0% found this document useful (0 votes)
18 views20 pages

Definitions: Specifications

Uploaded by

electro-ub ub
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
18 views20 pages

Definitions: Specifications

Uploaded by

electro-ub ub
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 20

SPECIFICATIONS

VB-8054
4-Channel, 500 MHz Bandwidth Oscilloscope VirtualBench™ (All-In-
One Instrument)

Definitions
Warranted specifications describe the performance of a model under stated operating
conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating
conditions but are not covered by the model warranty.
• Typical specifications describe the expected performance met by a majority of the
models.
• Nominal specifications describe parameters and attributes that may be useful in operation.

Specifications are Typical unless otherwise noted.

Conditions
Typical specifications are valid under the following conditions unless otherwise noted:
• 25 °C
• 30-minute warm-up time before operation

Warranted specifications are valid at Tcal ±5 °C. Temperature coefficients are calculated using
the temperature change from last external calibration.
Mixed Signal Oscilloscope
Analog Channels
Vertical System
Number of channels 4 single-ended, non-isolated
Bandwidth (-3 dB)1 500 MHz
Resolution 8 bits
Accuracy (warranted) ±2% of input ±1% full scale (V peak-to-peak)
Input coupling DC, AC
Vertical sensitivity (range) 5 mV/div (40 mV peak-to-peak)
10 mV/div (100 mV peak-to-peak)
20 mV/div (200 mV peak-to-peak)
50 mV/div (400 mV peak-to-peak)
100 mV/div (1 V peak-to-peak)
200 mV/div (2 V peak-to-peak)
500 mV/div (4 V peak-to-peak)
1 V/div (10 V peak-to-peak)
2 V/div (20 V peak-to-peak)
5 V/div (40 V peak-to-peak)
Input impedance (user selectable) 1 MΩ || 12 pF or 50 Ω2

Table 1. DC Offset Range


Range Programmable Offset Range

5 mV/div, 10 mV/div, 20 mV/div, 50 mV/div ±5 V

100 mV/div, 200 mV/div, 500 mV/div, ±20 V

1 V/div, 2 V/div, 5 V/div

Acquisition modes Sample, peak detect, averaging

Horizontal System
Maximum sample rate 2 GS/s/channel
Maximum record length 1 MS/channel

1 Bandwidth using 50 Ω mode or 1 MΩ mode with the accessory oscilloscope probe.


2 Maximum voltage when using 50 Ω input mode is 5 V RMS. For a periodic waveform with
frequency below 100 kHz, the maximum voltage is derated to 2.5 V RMS.

2 | ni.com | NI VirtualBench VB-8054 Specifications


Digital Channels/Logic Analyzer
Vertical System
Number of channels 34
Maximum input frequency 100 MHz
Input voltage 0 V to 5 V
Input current ≤50 μA

Note Mixed signal oscilloscope digital channels are designed to withstand


accidental overvoltage from signals on the VB-8054 or similar devices. They are not
recommended for use with signals likely to exceed 0 V to 5 V in normal operation.

Input threshold Programmable, 0 V to 2.0 V


Threshold accuracy 350 mV
Input impedance 100 kΩ || 7.5 pF pulled to -2.0 V to +6.5 V,
varies with the input threshold setting
Additional/internal channels Digital I/O lines, function generator start,
external trigger (TRIG), power line frequency

Horizontal System
Timing mode sample rate (warranted) 1 GS/s (down to ~15 kS/s)
Maximum external sample clock rate 100 MHz
Record length
Typical 1 MS
Minimum3 4 kS
Decimation External Sample Clock, 1:1, 2:1, and
n*4:1 where n is an integer
Maximum sample compression 215 to 1

3 Under most conditions, the logic analyzer can acquire 1 MS of data. Under some conditions with
very high sustained activity on multiple inputs, the logic analyzer may only capture 4 kS of data.

NI VirtualBench VB-8054 Specifications | © National Instruments | 3


Triggering
Trigger modes Normal, Auto, Force, Single4
Trigger sources Oscilloscope analog channels,
oscilloscope digital channels,
function generator start, digital I/O lines,
external trigger (TRIG), power line frequency
Trigger types
Analog Edge with hysteresis, pulse width
Digital Edge, pulse width, pattern, glitch5
Trigger resolution
Analog/oscilloscope 500 ps
Digital/logic analyzer 1 ns
Trigger export Available through external trigger (TRIG)

Waveform Measurements
Oscilloscope time6 Period, frequency, positive duty cycle,
negative duty cycle, positive pulse width,
negative pulse width, rise time, fall time,
rise rate, fall rate
Oscilloscope voltage6 High, low, amplitude, maximum, minimum,
peak-to-peak, overshoot, undershoot, RMS,
mean, cycle RMS, cycle mean
Logic analyzer time6 Period, frequency, positive duty cycle,
negative duty cycle, positive pulse width,
negative pulse width

Waveform Math
Operations7 A + B, A - B, A * B, A/B, FFT

4 Single trigger mode is only available in the VirtualBench application. For feature differences
between the VirtualBench application for Windows and iPad, go to ni.com/info and enter
vbfeatures.
5 Glitch triggers are only available with the NI VirtualBench driver.
6 Waveform measurements are only available in the VirtualBench application.
7 Waveform math is only available in the VirtualBench application.

4 | ni.com | NI VirtualBench VB-8054 Specifications


Function Generator (FGEN)
Waveforms Sine, square, ramp/triangle, DC, arbitrary
Update rate 200 MS/s
Resolution 14 bits
Number of channels 1
Output impedance 50 Ω
Switchable filter8 71 MHz lowpass, 7-pole, elliptical
Sine
Maximum frequency 40 MHz
Total Harmonic Distortion (THD)
1 MHz -55 dBc
10 MHz -50 dBc
Spurious Free Dynamic Range -70 dB at 1 MHz (non-harmonic)
(SFDR)
Phase noise (1 MHz) -125 dBc/Hz at 10 kHz offset
Square
Maximum frequency 5 MHz
Rise/fall time <20 ns (10% to 90%)
Overshoot <5%
Jitter 5 ns cycle-to-cycle
Ramp/triangle maximum frequency 1 MHz
Accuracy (with >10 kΩ load)
Amplitude (1 kHz sine) ±(1% of output value ± 5 mV)
DC ±(1% of output value ± 5 mV)
Output range
50 Ω ±6 V
Hi-Z (>10 kΩ) ±12 V
DC offset
50 Ω ±6 V
Hi-Z (>10 kΩ) ±12 V

8 Switchable filters are only available with the NI VirtualBench driver. The VirtualBench application
automatically enables the lowpass filter in sine mode.

NI VirtualBench VB-8054 Specifications | © National Instruments | 5


Note The combination of signal amplitude and DC offset cannot exceed the output
range specifications. The impedances listed are the loads applied by the user to the
FGEN output.

Frequency
Accuracy ≤100 ppm
Resolution 1 μHz
Arbitrary waveform
Points 1 MS
Sample rate 200 MS/s
Flatness ±0.3 dB to 40 MHz
Protection Short-circuit protected

Triggering
Trigger types Start of buffer9
Trigger resolution 5 ns
Trigger export Available through external trigger (TRIG)

Digital I/O
Number of channels 8
Direction control Input or output, software-selectable
Logic level 5 V compatible TTL input,
3.3 V LVTTL output
Drive strength 4 mA
Input voltage 0 V to 5 V

Note Digital I/O lines are designed to withstand accidental overvoltage from
signals on the VB-8054 or similar devices. They are not recommended for use with
signals likely to exceed 0 V to 5 V in normal operation.

DIO channel pull resistors


dig/<0..5> 10 kΩ, pull-down to 0 V
dig/<6,7> 10 kΩ, pull-down to 0 V (default) or
1.5 kΩ, configurable pull-up to 3.3 V

9 The function generator can only produce a trigger.

6 | ni.com | NI VirtualBench VB-8054 Specifications


External Power
3.3 V output
Voltage 3.3 V ±10%
Current 20 mA

Digital Multimeter
Functions DC voltage, AC voltage, DC current,
AC current, resistance, diode, continuity10
Resolution 5½ digits
Sample rate 5 S/s

Caution Do not use this device for connection to signals or for measurements
within Measurement Categories III or IV. For more information about Measurement
Categories, refer to the Safety Voltages section.

Input protection
Resistance, diode Up to 300 V DC
DC and AC voltage Up to 300 V DC or 265 V AC RMS,
400 V AC peak
DC and AC current
DMM A current connector fuse Internal ceramic fuse, 11 A, 1 kV AC,
10.3 × 38 mm, F 11A 1000V (SIBA part
number 5019906.11 at
www.siba-fuses.com)
DMM mA current Internal ceramic fuse, 1 A, 500 V AC,
connector fuse 5 × 20 mm, T 1A H 400V (Littelfuse part
number 0477001.MXP at
www.littelfuse.com)

Caution Fuses are located on bottom of device underneath door. Use Phillips #1
screwdriver for removal. Ensure all hazardous voltages are disconnected from the
device prior to removal of door.

Fuse When this fuse symbol is marked on a device, take proper precautions.

Maximum common-mode voltage 300 V DC or AC RMS

10 Continuity is only available in the VirtualBench application.

NI VirtualBench VB-8054 Specifications | © National Instruments | 7


DC
Table 2. DC Voltage Accuracy
1-Year Accuracy Temperature Coefficient
± (% of Reading + ± (% of Reading +
Range Input Impedance % of Range) (warranted) % of Range)/°C (warranted)

100 mV* >10 GΩ, 10 MΩ 0.015 + 0.005 0.001 + 0.0005

1V >10 GΩ, 10 MΩ 0.015 + 0.005 0.001 + 0.0005

10 V >10 GΩ, 10 MΩ 0.015 + 0.005 0.001 + 0.0005

100 V 10 MΩ 0.035 + 0.005 0.005 + 0.0005

300 V 10 MΩ 0.035 + 0.005 0.005 + 0.0005

* Add 15 µV if not immediately following offset null.

Table 3. DC Current Accuracy


1-Year Accuracy Temperature Coefficient
± (% of Reading + ± (% of Reading +
Range Burden Voltage % of Range) (warranted) % of Range)/°C (warranted)

10 mA <0.03 V 0.070 + 0.020 0.0035 + 0.0010

100 mA <0.3 V 0.070 + 0.003 0.0020 + 0.0010

1A <0.03 V 0.130 + 0.025 0.0065 + 0.0010

10 A* <0.3 V 0.130 + 0.004 0.0045 + 0.0010

* 30 seconds on, 30 seconds off. Add 300 ppm/A for currents >2.2 A. After measuring >5 A,
wait two minutes to get full accuracy in the 1 A range.

Table 4. DC Resistance Accuracy (2-Wire)*, 1 V Open Circuit Voltage


1-Year Accuracy Temperature Coefficient
Short-Circuit ± (% of Reading + ± (% of Reading +
Range Current % of Range) (warranted) % of Range)/°C (warranted)

100 Ω 170 μA 0.018 + 0.050 0.0010 + 0.0005

1 kΩ 170 μA 0.018 + 0.005 0.0010 + 0.0005

10 kΩ 70 μA 0.018 + 0.005 0.0010 + 0.0005

8 | ni.com | NI VirtualBench VB-8054 Specifications


Table 4. DC Resistance Accuracy (2-Wire)*, 1 V Open Circuit Voltage (Continued)
1-Year Accuracy Temperature Coefficient
Short-Circuit ± (% of Reading + ± (% of Reading +
Range Current % of Range) (warranted) % of Range)/°C (warranted)

100 kΩ 10 μA 0.018 + 0.005 0.0010 + 0.0005

1 MΩ 1.1 μA 0.035 + 0.005 0.0040 + 0.0005

10 MΩ 1.1 μA 0.150 + 0.005 0.0100 + 0.0005

100 MΩ 1.1 μA 1.3 + 0.005 0.1000 + 0.0005

* Perform offset nulling.

Caution The input terminals of the DMM are not protected for electromagnetic
interference. As a result, the DMM may experience reduced measurement accuracy
or other temporary performance degradation when connected to unshielded test leads
in an environment with radiated or conducted radio frequency electromagnetic
interference.

DC continuity accuracy range11 100 Ω


DC diode test range 2V
Effective Common-Mode Rejection Ratio >100 dB
(CMRR), 1 kΩ resistance in LO lead
Normal-Mode Rejection Ratio (NMRR), >100 dB
50/60 Hz ±0.1%
Overrange 105% of range except 300 V

11 DC continuity is only available in the VirtualBench application.

NI VirtualBench VB-8054 Specifications | © National Instruments | 9


AC
Table 5. AC Voltage Accuracy
Temperature
1-Year Accuracy Coefficient ± (% of
± (% of Reading Reading + % of
+ % of Range) Range)/°C
Range (rms) Peak Voltage Frequency (warranted) (warranted)

100 mV, 1 V, ±210 mV, 20 Hz to 45 Hz 0.91 + 0.10 0.01 + 0.005


10 V, 100 V, ±2.1 V, ±21 V,
45 Hz to 65 Hz 0.30 + 0.05 0.01 + 0.005
265 V ±210 V, ±400 V
65 Hz to 1 kHz 0.21 + 0.05 0.01 + 0.005

1 kHz to 5 kHz 0.12 + 0.05 0.01 + 0.005

5 kHz to 0.35 + 0.05 0.01 + 0.005


20 kHz

Table 6. AC Current Accuracy


Temperature
1-Year Accuracy Coefficient ± (%
Burden ± (% of Reading of Reading + %
Range Peak Voltage + % of Range) of Range)/°C
(rms) Current (rms) Frequency (warranted) (warranted)

5 mA ±10.5 mA <0.02 V 20 Hz to 1 kHz 0.20 + 0.01 0.01 + 0.005

1 kHz to 5 kHz 0.60 + 0.01

50 mA ±105 mA <0.2 V 20 Hz to 1 kHz 0.20 + 0.01 0.01 + 0.005

1 kHz to 5 kHz 0.50 + 0.01

500 mA ±1.05 A <0.02 V 20 Hz to 1 kHz 0.15 + 0.01 0.01 + 0.005

1 kHz to 5 kHz 0.50 + 0.01

5A ±10.5 A <0.2 V 20 Hz to 1 kHz 0.25 + 0.03 0.01 + 0.005

1 kHz to 5 kHz 0.60 + 0.03

Caution The input terminals of the DMM are not protected for electromagnetic
interference. As a result, the DMM may experience reduced measurement accuracy
or other temporary performance degradation when connected to unshielded test leads

10 | ni.com | NI VirtualBench VB-8054 Specifications


in an environment with radiated or conducted radio frequency electromagnetic
interference.

Input impedance 10 MΩ || 200 pF


CMRR, 1 kΩ resistance in LO lead >70 dB (DC to 60 Hz)

DC Power Supply
Outputs 0 V to +6 V/0 A to 3 A,
0 V to +25 V/0 A to 1 A (isolated),
0 V to -25 V/0 A to 1 A (isolated)

Note The +25 V and -25 V channels are bank isolated from ground but not from
each other.

Table 7. DC Accuracy/Resolution
Output Type +6 V +25 V -25 V

DC output (warranted) Voltage 0 V to +6 V 0 V to +25 V 0 V to -25 V

Current12 3A 1A 1A

Programming accuracy12 Voltage 0.1% + 5 mV 0.1% + 20 mV 0.1% + 20 mV


± (% of reading + offset)
Current 0.2% + 10 mA 0.15% + 4 mA 0.15% + 4 mA
(warranted)

Readback accuracy13 Voltage 0.1% + 5 mV 0.1% + 20 mV 0.1% + 20 mV


± (% of reading + offset)
Current 0.2% + 10 mA 0.15% + 4 mA 0.15% + 4 mA
(warranted)

Programming resolution Voltage 1.6 mV 6.6 mV 6.6 mV

Current 0.90 mA 0.30 mA 0.30 mA

12 Minimum programmable current limit is 1% of range.


13 Programming and readback accuracy specified at no load.

NI VirtualBench VB-8054 Specifications | © National Instruments | 11


Table 7. DC Accuracy/Resolution (Continued)
Output Type +6 V +25 V -25 V

Readback resolution Voltage 0.40 mV 1.7 mV 1.7 mV

Current 210 µA 70 µA 70 µA

Load regulation14 Voltage 0.01% + 25 mV 0.03% + 5 mV 0.03% + 5 mV


± (% of reading + offset)

Overvoltage protection 30 V (all channels)


Reverse voltage protection Reverse clamp diode, protected by
self-resetting fuse

External Trigger (TRIG)


Direction control Input or output, software-selectable
Logic level 5 V compatible TTL input,
3.3 V LVTTL output
Drive strength 4 mA
Input voltage 0 V to 5 V

Note The external trigger line is designed to withstand accidental overvoltage from
signals on the VB-8054 or similar devices. It is not recommended for use with
signals likely to exceed 0 V to 5 V in normal operation.

Connectivity
Wired USB Interface
USB specification USB 2.0 Hi-Speed

Wired Ethernet Interface


Network interface 1000 Base-TX, full-duplex;
100 Base-TX, full-duplex;
100 Base-TX, half-duplex;
10 Base-T, full-duplex; 10 Base-T, half-duplex
Communication rates 10/100/1000 Mbps, auto-negotiated

14 Change in output voltage for any load within range.

12 | ni.com | NI VirtualBench VB-8054 Specifications


Maximum cabling distance 100 m/segment
Network IP configuration IPv4, DHCP Client

Table 8. Network Protocols and Ports Used


Port Protocol Function

Port 80/TCP HTTP Device configuration (web, MAX)

Port 443/TCP HTTP Device configuration (web, MAX)

Port 3580/TCP Service locator Device configuration (web, MAX)

Port 9090/TCP Configuration only VirtualBench instrument protocol

Port 5353/UDP Multicast DNS Device discovery

Wireless Interface
Table 9. Network Protocols and Ports Used
Port Protocol Function

Port 80/TCP HTTP Device configuration (web, MAX)

Port 443/TCP HTTP Device configuration (web, MAX)

Port 3580/TCP Service locator Device configuration (web, MAX)

Port 9090/TCP Configuration only VirtualBench instrument protocol

Port 5353/UDP Multicast DNS Device discovery

Network IP configuration IPv4, DHCP Client/Server


Radio mode IEEE 802.11 b,g,n
Wireless modes AP mode (default), client mode
Frequency band 2.4 GHz ISM
Channel width 20 MHz
Channels USA 1-11, International 1-13
(12 and 13 client mode only)
TX power +10 dBm maximum (10 mW)
Security Open, WPA, WPA2, WPA2-Enterprise
Enterprise security EAP types EAP-TLS, EAP-TTLS/MS-CHAPv2,
PEAPv0/MS-CHAPv2
Antenna External RP-SMA omnidirectional dipole

NI VirtualBench VB-8054 Specifications | © National Instruments | 13


Software Compatibility
For information about operating system support for Windows and iPad, go to ni.com/info and
enter vbfeatures.

Power Requirements
Caution The protection provided by the VirtualBench hardware can be impaired if
it is used in a manner not described in the NI VirtualBench VB-8054 Safety,
Environmental, and Regulatory Information.

Voltage input range 100 V AC to 240 V AC, 50/60 Hz


Power consumption 150 W maximum
Power input connector IEC C13 power connector
Power disconnect The AC power cable provides main power
disconnect. Do not position the equipment so
that it is difficult to disconnect the power
cable. Depressing the front panel power button
does not inhibit the internal power supply.

Calibration
Calibration cycle (digital multimeter, 1 year
mixed signal oscilloscope,
function generator, DC power supply)
Specified temperature Tcal ± 5 °C
Warmup time 30 minutes

14 | ni.com | NI VirtualBench VB-8054 Specifications


Physical Characteristics
Dimensions
Enclosure 30.48 cm × 20.32 cm × 9.40 cm
(12.0 in. × 8.0 in. × 3.7 in.)
Enclosure with connectors and 30.48 cm × 25.40 cm × 16.00 cm
antenna (12.0 in. × 10.0 in. × 6.3 in.)

Note Use the VirtualBench instrument in a horizontal orientation. Allow at least


10.16 cm (4.0 in.) of clearance in front, on the sides, and behind the VirtualBench
instrument for airflow clearances, and USB, power, and common connector cabling.

Weight 3.130 kg (6 lb 4.4 oz)


Connectivity
Mixed signal oscilloscope 4, BNC
Logic analyzer 1, 2x20 shrouded IDC header
External trigger 1, BNC
Function generator 1, BNC
Digital I/O
Type 1, pluggable screw terminal, 3.5 mm
(14 position)
Screw terminal wiring 0.1 mm2 to 2.0 mm2 (30 AWG to 14 AWG)
Torque 0.25 N · m (2.2 lb · in.)
Digital multimeter 4, 4 mm banana jacks
DC power supply 6, 4 mm binding posts
Security cable slot 1, complies with Kensington security slot
dimensions

If you need to clean the device, wipe it with a dry towel.

Safety Voltages
Connect only voltages that are within these limits.

NI VirtualBench VB-8054 Specifications | © National Instruments | 15


DMM Isolation Voltages
Hazardous Voltage This icon denotes a warning advising you to take precautions
to avoid electrical shock.

Channel-to-earth ground
Continuous 300 V, Measurement Category II
Withstand 3,000 V RMS, verified by a 5 s dielectric
withstand test

Measurement Category II is for measurements performed on circuits directly connected to the


electrical distribution system. This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.

Caution Do not connect the VirtualBench hardware to signals or use for


measurements within Measurement Categories III or IV.

DC Power Supply Isolation Voltages


+25 V and -25 V-to-earth ground, 60 V DC, Measurement Category I
continuous

Note Measurement Categories CAT I and CAT O are equivalent. These test and
measurement circuits are for other circuits not intended for direct connection to the
MAINS building installations of Measurement Categories CAT II, CAT III, or
CAT IV.

Environmental
Operating temperature 0 °C to 40 °C
Storage temperature -20 °C to 70 °C
Operating humidity 10% to 90% RH, noncondensing
DMM full accuracy at 10% to 80%
Storage humidity 5% to 95% RH, noncondensing
Cooling Forced air circulation (negative pressurization)
through a fan. Fan speed automatically adjusts
according to operating conditions. Intake
locations are on the sides of device. Exhaust
location is on the rear of device. Ensure that
the intake and exhaust locations are not
obstructed.

16 | ni.com | NI VirtualBench VB-8054 Specifications


Pollution Degree 2
Maximum altitude 2,000 m

Indoor use only.

Shock and Vibration


Operational shock 30 g peak, half-sine, 11 ms pulse (Tested
in accordance with IEC 60068-2-27. Test
profile developed in accordance with
MIL-PRF-28800F.)
Random vibration
Operating 5 Hz to 500 Hz, 0.3 grms
Nonoperating 5 Hz to 500 Hz, 2.4 grms (Tested in accordance
with IEC 60068-2-64. Nonoperating test
profile exceeds the requirements of
MIL-PRF-28800F, Class 3.)

Safety
This product is designed to meet the requirements of the following electrical equipment safety
standards for measurement, control, and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA C22.2 No. 61010-1
Note For UL and other safety certifications, refer to the product label or the Online
Product Certification section.

Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment
for measurement, control, and laboratory use; for radio equipment; and for telecommunication
terminal equipment:
• EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
• EN 55011 (CISPR 11): Group 1, Class A emissions
• EN 55022 (CISPR 22): Class A emissions
• EN 55024 (CISPR 24): Immunity
• AS/NZS CISPR 11: Group 1, Class A emissions
• AS/NZS CISPR 22: Class A emissions

NI VirtualBench VB-8054 Specifications | © National Instruments | 17


• FCC 47 CFR Part 15B: Class A emissions
• ICES-001: Class A emissions

Note In the United States (per FCC 47 CFR), Class A equipment is intended for
use in commercial, light-industrial, and heavy-industrial locations. In Europe,
Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended
for use only in heavy-industrial locations.

Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical
equipment that does not intentionally generate radio frequency energy for the
treatment of material or inspection/analysis purposes.

Note For EMC declarations and certifications, and additional information, refer to
the Online Product Certification section.

CE Compliance
This product meets the essential requirements of applicable European Directives, as follows:
• 2014/35/EU; Low-Voltage Directive (safety)
• 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
• 2014/53/EU; Radio Equipment Directive (RED)

Online Product Certification


Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance
information. To obtain product certifications and the DoC for this product, visit ni.com/
certification, search by model number or product line, and click the appropriate link in the
Certification column.

Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible
manner. NI recognizes that eliminating certain hazardous substances from our products is
beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact
web page at ni.com/environment. This page contains the environmental regulations and
directives with which NI complies, as well as other environmental information not included in
this document.

18 | ni.com | NI VirtualBench VB-8054 Specifications


Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle, all NI products must be
disposed of according to local laws and regulations. For more information about
how to recycle NI products in your region, visit ni.com/environment/weee.

电子信息产品污染控制管理办法(中国 RoHS)
中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物
质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录
ni.com/environment/rohs_china。(For information about China RoHS
compliance, go to ni.com/environment/rohs_china.)

NI VirtualBench VB-8054 Specifications | © National Instruments | 19


Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for
information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their
respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your
software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find
information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer
to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how
to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS
TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S.
Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable
limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

© 2017 National Instruments. All rights reserved.

376473A-01 Mar17

You might also like