Technical Specifications

OPTICAL SYSTEM / PHOTOMETER

Optical Design Single beam with rotating filter wheel


Measurement Mode Monochromatic & Biochromatic
Light Source Quartz Halogen Lamp, 12 V, 20 W
Photometric Range -0.05 ~ 2.5 Absorbance
Standard wavelengths 340, 405,510, 546, 578, 630 nm + Two optional
Filter Type 4-Cavity interference, long life ion beam assisted deposition

AutoChem Ingeniious TM
Wavelength accuracy +/- 5nm
Sample Holder 18 μL flow cell, 10 mm square/semi-micro cuvette, Coagulation adaptor and cuvette

DATA PROCESSING

Calculation Modes Absorbance


End Point (with & without sample blank)
Semi Automated Clinical Chemistry Analyzer
Kinetic (with and without Reagent Blanking)
Fixed Time (with and without Reagent Blanking)
1-Point with Sample Blank
Coagulation
Profile (up to 8 tests/profile)
Calibration Factor mode
Single-standard mode
Multi-standard mode (up to 6 standard) (spline)
Multi-standard mode (up to 6 standard) (point to point)
Data storage capacity 144 test parameters
24 profiles
2000 (latest) samples, 200 calibrators/standards, 576 daily controls
10 (latest) graphs
Result retrieval Patient ID wise, Test wise, Date wise

TEMPERATURE CONTROL

Cuvette temperature Peltier based


Operating temperature 30 & 37 ºC (+/- 0.2 ºC)
Incubation block Inbuilt for 37°C ± 0.2°C
16 places for 12 mm diameter test tubes

FLUIDIC SYSTEM

Flow cell volume 18 µL


Aspiration volume 200 µL ~ 2000µL
Aspiration Stepper driven peristaltic pump
Data Management
Display type 240 x 128 resolution graphical LCD with backlight
Data input In-built 28 keys membrane keypad PHO
TO
Facility to connect to external standard computer keyboard M
Data Output Real time curve plotting of absorbance Vs. Time

OR

ET
INCUBAT
Internal thermal printer (speed up to 0.5 lines/second)

ER
Facility to connect to external 80 column dot matrix printer
Serial port for computer connectivity

POWER /DIMENSIONS
CO
AG
Power requirements Input Voltage : 90-260 VAC, 50/60 Hz ULO
METER
Power Consumption : 100 Watts Max
Operating Environment 20°C to 40°C,RH 85% max
Relative Humidity 10% to 85%, Non-condensing
Dimensions 35 cm (W) X 34 cm (D) X 26 cm (H)
Weight <10 kg

For more information contact:

022 61559355 | [email protected] www.arkray.co.in


ARKRAY Healthcare Pvt. Ltd. 701-702, Opulence, 6th Road, TPS III, Santacruz (E), Mumbai - 400055, Maharashtra, India.
Flexible Calibration Modes
Single & Multi-Point Calibration Possible 2.5

Flexibility to run routine as well as specialized parameters


In-built Peltier regulated incubation compartment
2

(Eg. Immunoturbidimetry) on same instrument 1.5

ABSORBANCE
Cost Efficient 1

Saves cost of buying additional Incubator (approx. Rs.5000) Auto-normalization of Calibration curve making 0.5
recalibration simple & cost effective
Saves space of approx. 1 Sq. Feet for placing the 0

additional incubator Recalibration can be done using any 1 Standard and the 0 200 400 600 800

instrument will automatically normalize the calibration CONCENTRATION


Flexible
curve
On/Off facility for using incubator when required
Save cost of expensive reagents and standards
Convenient
Integrated timer with buzzer for ensuring right incubation time
Easy Operation
Integrated Coagulation Testing Mode
1. Wash & Aspiration function can be controlled using
Cost Efficient External Keyboard (optional PS/2 keyboard)
Saves cost of buying additional 2. Direct access key to run tests
Coagulation Analyser (approx. Rs.75,000) Saves time by directly running 9 program tests with single
Saves space of approx. 1 Sq. Feet for key from main menu, without accessing detailed test menu
placing the additional incubator Reduces chance of error by eliminating the need to memorize
No manual calculation of INR test menu number

Convenient
No maintenance of additional coagulation
instrument & accessories
Special Coagulation
Cuvette
Coagulation Cuvette Enhanced Data Management Capability
Holder
Enhanced Patient Data Storage & Retrieval

Single & Profile Wise Testing Facility Stores data of latest 2000 samples, 200 calibrators/standards, 576
daily controls and 10 reaction graphs
Program to create 24 different profiles / panels & upto 8 different Flexibility to retrive Sample & Calibration results by – Patient ID, Date
tests in each profile & Test Name
Convenience of running multiple parameters on same sample without Facility to input patient information & demographic details
switching program Daily & monthlty QC program
Enables faster sample turn around Manage 2 levels of controls with Levey-Jennings Chart