array is set up
and front panel keys. The following control function circuitry such that when the valve is energized only channel 2 is used;
makes up the GFI® system: conversely, with the valve shut only channel 1 information is
used. The EPROM has been coded such that its 8-bit output
• Valve Control represents two 4-bit nibbles, each in binary coded decimal.
• Pressure Display Control These nibbles represent the MSB and LSB of the displayed
• Error, Tone Detection, and Pressure Sensing pressure and are routed to the programmable gate array which
• Power On 7-Segment Display Test and Display Flashing converts the binary coded decimal into 7-segment display
• Timer Control data. This data is routed to the GFI® Display PCB via
• Timer Display interconnect ribbon cable W1.
• Power Up
• Processor Interface Error, Tone Detection, and Pressure Sensing Circuitry
• +5.12 V Regulated Voltage Creation
Potentiometer/Upline
Unless otherwise specified, all components are located on the In order to ensure that the system is accurately detecting the
GFI® Control PCB. sensed pressure, two sets of backups have been designed into
the system. First, to ensure that the pressure regulator has not
Valve Control Circuitry failed wide open, upline transducer XD2 is compared against a
potentiometer. The potentiometer output voltage represents the
When in operation, the GFI® system transfers pressurized air regulated output pressure. If the sensed value of XD2 exceeds
to the environment. Valve K1 is placed in line to provide the ≥0.5 V more than the voltage level represented by the
capability of removing the pressurized air from the potentiometer’s position, the output of comparator U6 will
environment. When power is first applied to the Fragmatome change to a TTL high, sending an interrupt (SETERR) to the
system, the pressure pump is OFF. When power is first programmable gate array. This will set U5–61 high, putting a
applied to the pump, K1 is de-energized. This allows the +2 V signal at A2J1–8 (ERRTIMER). The signal present at
operator to adjust the pressure set point without regard to the A2J1–8 is sent to the system microprocessor via the ADC on
downstream load. The logic is designed such that consecutive the CPU PCB. The +2 V signal present at this ADC channel
presses of the valve control front panel momentary switch instructs the processor to sound the alarm tone.
activate or shut OFF K1. This toggling action is controlled by
programmable gate array U5. A green LED, located on the As a backup to the comparator, the output of differential
front panel, is illuminated whenever K1 is energized. amplifier U2 is digitized by the ADC (channel 4) and checked
by the programmable gate array against predefined high and low
To energize the valve, press the front panel valve key. This
limits. The digitized data becomes an address in EPROM U14.
action turns Q1 (TIP120) ON, thus supplying a ground
The information is sent to the programmable gate array for
potential to valve K1. U9 is an optocoupler, which is in place
processing. (Refer to Tables 2-1 and 2-2 for digital code data.)
to isolate the +24 V circuitry from the rest of the GFI® system
circuitry. When the optocoupler is OFF (signal VO to pin 3 is
POT/UL comparator circuitry generates two types of errors
high), the output is pulled high which in turn energizes the
with associated tone intervals. With valve K1 energized, if the
valve (a high signal on the base of a TIP120 puts the collector
failure is due to the potentiometer signal being greater than the
of Q1 at 24 VRTN potential).
UL signal, then the signal present at TP8 (ERRTIMER) will
be 3.0V (high level concern). Both U5-57 and U5-61 will be
Pressure Display Control Circuitry
set and a tone will sound every 2 seconds. If the failure is due
The pressure (in mmHg) is continuously displayed on the front to the UL signal being greater than the potentiometer signal,
panel through the use of a 2–digit, 7–segment display (U7 & then only U5-61 will be set (2.0V present at TP8, low level
U8 on the GFI® Display PCB). If K1 is shut OFF, the pressure concern) and a tone will sound every 30 seconds.
displayed represents the upline pressure detected by transducer
XD2 on the GFI® Control PCB. If K1 is energized, the Upline/Downline
pressure displayed represents the actual pressure detected by The signal from upline transducer XD2 is compared with the
transducer XD1 on the GFI® Control PCB. The signal signal from downline transducer XD1. Differential amplifier
detected by transducer XD1 is also used to control the timer. U7 calculates the difference between the two signals, then
comparator U6 compares the value against a fixed 2.0V level.
906-8400-501 2.29
Universal Fragmatome®
+5.12 V
32.768 KHz
CIRCUITS 32.768 KHz
+5.12 V OSCILLATOR
U10,U11
U16
Q2
POT/UPLINE
POT/UPLINE COMPARATOR
PRESSURE PRESSPOT
REGULATOR POT- UL>>POT
DIFF INTEG
POT SETTING ERROR SETERR
AMP
DETECT LED DRIVE DATA
2.5VREF U3 U6 (DIGØON*-DIG5ON*)
DIFF POT/UL
AMP
1.25VREF U2
GAIN UL
UP
ADJUST
U4
ADC
U13
UPLINE/DOWNLINE
DOWNLINE
GAIN DL
OFFSET TRANSDUCER
ADJUST
ADJUST XD1 DOWN
U4
+5.12 V
REF
+2.5 V CLOCKS
GND
ADRD
3 ADADRØ-ADADR2
Figure 2-7. GFI® CONTROL PCB – The Gas Forced Infusion system provides accurate, consistent pressure (0–99 mmHg)
for use in GFX and GFI® procedures. The GFI® system consists of the GFI® Control PCB, GFI® Display PCB,
GFI® air pump, GFI® pressure regulator, and front left overlay switches. The GFI® Control PCB is the control
center of the GFI® system, receiving data from, and issuing commands to, the other GFI® components.
2.30 906-8400-501
Universal Fragmatome®
GFI®
The offset of the differential amplifier is fixed at +2.5 V; As with the previously described circuit, to back up the
therefore, as soon as the downline transducer output voltage comparator the output of differential amplifier U8 is digitized
exceeds the upline transducer output voltage by at least 0.5 V, by the ADC (channel 3), becoming an address in the EPROM.
the comparator output changes to a TTL high, sending an The information is sent to the programmable gate array for
interrupt (ULDLERR) to the programmable gate array. This processing. (Refer to Tables 2-1 and 2-2 for digital code data.)
sets U5-61 high, putting a +2 V signal at A2J1–8. The signal
present at A2J1–8 is sent to the system microprocessor via the Note: If K1 is OFF, the upline transducer/downline transducer
ADC on the CPU PCB. The +2 V signal present at this ADC comparison circuitry will be disabled; i.e., the programmable
channel instructs the processor to sound the alarm tone. gate array will not process any errors from this circuitry.
The potentiometer/upline transducer comparison is
validated whenever power is applied to the pump.
ADDRESS ADDRESS
LOCATIONS LOCATIONS
0000 00FF Upline and downline pressure information (for pressure display). 1000 10FF Part of walking one's test (initialization routine)
0100 01FF Determines when delivered pressure rises above 60 mmHg. Verifies +5.12V, DGND.
0200 02FF Error checking for UL/DL signal. 1100 20FF Not used.
0300 03FF Error checking for POT/UL signal (with valve closed). 2100 21FF Part of walking one’s test (initialization routine).
0400 04FF Error checking for PRESSPOT signal. 2200 31FF Not used.
0500 05FF Error checking for +2.5V signal. 3200 32FF Part of walking one’s test (initialization routine).
0600 06FF Not Used. 3300 42FF Not used.
0700 07FF Error checking for POT/UL signal (with valve open). 4300 44FF Part of walking one’s test (initialization routine).
0800 08FF Part of walking one’s test (initialization routine – verifies +2.5V). 4500 7EFF Not used.
0900 09FF Determines when delivered pressure falls below 59 mmHg. 7F00 7FFF Part of walking one’s test (initialization routine).
0A00 0FFF Not used.
LOW HIGH
ERROR RANGE VALID RANGE ERROR RANGE
ADC INPUT DATA = FEH DATA = FCH DATA = FFH NOTATIONS
906-8400-501 2.31
Universal Fragmatome®
GFI®
ADC Integrity +5.12V/Ground Integrity
To ensure the ADC integrity, a +2.5 V reference is digitized at Finally, the ADC digitizes both a +5.12 V signal (scaled down
U13-1 (channel 6). The source of this reference voltage is to +2.28 V using 0.1% resistors R46 & R47) and a ground
independent of the +2.5 V ADC reference. The signal present signal. The +2.28 V signal is present at channel 8, and the
at channel 6 is +1.25 V (half of +2.5 V). The resistors used ground signal is present at channel 7. These signals are only
(R48, R49) are rated at 0.1% to maintain signal integrity. The checked during power on self test (POST).
digitized data becomes an address in the EPROM. The
information is sent to the programmable gate array for Power On 7-Segment Display Test
processing. If the reference is ever found to be outside the and Display Flashing Circuitry
tolerance window, an error flag will set and the alarm tone
will sound. (Refer to Tables 2-1 and 2-2 for digital code data.) Each time power is applied to the GFI® system or pump power
is applied, a blanking and lamp test occurs to ensure the
Error Detection integrity of each LED. The maximum displayed time is 99:59.
When an error is detected the microprocessor sounds the alarm When this time is reached, the timer stops and the display
tone until the error is cleared. If the error is due to the upline/ flashes. To reset the display to zero, press the Reset key. This
downline circuitry, the tone will stop as soon as the valve key only functions if the downline pressure sensed is
shuts because, as stated previously, the programmable gate <60 mmHg.
array does not process upline/downline comparison data with
the valve shut. If the error is due to the upline/potentiometer Timer Control Circuitry
circuitry, the tone will continue to sound until the error is
resolved or the pump is turned OFF. U16 output (32.768 KHz crystal oscillator) is the system host
frequency, which is then divided down using high speed CMOS
Over/Under Pressure 14-stage binary counter U15. This IC outputs the primary
When the delivered pressure meets or exceeds 60 mmHg a frequency divided by 2, 4, 8, 16 — 32768. Of the available
timer located on the front panel clears its display (to read frequencies, 2048 Hz, 1024 Hz, 512 Hz, 256 Hz, 128 Hz,
00:00), then begins incrementing the time (updated every 4 Hz, & 2 Hz are inputted into programmable gate array U5.
second). At each minute interval (“xx:00”) a signal is sent to
the microprocessor, which in turn generates a short, audible There are two methods of engaging the timer: press the Timer
Start/Stop key (manual, conditional), or increase the actual
tone.
delivered downline pressure to at least 60 mmHg (automatic,
conditional). The circuit is designed such that each press of the
Potentiometer Integrity
Timer Start/Stop key will start or stop the timer; although, this
To ensure regulator integrity, a potentiometer is mechanically is a conditional response. If the delivered pressure is
attached. The electrical signal is compared to the upline ≥60 mmHg, then pressing the key has no affect on system
transducer output signal (see NOTE below). To verify response. Increasing the delivered pressure to at least
potentiometer integrity, its output (PRESSPOT) is digitized by 60 mmHg automatically starts the timer from 00:00–in other
the ADC (U13-2, channel 5). The input to the ADC is buffered words, the display timer will reset to 00:00 and continue
through a differential amplifier (U21). With the cable incrementing at the time the delivered pressure crosses the
connected, the output of U21 will be within the range 0.2V to 60 mmHg threshold. Conversely, lowering the delivered
2.0V (approximate). If the cable is disconnected, U21 output pressure below 59 mmHg stops the timer at the displayed time.
will be 0.0V. If the potentiometer fails open, PRESSPOT will
rise well above the +2.0V (to +5.0V). The digitized data In order for the programmable gate array to determine what
becomes an address in the EPROM. The information is sent to the delivered pressure is, the signal derived by transducer XD1
the programmable gate array for processing. If the reference is is digitized by the ADC (U13, channel 2). The digitized data
ever found to be outside the tolerance window, an error flag becomes an address in the EPROM. The information is sent to
will set and the alarm tone will sound. Refer to Tables 2-1 and the programmable gate array for processing. Once the detected
2-2 for digital code data. pressure is determined to be ≥60 mmHg, a different set of 256
addresses in the EPROM (0900 to 09FF) is then used to
NOTE: R6 and R8 were selected such that when the determine when the pressure falls below 59 mmHg. Of these
potentiometer is calibrated, its range of voltage output at available 256 addresses, 0900 to 0984 represent a pressure of
the wiper (PRESSPOT) is ≈ 0.21 V to 2.0 V. This is less than 59 mmHg, and any address above 0984 represents a
approximately the range of the transducers used pressure of ≥59 mmHg. As before, the information is sent to
the programmable gate array for processing.
(900 mV/PSI + 250 mV offset). With a range of
0-100 mmHg (100 mmHg ≈ 2 PSI), the output range of the
To summarize:
transducer is 250 mV to ≈ 2.05 V. To properly calibrate
• The Timer Start/Stop key toggles the clock ON and OFF only
Ω (full range =
the potentiometer, its range is ≈ 6.75 KΩ if delivered pressure is <60 mmHg. If delivered pressure is
Ω Ω
10 KΩ). High end is 7.5 KΩ and low end is 0.75 KΩ Ω . Using ≥60 mmHg, pressing the key has no affect on the clock.
Ohm's Law: • If delivered pressure is raised to ≥60 mmHg, the timer
[5.12(7.51K)]/18.46K ≈ 2.0V and [5.12(.76K)]/18.46K ≈ 0.21V automatically begins incrementing from 00:00.
• If the delivered pressure is lowered to 58 mmHg, the timer
automatically stops incrementing.
2.32 906-8400-501
Universal Fragmatome®
GFI®
Timer Display Circuitry Processor Interface
This circuit is designed so that the display shows minutes and Outputs
seconds from 00:00 to 99:59. If 99:59 is reached, the display Four signals (U5 pins 61, 57, 58, & 60) perform all the
flashes. At power up, and after the lamp and blanking tests, communication between the GFI® Control PCB and the
00:00 is displayed. There are two ways to start the timer. Press microprocessor, located on the CPU PCB. The signals are ERR,
the Timer Start/Stop key, or raise the delivered pressure to at TONE, KV, and KIV. ERR generates the error tone, TONE
least 60 mmHg. The circuit design also stops the clock generates the high pressure timer tone, KIV generates the invalid
automatically when the detected pressure drops to 57 mmHg. key tone, and KV generates the valid key tone. In order to prevent
multiple configurations of the PCBs being used by both the
Located on the GFI® Display PCB are six 7-segment displays Universal II and the Universal Fragmatome®, the four signals are
(four timer displays, two pressure displays). These displays sent to the processor via an ADC on the CPU PCB. Two inputs to
are controlled by the programmable gate array. During normal the ADC are available (KEYTONE, ERRTIMER). The four
operation each display is multiplexed; in other words, they are signals have been split into pairs: KIV & KV, ERR & TONE.
not illuminated 100% of the time. This design lowers the Summing amplifier U4 is used to generate different output
current requirements while maintaining adequate visibility for voltages to represent the different messages (see Table 2-3).
the user. The multiplexed rate is 512 Hz for the time display,
and 256 Hz for the pressure display, which prevents the If TONE is high, +1 V is present at its channel input
viewer from seeing any flickering. A display is turned ON (ERRTIMER). If ERR is high, +2 V is present. If both TONE and
when its control signal (DIG0ON* - DIG5ON*) is set low. ERR are high, +3 V is present. No signal is represented by 0 V.
Only one of these signals may be low at a time. Using The range to the ADC is broken into four segments: 0.0V to 0.5V,
DIG4ON* as an example, when this signal is a TTL low, Q1 0.51V to 1.50V, 1.51V to 2.50V, and 2.51V to 3.50V.
turns ON (low signal present at the base of a PNP transistor),
bringing its collector to the emitter reference minus the C-E The programmable logic array provides circuitry to prevent
voltage drop of ≈0.7 V. The resultant +4.3 V is present at U7- multiple tones from occurring simultaneously. Also, if an error
12, which is the common anode for each segment of this tone is to be generated, it locks out any other tone. The software
display. Whichever segment driver signal (labeled A through has been set up to wait until each tone has been sounded before
G) is low while DIG4ON* is low will cause that particular determining whether or not to generate another tone.
segment to illuminate. All of the timing is handled by the
programmable gate array. As an example, if the pressure
VOLTS KEYTONE COMMENT ERRTONE COMMENT
displayed is currently 23 and DIG4ON* is low, the
programmable gate array sets signals A, B, D, E, & G low in 1V KIV Generates invalid key tone TONE Generates high pressure tone
every 60 seconds as long as
order to display a 2. delivered pressure remains
above 60mmHg
Power Up Circuitry 2V KV Generates valid key tone ERR Generates error tone every 2
seconds
906-8400-501 2.33
Universal Fragmatome®
GFI®
+5.12 V Regulated Voltage Supply Automatic Start/Stop
Whenever the output pressure exceeds 60 mmHg, the timer
The +5.12 V reference supply generated on the GFI® Control automatically resets to 00:00 and starts again. The clock
PCB originates from the +2.5 V output of precision low continues to increment until the output pressure drops below
voltage reference supply U10. U11, Q2, and the associated 60 mmHg. At this time, the clock stops and holds the
discrete components comprise the amplifier circuitry required displayed time. If the output pressure again rises above
to generate the +5.12 voltage. Note that this is an identical 60 mmHg, the timer clears and starts again. Note that
circuit to the one located on the CPU PCB. The reason a whenever the delivered pressure is ≥60 mmHg, pressing the
duplicate circuit has been generated is because a redesign of Timer Reset key or the Timer Start/Stop key has no effect on
the backplane would be necessary. the clock.
GFI® Timing
The GFI® PCB set has a dedicated timer that has been
designed to perform the following functions:
• Manual start/stop
• Automatic start/stop
• Manual reset (returns time displayed to 00:00). This is
conditional — for details, read the following:
Manual Start/Stop
The Timer Start/Stop key on the front panel allows the user to
start or stop the clock when the delivered pressure is less than
60 mmHg. If the delivered pressure is ≥60 mmHg, pressing
this key has no effect—the clock continues to increment.
2.34 906-8400-501
Universal Fragmatome®
SECTION THREE
PARTS LOCATION & DISASSEMBLY
This section of the service manual instructs the service 1 Remove Top Cover
engineer on how to locate, access, and remove various
subassemblies and parts from the Universal Fragmatome®. In 1.1 Turn system power OFF. This is done by turning the
addition to the written instructions, several drawings are ON/OFF switch to the OFF position and unplugging the
included to help perform the disassembly procedures. AC power cord from the power input module (both
located on the rear panel, see Figure 3-2).
When following these instructions, references to right and left
are made when facing the front of the console. NOTE: Throughout this document “Turn system
power OFF” refers to this method of removing AC
Replacement of the following items is performed by doing the power from the system. “Turn system power ON”
steps in reverse order. When connecting cables, refer to refers to applying AC power by use of the same
interconnection diagram on page 6.2. method in reverse order.
CAUTION
®
e
om
at
m
ag
Fr
al
ers
niv
U
By
Stn
g
RG
ICAL Coa
SU
e
Tun
!
et
Res
g
Fra
Hg
mm
n
sio
Infu
Air /Off
On
ff
n/O
: SEC pO
MIN Pum
Air
Figure 3-1. REMOVAL OF TOP COVER – To remove top cover from Universal Fragmatome® console simply loosen two
screws, lift rear of cover up, slide cover back, and lift cover up and off.
906-8400-501 3.1
Universal Fragmatome®
2.3 Remove three 2 mm allen screws from each side of rear 3.2 Unscrew captive screw at left end of PCB containment
panel (six total screws. see Figure 3-2). bracket, then lift bracket up and out of card cage frame
(see Figure 3-3).
2.4 Remove two 2 mm allen screws from lower section of
rear panel (see Figure 3-2). 3.3 Remove Front Panel Control PCB.
2.5 Disconnect fan cable connector A1P16 from the 3.3.1 Disconnect cable connectors A7P2 from Front
Backplane PCB. Panel Control PCB, P2 from Coagulation PCB,
and A4P2 from CPU PCB.
2.6 Disconnect Comm Port cable connector A1P11 from
the Backplane PCB. 3.3.2 Lift up on levers at ends of Front Panel Control
PCB to extract PCB from the Backplane PCB.
2.7 Lift rear panel up and away from console.
3.3.3 Lift Front Panel Control PCB up and out of card
cage.
Step 1.2
2.5 mm allen screws
(two total screws)
Step 2.3
2 mm allen screws
(six total screws)
Figure 3-2. REMOVAL OF REAR PANEL – Twelve screws must be removed to release rear panel from Universal Fragmatome®
console. In addition to the screws, electronic cable connectors must also be unplugged.
3.2 906-8400-501
Universal Fragmatome®
3.5.1 Disconnect cable connectors A7P2 from Front 3.6.1 Disconnect cable connectors A4P1 and A4P2
Panel Control PCB, P2 from Coagulation PCB, from CPU PCB.
and A4P2 from CPU PCB.
3.6.2 Lift up on levers at ends of CPU PCB to extract
3.5.2 Lift up on levers at ends of Coagulation PCB to PCB from the Backplane PCB.
extract PCB from the Backplane PCB.
3.6.3 Lift CPU PCB up and out of card cage.
3.5.3 Lift Coagulation PCB up and out of card cage.
REAR PANEL
Comm
Port
A1J16
A6J1
A7J2
A4J1
A1J11
A1J5
A2J1
Lamp PCB
Auxiliary Output PCB
A1J14
GFI CONTROL PCB
A1J13
J5
Control PCB
J12
Coagulation PCB
Front Panel PCB
Backplane PCB
GFI
A1J15
A1J10
Output PCB
CPU PCB
Air Pump
A2J2
A1J6
A2J3
A1J9
A1J8
A5J2
A4J2
Regulator
A1J7
FRONT PANEL
Figure 3-3. LOCATOR DIAGRAM FOR UNIVERSAL FRAGMATOME® SUBASSEMBLIES – As a visual aid you can refer to
this locator diagram to help remove and replace subassemblies in the Universal Fragmatome® console.
906-8400-501 3.3
Universal Fragmatome®
4 Remove Card Cage and Backplane PCB 5.9 Using 3 mm allen wrench, remove ground wire from
Backplane PCB that comes from U/S handpiece
4.1 Remove all PCB’s from card cage as instructed in step 3. connector.
4.2 Remove ten cable connectors from Backplane PCB (see 5.10 Disconnect GFI® pressure tube from front panel luer.
Figure 3–3. A1J10, A1J14, and A1J17 are not used).
5.11 Carefully lift front panel up and away from console.
4.3 Unlock three cable ties and remove cables from left
side of card cage frame.
6 Remove VFM (Vacuum Fluorescent Module)
4.4 Disconnect two ground straps from Backplane PCB Display Assembly
with 3 mm allen wrench. (One ground goes to
handpiece connector, the other to GFI® air pump 6.1 Remove card cage (steps 4.1 through 4.6) OR remove
bracket. See Figure 3-3.) front panel (step 5).
4.5 Loosen four captive screws at corners of card cage 6.2 Disconnect cable connectors DIS-CON (CN1) and DIS-
securing cage to bottom of console. PWR (CN2) from VFM Display PCB.
4.6 Carefully remove card cage assembly from console and 6.3 Remove one 2.5 mm allen screw from each corner of
set on flat surface. VFM Display PCB (four total screws).
4.7 Remove remaining 3 mm allen screws securing Backplane 6.4 Lift VFM Display Assembly up and away from front
PCB to the card cage (there are 17 total screws). panel.
5 Remove Front Panel 7.1 Remove card cage (steps 4.1 through 4.6) OR remove
front panel (step 5).
5.1 Remove top cover as instructed in step 1.
7.2 Remove seven 5.5 mm (7/32”) nuts and washers
5.2 In power supply assembly, disconnect three fiber optic securing front right panel overlay to front panel.
cables from Lamp PCB, and one fiber optic cable from
Output PCB (carefully disconnect with small flat head 7.3 With all four fiber optic cables disconnected from
screwdriver). power supply (see step 5.2), lift front right panel
overlay away from front panel, while carefully sliding
5.3 Disconnect tubing from input and output ports on GFI® fiber optic cables and ribbon cable through their front
pressure regulator. panel slots.
5.6 Unlock three cable ties and remove cables from left
side of card cage frame.
3.4 906-8400-501
Universal Fragmatome®
8.1 Remove top cover as instructed in step 1. 10.1 Turn infusion pressure adjust knob to mid point.
Remove end cap from infusion pressure adjust knob
8.2 Disconnect cable connectors A2P1, A2P2, and A2P3 and loosen 5/16" nut inside knob. Slide knob off shaft.
from GFI® Control PCB.
10.2 Remove front panel as instructed in step 5.
8.3 Disconnect tubing from upper port of transducer XD1,
and tubing from upper port of transducer XD2. 10.3 Remove three 3 mm allen screws securing regulator
assembly to front panel. The lower screw also secures a
NOTE: To help when installing GFI® Control PCB, ground strap to the side bracket.
label locations of tubing before disconnecting.
10.4 Lift GFI® pressure regulator assembly away from front
8.4 Disconnect tubing from inner and outer ports of panel.
solenoid valve K1. Label locations of tubing before
disconnecting.
11 Remove GFI® Display PCB
8.5 Unscrew captive screws at bottom of brackets at each
end of PCB, then lift PCB, with brackets attached, up 11.1 Remove front panel as instructed in step 5.
and out of console.
11.2 Disconnect cable connector A3P3 from GFI® Display
8.6 Squeeze plastic tabs on metal brackets and remove PCB.
brackets from GFI® Control PCB.
11.3 Remove six 2.5 mm screws securing GFI® Display PCB
NOTE: Refer to tubing diagram on console assembly to front panel, then lift GFI® Display PCB away from
drawing when installing GFI® Control PCB. front panel.
9 Remove GFI® Air Pump Assembly 12 Remove Front Left Panel Overlay
9.1 Remove top cover as instructed in step 1. 12.1 Remove GFI® Pressure Regulator Assembly as
instructed in step 10.
9.2 Disconnect pressure tubing from input port on GFI®
Pressure Regulator. 12.2 Remove GFI® Display PCB (steps 11.2 & 11.3).
9.3 Disconnect tubing from upper port of transducer XD1, 12.3 Disconnect W8 Coagulation cables from terminal lugs
and tubing from upper port of transducer XD2. on coag connectors, then remove spanner nuts securing
coag connectors to front panel overlay.
NOTE: To help when installing GFI® Control PCB,
label locations of tubing before disconnecting. 12.4 Remove large spanner nut securing U/S handpiece
connector to front panel overlay (nut is on front side of
9.4 Disconnect tubing from inner and outer ports of panel).
solenoid valve K1. Label locations of tubing before
disconnecting. 12.5 Remove 10 mm nut securing pressure output luer fitting
to front panel overlay.
9.5 Disconnect cable connector A3P4 from GFI® Display PCB.
12.6 Remove seven 7/32” nuts & washers securing front-left
9.6 Disconnect ground strap from air pump assembly (see panel overlay to front panel, then lift front-left panel
Figure 3-3). overlay away from front of console, while carefully
sliding fiber optic cable and ribbon cable through their
9.7 Cut tie wraps securing ferrite beads to cable clamps on front panel slots.
side of power supply.
CAUTION
9.8 Remove four 7 mm nuts securing base of GFI® air pump
assembly to chassis, then lift pump out of console.
Fiber optic cables are fragile and must be handled
with care to avoid damage. Do not apply undue
NOTE: Do not remove four nuts securing pump and
force when removing overlay.
its rubber mounts to base of GFI® air pump assembly.
906-8400-501 3.5
Universal Fragmatome®
13.4 Disconnect P5, P7, and P12 from Output PCB in power
supply assembly.
3.6 906-8400-501
Universal Fragmatome®
SECTION FOUR
SERVICE TEST PROCEDURE
GENERAL INFORMATION
906-8400-501 4.1
Universal Fragmatome®
4.2 906-8400-501
Universal Fragmatome®
SECTION FIVE
MAINTENANCE & TROUBLESHOOTING
INTRODUCTION REQUIREMENTS
The Maintenance & Troubleshooting section of this service • Universal Microtome Vitrectomy System (UMVS) with
manual contains the instructions used when making needed footswitch, gas supply hose, power cord, and UFrag/UMVS
adjustments to the Universal Fragmatome® System. These interconnect cable.
instructions can be used when the system fails to pass the • Standard service tool kit containing:
Service Test Procedure (STP), or following the replacement of • Combination wrenches (standard and metric).
parts that require calibration. These instructions are only • Ratchet with sockets (standard and metric).
written to calibrate the components within the Universal • Nut drivers (standard and metric) and nut driver extension.
Fragmatome® console. • Allen wrenches (ball end and straight, standard and metric).
• Screwdrivers (straight blade and Phillips).
DEFINITIONS • Hemostats.
• Potentiometer adjustment tool (tweaker).
ADC: Analog-to-Digital converter. • Fragmatome® Accessory Pak (1021 HP).
CW: Clockwise. • Vitrectomy Surgery Set (8065807101).
cc/min: Cubic centimeters per minute. • GFI® Tubing Set (8065808001).
CCW: Counterclockwise. • ScopeMeter® (Fluke 91, or equivalent). ScopeMeter® is a
Coag: Coagulation. registered trade mark of the John Fluke Mfg. Co., Inc.
DM: Default memory on front panel display. • Digital Pressure Meter (DPM, BIO-TEK DPM-III, or equivalent).
DPM: Digital pressure meter. • Cautery Load Box.
Display: An electronic vacuum fluorescent module
(VFM) located on the front panel. Its electronic SPARE PARTS
display is used for visual alpha-numeric
communications with the user. To properly service the Universal Fragmatome® it is
Frag: Fragmentation or Fragmatome®. recommended that you carry the spare parts listed in Table 5-1.
GFI®: Gas Forced Infusion These spare parts are intended to help you through the
GFX: Gas Fluid Exchange preventive maintenance procedure. This is a suggested list,
MIN:SEC Infusion timer LED display on front panel which although you should add to it as necessary. Diagnosing a
displays elapsed time in minutes (MIN:) and customer's problem over the telephone prior to going on-site
seconds (:SEC). will further dictate what parts to take on a service call.
mmHg: Air pressure measured in millimeters of mercury.
Mode keys: Keys pressed to enter the Frag and Coag modes.
p/sec: Pulses per second. Description Part Number QTY
Prime: Key pressed to fill the tubing set and handpiece
with fluid prior to the start of surgery. Gasket, EMI shield 775-014 12 inches
PSI: Pounds per square inch. Gasket, EMI shield 775-010 12 inches
Reset: Keys pressed to set ultrasonic elapsed time to 0.0 Filter, air pump 699-004 1
Fitting, nylon union 893-358 2
min, or to set infusion timer to 00:00.
Potentiometer, W14 840-1022-501 1
StnBy: Key pressed to quickly turn operating power Off.
Lamp, fiber optic illuminator 143-033 3
STP: Service test procedure.
Tune: Key pressed to tune Fragmatome® handpiece. Table 5-1. RECOMMENDED SPARE PARTS - This is a
TP: Test point on an electronic circuit for attaching a suggested spare parts list, although you should add to it as
test lead/Scopemeter probe. TP stands for necessary prior to going on-site to perform repairs.
trimpot on Coagulation PCB .
U/S: Ultrasonic. EMI SHIELD GASKETS
VFM: Vacuum Fluorescent Module (display).
EMI shield gaskets are designed to provide good ground
contact between parts that have variable spacings, and are
required for the Universal Fragmatome® to comply with 601-1
requirements. The gaskets are located inside the console (see
item 44 on page 7.8, and item 23 on page 7.12). The front
gasket has teeth to ensure good contact between the top cover
and front panel. These gaskets must be maintained; if they fall
off or become damaged, they must be replaced.
906-8400-501 5.1
Universal Fragmatome®
3. Disconnect P5 of cable W10 from Output PCB in 9. Connect jumper between J7 pins 3 and 2. Connect
power supply (see Figure 5-1). ScopeMeter® probes to J5 pins 11 (+) and 9 (-). Turn
system power ON and verify +85.0 VDC (±1.0 VDC).
4. Connect ScopeMeter® probes to J5 pins 4 (+) and 3 (-).
Turn system power ON and verify +5.0 VDC (±0.2 VDC). The +85.0 VDC is not adjustable.
The +5.0 VDC adjustment is located on the power 10. Turn system power OFF. Remove jumper and
supply's Control PCB at R23. ScopeMeter® probes. Connect P5 & P7 to Output PCB.
OUTPUT PCB
11
R23
P12/J12
1
J12
P7/J7
1
J7
R23
R17
R25
R18
Figure 5-1. POWER SUPPLY LOCATOR DIAGRAM - Shown above are the test points and adjustment points on the power
supply's Motherboard, Control PCB, and Output PCB.
5.2 906-8400-501
Universal Fragmatome®
DIGITAL
1.1.1 With system power OFF, use T–fitting and tubing to PRESSURE
METER
connect digital pressure meter (DPM) in line between (DPM)
1.1.2 Press Air Pump On/Off key to turn air pump ON, and
verify Infusion Valve LED is NOT illuminated. SURGICAL
®
Universal Fragmatome®
1.1.3 While observing front panel pressure display and DPM, Air Infusion
01 Tune
On/Off mmHg
shown below. The display and DPM should stay within Frag Coag StnBy
listed tolerances at each setting. 01:35
MIN : SEC Reset
1.1.5 Turn knob CW until the display reads 99. The DPM
TP2
should stay within 2 mmHg; if not, adjust R14 and
repeat step 1.1.3. R14
TP7
GFI CONTROL PCB
TP6
1.1.6 Press Air Pump On/Off key to turn air pump OFF. R13
TP1
TP5 TP4
TP4
continued on next page R12
TP3
R11
Coagulation PCB
Front Panel PCB
TP2 TP8
U/S Driver PCB
TP1
CPU PCB
TP9
TP3
906-8400-501 5.3
Universal Fragmatome®
1.2.1 Remove DPM, T-fitting, and tubing from between Y- 2.1 Set Universal Fragmatome® system securely on top of
fitting and transducer XD2. Re-connect tubing between UMVS, and connect together with interconnect cable.
Y-fitting and XD2. Connect DPM to front panel Verify footswitch and gas supply hose are properly
infusion port with tubing (see Figure 5-4). attached to UMVS.
1.2.2 Press Air Pump On/Off key to turn air pump ON, and 2.2 Plug cautery load box into coag handpiece connectors
press Air Infusion On/Off key to deliver air pressure to on Fragmatome® front panel. Connect ScopeMeter®
infusion port (LED illuminated). probes across load box terminals. Insert fluid collection
cassette into UMVS and turn both systems ON.
1.2.3 While observing front panel pressure display and DPM,
2.3 Press the Frag key on the UMVS, and press the Coag
turn Infusion Pressure Adjust Knob to the settings
key on the Fragmatome® to enter coagulation mode. Set
shown below. The display and DPM should stay within the power Limit to 100%.
listed tolerances at each setting.
2.4 Fully depress UMVS footpedal and quickly verify the
FRONT PANEL DIGITAL output is 38.7 ±3.8 Vrms (98.6 to 120.5 Vpp) and the
PRESSURE DISPLAY PRESSURE METER
period is 340 ±17 KHz (2.8 to 3.1 µS). Release
00 mmHg 00.0 ±1 mmHg footpedal. If these values are not within tolerance
proceed to next step.
19 mmHg 19.0 ±1 mmHg
If successful proceed to step 1.2.6. If not successful 2.5 To gain access to potientiometers on Coagulation PCB,
proceed to next step. remove PCB containment bracket from top of card
cage. Fully depress footpedal and quickly do the
1.2.4 Turn Infusion Pressure Adjust Knob fully CCW. following:
Adjust R13 on GFI® Control PCB (see Figure 5-3) until
front panel display flickers between 00 and 01, then • Adjust TP4 on Coagulation PCB (see Figure 5-3)
turn R13 one full turn CCW (360°). until front panel displayed power is less than
70 volts.
1.2.5 Turn knob CW until the display reads 99. The DPM
should stay within 2 mmHg; if not, adjust R11 and • Adjust TP2 on Coagulation PCB until the period on
repeat step 1.2.3. ScopeMeter® is 340 ±17 KHz (2.8 to 3.1 µS).
1.2.6 Press Air Infusion On/Off key to remove air pressure • Adjust TP1 on Coagulation PCB until voltage on
from infusion port (LED not illuminated), and press Air ScopeMeter® is 38.7 ±3.8 Vrms (98.6 to 120.5 Vpp).
Pump On/Off key to turn air pump OFF. Remove DPM
Release footpedal. Replace PCB containment bracket
and tubing from front panel infusion port. Turn system
onto top of card cage.
power OFF.
2.6 Disconnect cautery load box from coag handpiece
®
connectors, and ScopeMeter® from cautery load box.
SURGICAL Universal Fragmatome®
Air Infusion
01 Tune 2.7 Fully depress footpedal and adjust TP4 until front panel
mmHg
On/Off
Reset
displayed power reads 100 volts (-10V / +0V). Release
footpedal.
Frag Coag StnBy
01:35
MIN : SEC Reset
001
DIGITAL
PRESSURE
METER
(DPM)
5.4 906-8400-501
Universal Fragmatome®
3.0 U/S Driver Checks REMOVE AND REPLACE FIBER OPTIC LAMPS
The U/S Driver PCB output power cannot be calibrated
in the field. This PCB requires Automated Test Three lamps (5 V, 4.85 W, halogen) are used to illuminate the
Equipment (ATE) with highly precise meters to obtain front panel keys, directed to the front panel through focusing
an accurate calibration. If a problem with U/S power is lenses and fiber optics. The lamps and lenses are mounted on
suspected, and the Fragmatome® handpiece has been top of the Lamp PCB, built into the power supply (see Figure
eliminated as a possible problem, then the U/S Driver 3-3).
PCB must be replaced.
Remove and Replace Lamp
The following checks can be done at the system level.
1. Turn system power OFF.
3.1 Set Universal Fragmatome® system securely on top of
UMVS, and connect together with interconnect cable. 2. Remove top cover from console (refer to Section Three
Verify footswitch and gas supply hose are properly of this manual).
attached to UMVS.
3. Remove four screws securing Lamp PCB to brackets.
3.2 Turn system power OFF in both machines. Connect
ScopeMeter® positive lead to TP9 and negative lead to
TP8 on the U/S Driver PCB (see Figure 5-3). (TP9 may WARNING!
be difficult to reach. If so, slide PCB up from card cage, The three lamps on the Lamp PCB may be very hot.
connect positive lead, and reinsert.) Handle lamps carefully to avoid burning fingers.
3.4 After successfully tuning the handpiece, verify a 40.9 to 6. Insert two wires from new lamp into wiring block and
44.9 KHz readout. Remove ScopeMeter®. set new lamp in PCB bracket.
3.5 Set output power Limit to 100%. Press footswitch's 7. Tighten two set screws to secure lamp wires in wiring
right side switch to enable fragmentation (Frag: Avail.) block.
on UMVS.
8. Reattach Lamp PCB to brackets with four screws, and
3.6 Fully depress footpedal. Verify that the handpiece is replace top cover on console.
operating and that no error messages are displayed.
CAUTION
906-8400-501 5.5
Universal Fragmatome®
TROUBLESHOOTING CHART
Error message on display screen. Possible system failure. Refer to Power Up Self-Test Error Messages and
Run Time Error Messages in this troubleshooting
chart.
Frag Handpiece
Frag handpiece does not tune. 1. Handpiece may not be plugged in. 1. Plug handpiece cable into connector on front
of console.
4. Handpiece tuned while hot. 4. Allow handpiece to air cool for at least 15
minutes before tuning.
Intraoperative loss of ultrasonic power. 1. Worn handpiece tip. 1. Install new handpiece tip.
Low ultrasonic power.
Intermittent ultrasonic power. 2. Loose handpiece tip. 2. Loosen tip, re-tighten, and tune.
3. Handpiece tuned while hot. 3. Allow handpiece to air cool for at least 15
minutes before tuning.
Air pump turns OFF during self-test. Electrical or mechanical failure. Turn system power OFF, then ON, to repeat the
automatic self-test.
System stuck in self-test. Electrical or mechanical failure. Turn system power OFF, then ON, to repeat the
automatic self-test.
Error tone sounds every 2 seconds 1. GFX or GFI® tubing disconnected from front 1. Connect GFX or GFI® tubing to front panel.
during run time. panel.
Error tone sounds every 30 seconds Pressure transducer error. System still functional, although displayed pressure
during run time. may be invalid. At conclusion of procedure turn
system power OFF, then ON, to reset system.
Tone sounds every minute during Normal system operation. No correction required.
high pressure.
Table 5-2. TROUBLESHOOTING CHART - This table contains three columns of information. The first column labeled
Condition or Error Message lists the problem observed. The Possible Cause/Result column lists the possible
reasons for the problem. The last column, labeled Correction, suggests what you can do to correct the problem.
5.6 906-8400-501
Universal Fragmatome®
ERR 0 (RAM failure) Random Access Memory (RAM) failure. Turn system power OFF, then ON, to repeat the
automatic self-test.
ERR 1 (ROM failure) Erasable Programmable Read Only Memory Turn system power OFF, then ON, to repeat the
(EPROM) failure. (The contents of memory has automatic self-test.
changed.)
ERR 2 (CPU failure) Central Processing Unit operating code Turn system power OFF, then ON, to repeat the
(instruction) failure. automatic self-test.
ERR 3 (CPU/Memory failure) Central Processing Unit failure during Random Turn system power OFF, then ON, to repeat the
Access Memory test. (System unable to automatic self-test.
differentiate.)
EEPROM Failure Electrically Eraseable Programmable Read Only Turn system power OFF, then ON, to repeat the
Memory failure automatic self-test.
WDT Err Watchdog Timer. Turn system power OFF, then ON, to repeat the
automatic self-test.
Key Swtch Err Stuck arrow keys. Turn system power OFF, then ON, to repeat the
automatic self-test.
85V Err 85 volt power supply out of tolerance. Turn system power OFF, then ON, to repeat the
automatic self-test.
ADC Err Reference voltage out of tolerance. Turn system power OFF, then ON, to repeat the
automatic self-test.
906-8400-501 5.7
Universal Fragmatome®
Hdpc Err 1. Electrical or mechanical failure. Power to 1. Release footpedal and press Frag key to clear
handpiece is disabled. error message. Retune handpiece and
reengage ultrasonic power by depressing the
footpedal. If problem persists, replace
handpiece.
3. Handpiece tuned while hot. 3. Allow handpiece to air cool for at least 15
minutes before tuning.
Gnd Flt Err Ground fault. Power to handpiece is disabled. Release footpedal and press Frag key to clear
error message. Retune handpiece and reengage
ultrasonic power. If problem persists, replace
handpiece.
Coag Err Coagulation fault or voltage output fault. Release footpedal, press Coag key to clear error
Coagulation voltage disabled. message, and reengage coagulation by depressing
the footpedal.
Key Swtch Err Stuck keyboard membrane switch (non arrow Release all front panel keys. Press any mode key
key). Key depressed too long. to clear message.
U/S Err Ultrasonic power output fault. Power disabled. Release footpedal and press Frag key to clear
error message. Reengage ultrasonic power by
depressing the footpedal. If problem persists,
replace handpiece.
Pwr Sply Err Power supply fault. Press either mode key to clear the message. If
error persists turn system power OFF, then ON,
and continue if possible.
ADC Err Reference voltage out of tolerance. Press active mode key to clear error. If error
persists turn system power OFF, then ON, and
continue if possible.
U/S WDT Err U/S watchdog time out. Power to handpiece is To clear error message release footpedal and
disabled. press Frag key, or depress footpedal from position
0 to position 1. Retune handpiece and reengage
ultrasonic power. Call Alcon if problem persists.
CHECK UMVS CONNECTION 1. Communications error. 1. Check cable connection between Universal
Fragmotome® and UMVS. If problem persists,
recycle power on both units.
5.8 906-8400-501
Universal Fragmatome®
SECTION SIX
SCHEMATICS
TABLE OF CONTENTS
906-8400-501 6.1
Universal Fragmatome®
W1, Assy, Cable, Flat 26 Cond W2, Assy, Cable, U/S Handpiece
840-1035-501 394-1044-501
INTERCONNECTION CABLES
906-8400-501 6.3
Universal Fragmatome®
W5, Assy, Cable, Solenoid Transducer W6, Assy, Cable, CPU-Front Panel Controller
394-1046-501 394-1014-501
INTERCONNECTION CABLES
6.4 906-8400-501
Universal Fragmatome®
INTERCONNECTION CABLES
906-8400-501 6.5
Universal Fragmatome®
SECTION SEVEN
PARTS LISTS & DRAWINGS
TABLE OF CONTENTS
906-8400-501 7.1
Universal Fragmatome®
7.2 906-8400-501
Universal Fragmatome®
906-8400-501 7.3
Universal Fragmatome®
7.4 906-8400-501
Universal Fragmatome®
SECTION EIGHT
ADDITIONAL INFORMATION
Explanation
This section is flexible and can be used for any Universal Fragmatome®data that doesn't fit into the prior
sections of this manual. Examples of this data would be system accessories, technical bulletins, and personal
notes you collect to help perform maintenance on the Universal Fragmatome®.
906-8400-501 8.1
Universal Fragmatome®
8.2 906-8400-501