About This Manual
The Rconsole1 User Manual is designed for the appropriate use of digital radiography
operating console software developed by DRTECH Co, Ltd. This manual provides the
instructions for the proper use of Rconsole1.
Please contact us if there are any problems while using the product.
Website http://www.drtech.co.kr
Customer Support Team Email [email protected]
Tel+82 31 779 7448
Fax+82 31 779 7490
RConsole1_UM_COM_EN_02
About This Manual
Information contained in this manual is confidential and the property of DRTECH Co, Ltd.,
which is only provided to the authorized customers of DRTECH Co, Ltd.
The purpose of the manual is to assist the users to comprehensively utilize the product of
DRTECH Co, Ltd. All information contained in the manual may not be altered or
distributed without prior written consent of DRTECH Co, Ltd., and may not be disclosed
to unauthorized personnel.
Information contained in the manual may be subject to change without prior notice.
RConsole1_UM_COM_EN_02
Table of Contents
1. INTRODUCTION ..................................................................................................................... 6
1.1 INTRODUCTION OF RCONSOLE1 ................................................................................................................ 6
1.2 INTENDED USE .......................................................................................................................................... 6
2. SYSTEM REQUIREMENTS ........................................................................................................ 7
2.1 SUPPORTED DIGITAL X-RAY DETECTOR ..................................................................................................... 7
2.2 SYSTEM REQUIREMENTS FOR RCONSOLE1 .................................................................................................. 7
3. OVERVIEW .............................................................................................................................. 8
3.1 PROGRAM START-UP ................................................................................................................................. 8
3.2 MAIN SCREEN OF PROGRAM..................................................................................................................... 10
3.1.1. Function Tabs.................................................................................................................................... 10
3.1.2. Status Display ................................................................................................................................... 12
4. LIST ......................................................................................................................................... 14
4.1 WORKLIST ............................................................................................................................................... 15
4.1.1. Main Screen of WorkList .................................................................................................................. 15
4.1.2. Add Examination Window ............................................................................................................... 19
4.1.3. Edit Window...................................................................................................................................... 21
4.1.4. Load Worklist (An EXCEL File) .......................................................................................................23
4.2 STUDYLIST ............................................................................................................................................... 24
4.2.1. Main Screen of the Studylist ........................................................................................................... 24
4.2.2. Merge Window ................................................................................................................................. 29
4.2.3. Export Window ................................................................................................................................ 30
4.2.4. Edit Window..................................................................................................................................... 38
5. EXAM ..................................................................................................................................... 40
5.1 PROJECTION LIST .................................................................................................................................... 42
6. REVIEW .................................................................................................................................. 45
6.1 TOOLS, PROCESSING, AND TRUVIEW ART TAB ........................................................................................ 47
6.1.1. Tools Tab .......................................................................................................................................... 48
6.1.2. Image Processing Tab ...................................................................................................................... 56
6.1.3. Truview ART ..................................................................................................................................... 58
7. CONFIGURATION .................................................................................................................. 59
7.1 SYSTEM ................................................................................................................................................... 60
7.1.1. Language ........................................................................................................................................... 61
7.1.2. Format .............................................................................................................................................. 63
7.1.3. Institution ........................................................................................................................................ 65
7.1.4. Options ............................................................................................................................................. 66
7.1.5. Workflow .......................................................................................................................................... 67
7.1.6. Exposure Condition ......................................................................................................................... 69
7.1.7. Screening Bus .................................................................................................................................... 71
7.2 IMAGE ..................................................................................................................................................... 73
7.2.1. Save Option ...................................................................................................................................... 74
7.2.2. Delete Option ................................................................................................................................... 75
7.2.3. Overlay ............................................................................................................................................. 76
7.2.4. Import ............................................................................................................................................... 79
7.2.5. Export ............................................................................................................................................... 80
7.2.6. TruviewART ....................................................................................................................................... 81
7.2.7. Image Crop Option .......................................................................................................................... 82
RConsole1_UM_COM_EN_02
7.3 STORAGE ................................................................................................................................................. 83
7.3.1. Delete ................................................................................................................................................ 84
7.3.2. Backup .............................................................................................................................................. 86
7.4 ACCOUNT ................................................................................................................................................ 88
7.5 BLOCKED LIST ......................................................................................................................................... 89
7.6 DATASET ................................................................................................................................................. 90
7.7 NETWORK ................................................................................................................................................ 91
7.7.1. Local ................................................................................................................................................. 92
7.7.2. MWL ................................................................................................................................................. 93
7.7.3. PACS ................................................................................................................................................. 96
7.7.4. Printer............................................................................................................................................... 98
7.7.5. MPPS ............................................................................................................................................... 100
7.8 DETECTOR ............................................................................................................................................. 102
7.8.1. General ............................................................................................................................................ 103
7.8.2. Detailed Configuration for Panel ................................................................................................... 104
7.9 RIS CODE...............................................................................................................................................106
7.9.1. Edit RIS Code Window ................................................................................................................... 107
7.9.2. Add Marker Window ....................................................................................................................... 111
7.10 TRASH BIN .............................................................................................................................................. 112
7.10.1. Statistics ........................................................................................................................................... 114
APPENDIX A. GLOSSARY ........................................................................................................................... 116
APPENDIX B. GAIN CALIBRATION............................................................................................................ 118
PREPARATION ........................................................................................................................................... 118
CALIBRATION PROCEDURE ..................................................................................................................... 118
RConsole1_UM_COM_EN_02
1. Introduction
Rconsole1 is a complete digital image processing console software specialized for the
digital X-ray detector RSM series, developed by DRTECH Co, Ltd.
Rconsole1 provides the best image processing solution for fine tuned quality images by
integrating with the X-ray generator and the digital detector. Rconsole1 not only
processes the acquired images but also complies with DICOM standards which allow the
user to transmit and receive data with the PACS system and print images through the
DICOM printer.
The Rconsole1 is intended for use with mammographic images acquired using digital
breast x-ray systems.
RConsole1_UM_COM_EN_02 6 / 124
2. System Requirements
Rconsole1 supports the following Digital X-ray Detector manufactured by DRTECH Co,
Ltd.
RSM 1824C
RSM 2430C
RSM 1824TD
RSM 2430TDS
The table provides the minimum recommended specifications for the operating system
that runs Rconsole1. All specifications are subject to change without notice.
Table 1. Recommended Operating System Specifications
Items Minimum Recommended
CPU Intel i5-3470 Intel i7-4790
RAM 4 GB 8 GB
SSD 128GB
HDD 500 GB
500 GB
NVIDIA GeForce NVIDIA GeForce
VGA
GT630 1GB GT630 1GB
ODD N/A DVD Recorder
Windows 7 Pro Windows 7 Pro
OS
32 / 64bit 64bit
Display Size 23 inches 27 inches
Display Resolution 1600 x 900 1920 x 1080
* Optimal performance is not guaranteed for PC that does not match the recommended specifications.
RConsole1_UM_COM_EN_02 7 / 124
3. Overview
In order to run the program, double click the [Rconsole1] icon on the desktop or select
‘Rconsole1’ from ‘Windows Menu’ – ‘All Programs’ – ‘Rconsole1’.
A login window as seen in [Figure 3] will appear if the program is properly initiated.
Enter the ID and Password, and then click the [Login] icon.
RConsole1_UM_COM_EN_02 8 / 124
If you enter an invalid ID or Password, an ‘Error Window’ will appear as shown in [Figure
4] and you will not be able to log in.
RConsole1_UM_COM_EN_02 9 / 124
3.2 Main Screen of Program
You will see the Rconsole1 main screen as in [Figure 5] after logging in.
RConsole1_UM_COM_EN_02 10 / 124
The functions of these tabs are as follows.
RConsole1_UM_COM_EN_02 11 / 124
3.1.2. Status Display
Status display icons can be found on the lower left corner in the main screen as in [Figure
7].
RConsole1_UM_COM_EN_02 12 / 124
Display the connection status with the DICOM
Printer.
Refer to 7.7.4 Printer for more details.
Display the connection status with the Modality
Worklist Server.
Refer to7.7.2 WL Server for more details.
Display the connection status with the MPPS
server..
③ Logout Icon
④ Configuration Icon
RConsole1_UM_COM_EN_02 13 / 124
4. List
The List tab displays the lists of patients and studies, which can be divided into two
categories as described below.
Studylist (post-examination)
Display the examined studies.
You can edit/delete the examined studies on the list.
Note: The changes such as editing and deleting images on the list will not affect
data in the PACS server.
You can transmit examined images to the PACS server.
Refer to 4.2 Studylist for more details.
Please note!
The information on the Worklist and Studylist is stored on the Local Storage (HDD)
as default settings.
Every time the Rconsole1 runs, a new Worklist is retrieved from the PACS server
and stored on the Local Storage.
Only the contents on the list which are edited and deleted in the Rconsole1 affect
the Local Storage as default. However, edited records may be transmitted to the
PACS server using the ‘Send’ command.
RConsole1_UM_COM_EN_02 14 / 124
4.1 WorkList
Display the list of patients. The list is displayed by clicking the [WorkList] icon on the
upper left side of the List tab.
Figure 8. Worklist
RConsole1_UM_COM_EN_02 15 / 124
① List of Patients
The types of information are as noted in [Table 2].
The displayed information as well as its order can be set in ‘Configuration’.
Click or drag the name of the information type (ID, Name, and etc.) to change the
order.
If you run Rconsole1 for the first time, set the column size that fits your needs and
the setting will be saved automatically.
RConsole1_UM_COM_EN_02 16 / 124
② Search Bar & Refresh Icon
Search Bar
Search the entered value such as Patient, ID, Name, and Access Number in the
Worklist.
If you click the [Search] icon after entering the value to be searched, only
the relevant works will be displayed on the list.
Refresh Icon
Retrieve recently registered patients from the PACS server.
RConsole1_UM_COM_EN_02 17 / 124
③ Add/Edit/Delete/Load Icon
④ Exam
Switch to the Exam tab for examination of the selected patient.
Refer to 5. Exam for more details.
RConsole1_UM_COM_EN_02 18 / 124
4.1.2. Add Examination Window
An ‘Add Examination Window’ appears when clicking the icon, as shown in [Figure
10] below.
RConsole1_UM_COM_EN_02 19 / 124
① Patient Information
Patient information is entered according to the category described in [Table 4].
② Projection Information
Display the information related to projection.
Display registered procedures.
③ OK
Save the patient with the entered information and close the window.
④ Exam
Save the patient with the entered information and directly switch to the Exam tab for
examination.
Refer to 5. Exam for more details.
⑤ Cancel
Cancel the work and close the window.
RConsole1_UM_COM_EN_02 20 / 124
4.1.3. Edit Window
The ‘Edit Window’ will appear when selecting a study from the Worklist and then clicking
Note: The ‘Edit Window’ is not activated if the study information is retrieved from the
PACS server.
RConsole1_UM_COM_EN_02 21 / 124
Table 5. Study Information (Edit Window)
Category Detail
ID Patient ID
Name Patient name
② OK
Save changes and close the window.
③ Cancel
Cancel changes and close the window
RConsole1_UM_COM_EN_02 22 / 124
4.1.4. Load Worklist (An EXCEL File)
Load a Worklist in an EXCEL file format.
① Path
Locate the path where the EXCEL file is saved.
③ Open
Display the selected Worklist in the Worklist screen and close the window.
④ Cancel
Cancel and close the window.
RConsole1_UM_COM_EN_02 23 / 124
4.2 Studylist
Display the examined study list. When clicking the [Study List] icon on the upper side of
the screen, the study list will be activated.
RConsole1_UM_COM_EN_02 24 / 124
① List of the Examined Patients
Display the list of the examined patients’ information.
Each row represents the information of one patient, and a single ID can have
multiple studies.
The column of the list can be changed by dragging the names of the categories
(e.g. ID, Name, and etc.).
Height of the list is adjustable using [Ctrl key] + Mouse wheel or using [Ctrl key] +
[Arrow key].
Types of information are as noted in [Table 6].
Patient Selection
Patient can be selected by left-clicking.
The selected patient is highlighted in red. In the '① Examined Patient List' and '②
Examined Study List per Patient’, an activated study is highlighted in darker color,
while an inactivated study is highlighted in lighter color.
The selected patients can be edited or deleted.
RConsole1_UM_COM_EN_02 25 / 124
Table 7. Information Types of Examined Study List
Items Description
Accession Number.
Acc. No.
Registered Identification Number
Study ID RIS code
StudyDesc. Study Description.
Status [Number of images] / [Number of projections]
Date/Time Date/Time for an examined study
[Number of images transmitted] / [Number of
PACS
images]
[Number of images printed] / [Number of
Printer
images]
[Number of images exported] / [Number of
Export
images]
③ Thumbnail
RConsole1_UM_COM_EN_02 26 / 124
④ Search Bar & Refresh Icon
Search Bar
Search the values of Patient ID, Name, and Access Number within the Studylist.
If you click the search icon after entering the value to be searched, only
the relevant works will be displayed on the list.
Refresh Icon
Renew the list to show the latest results.
RConsole1_UM_COM_EN_02 27 / 124
⑤ Tool Icons
⑥ Send Icon
Transmit the selected images to the PACS server
RConsole1_UM_COM_EN_02 28 / 124
4.2.2. Merge Window
Merge the images and studies.
If a patient has multiple ‘Patient ID’, it is more convenient to group them by using this
function. Edit if the image and patient information do not match with each other.
Identical items selected in ① will be also located in ②if the same list is checked.
③ Merge Icon
Transmit the selected items.
Every change is saved while being transmitted.
④ Close
Close the ‘Merge Window’.
RConsole1_UM_COM_EN_02 29 / 124
4.2.3. Export Window
Save the images of the selected study from the Studylist in a USB, CD, DVD or as PDF,
print them with DICOM Printer, and create a report.
4.2.3.1. Drive
RConsole1_UM_COM_EN_02 30 / 124
4.2.3.2. Print
① Print Setting
Printer Selection
Select a printer from the drop-down menu.
Film orientation
Select a film orientation between portrait/landscape
Film size
Select the size of the film.
Smart Layout
Select the view position (CC, MLO, Left, Right) to display related images in the
screen. The screen is divided into the number of related images.
Layout
RConsole1_UM_COM_EN_02 31 / 124
Set the print layout.
Zoom
Perform zoom in/out of the image.
② Preview
Thumbnail
The image to be printed appears in the thumbnail list on the right.
Preview
If an image is selected in the thumbnail, a large image appears in the center as a
preview screen.
Merge
Merge the divided screen which has been separated by the ‘Divide’ function.
Only adjacent screens of the same sizes can be merged.
The merged shape must be a square.
Divide
Divide the screen by the value set in ‘Layout.’
Rotate
Rotate the screen 90 degrees clock-wise.
Flip
Reverse the up and down sides.
Print
Print out using general printer.
Edit Overlay
Edit overlay texts which are displayed in the Review screen.
③ DICOM Print
Perform printing with the DICOM Printer.
RConsole1_UM_COM_EN_02 32 / 124
4.2.3.3. CD/DVD
RConsole1_UM_COM_EN_02 33 / 124
Eject the tray of the CD / DVD writer.
③ Export Process
Estimated Time
Present the expected time needed for Export.
Time Left
Display estimated time left to the completion of Export.
④ Task Icons
Burn
Perform the Export task.
Cancel
Cancel the Export task.
RConsole1_UM_COM_EN_02 34 / 124
4.2.3.4. Report Window
Save or print the report which contains the information about the exam.
② Report Contents
Edit contents
Edit contents of the report.
Title of contents can be changed in ‘Setting’.
Image Preview
Show loaded images.
③ Task Icons
Print
Display preview of report and print.
RConsole1_UM_COM_EN_02 35 / 124
Save as PDF
Save the report as PDF format.
Save
Save this report temporally
Setting
Show the ‘Report Value Setting Window’.
RConsole1_UM_COM_EN_02 36 / 124
4.2.3.5 Report Value Setting Window
② Visibility Setting
Configure the visibility of items when using ‘Print’ and ‘Save as PDF’.
③ Set as default
Set the current setting as the default setting
Changed default setting will be applied from the next new reports.
RConsole1_UM_COM_EN_02 37 / 124
4.2.4. Edit Window
Only the patient information (ID, Name, Sex, Birth Date, and Age) can be edited,
when clicking the [Edit] icon in the patient list of the Studylist (upper part of the list).
OK
Save changes and close the window.
Cancel
Cancel changes and close the window.
RConsole1_UM_COM_EN_02 38 / 124
4.2.4.2. Edit Study Information
Only the study information (Accession Number, Study Description, Sex, and
Schedule date & time) can be edited when clicking the [Edit] icon in the projection list
(lower part of the list) of the Studylist.
OK
Save changes and close the window.
Cancel
Cancel changes and close the window.
RConsole1_UM_COM_EN_02 39 / 124
5. Exam
The Exam tab processes exam and edits images.
② Patient Information
Display ID, patient name, and study description.
④ Acquisition Settings
Auto Sending
RConsole1_UM_COM_EN_02 40 / 124
When the [Auto sending] icon is activated and every scheduled exam is
completed, send images to PACS server automatically.
Skip review
When the [Skip review] icon is activated, screen transition to Review screen does
not occur after the image acquisition.
Implant processing
When the [Implant processing] icon is activated, image processing parameter will
be changed to the one for implant only.
Suspend
Suspend current exam and transit to the Worklist screen.
Send
Transmit the images to the PACS server or folder.
With a long-clicking on this icon, all examined images will be transmitted.
Change detector
Connect to another detector that is currently connected when clicking the button.
RConsole1_UM_COM_EN_02 41 / 124
5.1 Projection List
① Time Machine
If there are remaining images of prior examinations in the Local DB, those images will
be displayed on the thumbnails.
② Projection Guide
Projection guidelines for the selected projection.
RConsole1_UM_COM_EN_02 42 / 124
Figure 25. Projection Guide
RConsole1_UM_COM_EN_02 43 / 124
Figure 26. Add View Position Window
After selecting position, a projection list is added by clicking [OK].
Delete
After being selected, a projection list is removed by clicking [Delete].
Edit
After selecting a list, the same screen as the ‘Add View Position’ appears when
clicking the [Edit] Icon.
Select the list that you want to add and the click the [OK] icon to apply the change.
④Projection List
Display previously examined list or projection list to be examined.
After selecting a projection list, examining X-ray shows the Review Screen and
displays the examined images on the Viewer.
Double-clicking on an image brings the Review Screen.
If the manual dose input option is checked, you can input through the dose setting
window at the bottom of each projection.
RConsole1_UM_COM_EN_02 44 / 124
6. Review
Perform editing the examined images.
② Patient Information
Display ID, patient name, and study description.
RConsole1_UM_COM_EN_02 45 / 124
④ Save / Pause / Send Icons
Save
Store the images in the Local Storage.
Pause
Pause the projection under progress and return to the Worklist screen.
The paused projection list is displayed in light brown color.
Send
Transmit the images to the PACS server or folder.
With a long-clicking on this icon, all examined images will be transmitted.
RConsole1_UM_COM_EN_02 46 / 124
6.1 Tools, Processing, and Truview ART Tab
The Tools, Processing, and Truview ART Tab can be used through the tab icons on the
upper right corner.
Figure 28.ToolsTab (Left), Processing Tab (Center), Truview ART Tab (Right)
RConsole1_UM_COM_EN_02 47 / 124
6.1.1. Tools Tab
If there are images examined in the past, they can be displayed according to the
selected date.
Display a list of studies categorized by projections.
Display a collection of image editing icons.
① Time Machine
Display the images examined in the past.
Note: There must be the images in the Local Storage.
RConsole1_UM_COM_EN_02 48 / 124
③ Toolbar
A collection of icons required for image analysis and editing.
Display the currently configured list of icons.
The icons can be edited on the ‘Edit Toolbar Window’.
The [Edit] Icon is used to edit the Toolbar window.
Refer to 6.1.1.1 Edit Toolbar for more details.
RConsole1_UM_COM_EN_02 49 / 124
6.1.1.1. Edit Toolbar
② Available Icons
All the available image editing icons are listed.
Add tools to ‘① Applied Icons Box’ through drag & drop.
Previously added icons are displayed in dark color.
RConsole1_UM_COM_EN_02 50 / 124
③ Add Shutter Icon
Add the icons with ROI Shutter Sizes specified by the User.
The following window appears when clicking the icon.
⑤ OK
Apply the modified tools arrangement and close the window.
⑥ Cancel
Cancel and close the window.
RConsole1_UM_COM_EN_02 51 / 124
6.1.1.2. Toolbar Icons
The functions of all the icons in the toolbar are described in [Table 8].
RConsole1_UM_COM_EN_02 52 / 124
Display the "↓" marker on the screen.
RConsole1_UM_COM_EN_02 53 / 124
Capture the image.
Ruler calibration.
Select annotation.
RConsole1_UM_COM_EN_02 54 / 124
Draw rectangular annotation.
RConsole1_UM_COM_EN_02 55 / 124
6.1.2. Image Processing Tab
* Function for advanced users *
It is utilized to acquire a more optimized image by adjusting the post-processing
parameters of the completed image.
RConsole1_UM_COM_EN_02 56 / 124
② Parameter Functions
Wizard
Open the multi-image viewer window.
8 types of images are shown based on the currently set parameter.
Apply
Apply the currently set parameters to the image.
Reset
Reset the parameter to its initial setting.
Save parameter
Save the currently set parameters and apply them when performing the projection
of the relevant projection-position.
RConsole1_UM_COM_EN_02 57 / 124
6.1.3. Truview ART
* Function for advanced users *
② Preview
Preview zooming in the selected part on the Viewer.
③ Compare
Compare the two images with and without applying Truview ART.
④ Save
Save Truview ART parameters.
RConsole1_UM_COM_EN_02 58 / 124
7. Configuration
Configure the user settings of Rconsole1.
RConsole1_UM_COM_EN_02 59 / 124
7.1 System
Subcategory Description
Language Select system language.
Format Select date format for presentation.
Institution Enter the information of the institution using the Rconsole1.
Options Change settings for options.
Workflow Change settings for work flow
Exposure Change settings for exposure condition.
Screening Bus Configure the options related to mobile screening bus
RConsole1_UM_COM_EN_02 60 / 124
7.1.1. Language
RConsole1_UM_COM_EN_02 61 / 124
Configure the system language.
Items Description
Korean Set when you use the Korean window.
Chinese Set when you use the Chinese window.
Japanese Set when you use the Japanese window.
Portuguese Set when you use the Portuguese window.
Spanish Set when you use the Spanish window.
German Set when you use the German window.
Turkish Set when you use the Turkish window.
Swedish Set when you use the Swedish window.
French Set when you use the French window.
Russian Set when you use the Russian window.
Ukrainian Set when you use the Ukrainian window.
Italian Set when you use the Italian window.
Polish Set when you use the Polish window.
Traditional
Set when you use the Traditional Chinese window.
Chinese
Finnish Set when you use the Finnish window.
RConsole1_UM_COM_EN_02 62 / 124
DICOM character set:
Set the DICOM character set. If language is selected in the program installation
sequence, this will be set automatically.
If this item is not correctly changed, some tag value of DICOM header will be not
correctly contained.
Language DICOM character set
Korean ISO IR 6\ISO 2022 IR 149
Chinese ISO IR 6\GB18030
Japanese ISO 2022 IR 13\ISO 2022 IR 87
Turkish ISO IR 148
Default ISO IR 100
Thai ISO 2022 IR 166
Ukrainian, Russian ISO IR 6\ISO IR 144
UTF-8 ISO IR 192
Japanese \ISO 2022 IR 87
7.1.2. Format
RConsole1_UM_COM_EN_02 63 / 124
① Date Format
Configure the date format.
Date
Y: Year , M: Month, D: Day
Delimiter
Set the symbol to be used between Y, M, and D.
For example, if today is May 31st, 2012,
Type Description
Alphabetic Alphabetic character Type
Ideographic Ideographic character Type
Phonetic Phonetic character Type
You can select the Family name, Given name, Middle name.
RConsole1_UM_COM_EN_02 64 / 124
7.1.3. Institution
RConsole1_UM_COM_EN_02 65 / 124
7.1.4. Options
Auto Start
If checked, Rconsole1 will automatically start when Windows starts.
Virtual Keyboard
If checked, virtual keyboard will be available for use.
RConsole1_UM_COM_EN_02 66 / 124
7.1.5. Workflow
RConsole1_UM_COM_EN_02 67 / 124
Exposure Condition Input Option
Select the exposure condition options.
The exam window composition is changed according to the selected option.
Option Description
RConsole1_UM_COM_EN_02 68 / 124
7.1.6. Exposure Condition
Delete the selected Grid from the Grid section using button.
RConsole1_UM_COM_EN_02 69 / 124
② Focal Spot, Thickness, Compression rate settings
Focal Spot Settings
Large FS : Set Large Focal Spot value.
Small FS : Set Small Focal Spot value.
Thickness
Set a default value for thickness.
Compression
Set a default value of compression rate.
RConsole1_UM_COM_EN_02 70 / 124
7.1.7. Screening Bus
Note: “Use a PID based on Mammo bus index” option must be selected in PID option in
Workflow in order to use this menu.
RConsole1_UM_COM_EN_02 71 / 124
Input station number.
Data index consists of 4-7 digits and it can be initialized by clicking the ‘Reset ID’
button.
Use '-'
When ‘Use ‘-‘’ is checked, ‘-‘is inserted between each item value.
Example) 20160514-1234-000001 / 201605141234000001
Setting the same accession number with patient ID
When this option is checked, the same accession number is generated as the
created patient’s ID.
When not using this option, an accession number is generated with the number
‘1’added after the created patient’s ID.
RConsole1_UM_COM_EN_02 72 / 124
7.2 Image
Subcategories Description
Image Folder Set the location to save the images.
Delete Option Set delete options for image files.
Overlay Set overlay on images.
Import Import images to the Studylist.
Export Set image export options
TRUVIEW ART Enable or disable the Truview ART.
Crop Option Set edge cut options.
RConsole1_UM_COM_EN_02 73 / 124
7.2.1. Save Option
RConsole1_UM_COM_EN_02 74 / 124
7.2.2. Delete Option
RConsole1_UM_COM_EN_02 75 / 124
② Reject Code
List of reasons for deletion.
Delete
Delete the selected Reject Code from the list.
Add
Add a new Reject Code.
Edit
Edit the selected Reject Code.
7.2.3. Overlay
Allow overlay configuration.
RConsole1_UM_COM_EN_02 76 / 124
Figure 45. Edit Overlay
② Edition
Add a new overlay to the list.
The following window as shown in [Figure 46] appears when clicking the icon.
RConsole1_UM_COM_EN_02 77 / 124
Figure 46. Overlay Option
The selected overlay is added after entering the size and clicking [OK].
Delete the selected annotation from the list.
The selected annotation moves up by one row.
The selected annotation moves down by one row.
RConsole1_UM_COM_EN_02 78 / 124
7.2.4. Import
RConsole1_UM_COM_EN_02 79 / 124
7.2.5. Export
RConsole1_UM_COM_EN_02 80 / 124
7.2.6. TruviewART
RConsole1_UM_COM_EN_02 81 / 124
7.2.7. Image Crop Option
RConsole1_UM_COM_EN_02 82 / 124
7.3 Storage
Configurations concerning the deletion and storage of data in the operating pc where
the images will be saved.
Details Description
Configure the deletion and automatic deletion function of
Delete
images stored in Local Storage.
Compress and backs-up the images stored in Local
Backup
Storage.
RConsole1_UM_COM_EN_02 83 / 124
7.3.1. Delete
Configure the deletion and automatic deletion function of images stored in Local
Storage.
① Storage Delete
The [Delete] Icon deletes all images stored in the Local Storage.
Deleted images cannot be restored, so please make sure to perform a backup prior to
deleting images!
② Auto Delete
Auto Delete
If checked, all images satisfying the pertinent conditions are automatically deleted
from the PC storage.
Disk Size
RConsole1_UM_COM_EN_02 84 / 124
If checked, images in the PC storage are automatically deleted once the used
disk space exceeds the set limit.
Max: maximum limit of disk space set for storage of images in the PC storage.
Period
If checked, images in the PC Storage are continuously deleted automatically,
during the set period below.
ex1) Set to "2", "Month": Delete all images that were stored 2 months earlier as of
today
ex2) Set to "1" "Day": Delete all images that were stored 1 day earlier as of today.
The cycle may be set by one of the following;Day/Week/Month/Year
Only one of the two auto deleting methods, ‘Disk Size’ and ‘Period’, may be selected
at a time
RConsole1_UM_COM_EN_02 85 / 124
7.3.2. Backup
RConsole1_UM_COM_EN_02 86 / 124
②Automatic Backup
Auto Backup Check
If checked, automatic backup is performed during the set period explained below.
Period
Set the automatic backup period.
Available units consist of Day/Week/Month.
RConsole1_UM_COM_EN_02 87 / 124
7.4 Account
RConsole1_UM_COM_EN_02 88 / 124
7.5 Blocked List
RConsole1_UM_COM_EN_02 89 / 124
7.6 Dataset
③ Series Order
Set the series storing method for the Study routine.
One series: Storing as one series.
Each series: Storing as individual series.
RConsole1_UM_COM_EN_02 90 / 124
7.7 Network
Subcategory Description
Network connection is required.
Configure the network information of the
Local
PC (or Work Station) currently operating
Rconsole1.
Network connection is required.
MWL Configure the network information of the
server storing the Modality Worklist.
Network connection is required.
Configure the network information of the
Storage
Storage Server, which stores the image
files.
Printer connection is required.
Printer Configure the connection information of
the printer.
Network connection is required.
MPPS Configure the network information of the
MPPS Server.
RConsole1_UM_COM_EN_02 91 / 124
7.7.1. Local
Configure the network information of the PC (or Work Station) currently operating
Rconsole1.
RConsole1_UM_COM_EN_02 92 / 124
7.7.2. MWL
RConsole1_UM_COM_EN_02 93 / 124
Edit the selected server information.
②Option
Time-out
Response standby time with the server.
It is regarded as connection failure, if the server is unresponsive longer than the
set period of time.
③ Query
Configure the automatic refresh settings of MWL.
Auto
MWL is automatically refreshed during the entered interval.
Manual
Set MWL to be refreshed manually.
Interval
Set the interval for automatic refresh.
④ Modality
Select the modality type of the imaging equipment.
E.g. CR, DR, ALL
Add
Add Modality.
Delete
Delete the selected Modality.
⑤ Date
Configure the date format for MWL
Date
Y: Year , M: Month, D: Day
Delimiter
Set the symbol used between Y, M, and D.
RConsole1_UM_COM_EN_02 94 / 124
For example, if today is May 31st, 2012,
RConsole1_UM_COM_EN_02 95 / 124
7.7.3. PACS
RConsole1_UM_COM_EN_02 96 / 124
Edit
Edit the selected server information.
② Option
Time-out
Response standby time with the server.
It is regarded as a connection failure if the server is unresponsive longer than the
set period of time.
③ DICOM Type
Configure the DICOM type
RConsole1_UM_COM_EN_02 97 / 124
7.7.4. Printer
RConsole1_UM_COM_EN_02 98 / 124
② Overlay Information
Select the additional annotation information added on the film.
The following pop-up window shown in [Figure 60] appears when clicking the [Edit]
icon.
RConsole1_UM_COM_EN_02 99 / 124
7.7.5. MPPS
Subcategory Description
Register and edit the model of panel
(detector) being used.
General
Execute detector software, RCali1 and
Terminal Emulator.
Detailed configuration screen related to
Panel A, registered in ‘General’ tab.
Panel A Detailed configuration items for panel are
created in accordance with the number of
panels registered in ‘General’.
② Serial Number
RIS Code
The [Edit] icon shows ‘RIS Code Edit Window’.
RIS Code Import from Excel
Add RIS Code values from an EXCEL file.
Note: Duplicate code values are overwritten with contents of the Excel file. (It
overwrites the existing data)
Detector Rotation
Rotate every image by 90 degrees.
Marker size change
Change the size of every marker.
3 task icons are available. The descriptions are noted in [Table 11].
Edit the code and descriptions of the RIS Code selected from
the list, with the information entered in the text box below the
icon.
[Editing Sequence]
Edit
1) Select the code to be edited from the list.
2) Enter new information in the textbox below the icon.
3) Click the [Edit] icon.
Add the RIS Code with the information entered in the textbox
below the icon.
[Adding Sequence]
Add
1) Enter new information in the textbox below the icon.
2)Click the [Add] icon.
[Deleting Sequence]
Delete 1)Select the code to be deleted from the list.
2)Click the [Delete] icon.
④ View Position
Set the position of the RIS Code projection.
Terms Description
Preparation
- Aging for 20 minutes before starting calibration
- Remove Compression Paddle.
- Fully open Collimator to cover X-ray on the whole detector area.
- Put an Acryl Phantom (4cm) on Bucky.
- Set Grid type as “None” in RCali1 configuration. (When Grid Line Pattern appears, set
Grid Type as”1824C” or “2430C” depending on the product model)
Calibration Procedure
* Check the radiation field and clear Acryl Phantom as well as Bucky before calibration.
* If the Mean values of every image have more than 300Gy difference, you have to re-
take images.
1. Run Rcali1
2. Select the detector to perform the Gain Calibration from the Detector Menu. (Tool bar
→ Detector)
6. A main screen will appear. Proceed with the X-ray projection at this point in order to
acquire an image.
10. If the X-ray radiation dose is appropriate, please select [OK] and proceed with the
projections.
13. The next projection will take place after the projection standby timer (60 seconds)
disappears. The projection standby time may be set in Configuration.
14. The Gain Calibration results are automatically saved as Gain Map (*.GMP) when all
required images are projected and the existing files are backed-up.
15. The existing GMP files are backed up in the [RCali1 Installation Folder]\MAP folder. The
file name is as follows in [Figure 83].