Document No.

: RConsole1_UM_COM_EN_02
Revision No.: Rev.02
Release date:2018-04-26
About This Manual
The Rconsole1 User Manual is designed for the appropriate use of digital radiography
operating console software developed by DRTECH Co, Ltd. This manual provides the
instructions for the proper use of Rconsole1.

This manual contains the following contents.

 Operation Environment, Installation Methods for Rconsole1


 Functions and Instructions for Rconsole1

Please contact us if there are any problems while using the product.

Website http://www.drtech.co.kr
Customer Support Team Email [email protected]
Tel+82 31 779 7448
Fax+82 31 779 7490

To Enhance The Quality Of Human Life!

RConsole1_UM_COM_EN_02
About This Manual

Copyright - Rconsole1 Manual

Information contained in this manual is confidential and the property of DRTECH Co, Ltd.,
which is only provided to the authorized customers of DRTECH Co, Ltd.

The purpose of the manual is to assist the users to comprehensively utilize the product of
DRTECH Co, Ltd. All information contained in the manual may not be altered or
distributed without prior written consent of DRTECH Co, Ltd., and may not be disclosed
to unauthorized personnel.

Information contained in the manual may be subject to change without prior notice.

Copyright © 2017 DRTECH Co, Ltd. All rights reserved.

RConsole1_UM_COM_EN_02
Table of Contents
1. INTRODUCTION ..................................................................................................................... 6
1.1 INTRODUCTION OF RCONSOLE1 ................................................................................................................ 6
1.2 INTENDED USE .......................................................................................................................................... 6
2. SYSTEM REQUIREMENTS ........................................................................................................ 7
2.1 SUPPORTED DIGITAL X-RAY DETECTOR ..................................................................................................... 7
2.2 SYSTEM REQUIREMENTS FOR RCONSOLE1 .................................................................................................. 7
3. OVERVIEW .............................................................................................................................. 8
3.1 PROGRAM START-UP ................................................................................................................................. 8
3.2 MAIN SCREEN OF PROGRAM..................................................................................................................... 10
3.1.1. Function Tabs.................................................................................................................................... 10
3.1.2. Status Display ................................................................................................................................... 12
4. LIST ......................................................................................................................................... 14
4.1 WORKLIST ............................................................................................................................................... 15
4.1.1. Main Screen of WorkList .................................................................................................................. 15
4.1.2. Add Examination Window ............................................................................................................... 19
4.1.3. Edit Window...................................................................................................................................... 21
4.1.4. Load Worklist (An EXCEL File) .......................................................................................................23
4.2 STUDYLIST ............................................................................................................................................... 24
4.2.1. Main Screen of the Studylist ........................................................................................................... 24
4.2.2. Merge Window ................................................................................................................................. 29
4.2.3. Export Window ................................................................................................................................ 30
4.2.4. Edit Window..................................................................................................................................... 38
5. EXAM ..................................................................................................................................... 40
5.1 PROJECTION LIST .................................................................................................................................... 42
6. REVIEW .................................................................................................................................. 45
6.1 TOOLS, PROCESSING, AND TRUVIEW ART TAB ........................................................................................ 47
6.1.1. Tools Tab .......................................................................................................................................... 48
6.1.2. Image Processing Tab ...................................................................................................................... 56
6.1.3. Truview ART ..................................................................................................................................... 58
7. CONFIGURATION .................................................................................................................. 59
7.1 SYSTEM ................................................................................................................................................... 60
7.1.1. Language ........................................................................................................................................... 61
7.1.2. Format .............................................................................................................................................. 63
7.1.3. Institution ........................................................................................................................................ 65
7.1.4. Options ............................................................................................................................................. 66
7.1.5. Workflow .......................................................................................................................................... 67
7.1.6. Exposure Condition ......................................................................................................................... 69
7.1.7. Screening Bus .................................................................................................................................... 71
7.2 IMAGE ..................................................................................................................................................... 73
7.2.1. Save Option ...................................................................................................................................... 74
7.2.2. Delete Option ................................................................................................................................... 75
7.2.3. Overlay ............................................................................................................................................. 76
7.2.4. Import ............................................................................................................................................... 79
7.2.5. Export ............................................................................................................................................... 80
7.2.6. TruviewART ....................................................................................................................................... 81
7.2.7. Image Crop Option .......................................................................................................................... 82

RConsole1_UM_COM_EN_02
7.3 STORAGE ................................................................................................................................................. 83
7.3.1. Delete ................................................................................................................................................ 84
7.3.2. Backup .............................................................................................................................................. 86
7.4 ACCOUNT ................................................................................................................................................ 88
7.5 BLOCKED LIST ......................................................................................................................................... 89
7.6 DATASET ................................................................................................................................................. 90
7.7 NETWORK ................................................................................................................................................ 91
7.7.1. Local ................................................................................................................................................. 92
7.7.2. MWL ................................................................................................................................................. 93
7.7.3. PACS ................................................................................................................................................. 96
7.7.4. Printer............................................................................................................................................... 98
7.7.5. MPPS ............................................................................................................................................... 100
7.8 DETECTOR ............................................................................................................................................. 102
7.8.1. General ............................................................................................................................................ 103
7.8.2. Detailed Configuration for Panel ................................................................................................... 104
7.9 RIS CODE...............................................................................................................................................106
7.9.1. Edit RIS Code Window ................................................................................................................... 107
7.9.2. Add Marker Window ....................................................................................................................... 111
7.10 TRASH BIN .............................................................................................................................................. 112
7.10.1. Statistics ........................................................................................................................................... 114
APPENDIX A. GLOSSARY ........................................................................................................................... 116
APPENDIX B. GAIN CALIBRATION............................................................................................................ 118
PREPARATION ........................................................................................................................................... 118
CALIBRATION PROCEDURE ..................................................................................................................... 118

RConsole1_UM_COM_EN_02
1. Introduction

1.1 Introduction of Rconsole1

Rconsole1 is a complete digital image processing console software specialized for the
digital X-ray detector RSM series, developed by DRTECH Co, Ltd.
Rconsole1 provides the best image processing solution for fine tuned quality images by
integrating with the X-ray generator and the digital detector. Rconsole1 not only
processes the acquired images but also complies with DICOM standards which allow the
user to transmit and receive data with the PACS system and print images through the
DICOM printer.

Figure 1. Rconsole1 Image Processing Workflow

1.2 Intended Use

The Rconsole1 is intended for use with mammographic images acquired using digital
breast x-ray systems.

RConsole1_UM_COM_EN_02 6 / 124
2. System Requirements

2.1 Supported Digital X-ray Detector

Rconsole1 supports the following Digital X-ray Detector manufactured by DRTECH Co,
Ltd.
 RSM 1824C
 RSM 2430C
 RSM 1824TD
 RSM 2430TDS

2.2 System Requirements for Rconsole1

The table provides the minimum recommended specifications for the operating system
that runs Rconsole1. All specifications are subject to change without notice.
Table 1. Recommended Operating System Specifications
Items Minimum Recommended
CPU Intel i5-3470 Intel i7-4790
RAM 4 GB 8 GB
SSD 128GB
HDD 500 GB
500 GB
NVIDIA GeForce NVIDIA GeForce
VGA
GT630 1GB GT630 1GB
ODD N/A DVD Recorder
Windows 7 Pro Windows 7 Pro
OS
32 / 64bit 64bit
Display Size 23 inches 27 inches
Display Resolution 1600 x 900 1920 x 1080
* Optimal performance is not guaranteed for PC that does not match the recommended specifications.

RConsole1_UM_COM_EN_02 7 / 124
3. Overview

3.1 Program Start-up

In order to run the program, double click the [Rconsole1] icon on the desktop or select
‘Rconsole1’ from ‘Windows Menu’ – ‘All Programs’ – ‘Rconsole1’.

Figure 2. Rconsole1 Shortcut Icon

A login window as seen in [Figure 3] will appear if the program is properly initiated.

Figure 3. Login Window

Enter the ID and Password, and then click the [Login] icon.

RConsole1_UM_COM_EN_02 8 / 124
If you enter an invalid ID or Password, an ‘Error Window’ will appear as shown in [Figure
4] and you will not be able to log in.

Figure 4. Login Error

RConsole1_UM_COM_EN_02 9 / 124
3.2 Main Screen of Program

You will see the Rconsole1 main screen as in [Figure 5] after logging in.

Figure 5. Main Screen of the Rconsole1


a) Function Tabs
b) Status Display

3.1.1. Function Tabs


On the upper left corner in the main screen(a), three tab icons can be found as in [Figure
6].

Figure 6. Tab Icons of the Rconsole1

RConsole1_UM_COM_EN_02 10 / 124
The functions of these tabs are as follows.

- Manage the Worklist and Studylist.


 Worklist: List of patients.
 Studylist: List of examined studies.
- Refer to4. List for more details.
- Perform X-ray projection of the selected patient in
the Worklist
- Review the examined study of the selected patient
in the Studylist.
- Refer to 5. Exam for more details.
- Display examined images including inverted
images and markers.
- Refer to 6. Review for more details.

RConsole1_UM_COM_EN_02 11 / 124
3.1.2. Status Display
Status display icons can be found on the lower left corner in the main screen as in [Figure
7].

Figure 7. Status Display

①Hardware & Connection Status Display


Display the status of hardware and the connection status between equipment.

Send the log files of the Rconsole1 to manufacturer


via E-mail. The log files do not contain any private
information.
Display connection status and the available space
left on the currently connected HDD.
The displayed number indicates the remaining HDD
capacity.
Display the connection status (On/Off)with the
detector and the temperature of the detector.
Display the connection status with the Storage
Server.
When you click this icon, a pop-up window will
appear. You can change the PACS settings or
conduct a test in the new window. The UI is very
similar to the PACS menu of Configuration.
Refer to 7.7.3 PACS for more details.

RConsole1_UM_COM_EN_02 12 / 124
Display the connection status with the DICOM
Printer.
Refer to 7.7.4 Printer for more details.
Display the connection status with the Modality
Worklist Server.
Refer to7.7.2 WL Server for more details.
Display the connection status with the MPPS
server..

② Current Time & Log-in Information Display

 Display the current date, time and user name.


 Click the user name to change without logging out

③ Logout Icon

 This icon terminates the program or switches.

④ Configuration Icon

 Open the ‘Configuration Window’ for various settings.


 Refer to7. Configuration for more details.

RConsole1_UM_COM_EN_02 13 / 124
4. List
The List tab displays the lists of patients and studies, which can be divided into two
categories as described below.

 Worklist (prior to examination)


 Display the list of patients to be examined.
 You can add/edit/delete patients on the list.
 You can select the patient on the list to proceed with the examination. .
 When you run the Rconsole1 for the first time, you will be asked to set the length
of the list. The setting will be saved automatically.
 Refer to 4.1 Worklist for more details.

 Studylist (post-examination)
 Display the examined studies.
 You can edit/delete the examined studies on the list.
Note: The changes such as editing and deleting images on the list will not affect
data in the PACS server.
 You can transmit examined images to the PACS server.
 Refer to 4.2 Studylist for more details.

 Please note!
 The information on the Worklist and Studylist is stored on the Local Storage (HDD)
as default settings.
 Every time the Rconsole1 runs, a new Worklist is retrieved from the PACS server
and stored on the Local Storage.
 Only the contents on the list which are edited and deleted in the Rconsole1 affect
the Local Storage as default. However, edited records may be transmitted to the
PACS server using the ‘Send’ command.

RConsole1_UM_COM_EN_02 14 / 124
4.1 WorkList

Display the list of patients. The list is displayed by clicking the [WorkList] icon on the
upper left side of the List tab.

4.1.1. Main Screen of WorkList

The main screen of the Worklist is as seen in [Figure 8].

Figure 8. Worklist

RConsole1_UM_COM_EN_02 15 / 124
① List of Patients
 The types of information are as noted in [Table 2].
 The displayed information as well as its order can be set in ‘Configuration’.

 Click or drag the name of the information type (ID, Name, and etc.) to change the
order.

 If you run Rconsole1 for the first time, set the column size that fits your needs and
the setting will be saved automatically.

Table 2. Information Type of the Worklist


Type Description
ID Patient ID

Name Patient name

Study Desc. Study description


Accession Number.
Acc. No.
X-ray study registration number
Date / Time Time of Registration
D.O.B. Date Of Birth.
Sex Sex of the patient
Status Status of study

 Patient selection in the Worklist.


 Click the patient to select.
 The selected patient is displayed in brown.
 The patient by manual input can be edited or deleted afterwards.

 Projection can be performed in the Exam tab by double-clicking on the selected


patient.
 You can also click [Exam] icon after selecting a specific patient to perform
projection. Refer to 5. Exam for more details.

RConsole1_UM_COM_EN_02 16 / 124
② Search Bar & Refresh Icon

 Search Bar
 Search the entered value such as Patient, ID, Name, and Access Number in the
Worklist.

 If you click the [Search] icon after entering the value to be searched, only
the relevant works will be displayed on the list.

 If you click the [Search] icon without entering a specific value, an


‘Advanced Search Window’ will appear.
 Advanced Search Window

Figure 9. Advanced Search Window


 Search by specifying multiple options to be searched simultaneously.
 Only the search results that satisfy the entered details will be displayed.
.

 Refresh Icon
 Retrieve recently registered patients from the PACS server.

RConsole1_UM_COM_EN_02 17 / 124
③ Add/Edit/Delete/Load Icon

Table 3. Add/Edit/Delete/Load Worklist Icon


Icon Description

Add a new patient.


When clicking this icon, an ‘Add Examination Window’ will appear.

Add Refer to 4.1.2 Add Examination Window for more details.

Edit the selected patient information in the Worklist.


This function is only available for the patient info which has been
added by Rconosole1.
When clicking this icon after selecting the patient, an ‘EditWindow’
Edit will appear.
Refer to 4.1.3 Edit Window for more details.

Delete the selected patient.


This function is only available for the patient info which has been

Delete added by Rconosole1.

Load the Worklist as an EXCEL file.


Load
Worklist

④ Exam
 Switch to the Exam tab for examination of the selected patient.
 Refer to 5. Exam for more details.

RConsole1_UM_COM_EN_02 18 / 124
4.1.2. Add Examination Window

An ‘Add Examination Window’ appears when clicking the icon, as shown in [Figure
10] below.

Figure 10. Add Examination Window

RConsole1_UM_COM_EN_02 19 / 124
① Patient Information
Patient information is entered according to the category described in [Table 4].

Table 4. Patient Information (Add Window)


Category Description
ID Patient ID.
Accession number Registered identification number
Name Patient name
Operator Radiographer name
Gender Gender of the patient
Birth Date Date of birth of the patient
Age Age of the patient

② Projection Information
 Display the information related to projection.
 Display registered procedures.

③ OK
 Save the patient with the entered information and close the window.

④ Exam
 Save the patient with the entered information and directly switch to the Exam tab for
examination.
 Refer to 5. Exam for more details.

⑤ Cancel
 Cancel the work and close the window.

RConsole1_UM_COM_EN_02 20 / 124
4.1.3. Edit Window

The ‘Edit Window’ will appear when selecting a study from the Worklist and then clicking

the [Edit] icon.

Note: The ‘Edit Window’ is not activated if the study information is retrieved from the
PACS server.

Figure 11. Edit Window


① Study Information
 Edit the information of radiographer(s) and patient(s).
 The study information described in [Table 5] below is entered.

RConsole1_UM_COM_EN_02 21 / 124
Table 5. Study Information (Edit Window)
Category Detail
ID Patient ID
Name Patient name

Accession number Registered Identification Number

Operator Radiographer name


Study description Study description
Sex Sex of the patient
Birth Date Date of Birth of the patient
Age Age of the patient
Schedule & time Scheduled date and time of imaging

② OK
 Save changes and close the window.

③ Cancel
 Cancel changes and close the window

RConsole1_UM_COM_EN_02 22 / 124
4.1.4. Load Worklist (An EXCEL File)
Load a Worklist in an EXCEL file format.

Figure 12. Load the Worklist

① Path
 Locate the path where the EXCEL file is saved.

② List of the files


 Show the files containing Worklists which you can select.

③ Open
 Display the selected Worklist in the Worklist screen and close the window.

④ Cancel
 Cancel and close the window.

RConsole1_UM_COM_EN_02 23 / 124
4.2 Studylist

Display the examined study list. When clicking the [Study List] icon on the upper side of
the screen, the study list will be activated.

4.2.1. Main Screen of the Studylist


The main screen of the Studylist is displayed as seen in [Figure 13].

Figure 13. Main Screen of the Studylist

RConsole1_UM_COM_EN_02 24 / 124
① List of the Examined Patients
 Display the list of the examined patients’ information.
 Each row represents the information of one patient, and a single ID can have
multiple studies.
 The column of the list can be changed by dragging the names of the categories
(e.g. ID, Name, and etc.).
 Height of the list is adjustable using [Ctrl key] + Mouse wheel or using [Ctrl key] +
[Arrow key].
 Types of information are as noted in [Table 6].

Table 6. Information Types Regarding Patient


Category Description
Name Patient name
ID Patient ID.
Sex Sex of the patient
DOB Date Of Birth
Date/Time Date/Time for recent examined study

 Patient Selection
 Patient can be selected by left-clicking.
 The selected patient is highlighted in red. In the '① Examined Patient List' and '②
Examined Study List per Patient’, an activated study is highlighted in darker color,
while an inactivated study is highlighted in lighter color.
 The selected patients can be edited or deleted.

②Examined Study List per Patient (Series)


 Display detailed study information of the selected patient on ‘① List of the Examined
Patients.’
 Types of information are noted in [Table 7].

RConsole1_UM_COM_EN_02 25 / 124
Table 7. Information Types of Examined Study List
Items Description
Accession Number.
Acc. No.
Registered Identification Number
Study ID RIS code
StudyDesc. Study Description.
Status [Number of images] / [Number of projections]
Date/Time Date/Time for an examined study
[Number of images transmitted] / [Number of
PACS
images]
[Number of images printed] / [Number of
Printer
images]
[Number of images exported] / [Number of
Export
images]

③ Thumbnail

 Display the images of the selected patient.


 A check mark is added in the thumbnail of the selected images, for opening or
transmitting the image.
 The thumbnail can be selected by double-clicking on it.
 Images transmitted to the PACS Server are shown in green outlines.
 1/1: Series number / Image number.
 Indicators on the image
 S : PACS Send
 P : Print
 E : Export
 These indicators can be also found in the exam tab as well as review tab.

RConsole1_UM_COM_EN_02 26 / 124
④ Search Bar & Refresh Icon

 Search Bar
 Search the values of Patient ID, Name, and Access Number within the Studylist.

 If you click the search icon after entering the value to be searched, only
the relevant works will be displayed on the list.

 Refresh Icon
 Renew the list to show the latest results.

RConsole1_UM_COM_EN_02 27 / 124
⑤ Tool Icons

Icon or Button Description


Open detailed overview of a single selected image from the list
of thumbnails.
Open Study Refer to 6. Review for more details.

Repeat exam for the selected study in the Exam tab.


Refer to 5. Exam for more details.
Repeat Exam

Perform merging two studies of patients.


Refer to 4.2.2 Merge Window for more details.
Merge
Store images on a USB, CD, or DVD, and print through the
DICOM Printer.
Export Refer to 4.2.3 Export Window for more details.
Edit the selected patient or study information in the Studylist.
The ‘Edit Window’ appears when clicking the [Edit] icon.
Edit Refer to 4.2.4 Edit Window for more details.

Delete the selected item.


Delete

Save the patient list in an EXCEL file format.


EXCEL Export

⑥ Send Icon
 Transmit the selected images to the PACS server

RConsole1_UM_COM_EN_02 28 / 124
4.2.2. Merge Window
 Merge the images and studies.
 If a patient has multiple ‘Patient ID’, it is more convenient to group them by using this
function. Edit if the image and patient information do not match with each other.

Figure 14. Merge Window

① & ② Merge Subject Imaging Information


 Study lists appear on each side.

 Required items can be easily searched by using the search tool.

 Identical items selected in ① will be also located in ②if the same list is checked.

③ Merge Icon
 Transmit the selected items.
 Every change is saved while being transmitted.

④ Close
 Close the ‘Merge Window’.

RConsole1_UM_COM_EN_02 29 / 124
4.2.3. Export Window
Save the images of the selected study from the Studylist in a USB, CD, DVD or as PDF,
print them with DICOM Printer, and create a report.

4.2.3.1. Drive

Figure 15. Export Drive Window


 Store the images in the local drive.
 Only the files in BMP, JPG, PNG, TIF and DCM format can be exported.
 If a local drive is selected, the current lists of stored files are displayed.
 Export File Format
 Select the file format for storage.
 Image is stored in a USB when clicking the [OK] icon
 File name is automatically designated and it will be stored in 'Selected
folder/Patient ID/Acc No/' folder.

RConsole1_UM_COM_EN_02 30 / 124
4.2.3.2. Print

Figure 16. Export - Print Window


 Print through the DICOM Printer
 The DICOM Printer must be connected.

① Print Setting
 Printer Selection
 Select a printer from the drop-down menu.
 Film orientation
 Select a film orientation between portrait/landscape
 Film size
 Select the size of the film.
 Smart Layout
 Select the view position (CC, MLO, Left, Right) to display related images in the
screen. The screen is divided into the number of related images.
 Layout

RConsole1_UM_COM_EN_02 31 / 124
 Set the print layout.
 Zoom
 Perform zoom in/out of the image.

② Preview
 Thumbnail
 The image to be printed appears in the thumbnail list on the right.
 Preview
 If an image is selected in the thumbnail, a large image appears in the center as a
preview screen.
 Merge
 Merge the divided screen which has been separated by the ‘Divide’ function.
 Only adjacent screens of the same sizes can be merged.
 The merged shape must be a square.
 Divide
 Divide the screen by the value set in ‘Layout.’
 Rotate
 Rotate the screen 90 degrees clock-wise.
 Flip
 Reverse the up and down sides.
 Print
 Print out using general printer.
 Edit Overlay
 Edit overlay texts which are displayed in the Review screen.

③ DICOM Print
 Perform printing with the DICOM Printer.

RConsole1_UM_COM_EN_02 32 / 124
4.2.3.3. CD/DVD

Figure 17. Export - CD / DVD


 Store the images on a CD or DVD.
 Only DICOM files can be exported.
 A CD/DVD writer is required.

①CD / DVD Media Configuration


 Select the target media to export the image.
 CDs, DVDs, Dual-Layer DVDs can be used.
 Volume Label
 Set the name of the target media.
 Total Space
 Display the total available space of the target media.
 Insert
 Close the CD / DVD writer.
 Eject

RConsole1_UM_COM_EN_02 33 / 124
 Eject the tray of the CD / DVD writer.

② Export File List


 Display the list of files to be exported.

③ Export Process
 Estimated Time
 Present the expected time needed for Export.
 Time Left
 Display estimated time left to the completion of Export.

④ Task Icons
 Burn
 Perform the Export task.
 Cancel
 Cancel the Export task.

RConsole1_UM_COM_EN_02 34 / 124
4.2.3.4. Report Window
Save or print the report which contains the information about the exam.

Figure 18. Export - Report

① Patient and Exam Information


 Display the Patient and Exam information to save or print.

② Report Contents
 Edit contents
 Edit contents of the report.
 Title of contents can be changed in ‘Setting’.
 Image Preview
 Show loaded images.

③ Task Icons
 Print
 Display preview of report and print.

RConsole1_UM_COM_EN_02 35 / 124
 Save as PDF
 Save the report as PDF format.
 Save
 Save this report temporally
 Setting
 Show the ‘Report Value Setting Window’.

④ Add Image Window

Figure 19. Export - Report Load Images

 Load images from the Studylist.

RConsole1_UM_COM_EN_02 36 / 124
4.2.3.5 Report Value Setting Window

Figure 20. Export - Report Value Setting


① Default Value
 Edit default values of reports such as ‘Address’ or ‘Extra Info’.

② Visibility Setting
 Configure the visibility of items when using ‘Print’ and ‘Save as PDF’.

③ Set as default
 Set the current setting as the default setting
 Changed default setting will be applied from the next new reports.

RConsole1_UM_COM_EN_02 37 / 124
4.2.4. Edit Window

4.2.4.1. Edit Patient Information

Figure 21. Edit Window on Studylist (Patient Information)

 Only the patient information (ID, Name, Sex, Birth Date, and Age) can be edited,
when clicking the [Edit] icon in the patient list of the Studylist (upper part of the list).
 OK
 Save changes and close the window.
 Cancel
 Cancel changes and close the window.

RConsole1_UM_COM_EN_02 38 / 124
4.2.4.2. Edit Study Information

Figure 22. Edit Window inStudylist (Study Information)

 Only the study information (Accession Number, Study Description, Sex, and
Schedule date & time) can be edited when clicking the [Edit] icon in the projection list
(lower part of the list) of the Studylist.
 OK
 Save changes and close the window.
 Cancel
 Cancel changes and close the window.

RConsole1_UM_COM_EN_02 39 / 124
5. Exam
The Exam tab processes exam and edits images.

Figure 23. Main Screen of Exam Tab


① Projection List
 Display the scheduled exam list.
 Refer to 5.1 Projection List for more details.

② Patient Information
 Display ID, patient name, and study description.

③ X-ray Detector Status Display


 Display the status of the detector being used in exam.
 Blue: Ready for X-ray projection.
 Orange: Currently busy, not ready for X-ray.

④ Acquisition Settings
 Auto Sending

RConsole1_UM_COM_EN_02 40 / 124
 When the [Auto sending] icon is activated and every scheduled exam is
completed, send images to PACS server automatically.
 Skip review
 When the [Skip review] icon is activated, screen transition to Review screen does
not occur after the image acquisition.
 Implant processing
 When the [Implant processing] icon is activated, image processing parameter will
be changed to the one for implant only.
 Suspend
 Suspend current exam and transit to the Worklist screen.
 Send
 Transmit the images to the PACS server or folder.
 With a long-clicking on this icon, all examined images will be transmitted.
 Change detector
 Connect to another detector that is currently connected when clicking the button.

RConsole1_UM_COM_EN_02 41 / 124
5.1 Projection List

Display previously examined lists or projection lists to be examined.

Figure 24. Projection List

① Time Machine
 If there are remaining images of prior examinations in the Local DB, those images will
be displayed on the thumbnails.

② Projection Guide
 Projection guidelines for the selected projection.

RConsole1_UM_COM_EN_02 42 / 124
Figure 25. Projection Guide

 The presented projection guide can be closed by left-clicking.

③ Add / Delete / Edit Projection List


 Add
 ‘Add View Position’ option appears when clicking the [Add] Icon, as shown in
[Figure 26] below.

RConsole1_UM_COM_EN_02 43 / 124
Figure 26. Add View Position Window
 After selecting position, a projection list is added by clicking [OK].
 Delete
 After being selected, a projection list is removed by clicking [Delete].
 Edit
 After selecting a list, the same screen as the ‘Add View Position’ appears when
clicking the [Edit] Icon.
 Select the list that you want to add and the click the [OK] icon to apply the change.

④Projection List
 Display previously examined list or projection list to be examined.
 After selecting a projection list, examining X-ray shows the Review Screen and
displays the examined images on the Viewer.
 Double-clicking on an image brings the Review Screen.
 If the manual dose input option is checked, you can input through the dose setting
window at the bottom of each projection.

RConsole1_UM_COM_EN_02 44 / 124
6. Review
Perform editing the examined images.

Figure 27. Main Screen of Review Tab

① Image Display Screen


 Area of displaying images.

② Patient Information
 Display ID, patient name, and study description.

③ Tools, Processing, and Truview ART Tab


 Display a series of images to be examined, with various tool icons.
 Set the post-processing parameters which will be applied to the images or the
Truview ART Application.
 Refer to 6.1 Tools, Processing, and Truview ART for more details.

RConsole1_UM_COM_EN_02 45 / 124
④ Save / Pause / Send Icons
 Save
 Store the images in the Local Storage.
 Pause
 Pause the projection under progress and return to the Worklist screen.
 The paused projection list is displayed in light brown color.
 Send
 Transmit the images to the PACS server or folder.
 With a long-clicking on this icon, all examined images will be transmitted.

RConsole1_UM_COM_EN_02 46 / 124
6.1 Tools, Processing, and Truview ART Tab

The Tools, Processing, and Truview ART Tab can be used through the tab icons on the
upper right corner.

Figure 28.ToolsTab (Left), Processing Tab (Center), Truview ART Tab (Right)

RConsole1_UM_COM_EN_02 47 / 124
6.1.1. Tools Tab
 If there are images examined in the past, they can be displayed according to the
selected date.
 Display a list of studies categorized by projections.
 Display a collection of image editing icons.

Figure 29.Tools Tab

① Time Machine
 Display the images examined in the past.
 Note: There must be the images in the Local Storage.

② Projection List or Projected List


 Additional tabs will be created if the number of procedures is more than two.
 Display only the position of the selected procedures.
 Verify thumbnails and positions.

RConsole1_UM_COM_EN_02 48 / 124
③ Toolbar
 A collection of icons required for image analysis and editing.
 Display the currently configured list of icons.
 The icons can be edited on the ‘Edit Toolbar Window’.
 The [Edit] Icon is used to edit the Toolbar window.
 Refer to 6.1.1.1 Edit Toolbar for more details.

RConsole1_UM_COM_EN_02 49 / 124
6.1.1.1. Edit Toolbar

Figure 30. Edit Toolbar Window

① Applied Icons Box


 The currently applied icons in the Exam tab are displayed in this box.
 Icons in this box can be added by dragging& dropping the icons into ‘② Tools
Collection on the right.’
 Icons can be added into this box by dragging icons into ‘② Tools Collection’ and
dropping them on any empty location.

② Available Icons
 All the available image editing icons are listed.
 Add tools to ‘① Applied Icons Box’ through drag & drop.
 Previously added icons are displayed in dark color.

RConsole1_UM_COM_EN_02 50 / 124
③ Add Shutter Icon
 Add the icons with ROI Shutter Sizes specified by the User.
 The following window appears when clicking the icon.

Figure 31. Add Shutter Size Window


 Enter the size (Width and Height) and click [OK]to add a new ROI.

④ Shutter Delete Icon


 Delete the Shutter size added by the User only.
 Delete when you drag and drop the added Shutter icon onto this Trash Bin-shaped
icon.

⑤ OK
 Apply the modified tools arrangement and close the window.

⑥ Cancel
 Cancel and close the window.

RConsole1_UM_COM_EN_02 51 / 124
6.1.1.2. Toolbar Icons
The functions of all the icons in the toolbar are described in [Table 8].

Table 8. Toolbar Icons


Icon Description

Display the text entered by the User on the screen.

Set the ROI size to a userspecified size.

Display the “RCC" marker on the screen.

Display the “LCC" marker on the screen.

Display the “RMLO" marker on the screen.

Display the “LMLO" marker on the screen.

Display the screen in 1×1.

Divide the screen horizontally, in 2×1.

Divide the screen vertically, in 1×2.

Divide the screen in 2×2.

Display the "↑" marker on the screen.

Display the "←" marker on the screen.

RConsole1_UM_COM_EN_02 52 / 124
Display the "↓" marker on the screen.

Display the "→" marker on the screen.

Verify the DICOM Header information.

Arrange windows side-by-side symmetrically.

Arrange windows up-and-down symmetrically.

Rotate the image 90 degrees counter-clockwise.

Rotate the image 90 degrees clockwise.

Fit image to screen.

Zoom-in and zoom-out the image.

Draw arrow annotation.

Partially expand the image.

Move the image up/down/left/right.

Duplicate the image.

Adjust the window level of the image.

Invert black and white of the image.

RConsole1_UM_COM_EN_02 53 / 124
Capture the image.

Delete the image.

Reset the edited image into its original state.

Set the ROI into a square shape.

Set the ROI into a circular shape.

Space for user-selected texts.

Show the ‘Export Window’. Refer to 4.2.3 Export Windowfor


more details.

Delete the selected annotation.

Measure the length of the image.

Ruler calibration.

Take the angle of the image.

Show location and pixel value for biospy.

Adjust mouse sensitivity.

Select annotation.

Draw free line annotation.

RConsole1_UM_COM_EN_02 54 / 124
Draw rectangular annotation.

Draw polygon annotation.

Draw ellipse annotation.

Display the present / the latest of the selected position.

Display RCC / LCC.

Display RMLO / LMLO.

Display RCC / RMLO.

Display LCC / LMLO.

Apply implant processing.

Show the ROI in the size of 18x24

Show the ROI in the size of 24x30

Save the location of market as default setting

Save the annotations applied in the image.

Save the location of the nipple by aligning it manually

Adjust the size of the text entered by the User.

RConsole1_UM_COM_EN_02 55 / 124
6.1.2. Image Processing Tab
* Function for advanced users *
 It is utilized to acquire a more optimized image by adjusting the post-processing
parameters of the completed image.

Figure 32.Image Processing Tab

① Image Enhancing Parameters


 Adjust the image enhancing parameters.
 The adjustable parameters are noted in [Table 9].

Table 9. Image Enhancing Parameters


Image Processing Description
Contrast Adjust the contrast level.
Sharpness Adjust the sharpness level.
Brightness Adjust the brightness level
Frequency Adjust the frequency
Fat Suppression Adjust the fat suppression level.

RConsole1_UM_COM_EN_02 56 / 124
② Parameter Functions
 Wizard
 Open the multi-image viewer window.
 8 types of images are shown based on the currently set parameter.
 Apply
 Apply the currently set parameters to the image.
 Reset
 Reset the parameter to its initial setting.
 Save parameter
 Save the currently set parameters and apply them when performing the projection
of the relevant projection-position.

RConsole1_UM_COM_EN_02 57 / 124
6.1.3. Truview ART
* Function for advanced users *

Figure 33.Truview ART Tab


① On / Off
 Configure whether to apply Truview ART.

② Preview
 Preview zooming in the selected part on the Viewer.

③ Compare
 Compare the two images with and without applying Truview ART.

④ Save
 Save Truview ART parameters.

RConsole1_UM_COM_EN_02 58 / 124
7. Configuration
Configure the user settings of Rconsole1.

Click on the bottom left to configure.

 Administrator privileges are required for configuration.


 Contents of configuration are saved and applied immediately following any changes
made.

Categories of Configuration are as noted in [Table 10].

Table 10. Configuration Category


Category Description
Configure language, date display, hospital information, and
System
etc.
Configure image storage folder, deletes policy, annotation
Image
information, and etc.
Storage Set configuration related to PC storage (HDD).
Account Manage the accounts of Rconsole1 users.
Configure unnecessary RIS code filtering subsequent to
Blocked List
ultrasound or other devices.
Dataset Configure the DICOM Header Format.
Configure the network information such as Local, PACS,
Network
Printer, and etc.
Configure the detector which will be connected with the
Detector
Rconsole1.
RIS Code Configure the RIS code.
Recycle Bin Manage deleted images

RConsole1_UM_COM_EN_02 59 / 124
7.1 System

 Perform configurations affecting the overall Rconsole1.

Subcategory Description
Language Select system language.
Format Select date format for presentation.
Institution Enter the information of the institution using the Rconsole1.
Options Change settings for options.
Workflow Change settings for work flow
Exposure Change settings for exposure condition.
Screening Bus Configure the options related to mobile screening bus

RConsole1_UM_COM_EN_02 60 / 124
7.1.1. Language

Figure 34. Language Configuration

RConsole1_UM_COM_EN_02 61 / 124
 Configure the system language.
Items Description
Korean Set when you use the Korean window.
Chinese Set when you use the Chinese window.
Japanese Set when you use the Japanese window.
Portuguese Set when you use the Portuguese window.
Spanish Set when you use the Spanish window.
German Set when you use the German window.
Turkish Set when you use the Turkish window.
Swedish Set when you use the Swedish window.
French Set when you use the French window.
Russian Set when you use the Russian window.
Ukrainian Set when you use the Ukrainian window.
Italian Set when you use the Italian window.
Polish Set when you use the Polish window.
Traditional
Set when you use the Traditional Chinese window.
Chinese
Finnish Set when you use the Finnish window.

RConsole1_UM_COM_EN_02 62 / 124
 DICOM character set:
 Set the DICOM character set. If language is selected in the program installation
sequence, this will be set automatically.
 If this item is not correctly changed, some tag value of DICOM header will be not
correctly contained.
Language DICOM character set
Korean ISO IR 6\ISO 2022 IR 149
Chinese ISO IR 6\GB18030
Japanese ISO 2022 IR 13\ISO 2022 IR 87
Turkish ISO IR 148
Default ISO IR 100
Thai ISO 2022 IR 166
Ukrainian, Russian ISO IR 6\ISO IR 144
UTF-8 ISO IR 192
Japanese \ISO 2022 IR 87

7.1.2. Format

Figure 35. Format Configuration

RConsole1_UM_COM_EN_02 63 / 124
① Date Format
 Configure the date format.
 Date
 Y: Year , M: Month, D: Day
 Delimiter
 Set the symbol to be used between Y, M, and D.
 For example, if today is May 31st, 2012,

Date Delimiter Final Notation


MM DD YYYY / 5/31/2012
MM DD YYYY - 5-31-2012
MM DD YYYY . 5.31.2012
YYYY MM DD / 2012/5/31
MM DD YY - 5-31-12
YY MM DD . 12.5.31
② Name Format
 Configures the name Format.
 If checked, name is displayed on the screen according to the name format of DICOM.
 You can select the type of name displayed on the screen

Type Description
Alphabetic Alphabetic character Type
Ideographic Ideographic character Type
Phonetic Phonetic character Type

 You can select the Family name, Given name, Middle name.

Figure 36. Add Examination

RConsole1_UM_COM_EN_02 64 / 124
7.1.3. Institution

Figure 37. Institution Configuration

 Input hospital/institution information using the Rconsole1.


 Hospital
 Hospital name
 Department
 Department name of the hospital
 Station
 Station name
 Referring physician’s name
 Set the physician’s name who recommended the exam.
 Performing physician’s name
 Set the physician’s name who performed the exam.

RConsole1_UM_COM_EN_02 65 / 124
7.1.4. Options

Figure 38. Options Configuration

 Auto Start
 If checked, Rconsole1 will automatically start when Windows starts.
 Virtual Keyboard
 If checked, virtual keyboard will be available for use.

RConsole1_UM_COM_EN_02 66 / 124
7.1.5. Workflow

Figure 39. Workflow Configuration


 Add Procedure Option
Option Description
Manual selection of exam
Add a manually selected procedure
procedure
Auto selection of exam
Add a automatically fixed procedure
procedure

 Set PID Option


Option Description
Create Patient ID based on
Use a PID based on Date
date/time.
Use a PID based on Mammo bus Create Patient ID based on
index index

RConsole1_UM_COM_EN_02 67 / 124
 Exposure Condition Input Option
 Select the exposure condition options.
 The exam window composition is changed according to the selected option.

Option Description

None Exposure condition is not used.

kVp mAs Use kVp and mAs value.

Use all exposure condition parameters (kVp, mAs, mA,


All
Thickness, AEC Mode, CompForce, Grid, and Target Filters)

 Send Option Selection


 Select between ‘Send PACS’ and ‘Send Folder’ after acquisition.
 The [Send] icon will be changed by the selection of this option.

Option Description Send Icon

Send PACS Send to PACS

Copy DICOM files to the designated


Send Folder
folder

 Send Folder Path


 Select the folder to send the files to.

RConsole1_UM_COM_EN_02 68 / 124
7.1.6. Exposure Condition

Figure 40. Exposure Condition

 Add / Edit / Delete types of Grid.


 Configure Focal Spot settings.
 Set a default value for object thickness and compression rate.

① Edit Grid List


 Configure Grid settings by typing a Grid in the text box.
 Add: Add the Grid to the Grid List.
 Edit: Change selected Grid name in Grid List to the name in the text box.
 Delete: Delete the selected Grid in Grid List.

 Select the Grid to display in exam window.

 Add the selected Grid to the Grid section using button.

 Delete the selected Grid from the Grid section using button.

RConsole1_UM_COM_EN_02 69 / 124
② Focal Spot, Thickness, Compression rate settings
 Focal Spot Settings
 Large FS : Set Large Focal Spot value.
 Small FS : Set Small Focal Spot value.

 Thickness
 Set a default value for thickness.

 Compression
 Set a default value of compression rate.

RConsole1_UM_COM_EN_02 70 / 124
7.1.7. Screening Bus

Configure options related to the screening bus.

Note: “Use a PID based on Mammo bus index” option must be selected in PID option in
Workflow in order to use this menu.

Figure 41. Screening Bus Setting


 ID setting
 Configure the date format
Date Format Example
YYYYMMDD 20160514
MMDDYYYY 05142016
DDMMYYYY 14052016
YYMMDD 160514
MMDDYY 051416
DDMMYY 140516

RConsole1_UM_COM_EN_02 71 / 124
 Input station number.
 Data index consists of 4-7 digits and it can be initialized by clicking the ‘Reset ID’
button.
 Use '-'
 When ‘Use ‘-‘’ is checked, ‘-‘is inserted between each item value.
 Example) 20160514-1234-000001 / 201605141234000001
 Setting the same accession number with patient ID
 When this option is checked, the same accession number is generated as the
created patient’s ID.
 When not using this option, an accession number is generated with the number
‘1’added after the created patient’s ID.

RConsole1_UM_COM_EN_02 72 / 124
7.2 Image

 Allow configuration relevant to examined images.

Subcategories Description
Image Folder Set the location to save the images.
Delete Option Set delete options for image files.
Overlay Set overlay on images.
Import Import images to the Studylist.
Export Set image export options
TRUVIEW ART Enable or disable the Truview ART.
Crop Option Set edge cut options.

RConsole1_UM_COM_EN_02 73 / 124
7.2.1. Save Option

Figure 42. Save Option Configuration

 Set the location to save the images in the operating PC


 Edit
 Open the folder selection window.
 Select the folder where images are to be stored.
 DCM Save Option
 When ‘Save raw DICOM automatically’ is checked, RAW DICOM file will be
saved together with general DICOM file.
 To send RAW DICOM to PACS server, ‘Save raw DICOM automatically’ option
must be checked.

RConsole1_UM_COM_EN_02 74 / 124
7.2.2. Delete Option

Figure 43. Delete Option Configuration

① Image Delete Option


 Use Password
 If checked, deletion of image requires a password.
 Use Reject Code
 If checked, a reject code (reason for deletion) is required whenever attempting to
delete an image file.
 Use Trash Bin
 If checked, deleted images are stored in the Trash Bin.

RConsole1_UM_COM_EN_02 75 / 124
② Reject Code
 List of reasons for deletion.
 Delete
 Delete the selected Reject Code from the list.
 Add
 Add a new Reject Code.
 Edit
 Edit the selected Reject Code.

7.2.3. Overlay
 Allow overlay configuration.

Figure 44. Overlay Configuration

 Editing types of overlay to be displayed on the Top and Bottom.


 The editing method is the same while the displaying locations differ.

RConsole1_UM_COM_EN_02 76 / 124
Figure 45. Edit Overlay

① Registered Overlay Type & Font Size


 Overlay: Information which will be displayed.
 Size: Font size of the relevant information.
 Information is displayed in accordance with the arrangement on the list.

② Edition


 Add a new overlay to the list.
 The following window as shown in [Figure 46] appears when clicking the icon.

RConsole1_UM_COM_EN_02 77 / 124
Figure 46. Overlay Option

 The selected overlay is added after entering the size and clicking [OK].


 Delete the selected annotation from the list.


 The selected annotation moves up by one row.


 The selected annotation moves down by one row.

RConsole1_UM_COM_EN_02 78 / 124
7.2.4. Import

Figure 47. Import Configuration

 Import an external image into the Studylist.


 Import
 Open the file selection window.
 Select the image to be imported.

RConsole1_UM_COM_EN_02 79 / 124
7.2.5. Export

Figure 48. Export Setting Screen


 Set the option to export DICOM images.
 If checked, DICOM Viewer is also created when exporting DICOM images.

RConsole1_UM_COM_EN_02 80 / 124
7.2.6. TruviewART

Figure 49. TRUVIEW ART Configuration

 Set the scope for applying TRUVIEW ART.


 Coverage
 Do not apply: Do not apply the TRUVIEW ART for image processing.
 Depends on parameter: The use of TRUVIEW ART depends on the value of post-
processing parameters.
 All position: TRUVIEW ART will be applied for image processing regardless of the
value of post-processing parameters.

RConsole1_UM_COM_EN_02 81 / 124
7.2.7. Image Crop Option

Figure 50. Image Crop Option Configuration

 Configure image crop option settings.


 Use EdgeCut Auto crop option
 Automatically crop the edges of an image to preset values.
 Refer to 7.8.2 ⑤ Edge Cut Value for EdgeCut value settings.
 Use Magnification paddle Auto crop option
 Crop image to magnification paddle ROI values.
(1 = 1 Pixel)

RConsole1_UM_COM_EN_02 82 / 124
7.3 Storage

 Configurations concerning the deletion and storage of data in the operating pc where
the images will be saved.

Details Description
Configure the deletion and automatic deletion function of
Delete
images stored in Local Storage.
Compress and backs-up the images stored in Local
Backup
Storage.

RConsole1_UM_COM_EN_02 83 / 124
7.3.1. Delete

 Configure the deletion and automatic deletion function of images stored in Local
Storage.

Figure 51. Delete Configuration

① Storage Delete
 The [Delete] Icon deletes all images stored in the Local Storage.
 Deleted images cannot be restored, so please make sure to perform a backup prior to
deleting images!

② Auto Delete
 Auto Delete
 If checked, all images satisfying the pertinent conditions are automatically deleted
from the PC storage.
 Disk Size

RConsole1_UM_COM_EN_02 84 / 124
 If checked, images in the PC storage are automatically deleted once the used
disk space exceeds the set limit.
 Max: maximum limit of disk space set for storage of images in the PC storage.
 Period
 If checked, images in the PC Storage are continuously deleted automatically,
during the set period below.
 ex1) Set to "2", "Month": Delete all images that were stored 2 months earlier as of
today
 ex2) Set to "1" "Day": Delete all images that were stored 1 day earlier as of today.
 The cycle may be set by one of the following;Day/Week/Month/Year
 Only one of the two auto deleting methods, ‘Disk Size’ and ‘Period’, may be selected
at a time

RConsole1_UM_COM_EN_02 85 / 124
7.3.2. Backup

 Compress and back-up the images stored in Local Storage.

Figure 52. Backup Configuration

① Database Backup Configuration


 Display the location of the folder used for Data Backup.
 Edit
 Open the selection window for choosing a folder to be used in backup.
 Restore
 Restore the selected database file and store it in the system
 Backup
 Back-up the currently used database in the selected folder.

RConsole1_UM_COM_EN_02 86 / 124
②Automatic Backup
 Auto Backup Check
 If checked, automatic backup is performed during the set period explained below.
 Period
 Set the automatic backup period.
 Available units consist of Day/Week/Month.

RConsole1_UM_COM_EN_02 87 / 124
7.4 Account

 Configure the account of the Rconsole1 user.

Figure 53. User Account Configuration

 The currently registered user information is available on the list.


 Configure the user ID, password, and authorization.
 Delete
 Delete the selected user account.
 Add
 Add a new user account.
 Edit
 Edit an existing user account.

RConsole1_UM_COM_EN_02 88 / 124
7.5 Blocked List

 Filter unnecessary RIS codessuch as codes related to ultrasound or other devices.

Figure 54. Blocked List Configuration

 The currently registered Blocked List information is available on the list.


 Delete
 Delete the selected Blocked List entry.
 Add
 Add a new entry to the Blocked List.
 Edit
 Edit the existing Blocked List entry.

RConsole1_UM_COM_EN_02 89 / 124
7.6 Dataset

 Configure the DICOM Header Format.

Figure 55. Dataset Configuration


① DICOM Header
 CR and DR are separately applied.

② Image Bits per pixel


 Set the BPP of the images.

③ Series Order
 Set the series storing method for the Study routine.
 One series: Storing as one series.
 Each series: Storing as individual series.

RConsole1_UM_COM_EN_02 90 / 124
7.7 Network

 Configure network information such as IP addresses.

Subcategory Description
Network connection is required.
Configure the network information of the
Local
PC (or Work Station) currently operating
Rconsole1.
Network connection is required.
MWL Configure the network information of the
server storing the Modality Worklist.
Network connection is required.
Configure the network information of the
Storage
Storage Server, which stores the image
files.
Printer connection is required.
Printer Configure the connection information of
the printer.
Network connection is required.
MPPS Configure the network information of the
MPPS Server.

RConsole1_UM_COM_EN_02 91 / 124
7.7.1. Local

 Configure the network information of the PC (or Work Station) currently operating
Rconsole1.

Figure 56. Local Configuration


 AE Title
 Enter name of the local PC (or Work Station) which will be displayed on the
Network.
 IP
 Enter the IPv4 Address of the PC (or Work Station) currently operating Rconsole1.
 Port
 Enter the Port number.
 It should be configured in accordance with the linked server.
 All entered information will be automatically saved and applied.

RConsole1_UM_COM_EN_02 92 / 124
7.7.2. MWL

 Modality Worklist Server.


 Set the server retrieving patient lists.

Figure 57. MWL Configuration


①MWL Server
 Manage the information of server being accessed.
 List of AE Title (Access Server Name), IP, and Port information
 Test
 Verify the connection with the selected server.
 Delete
 Delete the selected server from the list.
 Add
 Add an MWL server.
 Edit

RConsole1_UM_COM_EN_02 93 / 124
 Edit the selected server information.

②Option
 Time-out
 Response standby time with the server.
 It is regarded as connection failure, if the server is unresponsive longer than the
set period of time.

③ Query
 Configure the automatic refresh settings of MWL.
 Auto
 MWL is automatically refreshed during the entered interval.
 Manual
 Set MWL to be refreshed manually.
 Interval
 Set the interval for automatic refresh.

④ Modality
 Select the modality type of the imaging equipment.
 E.g. CR, DR, ALL
 Add
 Add Modality.
 Delete
 Delete the selected Modality.

⑤ Date
 Configure the date format for MWL
 Date
 Y: Year , M: Month, D: Day
 Delimiter
 Set the symbol used between Y, M, and D.

RConsole1_UM_COM_EN_02 94 / 124
 For example, if today is May 31st, 2012,

Date Delimiter Notation


MM DD YYYY / 05/31/2012
MM DD YYYY - 05-31-2012
MM DD YYYY . 05.31.2012
YYYY MM DD / 2012/05/31
MM DD YY - 05-31-12
YY MM DD . 12.05.31

RConsole1_UM_COM_EN_02 95 / 124
7.7.3. PACS

 Configure the Storage Server (PACS).

Figure 58. Storage Server Configuration


① Storage Server
 Manage the information of accessing storage server.
 It consists of AE Title (Access server name), IP, Port, and Selection information.
 DICOM images are sent to the PACS server if the selection item of the server is ‘O’.
More than one PACS server can be selected at a time.
 Test
 Verify the connection with the selected server.
 Delete
 Delete the selected server from the list.
 Add
 Add an access server.

RConsole1_UM_COM_EN_02 96 / 124
 Edit
 Edit the selected server information.

② Option
 Time-out
 Response standby time with the server.
 It is regarded as a connection failure if the server is unresponsive longer than the
set period of time.

③ DICOM Type
 Configure the DICOM type

DICOM Type Description


Send the DICOM file which is applied the image
processing.
DICOM If MG modality is selected at Dataset, DICOM Tag
Presentation Intent Type (0008:0068) will be set “FOR
PRESENTATION”.
Send the Raw DICOM file which is not applied the
image processing.
To send the Raw DICOM file correctly, “Save raw
DICOM automatically” option must be checked in
Raw DICOM
“Image” ->“Save Option”.
If MG modality is selected at Dataset, DICOM Tag
Presentation Intent Type (0008:0068) will be set “FOR
PROCESSING”.

RConsole1_UM_COM_EN_02 97 / 124
7.7.4. Printer

Figure 59. Printer Configuration

①Printer Information & Printer Selection for Connection


 Manage the information of the printer to be connected.
 Select the printer from the list.
 Add
 Add a printer.
 Edit
 Edit the selected printer information.
 Delete
 Delete the selected printer from the list.
 Test
 Verify the connection with the selected printer.

RConsole1_UM_COM_EN_02 98 / 124
② Overlay Information
 Select the additional annotation information added on the film.
 The following pop-up window shown in [Figure 60] appears when clicking the [Edit]
icon.

Figure 60. Printer Overlay Information Edit

RConsole1_UM_COM_EN_02 99 / 124
7.7.5. MPPS

 Configure the MPPS server to connect to.

Figure 61. MPPS Configuration


① MPPS Server
 Manage the information of the server to connect to.
 This section consists of AE Title(Access Server Name), IP, and Port information.
 Test
 Check the connection to the selected server.
 Delete
 Delete the selected server from the list
 Add
 Add the server to connect to.
 Edit
 Edit the information of the selected server

RConsole1_UM_COM_EN_02 100 / 124


② Option
 Station AE Title
 Internally used name of the MPPS server
 Timed out
 Waiting time for receiving response from the server
 It is considered to be connection failure if no response is found from the server
during the period that has been set.

RConsole1_UM_COM_EN_02 101 / 124


7.8 Detector

 Detector related configurations.


 Run Calibration Tool programs.

Subcategory Description
Register and edit the model of panel
(detector) being used.
General
Execute detector software, RCali1 and
Terminal Emulator.
Detailed configuration screen related to
Panel A, registered in ‘General’ tab.
Panel A Detailed configuration items for panel are
created in accordance with the number of
panels registered in ‘General’.

RConsole1_UM_COM_EN_02 102 / 124


7.8.1. General
 Register and edit the panel (detector) to be used.
 Execute calibration software, RCali1.

Figure 62. General Configuration

①Panel(Detector) Model Registration and Editing


 Register the panel to be connected.
 Add
 Register the detector selected from ‘Model Select’.
 Edit
 Replace the currently registered detector with the detector selected from the
‘Model Select’.
 Delete
 Delete the most recently registered detector.
② RCali1
 Press the [Run] icon to execute the RCali1 program.

RConsole1_UM_COM_EN_02 103 / 124


7.8.2. Detailed Configuration for Panel
 Perform detailed configuration for each panel (detector).
 A menu such as ‘Panel A’ and ‘Panel B’ are generated in accordance with the
number of panels registered in ‘General’.

Figure 63. Detector Detailed Configuration


① Model

 Automatically set by the value registered in ‘General’ tab.

② Serial Number

 Enter the serial number of the Detector.


 It should be identical to the Map file name.

RConsole1_UM_COM_EN_02 104 / 124


③ Grid Type

 Select type of grid.

④ Sensitivity and Offset

 Configure sensitivity and offset.

⑤ Edge Cut Value

 Set Edge Cut values.


 If Edge Cut Auto Crop function is activated, edges of an image are automatically
cropped to these values.
 Configure Edge Cut Auto Crop settings in 7.2.7 Image Crop Option.
(1 = 1 Pixel)

RConsole1_UM_COM_EN_02 105 / 124


7.9 RIS Code

 RIS (Radiology Information System) Code is a collection of projections.


 It regulates contents in areas. For example, “A, B, C regions are to be imaged for RIS
Codes for certain cancers.”
 It allows the rearrangement of sequence of projections.

Figure 64. RIS Code Configuration

 RIS Code
 The [Edit] icon shows ‘RIS Code Edit Window’.
 RIS Code Import from Excel
 Add RIS Code values from an EXCEL file.
 Note: Duplicate code values are overwritten with contents of the Excel file. (It
overwrites the existing data)
 Detector Rotation
 Rotate every image by 90 degrees.
 Marker size change
 Change the size of every marker.

RConsole1_UM_COM_EN_02 106 / 124


 Digital marker option
 Select whether to use digital marker.

7.9.1. Edit RIS Code Window

Figure 65. RIS Code Edit Window


① RIS Code List
 Manage the codes and provide description of RIS Codes.
 Add stars to the codes to designate them to be displayed in priority when adding
patients (Up to 6 items)
 Text Box
 Used to view the Code and Descriptions when adding or editing an RIS Code.

 3 task icons are available. The descriptions are noted in [Table 11].

RConsole1_UM_COM_EN_02 107 / 124


Table 11. RIS CodeListManagement Icons
Icon Description

Edit the code and descriptions of the RIS Code selected from
the list, with the information entered in the text box below the
icon.

[Editing Sequence]
Edit
1) Select the code to be edited from the list.
2) Enter new information in the textbox below the icon.
3) Click the [Edit] icon.

Add the RIS Code with the information entered in the textbox
below the icon.

[Adding Sequence]
Add
1) Enter new information in the textbox below the icon.
2)Click the [Add] icon.

Delete the RIS Code selected from the list.

[Deleting Sequence]
Delete 1)Select the code to be deleted from the list.
2)Click the [Delete] icon.

② Editing positions included in the RIS Code


 Manage the positions to be included in the selected RIS Code.
 Descriptions regarding the task icons are noted in [Table 12].

RConsole1_UM_COM_EN_02 108 / 124


Table 12. RIS CodePosition Management Icons
Icon Description

Add the position relevant to the selected Projection /


Position in the RIS Code.
Add

Delete the selected position from the RIS Code.


Delete

Move the selected position up by one row.


Move up

Move the selected position down by one row.


Move down

③ Imaging Environment Configuration


 Enter the automatically configured environment value when imaging of the pertinent
RIS Code is performed.
 Paddle Type
 Select the paddle type for the selected position.
 When the selected position used for magnification, please select “Magnification”
item.
 AEC
 Set AEC function.
 Grid
 Select the Grid Type being used.
 Focal Spot
 Set the focus. (Large / Small)
 Density
 Set the density.
 SID
 Set the distance between the Detector and X-ray Tube.
 kVp, mA, mAs(mSec)
 Set the X-ray Generator condition.

RConsole1_UM_COM_EN_02 109 / 124


 Parameter
 Locate the parameter file of image processing which will be applied after taking
image.
 Rotation
 Set the rotation rate of the image.
 Flip H, Flip V
 Select whether to flip the image horizontally or vertically.
 Shutter
 Select the ROI Size.
 Marker
 Set and managethe markers, which will be automatically inserted in the image.
 Descriptions regarding the Marker Icon are noted in [Table 13].

Table 13. Marker Icon


Icon Description

Add a new Marker.


Add

Delete the selected marker.


Delete

Move the selected Marker up.


Move up

Move the selected Marker down.


Move down

④ View Position
 Set the position of the RIS Code projection.

RConsole1_UM_COM_EN_02 110 / 124


7.9.2. Add Marker Window

Figure 66. Add Marker Window


 Location
 Select the location for the marker to appear on the image.
 Text
 Select the type of Marker.
 In addition to the texts provided by default, random texts can be used as a marker.
 User-registered markers can be selected from the list.
 The Marker list window can be edited using the [Add] Icon and [Delete] Icon.

RConsole1_UM_COM_EN_02 111 / 124


7.10 Trash Bin

 Manage deleted images.


 If “Use Recycle Bin”is checked in Image-Delete tab in Configuration setting, Trash Bin
setting is available.

Figure 67. Trash Bin Setting


 When deleting the image after selecting ‘Use Recycle Bin’ option, you can check the
information of the images from the Trash Bin Tab.

① Deleted image information


 List the deleted images
 Search by the date by clicking Search button.
 Select all the images in the list by clicking Select All button.
 Deselect all the selected list by clicking Deselect All button.

RConsole1_UM_COM_EN_02 112 / 124


② Trash Bin
 Delete
 Delete the selected images in the Trash Bin.
 Restore
 Restore the selected images in the Trash Bin.
 Statistics
 Check the statistics of deleting history with their reasons to be deleted.

RConsole1_UM_COM_EN_02 113 / 124


7.10.1. Statistics
 Statistics of deleted images are provided by using reject code..

Figure 68. Statistics


① Information of the rejected images
 List the statistics by reject code.
 Search the image by the date using Search button.
 Information such as deleted position, reasons for rejection, etc. is described.

RConsole1_UM_COM_EN_02 114 / 124


② Export statistics
 Export
 Save the current statistics list as Excel file..
 File Name: Rejected_Image_Statistics_Date of today_(Search Period).xls.
 In the excel file, 3 types of statistics are provided. .
 Statistics : Statistics list of all rejected images
 Reject Reason : Statistics list of reasons for rejection
 Reject Position : Statistics list of rejected positions

Figure 69. Reject Position Statistics Results


 Cancel
 Close statistics screen.

RConsole1_UM_COM_EN_02 115 / 124


Appendix A. Glossary

Terms Description

Digital Radiography Operating Console.


DROC
Software for the operations of a digital X-ray Detector.
Picture Archiving and Communications System.
PACS System for storing, interpreting and accessing medical
imaging information.
Digital Imaging and Communication in Medicine.
DICOM Standards and regulations regarding medical digital
imaging and communications.
DICOM Directory.
DICOMDIR DICOM directory information file is stored in the root
folder of the media (CD/DVD).
Hospital Information System.
HIS Information of registration/administrative system,
including OCS, PACS, etc.

OCS Order Communication System.

Radiology Information System.


RIS
Operated as part of HIS or independently.
Modality Worklist.
MWL DICOM standards for sharing scheduled examinations
information.
GPU Graphic Processing Unit.
Graphical User Interface.
GUI Computer interface for users, developed by graphic
elements.
Detector Flat Panel Digital X-ray Detector.
Program run for installation, maintenance and removal of
Installer
a specific application program.

RConsole1_UM_COM_EN_02 116 / 124


Process authenticating user for permission to access the
Login
system.
Texts and various shapes marked on the completed
Annotation
image.
Lookup Table.
LUT
Table of previously calculated values.
Region of Interest.
ROI
Actual region necessary in a completed image.
Calibration Task calibrating the quality of the detector image.
Generic term of data files used for the calibration of
Map File
detector image quality.
Pixel Map Data file used to calibrate bad pixels in detectors.
Data file used to calibrate the properties of the detector
Gain Map
subsequent to installation environment.
Data file used to calibrate the properties of the detector
Panel Map
panel itself.

RConsole1_UM_COM_EN_02 117 / 124


Appendix B. Gain Calibration
Please perform gain calibration every 6 months in order to avoid any abnormal effects.

Preparation
- Aging for 20 minutes before starting calibration
- Remove Compression Paddle.
- Fully open Collimator to cover X-ray on the whole detector area.
- Put an Acryl Phantom (4cm) on Bucky.
- Set Grid type as “None” in RCali1 configuration. (When Grid Line Pattern appears, set
Grid Type as”1824C” or “2430C” depending on the product model)

Calibration Procedure
* Check the radiation field and clear Acryl Phantom as well as Bucky before calibration.
* If the Mean values of every image have more than 300Gy difference, you have to re-
take images.

1. Run Rcali1

2. Select the detector to perform the Gain Calibration from the Detector Menu. (Tool bar
→ Detector)

Figure 70. Detector Selection

RConsole1_UM_COM_EN_02 118 / 124


3. Select Gain Map Calibration from the Tool bar or select the shortcut icon.

Figure 71. Starting Gain Calibration

4. When a notification message appears, select [OK].

Figure 72. Calibration Start Notification

RConsole1_UM_COM_EN_02 119 / 124


5. RCali1 alerts whether the radiation dose is appropriate or not and whether it should be
automatically calculated. Proceed by selecting [OK].

* Set OP console manually (approx. 28kVp 80mAs)


* Mean value should be 2,200 ~ 2,500Gy.

Figure 73. X-ray Radiation Dose Check Activation

6. A main screen will appear. Proceed with the X-ray projection at this point in order to
acquire an image.

Figure 74. Projection Standby Main Screen

RConsole1_UM_COM_EN_02 120 / 124


7. The program automatically determines the appropriateness of the X-ray radiation dose
when the image is acquired.

Figure 75. X-ray Radiation Dose Verification Notification


8. If provided with a notification of low radiation dose, please select [OK] then increase
the X-ray radiation dose and re-project.

Figure 76. Notification of Low X-ray Radiation Dose


9. A notification will also appear if the X-ray radiation dose is high. Please select [OK]
then reduce the X-ray radiation dose and re-project.

10. If the X-ray radiation dose is appropriate, please select [OK] and proceed with the
projections.

Figure 77. Notification for Appropriate X-ray Radiation Dose

RConsole1_UM_COM_EN_02 121 / 124


11. A notification informing the requirement of 7 images will appear. The required number
of images may be set in Configuration. Select [OK] and proceed with the projections of 7
images.

Figure 78. Notification Informing 7 Images Requirement

Figure 79. Starting the Gain Calibration Projection

Figure 80. Standby post Projection of 1 Image

RConsole1_UM_COM_EN_02 122 / 124


12. Wait until the next designated projection is ready after the initial projection.

Figure 81. Projection Standby Timer

13. The next projection will take place after the projection standby timer (60 seconds)
disappears. The projection standby time may be set in Configuration.

14. The Gain Calibration results are automatically saved as Gain Map (*.GMP) when all
required images are projected and the existing files are backed-up.

Figure 82. Notification for Gain Calibration Completion

15. The existing GMP files are backed up in the [RCali1 Installation Folder]\MAP folder. The
file name is as follows in [Figure 83].

Figure 83. Example of a Back-up File

RConsole1_UM_COM_EN_02 123 / 124


RConsole1_UM_COM_EN_02 124 / 124