Air Pollution PDF 3
Air Pollution PDF 3
• SO2 + UV SO2*
• SO2* SO2 + light
Therefore:
F = k [SO2]
Where: F = is the intensity of fluorescence
radiation; K = is the factor of
proportionality; [SO2] = concentration of
SO2
Procedure
• The air sample flows into the inlet of the analyser where it is
scrubbed to remove any interference by aromatic hydrocarbons that
may be present. A hydrocarbon scrubber device usually
accomplishes this.
• Then the air sample flows into a reaction chamber, where it is
irradiated by UV radiation with a wavelength range of (200-220) nm.
• The UV fluorescence light, in the wavelength range of (240-420) nm,
is optically filtered and then converted to an electrical signal by a UV
detector, for example, a photomultiplier tube.
• The response of the analyser is proportional to the number of SO2
molecules in the reaction chamber.
• Therefore, either temperature or pressure has to be kept constant,
or if variation of these parameters is expected, the measured values
have to be corrected.
• For this UV fluorescence method to yield accurate concentration
measurements, it must be calibrated against some primary standard
Calibration Procedure
Zero Calibration – Switch on the analyzer at ZERO mode and zero
gas from internal source will be measured by the analyzer. After the
reading has stabilized, check the display of the zero value. In case
of derivation, adjust the zero value.
Zero Gas: Zero air used in the calibration of the analyser should
not contain a concentration of SO2 detectable by the analyser under
calibration. The concentration of O2 in the zero air shall be within +/-
2% of the normal composition of air (20.9%).
Span Calibration – After the ZERO calibration has been done,
switch the analyzer at SPAN mode. SO2 span gas (pre-determined
concentration) from permeation tube, kept in permeation oven,
would be measured by the analyzer. In case of any deviation in the
displayed value and the span gas concentration adjust the reading
of analyzer to the span value. Repeat ZERO and SPAN calibration
for atleast three times or till stable and true values are indicated.
After ZERO and SPAN calibration, switch the analyzer at SAMPLE
mode. Now, analyzer will measure SO2 present in the ambient air.
Carbon Monoxide (CO) in ambient air
(Non-Dispersive Infrared Method)
• Non Dispersive Infra-Red (NDIR) photometry provides a method of utilising the
integrated absorption of infra-red energy over most of the spectrum for a given
compound to provide a quantitative determination of the concentration of Carbon
Monoxide (CO) in ambient air.
• The spectrometer measures the absorption by CO at 4.7 mm using two parallel
infrared beams through a sample cell, a reference cell and a selective detector.
• The detector signal is led to an amplifier control section and the analyser output
measured on a meter and recording system.
• Some instruments use gas filter correlation to compare the IR absorption spectrum
between the measured gas and other gases present in the sample, in a single
sample cell.
• These instruments utilize a highly concentrated sample of CO as a filter for the IR
transmitted through the sample cell, to yield a beam that cannot be further attenuated
by the CO in the sample and thus acts as a reference beam.
• The board-band radiation that passes through the sample cell and the CO filter is
filtered again by a narrow-band-pass filter that allows only the CO-sensitive portion of
the band to pass to the detector. The removal of wavelength sensitive to other gases
reduces interferences.
Air Quality Measurement
Gases – Oxides of Nitrogen
• determined using chemiluminescence
• specific for NO, but total oxides of nitrogen
determined by passing sample over a
catalyst to convert NO2 to NO
• suitable for ambient air containing NOx (NO
and NO2) at levels less than 1 mL/m3
Air Quality Measurement
Gases – Oxides of Nitrogen
• reaction of NO with ozone in a dark enclosed
chamber to produce light - detected by a pmt
• Provided the ozone is present in excess the light
output is directly proportional to the concentration
of NO
• NO + O3 NO2* + O2
• NO2* NO2 + hn (light)
Air Quality Measurement
Gases – Carbon Monoxide
• non-dispersive infra red (NDIR) devices,
suitable for detection from 0-500ppm by
volume
• sample through a flow cell in the instrument
where it is irradiated with infrared radiation
• essentially just a modified dual beam
infrared spectrophotometer
Air Quality Measurement
IR source
Chopper
CO free
air in Ambient sample in
H i g h Vo l u m e
Samplers for TSP
Measurement
Unlikely to collect
airborne particles with
aerodynamic diameter
>100 µm due to the
design of the air inlet
Sampling usually takes
place at 6-day intervals
because filters are
changed manually
Adapted from https://www.qld.gov.au/environment/pollution/monitoring/air/air-monitoring/measuring/samplers)
THAPAR INSTITUTE PES111 By Ashutosh Kumar, Ph.D.
OF ENGINEERING & TECHNOLOGY
(Deemed to be University) AQM&CC Assistant Professor, SEE 3
Air Quality Measurement (Cont.)
Principles:
To capture the particles passing through a defined area
for a defined time without disturbing their paths
The velocity of the air going into the sampling probe
nozzle is equal to the velocity of the undisturbed air at
that point (cross section of Duct/Stack)
The air is disturbed as little as possible so that the same
particles (no more and no less) go into the nozzle as
would have passed the area of the nozzle had it not been
there
Adapted from https://www.airflowsciences.com/services/field-testing/isokinetic)